Professional Documents
Culture Documents
C2 - mp7547 Sup 0004
C2 - mp7547 Sup 0004
1 Procedure failures
2.1 – 2.3 Equipment availabilit or Wrong dose distribution
effectiveness Wrong volumes
Delineate GTV/CTV 2. Excessive
2.5-2.6 User error
(MD) and other delineation errors
3.2 Desgin Bad drawing leads directly to geometric errors, wrong plan
59 7 - RTP Anatomy structures for resulting in <3*sigma 5.86 6.57 8.00 325.7 5
3.5 – 3.6 Assessment or or dose distribution.
planning and segmentation errors
programming
optimization
6.1 Inadequate training
Inattention, lack of time, failure to
review own work
3. Margin width
protocol for PTV
1 (all) Procedure failure
construction is Wrong dose distribution
5.2-5.4 Communication failure Very common, since many depts don’t assess consistency
inconsistent with actual Wrong volumes
65 7 - RTP Anatomy PTV construction 6.1Inadequate training 7.29 5.43 7.86 316.3 6 between margins and actual setup errors. Very difficult to
distribution of patient Or
Inattention, lack of time, failure to detect when discrepancies are modest.
setup errors. Suboptimal plan
review own work
5.4. Miscommunication; Physician reviewed one of the trial plans rather than the
inattention, final plan that dosimetrist intended for MD approval. The
1.1. lack of procedure Very wrong dose; Very trial plan may have severe problems that the MD does not
1. Plan OK to go to 1.2 Inadequateprocedure wrong dose distribution notice thinking that the plan meets prescription and
137 9 - Plan Approval 3. Bad plan approved 4.89 8.00 7.89 313.3 7
treatment 1.3 Procedure not followed Very wrong volume planning criteria.
2.5 Used incorrectly
6.1 Inadequate training/orientation
Example: For some kinds of breathing motion, phase-
gating on RPM can gate at the wrong phase of the
breathing cycle (inspiration rather than expiration). If this
is not recognized, for one or more sessions, treatment of
a patient with large breathing motion can miss parts of the
tumor because the beam is on in the wrong part of the
3.2. Poorly designed software for
breathing cycle. To add to the problem, the GE 4D CT
task at hand.
Special motion Wrong dose phase calculation algorithm differs from the RPM
3.2. Poorly designed hardware.
management methods Wrong dose distribution algorithm, so a phase gate chosen on the basis of a GE
12 - Day N Set treatment 6.1. Use by inexperienced personnel
205 (e.g. gating, breath- Wrong location 6.22 6.67 7.11 310.1 8 4D study doesn’t necessarily carry over even qualitatively
Treatment parameters 2 Operator doesn’t observe counter-
hold) not applied or Wrong volume to RPM treatment!
intuitive patterns on screen (6.1. lack
incorrectly applied Example: The patient’s breathing trace for amplitude
of training, Lack of attention)
gating can drift (presumably due to overall ant/post body
motion of < 1 cm- part of body with gating block “settling”)
causing the RPM to gate on the wrong phase of the
breathing cycle.
Effect: Radiation delivered during the wrong part of the
breathing cycle with anatomy in wrong position.
Lack of dept policies/documentation Pacemaker not noticed in treatment volume for radical
Special instructions not 8.2 (Lack of time) patient.
Specify special given 8. (Lack of attention to detail.)
6 - Initial Treatment instructions, viz. 5.4 ,5.1 (Miscommunication)
Non-radiation related
46 Planning Directive pacemaker, Wrong special 5.25 8.75 6.50 305.9 9
injury
(from MD) allergies, voiding, instruction (e.g., allergy,
bowel prep, etc pacemaker) 2.2 (Wrong or inadequate info
obtained)
Manual entry: Large error Very wrong dose Very Some planning systems may not upload all treatment
1. Incorrect Tx info, 2.
1.2 Inadequate procedure wrong dose distribution parameters digitally to R&V. Error >10% is made in
10 - Plan Wrong Rx,
168 9. Prepare e-chart 1.3 Procedure not followed Very wrong location 5.44 8.89 5.56 272.9 15 manual entry of fractions or dose carried.
Preparation 3. Wrong patient/plan
6.1 Inadequate training/orientation Very wrong volume
Incomplete/ incorrect Important OAR not contoured and not taken into account
6.1 (Ignorance of available imaging
list of specified in design of treatment plan
Specify protocol for studies)
structures and
delineating target 5.3 (Miscommunication)
6 - Initial Treatment corresponding image
and structure (which 1.2, 4.5 (Ambiguous labeling of
41 Planning Directive sets. Wrong volume 6.22 7.67 5.56 266.4 17
structures to contour image sets)
(from MD)
and CTV delineation 1.1 (Lack of eplicit protocol)
CTV maybe incorrectly
guidelines) 8. (User error)
contoured
2. Enter prescription 2. Incorrect dose Explanation in the above box (#17) applies but in this
8 - Treatment
95 + planning prescription and dose B. 8. Human error by the planner Very wrong dose 5.67 7.78 5.22 245.9 21 case instead of the physician making the error, the
Planning
constraints limit constraints planner makes the error.
1. Omission in entry Historical information about patients treated in the dept. is
maintained in a database. The archive is incomplete and
Entry of patient data
1- Patient 2. Incorrect or 8. (Human transcription error), 1.2, there are no other records (or no effort is made to get the
in electronic
3 Database incomplete previous. 1.3, 1.4 Very wrong dose 4.13 8.88 6.88 245.8 22 other records such as paper chart). The patient & family
database or written
Information treatment history don’t remember treatment (or details) and no-one notices
chart
2. Incomplete patient history physical changes from the prior treatment (tattoos). Part
5.3 (Source of data incomplete) of previously treated area is re-treated.
Specify inappropriate Omit 4D
protocol 6.1, 1.4 (Ignorance of available
Specify inappropriate methods, protocols, protocol
6 - Initial Treatment Specify image tolerances for requirements) Wrong location
43 Planning Directive registration goals – registration 5,3 (Miscommunication) Wrong volume 5.71 7.00 6.00 245.3 23
(from MD) 4D Specify wrong 1.1/1.2 (Lack of eplicit protocol)
reference image 8. (User error)
dataset (primary)
2. Wrong treatment 5.2 Poor communication With IMRT treatments, this failure is probably less likely
11 - Day 1 Set treatment accessories Human error Very wrong dose Very unless using a compensator-based approach.
189 5.56 8.00 5.33 240.0 24
Treatment parameters (Missing/incorrect Incorrect documentation wrong volume
bolus, blocks)
2. Wrong imaging study 2.5 User error
(correct patient) 1.1-1.6 Standards, procedures,
Wrong dose distribution Choose inhale rather than exhale CT study, or MR T2
Viz.; wrong phase of 4D protocols
Import images into Wrong volume rather than MR T1-gadolinium scans for anatomy and
52 7 - RTP Anatomy CT selected for 5.2-5.4: Communication (poor) 5.38 7.25 6.13 239.6 25
RTP data base target volume. Leads to wrong anatomical model (leading
planning; wrong MR for 6.1 Inadequate training
to systematic geometric and dosimetric errors)
target volume
delineation
3.2, 1.2, (Lack of dept Plan design based on incorrect tolerance dose for critical
policies/documentation) OAR
Inappropriate or
4.6 (Lack of time)
incomplete target
5.3 (Miscommunication)
6 - Initial Treatment Specify dose limits doses
6.1 (Inadequate training) Moderately suboptimal
47 Planning Directive goals, and and/or normal tissue 4.89 7.78 6.00 234.4 26
plan
(from MD) fractionation constraints specified or
5,1, 8.6 (Planner selects wrong
assumed.
protocol in absence of clear
physician directive)
Localization; (port
11 - Day 1 2. Images Human error;
187 films and/or other Wrong location 5.44 7.56 5.44 234.1 27
Treatment misinterpreted Inappropriate settings
localization devices)
2.2. Software error; All causes listed in the cause column may result in an
User selects wrong parameters for incorrect evaluation of DVH or IDLs etc.
DVH.
Wrong or inappropriate structure(s)
14. Evaluate plan
8 - Treatment Incorrect DVHs, IDLS, for DVH(s)
128 (DVH, isodose, dose Very wrong dose 5.00 8.11 5.44 230.7 28
Planning etc 3.2. Poor evaluation tools
tables, etc.)
2.5. Wrong normalization point
selected.
Inadequate commissioning
1 Procedure failures
2.1 – 2.3 Equipment availabilit or
effectiveness
2.5-2.6 User error
3.2 Desgin
Boolean Very wrong dose Plan based on DVH for incorrectly combined structs,
1. Wrong structures 3.5 – 3.6 Assessment or
66 7 - RTP Anatomy combination of distribution 4.75 7.88 5.75 230.0 29 easily results in wrong plan optimization decisions, and
combined programming
delineated structures Very wrong volumes very wrong dose distribution and/or volumes.
5.2 Lack of communication
5.4 Misunderstanding
6.1 Inadequate training
Inattention, lack of time, failure to
review own work
The cause may also be a “software limitation” – e.g., a
radiological path length correction may be correctly
implemented but may not be the best algorithm to use in a
2. or 3. Software or design error.
particular situation; it’s not a software error but rather a
3.2. Software limitation
8 - Treatment 6. Heterogeneity 1. Wrong or poor Wrong dose distribution; limitation. This situation probably occurs frequently – a
108 2.5. User choice of wrong correction 5.63 5.25 7.63 230.0 30
Planning correction algorithm Wrong dose heterogeneity correction algorithm that is not appropriate
model
to the situation is used for optimization and dose
calculation . Measurements in a situation duplicating the
treatment situation are difficult – most confirming
measurements are made in uniform phantom.
Inadequate characterization of Acquired data quality is not good. Input data is not
delivery system in the RTPS; measured accurately. This will result in poor beam
5. Calculate the
Absolute output modeling.
dose to the
Relative dose inside beam
8 - Treatment optimization points, Wrong dose; Wrong dose
107 2. Poor beam modeling Penumbra, leaf edge/end modeling 5.44 5.11 7.89 229.4 31
Planning and also calc the distribution
Outside beam
final dose
Buildup region
distribution
Transmission
Pacemaker, prior treatment, etc.. Very wrong dose Very Pacemaker avoided in the treatment plan but forgets to
1. Entry of 3. Critical patient 1.2 Inadequate procedure wrong dose distribution make note to restrict size of port film or MV cone beam;
10 - Plan
147 demographic information not 1.3 Procedure not followed Very wrong volume 3.89 8.56 6.56 213.4 49 therefore, device receives dose that makes it malfunction
Preparation
information recorded 6.1 Inadequate training/orientation Non-radiation dose injury
Wrong user input. User error results in setup DRRs at wrong angles. Patient
4. Prepare DRRs, Incorrect images 1.2 Inadequate procedure is shifted unnecessarily
10 - Plan
159 other localization (wrong angle, wrong 1.3 Procedure not followed Very wrong volume 4.78 8.33 5.11 212.8 50
Preparation
imaging data divergence, etc) 6.1 Inadequate training/orientation
5. Misleading or
1 Procedure failures
erroneous structure
2.1 – 2.3 Equipment availabilit or
names and/or other
effectiveness
Delineate GTV/CTV labels (e.g., color)
2.5-2.6 User error Very wrong dose
(MD) and other resulting in incorrect Wrong names or labels of structs can result in wrong use
3.2 Desgin distribution
62 7 - RTP Anatomy structures for correspondence of 5.00 7.00 5.86 210.0 55 of struct info, leading to overdose to normal tissue or
3.5 – 3.6 Assessment or Very wrong volumes
planning and treatment underdose/miss of target
programming
optimization goals/constraints and
6.1 Inadequate training
structures
Inattention, lack of time, failure to
review own work
Examples: Therapists not required or not trained to check
SSDs that would indicate a change in patient shape (e.g.
AP SSD for patient set up by lasers on side tattoos).
Loose mask (which might indicate tumor shrinkage or wt
Changes in patient loss) ignored on H&N patient.
geometry (thickness Effects: If patient thickness has changed by several cm,
12 - Day N Patient set up for changes due to wt gain 1.2. Inadequate department policy Wrong dose Wrong the dose/dose distribution is incorrect. If the
198 5.89 5.44 6.56 209.9 56
Treatment treatment or loss, tumor growth or Inattention dose distribution immobilization has gotten loose, patient motion can result
shrinkage) ignored or leading to wrong dose distribution.
not recognized. Secondarily –if the MD is not informed about unusual
changes in thickness, tumor growth or radiation-
complication induced weight loss can be neglected for
several days, leading to medical problems.
1. Incorrect plan info User error causes incorrect transfer of sequencing data
Very wrong dose Very
10. Download into delivery system : 2. and/or treatment parameters from R&V to LINAC.
3.2. Poor interface leads to user wrong dose distribution
10 - Plan complete delivery Connect wrong
171 error Very wrong location 3.89 8.67 5.33 187.7 76
Preparation plan to delivery patient/plan in RTP with
3.6 Inadequate programming Very wrong volume
system wrong patient/plan in Tx
Delivery system
11 - Day 1 Position patient for 1. Incorrect treatment C. 2.2 Device failure
183 Wrong location 4.78 7.00 5.56 186.7 77 Lasers misaligned. Incorrect use of IGRT system
Treatment treatment isocenter. 6.1 poorly trained personnel
5.4. Miscommunication; There are two sets of fields (initial and boost). Therapists
inattention, treat both sets of fields due to their approval and
2. Completion of 1.2 Inadequateprocedure availability in R&V system
5. Premature
142 9 - Plan Approval formal prescription 1.3 Procedure not followed Wrong dose fractionation 4.33 8.22 4.78 185.4 78
signaturization
after planning 2.5 Used incorrectly
6.1 Inadequate training/orientation
Pull up wrong patient’s record Patient treated on the basis of another patient’s volumes
{usually carelessness ! Not covered
in Canadian table} Very wrong dose
5 - Transfer images
Transfer primary Incorrect CT data set Closest categories are 2.5 distribution
32 and other DICOM 3.78 8.78 5.78 181.3 82
(CT) data set associated with patient (Materials/tools/equipment used Very wrong volume
Data
incorrectly), 6.1 Inadequate training
and 8.4 fatigue.
8 - Treatment 6. Heterogeneity Wrong dose; Wrong dose Planner wanted to incorporate heterogeneity correction
109 2. Wrong on/off choice Human error 5.00 6.00 5.89 179.8 87
Planning correction distribution but by error did not turn the heterogeneity correction on.
Immobilization device excluded from dose calculation and
Properties of device not Inadequate commissioning: device
Radiological attenuation through the device was omitted from patient
consistent with properties not assessed
properties of dose distribution.
accurate dose delivery .
positioning aids are
2- Immobilization or artifact-free Communication failure so not Wrong absolute dose;
8 known to treatment 6.22 4.67 5.78 179.6 88
and Positioning verification imaging included in treatment plan wrong dose distribution
planners. (e.g.
attenuation, skin-
Inadequate procedures 2.6: Inadequate assessment of
sparing)
material/tools/equipment
3. Setup treatment 5.3. Ambiguous written directive and A beam energy is chosen that dos not return optimal dose
fields (machine, or communication. distribution.
8 - Treatment 4. Poor selection of
99 energy, MLC, beam Human error by the Planner Poorly optimized plan 5.63 4.88 6.13 179.4 89
Planning beam energy
angle, , beamlet 8. Poor choice of beam energy
size, etc.)
Overlap regions not handled correctly because of
software error.
1. Classification of overlap or non-overlap structure is not
1. Incorrect
handled correctly. For example, define a structure (instead
classification of a
8 - Treatment 1. Specify ROI for of PTV) with a certain name as overlapping structure but
81 structure as B. 2.2. Software error Very wrong dose 3.44 8.11 6.22 179.3 90
Planning optimization process because of software error it does not treat it as
overlapping or non-
overlapping structure.
overlapping
2. Part of the region of an overlapping structure overlaps
with other structures even though it is defined as non-
overlapping structures and vice-versa.
6.1. Poor training. Similar to the last one.
Inattention.
8 - Treatment 11. Evaluate leaf 2. Incorrect evaluation 3.2. Poorly designed evaluation
125 Very wrong dose 3.78 7.89 5.89 179.3 91
Planning sequences (ie, ok when not, etc) tools,
2.2. Software error.
Inadequate commissioning Everything stated in the cause column can cause error in
12. Transfer
2.2. Incomplete or failed transfer. the final dose calculation.
sequencer results to RTP (Dose calculation
8 - Treatment 2.2. Software error
126 RTP ( should it not algorithm) gets wrong Very wrong dose 4.00 7.67 5.22 170.8 100
Planning Bad data
be TPS?) for final intensities.
Data handling error
dose calc (if needed)
Bad interface definition
2.5 Human error
6.1 Inadequate training and 8.4
fatigue.
But could also be.. Very wrong dose
1. Wrong patient’s 1.1-1.6 distribution Select images from wrong patient (name collision, human
Import images into
51 7 - RTP Anatomy images selected or Standards/procedures/protocols Very wrong volume 4.38 7.88 4.63 170.5 101 error). Causes incorrect patient anatomical info or mixes
RTP data base
imported (commissioning/acceptance testing) data from two patients.
2.2 Defective materials (software or
hardware)
3.2 Inadequate design specifications
3.6: Inadequate programming
3. Setup treatment Human error by the Planner. Very wrong dose Planned isocenter is different from marked isocenter for
fields (machine, 2.2. Software error in information distribution. Very whatever reason i.e., instruction is not clear,
8 - Treatment 6. Incorrect selection of
101 energy, MLC, beam transfer wrong location 4.22 7.78 4.78 169.3 102 communication is not established planner and therapists.
Planning isocenter
angle, , beamlet Wrong information input at sim Wrong dose
size, etc.)
Plan based on incorrect geometry – wrong treatment data
Documentation of
sent to machine (potential for left-right reversals). Patient
unusual simulation
treated incorrectly because simulation conditions are not
attributes (in print or
duplicated. Plan based on incorrect geometry – wrong
software), e.g., 1.2 (inadequate procedures or forms
treatment data sent to machine (potential for left-right
Unusual patient for capturing attributes)
Simulation attributes Very wrong location reversals).
positioning (prone or
17 3 - CT/simulation not correctly Wrong dose 4.22 8.33 4.89 168.8 103
feet-first) or patient 8. (Inattention to detail.)
documented.
preparation needs 6.1 (Poorly trained personnel.)
(e.g. simulated with
full bladder so need
to treat with full
bladder”)
1. Bad treatment The dose grid resolution may not be correct (viz. the
decisions due to bad dose policy may be to choose a dose grid of 2x2 mm but the
4. Setup dose
distribution. 2. Wasted planer chooses a grid of 5x5 mm. For small structure this
calculation
8 - Treatment 2. Poor parameters Human erro time and effort/ may lead to inaccurate dose calculation.)
104 parameters (eg 4.50 5.63 6.25 166.5 107
Planning selected 1.2. Wrong departmental policy 3. Patient treatment
calculation grid or
delayed
dose point density)
4. Undeliverable plan
5.4. Miscommunication Exported a treatment plan for wrong Mr. Jones to record
inattention and verify system. The error could be discovered during
1.1. Lack of procedure; Very wrong dose; Very the import into record and verify system or during plan
1. Plan OK to go to 1.2 Inadequate procedure wrong dose distribution QA.
133 9 - Plan Approval 1. Wrong patient 3.78 8.78 5.33 163.9 110
treatment 1.3 Procedure not followed Very wrong volume
2.5 Equipment used incorrectly
6.1 Inadequate training/orientation
2.5. Manual entry error Wrong ID is entered for a patient with the same name.
Very wrong dose Very
Mistaken info Plan is eventually exported to wrong patient and treated.
1. Entry of wrong dose distribution
10 - Plan 1.1, 1.3. Wrong identification
144 demographic 1. Bad info entered Very wrong volume 3.67 8.56 5.22 163.4 111
Preparation 1.2 Inadequate procedure
information Very wrong location
6.1 Inadequate training/orientation
Pull up wrong patient’s record Patient treated on the basis of another patient’s volumes
defined on secondary data sets
{usually carelessness ! Not covered Very wrong dose
5 - Transfer images Transfer secondary
in Canadian table} distribution
35 and other DICOM (MRI, PET) data 1. Incorrect patient 4.22 8.00 4.89 157.9 118
Closest categories are 2.5 Very wrong volume
Data sets
(Materials/tools/equipment used
incorrectly”), 6.1 Inadequate training
and 8.4 fatigue.
Very wrong dose Very
2.5. Wrong patient chosen;
1. Entry of wrong dose distribution
10 - Plan 2. Failure to link to 1.2 Inadequate procedure
146 demographic Very wrong volume 3.89 8.67 4.88 157.9 119
Preparation external database 6.1 Inadequate training/orientation
information Very wrong location
Two courses:
1. The fluence file is incomplete (not corrupt as in item
49). This is then input into the sequencer. An incomplete
10. Run leaf 3. Sequence file or fluence file input into sequencer will result into an
8 - Treatment
119 sequencing to create output incomplete or 2.2. Sequencer software error. Wrong dose distribution 3.67 6.67 5.78 148.7 133 incomplete output of the sequencer.
Planning
deliverable plan. incorrect
2. Fluence is correct but the software error may cause
wrong leaf sequence output.
5.2 Poor communication
Misinterpreted treatment plan
11 - Day 1 Position patient for 3. Incorrect patient Wrong location; Wrong
185 2.2 Incompatibility between 4.00 7.78 4.33 148.3 134
Treatment treatment orientation volume
treatment machine and R&V system
2, Patient medical Inappropriate dose for new condition. There may be some
condition changes events (e.g. drug interactions ) that would have more
11 - Day 1 Take patient into the between prescription serious consequences
175 5.2 Poor communication Wrong dose 4.33 6.44 5.56 147.3 135
Treatment treatment room and treatment (dental
situation, wt
changes)Com
Min: Confusion and
2.5 Human error inconvenience
Create case (define 1. Misidentification 6.1 Inadequate training Max: Very wrong dose Mistype patient name or ID. Causes wrong anatomical
49 7 - RTP Anatomy 4.75 7.63 4.25 147.0 136
patient in RTP db) (Wrong name) 8.4 fatigue. (distribution, absolute and model to be used for planning.
volume)
Lack of standardized procedures. Patient immobilized with dentures in and treated with
QA checks dentures removed.
Personnel inadequately trained
Suboptimal Poor communication as to goals of
immobilization method. treatment
Incorrect position.
Patient Suboptimal treatment
Wrong choice of 1 Standards/procedures/practices
2- Immobilization positioning/immobiliz Inconvenience (patient
5 materials or (all aspects listed 1.1-1.6) 5.22 4.78 5.00 140.4 139
and Positioning ation appropriate to and staff)
accessories. Poorly 2. Materials/Tools/Equipment (all
treatment
fitting mask, aspects 2.1-2.6)
Best/Typical Case 3.2 Inadequate design specification,
3.3 Design process not followed
5: Communications (5.1-5.4)
6: Inadequate training/orientation
1. Measurements not
made
2. Measurements are
Other setup data incorrect.
1.1, 1.2, 1.4 (Lack of regular
acquired and 3. Measurements are
procedures)
documented in chart not clearly Wrong location
6.1 (Inadequate staff training.)
22 3 - CT/simulation (e.g. caliper or ruler documented., e.g., Wrong dose 5.00 6.11 4.44 140.4 140
8. (Inattention to detail.)
measurements of (photos unclear or don’t
4.6, 4.1 (Rushed process)
“AP setup depth”, show key features of
setup photos) setup such as arm
position or setup mark
locations)
Human error Human error leads to initial fields being treated the
1.2 Inadequate procedure number of fractions prescribed for the boost.
10 - Plan 2. Specification of 4. Fields not ordered
154 1.3 Procedure not followed Wrong dose fractionation 3.89 6.89 4.78 135.0 150
Preparation treatment course correctly (initial/boost)
6.1 Inadequate training/orientation
Inadequate
quality/Incomplete
Volume needed for Wrong volume
Images 1 1.4 (General procedures not
14 3 - CT/simulation treatment planning Wrong dose 5.00 5.56 3.89 109.8 166
1. Wrong scan protocol documented)
scanned
used (e.g. wrong slice
thickness/separation)
3.5. Failure to recognize the
1. Treatment volume
8 - Treatment 15. Evaluate delivery limitations of the delivery system, Suboptimal plan
130 inappropriate for the 3.71 5.14 4.71 108.4 167
Planning system limitations 6.1. Inadequate training Poor treatment used
delivery system
Support of multiple 2.5. User error Wrong dose distribution
Scans for different Original CT scan set could be mixed up or combined with
anatomical models 2.1-2.3 Software errors Wrong volumes
patient representations the scan set from a new anatomical site – or the two sets
77 7 - RTP Anatomy (ie, different body 6.1. Inadequate training or 3.67 5.50 5.00 108.0 168
combined or mixed up can be incorrectly registered with each other, leading to
parts) for same understanding of algorithms
wrong anatomical model (and thus dose).
patient
Patient name is typed into the database incorrectly
Errors in manual entry, most likely (misspelled; e.g. Yorke is spelled without the ’e’)
Entry of patient data
1- Patient causes: Information is requested from another department for
in electronic 1. Incorrect Patient ID
1 Database 8. (Human transcription error; Very wrong dose 3.78 7.89 3.89 106.8 169 Ellen York, a different person who actually exists.
database or written data
Information 5,3 (wrong data communicated to Information regarding Ellen York is sent back to the
chart
RO data entry) requesting dept..Incorrect staging/labs transcribed to pt
chart from outside. Suboptimal dose prescribed
Human error Bolus mislabeled or paired with wrong patient’s
immobilization devices and used on wrong patient
Patient-specific
2- Immobilization Hardware used on 1 Standards/procedures/practices Wrong dose
9 hardware labeling 3.78 6.89 3.89 106.6 170
and Positioning wrong patient (1.1, 1.2, 1.3, 1.4, 1.6) and 5.2 Lack distribution/wrong volume
(name)
of communications
The optimizer is not a good quality optimizer because the
Poor optimizer/optimization. optimization algorithm is inferior in quality.
1. Treatment plan has
Conflicting planning objectives and The optimizer could not return desired dose distribution
8 - Treatment significant failure
111 8. Run optimization dose constraints. Inconvenience (staff) 6.11 4.89 4.11 106.4 171 because of the conflicting choice of dose constraints and
Planning meeting planning goals
Inadequate evaluation. planning objectives. The objectives and constraints were
and objectives
not evaluated correctly when they were entered into the
optimizer.
Incorrect delivery system definition in Explanation of 1 and 2. Certain gantry and table angles
the TPS. will result in collision but these angles are not defined in
Lack of machine limitations in TPS. the TPS. So, TPS will allow this but in reality this will
4.4. Failure to recognize the impact cause collisions.
8 - Treatment 15. Evaluate delivery 3. Patient and delivery of patient immobilization and/or Non-radiation related
132 3.50 9.00 3.36 106.3 172
Planning system limitations system collision positioning aids on collision free injury 3. For large patient or a patient in an immobilization
zone. system there will be collisions at certain gantry angles.
Poor knowledge of patient location + This is due to failure to recognize the limitations of large
positioning on table. patients or immobilization devices.
10 - Plan 2. Specification of 4. Fields not ordered 2.2. software error Software error leads to initial fields being treated the
155 Wrong dose fractionation 2.57 5.71 6.14 86.0 188
Preparation treatment course correctly (initial/boost) 3.6 Inadequate programming number of fractions prescribed for the boost
Delivery recorded in
12 - Day N Partial delivery
216 paper chart (if Human error/carelessness Wrong dose 4.89 4.44 4.11 84.9 189
Treatment incorrectly recorded.
applicable)
Bad parameters were chosen. Optimization fails because of all the causes listed in the
3.2. Software error (Wrong cost cause column.
function)
3. Optimization fails Impossible plan constraints (no
8 - Treatment (does not converge, or feasible solutions).
112 8. Run optimization Inconvenience (staff) 5.78 3.89 3.33 83.3 190
Planning gives clearly non- Inadequate search (not enough
optimal answer) iterations).
Inadequate beam technique
parameters
Sequencer may ask the user to enter the MLC limit at the
user level. At this level, suppose the user enters a limit of
10. Run leaf 6. Interdigitization limits
8 - Treatment Wrong machine configuration; 16 cm. The sequencer will generate leaf profiles that will
122 sequencing to create incorrect (Elekta, Inconvenience 4.33 3.67 3.22 65.7 197
Planning 8. wrong choice allow 16 cm of leaf motion. But hardware requirements
deliverable plan. Siemens)
may limit leaf motion to only 14 cm. Therefore, sequencer
will generate a leaf profile that results in an incorrect dose
distribution.
10 - Plan 7. Decide delivery 2. Unexpected changes 2.2. software error Potential table crash
165 2.00 4.25 3.25 64.5 198
Preparation ordering of fields in sequence of fields 3.6 Inadequate programming scenarios.
Wrong user input. 1. Delay or inefficient
5. Define imaging
1.2 Inadequate procedure setup. 2.
10 - Plan sequences to be
161 Wrong imaging planned 1.3 Procedure not followed Set patient at incorrect 3.80 3.80 2.40 64.2 199
Preparation used for localization
6.1 Inadequate training/orientation position. Geometrical
process.
miss. Failure.
Problem with file transfer Corrupt CT file on urgent patient requires beam definition
on the basis of orthogonal plain x-rays
Very wrong dose
5 - Transfer images 2.2 Defective materials (software or
Transfer primary File corrupted distribution
33 and other DICOM hardware) 3.89 5.78 2.33 62.7 200
(CT) data set Very wrong volume
Data 3.2 Inadequate design specifications
3.6: Inadequate programming
4. Patient “exceeds
scan diameter” and 4 2.6, 6.4, 5.2 (Equipment limitation
Volume needed for Wrong dose
volume of interest or Poor patient positioning. Failure to
16 3 - CT/simulation treatment planning Wrong location 4.33 5.44 2.67 61.9 201
volume traversed by communicate with planner when
scanned
one or more beams is problem is recognized.)
truncated.
1.2. Poor dept. policy This shouldn’t be possible with an R&V system but I’ve
Some beam(s) 2.1. No R&V seen instances of this on machines without an R&V. It’s
12 - Day N
212 Treatment delivered delivered twice in one 3.2. Poor R&V design Very wrong dose 2.56 5.67 3.44 59.7 202 most likely if therapists can “tag team” each other during a
Treatment
session Carelessness patient’s treatment session instead of waiting to trade off
jobs till that session is over.
1. Delay or inefficient
5. Define imaging
setup.
10 - Plan sequences to be 2.2. software error
160 Wrong imaging planned 2. Set patient at incorrect 2.80 3.80 2.60 59.4 203
Preparation used for localization 3.6 Inadequate programming
position. Geometrical
process.
miss. Failure.
8. Poor user choicet. Poor choice of beam sequencing requires the gantry or
1.2 Inadequate procedure collimator to rotate back and forth on itself
10 - Plan 7. Decide delivery 1. Poor choice of No effect
164 1.3 Procedure not followed 4.67 2.89 4.00 55.2 204
Preparation ordering of fields sequence Inefficient delivery
6.1 Inadequate training/orientation
5.4. Miscommunication; Physician does not sing the hard copy or electronic chart
inattention, indicating that the plan was approved for treatment.
1.1. lack of procedure
2. Completion of
6. Not signed when 1.2 Inadequateprocedure
143 9 - Plan Approval formal prescription No effect on patient 6.78 2.00 3.00 53.3 205
appropriate 1.3 Procedure not followed
after planning
2.5 Used incorrectly
6.1 Inadequate training/orientation
Problem with file transfer Corrupt secondary file requires plan based on primary
data set only.
Very wrong dose
5 - Transfer images 2.2 Defective materials (software or
Transfer secondary 2. File corrupted distribution
36 and other DICOM hardware) 4.00 5.67 1.89 43.7 211
(MRI, PET) data sets Very wrong volume
Data 3.2 Inadequate design specifications
3.6: Inadequate programming
8. (Inattention to detail)
Image set saved or
Simulation image set 4.6, 4.1 (rushed process) Inconvenience (staff and
23 3 - CT/simulation sent to treatment 3.56 2.89 1.44 18.9 216
accidentally deleted 1.2 (Inadequate backup procedures) patient)
planning