Professional Documents
Culture Documents
Installation Sheet PN 1245 CAPILLARYS 3 OCTA GB
Installation Sheet PN 1245 CAPILLARYS 3 OCTA GB
LABORATORY OR INSTITUTION:________________________________________________________
Address:__________________________________________________________________________
Country/Distributor:___________________________________________________________________
1. INSTALLATION
2. CHECKLIST
3. CAPILLARYS POWER-UP
4. VALIDATION
6. USER TRAINING
• Back up device settings using the script available on S.K.B online support
http://support.sebia.com
• Back up the database using the Phoresis "Advanced Backup" menu
8. DOCUMENT
• Instruction manual
9. ACCEPTANCE
For purposes of this Agreement, the Capillarys3 system will be deemed accepted when Sebia has completed the setup, installation
and start-up testing of the system and confirms the system is performing according to manufacturer’s performance specifications,
and the Customer has evidenced acceptance by execution of Sebia’s Installation Sheet for the serial number of the Capillarys3 System
(the “Acceptance Date”)
_______________________________ ________________________________
Name and signature
Name and signature
LABORATORY OR INSTITUTION:________________________________________________________
Address:__________________________________________________________________________
Country/Distributor:___________________________________________________________________
PROTEIN(E) 6 buffer
Wash solution
• After startup check the CMOS calibration using the OPTICAL menu
- Integration time t ≤ 20 ms
3. CONTROL ANALYSIS
• Using the rack “0”, run both Normal and Hypergamma controls 2 times in positions 1 and 8
(with barcode)
• Verify absence of error messages
• Using the statistic tool in Phoresis, complete the tables below:
• Report minimum OD obtained on all the measurements of the Normal Control: _____ > 0,35
_______________________________ ________________________________
Name and signature Name and signature
LABORATORY OR INSTITUTION:________________________________________________________
Address:__________________________________________________________________________
Country/Distributor:___________________________________________________________________
PROTEIN(E) 6 buffer
Solution de lavage
IMMUNOTYPING KIT
• After startup check the CMOS calibration using the OPTICAL menu
- Integration time t ≤ 20 ms
3. CONTROL ANALYSIS
TERA : Analyze two normal control sera on 2 different "standard" racks in position 1 (standard dilution) (with
barcode)
OCTA : Analyze a normal control serum on a standard rack (standard dilution) (with barcode)
_______________________________ ________________________________
Name and signature Name and signature
LABORATORY OR INSTITUTION:________________________________________________________
Address:__________________________________________________________________________
Country/Distributor:___________________________________________________________________
HEMOGLOBIN(E) buffer
HEMOGLOBIN(E) Hemolysing
solution
Wash solution
Hb A2 Normal Control
2. PHORESIS SETTING
NA
3. CMOS CALIBRATION
• After startup check the CMOS calibration using the OPTICAL menu
- Integration time t ≤ 10 ms and current in the LED ≤ 900.
4. CONTROL ANALYSIS
• Using the rack “0”, run Normal control 3 times (with barcode)
• Verify proper dilutions and absence of error messages
• Using the statistic tool in Phoresis, complete the tables below:
Hb A2 NORMAL CONTROL
Fraction values (%)
CVs (%)
C/NC (*)
Hb A2 _______ ≤3%
(*) : min/max values in the target values / see instructions for use of the HEMOGLOBIN(E) technique
• Report minimum OD across all Normal Control measures obtained: ______ > 0,11
• On the same analyzes, check the absence of noise on the profiles.
_______________________________ ________________________________
Name and signature Name and signature
LABORATORY OR INSTITUTION:________________________________________________________
Address:__________________________________________________________________________
Country/Distributor:___________________________________________________________________
Hb A1c buffer
Wash solution
2. PHORESIS SETTING
3. CMOS CALIBRATION
• After startup check the CMOS calibration using the OPTICAL menu
- Integration time t ≤ 10 ms and current in the LED ≤ 900.
3. CALIBRATION
• Analyze 1 rack "0" with calibrators level 1 and 2 in position 1 and 8 and with barcode
• Check Hb A0 peak positions on each profile and note the values in the table : values must be in the target
specifications 130-170 ; see the raw data on the profiles using “VIEW_RAW_DATA=1” in the Phoresys.ini file.
OCTA
TERA
If at least one value is out of the specified target, launch an activation cycle.
Pursue the Hb A1c calibrations, in order to obtain 3 successful calibrations, with no alert (yellow, purple or red symbols, or
“atypical profile” on the profiles) and with Hb A0 positions in the specified target.
Notes :
Repetitive alerts demonstrate a problem of instrument or sample.
Pursue calibrations only after having resolved the problem.
- yellow alert when Hb A0 or Hb A1c peaks are out of range;
- red alert when raw values of %Hb A1c are out of the expected targets;
- purple alert when OD is below 0.06;
- “atypical profile” on the display window and on the calibration list when a normal peak is missing (Hb A0, Hb A1c,
Hb A2 or other HbA), or supplementary peak.
4. CONTROLS ANALYSIS
Analyze 1 rack "0" with level 1 and 2 controls in position 1 and 8 and with barcode
• Verify absence of error messages ((yellow, purple or red symbols, or “atypical profile”).
• Using the statistic tool in Phoresis, complete the tables below:
Level 1 Level 2
CVs (%) CVs (%)
Hb A1c _______ ≤2% Hb A1c _______ ≤2%
OR Difference (max-min)% A1c ≤ 0.3point : _____
OR Difference (max-min)% A1c ≤ 0.6point : _____
• Report minimum OD across the 12 level 1 Control obtained: _____ > 0.11
• Report minimum OD across the 12 level 2 Control obtained: _____ > 0.11
The average value of each control must be in the target range specified in the instructions for use.
Using Hb A1c technique, the target range must then be customized as described in the controls instructions for use.
Customized values :
Level 1 Level 2
IF CAPILLARY SAMPLES MUST BE TESTED, REFER TO THE 3 TERA / OCTA CAPILLARYS SERVICE MANUAL AVAILABLE ON
THE S.K.B ONLINE SUPPORT (HTTP://SUPPORT.SEBIA.COM).
_______________________________ ________________________________
Name and signature Name and signature
LABORATORY OR INSTITUTION:________________________________________________________
Address:__________________________________________________________________________
Country/Distributor:___________________________________________________________________
PROTEIN(E) 6 buffer
Dialysis buffer
Wash solution
IMMUNOTYPING KIT
URINE Technique:
2. Perform the PROTEIN (E) 6 validation of the device if it has not already been performed.
3. CONTROL ANALYSIS
• Using the rack “0”, run Normal control diluted to 80th in dialysis buffer as for the normalization of
the capillaries.
• Verify absence of error messages
• On the same analyzes, check the absence of lifting of the profile.
Example :
IT URINE Technique:
2. Perform the URINE validation of the device if it has not already been performed.
_______________________________ ________________________________
Name and signature Name and signature
LABORATORY OR INSTITUTION:________________________________________________________
Address:__________________________________________________________________________
Country/Distributor:___________________________________________________________________
2. CMOS CALIBRATION
• After startup check the CMOS calibration using the OPTICAL menu
Integration time t ≤ 20 ms
3. CALIBRATION
• If calibrated, analyze 1 rack "0" 3 times with calibrators level 1 and 2 in position 1 and 8 with barcodes.
4. CONTROLS ANALYSIS
on 1 rack "0" Analyze the 2 levels of controls in position 1 and 8 with barcode.
• Verify proper dilutions and absence of error messages.
(yellow, purple or red symbols, or “atypical profile” on the profiles).
• Using the stastistic tool in Phoresis, complete the tables below:
• Report minimum OD across all Normal Control obtained : _____ > 0.07
_______________________________ ________________________________
Name and signature Name and signature