PNX.IM.033.0(E) (2013.05.

20)

INSTALLATION

M
A
N
U
A
L

※The device is being used for X-ray purpose.

Point 200HD
PNX.IM.033.0(E) (2013.05.20) CONTENTS

CONTENTS
1. OVERVIEW ..................................................................................................................................... 3
2. FEAUTRES AND PURPOSE ......................................................................................................... 4
1) FEATURES ................................................................................................................................... 4
2) PERPOSE ...................................................................................................................................... 4
3. STRUCTURE ................................................................................................................................... 5
4. SAFETY ............................................................................................................................................ 6
5. PRECAUTIONS .............................................................................................................................. 7
6. SPECIFICATIONS........................................................................................................................ 11
7. COMPOSITION ............................................................................................................................ 14
1) PANORAMA ............................................................................................................................... 14
2) CONTROL PANEL ..................................................................................................................... 16
3) SID DISTANCE MEASURE .................................................................................................... 18
4) ACCESSORY .............................................................................................................................. 19
8. INSTALLATION ........................................................................................................................... 20
1) HOW TO INSTALLATION ......................................................................................................... 20
2) CONNECTION BLOCK DIAGRAM ......................................................................................... 21
3) PROGRAM INSTALLATION..................................................................................................... 22
4) ALIGNMENT AND CALIBRATION ......................................................................................... 23
5) CDX-VIEW IMAGE ................................................................................................................... 27
9. TECHNIOCAL FEATURE ........................................................................................................... 28
1) BLOCK DIAGRAM.................................................................................................................... 28
2) CONNECTION DIAGRAM ....................................................................................................... 29
3) INFORMATION .......................................................................................................................... 33
4) GENERAL DATA ....................................................................................................................... 35
5) DISPERSION OF EXPOSURE DOSE & RECEPTION ............................................................ 37
6) SERVICE ..................................................................................................................................... 38
10. MAINTENANCE ......................................................................................................................... 55
1) USE, STORAGE, MANAGEMENT ........................................................................................... 55
2) MAINTENANCE, WARRANTY ............................................................................................... 56
3) WASTE/DELIVER...................................................................................................................... 58
11. INFORMATION .......................................................................................................................... 59

Installation Point 200HD 2 / 60


PNX.IM.033.0(E) (2013.05.20) 1. OVERVIEW
PointNix Co., Ltd provides this printed User Manual as a guide for the
operation of Point 200HD, the High Definition HD Digital Panorama
manufactured by PointNix Co., Ltd. This document is designed to help the users
for better understanding of the equipment in order to obtain satisfactory results
from the use of the equipment.

Point 200HD, the High Definition HD Digital Panorama, is representing the


latest technology in X-ray Digital Panorama field. Point 200HD features superior
imaging quality, and is designed to provide the X-ray professional with an
intuitive, easy-to-use interface. It converts X-ray information directly to digital
signal to acquire high definition images for X-ray purposes. It is consisted of X-
Ray tube, High Voltage Power supply, Panorama detector assembly, Head
Position, Main frame, Control Panel, Image-taking stand, Base plate , Column,
Imaging controlling device(PC, Monitor), and CDX-View Software(Ver. 1.93).

Point 200HD is an outcome of high technology that reflects the next


generation’s needs for X-ray Digital Panorama. The use of new material called
Cadmium Telluride (CdTe) to the Panorama sensor makes this X-ray equipment
acquire clear and accurate images without having image distortion problems.
Point 200HD maintains a low radiation dosage due to its high absorptance
capacity.

Installation Point 200HD 3 / 60


PNX.IM.033.0(E) (2013.05.20) 3. STRUCTURE

① Features
(1) High resolution HD sensor
(2) Secured X-ray exposure
(3) Clear and accurate images due to CsI sensor
(4) Low radiation dosage due to high absorptance
(5) Scanning and receiving images at the same time
(6) Centralized membrane switch
(7) Integrated software program, CDX-View
CDX-View: This is a tool of digital diagnosis system which implements all
images from Panorama, Cephalo, CT, Intra Oral Camera, Digital Camera,
and Intra Oral Sensor etc. This is represented as an integrated data system
of X-ray clinics on Network use.

② Purpose
The equipment is to acquire tomographic images of teeth, jaw, oral structure
and panoramic etc. by transforming x-ray exposed signals to digital signals
for diagnosis purposes.

Installation Point 200HD 4 / 60


PNX.IM.033.0(E) (2013.05.20) 3. STRUCTURE

■ Dimension : 870(W) X 1020(L) X 1620(2260) mm


■ Weight : 150 kg

Installation Point 200HD 5 / 60


PNX.IM.033.0(E) (2013.05.20) 4. SAFETY
1) Symbols

Symbols Description
Direct risk of death, serious adverse reaction, serious
damage to the properties, or fire.

Indirect risk of death, serious adverse reaction, serious


damage to the properties, or fire.

Indicate the dangers of risk of serious and slight injuries,


partial damages to the properties or data loss.

Warning, Electrical shock

Radiography exposure indicator

Indication mark for protective earth

Type B applied part

Indication mark for radiate position beam

Installation Point 200HD 6 / 60


PNX.IM.033.0(E) (2013.05.20) 5. PRECAUTIONS
1) Direction for the use

Please read carefully to prevent misoperation due to electricity damage and


electromagnetic waves.
(1) Be sure to pull the plug out the device before cleaning the device or not
using for long time.
(2) Always keep clean the device by wiping with soft and dry towel.
(3) Maintain clean and dry around X-ray room and around the device.
(4) Caution for fire from a sparkler to wood piece, sawdust, dissolve material.
(5) Do not place electric plug near hot machine or equipment.
(6) This device should be used at facility where protection equipment against
X-ray is provided.
(7) Modification by unauthorized person is prohibited. Only trained, qualified
and authorized engineers/technicians are allowed to service this device.
(8) Staff must use device in mentioned order at manual and be well-informed
of the content of manual for patient’s safety.
(9) Before use, check the device’s appearance, connected PC, connected cable
and etc.
(10) Staff must be able to communicate with patient until scanning is finished.
(11) Turn off the power when device is not using for more than 6 hours.
(12) Do not put any object over the device during the usage.
(13) Only qualified engineer is able to inspect and maintenance/repair of the
device.
(14) Please use humidity-resistance plug when the device is using outside. This
outlet for outside have to have Ground Fault Interrupter (G.F.I) function
and this type of circuit breaker prevents user’s electric shock.
(15) Extension code must use which is grounded and use when it is necessary.

Installation Point 200HD 7 / 60


PNX.IM.033.0(E) (2013.05.20) 5. PRECAUTIONS
(16) Caution for deterioration in quality and damage of product due to the usage
of near the high intensity electric and magnetic field.
(17) We recommend taking 6 minutes of rest between scanning.
(18) Please contact agent or head office if an error occurs during the device
usage.
This product is safe product which passed Certification Test of Electro
Magnetic Interference and Tolerance

2) Safety Regulations

(1) Keep the device away from water or humidity.


(2) Make sure patient is not moving while scanning.
(3) If there is a doubt the oil is leaking from X-Ray tube, please turn off the
device immediately and contact to agent or head office.
(4) Always equip fire distinguisher and staff must well-informed of its use.
(5) PC Hardware and the device must be installed at least 2 meter away from
each other to avoid unnecessary contact of patient to PC Hardware while
operating.
(6) When you clean the device, make sure to turn off the device and
disconnect all cables between the device and PC. Be careful not to spill
anything into the device.

3) Danger

(1) This device’s position beam can cause impaired vision. Do not aim position
beam direct to patient’s eyes and guide patient not to look at the position
beam.

Installation Point 200HD 8 / 60


PNX.IM.033.0(E) (2013.05.20) 5. PRECAUTIONS
(2) Do not put anything in radius of the device movement. That can cause a
danger or damage of device during scanning.
(3) Do not place flammable materials or volatile materials near the device.
(4) Stop the device by pressing emergency switch when it occurs problem
during scanning.

4) X-ray Protective Policy

(1) Staff must obligatorily provide protection material such as lead apron to
patient when they are scanning image.
(2) Frequent exposed to X-Ray can cause cancer and various diseased. So staff
must protect themselves by lead apron or lead protect.
(3) Staff must keep the distance at least 2meters (7ft.) from focus and X-Ray
to hand exposure switch.
(4) The device must locate inside of shield room and staff should keep eyes on
scanning through outside window.
(5) Person who needs special attention such as children, pregnant woman,
elderly and unhealthy patient must counsel with medical attendant before
scanning.
(6) During the scanning, staff must keep check the device and patient for in case
of emergency.
(7) During the scanning, stop the device immediately by pressing emergency
button if it works unusual.
(8) Patients and staffs are prohibited long-time expose to X-Ray and take
enough rest because excessive use of X-Ray can cause various diseases.
(9) After scanning, it should wait 6minutes of cool time for next scanning.

The law relate to Radiation Safety and Protect is different for each Country.
User must observe the Law of Radiation Safety and Protect for their own
country.

Installation Point 200HD 9 / 60


PNX.IM.033.0(E) (2013.05.20) 6. SPECIFICATIONS
No. Name Technical Specifications
■ X-ray Beam: Cone Beam
■ kV: 60~90kV
■ mA: 4~15mA
■ Focal spot: 0.5 X 0.5 mm
■ Target angle: 5 °
X-ray Tube ■ Inherent filtration: 0.8mmAl or higher (equivalent
D-054SB
at 50kV)
(Option1)
■ Cooling method: oil immersed(Max. 60℃) and convection
oil cooling
■ Anode heat storage capacity: 35kJ
■ Mas anode heat dissipation: 250W
■ Exposure time: Max. 20sec
1
■ X-ray Beam: Cone Beam
■ kV: 60~90kV
■ mA: 4~10mA
■ Focal spot: 0.5 X 0.5 mm
■ Target angle: 5 °
X-ray Tube ■ Inherent filtration: 0.8mmAl or higher (equivalent
PDM90P
at 50kV)
(Option2)
■ Cooling method: oil immersed(Max. 60℃) and convection
oil cooling
■ Anode heat storage capacity: 35kJ
■ Mas anode heat dissipation: 250W
■ Exposure time: Max. 20sec

Installation Point 200HD 10 / 60


PNX.IM.033.0(E) (2013.05.20) 6. SPECIFICATIONS

No. Name Technical Specifications


■ Sensor Material : Direct Emission Super AMOLED CdTe,
750 ㎛ thick
■ Active Area(mm) : 150 x 4.8
■ Pixel number of active area : 3125 x 100
■ Pixel size : 48 ㎛
■ Resolution :10.4 lp/mm
Panorama XID-
C15DP ■ Dynamic Range(dB) : ≥72
2
Detector
■ Frame Rate : 200fps
■ Data Bits per pixel : 12bit
■ Data I/O: GigE, Optical Link, or Camera Link
■ Command I/O : 3 TTL triggers
■ Power Supply Requirement : D +5V/1A
A +5V/1A

■ Version: Ver. 1.93


■ OS: Windows XP or higher
■ CPU: Dual Core Pentium 2.4 GHz or higher
■ Supported GPU: NVidia GPU 1GB or higher
3 CDX- ■ RAM : 3G or higher
Soft ware
VIEW
■ HDD : 500GB free hard disk space or higher
■ Network : Gigabit Ethernet Network card
■ Display : 32bit color display or higher
■ Supported resolution : 1,280 X 1,024 or higher

■ OS : Windows XP or higher
■ CPU : Intel Core i5-2400 or higher
Imaging ■ GPU : NVidia GPU 1GB or higher
control ■ RAM : 3G or higher
4 PC
device ■ HDD : 500GB free hard disk space or higher
(Option) ■ Network : Gigabit Ethernet Network card
■ Display : 32bit color display or higher
■ Resolutions : 1,280 X 1,024 or higher

Installation Point 200HD 11 / 60


PNX.IM.033.0(E) (2013.05.20) 6. SPECIFICATIONS
No. Name Technical Specifications
■ Screen size : True color 19 inch
■ Viewing angle : 170° Horizontal, 170° Vertical
Imaging ■ Brightness : up to 250 nits (cd/m2)
control
5 Monitor ■ Signal input : VGA, DVI-I
device
■ Contrast rate : 500 : 1
(Option)
■ Recommended resolution : 1,280 X 1,024 or higher
■ Height adjustment : Vertical tilt range of -5°~+25°

Installation Point 200HD 12 / 60


PNX.IM.033.0(E) (2013.05.20) 7. COMPOSITION
1) Panorama

4
3
5

6
8 1

11
9

10

Installation Point 200HD 13 / 60


PNX.IM.033.0(E) (2013.05.20) 7. COMPOSITION

No. Name Functions


1 Main Frame Rotation driving part of the rotation arm

2 Rotation Unit Rotation part following to the teeth locus while scanning

3 X-Ray unit Contains the X-ray tube refrigerated by oil

4 Mirror Patient’s position fixing mirror

5 Detector part Image detector included (detachable)

6 Head position Help patient’s head stable


Patient’s holding are to be properly positioned for
7 Handle
scanning

8 Control panel Operating control panel

9 Column Support Main frame part (UP/DOWN column)

10 Base plate Support and fix the whole body of the device

11 Exposure S/W Switch for exposure X-ray

Emergency switch button- Stop device automatically by pressing switch


when unexpected danger situation

Installation Point 200HD 14 / 60


PNX.IM.033.0(E) (2013.05.20) 7. COMPOSITION
2) Control Panel

No. Name Functions


Basic mode when power is on and also use this button to select
ⓐ Panorama mode
the Panorama mode

ⓑ TMJ mode Select the TMJ mode

ⓒ Cephalo mode Select the Cephalo mode

ⓓ CT mode Select the CT mode

Installation Point 200HD 15 / 60


PNX.IM.033.0(E) (2013.05.20) 7. COMPOSITION
No. Name Functions
Preparation process before taking an image.
Fix the patient’s position. Rotation part, detector part and
ⓔ READY
collimator part are moving depends on scanning mode. (LED is
on when it is ready to scan)

ⓕ X-ray lamp Indicator the means images are on the process of taking

ⓖ STOP lamp Indicator the means image taking is completed

Help to adjust the patient’s horizontal and vertical positions.


ⓗ Position beam (Automatically turned off in 30 seconds, and it is not operating
while scanning)

kV control key-Adjust kV based on patient and indicate error


ⓘ kV
mark

mA control key-Adjust mA based on patient and indicate error


ⓙ mA
mark

ⓚ Time Adjust image taking time and indicate error mark

Adjust numbers of ⓘⓙ and ⓚ.



Up, Down Move column to upward or downward. (Possible to operate

also in Ready condition.)

Import pre-set value of kV and mA for each ⓘ, ⓙ and ⓚ


ⓝ button which are set based on each patient’s info body type(ⓝ-
ⓞ Memory mode Children, ⓞ-Woman, ⓟ-Man) information.
ⓟ (Each button contains 2 types of memory function, and can
change function by one press.)

ⓠ Image taking Move rotation arm back or forth while image scanning
ⓡ location of
(ⓠ-Protrusion, ⓡ-Normal, ⓢ-Depression)
ⓢ rotation unit

Installation Point 200HD 16 / 60


PNX.IM.033.0(E) (2013.05.20) 7. COMPOSITION
3) SID Distance Measure

SID location
Condition SID(mm)

Panorama, Tmj 430mm

Condition SCAN TIME

PANORAMA 17s

Tmj 11s

Condition Rotation Angle

PANORAMA
270°
Tmj

Installation Point 200HD 17 / 60


PNX.IM.033.0(E) (2013.05.20) 7. COMPOSITION
4) Accessory
▶ Point 200HD
No. ACCESSORY
Exposure Switch

: X-Ray Scanning (3M)


LAN Card(Giga bit(Intel)-)

: Sensor and PC Data Communication Interface Board (CT )


Serial Cable

: Device and PC Interface Cable (5M)


LAN Cable(CAT.7)

: Panorama Image data Cable (10M)

Installation Point 200HD 18 / 60


PNX.IM.033.0(E) (2013.05.20) 8. INSTALLATION
1) How to installation
▶ Ready
① Unpack the system, check the following.
The state of the product is good, there should be no damage in transit.
Check that there are no missing parts installations
② Basic Components

Main Unit(200HD)
③ Detail explanation of installation step is as below.
Prepare Main Unit
Move the packed Main Unit to the location where you will install, and
adjust the level with adjusting screw.

Installation Point 200HD 19 / 60


PNX.IM.033.0(E) (2013.05.20) 8. INSTALLATION
2) Connection Block Diagram

Point 200HD

Point 200HD
CN18 Expos S/W

CN13 PANORAMA

USB or
Serial Cable

LAN Card
PC

Installation Point 200HD 20 / 60


PNX.IM.033.0(E) (2013.05.20) 8. INSTALLATION
3) Program Installation
1. CDX-VIEW Installation

 Install “CDX_Setup” at D Hard Disk Drive

2. CDX_HD_PANO_Setup Installation
Smart Sensor(PANORAMA)

Install at C Hard Disk Driver.


(Install at HD folder in the CDX-View folder which is installed in
previous steps)

During the installation, select correct device type at ‘Additional


Selection’ between ‘use Gigabit Interface (PANO only)’ and ‘use
Gigabit Interface (PANO/CEPH)’

Installation Point 200HD 21 / 60


PNX.IM.033.0(E) (2013.05.20) 8. INSTALLATION
4) Alignment and Calibration

A. Run “XI.exe” program to check Alignment and do


Calibration. It is installed at ‘PointNix\CDX-View\HD’

B. Default screen is like below picture.

I. Check Sensor’s connection status


▶ When the sensor is connected, it will show as ‘0xCA1’
II. Check Sensor’s temperature
▶ Recommend temperature is 26~33℃
III. Check Sensor’s board ID

Installation Point 200HD 22 / 60


PNX.IM.033.0(E) (2013.05.20) 8. INSTALLATION
C. Comm test mode
▶ This is communication test mode to check the communication
between sensor and main PC.

2. Click
‘Start
Testing’

When communication test is completed, it


will show as “Success communication test”

D. Align mode – Panorama


▶ This mode is for checking Alignment of Panorama or Cephalo

For Panorama Alignment,


Process the alignment with Service
Mode and change exposure value Check
as below: ‘Panorama’ is
selected
70 kV, 4 mA, 3 Sec

Installation Point 200HD 23 / 60


PNX.IM.033.0(E) (2013.05.20) 8. INSTALLATION
E. Align mode – Cephalo
▶ This mode is for checking Alignment of Panorama or Cephalo.

Check ‘Cephalo’ is
selected

Displays detected X-
ray at Active Area of
sensor in real-time

F. Extraction mode – Calibration


▶ This mode is for Calibration work

Installation Point 200HD 24 / 60


PNX.IM.033.0(E) (2013.05.20) 8. INSTALLATION
▶ Panorama Mode

Change the device to Service


Mode, and attach 26mm
Aluminum Filter then expose
correct X-ray for each Level

Level1: Dark (X-ray off)


Level2: 65kV, 4mA, 19Sec
Level3: 71kV, 10mA, 19Sec

(Create

Calibration file)

Installation Point 200HD 25 / 60


PNX.IM.033.0(E) (2013.05.20) 8. INSTALLATION
5) CDX-VIEW Image
1. Run Exposure program of HD+ Panorama at CDX-View.

2. It displays ‘Device 311 error’ error message at the first time when you
run exposure program. This error code is due to no applied calibration
file. Enter configuration window, and setup the saved calibration file as
below picture. Then save and exit the configuration to scan
Panorama/Cephalo image. (Panorama & Cephalometric are the same)

Installation Point 200HD 26 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
1) Connection diagram
1. Main Control Board Pin Map

Installation Point 200HD 27 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
2. Name of Main Control Board Connector

No. Connector Name


1 CN1 Head rotating Motor Driver
2 CN2 AC Motor Control
3 CN3 Limit S/W (Y Motor )
4 CN4 Y Motor Driver
5 CN5 Limit S/W (Y Motor)
6 CN6 X-ray on signal
7 CN8 Ceph Motor Driver
8 CN10 Limit S/W (Y Motor)
9 CN11 Collimator Motor Driver
10 CN12 Limit S/W (Ceph Motor)
11 CN13 PC
12 CN14 Detector Motor Driver
13 CN15 Limit S/W (Ceph Motor)
14 CN16 Tube Motor Driver
15 CN17 Limit S/W (Collmator Motor)
16 CN18 LED Board
17 CN19 Limit S/W (Collmator Motor)
18 CN20 Photosensor 1(Start)
19 CN21 Limit S/W (Up)
20 CN22 Photosensor 2(Ready)
21 CN23 Control Panel
22 CN25 Limit S/W (Down)
23 CN28 INVERTER Board
24 CN29 Limit S/W (Detector Motor)
25 CN31 Photosensor 5(Start & End)
26 CN33 Limit S/W (Detector Motor)
27 CN34 Photosensor SCAN(PA, AP Mode)
28 CN35 UP/DOWN Brake
29 CN36 Photosensor CP(Ceph Mode)
30 CN37 Limit S/W (Tube Motor)
31 CN38 Limit S/W (Tube Motor)
32 CN39 Laser Diode 1
33 CN40 Laser Diode 2
34 CN41 Power IN(DC12V)
35 CN43 Ceph Hand S/W
36 CN44 Ceph X-ray Gen
37 CN50 Laser Diode 3
38 CN60 Exposure S/W or Main Board S/W upgrade
39 CN50 Laser Diode 3
40 TP1 MA_REF
41 TP3 KV_REF
42 TP5 Ground
43 TP7 Ground
44 SW1 Mode Setting

Installation Point 200HD 28 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
3. How to connect
▶ How to connect Main Control Board & Inverter Board
Main Control Board Inverter Board
CN28 CN1

▶ Name of PWR Board & INVERTER Board connector/How to


connect
PWR Board INVERTER Board
CN1 Data In/Out
CN1 DC 300V Output CN5 DC 300V Input
CN2 110-120/220-240V~ Input CN4 X-ray tube(Feedback)
CN7 X-ray tube(High voltage)
CN7 DC Output CN3 DC Input
CN10
CN11
CN12
DC 24V, 10A Output
CN13
CN14
CN15
CN16 DC 12V, 6A Output
CN17 DC 5V, 6A Output
F1 Fuse 250V, 6.3A F1 Fuse 250V, 15A
F2 Fuse 250V, 2A
F3 Fuse 250V, 10A

▶ How to connect PWR Board & Detector (Pano) & Detector (Ceph)
& Main Control Board
PWR Board Detector(Pano) Detector(Ceph) Main Control
Board
CN16 DC 12V DC 12V Input CN41(DC 12V
Output Input)
CN17 DC 5V DC 5V Input
Output
X-ray On Signal CN13
X-ray Gen CN44
Hand S/W CN43

Installation Point 200HD 29 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
▶ How to connect PWR Board & Motor Driver & Main Control Board
PWR Board Motor Driver Main Control Board
CN10 DC 24V Head rotating Motor CN1
~CN1 Output Driver(DC)
5 Motor Driver CN4, CN8, CN11
CN14 CN16
Motor(Stepping Motor)
CN16 DC 12V CN41

▶ How to connect Main Control Board & AC Motor Inverter (Tilt


Motor Driver) & Up/Down AC Motor

Main Control Board AC Motor Inverter Up/Down AC Motor


CN35 Brake L1(100~240V~) Orange
L2 Orange
Protective earth
U Black
V White
W Gray
CN2 Up/Down STF,STR,RH,RM,RL,SD

Installation Point 200HD 30 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

2) Information

1. Nominal X-Ray tube voltage


Maximum X-Ray tube voltage of High Frequency Generator when it is
operating in X-Ray tube voltage
(Max current: 15mA Voltage: 65kVp)

2. Highest electric output power


Combination of X-Ray tube voltage in maximum output and X-Ray tube
current
(65KV, 15mA)

3. Nominal electric power


The maximum output, Nominal electric power, is presented as combination
of X-Ray tube voltage, X-Ray tube current and loading time during 0.1
second’s loading time at 100kV’s X-Ray tube voltage.
(100W = 100kv1mA)-> Withstanding voltage’s specification. 100KV print
is unavailable.

4. Lowest current time product 3


If the value of the lowest current time product is from the combination of
X-Ray tube voltage and X-Ray tube loading factor, the lowest current time
product is presented as chart or graph.

5. Maximum Deviation from indicated values


Parameter Indicated value Deviation
Tube voltage 60 – 90kVp ±10%
Tube current 4 – 15mA ±20%
Exposure time(Pano/CT) 17.0s/19.0s
Exposure time(TMJ) 17.0s ±10%
CT 19.0s
inearity 0.9446HU 0±30HU

Installation Point 200HD 31 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
6. Nominal shortest irradiation time
If High Frequency Generator with automatic exposure control system is
from loading factor such as Nominal shortest irradiation time X-Ray tube
voltage and X-Ray tube current etc, the loading factor’s range of nominal
shortest irradiation time shall be indicated and also maximum declination of
X-Ray tube voltage and X-Ray tube current shall also be indicated.

7. kVp range, mA rating chart


mA kVp
4 – 10mA 60 – 90kVp
12mA 60 - 80kVp
14mA 60 – 75kVp
15mA 60 – 65kVp

8. Stationary Anode X-Ray Tube


A. Especially designed for X-ray radiography unit.
B. Low target angle adaptive for X-ray tomography.
C. Provided with an insulation cylinder (D-054S) and lead cylinder. (D-
054SB)
D. These tubes have 0.5 mm focus, and are available for maximum tube
voltage 90kV with Single-phase full-wave rectified or DC circuit.
E. Installed in the same enclosure with the high tension transformer.

Installation Point 200HD 32 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

3) General Data
1. Electrical
A. Circuit (Center-Grounded) : Single-phase full-wave rectified or DC
B. Operating Tube Voltage : 50 ~ 90kV
C. Focal Spot : 0.5mm
D. Input Energy (at 1.0s) : See Rating Chart
E. Single-phase full-wave rectified: 840W
F. DC : 1750W

2. Mechanical
A. Dimensions : See dimensional outline
B. Overall Length: D-054SB - 146 mm
C. Maximum Diameter: 054SB - 58 mm
D. Target Angle : 5 degrees
E. Inherent Filtration : At least 0.8 mm Al equivalent at 50 kV
F. X-ray Coverage :75 × 380 mm at SID 550 mm
G. Weight:D-054SB - Approx. 780 g
H. Cooling Method :Oil immersed (60℃Max.) and convection oil
cooling
I. Tube Holding : D-054SB - Holding the lead cylinder

3. Maximum Rating Chart(Absolute maximum rating charts)


A. DC (Center-Grounded), Focal Spot : 0.5 mm

Installation Point 200HD 33 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
B. Single-Phase Full-Wave, Focus Spot : 0.5 mm

4. Emission & Filament Characteristics


A. DC (Center-Grounded), Focal Spot : 0.5 mm

B. Single-Phase Full-Wave, Focus Spot : 0.5 mm

Installation Point 200HD 34 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
5. Anode Thermal Characteristics

4) Dispersion of Exposure Dose & Reception


1. Available Energy &Coefficient of Energy Absorption
Available Coefficient of Energy Absorption (μen/ρ):cm^2/g
Energy (keV) (1) (2) (3) (4)
10 4.7420 4.9870 4.9440 3.0260
15 1.3340 1.4020 1.3740 0.8324
20 0.5389 0.5663 0.5503 0.3328
30 0.1589 0.1616 0.1553 0.0965
40 0.0683 0.0722 0.0695 0.0460
50 0.0410 0.0436 0.0422 0.0307
60 0.0304 0.0326 0.0319 0.0253
80 0.0241 0.0262 0.0260 0.0230
A. Air: AIR,DRY(NEAR SEA LEVEL) density ρ: 0.001205g/㎤
B. Soft Tissue: TISSUE,SOFT(ICRU 44) density ρ:1.060g/㎤
C. Water: WATER,LIQUID density ρ:1.000g/㎤
D. Acrylic: POLYMETHYL METHACRYLATE density ρ:1.190g/㎤

Installation Point 200HD 35 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
2. Changing Volume Coefficient to Dose Absorption from Energy
Available Energy Changing Volume CoefficientC/kg---Gy
(keV) (2)/(1) (3)/(1) (4)/(1)
0 35.7251 35.4171 21.6772
15 35.7016 34.9886 21.1969
20 35.6972 34.6886 20.9783
30 34.5472 33.2004 20.6300
40 35.9097 34.5668 22.8788
50 36.1242 34.9642 25.4361
60 36.4284 35.6462 28.2711
80 36.9300 36.6481 32.4195

5) Service
1. How to set up Test Mode
This is a mode to test whether device is operating or not under the
condition that X-Ray is not exposed.
▶ Start Test Mode
Before power is on, press Down and Right button at the same time then
turn the power switch on.
▶ How to operate
This is same as user manual but X-Ray is not exposed.
2. How to set up Service Mode
X-Ray is only detected and rotation mechanism is not operating.

Installation Point 200HD 36 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
A. Starting Service Mode
Turn power switch on while pressing UP (+) and LEFT (<)
buttons together before turning power on.
B. How to operate(control)
Press Exposure S/W after adjusting kVp, mA, and Sec values.
C. Keys available to use
kVp, mA, Sec, Laser, Memory function(C, W, M), UP/Down,
Stop
3. How to set up Setting Mode
① Entering a setting mode

Before turning power ON, press (READY) Key and


(LASER) Key at the same time, and then turn main switch on.
② Setting up a setting mode and how to control

Installation Point 200HD 37 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

Mode
Key Sequence For details, refer to
Setting

#login Main Power


ON Device version display
Screen + +

A1 Heating time adjustment

A2 Auto ready time

A3 Buzzer sound of shot switch

A4 Error check(mA_OK,kV_ok)

A5 KV Calibration adjustment
Use buttons on log in
A6 MA Calibration adjustment

A7 screen of Setting mode to adjust each Maximum value of kV

A8 setting mode Minimum value of kV

A9 Maximum value of mA

AA Minimum value of mA

b1 Initial value of pano scanning rail


Collimator’s starting pointbefore a
b2
Panoramic

Installation Point 200HD 38 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

Mode
Key Sequence For details, refer to
Setting

#login Main Power


ON Device version display
Screen + +

b3 Adjust mA Target value while PA


scanning.
b4 Panorama scanning time

c1 Initial value of ‘Z’ axis


Collimator’s starting point before a
c2 CT
Use buttons on log in scanning
c3 Control the rotation speed of CT
screen of Setting mode to adjust each tarting point of ‘Z’ axis for Ceph
d1
scanning
Panorama Collimator location
d2 setting mode. control
Angle of rotating unit during a ceph
d3
scanning.
Starting point of LA(Lateral) scan
d4
cephalomtric before ceph scanning.
The starting point of
PA(Posterior/Anterior)
d5
AP(Anterior/Posterior) scan
cephalomtric before ceph scanning.

Installation Point 200HD 39 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

Mode
Key Sequence For details, refer to
Setting

#login Main Power


ON Device version display
Screen + +
Scanning method to setup one
E1 direction scanning
or bi-direction scanning for TMJ
Use buttons on log in Initial angle in Point 3D Combi
F1
500S
F2 Initial angle of Tube in Point 3D
screen of Setting mode to adjust each Combi 500S
F3 Check an equipment error history
setting mode.
G1 Collimator assy’ of old or new type

G2 Test a tube for ageing

Installation Point 200HD 40 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

③ Login Screen
(1) Checking the Firmware version.
Anterior number is the Main Board(MCU board) version.
Posterior number is the Program version.

(2) Mode A1 is to control heating time of HFG. Setup range is


between 30 seconds and 99 seconds and the values are
converted by 5 seconds. If scanning is processed with 6
minutes’ interval, heating time is not operated but it is operated
with 5 minutes’ interval. For the three times of initial scanning,
it will be operated for every scanning in a row, after 99 seconds’
heating time.
ex) 1—99
Warning: Scanning condition for this device is as below.
CT - Take at least 6 minutes break after scanning.

(3) Mode A2 is to adjust automatic READY waiting time. Setup


range is from 30 seconds to 90 seconds, and converting is
possibly made by 10 seconds

(4) Mode A3 is to turn buzzer sound for ON/OFF while X-Ray


scanning. RIGHT key - ON, LEFT-key OFF
ex) 1 – ON/ 2 – OFF

(5) Mode A4 is to set error check (mA OK, kV OK). If you press
RIGHT key, error is going to be checked, and if you check
LEFT key, error is not going to be checked.

Installation Point 200HD 41 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

(6) Mode A5 is a KV Calibration mode.


Setup range is changeable from 15 to 25, and kV value is
lowered when the value is getting higher

(7) Mode A6 is a mA Calibration mode.


Setup range is from 50 to 150, and mA value is lowered when
the value is getting higher.

(8) Mode ‘A7’ is to setup the maximum value of kV.


Setup range is between 50Kvp and 90Kvp.
It can save the maximum kV value, if the kV value is more
than 80kV.

(9) Mode ‘A8’ is to setup the minimum value of kV.


Setup range is between 50Kvp and 90Kvp.
It can save the minimum kV value, if the value is lower than
60kV.

(10) Mode ‘A9’ is to setup the maximum value of mA.


Setup range is between 4mA and 16mA.
It can save the maximum mA value, if the kV value is more
than 10mA.

Installation Point 200HD 42 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
(11) The Mode ‘AA’ is to setup the minimum value of mA.
Setup range is between 4mA and 16mA.
It can save the maximum mA value, if the kV value is lower
than 7mA.

(12) Mode B1 is to adjust initial value of starting point for


Panorama/TMJ scanning rail. Setup range is between -15 and
+15.

(13) Mode B2 is to set the Collimator location for Panorama


scanning. Setup ranges are between 0 and 50 by 0.1 mm which
are variable. (Collimator is screw type.)

(14) Mode B3 is to set the control of mA target value while


Panorama scanning.
Setup 0: mA Non-Adjustable Output
Setup 1: Adjustable Spine part of increased output 2Ma

(15) Mode B4 set Panorama scanning time.


Press RIGHT button to set '1' and scanning time to be 14sec.
Press LEFT button to set '0' and scanning time to be 17sec.

(16) Mode C1 is to adjust initial value of starting point for CT


scanning rail. Setup range is from 0 to 30.

Installation Point 200HD 43 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
(17) Mode C2 set a scanning position value of Collimator,
when take CT.
The setting range is 0 to 90, and the measured unit is 0.1mm.
Set the value maximum 90mm when update F/W from existing
Collimator, Because Collimator should be full opened

(18) Mode C3 is to change rotation speed of CT mode. Setup


range is between 30 and 50 and lower value means fast rotation
speed.
Warning: This is not related with Panorama scanning.

(19) Mode D1 is to adjust initial value of starting point for


Cephalo scanning. Setup range is from 0 to 30

(20) Mode D2 is to set the starting point of Collimator for


SCAN Cephalo scanning.
Setup ranges are between 18 and 50 by 0.1 mm which are
variable.

(21) Mode D3 is to set the angle value of Arm for Cephalo


scanning. Setup ranges are between 0 and 50 degree by 0.5
degree which are variable.

(22) Mode D4 is to control the starting point for LA scanning


of the sensor for SCAN Cephalo scanning.
Setup ranges are between 0 and 30 by 1mm which are variable.

Installation Point 200HD 44 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

(23) Mode D5 is to control the starting point for AP/PA


scanning of the sensor for SCAN Cephalo scanning.
Setup ranges are between 0 and 30 by 1mm which are variable.

(24) D6 mode is sensitive to the kind of CEPH SENSOR '0 'if


the CSI sensor and 1' If you choose to SMART sensor. It was a
little wrong to discriminate X-RAY-driven.

(25) Mode E1 is set up scanning methods for TMJ scanning.


If you press RIGHT key, scanning is made to one direction and
if you press LEFT key, scanning is going to be made to both
directions.

(26) Mode F1 set Tube initial moving angle of PANO, TMJ,


CT in 500s.
Press RIGHT to increase a value in range 150, Press LEFT
button to decrease a value in range 30.

(27) Mode F2 set Tube initial moving angle of SEPH in 500s.


Press RIGHT button to increase a value in range 50.0 ˚,
Press LEFT button to decrease value in 40.0 ˚.

Installation Point 200HD 45 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
(28) The Mode ‘F3’ is to check an equipment error history.
mA FND is displayed room numbers.
Each room has error code, and it is displayed on SEC FND.
The number ‘0’ means the equipment has no error history.

(29) Mode G1 is to set the Collimator type.


Setup 0: Timing Belt.
Setup 1: Screw

(30) Mode G2 set to Ageing test mode from HFG SET.


Press RIGHT button to set '1' that is set Ageing Test mode,
Press LEFT button to set '0' that is set normal mode.

4. Test Mode
This mode is to check operation of the device rotating without X-ray
exposure. Therefore, it is possible to check device’s rotation and all
other conditions include whether any parts has problem or not. Also it is
possible to understand the distance of patient and sensor during the
device’s rotation
① Entering Test Mode

Click button than click button and use as CW/CCW when


turn ON the device and it enters CT mode.

② Controlling method
Same as the User Manual accept no X-ray exposure during rotation.

5. Service Mode
Device exposes X-ray without rotation.

Installation Point 200HD 46 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

In service mode, LED of button is ON and use it to keep


the fixed position during alignment operation.
① Entering Service Mode

Press UP and + button at the same time then turn ON the device.

② Controlling method
Adjust kV, mA, Sec values as wanted then press exposure S/W.

③ Usable keys
kV, mA, Sec, Laser, memory function(C, W, M), UP/Down, Stop

6. Collimator location value change and save for panorama, ceph


mode (screw type)Measure Condition
In this mode, it is possible to check, adjust(left or right) and save the
collimator’s location easily by using control panel. Also this mode helps
to do alignment operation more easily.
① Entering panorama collimator check mode

At the panorama mode, press Left and Right button at the


same time for 3 seconds. (It will display current collimator location
value)

② Control method

On Collimator check mode, set the collimator location by using

Left and Right button and press button for three senconds. Then
displayed value will be saved with beep sound.

③ Getting out from Panorama, CT and Ceph collimator check mode

Press Left and Right button at the same time for three seconds
then it will exit collimator check mode.

Installation Point 200HD 47 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
7. Change default value of kV, mA
This setting is related to memory functions by saving basic value of kV,
mA of child, woman and man, and this function provides convenience
for users making by not to input different values every single time.
① Select scanning mode(PA, TMJ, CT, CEPH)

② Select or or

③ Select ,

④ Use , button to adjust

⑤ Check LED by pressing button

⑥ Press (OK) button for 3 seconds to save(buzzer sound)

Installation Point 200HD 48 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
8. ERROR Massages & Code
▶ HARDWARE
The purpose of Error message is to find problem of device easily and
listed 19 errors will help to find problem of device. Each error message
is displayed at control panel and the device will not temporary operate
when error message is displayed.
◈Error Code
(1) HFG Error(feedback)
① ER PS 1
A. Description: Tube Fail error (kV feed value is not
entering from HFG)
B. Follow up action: X-ray is disconnected with hardware
This is occurred disconnection of 300V between inverter
board and power board
C. Checklis
- Feedback line
- Feedback connection line (Check tank, inverter, flat
cable)
- 300V line of inverter board and power board (1:1)
② ER PS2
A. Description: Temp Fail Error (HFG Tank is warning
overheating problem)
B. Follow up action: X-Ray is disconnected with software.
C. Debug: Check HFG board’s LED1 is ON.
f it is ON, please use it again after certain cooling time is
passed.
D. Checklist
- Pull out feedback cable’s connecter in inverter board
while device’s power is ON, and then check if LED1 is
ON (Normal condition is LED ON.)
- Check disconnection of flat cable or connector’s
connection status.
③ ER PS3
A. Description: kV OK Fail Error (HFG kV feed value is
different from kV reference value)
B. Follow up action: X-Ray is disconnected with software.

Installation Point 200HD 49 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
- Due to disconnection or short of feedback line, check
zener diode of inverter board or check X-Ray tube
- This is occurred when tube, inverter is damaged or main
board’s feedback chip is damaged.
C. Checklist
When you expose beam while device’s power is ON,
check status of LED 15 of inverter board.
- LED15 is OFF: Check tube, inverter, and power board.
- LED15 is ON: Check main board.
④ ER PS 4
A. Description: mA OK Fail Error (HFG mA feed value is
different from mA reference value)
B. Follow up action: X-Ray is disconnected with software.
- This is occurred when feedback line which is inserted
from tube is old version.
C. Check list
1. Check feedback line’s version1
2. When you expose beam while device’s power is ON,
check status of LED10 of inverter board.
- LED10 is OFF: Check tube, inverter, and power board.
- LED10 is ON: Check main board.

(2) Sensor, micro switch error


Error is indicated by detecting malfunction of sensor and micro
switch and judging incorrect location. Check error list of sensor and
micro switch and prevent malfunction of the device.
① Rotation Unit Motor Photo Sensor Error

A. (ARM Motor photo sensor1 Error)

B. (ARM Motor photo sensor2 Error)

C. (ARM Motor photo sensor5 Error)


Symptom: Fail to initialize the device.
② Y axis Motor Micro Switch

A. (Rail Motor micro switch1 Error)

B. (Rail Motor micro switch2 Error)

Installation Point 200HD 50 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE

C. (Rail Motor micro switch3 Error)


Symptom: Fail to initialize the device It is not operating
when the error displays.
③ Detector Motor Micro Switch Error

A. (Detector Motor micro switch1


Error)

B. (Detector Motor micro switch2


Error)
Symptom: Fail to initialize the device.
④ Collimator Micro Switch Error

A. (Collimator Motor micro switch1


Error)

B. (Collimator Motor micro switch2


Error)
Symptom: Fail to initialize the device.
⑤ TANK Micro Switch Error

A. (HFG Motor micro switch1 Error)

B. (HFG Motor micro switch2 Error)


Symptom: Fail to initialize the device. It is not operating
when the error displays.
⑥ Up/Down Motor Micro Switch Error

A. (Up/Down Motor micro switch1


Error)

B. (Up/Down Motor micro switch2


Error)
Symptom: Fail to detect micro switch of Up/Down part or
the location is incorrect.

Installation Point 200HD 51 / 60


9. TECHNICAL
PNX.IM.033.0(E) (2013.05.20) FEATURE
⑦ ER SR CP
A. Description: Ceph sensor check Error
(Sensor is not checked when scanning the Cephalo)
B. Follow up action: X-Ray is disconnected with
software.
- Tube is not in the correct location.
- Sensor 7 is not detected.
Check list: Check tube’s location and photo sensor’s
operation.

Installation Point 200HD 52 / 60


PNX.IM.033.0(E) (2013.05.20) 10. MAINTENANCE
1) Use, Storage, Management
(1) Use and Environmental requirements
① Measurement spaces for patient: To properly adjust the device and
patient's positioning, the installation of inner parts of the device has
to keep at least 200cm (W) & 155 cm distances.

② Width of door: Minimum 120cm

③ Height: minimum 240cm

④ Floor: It has to support at least 500 Kg/m

⑤ Operation Requirements
Choose a site falling within the following temperature and humidity
ranges.
■ Temperature: 10 ~ 40℃
■ Humidity: 95%

⑥ Storage Requirements
■ Temperature: -10 ~ 50℃
■ Humidity: 20 ~ 95%
■ Atmospheric pressure: 600~1060Hpa
(2) Maintenance after use
① The device should be cleaned for the next use. Contaminated
materials are wiped with chemical solvents.
Allowed chemical solvents for disinfection
: Acetylene, Butyl Alcohol, Ethanol 96%, Methanol, Soap
Restricted chemical solvents for disinfection
: Benzene, Acetone, Acetic Ether

Installation Point 200HD 53 / 60


PNX.IM.033.0(E) (2013.05.20) 10. MAINTENANCE
2) Maintenance, Warranty

(1) Maintenance
This device is designed to improve customer satisfaction with supreme
functions. To makre sure its safe operation, we recomment a periodic
inspection at least for one time per year.

① User managment
If problems are occurred, please immediate contact your local distributor
of PointNix Co., Ltd for further instructions.

② User inspection
User is asked to inspect these details in below every month.(Monthly
inspection is required for this information in below.)
■ Check if Ready button should be ON within two minuted after power
is connected.
■ Check signal sound is normally working while scanning.
■ Check all lampes in control panel are ON.
■ Check all buttons in control panel are working properly.

Installation Point 200HD 54 / 60


PNX.IM.033.0(E) (2013.05.20) 10. MAINTENANCE
(2) Warranty
PointNix Co., Ltd guarantees that the device is free from the manufacturing
problems for a year from shipping date. For X-ray tube, PointNix will
provide the warranty for 1 year or 10,000 times of scanning which comes
first. Warranty period is valid for one year from the shipping date. Any
repair work or replacement that is occurred after this period is not
considered as warranty and additional expenses are charged for these
works after warranty period.
Warranty shall not apply to natural disasters (including fire, earthquake and
other acts of GOD), the neglect, misuse or other-than-ordinary use of the
device, and attempted repair or alteration.

Installation Point 200HD 55 / 60


PNX.IM.033.0(E) (2013.05.20) 10. MAINTENANCE
3) Waste/Deliver

(1) Waste
This device is basically designed to saftely use and dispose of device
to reduce the environmental pollution. Except X-ray tube and such a
several part, exterior and interior part is possible to recycle as
following use.
All the part or component that inculdes dangerous substance must be
seperatly disposed by regulating laws.

Main part need to Dangerous,


Part Recycle Waste disposal
dispose Separation
Aluminum,
Frame, Cover o
Plastic
Motor o
Board o
Cable Cop, Iron o
Wood, Board,
Package o
Paper
X-RAY Tube o
Sensor head Please return to PointNixhead office.
Other parts o
Ask detail instructions to your local distributor.

(2) Ways to be delivered/transported


Air, Cargo, Ocean

Installation Point 200HD 56 / 60


PNX.IM.033.0(E) (2013.05.20) 11. INFORMATION
■ Product Name
x-ray system, panoramic/tomographic, digital

■ Model
Point 200 HD

■ Input Voltage
100~120V/220-240V~

■ Input Frequency
50/60Hz

■ Input Current
15/8A
■ Fuse
T25AL 250V~

■ Filter
X-ray own filter 0.8mmAl
Aluminum
Minimum total filter 2.7mmAl

Installation Point 200HD 57 / 60


PNX.IM.033.0(E) (2013.05.20) 11. INFORMATION
■ Maximum allowed rating
2.0KVA
■ Device Type
Fixed Type
■ Operating mode
Discontinuous behavior
■ Apparent resistance
<1Ω
■ Applied standard
IEC 60601-1
IEC 60601-1-3
IEC 60601-2-7
IEC 60601-2-28
IEC 60601-2-32
IEC 60601-1-2
■ Class
According to the type of protecting against electric shock
CLASS Ⅰ EQUIPENT
According to the degree of protecting against electric shock
TYPE B APPLIED PART
■ EC Representative
DTF TECHNOLOGY Srl
ADDRESS : Via Castel Morrone 18, 20219 Milano, Italy

TEL : +39-02-84893641, Fax : +39- 02-8471894

Installation Point 200HD 58 / 60


PNX.IM.033.0(E) (2013.05.20) 11. INFORMATION

Copyright ⓒ 2013
All rights reserved by PointNix Co., Ltd.

The information contained in this manual, the copyright laws of the


Republic of Korea, and computer programs are protected by the Act.
Contents of the manual in any form without prior written consent of
the manufacturer to modify, change, or cannot reproduce.

Information of the manual can be changed without consent of the user


according to improvement and complement of the product. For more
information about the product or manual, or any questions, please
contact our nearest dealer or head office.