CALIBRATION CERTIFICATE OF DIGITAL EARTH RESISTANCE TESTER
LAB:- 8100143376, E-mail: measuretechno@yahoo.co.in
cass
[CALIBRATION CERTIFICATE NO:MTL/ ET / DERT/RO1/ 03-23,
ULR -€C254523000000100F Page: 1 of
1.0 Service Request No.: Verbal
1.1 Issued to: M/s. Measure Techno Lab,
1.2 Description &
2, BLT. Road (Jayant Cinema Complex),
Barrackpore, Kolkata - 700120, W.B.
Digital Earth Resistance Tester b) CodeNo.: MTL /DERT/ 01
Identification of item 63034 d)Make: = Waco
to be Calibrated: €)Model /Type: N Range: —0to10N_-&
Oto 10009
g) Sensor: NA. h) Resolutior
4) End User: Electro Technical Lab §) Accuracy:
k) Calibration done at: On Site / V In House
1.3 Date of receipt of item : 01-03-23, 1.4 Physical Condition of DUC: OK
1.5 Date of calibration : 01-03-23 1.6 Recommended date of next calibration: 01-03-24
1.7 Date of Issue : 01-03-23
1.8 Environmental Conditions During Calibration: Temperature: 25°C#4°C
Humidity: 30 % RH to 75 % RH
1.9 Method of Calibration: MTL /ET /05 &MTL/ET /CAL/01
2.0 Traceability :
a)Standards used for calibration are traceable to National standards through NABL. Accredited Laboratory.
'b)The following standards / Equipment have been used.
1 Preckion Decade Resistance Box Cal. Cartels No 2158 (2022-2023) (EAL, ne}, (Cal. Date: 28/1722, Ove Cat 2812723)
2.4 Result :
Electrotechnical Calibration
SL] Parameter7 | Nominal Value Std Value Error _ | Measurement Expanded
No. Range on auc a a Uncertainty
a 20
1 | Resistance 045 Os or ‘on104
099 1 0.01 0.0062
1.98 2 0.02 0.0062
3.96 4 0.04 0.0074
ote reo 591 6 -0.09 0.0090
779 8 021 0.0109
9.76 10 0.24 0.0129
201 200 1 06218
402 400 2 0.7394
0to 1000.0 603 600 3 0.9019
805 800 5 1.0894
907 900 z 1.1889
Remarks: f) This result has an expanded uncertainty with a coverage factor k=2 at approximately 95% confidence level.
i) The calibration certificate issued for this instrument isto be used for scientific or industrial purposes only.
lil) Error = DUC Reading - Standard Reading.
DUC-Device Under Calibration __—_NS.-Not Specified Std.-Standard —__NA.-Not Applicable
[Opinions and interpretations
calibrated ZT [Accepted [Val for use
[Limited Use Rejected / Outof use
Calibration Engineer