Effectivity: Performix Ultra/ADV Tubes D3186T, D3185T, D3176T, D3152T and D3132T
Performix PRO100/VCT Tubes D3194T and D3193T
Perseus Tube D3702T and D3701T
Details:
The CT tube crate was redesigned several years ago using a
wheelbarrow-concept, which allows the FE to easily move the
heavy crate down a hallway using handles and wheels that are
permanently mounted on the crate. The wheelbarrow-concept
has been very effective in open areas such as receiving docks
and hallways, but in situations where a doorway or room does
not have adequate space to work with, the wheelbarrow is
awkward and sometimes too bulky when attempting to
maneuver in and out of such areas.
Figure 1 Wheelbarrow
Resolution:
A 3-wheeled dolly has been designed to fit under the tube crate opposite of the current stationary
wheels. The dolly raises the crate high enough to ensure the blue shipping feet are clear of the
floor, thus allowing the crate to be moved on the two original wheels and on the dolly wheels. This
design makes the crate as simple to push as a department store “shopping cart”, which now
facilitates movement inside small rooms. NOTE: Dolly ships directly inside the Tube crate.
Figure 2
Figure 3
DOC0359256
Page 1 of 2
GE Healthcare MI&CT Class M eService Notes are protected by copyright (c) to General Electric Company Inc with all rights
reserved. Class M eService Notes issued by GE Healthcare are proprietary to GE Healthcare. Disclosure to persons not employed
by GE Healthcare prohibited. Duplication and distribution of any class of MI&CT eService Notes, in any form by entities other
than GE Healthcare is strictly prohibited.
Step 2 – Figure 4
Place the three-wheel dolly in front of the tube crate.
Position the dolly channel parallel to the board on the
bottom of the tube crate.
Figure 4
Step 3 – Figure 5
Pull out the yellow crate handles. Lift crate using proper
lifting technique. Safely lower crate into dolly channel.
Put handles back into stored / shipping position.
Figure 5
Step 4 – Figure 6
Push the crate using proper ergonomic techniques
Figure 6