GC85
EQUIPMENT MAY BE INSTALLED IN AN KANSAS CITY, MISSOURI 64153
EXISTING ROOM OF SIMILAR FUNCTION,
REQUIREMENTS STILL NEED TO BE
CHECKED.
PROJECT
~ DRAWING INDEX ~
SIOUX FALLS VA MEDICAL CENTER INSTALLED), ADDITIONAL VISITS MAY BE
REQUIRED
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105 ARCHITECT:
A1.1 EQUIPMENT LAYOUT ROOM 146 N/A
p.N/A
A1.2 EQUIPMENT LAYOUT ROOM 164
RECEIPT OF DRAWING FINAL/REVIEW N/A
AN NOTES
E1 ELECTRICAL LAYOUT GC85 THIS SIGNATURE PRINTED OR OTHERWISE ELECTRICAL CONTRACTOR:
ROOM# A21 REPRESENTS RECEIPT OF THIS SET OF XXX
E2 ELECTRICAL LAYOUT GC85 PLANS, IT IS UNDERSTOOD THAT ANY p.XXX-XXX-XXXX
DEVIATION FROM THESE DRAWINGS, m.XXX-XXX-XXXX
ROOM# A20
DETAILS AND SPECIFICATIONS MAY .NET/.COM
E1.1 ELECTRICAL DETAILS ENCROACH UPON THE EQUIPMENT
E1.3 MAIN PANEL DETAILS OPERATION, SERVICEABILITY, OR SAFETY GENERAL CONTRACTOR:
S1 OVERHEAD STRUCTURAL GUIDELINES. ALSO I AM AWARE THAT ANY N/A
LAYOUT ROOM #A21 CHANGES MADE AFTER 16 June 2017 p.XXX-XXX-XXXX
S2 OVERHEAD STRUCTURAL COULD RESULT IN ADDITIONAL EXPENSES m.XXX-XXX-XXXX
BEING INCURRED IF RADSOURCE IS NOT N/A
LAYOUT ROOM #A20
NOTIFIED IN WRITING i.e. E-MAIL ETC...
S1.1 STRUCTURAL NOTES GENERAL NEW SHEETS WILL BE ISSUED AS
CONNECTION DETAILS PROJECT MANAGER FOR CUSTOMER:
ADDITIONS AND REVISIONS WILL HAVE
KERRY NILSON
Q1 EQUIPMENT DETAILS GC85 CURRENT DATES ADDED. p.XXX
QE1 EQUIPMENT CABLE LENGTH GC85 KERRY.NILSON@VA.GOV
NAME:_KERRY NILSONDATE:_______
IT MANAGER
PRINTED:________________________________ TROY MERGEN
p. XXX
PROJECT MANAGER: SHAWN SALANDER COMPANY:______________________________ TROY.MERGEN@VA.GOV
5257 27TH ST S. SUITE 101
RADIOLOGY MANAGER
FARGO,NORTH DAKOTA ERIN HEWER
NAME:_TROY MERGENDATE:_______
58104 p.XXX
PRINTED:________________________________ ERIN.HEWER@VA.GOV
PRINTED:________________________________
MAKE ARRANGEMENTS FOR ANY RIGGING, SPECIAL HANDLING, OR FACILITY COMPANY:______________________________ SCALE: NOTED
MODIFICATIONS THAT MUST BE MADE IN ORDER FOR THE EQUIPMENT TO BE
DATE:
DELIVERED TO THE INSTALLATION SITE. IF DESIRED, YOUR LOCAL RADSOURCE 6/16/2017 10:43:52 AM
TEAM REPRESENTATIVE CAN SUPPLY A REFERENCE LIST OF RIGGERS.
MAKE SURE A DUST FREE, TEMPERATURE AND HUMIDITY CONTROLLED
FILE NAME:
ENVIRONMENT IS AVAILABLE FOR STORING THE EQUIPMENT IF YOUR SITE Sioux Falls F1 06-16-2017.dwg
IS NOT READY FOR INSTALLATION AT THE TIME OF DELIVERY. ONCE THE
SITE IS PREPARED, YOU ARE THEN RESPONSIBLE FOR DELIVERING THE DRAWING SHEET #
EQUIPMENT TO THE SITE.
SUPPLEMENTAL NOTES:
ALL ITEMS LISTED BELOW ARE TO BE PROVIDED SAMSUNG GC85
/ DESIGNED BY THE ARCHITECT N/A. THE
4'-2" 12'-3" ARCHITECT N/A SHALL CONSULT WITH SIOUX SIOUX FALLS VA MEDICAL
2'-0"
FALLS VA MEDICAL CENTER FOR ANY CENTER
ADDITIONAL INFORMATION.
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
3'-
2 N/A
0
DISTRIBUTED AROUND THE ROOM AND
" 1
p.N/A
SHOULD BE PROVIDED WITH DIMMER N/A
1143"
10'-1"
SUPPLEMENTAL NOTES:
ALL ITEMS LISTED BELOW ARE TO BE PROVIDED SAMSUNG GC85
11'-2" 6'-4" / DESIGNED BY THE ARCHITECT N/A. THE
2'-0"
22'-11"
1
"
2 ARCHITECT:
FIXTURES. THE LED FIXTURES ARE TO BE
0
3'-
N/A
DISTRIBUTED AROUND THE ROOM AND
p.N/A
SHOULD BE PROVIDED WITH DIMMER N/A
3'-821" CONTROLS. THE DIMMER CONTROLS
SHOULD BE LOCATED NEAR THE ELECTRICAL CONTRACTOR:
XXX
ENTRANCES AND IN THE CONTROL AREAS. p.XXX-XXX-XXXX
6" m.XXX-XXX-XXXX
6" · IT IS RECOMMENDED THAT ALL DIAGNOSTIC .NET/.COM
ROOMS SHOULD HAVE TILE OR VINYL
FLOORING. GENERAL CONTRACTOR:
N/A
p.XXX-XXX-XXXX
4 2 · A TILE DROP CEILING IS RECOMMENDED
4'-8"
m.XXX-XXX-XXXX
FOR ANY DIAGNOSTIC ROOMS. MINIMUM N/A
CEILING HEIGHT MIN 9'-4" / MAX 9'-6.0"
PROJECT MANAGER FOR CUSTOMER:
KERRY NILSON
· ALL CABINETRY WORK IS TO BE DESIGNED
p.XXX
BY THE ARCHITECT AND PROVIDED BY
12'-6"
KERRY.NILSON@VA.GOV
SIOUX FALLS VA MEDICAL CENTER AND OR
11'-4"
N/A . RECOMMEND ALL COUNTERS SHALL IT
TROY MERGEN
HAVE ROUNDED CORNERS. p. XXX
1 · PROVIDE FOR REFUSE REMOVAL AND
TROY.MERGEN@VA.GOV
FILE NAME:
Sioux Falls F1 06-16-2017.dwg
DRAWING SHEET #
AN
ELECTRICAL NOTES
ELECTRICAL LEGEND
MAIN DISCONNECT CONTRACTOR SUPPLIED, PART # ALL WIRES SPECIFIED SHALL BE COPPER STRANDED,
( NEA-43080-4W-ST24DC-SPD-WLC-FC) SEE SHEET (E1.3) CAN BE FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT
CBE1
OBTAINED FROM RADSOURCE IMAGING, CONTACT GENE LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR
NOWACZYK AT 816-223-6378 STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER, GENE NOWACZYK
480VAC @ 80 AMP FRAME, UL LISTED WITH MAGNETIC 4 SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR
MEP
SITE PLANNING &
CONTACTOR, SEMI FLUSH MOUNTED IN X-RAY EXAM ROOM. DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING PROJECT DEVELOPMENT
FEED THREE (3) CONDUCTORS #___ AWG ,ONE (1) OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS
NEUTRAL#___AWG AND ONE (1) #1__ GROUND TO "XG" FROM MUST BE CONTINUOUS COPPER STRANDED AND FREE OFFICE: 816-741-5558
BREAKER PER N.E.C. FROM SPLICES. MIN 2 PULL STRINGS PER CONDUIT RUN. CELL: 816-223-6378
FAX: 816-587-2423
INSTALL A 5.0' TAIL IN METAL FLEX ON SPLIT COVER LEAVING 2.0'
EMAIL: GNOWACZYK@RADSOURCE.NET
OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER
8121 N.W. 97TH TERR
CONTRACTOR TO SUPPLY AND INSTALL MAIN DISCONNECT SIZES MAY BE REQUIRED BY LOCAL CODES.
KANSAS CITY, MISSOURI 64153
LISTED ALONG WITH FLEX AND CONNECTORS AS REQUIRED,
TERMINATE FLEX WITH A 90DEG BOTH ENDS. ST1 2
IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED,
AS REQUIRED IN ACCORDANCE WITH NATIONAL AND PROJECT
SEE SHEET E-3 FOR DETAILS
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT BOTTOM
OC LOCAL ELECTRICAL CODES.
UB OF BOX TO FINISHED FLOOR OR AT MIN DIRECTLY ABOVE COVE LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO
BASE THE SYSTEM CONTROL, THE POWER DISTRIBUTION UNIT SAMSUNG GC85
1 @ COVER AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT, HOSPITAL APPROVED OUTLET OR EQUIVALENT. SIOUX FALLS VA MEDICAL
INSTALLED BY CONTRACTOR (CENTER IN COVER)
CENTER
1 @ 6 x 6 x 4 JUNCTION BOX, PART # (JB060604) MOUNT CENTER GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED.
CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT
OC OF BOX TO 18.0" A.F.F.
1 @ COVER FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
8'-11"
6'-634" p.N/A
INSTALL A 5.0' TAIL IN METAL FLEX ON ON SMALL SIDE OF SPLIT COVER ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST N/A
LEAVING 2.0' OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL
CONTRACTOR TO SUPPLY AND INSTALL JUNCTION BOX, FLEX AND
2 7 RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE
NEED FOR GREATER THAN STANDARD CABLE LENGTHS ELECTRICAL CONTRACTOR:
CONNECTORS AS REQUIRED, TERMINATE FLEX WITH A 90DEG BOTH ENDS.
E1.1 FB (REFER TO THE INTERCONNECTION DIAGRAM FOR XXX
MAXIMUM USABLE LENGTHS POINT TO POINT). SHEET QE1 p.XXX-XXX-XXXX
1 @ 10 x 3 x 4 RECESSED MOUNTED WALL DUCT WITH COVERS, MOUNT
m.XXX-XXX-XXXX
10'-1141"
XT MOUNT ONTOP OF "XG" WITH SUPPLIED CONNECTORS IN KIT. SEE SHEET
(E1.3) CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH
.NET/.COM
MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND
GENERAL CONTRACTOR:
LOCAL ELECTRICAL CODES.
SINGLE GANG 2-1/2" DEEP FLUSH MOUNTED JUNCTION BOX, N/A
ST1 SUPPLY AND INSTALL MUSHROOM HEAD "EMERGENCY STOP
8 6 PHYSICAL CONNECTION OF PRIMARY POWER TO SAMSUNG
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
3'-821"
BUTTON" 60.0" A.F.F.. WIRE TO "CBE1" MAIN DISCONNECT CBE1 EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL N/A
MAGNETIC CONTACTOR PER N.E.C. SUPPLIED AND INSTALLED BY
CONTRACTOR XXX WITH THE SUPERVISION OF A
ELECTRICAL CONTRACTOR XG XT RADSOURCE IMAGING REPRESENTATIVE. THE
PROJECT MANAGER FOR CUSTOMER:
1'-421" 1'-4" RADSOURCE IMAGING REPRESENTATIVE WOULD BE
WARNING LIGHT (X-RAY ON), LED LIGHT FIXTURE SUPPLIED AND KERRY NILSON
WL INSTALLED BY CONTRACTOR REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION p.XXX
LOCATION, AND INSURE PROPER HANDLING OF SAMSUNG KERRY.NILSON@VA.GOV
6'-441" EQUIPMENT.
1 @ 120 VAC RECEPTACLE MOUNTED ABOVE FINISHED CEILING, IT
ELECTRICAL LAYOUT SAMSUNG GC85
ACR PLACE TAG IN CEILING GRID DIRECTLY BELOW INDICATING
ABOVE CEILING OUTLET. SCALE: 3/8" = 1'-0"
ALL ELECTRICAL CONVEINCE, NETWORKIING AND PHONE
OUTLETS TO BE 18.0' A.F.F. UNLESS OTHER WISE NOTED
TROY MERGEN
p. XXX
MINIMUM CEILING HEIGHT: MIN 9'-4" / MAX 9'-6.0" TROY.MERGEN@VA.GOV
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT TOP OF FEILD VERIFY EXACT LOCATIONS OF EXISTING ELECTRICAL
FB BOX FLUSH TO FINISHED FLOOR
1 @ COVER
PANELS, CONDUITS, JUNCTION BOXES ETC FOR NEW
INFECTIOUS DISEASE
ERIN HEWER
CONNECTIONS. p.XXX
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT, ERIN.HEWER@VA.GOV
INSTALLED BY CONTRACTOR (CENTER IN COVER) REQUIRED CONDUIT RUNS FOR BASE SYSTEM BY CONTRACTOR LENGTH OF FLEXIBLE CONDUIT SHALL NOT EXCEED 6FT
JB060604-FC 6.0" X 6.0" X 3.5" JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL 3 x 10 HORIZONTAL WALL DUCT DRAWING SHEET #
INCOMING
8 TO CBE1 CONDUIT AS REQUIRED
E1
JB060604-FB 6.0" x 6.0" x 3.5" WELDED JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL MAINS
3 x 10 FLUSH TO FINISHED FLOOR DUCT
ELECTRICAL NOTES
ELECTRICAL LEGEND
MAIN DISCONNECT CONTRACTOR SUPPLIED, PART # ALL WIRES SPECIFIED SHALL BE COPPER STRANDED,
( NEA-43080-4W-ST24DC-SPD-WLC-FC) SEE SHEET (E1.3) CAN BE FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT
CBE1
OBTAINED FROM RADSOURCE IMAGING, CONTACT GENE LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR
NOWACZYK AT 816-223-6378 4 STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER, GENE NOWACZYK
MEP
480VAC @ 80 AMP FRAME, UL LISTED WITH MAGNETIC SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR SITE PLANNING &
CONTACTOR, SEMI FLUSH MOUNTED IN X-RAY EXAM ROOM. DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING PROJECT DEVELOPMENT
FEED THREE (3) CONDUCTORS #___ AWG ,ONE (1) OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS
NEUTRAL#___AWG AND ONE (1) #1__ GROUND TO "XG" FROM MUST BE CONTINUOUS COPPER STRANDED AND FREE OFFICE: 816-741-5558
BREAKER PER N.E.C. FROM SPLICES. MIN 2 PULL STRINGS PER CONDUIT RUN. CELL: 816-223-6378
FAX: 816-587-2423
INSTALL A 5.0' TAIL IN METAL FLEX ON SPLIT COVER LEAVING 2.0'
EMAIL: GNOWACZYK@RADSOURCE.NET
OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER
8121 N.W. 97TH TERR
CONTRACTOR TO SUPPLY AND INSTALL MAIN DISCONNECT SIZES MAY BE REQUIRED BY LOCAL CODES.
KANSAS CITY, MISSOURI 64153
LISTED ALONG WITH FLEX AND CONNECTORS AS REQUIRED,
TERMINATE FLEX WITH A 90DEG BOTH ENDS. ST1 IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED,
SEE SHEET E-3 FOR DETAILS 2 AS REQUIRED IN ACCORDANCE WITH NATIONAL AND PROJECT
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT BOTTOM OC LOCAL ELECTRICAL CODES.
UB OF BOX TO FINISHED FLOOR OR AT MIN DIRECTLY ABOVE COVE LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO
BASE THE SYSTEM CONTROL, THE POWER DISTRIBUTION UNIT SAMSUNG GC85
1 @ COVER AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT, HOSPITAL APPROVED OUTLET OR EQUIVALENT. SIOUX FALLS VA MEDICAL
INSTALLED BY CONTRACTOR (CENTER IN COVER)
CENTER
1 @ 6 x 6 x 4 JUNCTION BOX, PART # (JB060604) MOUNT CENTER GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED.
CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT
OC OF BOX TO 18.0" A.F.F.
UB
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
1 @ COVER UB 3 FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO
CEILING MOUNTING COMPONENTS AND WIRING IF HIGH
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,
INSTALLED BY CONTRACTOR (CENTER IN COVER)
1 WATTAGE BULBS ARE USED. RECOMMEND LED BULBS,
E1.1 6'-5"
USE DIMMER CONTROLS (EXCEPT MR). DO NOT MOUNT
1 @ 14.5 x 12 x 4 RECESSED MOUNTED JUNCTION BOX WITH SPLIT LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING ARCHITECT:
XG COVERAND DIVIDER, PART # (JB14.51204-SC-SP-DV) CONNECT TO "CBE1"
PER N.E.C. MOUNT CENTER 18.0" A.F.F. SEE SHEET (E1.3)
ACR MOUNTED ACCESSORIES WILL BE PARKED. N/A
p.N/A
INSTALL A 5.0' TAIL IN METAL FLEX ON ON SMALL SIDE OF SPLIT COVER ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST N/A
LEAVING 2.0' OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL 7 RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE
CONTRACTOR TO SUPPLY AND INSTALL JUNCTION BOX, FLEX AND 2 FB NEED FOR GREATER THAN STANDARD CABLE LENGTHS ELECTRICAL CONTRACTOR:
CONNECTORS AS REQUIRED, TERMINATE FLEX WITH A 90DEG BOTH ENDS. E1.1 (REFER TO THE INTERCONNECTION DIAGRAM FOR XXX
MAXIMUM USABLE LENGTHS POINT TO POINT). SHEET QE1 p.XXX-XXX-XXXX
1 @ 10 x 3 x 4 RECESSED MOUNTED WALL DUCT WITH COVERS, MOUNT
m.XXX-XXX-XXXX
XT 6'-5"
8'-241"
MOUNT ONTOP OF "XG" WITH SUPPLIED CONNECTORS IN KIT. SEE SHEET
(E1.3) .NET/.COM
CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH
MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND
GENERAL CONTRACTOR:
LOCAL ELECTRICAL CODES.
SINGLE GANG 2-1/2" DEEP FLUSH MOUNTED JUNCTION BOX, 6 N/A
ST1 SUPPLY AND INSTALL MUSHROOM HEAD "EMERGENCY STOP 8 PHYSICAL CONNECTION OF PRIMARY POWER TO SAMSUNG
p.XXX-XXX-XXXX
BUTTON" 60.0" A.F.F.. WIRE TO "CBE1" MAIN DISCONNECT CBE1 m.XXX-XXX-XXXX
EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL
3'-841"
N/A
MAGNETIC CONTACTOR PER N.E.C. SUPPLIED AND INSTALLED BY
ELECTRICAL CONTRACTOR
XG XT CONTRACTOR XXX WITH THE SUPERVISION OF A
1'-421" 1'-4"
RADSOURCE IMAGING REPRESENTATIVE. THE
PROJECT MANAGER FOR CUSTOMER:
WARNING LIGHT (X-RAY ON), LED LIGHT FIXTURE SUPPLIED AND RADSOURCE IMAGING REPRESENTATIVE WOULD BE KERRY NILSON
WL INSTALLED BY CONTRACTOR REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION p.XXX
6'-541" LOCATION, AND INSURE PROPER HANDLING OF SAMSUNG KERRY.NILSON@VA.GOV
EQUIPMENT.
1 @ 120 VAC RECEPTACLE MOUNTED ABOVE FINISHED CEILING, IT
ACR PLACE TAG IN CEILING GRID DIRECTLY BELOW INDICATING ELECTRICAL LAYOUT SAMSUNG GC85 ALL ELECTRICAL CONVEINCE, NETWORKIING AND PHONE TROY MERGEN
ABOVE CEILING OUTLET. SCALE: 3/8" = 1'-0" OUTLETS TO BE 18.0' A.F.F. UNLESS OTHER WISE NOTED p. XXX
MINIMUM CEILING HEIGHT: MIN 9'-4" / MAX 9'-6.0" TROY.MERGEN@VA.GOV
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT TOP OF FEILD VERIFY EXACT LOCATIONS OF EXISTING ELECTRICAL
FB BOX FLUSH TO FINISHED FLOOR
1 @ COVER
PANELS, CONDUITS, JUNCTION BOXES ETC FOR NEW
INFECTIOUS DISEASE
ERIN HEWER
CONNECTIONS. p.XXX
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT, ERIN.HEWER@VA.GOV
INSTALLED BY CONTRACTOR (CENTER IN COVER) REQUIRED CONDUIT RUNS FOR BASE SYSTEM BY CONTRACTOR LENGTH OF FLEXIBLE CONDUIT SHALL NOT EXCEED 6FT
JB060604-FC 6.0" X 6.0" X 3.5" JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL 3 x 10 HORIZONTAL WALL DUCT DRAWING SHEET #
INCOMING
8 TO CBE1 CONDUIT AS REQUIRED
E2
JB060604-FB 6.0" x 6.0" x 3.5" WELDED JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL MAINS
3 x 10 FLUSH TO FINISHED FLOOR DUCT
3.5" 10."
6"
FINISHED CEILING
DAT. 9'-4"
GENE NOWACZYK
XT MEP
XT SITE PLANNING &
1'-0"
XG SAMSUNG GC85
3.5"
14.5" 1'-4"
1'-0"
FINISHED FLOOR
SIOUX FALLS VA MEDICAL
DAT. 0'-0" CENTER
JUNCTION BOX/DUCT DETAILS 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
SCALE: NTS
GENERAL CONTRACTOR:
N/A
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
N/A
IT
FINISHED FLOOR TROY MERGEN
DAT. 0'-0" p. XXX
TROY.MERGEN@VA.GOV
INFECTIOUS DISEASE
ERIN HEWER
2 UPRIGHT ELEVATION p.XXX
ERIN.HEWER@VA.GOV
SCALE: 3/8" = 1'-0"
REVISION: DATE:
SCALE: NOTED
DATE:
6/16/2017 10:43:52 AM
FILE NAME:
Sioux Falls F1 06-16-2017.dwg
DRAWING SHEET #
E1.1
WIRING DIAGRAM KNOCKOUTS
AA
.38 SYMBOL
CONDUIT SIZE DIAMETER
.25 DIA [10] IN MM IN MM
[6]
C 1.00 25 1.38 35 GENE NOWACZYK
5.96 A
S
/ B
S
/ E 1.50 38 2.00 51
MEP
.50 DIA N N
SITE PLANNING &
[12] F 2.00 51 2.50 64 PROJECT DEVELOPMENT
(2 HOLES) 1.85
J250S G 2.50 64 3.00 76 OFFICE: 816-741-5558
14.35 2.85 H 3.00 76 3.63 92 CELL: 816-223-6378
H150 F OR S FAX: 816-587-2423
DASHED LINES
J250 F OR S AND CJ250NS EMAIL: GNOWACZYK@RADSOURCE.NET
INDICATE CIRCUIT BREAKER TERMINAL LUG DATA
SURFACE COVER NEMA/EEMAC TYPE 1 8121 N.W. 97TH TERR
CIRCUIT WIRE SIZE KANSAS CITY, MISSOURI 64153
BREAKER AMPERE CONDUCTORS (AWG/KCMIL)
TYPE
1.25 CATALOG NO. RATING PER PHASE
INFECTIOUS DISEASE
ERIN HEWER
p.XXX
A A ERIN.HEWER@VA.GOV
Main
X-RAY
Circuit
Contractor Supplied B B Generator
480V ac, 3 Phase + Digital
C Breaker
GND. C X-RAY REVISION: DATE:
ST24DC
50/60Hz PDU
GND GND
AC 3
Phase
DC Type
2
Ethernet
In Use Sign
SPD 24VDC DRAWN BY: GENE N.
CHK'D BY: A.W.
CUSTOMER :
Imaging
Class 2 Interface
Internal Class 2
Cabinet SCALE: NOTED
Emergency 24VDC Controls 24VDC DATE:
(Dry Contact)
Power Local 6/16/2017 10:43:52 AM
Off(EPO) Indication
FILE NAME:
Sioux Falls F1 06-16-2017.dwg
MAIN DISCONNECT ( NEA-43080-4W-ST24DC-SPD-WLC-FC)
DRAWING SHEET #
E1.3
CABINET CEILING
GENE NOWACZYK
MEP
WIRING : WALL AND CEILING SITE PLANNING &
PROJECT DEVELOPMENT
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
7.85 m
KANSAS CITY, MISSOURI 64153
POWER
DISTRIBUTION PROJECT
SAMSUNG GC85
SIOUX FALLS VA MEDICAL
CENTER
CABINET 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
CABINET WORK-
STATION ARCHITECT:
N/A
p.N/A
N/A
WIRING : FLOOR
CEILING ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
.NET/.COM
GENERAL CONTRACTOR:
13.1 m N/A
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
WALL STAND EXAM ROOM N/A
TABLE
PROJECT MANAGER FOR CUSTOMER:
KERRY NILSON
p.XXX
KERRY.NILSON@VA.GOV
IT
TROY MERGEN
p. XXX
TROY.MERGEN@VA.GOV
INFECTIOUS DISEASE
ERIN HEWER
p.XXX
POWER
DISTRIBUTION CABINET WALLSTAND CABINET TABLE CABINET ERIN.HEWER@VA.GOV
& CIB
DRAWING SHEET #
QE1
21.97
95.73 31.73
GENE NOWACZYK
35.98
MEP
MAX SITE PLANNING &
21.46
PROJECT DEVELOPMENT
86.44
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
67.52
10.71
MIN MAX EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
70.53
PROJECT
MIN °
90
FRONT VIEW SIDE VIEW 20° SAMSUNG GC85
11.12
SIOUX FALLS VA MEDICAL
95.58 18.98
CENTER
5.40
FRONT VIEW 31.74 SIDE VIEW
30.69
ARCHITECT:
N/A
21.94
p.N/A
N/A
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
.NET/.COM
IT
TROY MERGEN
p. XXX
TROY.MERGEN@VA.GOV
36.31
INFECTIOUS DISEASE
49.30
ERIN HEWER
Rotation Conter p.XXX
ERIN.HEWER@VA.GOV
29.53
118.50
42.72
42.72
17.72
DRAWN BY: GENE N.
CHK'D BY: A.W.
CUSTOMER :
FILE NAME:
DRAWING SHEET #
Q1
SPANNING MEMBERS: GENERAL STRUCTURAL NOTES
1 STEEL CHANNEL STRUT FLUSH WITH FINISHED
FLOOR SLABS ON WHICH EQUIPMENT IS TO BE INSTALLED MUST
CEILING. SPACED AS INDICATED ON SHEET "S1"
BE LEVEL TO 1MM IN 10.0'.
RECOMMENDED FABRICATION DETAILS ON SHEET
3'-10" "S2" LOADS SHOULD BE CALCULATED FROM SHEET DIMENSIONS ARE TO FINISHED SURFACES OF ROOM.
G1-1 (EQUIPMENT LEGEND ITEMS 4 & 5) GENE NOWACZYK
MEP
CUSTOMERS CONTRACTOR N/A MUST PROVIDE ALL SITE PLANNING &
AREA OF EXAM ROOM TO BE LEVELED FOR X-RAY
2 EQUIPMENT, CONTRACTOR TO VERIFY MINIMUM
PENETRATIONS IN POST TENSION FLOORS. PROJECT DEVELOPMENT
TABLE MOUNTING BOLT PATTERN AND LOCATION IT IS THE RESPONSIBILITY OF THE CUSTOMER SIOUX FALLS VA PROJECT
4 MEDICAL CENTER/CONTRACTOR N/A TO PERFORM ANY FLOOR
OR WALL PENETRATIONS THAT MAY BE REQUIRED. THE
CUSTOMER SIOUX FALLS VA MEDICAL CENTER/CONTRACTOR
N/A IS ALSO RESPONSIBLE FOR ENSURING THAT NO SAMSUNG GC85
SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING OR ANY
OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT WORK OR
SIOUX FALLS VA MEDICAL
STRUCTURAL SUPPORTS E.G. POST TENSION CABLES OR REBAR
WILL INTERFERE OR COME IN CONTACT WITH SUBSURFACE CENTER
PENETRATION OPERATIONS, E.G. DRILLING AND INSTALLATION
OF ANCHORING PERFORMED DURING THE INSTALLATION 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
1
5"
ARCHITECT:
2
9"
9"
MORE THEN 1/4" BELOW THE FINISHED CEILING. N/A
ALL STEEL WORK AND PARTS NECESSARY TO SUPPORT CEILING ELECTRICAL CONTRACTOR:
XXX
4
3
TO BE PROVIDED WITH SUPPORTS WHERE NECESSARY. WALL
6'-0"
A
CUSTOMER SIOUX FALLS VA MEDICAL CENTER OR THEIR N/A
721"
p.XXX-XXX-XXXX
CONTRACTOR N/A.
m.XXX-XXX-XXXX
N/A
721" 4'-141" 6'-1021" GENERAL STRUCTURAL CEILING SUPPORT REQUIREMENTS
4'-8"
UNITS. IT'S TOP MEMBER IS FASTENED TO THE FLOOR SLAB (OR BUILDING STEEL) ABOVE THE EXAM KERRY.NILSON@VA.GOV
ROOM. THE BOTTOM OF IT'S LOWEST SPANNING MEMBER IS FLUSH WITH THE FINISHED CEILING TO
WHICH THE STATIONARY CEILING TRACKS ARE FASTENED. IT
GENERAL
TROY STRUCTURAL
MERGEN NOTES
2'-2" 2'-2" 2'-2" 2'-2" 2'-2" 2'-941" THE DETAILS PROVIDED HERE INDICATE A BASIC SUPPORT METHOD THAT PERMITS TRANSVERSE AND p. XXX
LONGITUDINAL ADJUSTMENT OF BOLT CENTERS AT THE INSTALLATION SITE. IT CAN BE ADAPTED TO FLOORTROY.MERGEN@VA.GOV
SLABS ON WHICH EQUIPMENT IS TO BE INSTA
ACCOMMODATE THE VARIOUS ARCHITECTURAL CIRCUMSTANCES ENCOUNTERED AT EACH BE LEVEL TO 1MM IN 10.0'.
STRUCTURAL LAYOUT SAMSUNG GC85 INSTALLATION.
INFECTIOUS DISEASE
SCALE: 3/8" = 1'-0" ERIN HEWER
DIMENSIONS ARE TO FINISHED SURFACES OF ROOM
p.XXX
RECOMMENDED CEILING HEIGHT 9'-4.0" THE SUPPORT MUST BE RIGID, SECURELY BRACED AND LEVEL IN BOTH DIRECTIONS AT THE FINISHED ERIN.HEWER@VA.GOV
CEILING LINE. LENGTH AND LOADING CAPACITY WILL DEPEND ON THE SELECTED CEILING UNIT AND CUSTOMERS CONTRACTOR
NUMBER OF CARRIAGE ASSEMBLIES TO BE BORNE. ACTUAL EQUIPMENT LOADS FOR EACH ROOM ARE PENETRATIONS IN POST TENSION FLOORS.
SPECIFIED ON SITE PLANNING DRAWINGS UNDER THE STRUCTURAL DETAIL SPECIFICATIONS.
CUSTOMERS CONTRACTORS
IT IS SUGGESTED THAT THROUGH BOLTING THE SUPPORT STRUCTURE TO BUILDING STRUCTURE ALL HARDWARE
REVISION:FOR "THROUGH THE FLOOR" ANCHO
DATE:
UPPER SLAB CUSTOMER OR CONTRACTOR'S PREPARATION OCCUR WHENEVER POSSIBLE. APPLICABLE CONDITIONS WILL ALSO ACCEPT THE APPLICATION OF AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS
HILTI EXPANSION ANCHORS (OR EQUAL) IN LIEU OF THROUGH BOLTING. ACTUAL METHODS AND THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRIL
CEILING SUPPORT UNISTRUT (CEILING SUPPORT STRUT) MATERIALS SHALL BE DETERMINED BY THE ENGINEER OF RECORD. ALL SUPPORT MEMBERS, CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUC
ANCHOR BOLTS AND HARDWARE SHALL BE PROVIDED AND INSTALLED BY THE ENCOUNTERED BY THE INSTALLER SUCH AS REBAR E
CUSTOMER OR CONTRACTOR'S PREPARATION CUSTOMER/CONTRACTOR. THE LENGTH OF BOLTS SHALL BE DETERMINED TO MEET SPECIFIC SITE
REQUIREMENTS BY THE ENGINEER OF RECORD. IT IS THE RESPONSIBILITY OF THE CUSTOMER
(FINISHED CEILING)
CENTERLINES OF THE TWO LONGITUDINAL RAILS HOLDING THE SUSPENDED WEIGHTS ARE SHOWN. OR WALL PENETRATIONS THAT MAY BE REQUIRED. TH
CEILING FINISH STEEL CHANNEL STRUT FLUSH FINISHED CEILING SHALL BE IDENTIFIED AS THE SPANNING MEMBERS.
SAMSUNG
THESE MEMBERS MUST BE LEVEL IN BOTH DIRECTIONS TO WITHIN 1MM AND BRACED TO PREVENT
(X-RAY CEILING RAIL INSTALL) MOVEMENT IN ANY DIRECTION. THEY MUST BE SECURED TO THE SLAB ABOVE AND/OR TO THE SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING O
DRAWN BY: GENE N.
BUILDING STRUCTURE TO HOLD THE DEAD WEIGHTS INDICATED. see EQUIPMENT LEGEND "SHEET A1" OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT W
X-RAY CHK'D BY:
STRUCTURAL SUPPORTS E.G. POST TENSION
A.W. CABLES
INSTALLATION OF THE SUPPORT MAY BE AFFECTED BY OTHER ARCHITECTURAL OR MECHANICAL WILL INTERFERE
CUSTOMEROR : COME IN CONTACT WITH SUBSU
PROVISIONS WHICH MAY BE LOCATED IN THE SAME AREA. IN THIS CASE, COORDINATION OF THE PENETRATION OPERATIONS, E.G. DRILLING AND INST
REPRESENTATIVE MUST BE CONSULTED BEFORE ANY DEVIATION FROM THE SITE PLANNING OF ANCHORING PERFORMED DURING THE INSTALLAT
DRAWINGS IS UNDERTAKEN. PROCESS. TO ENSURE WORKERS SAFETY
SCALE: NOTED INSTALLER
ONLY PERFORM
DATE: SURFACE PENETRATION OPERATION
1. CUSTOMER SIOUX FALLS VA MEDICAL CENTER CAN ALSO UTILIZE AN EXISTING SUPPORT 6/16/2017CONTRACTOR
AFTER THE CUSTOMERS 10:43:52 AM HAS VALIDAT
SECTION A-A STRUCTURE, ONLY IF THE STRUCTURE IS CERTIFIED BY A SURFACE TO BE PENETRATED.
TABLE LICENSED/PROFESSIONAL STRUCTURAL ENGINEER.
FILE NAME:
ALL CEILING MOUNTED FIXTURES SUCH AS AIR VEN
2. SUPPORT STRUCTURE MUST MEET ALL REQUIRED EQUIPMENT SUPPORT NEEDS FOR WIDTH, Sioux Falls F1 06-16-2017.dwg
FLOOR FINISH DEPTH, STRENGTH, SPRINKLERS etc. TO BE FLUSH MOUNTED OR NOT T
ETC. MORE THEN 1/4" BELOW THE FINISHED CEILING.
DRAWING SHEET #
3. CUSTOMER SIOUX FALLS VA MEDICAL CENTER IS RESPONSIBLE FOR MEETING SITE CONDITIONS
S1
FOR STATIC LOADS, LEVELNESS, ETC. ALL STEEL WORK AND PARTS NECESSARY TO SUPPO
MOUNTED EQUIPMENT IS TO BE SUPPLIED BY THE CU
SIOUX FALLS VA MEDICAL CENTER
SPANNING MEMBERS: GENERAL STRUCTURAL NOTES
1 STEEL CHANNEL STRUT FLUSH WITH FINISHED
FLOOR SLABS ON WHICH EQUIPMENT IS TO BE INSTALLED MUST
CEILING. SPACED AS INDICATED ON SHEET "S1"
BE LEVEL TO 1MM IN 10.0'.
RECOMMENDED FABRICATION DETAILS ON SHEET
3'-10"
"S2" LOADS SHOULD BE CALCULATED FROM SHEET DIMENSIONS ARE TO FINISHED SURFACES OF ROOM.
G1-1 (EQUIPMENT LEGEND ITEMS 4 & 5) GENE NOWACZYK
MEP
CUSTOMERS CONTRACTOR N/A MUST PROVIDE ALL SITE PLANNING &
AREA OF EXAM ROOM TO BE LEVELED FOR X-RAY
2 EQUIPMENT, CONTRACTOR TO VERIFY MINIMUM
PENETRATIONS IN POST TENSION FLOORS. PROJECT DEVELOPMENT
TABLE MOUNTING BOLT PATTERN AND LOCATION IT IS THE RESPONSIBILITY OF THE CUSTOMER SIOUX FALLS VA PROJECT
4 MEDICAL CENTER/CONTRACTOR N/A TO PERFORM ANY FLOOR
OR WALL PENETRATIONS THAT MAY BE REQUIRED. THE
CUSTOMER SIOUX FALLS VA MEDICAL CENTER/CONTRACTOR
N/A IS ALSO RESPONSIBLE FOR ENSURING THAT NO SAMSUNG GC85
SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING OR ANY
OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT WORK OR
SIOUX FALLS VA MEDICAL
STRUCTURAL SUPPORTS E.G. POST TENSION CABLES OR REBAR
WILL INTERFERE OR COME IN CONTACT WITH SUBSURFACE CENTER
PENETRATION OPERATIONS, E.G. DRILLING AND INSTALLATION
1'-021"
1
OF ANCHORING PERFORMED DURING THE INSTALLATION 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
SURFACE TO BE PENETRATED.
2
11'-1043"
ARCHITECT:
ALL CEILING MOUNTED FIXTURES SUCH AS AIR VENTS, N/A
9"
9"
ELECTRICAL CONTRACTOR:
4
GENERAL CONTRACTOR:
A
SUPPORTS ARE TO BE SUPPLIED AND INSTALLED BY THE
721"
A WELDED OR FABRICATED STRUCTURAL SUPPORT IS REQUIRED FOR CEILING MOUNTED X-RAY p.XXX
UNITS. IT'S TOP MEMBER IS FASTENED TO THE FLOOR SLAB (OR BUILDING STEEL) ABOVE THE EXAM KERRY.NILSON@VA.GOV
ROOM. THE BOTTOM OF IT'S LOWEST SPANNING MEMBER IS FLUSH WITH THE FINISHED CEILING TO
WHICH THE STATIONARY CEILING TRACKS ARE FASTENED. IT
2'-941" 2'-2" 2'-2" 2'-2" 2'-2" 2'-2" 3'-11" GENERAL
TROY STRUCTURAL
MERGEN NOTES
THE DETAILS PROVIDED HERE INDICATE A BASIC SUPPORT METHOD THAT PERMITS TRANSVERSE AND p. XXX
LONGITUDINAL ADJUSTMENT OF BOLT CENTERS AT THE INSTALLATION SITE. IT CAN BE ADAPTED TO FLOORTROY.MERGEN@VA.GOV
SLABS ON WHICH EQUIPMENT IS TO BE INSTA
ACCOMMODATE THE VARIOUS ARCHITECTURAL CIRCUMSTANCES ENCOUNTERED AT EACH BE LEVEL TO 1MM IN 10.0'.
STRUCTURAL LAYOUT SAMSUNG GC85 INSTALLATION.
INFECTIOUS DISEASE
SCALE: 3/8" = 1'-0" ERIN HEWER
DIMENSIONS ARE TO FINISHED SURFACES OF ROOM
p.XXX
RECOMMENDED CEILING HEIGHT 9'-4.0" THE SUPPORT MUST BE RIGID, SECURELY BRACED AND LEVEL IN BOTH DIRECTIONS AT THE FINISHED ERIN.HEWER@VA.GOV
CEILING LINE. LENGTH AND LOADING CAPACITY WILL DEPEND ON THE SELECTED CEILING UNIT AND CUSTOMERS CONTRACTOR
NUMBER OF CARRIAGE ASSEMBLIES TO BE BORNE. ACTUAL EQUIPMENT LOADS FOR EACH ROOM ARE PENETRATIONS IN POST TENSION FLOORS.
SPECIFIED ON SITE PLANNING DRAWINGS UNDER THE STRUCTURAL DETAIL SPECIFICATIONS.
CUSTOMERS CONTRACTORS
IT IS SUGGESTED THAT THROUGH BOLTING THE SUPPORT STRUCTURE TO BUILDING STRUCTURE ALL HARDWARE
REVISION:FOR "THROUGH THE FLOOR" ANCHO
DATE:
UPPER SLAB CUSTOMER OR CONTRACTOR'S PREPARATION OCCUR WHENEVER POSSIBLE. APPLICABLE CONDITIONS WILL ALSO ACCEPT THE APPLICATION OF AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS
HILTI EXPANSION ANCHORS (OR EQUAL) IN LIEU OF THROUGH BOLTING. ACTUAL METHODS AND THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRIL
CEILING SUPPORT UNISTRUT (CEILING SUPPORT STRUT) MATERIALS SHALL BE DETERMINED BY THE ENGINEER OF RECORD. ALL SUPPORT MEMBERS, CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUC
ANCHOR BOLTS AND HARDWARE SHALL BE PROVIDED AND INSTALLED BY THE ENCOUNTERED BY THE INSTALLER SUCH AS REBAR E
CUSTOMER OR CONTRACTOR'S PREPARATION CUSTOMER/CONTRACTOR. THE LENGTH OF BOLTS SHALL BE DETERMINED TO MEET SPECIFIC SITE
REQUIREMENTS BY THE ENGINEER OF RECORD. IT IS THE RESPONSIBILITY OF THE CUSTOMER
(FINISHED CEILING)
CENTERLINES OF THE TWO LONGITUDINAL RAILS HOLDING THE SUSPENDED WEIGHTS ARE SHOWN. OR WALL PENETRATIONS THAT MAY BE REQUIRED. TH
CEILING FINISH STEEL CHANNEL STRUT FLUSH FINISHED CEILING SHALL BE IDENTIFIED AS THE SPANNING MEMBERS.
SAMSUNG
THESE MEMBERS MUST BE LEVEL IN BOTH DIRECTIONS TO WITHIN 1MM AND BRACED TO PREVENT
(X-RAY CEILING RAIL INSTALL) MOVEMENT IN ANY DIRECTION. THEY MUST BE SECURED TO THE SLAB ABOVE AND/OR TO THE SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING O
DRAWN BY: GENE N.
BUILDING STRUCTURE TO HOLD THE DEAD WEIGHTS INDICATED. see EQUIPMENT LEGEND "SHEET A1" OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT W
X-RAY CHK'D BY:
STRUCTURAL SUPPORTS E.G. POST TENSION
A.W. CABLES
INSTALLATION OF THE SUPPORT MAY BE AFFECTED BY OTHER ARCHITECTURAL OR MECHANICAL WILL INTERFERE
CUSTOMEROR : COME IN CONTACT WITH SUBSU
PROVISIONS WHICH MAY BE LOCATED IN THE SAME AREA. IN THIS CASE, COORDINATION OF THE PENETRATION OPERATIONS, E.G. DRILLING AND INST
REPRESENTATIVE MUST BE CONSULTED BEFORE ANY DEVIATION FROM THE SITE PLANNING OF ANCHORING PERFORMED DURING THE INSTALLAT
DRAWINGS IS UNDERTAKEN. PROCESS. TO ENSURE WORKERS SAFETY
SCALE: NOTED INSTALLER
ONLY PERFORM
DATE: SURFACE PENETRATION OPERATION
1. CUSTOMER SIOUX FALLS VA MEDICAL CENTER CAN ALSO UTILIZE AN EXISTING SUPPORT 6/16/2017CONTRACTOR
AFTER THE CUSTOMERS 10:43:52 AM HAS VALIDAT
SECTION A-A STRUCTURE, ONLY IF THE STRUCTURE IS CERTIFIED BY A SURFACE TO BE PENETRATED.
TABLE LICENSED/PROFESSIONAL STRUCTURAL ENGINEER.
FILE NAME:
ALL CEILING MOUNTED FIXTURES SUCH AS AIR VEN
2. SUPPORT STRUCTURE MUST MEET ALL REQUIRED EQUIPMENT SUPPORT NEEDS FOR WIDTH, Sioux Falls F1 06-16-2017.dwg
FLOOR FINISH DEPTH, STRENGTH, SPRINKLERS etc. TO BE FLUSH MOUNTED OR NOT T
ETC. MORE THEN 1/4" BELOW THE FINISHED CEILING.
DRAWING SHEET #
3. CUSTOMER SIOUX FALLS VA MEDICAL CENTER IS RESPONSIBLE FOR MEETING SITE CONDITIONS
S2
FOR STATIC LOADS, LEVELNESS, ETC. ALL STEEL WORK AND PARTS NECESSARY TO SUPPO
MOUNTED EQUIPMENT IS TO BE SUPPLIED BY THE CU
SIOUX FALLS VA MEDICAL CENTER
STRUT CHANNEL SUPPORT FRAME NOTES
RECOMMENDED UNIVERSAL STEEL STRUT
CHANNEL SUPPORT FRAME.
LONGITUDINAL
EQUIPMENT RAIL REQUIRED FITTING
FILE NAME:
Sioux Falls F1 06-16-2017.dwg
GENERAL CEILING DETAIL GENERAL CONNECTION DETAIL
NO SCALE NO SCALE DRAWING SHEET #
S2