Professional Documents
Culture Documents
Albane Fery - Lab2
Albane Fery - Lab2
Introduction
The purpose of this experiment is to use varying concentrations of NaCl to separate plastics based on their
densities.
By changing the amount of NaCl dissolved in solution, it is possible to change the density of the solution.
Therefore, by selecting a solution density that lies between that of the plastics, we are able to separate the
samples (one floats while the other, denser sample, sinks).
Lab safety
The pieces of plastic do not pose a chemical hazard – but they can be sharp so are cutting hazards.
Glassware used is fragile – if glassware breaks, broken pieces can be hazardous.
Materials
Plastics:
o PS (Polystyrene)
o LDPE (Low density polyethylene)
o PET (Polyethylene terephthalate)
Distilled water
NaCl 25% stock solution
Beaker (sink-float tank)
Graduated cylinder
Forceps
Glass stirring rod.
Strainer
Laptop with Excel software to record data.
Methods
0 0.998
10 1.071
20 1.148
25 1.202
Slope – 0.008023729
R2 – 0.9948647
PS 1.047 1.040
Fig. 3. Table comparing the class averages for plastic densities vs. literature values.
Fig. 4. Table demonstrating the calculations involved in determining the target NaCl %wt to separate LDPE from PS and PET.
Fig. 5. Table demonstrating the calculations involved in determining the target NaCl %wt to separate PS from PET.
Dilution equations
V1C1 = V2C2
Stock conc. (%wt) Stock volume (mL) Target conc. (%wt) Target volume (mL)
Fig. 6. Table demonstrating the calculation involved in determining the stock volume and target volume required to separate PS from PET.
Results
Discussion
Fig. 7. Picture demonstrating the separation of LDPE from PS and PET. Fig. 8. Picture demonstrating the separation of PS from PET.
By
Whencomparing theseLDPE
separating qualitative
from results
PET and to thePS,
literature
LDPE density values
Whenand the targetPET
separating density
fromof the
PS,solutions in bothhard
PET (white,
sink float
(clear, tanks,
hard we floated
plastic) can conclude
whilst that these two
the other results are most likely
sank. accurate.
plastic) sank, In
andthe
PSfirst sink-float
floated. tank, the as
This occurred target
PET is
density of the solution
This occurred becausewas
PS 0.998 (pureare
and PET water), meaning
denser than that LDPE with
denser thethe
than density below
solution this would
density float,
of 1.14 whilst
g/mL, with a
PET and PS with densities greater than that of
water, with densities 1.040 and 1.302 respectively, water, should sink. As shown in the picture above, this is what
density of 1.3. PS, however, has a density of 1.040,
occurred, successfully
whilst LDPE separating
has a density smallerLDPE from at
than water the0.866.
two other plastics.
which In isthesmaller
secondthan
sink-float
that oftank, the target
the solution anddensity
therefore
floated.
of the solution was 1.14, lying almost exactly between the densities of PS and PET (1.040 and 1.302 respectively).
This meant that if volumes were measured out precisely, PS with a density below that of the solution should float,
whilst PET with a larger density than that of the solution should sink. As demonstrated in the picture above, this also
occurred, successfully separating PS from PET.
Although the qualitative results match with the data values and what we predicted would occur, there could still
have been issues with accuracy at multiple occasions that may have led to errors. Most importantly, the graduated
cylinder only measured liquids in 2mL increments, whereas in our calculations, target volumes for the NaCl solution
were precise to 0.1mL. This meant that when measuring out the volume of NaCl stock solution for the sink-float
tank, we had to estimate how close we thought the bottom of the meniscus was to our target value, as we could not
be accurate to 0.1mL. Therefore, it is highly unlikely that our volumes were exact, as the uncertainty of our
measurement was too large, and the NaCl solution density in the second sink-float tank may be different to what was
intended. However, despite this, the plastics separated as expected, suggesting that this error did not have a large
impact on our results.
The experiment’s primary objective was to separate the plastics based on their densities, using the sink-float tank
method. This objective was met as we were successfully able to first separate LDPE from PES and PET, and then
use a second sink-float tank to separate PS from PET.