Professional Documents
Culture Documents
HM 8018
HM 8018
HM 8018
HM8018
}
% Value Calculation from measurement
L: 0.01 μH – 9999 H
Deviation value and reference value stored
Q: 0.0001 – 99.9
% Offset
D: 0.0001 – 9.9999
Θ: -180.00° – +180.00°
Miscellaneous
Basic accuracy: 0.2 %
The inputs are short-circuit-proof and overvoltage protected up to 100 VDC
Measuring frequencies: 100 Hz, 120 Hz, 1 kHz, 10 kHz, 25 kHz
with a maximum energy consumption of 1 J.
Freq. Accuracy: ± 100 ppm
One configuration can be saved.
(except 120 Hz: 120.2 Hz ± 100 ppm)
Operating temperature: +10 °C … 40 °C
Measuring voltage: 0.5 Veff ±10 % (unloaded)
Max. relative humidity: 80 %
Measuring rate: 2 measurements/second
Power supply
Range changing: automatic, manual
(from mainframe): +5 V/300 mA
DC Bias voltage: 1 V ±10 %
+5.2 V/50 mA
Zero setting: Open/short circuit compensation
-5.2 V/50 mA
Compensation limits: Short: R ‹ 10 Ω
(∑ = 2 W)
Z ‹ 15 Ω
Dimensions (W x H x D) (without 22-pole flat plug):
Open: Z › 10 kΩ
135 x 68 x 228 mm
Measurement accuracy
Weight: approx. 0.5 g
with D ‹ 0.1 or Q ›10: C: Ae = Af (1+Cx /Cmax + Cmin /Cx)
L: Ae = Af (1+Lx /Lmax + Lmin /Lx)
Z: Ae = Af (1+Zx /Zmax + Zmin /Zx)
R: Ae = Af (1+Rx /Rmax + Rmin /Rx)
with D≥0.1:
with the parameters: Cx = Measurement value
Af = 0.2 % at f = 100 Hz, 120 Hz, 1 kHz
Af = 0.3 % at f = 10 kHz
Af = 0.5 % at f = 25 kHz
w w w. h a m e g . co m
HM8018E/180107/ce · Subject to alterations · © HAMEG Instruments GmbH · ® Registered Trademark · DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 800 0 · Fax +49 (0) 6182 800 100 · www.hameg.com · info@hameg.com
A Rohde & Schwarz Company