Professional Documents
Culture Documents
Insert - Elecsys Folate IIIElecsys Folate III.07027290501.V8.En
Insert - Elecsys Folate IIIElecsys Folate III.07027290501.V8.En
Insert - Elecsys Folate IIIElecsys Folate III.07027290501.V8.En
cobas e 402
07027290190 07027290501 300
cobas e 801
For optimum performance of the assay follow the directions given in this Compound Concentration tested
document for the analyzer concerned. Refer to the appropriate operator’s
manual for analyzer‑specific assay instructions. Biotin ≤ 86.1 nmol/L or ≤ 21 ng/mL
Resuspension of the microparticles takes place automatically prior to use. Rheumatoid factors ≤ 1000 IU/mL
Place the cooled (stored at 2‑8 °C) cobas e pack on the reagent manager. IgG ≤ 1.6 g/dL
Avoid foam formation. The system automatically regulates the temperature
of the reagents and the opening/closing of the cobas e pack. IgA ≤ 0.4 g/dL
corresponds to the lowest analyte concentration which can be detected cobas e 801 and cobas e 402 analyzers
(value above the Limit of Blank with a probability of 95 %).
Repeatability Intermediate
The Limit of Quantitation is the lowest analyte concentration that can be precision
reproducibly measured with an intermediate precision CV of ≤ 30 %.
It has been determined using low concentration folate samples. Sample Mean SD CV SD CV
nmol/L nmol/L % nmol/L %
Dilution
Hemolysate samples with folate concentrations above the measuring range Hemolysate 1 291 21.1 7.3 22.7 7.8
can be diluted manually with Elecsys Folate RBC Hemolyzing Reagent Hemolysate 2 411 23.4 5.7 23.6 5.8
(ascorbic acid solution, 0.2 %). The recommended dilution is 1:2. The
concentration of the diluted sample must be ≥ 265 ng/mL or ≥ 602 nmol/L. Hemolysate 3 443 26.8 6.1 28.1 6.4
After manual dilution, multiply the results by the dilution factor 2. Hemolysate 4 792 23.6 3.0 32.0 4.0
Expected values Hemolysate 5 1255 44.7 3.6 53.1 4.2
The values shown below were measured on samples from an apparently
healthy population, using the Elecsys Folate III/RBC application. The values Method comparison
can be applied for the folate RBC application on all Elecsys and cobas e a) A comparison of the Elecsys Folate RBC assay, 05944295190 (y)
analyzers. The calculation is based on 290 sera (96 men, 194 women) from with the Elecsys Folate III/RBC application, 04476433190 (x) using
an European population. The age range was between 18 and 65 years. hemolyzed clinical samples gave the following correlations (ng/mL). Results
Pregnant or lactating women were excluded. The reference population was are given as whole blood folate (from hemolysate sample):
selected according to normal homocysteine values. The following values Number of samples measured: 187
were obtained:
Passing/Bablok9 Linear regression
Whole blood folate (from hemolysate samples)
y = 1.02x - 14.1 y = 1.00x - 12.0
N Median 2.5th‑97.5th percentile
τ = 0.869 r = 0.985
nmol/L ng/mL nmol/L ng/mL
The sample concentrations were between 151 and 551 ng/mL (343 and
Europe 290 673 296 481-1212 212-534 1251 nmol/L).
The measured hematocrit value in this study showed a range from b) A comparison of the Elecsys Folate III/RBC application,
37.1‑46.1 %. 07027290190 (cobas e 801 analyzer; y) with the Elecsys Folate RBC
assay, 05944295190 (cobas e 601 analyzer; x) gave the following
RBC folate (folate in erythrocyte fraction) correlations (ng/mL):
N Median 2.5th‑97.5th percentile Number of samples measured: 122
nmol/L ng/mL nmol/L ng/mL Passing/Bablok9 Linear regression
Europe 290 1657 730 1187-2854 523-1257 y = 0.979x - 3.06 y = 0.982x - 2.51
If pathologically low hematocrit values are considered for calculation of τ = 0.902 r = 0.984
RBC folate in the erythrocyte fraction, elevated RBC folate concentrations
may be observed. No medical conclusion should be based on the The sample concentrations were between 125 and 620 ng/mL (284 and
calculation considering hematocrit values in such cases. Instead, whole 1407 nmol/L).
blood folate results (from hemolysate samples) and suitable expected c) A comparison of the Elecsys Folate III/RBC application,
values may be used. 07027290190 (cobas e 402 analyzer; y) with the Elecsys
Each laboratory should investigate the transferability of the expected values Folate III/RBC application, 07027290190 (cobas e 801 analyzer; x)
to its own patient population and if necessary determine its own reference gave the following correlations (ng/mL):
ranges. Number of samples measured: 110
Specific performance data Passing/Bablok9 Linear regression
Representative performance data on the analyzers are given below.
Results obtained in individual laboratories may differ. y = 0.969x - 2.00 y = 0.971x - 1.17
Precision τ = 0.913 r = 0.990
Precision was determined using Elecsys reagents and hemolysate samples The sample concentrations were between 129 and 609 ng/mL (293 and
in a protocol (EP05‑A3) of the CLSI (Clinical and Laboratory Standards 1382 nmol/L).
Institute): 2 runs per day in duplicate each for 21 days (n = 84). Results are
given as whole blood folate (from hemolysate sample). The following results References
were obtained: 1 Nazki FH, Sameer AS, Ganaie BA. Folate: Metabolism, genes,
polymorphisms and the associated diseases. Gene 2014;533(1):11-20.
cobas e 801 and cobas e 402 analyzers 2 Scaglione F, Panzavolta G. Folate, folic acid and
Repeatability Intermediate 5-methyltetrahydrofolate are not the same thing. Xenobiotica
precision 2014;44(5):480-488.
Sample Mean SD CV SD CV 3 Reynolds EH. The neurology of folic acid deficiency. Handb Clin Neurol
2014;120:927-43.
ng/mL ng/mL % ng/mL %
4 Wick M, Pinggera W, Lehmann P. Clinical Aspects and Laboratory. Iron
Hemolysate 1 128 9.31 7.3 10.0 7.8 metabolism, Anemias. Springer Verlag, Wien, New York, 6th edition
Hemolysate 2 181 10.3 5.7 10.4 5.8 2011:41-42.
Hemolysate 3 195 11.8 6.1 12.4 6.4 5 Yetley EA, Pfeiffer CM, Phinney KW, et al. Biomarkers of folate status
in NHANES: a roundtable summary. Am J Clin Nutr
Hemolysate 4 349 10.4 3.0 14.1 4.0 2011;94(1):303S-312S.
Hemolysate 5 553 19.7 3.6 23.4 4.2 6 Occupational Safety and Health Standards: Bloodborne pathogens.
(29 CFR Part 1910.1030). Fed. Register.
COBAS, COBAS E and ELECSYS are trademarks of Roche. INTRALIPID is a trademark of Fresenius Kabi AB.
All other product names and trademarks are the property of their respective owners.
Additions, deletions or changes are indicated by a change bar in the margin.
© 2021, Roche Diagnostics