Product Data Sheet

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Product Data Sheet

www.amcor.com
PerfecForm® 35850 - W
Description: Ionomer-based Thermoforming Film
Film Composition Component Materials

Exterior Side

EVA
Ionomer
EVA

Sealant Side

Layer Construction
EVA / Ionomer / EVA (50% Ionomer)

Typical Applications
- Thermoforming film for protective packaging of medical products
- Ideal for devices such as syringes, gowns, drapes, OR towels, tubing sets, catheters
- Excellent formability for contoured tooling
- Wide thermoforming window

Material Attributes
- Excellent mechanical strength properties (puncture, tear, abrasion)
- Offers cost advantages through downgauging
- Compatible with electron beam and gamma radiation
- Good clarity

Machinery
- Horizontal Form, Fill, and Seal (HFFS)
Technical Information
PerfecForm® 35850 - W
Typical
Values Units Test Methods
Physical Properties
Basis Weight: 86.4 lbs./3000 sq.ft. TM #3001.00 / ASTM D-4321
140.6 g/m2
Gauge: 6.0 mils TM #3306.00 / ASTM D-2103
152 microns
Yield: 5000 in2/lb TM #3001.00 / ASTM D-4321
7 m2/kg
Haze: 12 % ASTM D-1003

Mechanical Properties
Tensile Strength MD: 22 lb.f./in. ASTM D-882

97.9 N/25.4mm

Tensile Strength TD: 17 lb.f./in. ASTM D-882

75.6 N/25.4mm

Elongation MD: 450 % ASTM D-882

Elongation TD: 600 % ASTM D-882

Elmendorf Tear MD: 200 grams ASTM D-1922

Elmendorf Tear TD: 200 grams ASTM D-1922

Slow Impact 23 Newtons ASTM F-1306

Permeation Properties
MVTR: <1.0 gms H2O/100 sq in/24 hours ASTM F-1249
(Moisture)
<16 gms H2O/SM/24 hours
Testing Conditions: 100 F, 90% RH (37.8 C, 90% RH)

OTR: <350 cc/100 sq in/24 hours ASTM D-3985


(Oxygen)
<5425 cc/SM/24 hours

Testing Conditions: 73 F, 0% RH (23C 0%)

This document is in compliance with IoPP Medical Packaging Technical Committee Guidelines for Data Sheets

Recommended Storage Conditions: Bemis Healthcare Packaging recommends all material be stored in a clean dry environment
and away from any direct heat. Storage conditions should be between 65F (18C) and 75F (24C) at 40 to 55% relative humidity.
Keep in original packaging until ready for use.

Revision Level: 2 Revision Date: 6/24/2005

IMPORTANT NOTICE TO PURCHASER: The Typical Values documented were determined by statistical analyses of limited
sample size. The Typical Data is not meant for customer specifications and is meant for comparative purposes only.
Customers who establish specifications based on Typical Data assume all liability in connection with the establishment of such
specifications. Customers are solely responsible for determining the suitability of this product for their intended use.