SIREMOBIL Compact / L
SP
© Siemens, 2008
03100497
© Siemens, 2008
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.
2008
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
Copyright
“© Siemens, 2008“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Mindit Magnetic Resonance Ltd. - China, Siemens Shang-
hai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or
Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Information on the protective conductor resistance test . . . . . . . . . . . . . . . . . . . . . . . 5
Additional information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Required measuring and auxiliary equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Required tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Prerequisite for retrofitting a Sony UP 970 or Sony UP 990 video printer to a SIREMO-
BIL Compact L with Memoskop CX (material no. 7142446) . . . . . . . . . . . . . . . . . . . 7
With a Memoskop C / C-E /C-SUB or Memoskop CX (material no. 7142446) . . . . . . . . . 9
With Multiformat Camera MS2000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Without Multiformat Camera MS2000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
With Memoskop CX (material no. 7142537). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Power cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
USB cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Function Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Safety Instructions 0
Note
¹ the product-specific safety notes in these instructions,
¹ the general safety information in TD00-000.860.01...
and
¹ the safety information in accordance with ARTD part 2.
NOTE The following measurements are based on the DIN VDE 0751,
Part 1 standard.
As of October 1, 2010, the DIN VDE 0751 standard will be
replaced with the IEC 62353 standard. The differences
between the standards will be explained in "Performance of
tests according to IEC 62353," print number
SPR2-000.860.01.xx.xx. As of that date, the guidelines in this
document must be observed.
Observe the instructions in the safety rules for installation and repair (ARTD-002.731.17
...).
The protective conductor resistance must be measured after every intervention in the sys-
tem.
However, documentation of the measured values is required only during periodic safety
checks.
If parts / components that can decisively influence the ground wire resistance are replaced
(e.g. replacement of the mains cable, replacement of the ON/OFF assembly, replacement
of multi-core connection cables, which also make the ground wire connection between
parts of the system (e.g. monitor cable or C-arm cable)), or if ground wire connections have
been repaired, then the ground wire resistance must be measured. The values must be
documented and assessed in the ground wire resistance report.
The measurement must be performed in accordance with DIN VDE 0751, Part 1 (refer to
ARTD Part 2). The protective conductor resistance for all touchable conductive parts must
be measured during the normal operating state of the system.
Make sure that control cables or data cables between the components of the system are
not mistaken for protective conductor connections.
During the measurement, the power cable and additional connection cables that also cre-
ate the protective conductor connection between system parts (e.g. monitor cable between
the basic unit and monitor trolley) must be moved section by section to detect cable breaks.
The protective conductor resistance must not exceed 0.2 ohms.
Fig. 1: Measuring circuit for measuring the protective ground wire resistance for units that are
disconnected from power, in compliance with DIN VDE 0751-1 2001-10, Fig. C2.
Pos. 1 System
Pos. 2 Application part (not available)
Pos. 3 Measurement setup (integrated into measuring device)
Additional information 0
These instructions apply only for retrofitting a video printer to a SIREMOBIL Compact / L.
Required Parts 0
• Service PC
• Service software (from the logbook or system binder)
• Cable for connecting service PC to host, e.g. 99 00 440
• Protective conductor test device, e.g. UNIMET 1100 safety tester 51 38 727
Required tools 0
Prerequisite for retrofitting a Sony UP 970 or Sony UP 990 video printer to a SIRE-
MOBIL Compact L with Memoskop CX (material no. 7142446) 0
The Memoskop CX requires the internal firmware version "46D" or higher. The retrofit is
not authorized for older firmware versions ("464" or "468" - the video signal is not properly
detected. The row structure in the screen is shifted).
The internal firmware of the Memoskop CX cannot be retrofitted higher in the field. If nec-
essary, the Memoskop CX (material no. 7142446) has to be replaced.
Since August 2006, the Memoskop CX has been delivered with internal firmware version
"46D". The inventory of the replacement part warehouse was updated.
• Pull out the power supply plug for the SIREMOBIL Compact / L.
With video printer available for 100V system voltage:
• Connect the power cable of the video printer to the T1 transformer, TI-34 and TI - 36 ter-
minals.
• Connect the protective conductor to the protective conductor terminal strip (ye/gn).
With video printer available for 120V system voltage:
• Connect the power cable of the video printer to the T1 transformer, TI-35 and TI - 36 ter-
minals.
• Connect the protective conductor to the protective conductor terminal strip (ye/gn).
With video printer available for 230V system voltage:
• Connect the power supply of the video printer to terminals X2.1 and X2.2.
• Connect the protective conductor to the protective conductor terminal strip (ye/gn).
Video cable
• Remove the 75 Ohm load resistor from the video output of the multiformat camera.
• Connect the video line and place towards the video printer compartment.
• Connect the video printer.. Make sure the video line is correctly connected (75 Ohm).
• Only for video printer Sony UP 970 or Sony UP 990:
Set the "mode" switch on the rear of the video printer to "analog".
Function Check
• Call up an x-ray image on Monitor A or monitor B.
• Trigger a photo on the operating panel of the basic unit and on the keyboard of the
visual device cart.
¹ The video printer prints a photo for each.
• Pull out the power supply plug for the SIREMOBIL Compact / L.
With video printer available for 100V system voltage:
• Connect the power cable of the video printer to the T1 transformer, TI-34 and TI - 36 ter-
minals.
• Connect the protective conductor to the protective conductor terminal strip (ye/gn).
With video printer available for 120V system voltage:
• Connect the power cable of the video printer to the T1 transformer, TI - 35 and TI - 36
terminals.
• Connect the protective conductor to the protective conductor terminal strip (ye/gn).
With video printer available for 230V system voltage:
• Connect the power supply of the video printer to terminals X2.1 and X2.2
• Connect the protective conductor to the protective conductor terminal strip (ye/gn).
Video cable
• Connect the video cable to the Memoskop "Ou3" plug.
• Lay the video line toward the video printer compartment.
• Connect the video printer.. Make sure the video line is correctly connected (75 Ohm).
• Only for video printer Sony UP 970 or Sony UP 990:
Set the "mode" switch on the rear of the video printer to "analog".
Settings
For Memoskop C (with Keyboard)
• Switch the system on.
• Using the keyboard of the memory, simultaneously press [CTRL] and [U].
¹ This calls the "User Setup" of the Memoskop.
• Call the submenu, "C Documentation".
• For SIREMOBIL Compact/ L with a monitor, configure monitor A.
• For SIREMOBIL Compact/ L with 2 monitors, configure monitor B.
• With an available surface dosing product measuring unit (Diamentor), also call the sub-
menu "K Documentation of the dose" and configure "Yes".
• Close the "User menu"
• After shutting down and switching the SIREMOBIL Compact/ L back on, the new config-
uration is valid.
Function Check
• Call up an x-ray image on Monitor A or monitor B.
• Trigger a photo on the operating panel of the basic unit and on the keyboard of the
visual device cart.
¹ The video printer prints a photo for each.
NOTE The installation kit for the video printer includes a control
cable with a 3.5 mm jack plug. This control cable is not used
with the existing Memoskop CX (Material No. 7142537).
Power cable 0
• Pull out the power supply plug for the SIREMOBIL Compact L.
• Connect the video printer power cable (included in the "printer installation kit" delivery
volume) at terminals X2.1 and X2.2 (on the ON/OFF assembly).
• Connect the protective conductor to the protective conductor terminal strip (green and
yellow).
• Route the power cable toward the video printer compartment and connect it to the video
printer.
• Switch on the video printer.
USB cable 0
• Connect the USB cable (included in the video printer delivery volume) at the Memoskop
“USB1” connector.
• Route the USB cable toward the video printer compartment and connect it to the video
printer.
• Set the "Mode" switch at the back of the video printer to "digital".
Settings 0
• Connect the Service PS2 keyboard provided to the Memoskop “PC Keyboard” connec-
tor.
NOTE The PS2 keyboard is kept in the compartment for the system
binder at the back of the monitor cart.
Once all settings have been programmed, the PS2 keyboard
is disconnected at stored back in the cart.
• Enter the following string in the command line of the "Command" window:
CD %WINDIR%\SYSTEM32
Then press Return .
¹ Change to the “C:\WINDOWS\system32” directory.
• Depending on which video printer is present, enter the following string in the command
line of the "Command" window:
InstallUP970AD.bat
or
InstallUP990AD.bat
NOTE If the Sony UP 970 video printer is present, enter the string
"InstallUP970AD.bat".
If the Sony UP 990 video printer is present, enter the string
"InstallUP990AD.bat".
NOTE If the video printer is not connected (power cable and USB
cable) or is not switched on, the message appears:
"Operation could not be completed. The specified port is
unknown."
In this case, confirm the message by pressing the "Return"
key and wait for the automatic restart of the Memoskop CX.
Finally, connect the video printer (power cable and USB
cable), switch it on, and begin the installation again.
• Using the Tab key, select the Browse button and activate it by pressing theReturn key.
¹ The “Locate File” window opens (Fig. 4 / p. 16).
• Use the Tab key to select the file UPXXXAD.UPD in the large window (1/Fig. 5 / p. 16),
and then mark the file by pressing an arrow key. Then, activate the Open button by
pressing the Return key.
¹ The correct path is displayed in the “Files Needed” window (Fig. 6 / p. 17).
• If there is a dose area product measuring device, also select check boxes "Dose on
hardcopy output" and "Storage of dose on disk".
• Close the "System Configuration" sub-menu (save changes).
• Close "User Setup."
• Simultaneously press the Ctrl + T keys at the Memoskop CX keyboard and then enter
the password.
¹ The Memoskop "Technical Setup" is selected.
• Remove the backup CD-R from the system binder and insert it in the CD drive of the
Memoskop CX.
• Select the "Save Settings" sub-menu and follow the menu instructions.
• After saving the parameters, remove the CD-R from the drive and place it back in the
system binder.
• Close "Technical Setup".
Function Check 0
NOTE The customer can change the default settings of the video
printer directly on the video printer, by following the instruc-
tions in the Operator Manual (included in the delivery volume
of the video printer).