Professional Documents
Culture Documents
Manual - 20kHz - SVC - Before P47 2001 (L261d1)
Manual - 20kHz - SVC - Before P47 2001 (L261d1)
HIGH FREQUENCY
X-RAY GENERATOR
Rev D
WARNING
Adequate precautions must be taken to prevent unauthorized or unqualified persons from operating
this equipment or exposing themselves or others to its radiation. Only qualified personnel should
install, set up, maintain, and operate this equipment. Only qualified service personnel should
remove electrical covers.
Before operation, persons qualified and authorized to operate this equipment should be familiar with
the Recommendations of the International Commission on Radiological Protection, contained in
Annals Number 26 of the ICRP, and with the applicable national and local standards.
The equipment described in this manual will perform reliably when installed, maintained, and
operated, in accordance with the instructions of this manual by qualified personnel. This equipment
is sold with the understanding that the user assumes sole responsibility for radiation safety and that
the manufacturer does not accept any responsibility for the following:
We are proud of our products and are confident they will provide many years of useful and
enjoyable service.
Summit Industries
-i-
SynergyTM Service Manual
- ii -
SynergyTM Service Manual
TABLE OF CONTENTS
1.0 PRE-INSTALLATION...................................................................................... 1 - 1
1.1 General ................................................................................................. 1 - 1
1.2 Use of This Manual ............................................................................... 1 - 1
1.3 Equipment Description.......................................................................... 1 - 2
1.3.1 General ...................................................................................... 1 - 2
1.3.2 Optional Automatic Exposure Control (AEC).............................. 1 - 2
1.3.3 Safety......................................................................................... 1 - 3
1.3.4 Manufacturer’s Responsibility .................................................... 1 - 4
1.3.5 Radiation Protection................................................................... 1 - 4
1.3.6 Monitoring of Personnel ............................................................. 1 - 5
1.4 Compatibility Statement ........................................................................ 1 - 5
1.5 Accuracy Statement.............................................................................. 1 - 6
1.5.1 Exposure kVp............................................................................. 1 - 6
1.5.2 Exposure mAs............................................................................ 1 - 6
1.5.3 Duty Cycle.................................................................................. 1 - 6
1.6 Maximum Outputs................................................................................. 1 - 7
1.7 Maximum Momentary Line Current....................................................... 1 - 7
1.8 Technique Factors That Constitute Maximum Line Current.................. 1 - 8
1.9 Power Line Requirements..................................................................... 1 - 8
1.10 Conduit for Communication Cable ........................................................ 1 - 8
1.11 Space Requirements ............................................................................ 1 - 8
- iii -
SynergyTM Service Manual
TABLE OF CONTENTS
- iv -
SynergyTM Service Manual
9.0 SCHEMATICS................................................................................................. 9 - 1
v
SynergyTM Service Manual
TABLES:
1-1 Maximum Power Rating........................................................................ 1 - 7
1-2 Maximum Line Current.......................................................................... 1 - 7
2-1 Line Matching Transformer Connections ............................................ 2 - 10
2-2 Tube Stator Cable Connections .......................................................... 2 - 11
2-3 Bucky Connections ............................................................................. 2 - 12
2-4 Collimator and DC Lock Supply Connections ..................................... 2 - 13
3-1 DIP1 Functions (System Controller PCB) ............................................. 3 - 2
3-2 Fusing for DC Voltages......................................................................... 3 - 4
3-3 Fusing for AC Voltages ......................................................................... 3 - 4
4-1 Button Functions in Programming Area ................................................ 4 - 3
4-2 Programming Stations, Functions and Values ...................................... 4 - 4
4-3 Typical Filament Values for a
Toshiba 1 x 2 mm x-ray Tube ............................................................. 4 - 14
4-4 Filament Value Mode Button Functions .............................................. 4 - 15
4-5 For Recording Actual Filament Current Settings
at this Installation ................................................................................ 4 - 21
5-1 Table of Error Codes............................................................................. 5 - 1
6-1 Sample of Reproducibility Calculation................................................... 6 - 4
6-2 Table for Recording Reproducibility Values .......................................... 6 - 4
6-3 Sample of Linearity Calculation ............................................................ 6 - 7
6-4 Table for Recording Linearity Values .................................................... 6 - 8
7-1 Required Maintenance Log................................................................... 7 - 2
FIGURES:
1-1 Component Physical Dimensions ......................................................... 1 - 9
2-1 Power Module Layout: Hinged Panel Closed........................................ 2 - 5
2-2 Power Module Layout: Hinged Panel Open ......................................... 2 - 6
2-3 Midwest 14 x 36 Bucky Interface ........................................................ 2 - 10
vi
SynergyTM Service Manual
1.0 PRE-INSTALLATION
1.1 General
This manual contains installation, calibration, acceptance and service information for the
SYNERGY™ Generator. The generator is part of a system which contains other equipment
not described in this manual. To gain an understanding of the complete system, study the
operator’s and service manuals for all associated equipment.
Throughout the manual there are indented statements which are preceded by alert words.
These alert words and the following statements may direct the user in the performance of a
task, or inform the user of a potentially hazardous situation. Each alert word is always used
with only one specific type of information. The alert words and the information they
reference are as follows:
NOTE: Information that assists the user of the manual in the performance of a task.
It may provide the user with better methods of conducting the task, or it may point
out conditions that could cause the system to fail to operate properly.
! CAUTION: Information that alerts the user of the manual to situations that could
result in damage to the equipment.
WARNING: Information that alerts the user of the manual to situations that could be
hazardous to the health of the patient, the operator, the maintenance person, or
others.
1-1
SynergyTM Service Manual
1.3.1 General
The SYNERGY™ AEC system accommodates AID ion chambers. A special Chiropractic
AEC system, CAEC, is also available with the SYNERGY™ generator. Refer to the
AEC/CAEC Manual L323 for Instructions on installation and calibration of this feature.
1-2
SynergyTM Service Manual
1.3.3 Safety
It is important that everyone associated with x-ray work is familiar with the recommen-
dations of the Department of Health, National Bureau of Standards, and the National
Council on Radiation Protection. The control of diagnostic medical x-ray equipment varies
in detail from state to state. However, in general, all of the states adhere strictly to the
established recommendations of the NCRP. Prior to operation, be sure that all personnel
who are authorized to operate the x-ray system are familiar with the established regulations
of the authorities named above. Also, they should be monitored to assure that they
conform to the recommendations.
Although x-ray radiation is hazardous, x-ray equipment does not pose any danger when it is
properly used. It is the responsibility of all service and operating personnel to be properly
informed on the hazards of radiation. Also, those responsible for the system must
understand the safety requirements for x-ray operation. Study this manual and the
manuals for each component in the system to become aware of all the safety and
operational requirements.
1-3
SynergyTM Service Manual
Although this equipment incorporates protection against x-ray radiation other than the
useful beam, practical design cannot provide complete protection. Equipment design does
not compel the operator or his assistants to take adequate precautions. Nor does it prevent
the possibility of improper use which results in authorized or unauthorized persons from
carelessly, unwisely, or unknowingly exposing themselves or others to direct or secondary
radiation. Allow only authorized, properly trained personnel to operate this equipment.
Be certain that all personnel authorized to use the equipment are aware of the danger of
excessive exposure to x-ray radiation.
This equipment is sold with the understanding that the manufacturer, its agents, and
representatives do not accept any responsibility for over-exposure of patients or personnel
to x-ray radiation. Furthermore, the manufacturer does not accept any responsibility for
over-exposure of patients or personnel to x-ray radiation generated by this equipment
which is a result of poor operating techniques or procedures.
Also, no responsibility will be assumed for any machine that has not been serviced and
maintained in accordance with the system technical manual, or which has been modified or
tampered with in any way.
Because exposure to x-ray radiation may be damaging to health, use great care to provide
protection against exposure to the primary beam. Some of the effects of x-ray radiation are
cumulative and may extend over a period of months or years.
1-4
SynergyTM Service Manual
Monitoring of personnel to determine the amount of radiation to which they have been
exposed provides a valuable cross-check to determine whether or not safety measures are
adequate. The most effective method of determining whether or not the existing protective
measures are adequate is the use of instruments to measure the exposure in rads. This
measurement should be taken at all locations where the operator or any portion of his body
may be during exposure.
X-RAY TUBES:
Toshiba E7239 (1.0/2.0 mm focal spots), Toshiba E7242 (0.6/1.5 mm focal spots), Varian
RAD 13 (1.0/2.0 mm focal spots), Varian RAD 14 (0.6/1.2 mm focal spots).
Expansion of the compatible tube list is possible as need dictates.
AEC DEVICES:
An integrated AEC option is available to allow interfacing with ion chambers manufactured
by Advanced Instrument Development of Melrose Park, Illinois.
BUCKY:
Summit J500, Midwest 7440 and 1436, Liebel 8000 and 9000.
COLLIMATOR:
Summit G800 and D800, Progeny MC150 and Linear II or Linear IV.
1-5
SynergyTM Service Manual
All measurements are made in compliance with the directions of the SYNERGY™ Service
Manual in the Installation, Acceptance, and Calibration procedures sections.
NOTE: Specified accuracy for kVp and mA does not include test equipment
accuracy. Acceptance testing must take test equipment error into account. For
additional information on the basis for the accuracy statements, refer to
Acceptance Procedure (Section 5.0).
The duty cycle is continuous for radiographic use. However, operation should be limited to
the capacity of the x-ray tube.
1-6
SynergyTM Service Manual
The maximum rating for the 325 is listed in Table 1-1. The maximum line current is listed in
Table 1-2.
FACTOR MAX VALUE
kVp 125 kVp
mA 300 mA
Power 30 kW
Maximumg Line current will occur during exposures of 100 kVp, 300 mA.
1-7
SynergyTM Service Manual
The line currents given in Table 1-2 (for the 208 to 240 volt line) are the momentary line
currents during an exposure. This is not the recommended circuit breaker rating. The
recommended supply power requirements are:
These wire sizes are for copper conductors between the distribution transformer and the
100 Amp service disconnect. The 100 Amp breaker listed for 240 VAC operation must be
upgraded to 125 Amp for operation below 240 VAC as indicated by the line currents listed
in Table 1-2 above.
If the two communication cables between the Power Module and Operator Console are to
be run through conduit, the minimum conduit size is 1-1/2” internal diameter due to the non-
removable DB9 connectors on the ends of the cables.
Figure 1-1 shows the physical dimension and cable access points of the Power Module, the
High Voltage Transformer, and the Operator’s Console.
1-8
SynergyTM Service Manual
⌧ ⌧
2-1
SynergyTM Service Manual
2.0 INSTALLATION
2.1 Introduction
This section describes procedures necessary for the installation of the following units:
The following miscellaneous items are required or may prove useful for the installation:
• Silicone Insulating Grease - DC4 vapor-proofing compound, or equivalent.
• Alcohol
• Standard Hand Tools
• Victorian mAs meter or equivalent
• Kiethley Non-invasive kVp meter or equivalent
2-2
SynergyTM Service Manual
Prior to beginning installation, inspect the x-ray room to verify compliance with
specifications for the following:
2.4 Unpacking
Upon receipt of the x-ray generator and associated equipment, inspect all shipping
containers for signs of damage. If damage is found, notify the carrier or his agent imme-
diately. When the equipment is unpacked, inspect all pieces for visible damage. If any
damaged parts are detected, repair or order replacements to prevent unnecessary delay in
installation.
2-2
SynergyTM Service Manual
Access to the Operator Console pcb is achieved by removing the five screws from the
bottom of the Console Enclosure. The electronics are contained in the top section of the
console.
Access to the Power Module is achieved by removing the two screws from each side of the
cover, lifting the cover up slightly and forward from the Power Module frame.
Terminal connections P1, P2, and GND can be accessed by removing the two screws at
the terminal cover on top of the unit.
2-3
SynergyTM Service Manual
<O> 03035 : KIT, FILAMENT DRIVER BRD REPLACEMENT (compatible replacement For ALL)
<O> 01985-000 : KIT, RELAY BOARD K660 REPLACEMENT ( includes required wiring adjustment ).
<O>
<O>
<O>
2-4
SynergyTM Service Manual
2-5
SynergyTM Service Manual
2-6
SynergyTM Service Manual
Measure and note the incoming line voltage. 240 VAC is required for full 30 kW output. If
30 kW output is desired and the line voltage is below 240 VAC, a line matching transformer
kit (part number L246) will be required.
NOTE: If the line voltage is below 240 VAC (i.e., 208 VAC), it will not be possible
to produce the full output capability of the generator without a line matching
transformer.
At this time, place the Power Module close to the final location, leaving access on all sides.
This allows the installer to establish the proper wiring, routing and lengths for all cables.
Place the operator console in the operator booth and run the two communication cables
from the console to the Power Module. The cables are keyed so that they will only plug into
the proper receptacles. At the power module the cables will plug into H13 and H14 on the
System Controller pcb.
Place the HV Transformer in front of the bottom shelf of the Power Module. The high
voltage cables cannot be inserted or removed when the transformer is inside the cabinet.
NOTE: Release the 8-32 screw in the oil fill plug 1 full turn counter-clockwise to
allow the oil level to expand with temperature and barometric changes.
2-7
SynergyTM Service Manual
Inspect all wiring in the Power Module. At this time, the Power Module can be connected to
the Operator’s Console and the HV Transformer.
The Communication Cable for the Operator’s Console connects to the Power Module as
described in Section 2.7.3 above. This shielded cable routes from the control room to the
Power Module.
Route the Filament/Feedback cable to the HV Transformer and plug it into the mate-n-lock
receptacle at the top of the transformer.
Three leads marked “P1, P2 and GND” should be routed from the bottom of the Power
Module to the HV Transformer and terminated at the connections labeled “P1, P2 and
GND”.
208 or 220 VAC Input with Line Matching Transformer (Field Installation Only)
Single phase power below 230 VAC will require a line matching transformer. See the
transformer kit, part # L246 for further details or section 2.7.7 below. Line voltages below
230 VAC will only support very low power exposures for light filed to x-ray field congruence
and system alignmments. This is only a stop-gap measure. Obtain a line match
transformer kit immediately!
2-8
SynergyTM Service Manual
Warning: Make absolutely certain that the AC mains supply is off and that
the DC capacitor bank voltage within the generator is fully discharged before
beginning this procedure. The capacitor bank voltage requires a minimum of 10
minutes to discharge. Verify that no hazzardous voltages are present.
1. At the bottom right rear of the Power Module is a saddle bracket with a 3/8-16 x 4.5
inch hex bolt and two nuts with lockwashers. Note that the hex bolt is isolated from the
saddle bracket with two plastic bushings.
3. Carefully place the toroid line matching transformer within the saddle bracket so as not
to scratch the transformer insulation on the bracket. Orient the transformer so that the
leads exit the saddle bracket to the top and to the right.
4. Secure the transformer to the saddle bracket using the long hex bolt. Be certain to use
the plastic bushings between the bolt and the brackets so that the bolt is electrically
isolated from the bracket. The bolt is to be snug, but not tightened so excessively that
it crushes the epoxy center of the toroid style transformer. Use the second hex nut as
a “jam nut” to prevent the first nut from loosening.
5. Connect the transformer leads to the right side of TB1 in the Power Module as shown
in Table 2 -1 below.
2-9
SynergyTM Service Manual
6. Also on TB1 there are three wires from the Power Module which should be in the
following locations. Some of these locations are different when using a line matching
transformer than they are when the unit is wired for 240 VAC operation at the factory.
Verify that the following is true:
The Power Module wire labeled “1B” should be on TB1-1 (280 V), along with the
blue transformer wire.
The Power Module wire labeled “2A” should be on TB1-2 (240 V), along with the
yellow transformer wire.
The Power Module wire labeled “1C” should be on TB1-6 (0 V), along with the
black transformer wire.
On TB1, locate the wire labeled “1”. This is the line voltage adjust tap. Move wire “1”
to the TB1 tap which most closely matches the incoming line voltage (208 VAC
or 220 VAC).
7. On TB2, located just in front and to the right of TB1, verify the following connection.
The wire labeled “2J” should be on TB2-1, the terminal labeled “240”.
8. When all connections have been verified and the mounting bolt is known to be isolated
from chassis ground the installation of the transformer is complete.
2 - 10
SynergyTM Service Manual
Connect the high voltage cables from the HV Transformer to the x-ray tube. With the vent
to the front, the cathode receptacle is on the left, by the red and black mAs meter jumpers.
The anode receptacle is on the right. Verify polarity and that all connections are made
correctly.
! CAUTION: The high voltage cable terminal pins are delicate. Take particular
care to handle them carefully.
NOTE: The HV Cable rings/nuts are delicate and can be damaged by dropping
or banging on the floor. This can deform the threads, making it difficult to
install them fully into the HV receptacles
Use a clean, dry, lint-free cloth and alcohol to clean the insulating surfaces of the terminals
and receptacles. Use particular care in cleaning the flat insulating surface. Do not touch
the insulating surface after cleaning. Coat with vapor proofing compound.
2 - 11
SynergyTM Service Manual
For the Summit J500 or equivalent Bucky, connect as shown in the table below.
Origin B1 B2 B3 B4 B6 B8
Function bucky bucky switched bucky power bucky
motion motion 110 V 110 V ground 110 V
return return hot return
Power Module TB4-1 TB4-2 TB4-3 TB4-9 TB3-15 TB4-8
Destination:
(Table bucky)
Power Module TB4-4 TB4-5 TB4-6 TB4-9 TB3-15 TB4-8
Destination:
(Wall bucky)
CAUTION: If 115 VAC is used as the Bucky Motion Signal (B1/ B2 signal to
2 - 12
SynergyTM Service Manual
The SYNERGY™ generator comes with power supply connections to run most commonly
available collimators and electric locks. The available supplies are a switched 24 VAC/7
Amps for collimation and a continuous 24 VDC/4 Amps (or 24 VDC/5 Amps at 65% duty
cycle) for locks.
NOTE: There is NO other power available for other accessory items requiring 110
VAC, 5 or 12 VDC. Connections of ancillary devices requiring other than the 24
Volt supplies listed above, or requiring more current than that listed above, will
cause component damage not covered under warranty.
2.7.12 Interlocks
There are two possible interlocks besides the thermal interlock from the x-ray tube.
Terminals TB3-6 and TB3-8 are provided for connection of an isolated switch contact for
collimator interlock. If the collimation is manual, jumper these two points.
Terminals TB3-5 and TB3-8 are provided for connection of an isolated switch contact for
door interlock. If no door interlock is being used, jumper these two points.
TB3-8 and TB3-9 are electrically the same point and can be used interchangeably.
2 - 13
SynergyTM Service Manual
3.1 General
The SYNERGY™ system should come in a standard configuration with typical default
calibration values stored in memory. After initial inspection and interconnections have been
made, the system will be ready for power up, voltage checks and calibration.
WARNING: The main storage capacitors retain their charge for some time after
the unit is turned off. Although the areas where this voltage may be present are
covered with a protective layer of Lexan, it is imperitive that all service personnel
are cognizant of the significant shock hazard the capacitor bank voltage
represents. Prior to servicing the power components, ensure that the LED on the
Charge Sense PCB is not lit, and verify there is no capacitor bank voltage with a
DC voltmeter.
Observe SW1 on the System Controller pcb in the Power Module and verify that all
switches are set to their default setting . Refer to the chart below for information on the
functions of the switches.
3-1
SynergyTM Service Manual
3-2
SynergyTM Service Manual
3.3 Power-Up
3.3.1 Start-Up
Turn the main disconnect switch ON. Press the Operator Console ON key and observe the
kVp Display and the mAs Display during the start-up phase. The kVp Display should
momentarily display the software version number for the System Controller and the mAs
Display should momentarily display the software version number for the Operator’s
Console.
Within five seconds of power up the kVp and mAs windows should display valid technique
selections.
Upon power up, there are a number of LEDs which will illuminate on boards within the
power module. Verify that the following are true.
L050 System Controller pcb: LED1 lights will strobe back and forth across the bar.
3-3
SynergyTM Service Manual
Verify the following output voltages from the Power Supplies referenced to ground.
3-4
SynergyTM Service Manual
4.1 General
Calibrations will be done completely through software adjustments. The Calibration Mode
has two sections: the Programming Section and the Filament Value Section. The
Programming Section controls system configuration selections such as tube type and
image receptors, as well as kVp and automatic exposure control adjustments. The
Filament Value Section allows precise control of the filament current and thus the leading
edge for each mA station across the kVp range.
NOTE: It is best to view both the kVp and the mA waveforms simultaneously
when calibrating the system. This way initial adjustments can be made to either
kVp or mA if one is dramatically in error. While both are on closed loop feedback
systems, they should each be reasonably close to the desired level before fine
adjustments are made.
The installer is responsible for achieving waveforms which are both of the proper amplitude
and of the proper “square” waveshape, where the desired amplitude is maintained
throughout the entire exposure. A Machlett Dynalyzer (or equivalent high voltage
divider) and a storage oscilloscope are required for calibration. The Dynalyzer will
indicate total kVp and mA, while the scope allows a view of the waveform shape. It is
preferable for the scope to be dual trace, 100 MHz response as some hand-held
scope/meters have a frequency response and pixel resolution that do not give a reliable
and discernable view of the output waveforms.
Insert the Dynalyzer between the HV transformer and x-ray tube to measure output levels.
Connect the scope to the kVp and mA outputs of the divider tank of the Dynalyzer to view
the leading edge and stabilized levels of kVp and mA.
4-1
SynergyTM Service Manual
To enter Calibration mode, turn the control OFF. When the control is turned back
ON, the kVp and mAs displays will momentarily indicate the software version (for
example, “1.0” and “1.1”). When this is displayed, press and hold the mA button
(located to the right of the mAs display) until the kVp display changes to “CAL”, then
release the button.
The control will automatically boot into the Programming Section of the
Calibration Mode. When in the Programming Section, the kVp display will show
the location (P1 through P46), and the mAs display will indicate the value stored in
this location. A typical display upon entering the Programming Section will be “P1”
and “1”, indicating a value of “1” at programming point #1.
To move from the Programming Section to the Filament Value Section, press
and hold the mA button, momentarily press the “LG” focal spot button, then release
both. When in the Filament Value Section, the kVp display will show a kVp value,
and the mAs display will show a number which represents actual filament amps.
To move back to the Programming Section, press and hold the mA button, press
the SM focal spot button, then release both buttons. The normal “P#” and “value #”
displays of the Programming Section will again be displayed.
See Section 4.4 for a more complete explanation of button functions when in the
Calibration Mode of operation.
4-2
SynergyTM Service Manual
When in this mode, the kVp window displays the P#, which is the
programming station. The mAs window displays the # value at this location.
Pressing this button . . . will perform this function . . .
kVp up increment to next P#
kVp down decrement to next P#
mAs up increment the value at this location
mAs down decrement the value at this location
SM focal spot save a change to the value at this P#
LG focal spot displays exposure count
prep or exposure are not allowed
mA and LG focal spot together go to Filament Value area (exposures)
mA, LG and SM focal spot together exit Calibration Mode
4-3
SynergyTM Service Manual
4-4
SynergyTM Service Manual
Programming
Function Range of Values within this station
Station (P#)
P19 Enables table AEC OFF=no AEC, 1=Standard AEC,
2=CAEC (chiropractic AEC)
(If OFF, no access to P20 - 24)
P20 Table AEC config Sets field selection. Typically 2 if AEC,
inactive if CAEC
P21 Table 50 kVp AEC Set for optical density per AEC manual
comp L323
P22 Table 70 kVp AEC Set for optical density per AEC manual
comp L323
P23 Table 100 kVp AEC Set to 1.5 per AEC manual
comp
P24 Table 125 kVp AEC Set for optical density
comp
OFF=no AEC, 1=AEC, 2=CAEC
P25 Wall AEC enables
P26 Sets field selection. Typically 2 if AEC,
AEC config inactive if CAEC
P27 Wall 50 kVp comp Set for density per AEC manual
P28 Wall 70 kVp comp Set density per AEC manual
P29 Wall 100 kVp comp Set to 1.5 per AEC manual
Set for optical density per AEC manual
P30 Wall 125 kVp comp
The following programming stations are the kVp offset and slope adjustments
for each mA station.
4-5
SynergyTM Service Manual
Programming
Function Range of Values within this station
Station (P#)
P35 40 kVp at 75 mA -40 to +40, 1.0 kVp steps
4-6
SynergyTM Service Manual
NOTE: If the x-ray tube you are using is not listed in Table 3-3, you may select a
tube with similar filament characteristics. Compare the tube rating charts of the
tube you have with one of the tubes listed and select the tube that most closely
matches your tube.
P02 Max kW
With P2 displayed in the kVp window, enter the number for kW output of the generator.
This will be preset to 30 kW. If the main power supply cannot maintain a load voltage
within 5% of the no-load voltage, this value can be reduced without affecting the maximum
available mAs. Reducing this number will change maximum mA.
4-7
SynergyTM Service Manual
4-8
SynergyTM Service Manual
4-9
SynergyTM Service Manual
4 - 10
SynergyTM Service Manual
4 - 11
SynergyTM Service Manual
The following Programming Stations control kVp. They are the slope
and offset kVp adjustments for each individual mA station. They control
the stabilized level of kVp.
4 - 12
SynergyTM Service Manual
With 40 kVp selected, this station will control accuracy at the lower end of kVp. The value
range is from –40 to +40, with steps of 1 kVp.
4 - 13
SynergyTM Service Manual
The chart below is a matrix of kVp and mA, showing typical filament current values which
result in a square mA waveform. Your values may be different, but the effect of reduced
filament current at higher kVp and increased filament current at lower kVp will still hold true.
* Exposure is not possible at these techniques. Enter values which follow the
pattern established at lower kVp for the mA station.
4 - 14
SynergyTM Service Manual
NOTE: Before the Filament Value Section is used to calibrate the mA, the
Program Section must be used to set up basic system configurations. See the
Programming Sections 4-4 and 4-5 above to configure your system. Once this
has been achieved, refer to section 4-8 for a complete explanation of mA
calibration using the Filament Value Section of Calibration Mode. At the end of
this section is a chart to record the actual filament current required to achieve the
desired outputs at this installation.
In this area, the kVp window displays kVp in steps of 40, 50, 70, 90, 110, and
125 kVp. The mAs window displays the filament current value for this
combination of kVp and mA.
Pressing this button . . . will perform this function . . .
kVp up increment kVp to next step
kVp down decrement kVp to next step
mAs up increment filament current value
mAs down decrease filament current value
SM focal spot save change in filament current value
LG focal spot performs no function
prep and expose are allowed
mA and SM focal spot go to Programming area
mA, SM and LG focal spot exit Calibration Mode
4 - 15
SynergyTM Service Manual
If the mA button is pressed and held, some button functions and displays are
different than as shown above. If the mA button is held down . . .
The kVp window shows the time setting in milliseconds, and the mAs window
shows the mA station selected. Pressing kVp up/down increases/decreases
time. Pressing mAs up/down increases/decreases the mA station. This allows
the installer to adjust the mA and time of exposure to obtain a good picture of
the mA waveform on their test equipment.
Using a dual trace storage oscilloscope, connect to the Dynalyzer kV and mA output. If no
Dynalyzer is available, place one probe on the kVp feedback test point TP8, and the other
on the mA test point TP9. TP2 is ground reference. The test points are located on the
System Controller pcb in the Power Module, upper left hand corner. At the kVp test point,
one volt is equal to 33.3 kVp, and at the mA test point one volt is equal to 100 mA.
Make exposures, monitoring and adjusting the leading edge and stabilized levels of mA and
kVp until satisfactory outputs have been achieved across the operating range.
4 - 16
SynergyTM Service Manual
A mAs meter can be put between the red and black terminals behind the cathode
receptacle (remove the jumper!)
Note: If the kVp or mA are dramatically inacurate, one will affect the other.
When making adjustments to mA leading edge or stabalized level, it may be
necessary to adjust kVp as well. This may be a little "out of sequence" but be
aware that an mA may change if kVp is off by a large amount. This is why
monitoring both waveforms at all times is a good idea. Making changes in the
kV controlling numbers (P08, P31 - P46) in the course of mA calibration is
sometimes required to prevent having to go back to readjust filament value or
feedback numbers.
First, the leading edge of 25 mA and 300 mA is set with feedback disabled.
Second, mA feedback is enabled. P10 & P11 are adjusted so that long exposures at 25
mA and 300 mA result in the proper stabilized mA.
Third, at the kVp/mA points for all mA stations, the filament value is adjusted to achieve a
square waveform.
Fourth, the leading edge of kVp is adjusted at a mid range setting of 70 kVp, 200 mA.
Fifth, kV slope and offset are adjusted as needed for each mA station for stabilized kV level
accuracy.
4 - 17
SynergyTM Service Manual
4. Select 50 msec.
Press and hold the mA button, and use the kVp up/kVp down keys to select time.
Using a storage oscilloscope set to 10 msec. sweep, this might be an exposure of
50 milliseconds. Scope both kVp and mA at the same time, triggering from the kVp
waveform. The kVp test point is TP8, and one volt equals 33.3 kVp.
4 - 18
SynergyTM Service Manual
4 - 19
SynergyTM Service Manual
Now that feedback has been set in step 9, the mAs for each other kVp/mA point will
be correct if the leading edge is adjusted to achieve a square waveform. In the
filament value mode, select 40 kVp, 100 mA and 50 msec.
Increase or decrease the factory's filament value number by the "filament value
correction factor" determined in step 7 above. Take exposures, making further
adjustments in the filament value to achieve a square waveform. Repeat this
process for the other kVp/mA points of large focal spot.
! IMPORTANT: Exposures at 125 kVp/250 mA, 110 kVp/300 mA and 125 kVp/300
mA are not possible. Change the filament value numbers at these points to
reflect the pattern of values established at other points. Example: 110
kVp/300 mA will be .02 less than the value at 90 kVp/300.
4 - 20
SynergyTM Service Manual
Table 4-5: For Recording Actual Filament Current Settings At This Installation
50 mA
75 mA
100 mA
(small)
100 mA
(small)
200 mA
250 mA
300 mA
4 - 21
SynergyTM Service Manual
With a storage scope probe on TP8 of the System Controller pcb and a Dynalyzer or non-
invasive kVp meter in the beam, make exposures to determine the shape and amplitude of
the kVp waveform. At TP8, one volt equals 33.3 kVp. Set scope to 10 ms sweep.
NOTE: This test point gives a good look at waveshape, but due to the
voltage/kVp ratio it is difficult to accurately determine the total kVp output. This is
why the Dynalyzer or kVp meter is required.
Set techniques to 200 mA, Large focal spot, and 50 to 100 milliseconds. Take exposures
at 40 kVp, 70 kVp, and 120 kVp. Note the output shape and amplitude of kVp.
Cycle power. Press the mA button when the software version first appears in the kVp/mAs
windows, and hold it until “CAL” appears in the kVp window. Once the displays indicate
“P1” and “1” the unit is in Programming mode. Increment to P08 and change the value to
achieve a good leading edge. P08 will control the leading edge of kVp for all mA and kVp
stations.
The stabilized level of kVp is adjusted at the lower and higher levels of kVp using P31
through P46. Each mA station has an offset and slope adjustment. for example, for 25 mA
P31 is the offset and P32 is the slope. It is not necessary to use 40 kVp, just the lowest
kVp your test equipment can measure. At some mA stations 110 or 125 kVp is not
possible, so the highest selectable kVp is acceptable.
Toggle between the programming mode station to change and save the value, and the
filament value mode to select techniques and make exposures.
4 - 22
SynergyTM Service Manual
Repeat the exposures taken in Step 3 above, and compare the results to the desired
output. If further adjustment is required, repeat the steps above until satisfied with the
results.
Note: It is a good idea to monitor mA during this step to fine tune the leading
edge or stabilized levels as needed.
The option of Automatic Exposure Control has its own complete manual, part
#L323, located in this manual package.
4 - 23
SynergyTM Service Manual
5.0 TROUBLESHOOTING
5.1 Overview
The Synergy™ generator has an extensive internal monitoring system. A number of error
codes are displayed to alert the operator or service engineer to the status of the system.
Some error codes are momentary alerts, such as E07. This error message indicates that
the operator has attempted an invalid selection, such as more than 125 kVp, and will
automatically reset once the offending button has been released. Other error codes
indicate simple or serious problems with the system. The chart below shows all error
codes, what they mean, and how to correct them.
7-2
SynergyTM Service Manual
7-2
SynergyTM Service Manual
7-2
SynergyTM Service Manual
7-2
SynergyTM Service Manual
E13 or E14.
There is a problem with the EEPROM memory. This can be reset to factory defaults
if the chip is still functional. To reset the factory defaults, set SW1-8 on the System
Controller pcb to ON, and boot up into Calibration Mode. This will reburn the factory
defaults into the Programming Stations. All existing calibration data will be lost.
Record the proper data (if possible) before resetting the defaults.
E32.
There is little or no filament current. Verify that filament supply voltage is present at
the filament pcb and that there is continuity in the high voltage secondary.
The SYNERGY™ has filament current feedback, which can be bypassed to allow for
shooting the system with an open secondary (no tube or HV cables) to investigate
the source of secondary arcing. This procedure should only be performed by
personnel familiar with high voltage tests and/or under the guidance of Summit
Technical Support.
When SW1-3 is in the CLOSED (ON) position, the rotor and filament interlocks and
feedbacks are disabled. In this condition, the generator will make kVp, but no mA
and will not spin the anode of the tube. This allows potential to be placed on the
high tension transformer.
Monitor the kVp feedback test point TP8 on the System Controller pcb to look for
evidence of high voltage breakdown, such as spikes to high potential or spikes to
ground. If none are present, the cables or tube would be suspect. When finished
with this test, ensure that SW1-3 is returned to the OPEN (OFF) position.
7-2
SynergyTM Service Manual
6.1 Overview
Once the SYNERGY™ has been installed and calibrated, the outputs must be verified.
There are two parts to this testing. One is to record the kVp and mAs values, ensuring that
they are compliant with the accuracies stated in this manual, as well as federal and local
guidelines. The second is to measure the radiation output to verify that it is both linear and
reproducible.
Note: When calculating kVp or mAs accuracy, the test equipment accuracy must be
subtracted from the stated generator accuracy. If the kVp output is within this new
performance standard then the kVp is certain to be within the generator
specifications, even if the test equipment is at the maximum of it’s error range.
SAMPLE CALCULATIONS:
For 40 kVp, the output must be within ± (40 x 3%) + 1 = (1.2) + 1 = 40 kVp ± 2.2 kVp.
For 100 kVp, the output must be within ±(100 x 3%) +1 = ±(3) + 1 = 100 kVp ± 4 kVp
These sample calculations are for kVp with a 2% test equipment error. The kVp or mAs
output measured and the accuracy of the test equipment used will be different than in these
samples, but the method of generating the system accuracy will be the same.
7-2
SynergyTM Service Manual
NOTE: If the unit does not meet the accuracy specifications, view the mA and
kVp waveforms, verify calibration, and check for line voltages below 230 VAC in
either idle or full load exposure mode. Correct the problem and retest the
system.
7-2
SynergyTM Service Manual
6.3 Reproducibility
To perform this testing, a non-invasive mR meter is required. Dose (mR) will be measured
and verified to be repeatable and predictable within standards set by the FDA. This test
verifies that the mR output remains constant within a prescribed limit over several
exposures at the same technique selection.
1. Select a technique, and make an exposure. Change the technique, return to the
original technique and make another exposure. Repeat until five exposures have been
made. Often this test is done at a technique commonly used by the equipment
operator.
4. Calculate the difference between the average mR, and mAs for each exposure.
7-2
SynergyTM Service Manual
S= 0.83666
7-2
SynergyTM Service Manual
S=
7-2
SynergyTM Service Manual
6.4 Linearity
To perform this test, a non-invasive radiation meter is required. This test verifies that the
dose is consistent and predictable as techniques such as mA or mAs are changed.
2. Record the mR and mAs for each exposure in the table below.
3. Select the next adjacent mAs station, keeping the kVp the same.
4. Record the mR and mAs for each exposure in the table below.
5. For series one, calculate the average mR/mAs. This value is X1.
6. For series two, calculate the average mR/mAs. This value is X2.
7. Calculate Linearity. Linearity is defined as the difference of these two values divided
by their sum, where the two values are mR/mAs of adjacent mAs stations. Ignore
negative numbers—calculate all values as positive integers. This number must be less
than 0.095. Use Table 6-2 below as a guide.
8. Mathematically, these two formulas are identical. Use whichever formula is easiest for
you to use. Coefficient of Linearity (CL) is defined as:
or
7-2
SynergyTM Service Manual
Exposure
Exposure series one Exposure series two (next adjacent
number
mAs station)
Dose (mR) mAs mR/mAs Dose (mR) mAs mR/mAs
1 35 10.4 3.37 42 12.1 3.47
2 33 10.2 3.24 42 12.3 3.41
3 33 10.3 3.24 40 12.0 3.33
4 34 10.3 3.30 41 12.2 3.36
5 34 10.4 3.30 42 12.1 3.47
7-2
SynergyTM Service Manual
Exposure
Exposure series one Exposure series two (next adjacent
number
mAs station)
Dose (mR) mAs mR/mAs Dose (mR) mAs mR/mAs
1
2
3
4
5
7-2
SynergyTM Service Manual
7-2
SynergyTM Service Manual
7.1 Overview
Routine maintenance is to be performed 30 to 60 days after the initial installation and every
six months thereafter. It is the responsibility of the equipment owner to see that this
maintenance is performed as scheduled to meet warranty obligations. The installing dealer
or other factory authorized service organization will be able to perform this maintenance.
The following list is a minimum checklist: your maintenance procedure may be more
extensive than that shown below.
All system cables inspected for wear, binding, and tightness of connections.
High Voltage cables inspected for signs of breakdown, abrasion, or wear; receptacles
regreased and tightened into the HV transformer and x-ray tube.
Line voltage checked, and TB1/TB2 verified to be set properly for the supply.
Power module internal connections checked for wear, binding and tight connections.
Operate all of the switches on the console and verify that the appropriate switches
function properly. Some switches (such as AEC controls) may not be enabled.
Note that all segments on the kVp and mAs windows illuminate properly.
7-2
SynergyTM Service Manual
Date of
Performed by Notes and results
maintenance
7-2
SynergyTM Service Manual
The Power Module contains microprocessor-driven control circuitry to spin the anode of the
x-ray tube, generate filament supply voltages, supply power for locks and collimation,
maintain monitoring of system status, and retain calibration information. Primary voltage for
the high voltage transformer is supplied by IPMs operating in a pulse width modulated
resonant circuit.
The Operator Console is a microprocessor controlled entry/display port connected with the
System Controller through serial communications.
The High Voltage Transformer steps up the voltage received at P1 and P2 to levels
required to make the x-ray tube conduct, ranging from 40 to 125 thousand volts. Filter
capacitors within the tank are part of the tuned resonant circuit. The assembly is filled with
Shell Diala AX, a highly refined non-PCB dielectric oil.
8-1
SynergyTM Service Manual
9.0 SCHEMATICS
The following prints are included in the schematic package part # L269
9-1