AXIOM Aristos Image System


AX

Software
AXIOM Aristos Image System

Retrofit of VB 20 & MX, VX, TX, FX System SW

© Siemens, 2006

10093898

© Siemens, 2006
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.

Print No.: AXB4-320.816.06.02.02 English


Replaces: AXB4-320.816.06.01.02 Doc. Gen. Date: 07.06
CS PS AX RF
n.a.

2006
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer

Copyright
“© Siemens, 2006“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA,
Siemens Healthcare Diagnostics Inc. - USA and/or Siemens Healthcare Diagnostics Prod-
ucts GmbH - Germany.

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 2 of 48 © Siemens, 2006


07.06 CS PS AX RF
Table of Contents 3
0Table of Contents

1 _______ Requirements___________________________________________________ 4

Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Protection of Electrostatically-sensitive Components . . . . . . . . . . . . . . . . . . . . . . . . . 4
Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

2 _______ SW Installation on the Celsius 400/460 & M410 _______________________ 6

VB20 Image System Installation on the Celsius 400/460 & M410 . . . . . . . . . . . . . . . . . . 6


General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

3 _______ SW Installation on an Celsius M430/M440___________________________ 19

Installation on M430/M440 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

4 _______ Updating the Aristos FX System SW _______________________________ 28

Downloading the FX VA03 SW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

5 _______ Remaining Work Steps and Correction of Errors _____________________ 32

Remaining Work Steps and Correction of Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32


Upgrading the Patient Display(s) (if available in MX/VX/TX only) . . . . . . . . . . . . . . . 32
Checking and Activating the System Management (only if remote is present) . . . . . 32
Resetting the Customer OGPs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
OGP Proposals for FX Ortho . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Resetting the Regional and Language Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Adjusting the Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
RADIS Japanese Configuration, Optional, only if needed . . . . . . . . . . . . . . . . . . . . 33
Subsequent Configuration of DICOM DX / CR Object, Optional, only if needed . . . 34
Creating a SW Version Folder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Installing a Paper Printer (Option). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Generating Backups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
For Information, Answers to FAQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
For Information only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

6 _______ Remaining Work Steps and Correction of Errors _____________________ 44

Important Settings for Monitoring Sites only! . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44


Reporting Errors by the Customer or CSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

7 _______ Changes to Previous Version _____________________________________ 47

© Siemens, 2006 AXB4-320.816.06.02.02 Page 3 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
4 Requirements
1-
Requirements
1Requirements

General Remarks 0

This procedure describes the steps necessary to upgrade or reinstall the Radis SW on the
AXIOM Aristos Image System and to update the system SW.

Required Tools 0

• Standard service tools


• SERVICE KEY (20 Characters); can be recorded from system before upgrade
• Syngo License; already available with the system

Safety Information 0

The “Safety Information“ contained in the system manual and in the particular unit instruc-
tions must be observed.

Protection of Electrostatically-sensitive Components 0

Integrated circuits and circuit boards on which MOS components are installed call for par-
ticularly careful handling because of their highly sensitive electrical structures and because
of their extremely high input resistance. The ESD regulations that apply for electrostati-
cally-sensitive components must be observed.

Required Parts 0

Qua
Description
ntity
FD Image System VB30F O. S. or higher (for Celsius 400/460 or M410 only) 1
FD VB30F O. S. or higher Japanese MUI, if this option is needed (for Celsius
1
400/460 or M410 only)
FD Image System Applic. VB20B or higher, CD1 (for Celsius 400/460 or M410
1
only)
FD I.S. Applic. VB20B or higher, CD2 (Syngo) (for Celsius 400/460 or M410
1
only)
Managed Node Package VF10D or higher (for Celsius 400/460 or M410 only) 1
Aristos I.S. VB20B or higher, M430 Clone (for Celsius M430 only) 1
Aristos I.S. VB20B or higher, M440 Clone (for Celsius M440 only) 1

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 4 of 48 © Siemens, 2006


07.06 CS PS AX RF
Requirements 5

Qua
Description
ntity
Aristos System SW CD (MX/VX/TX/FX), Latest Version according to the Com-
1
patibility List
Backup Customer Organ Programs (on Diskette) 1
Backup SW Settings 02 (on Diskette or CD) 1
Backup IQAP (on Diskette or CD) 1
Backup SW Settings, Calibration (on CD) 1
License.dat (also part of backup file “SW Setting 02” for updated systems to
VB10; normally on CD “Aristos License Files VA00A 5917245 for factory 1
shipped VB10, normally on ”Site Data Disk 7728111”).
EPROM for patient display 07049237 (not used for FX display) 2

© Siemens, 2006 AXB4-320.816.06.02.02 Page 5 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
6 SW Installation on the Celsius 400/460 & M410
2-
VB20 Image System Installation on the Celsius 400/460 & M410
2SW Installation on the Celsius 400/460 & M410

General Remarks 0

NOTE This chapter describes the installation on the Celsius


400/460 (green) and M410 (silver without DVD burner). It can
also be used for M430, but in the interest of saving time it is
recommended to use the next chapter describing the DVD
clone installation.

Preparation 0

NOTE Prior to a new installation, all patients must be archived from


the local database because the database is over-written dur-
ing the new installation.

NOTE Before installing the software, a backup of the configuration


(SW Setting 02, IQAP and SW Settings, Calibration) and an
Organ Program (OGP) backup is imperative.

Backup
• Save the host name and IP address:
- > Options > Service > Local Service... > Configuration > TCP/IP LAN
- Computer name = Host name.
- Save the host name and IP address.
- Home > close window.

Host name

IP Address
Subnet mask
Default gateway

Save the Service key:


- > Options > Service > Local Service...
- Close the window.

Service key

Customer Configuration:

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 6 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on the Celsius 400/460 & M410 7

- > Options > Configuration.. > Regional and Language...


- Read out Date format and Time format from “Regional Options” .
- Read out Language and Keyboard from “ Languages > Details...”

Date format
Time format
Language
Keyboard (Input Locales)

• Generate the following backups:


- SW Settings 02 on CD
- SW Settings, Calibration on CD
- IQAP on CD
- Customer OGPs on disk
• To do this:
> Options > Service > Local Service > Backup & Restore
- Insert the Backup CD.
- Select in Drives: [-R-]CD-R; select in Packages: SW Settings 02. Press GO and
wait for the Ready message.
- Insert the Backup CD.
- Select in Drives: [-R-]CD-R; select in Packages: SW Settings, Calibration. Press
GO and wait for the Ready message.
- Insert the Backup CD.
- Select in Drives: [-R-]CD-R; select in Packages: IQAP_Database. Press GO and
wait for the Ready message.
- Insert a new blank diskette, labeled "Customer OGP Diskette VB10 + Date". Go back
to the Local Service main menu. Select Radis Tools > Organ Programs > Backup.
Wait for the "Successfully..." message.
- Shut down the PC.

NOTE Prior to beginning the installation, all network, service and


fiber optic cables must be disconnected. Before disconnect-
ing the cables and fiber optic cables from the old PC, make
sure in particular that the fiber optic cables are properly
labeled.

NOTE Low level formatting of the hard drives is not required!


Low level formatting of both disks could help only if the
installation fails.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 7 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
8 SW Installation on the Celsius 400/460 & M410

Work Steps 0

OS Installation
• Switch on and insert the CD labeled “O.S. AXIOM Aristos Image System VB30F
(10093849 ) CD1” or higher into the CD drive. To restart, switch the power off and on
again.

NOTE SCSI drives can be recognized by an "S" at the end of the


model name, e.g. PlexWriter 12/10/32S.
If a Celsius 400/460 (green) PC will not reboot, a SCSI
CD-Rom drive may be installed but not is not recognized by
the “S”, however it will be listed in the boot menu under
“SCSI drives”. In this case, continue with a. and b.

• If installation is on the Celsius M410 with a SCSI CD drive and Celsius 400/460/470 ,
follow steps a. and b. below.
If M410 is with an IDE CD drive or another PC such as Celsius M430/M440, proceed to
the next step, skip a. and b.):
• a.Invoke the boot menu by pressing F12 immediately on the initial power-up screen.
For Fujitsu Siemens screens, you have to press the key before the screen displays the
Startup menu.
b. When the boot menu is shown, select SCSI CD Drive and <Enter>.
• When prompted, “Confirm to DESTROY DATA on first...”, confirm with <Enter> . A
reboot will occur and with a M410 with a SCSI CD drive (if IDE CD drive, proceed to
next step, skip a. and b.):
a. Invoke the boot menu by pressing F12 immediately on the initial power-up screen .
For Fujitsu Siemens screens, you have to press the key before the screen shows the
startup menu.
b. When the boot menu is shown, select SCSI CD Drive again and <Enter>.
- If you fail to press the "F12" key at the right moment, there will be an error message
saying "partition not found". Restart the RADIS PC and perform this step again.
• When prompted, set the following values. To navigate, use the arrow keys.
• "Distribution type: Customer"; keep the default value.
• "UDB file: on CD-Rom"; keep the default value.
• Set the correct machine name as recorded (computer name = host name).
Press <Enter>, type in the machine name and press <Enter> again.
• Go to the next line with the arrow keys "Keep data on other partition(s)" and change
the selection to "Create partitions on hard drive 2" by pressing <ENTER> (the
ENTER key will toggle between "Keep data..." and "Clear partitions...").
• Go to "Continue Installation" and confirm with > Enter > Enter.
• Wait until file copying of the installation file starts.
• At this point, the OS will run unattended for approximately 45 min.
• In some cases, a message ““Matrox PowerDesc..” will come up. You can ignore it by
checking “Do not display .”

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 8 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on the Celsius 400/460 & M410 9

• At the end of the installation, the unattended installation will stop and a black command
message box will be displayed and will indicate "Warning, no package was found on the
CD. Please install the package manually and press any key to close partially unat-
tended installation …". At this point, click inside the message box and press any key
(except the Enter key - if <Enter> is pressed, no test images will be loaded.).
• Another black command message box will open. Click inside the message box and
press any key (except the Enter key). Deletion of the OEM setup files will now start.
This will take about 1 min. No progress bar will be displayed! A reboot will occur auto-
matically.
• After this reboot, you will be asked for the administrator password. Enter the following:
"adm$pwd$4$med".

NOTE At this point, the keyboard layout is set to English (US).


During the installation, keep the language settings as they
are. They will be changed later (by "Restore SW Settings).

Leave the cables disconnected.

Installation of OS Patches
• Insert the AXIOM Aristos Image System VB30F O.S. or higher CD2 (10093849 ) into
the CD drive.
• Open Windows Explorer and double-click on the file
R:\OEMSetup\WIN_UPDT\ inswinup.bat .
• The installation will run unattended for approximately 5 min.
• Double-click on the file
R:\OEMSetup\WIN_UPDT\ insxmlup.bat .
• Wait for the end of the installation. (less than 2min.)
• Remove the CD.

Installation of Japanese Support (MUI), Optional, only if needed


• Insert the Japanese MUI CD into the drive.
• Open Windows Explorer and double-click on the file R:\Radis_Japanese_install.bat.
• The installation will run unattended for approximately 5 min. Then an Information dialog
"Multilingual file installation was completed successfully" will appear. Confirm this dia-
log with “OK”.
• Remove the CD.

Checking the Time Zone and Time

NOTE At this point in the installation, only the time settings may be
checked or changed. All other customer configurations must
be performed later in > Options > Configurations...

• Go to Start > Settings > Control Panel > Date and Time.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 9 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
10 SW Installation on the Celsius 400/460 & M410

• Go to the tab card “Time Zone” and set the correct Time Zone.
• Check “Automatically adjust clock for daylight saving changes”.
• Confirm by clicking on “Apply”.
• Go to the “Date & Time” tab card and set the correct Date and Time.
• Confirm by clicking on “OK” and close the Control Panel.

Installing the Aristos Service SW on the Radis PC

NOTE The service SW must be installed to Radis PC or else the sys-


tem will not boot up in application.

• Insert the Software CD-Rom according the Compatibility List for the AXIOM Aristos
MX/VX, TX or FX into the CD drive.
• Open “Windows Explorer”, select the CD-Rom drive and double-click on the
“start.exe” file. The Installer window opens.
• Click on “OK”.
• Click on the link "Service SW overall installation".
• Choose Full Install.
• Choose (Do not “Create”) > Skip.

NOTE Please choose “skip” for Installation. Otherwise an error


message will pop up. (You can ignore this error message
without any problems.)

• Choose Continue > OK.

NOTE If the Service software has to be installed again, the follow-


ing batch file has to be run after reinstallation. > Start > Run...
and type
C:\Radis\RadisRemote\Utils\RadisRemoteInstall.bat

Generating the new XCU Floppy


• For Aristos FX: Insert a blank, formatted floppy into the floppy drive of the RADIS PC
and click on "VA03A" (or higher) under "XCU floppy Software of Aristos FX".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.
• For Aristos MX/VX: Insert a blank, formatted floppy into the floppy drive of the RADIS
PC and click on "VD01A" (or higher) under "XCU floppy Software of Aristos MX".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 10 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on the Celsius 400/460 & M410 11

• For Aristos TX only: Insert a blank, formatted floppy into the floppy drive of the RADIS
PC and click on "VA02A" (or higher) under "XCU floppy Software of Aristos TX".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.
• Label the created floppy appropriately. Keep it for later use.
• Close the Service SW windows and remove CD.

Downloading the new XCU Floppy


• Insert the floppy just created into the XCU drive.
• Reboot the XCU by unplugging and plugging in the "XCU X1" power connector.
• Wait until the SW has downloaded. After the program has downloaded, an audible sig-
nal is sounded (TOOT-TA-TOOT) (three beeps).
• Remove the disk.
• Reboot the XCU by unplugging and plugging in the "XCU X1" power connector.

Deactivating the Serial Mouse at COM 1


In some cases, the mouse cursor sticks in the top right corner when the RADIS PC is con-
nected to the XCU; this is because the RADIS PC thinks that an RS232 mouse is con-
nected to its RS 232 interface and interprets the signals from the XCU as a "mouse
command". To prevent this, the RS 232 mouse must be disabled on the RADS PC.
• Select Start > Run > "devmgmt.msc", using "Windows key", arrow keys and Enter
key.

Fig. 1:
• After starting the Device Manager, select the "Mice and other pointing devices"
hardware components and select the context menu with the right mouse button.
• Connect the serial XCU service cable to the Radis PC.
• Select "Scan for hardware changes".
- If the serial cable is connected, the serial mouse driver will be found and the mouse
cursor will begin to move by itself.
- If the “mice and other pointing devices” was not found, repeat “scan for hard-
ware changes” several times.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 11 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
12 SW Installation on the Celsius 400/460 & M410

Fig. 2:
• Press Alt + Tab to activate the Device Manager window.
• Use the up/down arrow keys to navigate to "Mice and other pointing devices" (use
left/right arrow keys to open the submenus).

Fig. 3:
• Select "Microsoft Serial BallPoint".
• Press Enter.
• Press Tab until "Device Usage" is selected.
• Use the up/down arrow keys to select "Do not use this device (disable)".

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 12 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on the Celsius 400/460 & M410 13

• Use Tab until OK is activated; press Enter.

Fig. 4:
• The mouse for this hardware profile will now be deactivated for COM 1 and the Device
Manager will display a red cross in front of the deactivated component.

Fig. 5:

© Siemens, 2006 AXB4-320.816.06.02.02 Page 13 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
14 SW Installation on the Celsius 400/460 & M410

• Close all windows.

Installing AXIOM Aristos VB20 Image System Application

NOTE The fiber optic cables from Radis to XCU and detector plus
the network cable must be removed until instructed to recon-
nect them.

• Insert the first application CD (RADIS_AP cd 1) in CD drive and open the Windows
Explorer (Win+ E) and double-click on "R:setup.exe".
• Confirm (2x) with Next >, Installation start of FD-X
• When prompted, “Please enter RADIS-AP Disk 2” remove the RADIS_AP CD1 from the
CD drive and replace it with the RADIS_AP CD 2. In the dialog box, acknowledge the
CD drive path of (R:\) by selecting OK.
• The installation will continue unattended for approximately 20 min., and there will be a
request for a reboot now or later. Choose “Yes, I want to restart my computer now” >
Finish.
Any alarm messages from the firewall that occur in the interim should be ignored with
“Unblock”.
After reboot, the Syngo count-up screen will automatically start and count up to "99".
The windows desktop will now be visible.
• Remove the CD from the drive.

System Management Installation


• Insert "Managed Note Package MNP VF10D (P/N 3223026)” or higher.
• In Windows Explorer, select CD-ROM (R).
• Double-click on <install.bat>
Once the Command window closes, the System Management installation is completed.
• Remove the CD from drive.
• After "Restore" of SW Settings at the end of installation, please check the correct func-
tions according the "RDIAG and Remote Service Installation and Startup Instruc-
tions", AXB4-000.814.03.02.02 or higher.

Installing the Virus Scanner

NOTE For automatic update of the virus scanner (automatic disk


update), a valid license code is required.

• Insert the Software CD-Rom according the Compatibility List for the AXIOM Aristos
MX/VX, TX or FX into the CD drive.
• Open “Windows Explorer”, select the CD-Rom drive and double-click on the
“start.exe” file. The Installer window opens.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 14 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on the Celsius 400/460 & M410 15

• Under “Virus Scanner for RADIS”, click on the link “e.g. V7.0(UI) / V7.510
(ScanEngine)” or higher and wait for installation window closes.
• Close the service window.
• Remove the CD from the drive.

NOTE If the customer has no license for the virus scanner, the virus
scanner must be deactivated for legal licensing reasons. For
a customer with a valid license, the following point, “Deacti-
vating the Virus Scanner and Deleting the Virus Pattern”
must be ignored.

• Deactivating the Virus Scanner and Deleting the Virus Pattern:


Start > Run > services.msc > “OK”
In the window “Service”, double-click on the service “OfficeScanNT RealTime Scan” .
Change “Startup type:” to “Disabled”.
Confirm with “OK”
Open “Windows Explorer”, select C:\Program Files\OfficeScan NT
Delete all files beginning with LPT$VPN. E.g. LPT$VPN.220 and LPT$VPN.749...

DICOM Attribute Configurator (DAC), Optional, only if required


• For installation of DAC, use the Installation Instructions (for DAC Version VA20 or
higher), AX00-000.816.01.01.02 or higher.

RADIS, Initial Setup


• Restart the PC from the "Start Menu".
After the reboot, the Syngo count-up screen will start. Syngo will require you to complete
initial configuration in local service.
• Follow the links to local service and enter the password appropriate for your system.
• Use the temporary password: 1D328F7ECC927C F46C1F, which is valid until Aug.
30, 2006 (later, a new password for Vendor ID: "1234-SMS -AX-5678" will be needed).
Check “Set as Default”.
• If the password does not work, the fiber optic cable may be connected.
• A dialog box will ask if you want to use Internet Explorer for your default browser.
• Select the “Configuration” button.
• Select the correct “System type” (e.g Aristos MX_DX).

NOTE If the option "DX object" was acquired simultaneously with


the upgrade from VB10 to VB20, the correct system type
(Aristos MX_DX, Aristos TX_DX or Aristos FX_DX) should
have already been set. It is also possible to do this at a later
time, but the DB must then be deleted and regenerated.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 15 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
16 SW Installation on the Celsius 400/460 & M410

• Continue to press Next until you get to the page with a system serial number:
Enter "1" or enter the correct system Serial No., if available, and save.

NOTE It is very important to set a Serial No. different to "0" to force


the function "restart pending" and auto restart at the end.

• A dialog box will state that the system type has been changed, save new values?
Respond with OK.
• Continue to press Next or > until the configuration page is displayed for users: Under
Account for Administrator, enter the password adm$pwd$4$med twice.
Select "save".
• Continue to select Next through the remainder of the configuration pages until “Finish”
and press “finish”.
Do not enter any additional data in the configuration pages during this initial configura-
tion session.
• Select Home when the last page is displayed.
• While the window with a restart request is opening, select OK to restart.

NOTE The restart may take several minutes and may not be inter-
rupted by the user.
Do not shut down the PC manually.

• After the reboot, Syngo will start its count-up screen. Note that the Syngo databases will
be created during this count up.
The Radis Viewer will then come up.
• > Options > Service > Local Service...
• Use the temporary password: 1D328F7ECC927C F46C1F, which is valid until Aug.
30, 2006 (later, a new password for Vendor ID: "1234-SMS -AX-5678" will be needed).
Check “Set as Default”.
• Select "Backup & Restore" > "Command" >"Restore"
• Insert the backup medium.
• Under Drives, select [-R-] CD-R (or [-A-] Floppy Disk).
• Under "Archive", select "..._SW-Settings02_...".
• Under "Groups", select all entries.
• Select "Go" and wait for ready.
• Under "Archive", select "..._SW Settings Calibration....".
• Under "Groups", select all entries.
• Select "Go" and wait for ready.
• Under "Archive", select "..._IQAP_Database_...".
• Under "Groups", select "all".
• Select "Go" and wait for ready.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 16 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on the Celsius 400/460 & M410 17

• Select Home when the last page is displayed.


• Close window.
• Shut down Radis PC:
> Start > Shutdown > Restart System > Yes
• During restart connect all FOC and network cable.

Import License for Additional Functions

NOTE If new functionalities are also to be used with the VB20


upgrade, such as: - Virus scanner - DX Object - Ortho for
Aristos FX, a download is necessary to supplement the new
functions with licenses.
If no new functionalities are used, the original License.dat is
already contained in the “Restore” and does not have to be
additionally downloaded. Skip license import and go to
“Radis, Final Setup” (next page).

• To do this:
> Options > Service > Local Service... > Configuration
- > Next
- > Licensing
- Insert the data medium with the new License.dat (it is possible that the new license file
still needs to be renamed to “license.dat”.
- Click on “Browse..” and go to new “license.dat”.
- Click on “Open” > “Import License file” and confirm with “OK”.
- Select “Home” and close the window.

RADIS, Final Setup, Setting Firewall and Security


• Reboot the Radis PC
> Options > End Session... > Restart System... > Yes
• During restart, log in as Administrator
(During reboot hold the left <Shift> key)
User: Administrator
Password: adm$pwd$4$med
• Perform > Start > Run > C:\post_clone_tools\SetFirewallOn.bat
> OK.
• Perform > Start > Run > C:\post_clone_tools\SetSecurity.bat
> OK.
• Perform > Start > Run > C:\RADIS\Bin\Utilities\SetRadisLogDir.bat
> OK.
• Restart the Radis PC.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 17 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
18 SW Installation on the Celsius 400/460 & M410

• Continue with chapter “Updating Aristos FX System SW” if it is a FX or


“Remaining work steps and correction of errors”.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 18 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on an Celsius M430/M440 19
3-
Installation on M430/M440
3SW Installation on an Celsius M430/M440

General Remarks 0

NOTE This chapter describes the installation with a provided clone


DVD (Ghost image) for M430/M440 only. Please use this way
in compare to CD set installation (chapter 2) because of time
saving.
If there are problems, an installation according chapter 2 is
also possible with M430.

Preparation 0

NOTE Prior to a new installation, all patients must be archived from


the local database because the database is over-written dur-
ing the new installation.

NOTE Before installing the software, a backup of the configuration


(SW Setting 02, IQAP and SW Settings, Calibration) and an
Organ Program (OGP) backup is imperative.

Backup
• Save the host name and IP address:
- > Options > Service > Local Service... > Configuration > TCP/IP LAN
- Computer name = Host name.
- Save the host name and IP address.
- Home > close window.

Host Name

IP Address
Subnet mask
Default gateway

Save the Service key:


- > Options > Service > Local Service...
- Close the window.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 19 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
20 SW Installation on an Celsius M430/M440

Service key

Customer Configuration:
- > Options > Configuration.. > Regional and Language...
- Read out the date format and time format from “Regional Options”.
- Read out the language and keyboard from “ Languages > Details...”

Date format
Time format
Language
Keyboard (Input Locales)

• Generate the following backups:


- SW Settings 02 on CD
- SW Settings, Calibration on CD
- IQAP on CD
- Customer OGPs on disk
• To do this:
> Options > Service > Local Service > Backup & Restore
- Insert the Backup CD.
- Select in Drives: [-R-]CD-R; select in Packages: SW Settings 02. Press GO and
wait for the Ready message.
- Insert the Backup CD.
- Select in Drives: [-R-]CD-R; select in Packages: SW Settings, Calibration. Press
GO and wait for the Ready message.
- Insert the Backup CD.
- Select in Drives: [-R-]CD-R; select in Packages: IQAP_Database. Press GO and
wait for the Ready message.
- Insert a new blank diskette, labeled "Customer OGP Diskette VB10 + Date". Go back
to the Local Service main menu. Select Radis Tools > Organ Programs > Backup.
Wait for the "Successfully..." message.
- Shut down the PC.

NOTE Prior to beginning the installation, all network, service and


fiber optic cables must be disconnected. Before disconnect-
ing the cables and fiber optic cables from the old PC, in par-
ticular make sure that the fiber optic cables are properly
labeled.

NOTE Low level formatting of the hard drives is not required!


Low level formatting of both disks could help only if the
installation fails.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 20 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on an Celsius M430/M440 21

Work Steps 0

Clone Installation
• Switch on and insert the DVD labeled “Aristos I.S. VB20 M430 Clone” or higher for
M430 or “Aristos I.S. VB20 M440 Clone” or higher for M440.
To restart, switch the power off and on again.
• A menu appears. Chose “3” (Restore).
• Confirm “Are you Sure (Y/N)” with “Y”.
• Ghost starts.
Set highlighting with the cursor key to the button “Continue without marking drives”
and confirm by pressing “Enter”.
• The DVD image is installed. Wait for installation to finish (approx. 15 min.)
• Remove the DVD.
• Restart the PC by pressing “Ctrl.+Alt+Del.”.
• After the restart, enter the host name that was recorded (computer name = host name).
After confirmation with “Next”, the PC will restart automatically.
• After restart, a text file opens. Close the file without any changes and wait for the auto-
matic restart (this may take 5 min.)
• After restart, log in as Administrator.
User: administrator
Password: adm$pwd$4$med
Leave the cables disconnected.

Checking the Time Zone and Time

NOTE At this point in the installation, only the time settings may be
checked or changed. All other customer configurations must
be performed later in > Options > Configurations...

• Go to Start > Settings > Control Panel > Date and Time.
• Go to the “Time Zone” tab card and set the correct Time Zone.
• Check “Automatically adjust clock for daylight saving changes”.
• Confirm by clicking on “Apply”.
• Go to the “Date & Time” tab card and set the correct Date and Time.
• Confirm by clicking on “OK” and close the Control Panel.

Installing the Aristos Service SW on the Radis PC


• Insert the software CD-Rom according the Compatibility List for the AXIOM Aristos
MX/VX, TX or FX into the CD drive.
• Open “Windows Explorer”, select the CD-Rom drive and double-click on the
“start.exe” file. The Installer window opens.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 21 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
22 SW Installation on an Celsius M430/M440

• Click on “OK”.
• Click on the link "Service SW overall installation".
• Choose Full Install.
• Choose (Do not “Create”) > Skip.

NOTE Please choose “skip” for Installation. Otherwise an error


message will pop up. (You can ignore this error message
without any problems.)

• Choose Continue > OK.


• Go to > Start > Run... and type
C:\Radis\RadisRemote\Utils\RadisRemoteInstall.bat manual (there is a space in
front of <manual>).

Generating the new XCU Floppy


• For Aristos FX: Insert a blank, formatted floppy into the floppy drive of the RADIS PC
and click on "VA03A" (or higher) under "XCU floppy Software of Aristos FX".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.
• For Aristos MX/VX: Insert a blank, formatted floppy into the floppy drive of the RADIS
PC and click on "VD01A" (or higher) under "XCU floppy Software of Aristos MX".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.
• For Aristos TX only: Insert a blank, formatted floppy into the floppy drive of the RADIS
PC and click on "VA02A" (or higher) under "XCU floppy Software of Aristos TX".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.
• Label the created floppy appropriately. Keep it for later use.
• Close the Service SW windows and remove the CD.

Downloading the new XCU Floppy


• Insert the floppy just created into the XCU drive.
• Reboot the XCU by unplugging and plugging in the "XCU X1" power connector.
• Wait until the SW has downloaded. After the program has downloaded, an audible sig-
nal is sounded (TOOT-TA-TOOT) (three beeps).
• Remove the disk.
• Reboot the XCU by unplugging and plugging in the "XCU X1" power connector.

Deactivating the Serial Mouse at COM 1


In some cases, the mouse cursor sticks in the top right corner when the RADIS PC is con-
nected to the XCU; this is because the RADIS PC thinks that an RS232 mouse is con-
nected to its RS 232 interface and interprets the signals from the XCU as a "mouse
command". To prevent this, the RS 232 mouse must be disabled on the RADS PC.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 22 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on an Celsius M430/M440 23

• Select Start > Run > "devmgmt.msc", using the "Windows key", arrow keys and Enter
key.

Fig. 6:
• After starting the Device Manager, select the "Mice and other pointing devices"
hardware components and select the context menu with the right mouse button.
• Connect the serial XCU service cable to the Radis PC.
• Select "Scan for hardware changes".
- If the serial cable is connected, the serial mouse driver will be found and the mouse
cursor will begin to move by itself.
- If the “mice and other pointing devices” was not found, repeat “scan for hard-
ware changes” several times.

Fig. 7:
• Press Alt + Tab to activate the Device Manager window.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 23 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
24 SW Installation on an Celsius M430/M440

• Use the up/down arrow keys to navigate to "Mice and other pointing devices" (use
left/right arrow keys to open the submenus).

Fig. 8:
• Select "Microsoft Serial Ball Point".
• Press Enter.
• Press Tab until "Device Usage" is selected.
• Use the up/down arrow keys to select "Do not use this device (disable)".
• Use Tab until OK is activated; press Enter.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 24 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on an Celsius M430/M440 25

Fig. 9:
• The mouse for this hardware profile will now be deactivated for COM 1 and the Device
Manager will display a red cross in front of the deactivated component.

Fig. 10:
• Close all windows.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 25 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
26 SW Installation on an Celsius M430/M440

RADIS, Initial Setup


• Start the “Internet Explorer”.
• Use the temporary password: 1D328F7ECC927C F46C1F, which is valid until Aug.
30, 2006 (later, a new password for Vendor ID: "1234-SMS -AX-5678" will be needed).
Check “Set as Default”.
• Select "Backup & Restore" > "Command" >"Restore"
• Insert the backup medium.
• Under Drives, select [-R-] CD-R (or [-A-] Floppy Disk).
• Under "Archive", select "..._SW-Settings02_...".
• Under "Groups", select all entries.
• Select "Go" and wait for ready.
• Under "Archive", select "..._SW Settings Calibration....".
• Under "Groups", select all entries.
• Select "Go" and wait for ready.
• Under "Archive", select "..._IQAP_Database_...".
• Under "Groups", select "all".
• Select "Go" and wait for ready.
• Select Home.
• Close the window.
• Shut down the Radis PC:
> Start > Shutdown > Restart System > Yes
• After the reboot, syngo will start its count-up screen.
• Connect the Radis PC to the lab with fiber optic cables from the detector and XCU.

Import License for Additional Functions

NOTE If new functionalities are also to be used with the VB20


upgrade, such as: - Virus scanner, DX Object, Ortho for Aris-
tos FX, a download is necessary to supplement the new func-
tions with licenses.
If no new functionalities are used, the original License.dat is
already included in the “Restore” and does not have to be
downloaded again. Skip license import and go to “Radis,
Final Setup”.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 26 of 48 © Siemens, 2006


07.06 CS PS AX RF
SW Installation on an Celsius M430/M440 27

• To do this:
> Options > Service > Local Service... > Configuration
- > Next
- > Licensing
- Insert the data medium with the new License.dat (it is possible that the new license file
still needs to be renamed to “license.dat”.
- Click on “Browse..” and go to new “license.dat”.
- Click on “Open” > “Import License file” and confirm with “OK”.
- Select “Home” and close the window.

RADIS, Final Setup, Setting Firewall and Security


• Reboot the Radis PC
> Options > End Session... > Restart System... > Yes
• During restart, log in as Administrator
(During reboot hold the left <Shift> key)
User: Administrator
Password: adm$pwd$4$med
• Perform > Start > Run > C:\post_clone_tools\SetFirewallOn.bat
> OK.
• Perform > Start > Run > C:\post_clone_tools\SetSecurity.bat
> OK.
• Perform > Start > Run > C:\RADIS\Bin\Utilities\SetRadisLogDir.bat
> OK.
• Restart the Radis PC.
• Continue with the chapter “Updating Aristos FX System SW” if it is a FX or
“Remaining work steps and correction of errors”.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 27 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
28 Updating the Aristos FX System SW
4-
Downloading the FX VA03 SW
4Updating the Aristos FX System SW

NOTE This update takes approx. 30...45 min. and should be done
parallel to the Radis update with the Service PC.

• Connect the Service PC to the gateway master (use the short adapter cable) or use the
Radis PC: Options > Local Service > RADIS Tools > Service Software ...
• Start the Aristos FX service SW.
- Log in with your name. The password is "_Siemens" (case-sensitive!).
• Select Aristos FX SSW >Backup/Restore (1/Fig.1).

Fig. 11:
• Select Update (2/Fig. 11 / p. 28) and check the status (software version) of the compo-
nents and the "interface tasks" in the status window.
• Select Aristos FX SSW >Backup/Restore>Download (1/Fig. 12 / p. 29).
• Choose Select version (2/Fig. 12 / p. 29).
• In the window that comes up, click on VA03A or higher (3/Fig. 12 / p. 29) and confirm
with “OK”.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 28 of 48 © Siemens, 2006


07.06 CS PS AX RF
Updating the Aristos FX System SW 29

• Click on Select target (4/Fig. 12 / p. 29). The "Download targets" window


(Fig. 13 / p. 30) will open.

Fig. 12:
• Click on Select all.
• Click on Download selected (2/Fig. 13 / p. 30).
• Wait until the download has successfully finished (observe the status line and the
dimmed buttons).

© Siemens, 2006 AXB4-320.816.06.02.02 Page 29 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
30 Updating the Aristos FX System SW

NOTE The Aristos FX stand update should be performed simulta-


neously with other work steps. The download will take some
time.

Fig. 13:

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 30 of 48 © Siemens, 2006


07.06 CS PS AX RF
Updating the Aristos FX System SW 31

• Check the software version of all components (Fig. 14 / p. 31) (click on Update); it must
be VA03A (or higher) for all components. (1. in (Fig. 14 / p. 31) means table handle bar)

Fig. 14:
• Quit the service SW and disconnect the Service PC.
• Switch the system off completely (switch at the control cabinet) and on again.
• Continue with the chapter “Remaining work steps and correction of errors”.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 31 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
32 Remaining Work Steps and Correction of Errors
5-
Remaining Work Steps and Correction of Errors
5Remaining Work Steps and Correction of Errors

Upgrading the Patient Display(s) (if available in MX/VX/TX only) 0

• For 3D Top only:


For better access to the housing cover, it is recommended that the 2 screws on the
securing cover be slightly loosened (do not completely remove them) and pull the
bracket forward.
• For all Systems (MX/VX/TX):
- Open the housing on the patient display(s) by removing the four screws on the back.
- Replace the EPROM. When installing the EPROM, do not bend any pins. Pay atten-
tion to the orientation of the IC!
- Close the housing and install the display.

Checking and Activating the System Management (only if remote is present) 0

• The check and activation is performed per the instructions “RDIAG and Remote Ser-
vice beginning with SW VB10” AXB4-000.814.03.01 or higher".

Resetting the Customer OGPs 0

• > Options > Service > Local Service...


• Go to “Radis Tools” > “Organ Programs”.
• Insert the floppy with the Customer OGP.
• Under “Drive Select”, select “A”.
• Click on “Scan Drive”.
• Click on “Replace”
• Close the windows.

Checking RIS Match in the OGP Editor


OGP Proposals for FX Ortho 0

• Generate OGPs in discussion with the customer


Abbreviations:
HA KS = Harmonization Kernel
HA Gain = Harmonization Gain
EE KS = Edge Enhancement Kernel (Spatial Filter)
EE Gain = Edge Enhancement Gain (Spatial Filter Gain)

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 32 of 48 © Siemens, 2006


07.06 CS PS AX RF
Remaining Work Steps and Correction of Errors 33

Sens M CU Format Format Grid


OGP Name kV mAs Focus
. field filter Width Height cm

O150 Long Leg 77 40 large 800 R&L Y 3810 11430 300


O151 Full Spine pa 77 40 large 800 all Y 3810 11430 300
O152 Full Spine lat Adult 83 28 large 800 all Y 3810 11430 300
O153 Full Spine lat Child 83 10 large 800 all Y 3810 11430 300
O154 Full Spine lat Obese 90 40 large 800 all Y 3810 11430 300

Resetting the Regional and Language Settings 0

> Options > Configuration.. > Regional and Language...


Use the recorded data from the chapter “Preparation” .
- Reset the date format and time format from “Regional Options”.
- Reset the language and keyboard from “ Languages > Details...”

Adjusting the Volume 0

• Unlock the window keys by pressing "Tab" + "Del" + "+".


• Start > Settings > Control Panel > Switch to Classical View > Sounds and Audio
Devices.
• Set “Device Volume” to high (move it to the right).
• Press the "Speaker volume..." button under “Speaker Setting”.
(another dialog opens), move the volume slider to High (all the way to the right).
• Close all windows with OK.

RADIS Japanese Configuration, Optional, only if needed 0

Syngo Settings
> Options > Service > Local Service... > Configuration > Character Set
• Choose the option “allow multiple character sets”. Select in the drop-down list” ISO
2022 IR 13” (Japanese).
• Select “Save” and confirm “OK”.
• Select “Finish” and “Home”. Avoid the message about the required restart with Cancel.
• Select “Finish” and “Home” .
• Close the service window.

Regional Settings
> Options > Configuration > Regional and Language Options

© Siemens, 2006 AXB4-320.816.06.02.02 Page 33 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
34 Remaining Work Steps and Correction of Errors

• Under “Standards and formats”, choose “Japanese”.


If necessary, change the appropriate formats such as date and units using the button
“Customize...”.”location”.
Note: Currently only two characters are permitted for the month. (YYYY\MM\DD).
Under “Location”, select the location “Japan”.
• Confirm with “Apply”.
• Change to the “Languages” tab card.
• Under “Languages used in menus and dialogs”, change to “Japanese”. Select “Fin-
ish” and “Home”.
• Click on the “details..” button and select “Japanese - Microsoft IME Standard 2002
ver. 8.1”.
• Confirm with “Apply”.
• Click on the “Language Bar...” and mark it with a checkmark:
“Show the language bar on the desktop”
“Show additional Language bar icons in the taskbar”

• Close this window with “OK”.


• Select “Apply” and confirm the message for a restart.
• Confirm “logOff” with “OK”.
• Wait a few seconds and confirm the Restart with “OK”.

Subsequent Configuration of DICOM DX / CR Object, Optional, only if needed 0

NOTE Use of the "DICOM DX Object" is a option which can be used


only with a valid license. (Customer has to pay for this fea-
ture).

These steps are required only if patients were already generated in the database under
another “System type” (e.g. CR).
> Options > Service > Local Service... > Configuration
• Select the correct system type under "System type”, .
• Select “Next” and confirm the necessary restart by clicking on “OK”.
• During restart, log in as Administrator
(During reboot, hold the left <Shift> key)
User: Administrator
Password: adm$pwd$4$med
• Start > Run... and type C:\RADIS\bin\Utilities\DeleteDB.bat manual (there is a
space in front of <manual>)
• Confirm the message “Are you sure...” with “Y”. Go to Start > Shutdown > Restart.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 34 of 48 © Siemens, 2006


07.06 CS PS AX RF
Remaining Work Steps and Correction of Errors 35

• The RADIS bootup can take longer because the databases and the service patients are
regenerated (up to 20 min.).

Creating a SW Version Folder 0

To obtain quick access to the exact SW status of all system components, e.g. for service
escalations, this SW Info is copied into the Radis PC, where it can be accessed remotely.
> Options > Service > Local Service... > Utilities
• Under "Source", select "Escape to OS"
• Type < EXPLORER > into the "Parameter" mask (a Windows Explorer will open).
• Go to the folder MED_Data (D): and create a new folder with the name "Service" (right
mouse-click in D:\ folder > New > Folder > rename to "Service")
• Close the Windows Explorer.
• Click on "Home" in the "Utilities" window.
• Go to "Radis Tools".
• Go to "Service Software" and confirm the message with "OK".
• Log into the XCS SSW.
• In the XCS SSW window, click on Info > SW/FW Version. An information window will
open.
• Click on "Print to File"
• In the "Save As" window, browse to D:\Service.
• Do not change the file name.
• Click on "OK".
The SW Info file is now saved in D:\Service and is accessible remotely.
• Close all windows and log off from the XCU.

Installing a Paper Printer (Option) 0

• Go to Options > Service > Local Service > (enter the password) > Utilities > Escape
to OS > (enter "explorer").
• Click on My Computer > Control Panel > Printers and Faxes > Add Printer.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 35 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
36 Remaining Work Steps and Correction of Errors

• Follow the self-explanatory screen shots.

Fig. 15: Fig. 16:

• Configure a network printer (Fig. 16 / p. 36) only if you want to print to a print server
(Fig. 35 / p. 41) and (Fig. 36 / p. 41).

Fig. 17: Fig. 18:

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 36 of 48 © Siemens, 2006


07.06 CS PS AX RF
Remaining Work Steps and Correction of Errors 37

• Select Create a new port and Standard TCP/IP Port (Fig. 17 / p. 36).

Fig. 20:

Fig. 19:

• Enter the IP address of the printer (Fig. 19 / p. 37).

NOTE If the printer can be allocated, continue with (Fig. 23 / p. 38).


If the printer cannot be allocated (e.g. printer is switched off),
continue with (Fig. 21 / p. 37)

Fig. 21: Fig. 22:

© Siemens, 2006 AXB4-320.816.06.02.02 Page 37 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
38 Remaining Work Steps and Correction of Errors

Fig. 23: Fig. 24:

• Select printer manufacturer and type, or have the printer’s floppy to load the driver.

NOTE Configure postscript printers (suffix "PS") only!

Fig. 25: Fig. 26:

NOTE Do NOT print a test page (Fig. 26 / p. 38).

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 38 of 48 © Siemens, 2006


07.06 CS PS AX RF
Remaining Work Steps and Correction of Errors 39

Fig. 28:

Fig. 27:

• Click on Finish.
• For printer setup, click on Properties (Fig. 28 / p. 39).

Fig. 29: Fig. 30:

© Siemens, 2006 AXB4-320.816.06.02.02 Page 39 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
40 Remaining Work Steps and Correction of Errors

• In the Properties menu, select Printing Preferences (Fig. 29 / p. 39).

Fig. 31: Fig. 32:

• Select Paper Size, e.g. A4 (Fig. 31 / p. 40).


• In Device Settings, select the paper size for Tray 1(Fig. 32 / p. 40).

Fig. 33: Fig. 34:

• Select Ports (Fig. 32 / p. 40).


• Select radio button LPR (Fig. 33 / p. 40) and enter any Queue Name, e.g. LPT
(Fig. 34 / p. 40).
• Close the printer’s properties dialogue with OK and print a test page.
• To configure a printer connected through a print server, select A network printer...
(Fig. 35 / p. 41) (corresponds to (Fig. 16 / p. 36)).

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 40 of 48 © Siemens, 2006


07.06 CS PS AX RF
Remaining Work Steps and Correction of Errors 41

• Click on Next.

Fig. 35: Fig. 36:

• In the menu that comes up, enter the printer’s name and path ((Fig. 36 / p. 41) (the
screen shot is in German, please apply accordingly).
• Continue with (Fig. 17 / p. 36).

Generating Backups 0

• If settings were changed, generate the necessary backups:


- SW Settings 02 on CD
- SW Settings, Calibration on CD
- IQAP on CD
- Customer OGPs on disk
• In the application window, enter Options > Service > Local Service > Backup &
Restore
- Select Backup; select in Drives: [-R-]CD-R; select in Packages: SW Settings 02;
press GO and wait for the Ready message.
- Select in Drives: [-R-]CD-R; select in Packages: SW Settings Calibration; press
GO and wait for the Ready message.
- Select in Drives: [-R-]CD-R; select in Packages: IQAP_Database; press GO and
wait for the Ready message.
- Go back to the Local Service main menu and select Radis Tools > Organ Pro-
grams > Backup. Wait for the "completed" message.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 41 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
42 Remaining Work Steps and Correction of Errors

For Information, Answers to FAQ 5.1

For Information only 0

Display Problems with Eizo Monitor


If there are display problems with an EIZO monitor connected to Radis via the Digital Video
Interface (DVI) in the Celsius M410 (gray housing), carry out these settings:
• Go to Local Service Home Menu > Utilities > select: Escape to OS > Parameters:
enter: control panel; press GO; if necessary, select "classical view" in the window
that comes up; select: Display > Settings tab card.
- Color quality: Highest (32 bit)
- Screen resolution: 1280 x 1024 pixels
- Click on "Advanced" > Monitor > Set Screen refresh rate: 60Hz
- Close all windows

Using XP Explorer
In Local service:
select > Utilities > Escape to OS > type in Explorer and click on GO. Windows Explorer
opens

REBOOT of RADIS required


This is the case for
- Printer configuration
- CD burning for copying log files
- Calibration data (backup)
- IQAP DB
- Copying 14 bit files

BIOS Update for Radis PC Celsius 400/460 (green housing)


• Switch off the RADIS PC.
• Insert the previously generated "BIOS update" floppy.
• Switch on the RADIS PC.
• Wait until the update is completed (acoustical signal: "toot-a-toot"), and remove the
floppy.
• Before you switch off, insert the RADIS OS CD and continue with the next steps.
If there is a problem with the boot order:

When starting up the PC, press F12 as mentioned on the screen.

In the menu, the "Diskette" drive has to be first.

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 42 of 48 © Siemens, 2006


07.06 CS PS AX RF
Remaining Work Steps and Correction of Errors 43

Saving 14-Bit Images


In some cases, it is necessary to save the retired image(s).
There is an additional package in Local Service under Backup beginning with VB10.
• Preparation:
Open the Explorer window (Tab+Del.+ “+” ) and go to C:\syngo\CDR_OFFLINE and
delete all images.
Go to C:\radis\Exported14bitImages and delete all images.
• In the Patient browser, select the patient image.
• Select >Transfer > Export to Offline.
• Go to Option > Local Service > Backup&Restore.
• Under "Command", select "Backup" and under Drives, select "CD-R".
• In "Packages", select the package "Exported 14Bit Images".
• Insert a blank CD und press "GO".
• Wait until it is completed and remove the CD.
• Select "Home".
There is one final step that the user must perform: he/she MUST manually delete ALL the
files from the Exported 14-bit Images once the backup operation has been successfully
completed.
• Select "Utilities"
• Under "Source", select "Escape to OS" and type in EXPLORER under "Parameter"
• Go to "C:\Radis\Exported14bitImages”
• Delete all files in this directory.
• Close all windows.

Low Level Formatting after a Failed Installation


The reason for a failed installation could be that fragments from the old installation were not
completely deleted. Following a failed installation, a low level formatting of both hard drives
should be performed prior to making a second attempt. To do this:
• During the reboot, press the "CTRL + A" keys.
• Confirm the selection.
• Under "Options", select "SCSI Disk Utilities" with the arrow keys.
• Confirm the selection with Enter.
• Select the 1st hard drive (SCSI ID #0) in the table.
• Confirm the "Format Disk" selection with Enter.
• Confirm the safety inquiry with "Yes".
• Repeat the procedure for the 2nd hard drive (SCSI ID #1) and then exit the Adaptec
tool.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 43 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
44 Remaining Work Steps and Correction of Errors
6-
Important Settings for Monitoring Sites only!
6Remaining Work Steps and Correction of Errors

• Installation of auto report settings (with System Management only)


Prerequisites: The System Management has to be activated (see above).
- Take the “SIEMENS Remote Service Aristos RADIS VB20B disk and insert it into the
Radis PC
- Go to Options > Service > Local Service > Backup & Restore > Restore, select the
drive with the “RadisVB20ProductMonitoring.ar” file and select this file under
Archive.
- Under Groups, select all (Auto Start and Auto Report).
- Select “Go” and wait for “Ready”
- Remove the disk.
- Select “Home” > “Configuration” > > Mail >
- Under “Local Name”, type in a Logical Name, e.g. “RdiagServer”.
- Under “Mail Server Host Selection”, select the Mail Server lux09505 already config-
ured (for Europe) sgpt806x (for Asia) srsacc1 (for America)
- Under E-mail address”, type in < Aristos-RemoteDiag.MED@siemens.com >
- Click on “Save”.
- Select “Home” > “Auto Report”
- Under Report, select “RADISLogReport” and check whether the checkmark is set for
“Active:” and the following settings should appear.

Fig. 37:

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 44 of 48 © Siemens, 2006


07.06 CS PS AX RF
Remaining Work Steps and Correction of Errors 45

- The Auto Report for Monitoring is now configured.

© Siemens, 2006 AXB4-320.816.06.02.02 Page 45 of 48 AXIOM Aristos Image System


07.06 CS PS AX RF
46 Remaining Work Steps and Correction of Errors

Reporting Errors by the Customer or CSE 6.1

• The customer can easily report problems with the system online.
Please show this to the customer!
- Under “Help”, select “Report Error”

Fig. 38:
The following Report window appears.

Fig. 39:
- Please enter the problem here and the name of a contact person and confirm this with
“OK”. The reported information will now be saved in C:\RADIS\Log.
• If there is no SRS, the CSE has to create an ”Error Report" once a week as described
below and to save the log file from C:\Radis\Log and send it to mail to:
td.hsc21@med.siemens.de

AXIOM Aristos Image System AXB4-320.816.06.02.02 Page 46 of 48 © Siemens, 2006


07.06 CS PS AX RF
Changes to Previous Version 47
7Changes to Previous Version 7-
All chapters revised.