PNX.IM.033.0(E) (2013.05.20) CONTENTS
CONTENTS
1. OVERVIEW ..................................................................................................................................... 3
2. FEAUTRES AND PURPOSE ......................................................................................................... 4
1) FEATURES ................................................................................................................................... 4
2) PERPOSE ...................................................................................................................................... 4
3. STRUCTURE ................................................................................................................................... 5
4. SAFETY ............................................................................................................................................ 6
5. PRECAUTIONS .............................................................................................................................. 7
6. SPECIFICATIONS........................................................................................................................ 11
7. COMPOSITION ............................................................................................................................ 14
1) PANORAMA ............................................................................................................................... 14
2) CONTROL PANEL ..................................................................................................................... 16
3) SID DISTANCE MEASURE .................................................................................................... 18
4) ACCESSORY .............................................................................................................................. 19
8. INSTALLATION ........................................................................................................................... 20
1) HOW TO INSTALLATION ......................................................................................................... 20
2) CONNECTION BLOCK DIAGRAM ......................................................................................... 21
3) PROGRAM INSTALLATION..................................................................................................... 22
4) ALIGNMENT AND CALIBRATION ......................................................................................... 23
5) CDX-VIEW IMAGE ................................................................................................................... 27
9. TECHNIOCAL FEATURE ........................................................................................................... 28
1) BLOCK DIAGRAM.................................................................................................................... 28
2) CONNECTION DIAGRAM ....................................................................................................... 29
3) INFORMATION .......................................................................................................................... 33
4) GENERAL DATA ....................................................................................................................... 35
5) DISPERSION OF EXPOSURE DOSE & RECEPTION ............................................................ 37
6) SERVICE ..................................................................................................................................... 38
10. MAINTENANCE ......................................................................................................................... 55
1) USE, STORAGE, MANAGEMENT ........................................................................................... 55
2) MAINTENANCE, WARRANTY ............................................................................................... 56
3) WASTE/DELIVER...................................................................................................................... 58
11. INFORMATION .......................................................................................................................... 59
① Features
(1) High resolution HD sensor
(2) Secured X-ray exposure
(3) Clear and accurate images due to CsI sensor
(4) Low radiation dosage due to high absorptance
(5) Scanning and receiving images at the same time
(6) Centralized membrane switch
(7) Integrated software program, CDX-View
CDX-View: This is a tool of digital diagnosis system which implements all
images from Panorama, Cephalo, CT, Intra Oral Camera, Digital Camera,
and Intra Oral Sensor etc. This is represented as an integrated data system
of X-ray clinics on Network use.
② Purpose
The equipment is to acquire tomographic images of teeth, jaw, oral structure
and panoramic etc. by transforming x-ray exposed signals to digital signals
for diagnosis purposes.
Symbols Description
Direct risk of death, serious adverse reaction, serious
damage to the properties, or fire.
2) Safety Regulations
3) Danger
(1) This device’s position beam can cause impaired vision. Do not aim position
beam direct to patient’s eyes and guide patient not to look at the position
beam.
(1) Staff must obligatorily provide protection material such as lead apron to
patient when they are scanning image.
(2) Frequent exposed to X-Ray can cause cancer and various diseased. So staff
must protect themselves by lead apron or lead protect.
(3) Staff must keep the distance at least 2meters (7ft.) from focus and X-Ray
to hand exposure switch.
(4) The device must locate inside of shield room and staff should keep eyes on
scanning through outside window.
(5) Person who needs special attention such as children, pregnant woman,
elderly and unhealthy patient must counsel with medical attendant before
scanning.
(6) During the scanning, staff must keep check the device and patient for in case
of emergency.
(7) During the scanning, stop the device immediately by pressing emergency
button if it works unusual.
(8) Patients and staffs are prohibited long-time expose to X-Ray and take
enough rest because excessive use of X-Ray can cause various diseases.
(9) After scanning, it should wait 6minutes of cool time for next scanning.
The law relate to Radiation Safety and Protect is different for each Country.
User must observe the Law of Radiation Safety and Protect for their own
country.
■ OS : Windows XP or higher
■ CPU : Intel Core i5-2400 or higher
Imaging ■ GPU : NVidia GPU 1GB or higher
control ■ RAM : 3G or higher
4 PC
device ■ HDD : 500GB free hard disk space or higher
(Option) ■ Network : Gigabit Ethernet Network card
■ Display : 32bit color display or higher
■ Resolutions : 1,280 X 1,024 or higher
4
3
5
6
8 1
11
9
10
2 Rotation Unit Rotation part following to the teeth locus while scanning
10 Base plate Support and fix the whole body of the device
ⓕ X-ray lamp Indicator the means images are on the process of taking
ⓠ Image taking Move rotation arm back or forth while image scanning
ⓡ location of
(ⓠ-Protrusion, ⓡ-Normal, ⓢ-Depression)
ⓢ rotation unit
SID location
Condition SID(mm)
PANORAMA 17s
Tmj 11s
PANORAMA
270°
Tmj
Main Unit(200HD)
③ Detail explanation of installation step is as below.
Prepare Main Unit
Move the packed Main Unit to the location where you will install, and
adjust the level with adjusting screw.
Point 200HD
Point 200HD
CN18 Expos S/W
CN13 PANORAMA
USB or
Serial Cable
LAN Card
PC
2. CDX_HD_PANO_Setup Installation
Smart Sensor(PANORAMA)
2. Click
‘Start
Testing’
Check ‘Cephalo’ is
selected
Displays detected X-
ray at Active Area of
sensor in real-time
(Create
Calibration file)
2. It displays ‘Device 311 error’ error message at the first time when you
run exposure program. This error code is due to no applied calibration
file. Enter configuration window, and setup the saved calibration file as
below picture. Then save and exit the configuration to scan
Panorama/Cephalo image. (Panorama & Cephalometric are the same)
▶ How to connect PWR Board & Detector (Pano) & Detector (Ceph)
& Main Control Board
PWR Board Detector(Pano) Detector(Ceph) Main Control
Board
CN16 DC 12V DC 12V Input CN41(DC 12V
Output Input)
CN17 DC 5V DC 5V Input
Output
X-ray On Signal CN13
X-ray Gen CN44
Hand S/W CN43
2) Information
3) General Data
1. Electrical
A. Circuit (Center-Grounded) : Single-phase full-wave rectified or DC
B. Operating Tube Voltage : 50 ~ 90kV
C. Focal Spot : 0.5mm
D. Input Energy (at 1.0s) : See Rating Chart
E. Single-phase full-wave rectified: 840W
F. DC : 1750W
2. Mechanical
A. Dimensions : See dimensional outline
B. Overall Length: D-054SB - 146 mm
C. Maximum Diameter: 054SB - 58 mm
D. Target Angle : 5 degrees
E. Inherent Filtration : At least 0.8 mm Al equivalent at 50 kV
F. X-ray Coverage :75 × 380 mm at SID 550 mm
G. Weight:D-054SB - Approx. 780 g
H. Cooling Method :Oil immersed (60℃Max.) and convection oil
cooling
I. Tube Holding : D-054SB - Holding the lead cylinder
5) Service
1. How to set up Test Mode
This is a mode to test whether device is operating or not under the
condition that X-Ray is not exposed.
▶ Start Test Mode
Before power is on, press Down and Right button at the same time then
turn the power switch on.
▶ How to operate
This is same as user manual but X-Ray is not exposed.
2. How to set up Service Mode
X-Ray is only detected and rotation mechanism is not operating.
Mode
Key Sequence For details, refer to
Setting
A4 Error check(mA_OK,kV_ok)
A5 KV Calibration adjustment
Use buttons on log in
A6 MA Calibration adjustment
A9 Maximum value of mA
AA Minimum value of mA
Mode
Key Sequence For details, refer to
Setting
Mode
Key Sequence For details, refer to
Setting
③ Login Screen
(1) Checking the Firmware version.
Anterior number is the Main Board(MCU board) version.
Posterior number is the Program version.
(5) Mode A4 is to set error check (mA OK, kV OK). If you press
RIGHT key, error is going to be checked, and if you check
LEFT key, error is not going to be checked.
4. Test Mode
This mode is to check operation of the device rotating without X-ray
exposure. Therefore, it is possible to check device’s rotation and all
other conditions include whether any parts has problem or not. Also it is
possible to understand the distance of patient and sensor during the
device’s rotation
① Entering Test Mode
② Controlling method
Same as the User Manual accept no X-ray exposure during rotation.
5. Service Mode
Device exposes X-ray without rotation.
Press UP and + button at the same time then turn ON the device.
② Controlling method
Adjust kV, mA, Sec values as wanted then press exposure S/W.
③ Usable keys
kV, mA, Sec, Laser, memory function(C, W, M), UP/Down, Stop
② Control method
Left and Right button and press button for three senconds. Then
displayed value will be saved with beep sound.
Press Left and Right button at the same time for three seconds
then it will exit collimator check mode.
② Select or or
③ Select ,
⑤ Operation Requirements
Choose a site falling within the following temperature and humidity
ranges.
■ Temperature: 10 ~ 40℃
■ Humidity: 95%
⑥ Storage Requirements
■ Temperature: -10 ~ 50℃
■ Humidity: 20 ~ 95%
■ Atmospheric pressure: 600~1060Hpa
(2) Maintenance after use
① The device should be cleaned for the next use. Contaminated
materials are wiped with chemical solvents.
Allowed chemical solvents for disinfection
: Acetylene, Butyl Alcohol, Ethanol 96%, Methanol, Soap
Restricted chemical solvents for disinfection
: Benzene, Acetone, Acetic Ether
(1) Maintenance
This device is designed to improve customer satisfaction with supreme
functions. To makre sure its safe operation, we recomment a periodic
inspection at least for one time per year.
① User managment
If problems are occurred, please immediate contact your local distributor
of PointNix Co., Ltd for further instructions.
② User inspection
User is asked to inspect these details in below every month.(Monthly
inspection is required for this information in below.)
■ Check if Ready button should be ON within two minuted after power
is connected.
■ Check signal sound is normally working while scanning.
■ Check all lampes in control panel are ON.
■ Check all buttons in control panel are working properly.
(1) Waste
This device is basically designed to saftely use and dispose of device
to reduce the environmental pollution. Except X-ray tube and such a
several part, exterior and interior part is possible to recycle as
following use.
All the part or component that inculdes dangerous substance must be
seperatly disposed by regulating laws.
■ Model
Point 200 HD
■ Input Voltage
100~120V/220-240V~
■ Input Frequency
50/60Hz
■ Input Current
15/8A
■ Fuse
T25AL 250V~
■ Filter
X-ray own filter 0.8mmAl
Aluminum
Minimum total filter 2.7mmAl