CUSTOMER

ALL REQUIREMENTS FOR THIS


EQUIPMENT ARE NOT NOTED ON THIS
SHEET. IT IS SUGGESTED THAT THESE GENE NOWACZYK
DRAWINGS BE REVIEWED BY QUALIFIED MEP
SITE PLANNING &
PROFESSIONALS WHO CAN ASSIST PROJECT DEVELOPMENT
WITH MAKING DECISIONS REGARDING
RADIATION CONTAINMENT, MAGNETIC OFFICE: 816-741-5558
FIELD CONTAINMENT, ELECTRICAL, CELL: 816-223-6378

SAMSUNG STRUCTURAL AND MECHANICAL


REQUIREMENTS. ALTHOUGH THE
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR

GC85
EQUIPMENT MAY BE INSTALLED IN AN KANSAS CITY, MISSOURI 64153
EXISTING ROOM OF SIMILAR FUNCTION,
REQUIREMENTS STILL NEED TO BE
CHECKED.
PROJECT

SITE VISIT SAMSUNG GC85


NORMAL PROCEDURE IS TO ATTEND AN
INITIAL CONSTRUCTION MEETING AND SIOUX FALLS VA MEDICAL
THEN PERFORM AN INSPECTION OF CENTER
THE ELECTRICAL BUILD OUT AT OR
ABOUT THE SAME TIME AS THE 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
ELECTRICAL ROUGH IN (PRIOR TO
SHEETROCK OR LEAD BEING

~ DRAWING INDEX ~
SIOUX FALLS VA MEDICAL CENTER INSTALLED), ADDITIONAL VISITS MAY BE
REQUIRED
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105 ARCHITECT:
A1.1 EQUIPMENT LAYOUT ROOM 146 N/A
p.N/A
A1.2 EQUIPMENT LAYOUT ROOM 164
RECEIPT OF DRAWING FINAL/REVIEW N/A
AN NOTES
E1 ELECTRICAL LAYOUT GC85 THIS SIGNATURE PRINTED OR OTHERWISE ELECTRICAL CONTRACTOR:
ROOM# A21 REPRESENTS RECEIPT OF THIS SET OF XXX
E2 ELECTRICAL LAYOUT GC85 PLANS, IT IS UNDERSTOOD THAT ANY p.XXX-XXX-XXXX
DEVIATION FROM THESE DRAWINGS, m.XXX-XXX-XXXX
ROOM# A20
DETAILS AND SPECIFICATIONS MAY .NET/.COM
E1.1 ELECTRICAL DETAILS ENCROACH UPON THE EQUIPMENT
E1.3 MAIN PANEL DETAILS OPERATION, SERVICEABILITY, OR SAFETY GENERAL CONTRACTOR:
S1 OVERHEAD STRUCTURAL GUIDELINES. ALSO I AM AWARE THAT ANY N/A
LAYOUT ROOM #A21 CHANGES MADE AFTER 16 June 2017 p.XXX-XXX-XXXX
S2 OVERHEAD STRUCTURAL COULD RESULT IN ADDITIONAL EXPENSES m.XXX-XXX-XXXX
BEING INCURRED IF RADSOURCE IS NOT N/A
LAYOUT ROOM #A20
NOTIFIED IN WRITING i.e. E-MAIL ETC...
S1.1 STRUCTURAL NOTES GENERAL NEW SHEETS WILL BE ISSUED AS
CONNECTION DETAILS PROJECT MANAGER FOR CUSTOMER:
ADDITIONS AND REVISIONS WILL HAVE
KERRY NILSON
Q1 EQUIPMENT DETAILS GC85 CURRENT DATES ADDED. p.XXX
QE1 EQUIPMENT CABLE LENGTH GC85 KERRY.NILSON@VA.GOV
NAME:_KERRY NILSONDATE:_______
IT MANAGER
PRINTED:________________________________ TROY MERGEN
p. XXX
PROJECT MANAGER: SHAWN SALANDER COMPANY:______________________________ TROY.MERGEN@VA.GOV
5257 27TH ST S. SUITE 101
RADIOLOGY MANAGER
FARGO,NORTH DAKOTA ERIN HEWER
NAME:_TROY MERGENDATE:_______
58104 p.XXX
PRINTED:________________________________ ERIN.HEWER@VA.GOV

OFFICE PHONE 701-297-0305 COMPANY:______________________________


CELL PHONE 701-371-0505
E-MAIL SSALANDER@IMAGINGSOLUTIONSINC.COM NAME:ERIN HEWERDATE:_______ REVISION: DATE:

PRINTED:________________________________

- SITE PROGRESS CHECKLIST - COMPANY:______________________________

REVIEW EQUIPMENT ORDER FOR EXACT ITEMS PURCHASED. OPTIONAL/- NAME:______________________DATE:_______


FUTURE ITEMS NOT ON ORDER MAY BE INDICATED ON THESE PLANS.
PRINTED:________________________________
ALL ROOM DIMENSIONS ARE CRITICAL! IMMEDIATELY CONTACT GENE NOWACZYK
AT RADSOURCE IF CHANGES OCCUR OR DIMENSIONS ARE NOT CORRECT. COMPANY:______________________________
CONTACT A RADIATION PHYSICIST OR CONSULTANT TO SPECIFY DRAWN BY: GENE N.
REQUIREMENTS FOR RADIATION CONTAINMENT. CHK'D BY:
NAME:______________________DATE:_______ A.W.
PROVIDE A LOCKABLE EQUIPMENT HOLDING AREA CLOSE TO THE CUSTOMER :
INSTALLATION FOR STORING TOOLS AND TEST EQUIPMENT. PRINTED:________________________________

MAKE ARRANGEMENTS FOR ANY RIGGING, SPECIAL HANDLING, OR FACILITY COMPANY:______________________________ SCALE: NOTED
MODIFICATIONS THAT MUST BE MADE IN ORDER FOR THE EQUIPMENT TO BE
DATE:
DELIVERED TO THE INSTALLATION SITE. IF DESIRED, YOUR LOCAL RADSOURCE 6/16/2017 10:43:52 AM
TEAM REPRESENTATIVE CAN SUPPLY A REFERENCE LIST OF RIGGERS.
MAKE SURE A DUST FREE, TEMPERATURE AND HUMIDITY CONTROLLED
FILE NAME:
ENVIRONMENT IS AVAILABLE FOR STORING THE EQUIPMENT IF YOUR SITE Sioux Falls F1 06-16-2017.dwg
IS NOT READY FOR INSTALLATION AT THE TIME OF DELIVERY. ONCE THE
SITE IS PREPARED, YOU ARE THEN RESPONSIBLE FOR DELIVERING THE DRAWING SHEET #
EQUIPMENT TO THE SITE.

INSTALL LOCATION MAP LAST PLOT DATE: 16 June 2017


LAST REVIEW: 6/16/2017 12:23:22 PM
COVER
NOTE: NOTE:
THIS SAMSUNG GC85 ROOM CONFIGURATION ARCHITECT / CONTRACTOR IS TO VERIFY ALL
UTILIZES A STANDARD THREE (3) METER TRANSVERSE DIMENSIONS ON THIS DRAWING AGAINST AS
BRIDGE AND STANDARD FOUR (4) METER LONGITUDE BUILT. DIMENSIONS ILLUSTRATED WITH IN HAVE
RAILS OTHER VARIATIONS ARE AVAILABLE AT AN BEEN USED FOR PROPER EQUIPMENT GENE NOWACZYK
3'-10" ADDITIONAL EXPENSE BUT NOT REQUIRED. CONTACT PLACEMENT AND MAY BE USED IN THE FUTURE MEP
SITE PLANNING &
YOUR SALES PERSONAL FOR INFORMATION. FOR RADIATION SHIELDING PROTECTION BY THE PROJECT DEVELOPMENT
CUSTOMERS SELECTED PHYSICIST.
OFFICE: 816-741-5558
COUNTER TOP IN OPERATOR CONTROL AREA SHOULD CELL: 816-223-6378
BE AT 40.0"-42.0" A.F.F. FOR STANDING OPERATIONS EACH RESPECTIVE ROOM WILL HAVE AN FAX: 816-587-2423
ENLARGED VIEW ILLUSTRATING THE PLACEMENT EMAIL: GNOWACZYK@RADSOURCE.NET
OF THE RADIOLOGY EQUIPMENT FOR THE 8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
ASSOCIATED ROOM IDENTIFICATION AS SHOWN
ON SHEETS LABELED WITH THE LETTER "AQ-?".
PROJECT
10'-1"

SUPPLEMENTAL NOTES:
ALL ITEMS LISTED BELOW ARE TO BE PROVIDED SAMSUNG GC85
/ DESIGNED BY THE ARCHITECT N/A. THE
4'-2" 12'-3" ARCHITECT N/A SHALL CONSULT WITH SIOUX SIOUX FALLS VA MEDICAL

2'-0"
FALLS VA MEDICAL CENTER FOR ANY CENTER
ADDITIONAL INFORMATION.
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105

· ALL DIAGNOSTIC ROOMS INCLUDING X-RAY,


MAMMO, R/F, CT, ETC., SHALL HAVE " LED"
1'-6"

CAN LIGHTING AND/OR FLORESCENT LIGHT


22'-11"

FIXTURES. THE LED FIXTURES ARE TO BE ARCHITECT:

3'-
2 N/A

0
DISTRIBUTED AROUND THE ROOM AND

" 1
p.N/A
SHOULD BE PROVIDED WITH DIMMER N/A
1143"

3'-821" CONTROLS. THE DIMMER CONTROLS


SHOULD BE LOCATED NEAR THE ELECTRICAL CONTRACTOR:
XXX
ENTRANCES AND IN THE CONTROL AREAS.
1'-6" p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
6" 1'-443" · IT IS RECOMMENDED THAT ALL DIAGNOSTIC .NET/.COM
ROOMS SHOULD HAVE TILE OR VINYL
FLOORING. GENERAL CONTRACTOR:
N/A
p.XXX-XXX-XXXX
· A TILE DROP CEILING IS RECOMMENDED
2 4 m.XXX-XXX-XXXX
4'-8"

FOR ANY DIAGNOSTIC ROOMS. MINIMUM N/A


CEILING HEIGHT MIN 9'-4" / MAX 9'-6.0"
PROJECT MANAGER FOR CUSTOMER:
KERRY NILSON
· ALL CABINETRY WORK IS TO BE DESIGNED
p.XXX
BY THE ARCHITECT AND PROVIDED BY KERRY.NILSON@VA.GOV
SIOUX FALLS VA MEDICAL CENTER AND OR
12'-6"
11'-4"

N/A . RECOMMEND ALL COUNTERS SHALL IT


TROY MERGEN
HAVE ROUNDED CORNERS. p. XXX
1 · PROVIDE FOR REFUSE REMOVAL AND
TROY.MERGEN@VA.GOV

6'-5" DISPOSAL (E.G. CRATES, CARTONS, RADIOLOGY MANAGER


ERIN HEWER
PACKING) p.XXX
ERIN.HEWER@VA.GOV
· PROVIDE A WORKING RESTROOM WITH
SUPPLIES WITH IN THE FACILITY
4'-8"

NOTE: ENVIRONMENT REVISION: DATE:


3 16'-5"
SPECIAL DELIVERY TO BUILDING NOT
COVERED BY SAMSUNG , TO BE AMBIENT OPERATION TEMPERATURE: 55-75° (F)
COORDINATED BETWEEN "SIOUX FALLS ALLOWABLE TEMPERATURE CHANGE: 15° (F) PER HOUR
VA MEDICAL CENTER AND/OR N/A" AND HUMIDITY: 20-80 PERCENT NON CONDENSING
3"

SAMSUNG ALLOWABLE HUMIDITY CHANGE: 10 PERCENT PER HOUR

ROOM LAYOUT - RADIOLOGY A20


SCALE: 1/2" = 1'-0"
MINIMUM CEILING HEIGHT: MIN 9'-4" / MAX 9'-6.0" EQUIPMENT LEGEND SAMSUNG XGEO-GC85
OWNER: SIOUX FALLS VA MEDICAL CENTER
DRAWN BY: GENE N.
ITEM QUANTITY DESCRIPTION WEIGHT HEAT OUTPUT PROVIDED BY INSTALLED BY CHK'D BY: A.W.
1 1 PATIENT TABLE 440LBS 2,046 BTU OWNER SAMSUNG CUSTOMER :

2 1 WALL / CHEST STAND 330 LBS 511 BTU OWNER SAMSUNG


SCALE: NOTED
3 1 GENERATOR CABINET 220 LBS 11,798 BTU OWNER SAMSUNG DATE:
6/16/2017 10:43:52 AM
4 1 CEILING RAIL 140 LBS 3400 BTU OWNER SAMSUNG
FILE NAME:
5 1 CEILING MOUNTED TUBE HEAD 616 LBS 1,909 BTU OWNER SAMSUNG Sioux Falls F1 06-16-2017.dwg
0 1' 2' 3' 4' 5' 6' 7' 8' 9' 10'
6 1 WORK STATION 35 LBS 1,364 BTU OWNER SAMSUNG DRAWING SHEET #
0 1m 2m 3m
7 1 STITCHING STAND 235 LBS N/A BTU OWNER SAMSUNG
A1.2
NOTE: NOTE:
THIS SAMSUNG GC85 ROOM CONFIGURATION ARCHITECT / CONTRACTOR IS TO VERIFY ALL
UTILIZES A STANDARD THREE (3) METER TRANSVERSE DIMENSIONS ON THIS DRAWING AGAINST AS
BRIDGE AND STANDARD FOUR (4) METER LONGITUDE BUILT. DIMENSIONS ILLUSTRATED WITH IN HAVE
3'-10" RAILS OTHER VARIATIONS ARE AVAILABLE AT AN BEEN USED FOR PROPER EQUIPMENT GENE NOWACZYK
3'-10" MEP
ADDITIONAL EXPENSE BUT NOT REQUIRED. CONTACT PLACEMENT AND MAY BE USED IN THE FUTURE SITE PLANNING &
YOUR SALES PERSONAL FOR INFORMATION. FOR RADIATION SHIELDING PROTECTION BY THE PROJECT DEVELOPMENT
CUSTOMERS SELECTED PHYSICIST.
OFFICE: 816-741-5558
COUNTER TOP IN OPERATOR CONTROL AREA SHOULD CELL: 816-223-6378
BE AT 40.0"-42.0" A.F.F. FOR STANDING OPERATIONS EACH RESPECTIVE ROOM WILL HAVE AN FAX: 816-587-2423
ENLARGED VIEW ILLUSTRATING THE PLACEMENT EMAIL: GNOWACZYK@RADSOURCE.NET
OF THE RADIOLOGY EQUIPMENT FOR THE 8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
ASSOCIATED ROOM IDENTIFICATION AS SHOWN
ON SHEETS LABELED WITH THE LETTER "AQ-?".
PROJECT

10'-1"
SUPPLEMENTAL NOTES:
ALL ITEMS LISTED BELOW ARE TO BE PROVIDED SAMSUNG GC85
11'-2" 6'-4" / DESIGNED BY THE ARCHITECT N/A. THE
2'-0"

ARCHITECT N/A SHALL CONSULT WITH SIOUX SIOUX FALLS VA MEDICAL


FALLS VA MEDICAL CENTER FOR ANY CENTER
ADDITIONAL INFORMATION.
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105

· ALL DIAGNOSTIC ROOMS INCLUDING X-RAY,


MAMMO, R/F, CT, ETC., SHALL HAVE " LED"
CAN LIGHTING AND/OR FLORESCENT LIGHT

22'-11"
1
"

2 ARCHITECT:
FIXTURES. THE LED FIXTURES ARE TO BE
0
3'-

N/A
DISTRIBUTED AROUND THE ROOM AND
p.N/A
SHOULD BE PROVIDED WITH DIMMER N/A
3'-821" CONTROLS. THE DIMMER CONTROLS
SHOULD BE LOCATED NEAR THE ELECTRICAL CONTRACTOR:
XXX
ENTRANCES AND IN THE CONTROL AREAS. p.XXX-XXX-XXXX
6" m.XXX-XXX-XXXX
6" · IT IS RECOMMENDED THAT ALL DIAGNOSTIC .NET/.COM
ROOMS SHOULD HAVE TILE OR VINYL
FLOORING. GENERAL CONTRACTOR:
N/A
p.XXX-XXX-XXXX
4 2 · A TILE DROP CEILING IS RECOMMENDED
4'-8"

m.XXX-XXX-XXXX
FOR ANY DIAGNOSTIC ROOMS. MINIMUM N/A
CEILING HEIGHT MIN 9'-4" / MAX 9'-6.0"
PROJECT MANAGER FOR CUSTOMER:
KERRY NILSON
· ALL CABINETRY WORK IS TO BE DESIGNED
p.XXX
BY THE ARCHITECT AND PROVIDED BY
12'-6"

KERRY.NILSON@VA.GOV
SIOUX FALLS VA MEDICAL CENTER AND OR

11'-4"
N/A . RECOMMEND ALL COUNTERS SHALL IT
TROY MERGEN
HAVE ROUNDED CORNERS. p. XXX
1 · PROVIDE FOR REFUSE REMOVAL AND
TROY.MERGEN@VA.GOV

6'-221" DISPOSAL (E.G. CRATES, CARTONS, RADIOLOGY MANAGER


ERIN HEWER
PACKING) p.XXX
ERIN.HEWER@VA.GOV
· PROVIDE A WORKING RESTROOM WITH
SUPPLIES WITH IN THE FACILITY
4'-8"

ENVIRONMENT REVISION: DATE:


17'-6" 3 3 NOTE:
SPECIAL DELIVERY TO BUILDING NOT
3"

COVERED BY SAMSUNG , TO BE AMBIENT OPERATION TEMPERATURE: 55-75° (F)


COORDINATED BETWEEN "SIOUX FALLS ALLOWABLE TEMPERATURE CHANGE: 15° (F) PER HOUR
VA MEDICAL CENTER AND/OR N/A" AND HUMIDITY: 20-80 PERCENT NON CONDENSING
SAMSUNG ALLOWABLE HUMIDITY CHANGE: 10 PERCENT PER HOUR

ROOM LAYOUT - RADIOLOGY A21


SCALE: 1/2" = 1'-0"
MINIMUM CEILING HEIGHT: MIN 9'-4" / MAX 9'-6.0"
EQUIPMENT LEGEND SAMSUNG XGEO-GC85
OWNER: SIOUX FALLS VA MEDICAL CENTER
DRAWN BY: GENE N.
ITEM QUANTITY DESCRIPTION WEIGHT HEAT OUTPUT PROVIDED BY INSTALLED BY CHK'D BY: A.W.
1 1 PATIENT TABLE 440LBS 2,046 BTU OWNER SAMSUNG CUSTOMER :

2 1 WALL / CHEST STAND 330 LBS 511 BTU OWNER SAMSUNG


SCALE: NOTED
3 1 GENERATOR CABINET 220 LBS 11,798 BTU OWNER SAMSUNG DATE:
6/16/2017 10:43:52 AM
4 1 CEILING RAIL 140 LBS 3400 BTU OWNER SAMSUNG
FILE NAME:
5 1 CEILING MOUNTED TUBE HEAD 616 LBS 1,909 BTU OWNER SAMSUNG Sioux Falls F1 06-16-2017.dwg
0 1' 2' 3' 4' 5' 6' 7' 8' 9' 10'
6 1 WORK STATION 35 LBS 1,364 BTU OWNER SAMSUNG DRAWING SHEET #
0 1m 2m 3m
7 1 STITCHING STAND 235 LBS N/A BTU OWNER SAMSUNG
A1.2
MINIMUM SITE PREPARATION REQUIREMENTS PREFACE
NOTICE:
WALLS TO BE PAINTED OR COVERED, BASEBOARDS INSTALLED, FLOORS TO BE THIS DRAWING SET IS THE SOLE PROPERTY OF THESE DRAWINGS HAVE BEEN PREPARED BY
TITLED AND OR COVERED, CEILING SHALL HAVE GRID TILES AND LIGHTING NOWACZYK / PEMBER & ASSOCIATES NEUROLOGICA , THEIR PURPOSE IS TO PROVIDE THE
FIXTURES INSTALLED. MEP / ARCHITECTURAL ENGINEERS SITING, ELECTRICAL, MECHANICAL AND ENVIRONMENTAL GENE NOWACZYK
SPECIFICATIONS REQUIRED TO ACCOMMODATE THE MEP
DOORS AND WINDOWS, ESPECIALLY RADIATION PROTECTION BARRIERS, INSTALLATION AND OPERATION OF THE DIAGNOSTIC SITE PLANNING &
ITS USE IS AUTHORIZED ONLY FOR THE CUSTOMER SIOUX
INSTALLED AND FINISHED WITH LOCK SETS OPERATIONAL. IMAGING EQUIPMENT AND SUB-COMPONENTS PROJECT DEVELOPMENT
FALLS VA MEDICAL CENTER/CONTRACTOR N/A TO DESIGN
AND INCORPORATE OUR CONCEPT INTO CONSTRUCTION ILLUSTRATED. OFFICE: 816-741-5558
ALL ELECTRICAL CONVENIENCE, CONDUIT, RACEWAY AND JUNCTION BOXES
AND PREPARATION FOR IMAGING EQUIPMENT CELL: 816-223-6378
INSTALLED. THE LAYOUT(S) PROVIDED FOR ALL COMPONENTS
INSTALLATION. THESE DRAWINGS SHALL NOT BE USED FOR FAX: 816-587-2423
ANY PURPOSE OTHER THAN THE AGREED UPON DESIGN PURCHASED FROM AND/OR PROVIDED BY THE EQUIPMENT EMAIL: GNOWACZYK@RADSOURCE.NET
INCOMING MAINS POWER OPERATIONAL AND CONNECTED TO X-RAY ROOM MANUFACTURER SHALL SERVE AS A GUIDE FOR 8121 N.W. 97TH TERR
BETWEEN NEUROLOGICA AND THEIR CUSTOMER.
BREAKER. INSTALLATION BY THE LOCAL SERVICE/INSTALLATION KANSAS CITY, MISSOURI 64153
REPRESENTATIVES.
115 VOLTS CONVENIENCE OUTLETS OPERATIONAL. THE CUSTOMER SIOUX FALLS VA MEDICAL CENTER IS
RESPONSIBLE FOR ALL ROOM PREPARATION COSTS, FEES, PROJECT
INFORMATION IN THESE DRAWINGS RELATING TO
ALL SUPPORT STRUCTURES CORRECTLY INSTALLED. PERMITS AND INSPECTIONS UNLESS OTHERWISE
BUILDING/FACILITY SPECIFICATIONS THAT WILL SUPPORT
SPECIFIED IN THE GENERAL ORDER FOR THE EQUIPMENT
IMAGING EQUIPMENT COMPONENTS SUCH AS ELECTRICAL,
ALL CHANNELS, PIPES, BEAM'S AND OR OTHER SUPPORTING DEVICES SHOULD PURCHASE. SAMSUNG GC85
STRUCTURAL, MECHANICAL AND ENVIRONMENTAL
BE LEVEL, PARALLEL AND FREE OF LATERAL OR LONGITUDINAL REQUIREMENTS SHALL BE UTILIZED AS A GUIDE BY THE
MOVEMENTS. RADIATION SHIELDING IS NOT SHOWN ON THESE PLANS. CUSTOMER'S ARCHITECT N/A AND/OR CONTRACTOR N/A. SIOUX FALLS VA MEDICAL
HOWEVER, THE CUSTOMER SIOUX FALLS VA MEDICAL FACILITY CONDITIONS MAY DICTATE ACTUAL
ALL CONTRACTOR SUPPLIED CABLES PULLED AND TERMINATED. CENTER SHALL, AT THEIR OWN EXPENSE, HAVE SHIELDING
CENTER
CONSTRUCTION METHODS AND MATERIALS APPLIED.
CALCULATIONS FOR THE ROOM PREPARED BY A LICENSED HOWEVER, ALL METHODS AND MATERIALS MUST COMPLY 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
A DUST FREE ENVIRONMENT IN AND AROUND THE PROCEDURE ROOM. RADIATION PHYSICIST. WITH EQUIPMENT MANUFACTURER SPECIFICATIONS, AS
WELL AS LOCAL AND/OR NATIONAL BUILDING CODES.
ALL HEATING AND VENTILATION / AIR-CONDITIONING INSTALLED AND
OPERATIONAL. THE REQUIRED CEILING HEIGHT OF MIN 9'-4" / MAX 9'-6.0"
INDICATED ON THESE PLANS IS TO INSURE EQUIPMENT
A CLEAR DOOR OPENING FOR MOVING EQUIPMENT INTO THE BUILDING MUST ARCHITECT:
FUNCTION IS NOT INHIBITED. CONSULT WITH YOUR N/A
BE 42 IN. X 82 IN. OR LARGER CONTINGENT ON AN 8' CORRIDOR WIDTH. EQUIPMENT INSTALLATION SPECIALIST REGARDING p.N/A
ACCEPTABILITY OF THE OTHER CEILING HEIGHTS. CHECK N/A
ALL DOOR OPENINGS FROM DELIVERY LOCATION TO
WHERE EQUIPMENT IS TO BE INSTALLED. ELECTRICAL CONTRACTOR:
XXX
THE QUALITY OF CONSTRUCTION METHODS, MATERIALS p.XXX-XXX-XXXX
AND CONFORMANCE TO EQUIPMENT MANUFACTURER m.XXX-XXX-XXXX
SPECIFICATIONS AND TOLERANCES, WILL AFFECT .NET/.COM
EQUIPMENT PERFORMANCE.
GENERAL CONTRACTOR:
N/A
MAGNETIC INTERFERENCE SPECIFICATIONS p.XXX-XXX-XXXX
DIGITAL FLAT PANEL MUST BE LOCATED IN AMBIENT STATIC m.XXX-XXX-XXXX
MAGNETIC FIELDS OF LESS THAN ONE GAUSS TO N/A
GUARANTEE SPECIFIED IMAGING PERFORMANCE.
PROJECT MANAGER FOR CUSTOMER:
X-RAY TUBES AND EQUIPMENT MUST BE LOCATED IN KERRY NILSON
AMBIENT STATIC MAGNETIC FIELDS OF LESS THAN TEN p.XXX
GAUSS TO GUARANTEE SPECIFIED PERFORMANCE DATA KERRY.NILSON@VA.GOV
INTEGRITY
IT
SYSTEM ELECTRONICS / EQUIPMENT MUST BE LOCATED IN TROY MERGEN
AMBIENT STATIC MAGNETIC FIELDS OF LESS THAN TEN p. XXX
TROY.MERGEN@VA.GOV
GAUSS TO GUARANTEE DATA INTEGRITY
CONSOLE EQUIPMENT MUST BE LOCATED IN AMBIENT RADIOLOGY MANAGER
STATIC MAGNETIC FIELDS OF LESS THAN ONE GAUSS TO ERIN HEWER
OBTAIN SPECIFIED GEOMETRIC LINEARITY. p.XXX
ERIN.HEWER@VA.GOV

SITE ENVIRONMENT SPECIFICATIONS


AMBIENT OPERATING TEMPERATURE: 55⁰ TO 75⁰ F (16⁰ TO
24⁰ C) MAXIMUM REVISION: DATE:
ALLOWABLE TEMPERATURE CHANGE OF 15⁰ F, (5⁰ C) /
HOUR MAXIMUM ROOM
TEMPERATURE GRADIENT 5⁰ F, (3⁰ C)
HUMIDITY: 20 TO 80 PERCENT NON-CONDENSING, MAXIMUM
ALLOWABLE CHANGE OF 5 PERCENT / HOUR
ALTITUDE: NOT TO EXCEED 10,000 FT (3050M) ABOVE SEA
LEVEL
THE ENVIRONMENT FOR THE ELECTRONICS CABINET MUST
BE CONTROLLED SO THE ABOVE RESTRICTIONS ARE NOT DRAWN BY: GENE N.
EXCEEDED. CHK'D BY: A.W.
DO NOT RESTRICT THE AIR INTAKE OR AIR EXHAUST OF THE CUSTOMER :
SYSTEM COMPONENTS.
ENVIRONMENTAL CONDITIONS LISTED ABOVE MUST BE SCALE: NOTED
MAINTAINED AT ALL TIMES, INCLUDING OVERNIGHT AND
DATE:
HOLIDAYS. 6/16/2017 10:43:52 AM

FILE NAME:
Sioux Falls F1 06-16-2017.dwg

DRAWING SHEET #

AN
ELECTRICAL NOTES
ELECTRICAL LEGEND
MAIN DISCONNECT CONTRACTOR SUPPLIED, PART # ALL WIRES SPECIFIED SHALL BE COPPER STRANDED,
( NEA-43080-4W-ST24DC-SPD-WLC-FC) SEE SHEET (E1.3) CAN BE FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT
CBE1
OBTAINED FROM RADSOURCE IMAGING, CONTACT GENE LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR
NOWACZYK AT 816-223-6378 STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER, GENE NOWACZYK
480VAC @ 80 AMP FRAME, UL LISTED WITH MAGNETIC 4 SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR
MEP
SITE PLANNING &
CONTACTOR, SEMI FLUSH MOUNTED IN X-RAY EXAM ROOM. DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING PROJECT DEVELOPMENT
FEED THREE (3) CONDUCTORS #___ AWG ,ONE (1) OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS
NEUTRAL#___AWG AND ONE (1) #1__ GROUND TO "XG" FROM MUST BE CONTINUOUS COPPER STRANDED AND FREE OFFICE: 816-741-5558
BREAKER PER N.E.C. FROM SPLICES. MIN 2 PULL STRINGS PER CONDUIT RUN. CELL: 816-223-6378
FAX: 816-587-2423
INSTALL A 5.0' TAIL IN METAL FLEX ON SPLIT COVER LEAVING 2.0'
EMAIL: GNOWACZYK@RADSOURCE.NET
OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER
8121 N.W. 97TH TERR
CONTRACTOR TO SUPPLY AND INSTALL MAIN DISCONNECT SIZES MAY BE REQUIRED BY LOCAL CODES.
KANSAS CITY, MISSOURI 64153
LISTED ALONG WITH FLEX AND CONNECTORS AS REQUIRED,
TERMINATE FLEX WITH A 90DEG BOTH ENDS. ST1 2
IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED,
AS REQUIRED IN ACCORDANCE WITH NATIONAL AND PROJECT
SEE SHEET E-3 FOR DETAILS
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT BOTTOM
OC LOCAL ELECTRICAL CODES.

UB OF BOX TO FINISHED FLOOR OR AT MIN DIRECTLY ABOVE COVE LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO
BASE THE SYSTEM CONTROL, THE POWER DISTRIBUTION UNIT SAMSUNG GC85
1 @ COVER AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT, HOSPITAL APPROVED OUTLET OR EQUIVALENT. SIOUX FALLS VA MEDICAL
INSTALLED BY CONTRACTOR (CENTER IN COVER)
CENTER
1 @ 6 x 6 x 4 JUNCTION BOX, PART # (JB060604) MOUNT CENTER GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED.
CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT
OC OF BOX TO 18.0" A.F.F.
1 @ COVER FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105

3 UB CEILING MOUNTING COMPONENTS AND WIRING IF HIGH


1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,
INSTALLED BY CONTRACTOR (CENTER IN COVER)
1 WATTAGE BULBS ARE USED. RECOMMEND LED BULBS,
E1.1
USE DIMMER CONTROLS (EXCEPT MR). DO NOT MOUNT
1 @ 14.5 x 12 x 4 RECESSED MOUNTED JUNCTION BOX WITH SPLIT ACR LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING ARCHITECT:
XG COVERAND DIVIDER, PART # (JB14.51204-SC-SP-DV) CONNECT TO "CBE1"
PER N.E.C. MOUNT CENTER 18.0" A.F.F. SEE SHEET (E1.3)
MOUNTED ACCESSORIES WILL BE PARKED. N/A

8'-11"
6'-634" p.N/A
INSTALL A 5.0' TAIL IN METAL FLEX ON ON SMALL SIDE OF SPLIT COVER ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST N/A
LEAVING 2.0' OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL
CONTRACTOR TO SUPPLY AND INSTALL JUNCTION BOX, FLEX AND
2 7 RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE
NEED FOR GREATER THAN STANDARD CABLE LENGTHS ELECTRICAL CONTRACTOR:
CONNECTORS AS REQUIRED, TERMINATE FLEX WITH A 90DEG BOTH ENDS.
E1.1 FB (REFER TO THE INTERCONNECTION DIAGRAM FOR XXX
MAXIMUM USABLE LENGTHS POINT TO POINT). SHEET QE1 p.XXX-XXX-XXXX
1 @ 10 x 3 x 4 RECESSED MOUNTED WALL DUCT WITH COVERS, MOUNT
m.XXX-XXX-XXXX
10'-1141"
XT MOUNT ONTOP OF "XG" WITH SUPPLIED CONNECTORS IN KIT. SEE SHEET
(E1.3) CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH
.NET/.COM
MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND
GENERAL CONTRACTOR:
LOCAL ELECTRICAL CODES.
SINGLE GANG 2-1/2" DEEP FLUSH MOUNTED JUNCTION BOX, N/A
ST1 SUPPLY AND INSTALL MUSHROOM HEAD "EMERGENCY STOP
8 6 PHYSICAL CONNECTION OF PRIMARY POWER TO SAMSUNG
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX

3'-821"
BUTTON" 60.0" A.F.F.. WIRE TO "CBE1" MAIN DISCONNECT CBE1 EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL N/A
MAGNETIC CONTACTOR PER N.E.C. SUPPLIED AND INSTALLED BY
CONTRACTOR XXX WITH THE SUPERVISION OF A
ELECTRICAL CONTRACTOR XG XT RADSOURCE IMAGING REPRESENTATIVE. THE
PROJECT MANAGER FOR CUSTOMER:
1'-421" 1'-4" RADSOURCE IMAGING REPRESENTATIVE WOULD BE
WARNING LIGHT (X-RAY ON), LED LIGHT FIXTURE SUPPLIED AND KERRY NILSON
WL INSTALLED BY CONTRACTOR REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION p.XXX
LOCATION, AND INSURE PROPER HANDLING OF SAMSUNG KERRY.NILSON@VA.GOV
6'-441" EQUIPMENT.
1 @ 120 VAC RECEPTACLE MOUNTED ABOVE FINISHED CEILING, IT
ELECTRICAL LAYOUT SAMSUNG GC85
ACR PLACE TAG IN CEILING GRID DIRECTLY BELOW INDICATING
ABOVE CEILING OUTLET. SCALE: 3/8" = 1'-0"
ALL ELECTRICAL CONVEINCE, NETWORKIING AND PHONE
OUTLETS TO BE 18.0' A.F.F. UNLESS OTHER WISE NOTED
TROY MERGEN
p. XXX
MINIMUM CEILING HEIGHT: MIN 9'-4" / MAX 9'-6.0" TROY.MERGEN@VA.GOV
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT TOP OF FEILD VERIFY EXACT LOCATIONS OF EXISTING ELECTRICAL
FB BOX FLUSH TO FINISHED FLOOR
1 @ COVER
PANELS, CONDUITS, JUNCTION BOXES ETC FOR NEW
INFECTIOUS DISEASE
ERIN HEWER
CONNECTIONS. p.XXX
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT, ERIN.HEWER@VA.GOV
INSTALLED BY CONTRACTOR (CENTER IN COVER) REQUIRED CONDUIT RUNS FOR BASE SYSTEM BY CONTRACTOR LENGTH OF FLEXIBLE CONDUIT SHALL NOT EXCEED 6FT

ALL CONDUITS & CONDUCTERS SHALL BE INSTALLED


CONDUIT ABOVE CEILING PER N.E.C., PULL 3 @ #3AWG; PULSE 1 @ INSIDE WALLS OR ABOVE CEILING, NO EXPOSED CONDUIT
1 CBE1 TO XG #3 AWG GROUND. CONNECT WIRE TO BREAKER AND LEAVE 6' TAILS OR CONDUCTORS WILL BE ALLOWED. REVISION: DATE:
AT "XG". LABEL "FOR X-RAY POWER"
ALL WIRE RUNS WILL BE A MINIMUM OF FLEXIBLE METAL
CONDUIT AS REQUIRED PER N.E.C. ABOVE CEILING, PULL 2 @ #__ CONDUIT (FMC)
2 CBE1 T0 ST1
CONNECT TO SHUNT TRIP EMERGENCY STOP DEVICE
SAMSUNG GC 85 ELECTRICAL INSTALLATION KIT CONTRACTOR SUPPLIED
2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW
3 XG TO OC
GRAD IS ACCEPTABLE
MAIN DISCONNECT , 3 PHASE, 480VAC, 80 AMP, 5 WIRE, 24VDC SHUNT SYMBOLS LEGEND
TRIP, INTEGERAL 24VDC CONTROL VOLTAGE, TYPE 2 SP200 200KA
NEA-43080-5W-ST24DC-SPD-WLC-FC 4 CBE1 TO WL NO CONDUIT REQUIRED WITH KIT, WITH OUT KIT USE 3/4" CND
SURGE PROTECTION, CONTROLS FOR X-RAY IN USE SIGNAGE, FLUSH GROUND FAULT INTERRUPT RECEPTACLE
G.F.I. DRAWN BY: GENE N.
COVER 15.4"w x 32.4"h
EMERGENCY STOP PUSH BUTTON / 2NO-2NC CONTACTS 110/115VAC 20 AMP CHK'D BY:
NEA-ESTOP22 A.W.
1/2" CNDUIT ABOVE CEILING, PULL 18-2 AWG THEROMOSTATE WIRE,
5 N/A TO N/A LEAVE 3.0' AT CBE1 AND 12.0' AT XG, LABLE BOTH ENDS "FOR PHONE / DATA CUSTOMER :
NEA-ESTOPCV EMERGENCY STOP COVER PVC
WARNING LIGHT"
PHONE
NEAX-ON-24DC-E X-RAY IN USE LED WARNING LIGHT, 24VDC, ETHERNET CONNECTIVITY SCALE: NOTED
DATA
XIU-CABLE 50.0' ETHERNET CABLE FOR X-RAY IN USE LIGHT 6 XG TO FB 2.0" CONDUIT BELOW GRAD DATE:
W
WALL MOUNTED PHONE 48"A.F.F. 6/16/2017 10:43:52 AM
WIRE DUCT KIT FOR ALL SAMSUNG X-RAY INSTALLATIONS. INCLUDES
W1035-KIT-GC 10.0" X 3.5" VERTICAL RISER WITH 2 DIVIDERS, TERMINAL BOX 14.0" X ABOVE CEILING CONDUIT
2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW FILE NAME:
12.0" X 3.5" AND RELATED ACCESSORIES, FLUSH MOUNTED 7 XG TO UB Sioux Falls F1 06-16-2017.dwg
GRAD IS ACCEPTABLE BELOW GRAD CONDUIT

JB060604-FC 6.0" X 6.0" X 3.5" JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL 3 x 10 HORIZONTAL WALL DUCT DRAWING SHEET #
INCOMING
8 TO CBE1 CONDUIT AS REQUIRED

E1
JB060604-FB 6.0" x 6.0" x 3.5" WELDED JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL MAINS
3 x 10 FLUSH TO FINISHED FLOOR DUCT
ELECTRICAL NOTES
ELECTRICAL LEGEND
MAIN DISCONNECT CONTRACTOR SUPPLIED, PART # ALL WIRES SPECIFIED SHALL BE COPPER STRANDED,
( NEA-43080-4W-ST24DC-SPD-WLC-FC) SEE SHEET (E1.3) CAN BE FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT
CBE1
OBTAINED FROM RADSOURCE IMAGING, CONTACT GENE LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR
NOWACZYK AT 816-223-6378 4 STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER, GENE NOWACZYK
MEP
480VAC @ 80 AMP FRAME, UL LISTED WITH MAGNETIC SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR SITE PLANNING &
CONTACTOR, SEMI FLUSH MOUNTED IN X-RAY EXAM ROOM. DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING PROJECT DEVELOPMENT
FEED THREE (3) CONDUCTORS #___ AWG ,ONE (1) OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS
NEUTRAL#___AWG AND ONE (1) #1__ GROUND TO "XG" FROM MUST BE CONTINUOUS COPPER STRANDED AND FREE OFFICE: 816-741-5558
BREAKER PER N.E.C. FROM SPLICES. MIN 2 PULL STRINGS PER CONDUIT RUN. CELL: 816-223-6378
FAX: 816-587-2423
INSTALL A 5.0' TAIL IN METAL FLEX ON SPLIT COVER LEAVING 2.0'
EMAIL: GNOWACZYK@RADSOURCE.NET
OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER
8121 N.W. 97TH TERR
CONTRACTOR TO SUPPLY AND INSTALL MAIN DISCONNECT SIZES MAY BE REQUIRED BY LOCAL CODES.
KANSAS CITY, MISSOURI 64153
LISTED ALONG WITH FLEX AND CONNECTORS AS REQUIRED,
TERMINATE FLEX WITH A 90DEG BOTH ENDS. ST1 IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED,
SEE SHEET E-3 FOR DETAILS 2 AS REQUIRED IN ACCORDANCE WITH NATIONAL AND PROJECT
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT BOTTOM OC LOCAL ELECTRICAL CODES.

UB OF BOX TO FINISHED FLOOR OR AT MIN DIRECTLY ABOVE COVE LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO
BASE THE SYSTEM CONTROL, THE POWER DISTRIBUTION UNIT SAMSUNG GC85
1 @ COVER AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT, HOSPITAL APPROVED OUTLET OR EQUIVALENT. SIOUX FALLS VA MEDICAL
INSTALLED BY CONTRACTOR (CENTER IN COVER)
CENTER
1 @ 6 x 6 x 4 JUNCTION BOX, PART # (JB060604) MOUNT CENTER GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED.
CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT
OC OF BOX TO 18.0" A.F.F.
UB
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
1 @ COVER UB 3 FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO
CEILING MOUNTING COMPONENTS AND WIRING IF HIGH
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT,
INSTALLED BY CONTRACTOR (CENTER IN COVER)
1 WATTAGE BULBS ARE USED. RECOMMEND LED BULBS,
E1.1 6'-5"
USE DIMMER CONTROLS (EXCEPT MR). DO NOT MOUNT
1 @ 14.5 x 12 x 4 RECESSED MOUNTED JUNCTION BOX WITH SPLIT LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING ARCHITECT:
XG COVERAND DIVIDER, PART # (JB14.51204-SC-SP-DV) CONNECT TO "CBE1"
PER N.E.C. MOUNT CENTER 18.0" A.F.F. SEE SHEET (E1.3)
ACR MOUNTED ACCESSORIES WILL BE PARKED. N/A
p.N/A
INSTALL A 5.0' TAIL IN METAL FLEX ON ON SMALL SIDE OF SPLIT COVER ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST N/A
LEAVING 2.0' OF EXPOSED CONDUCTOR OUTSIDE THE FLEX, ELECTRICAL 7 RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE
CONTRACTOR TO SUPPLY AND INSTALL JUNCTION BOX, FLEX AND 2 FB NEED FOR GREATER THAN STANDARD CABLE LENGTHS ELECTRICAL CONTRACTOR:
CONNECTORS AS REQUIRED, TERMINATE FLEX WITH A 90DEG BOTH ENDS. E1.1 (REFER TO THE INTERCONNECTION DIAGRAM FOR XXX
MAXIMUM USABLE LENGTHS POINT TO POINT). SHEET QE1 p.XXX-XXX-XXXX
1 @ 10 x 3 x 4 RECESSED MOUNTED WALL DUCT WITH COVERS, MOUNT
m.XXX-XXX-XXXX
XT 6'-5"

8'-241"
MOUNT ONTOP OF "XG" WITH SUPPLIED CONNECTORS IN KIT. SEE SHEET
(E1.3) .NET/.COM
CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH
MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND
GENERAL CONTRACTOR:
LOCAL ELECTRICAL CODES.
SINGLE GANG 2-1/2" DEEP FLUSH MOUNTED JUNCTION BOX, 6 N/A
ST1 SUPPLY AND INSTALL MUSHROOM HEAD "EMERGENCY STOP 8 PHYSICAL CONNECTION OF PRIMARY POWER TO SAMSUNG
p.XXX-XXX-XXXX
BUTTON" 60.0" A.F.F.. WIRE TO "CBE1" MAIN DISCONNECT CBE1 m.XXX-XXX-XXXX
EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL

3'-841"
N/A
MAGNETIC CONTACTOR PER N.E.C. SUPPLIED AND INSTALLED BY
ELECTRICAL CONTRACTOR
XG XT CONTRACTOR XXX WITH THE SUPERVISION OF A
1'-421" 1'-4"
RADSOURCE IMAGING REPRESENTATIVE. THE
PROJECT MANAGER FOR CUSTOMER:
WARNING LIGHT (X-RAY ON), LED LIGHT FIXTURE SUPPLIED AND RADSOURCE IMAGING REPRESENTATIVE WOULD BE KERRY NILSON
WL INSTALLED BY CONTRACTOR REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION p.XXX
6'-541" LOCATION, AND INSURE PROPER HANDLING OF SAMSUNG KERRY.NILSON@VA.GOV
EQUIPMENT.
1 @ 120 VAC RECEPTACLE MOUNTED ABOVE FINISHED CEILING, IT
ACR PLACE TAG IN CEILING GRID DIRECTLY BELOW INDICATING ELECTRICAL LAYOUT SAMSUNG GC85 ALL ELECTRICAL CONVEINCE, NETWORKIING AND PHONE TROY MERGEN
ABOVE CEILING OUTLET. SCALE: 3/8" = 1'-0" OUTLETS TO BE 18.0' A.F.F. UNLESS OTHER WISE NOTED p. XXX
MINIMUM CEILING HEIGHT: MIN 9'-4" / MAX 9'-6.0" TROY.MERGEN@VA.GOV
1 @ 6 x 6 x 4 JUNCTION BOX PART # (JB060604) MOUNT TOP OF FEILD VERIFY EXACT LOCATIONS OF EXISTING ELECTRICAL
FB BOX FLUSH TO FINISHED FLOOR
1 @ COVER
PANELS, CONDUITS, JUNCTION BOXES ETC FOR NEW
INFECTIOUS DISEASE
ERIN HEWER
CONNECTIONS. p.XXX
1 @ 2-1/2 CHASE NIPPLE WITH BUSHING AND LOCKNUT, ERIN.HEWER@VA.GOV
INSTALLED BY CONTRACTOR (CENTER IN COVER) REQUIRED CONDUIT RUNS FOR BASE SYSTEM BY CONTRACTOR LENGTH OF FLEXIBLE CONDUIT SHALL NOT EXCEED 6FT

ALL CONDUITS & CONDUCTERS SHALL BE INSTALLED


CONDUIT ABOVE CEILING PER N.E.C., PULL 3 @ #3AWG; PULSE 1 @ INSIDE WALLS OR ABOVE CEILING, NO EXPOSED CONDUIT
1 CBE1 TO XG #3 AWG GROUND. CONNECT WIRE TO BREAKER AND LEAVE 6' TAILS OR CONDUCTORS WILL BE ALLOWED. REVISION: DATE:
AT "XG". LABEL "FOR X-RAY POWER"
ALL WIRE RUNS WILL BE A MINIMUM OF FLEXIBLE METAL
CONDUIT AS REQUIRED PER N.E.C. ABOVE CEILING, PULL 2 @ #__ CONDUIT (FMC)
2 CBE1 T0 ST1
CONNECT TO SHUNT TRIP EMERGENCY STOP DEVICE
SAMSUNG GC 85 ELECTRICAL INSTALLATION KIT CONTRACTOR SUPPLIED
2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW
3 XG TO OC
GRAD IS ACCEPTABLE
MAIN DISCONNECT , 3 PHASE, 480VAC, 80 AMP, 5 WIRE, 24VDC SHUNT SYMBOLS LEGEND
TRIP, INTEGERAL 24VDC CONTROL VOLTAGE, TYPE 2 SP200 200KA
NEA-43080-5W-ST24DC-SPD-WLC-FC 4 CBE1 TO WL NO CONDUIT REQUIRED WITH KIT, WITH OUT KIT USE 3/4" CND
SURGE PROTECTION, CONTROLS FOR X-RAY IN USE SIGNAGE, FLUSH GROUND FAULT INTERRUPT RECEPTACLE
G.F.I. DRAWN BY: GENE N.
COVER 15.4"w x 32.4"h
EMERGENCY STOP PUSH BUTTON / 2NO-2NC CONTACTS 110/115VAC 20 AMP CHK'D BY:
NEA-ESTOP22 A.W.
1/2" CNDUIT ABOVE CEILING, PULL 18-2 AWG THEROMOSTATE WIRE,
5 N/A TO N/A LEAVE 3.0' AT CBE1 AND 12.0' AT XG, LABLE BOTH ENDS "FOR PHONE / DATA CUSTOMER :
NEA-ESTOPCV EMERGENCY STOP COVER PVC
WARNING LIGHT"
PHONE
NEAX-ON-24DC-E X-RAY IN USE LED WARNING LIGHT, 24VDC, ETHERNET CONNECTIVITY SCALE: NOTED
DATA
XIU-CABLE 50.0' ETHERNET CABLE FOR X-RAY IN USE LIGHT 6 XG TO FB 2.0" CONDUIT BELOW GRAD DATE:
W
WALL MOUNTED PHONE 48"A.F.F. 6/16/2017 10:43:52 AM
WIRE DUCT KIT FOR ALL SAMSUNG X-RAY INSTALLATIONS. INCLUDES
W1035-KIT-GC 10.0" X 3.5" VERTICAL RISER WITH 2 DIVIDERS, TERMINAL BOX 14.0" X ABOVE CEILING CONDUIT
2.0" CONDUIT, ABOVE CEILING IS ILLUSTRATED HOWEVER BELOW FILE NAME:
12.0" X 3.5" AND RELATED ACCESSORIES, FLUSH MOUNTED 7 XG TO UB Sioux Falls F1 06-16-2017.dwg
GRAD IS ACCEPTABLE BELOW GRAD CONDUIT

JB060604-FC 6.0" X 6.0" X 3.5" JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL 3 x 10 HORIZONTAL WALL DUCT DRAWING SHEET #
INCOMING
8 TO CBE1 CONDUIT AS REQUIRED

E2
JB060604-FB 6.0" x 6.0" x 3.5" WELDED JUNCTION BOX WITH FLUSH COVER, PAINTED STEEL MAINS
3 x 10 FLUSH TO FINISHED FLOOR DUCT
3.5" 10."

6"
FINISHED CEILING
DAT. 9'-4"
GENE NOWACZYK
XT MEP
XT SITE PLANNING &

TO ABOVE FINISHED CEILIG


PROJECT DEVELOPMENT
CBE1
OFFICE: 816-741-5558
SUPPLIED CELL: 816-223-6378
DIVIDER FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153

CONNECT 2.0" CND FIT INTO NEAREST


SUPPLIED STUD POCKET PROJECT
FOR INCOMING LINE
DIVIDER
FEED TO CBE1 XG

1'-0"
XG SAMSUNG GC85
3.5"
14.5" 1'-4"

1'-0"
FINISHED FLOOR
SIOUX FALLS VA MEDICAL
DAT. 0'-0" CENTER
JUNCTION BOX/DUCT DETAILS 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
SCALE: NTS

1 JUNCTION BOX ELEVATION ARCHITECT:


SCALE: 3/8" = 1'-0" N/A
p.N/A
FINISHED CEILING N/A
DAT. 9'-4"
ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
.NET/.COM

GENERAL CONTRACTOR:
N/A
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
N/A

PROJECT MANAGER FOR CUSTOMER:


KERRY NILSON
UB p.XXX
1'-1178" KERRY.NILSON@VA.GOV

IT
FINISHED FLOOR TROY MERGEN
DAT. 0'-0" p. XXX
TROY.MERGEN@VA.GOV

INFECTIOUS DISEASE
ERIN HEWER
2 UPRIGHT ELEVATION p.XXX
ERIN.HEWER@VA.GOV
SCALE: 3/8" = 1'-0"

REVISION: DATE:

DRAWN BY: GENE N.


CHK'D BY: A.W.
CUSTOMER :

SCALE: NOTED
DATE:
6/16/2017 10:43:52 AM

FILE NAME:
Sioux Falls F1 06-16-2017.dwg

DRAWING SHEET #

E1.1
WIRING DIAGRAM KNOCKOUTS
AA
.38 SYMBOL
CONDUIT SIZE DIAMETER
.25 DIA [10] IN MM IN MM
[6]
C 1.00 25 1.38 35 GENE NOWACZYK
5.96 A
S
/ B
S
/ E 1.50 38 2.00 51
MEP
.50 DIA N N
SITE PLANNING &
[12] F 2.00 51 2.50 64 PROJECT DEVELOPMENT
(2 HOLES) 1.85
J250S G 2.50 64 3.00 76 OFFICE: 816-741-5558
14.35 2.85 H 3.00 76 3.63 92 CELL: 816-223-6378
H150 F OR S FAX: 816-587-2423
DASHED LINES
J250 F OR S AND CJ250NS EMAIL: GNOWACZYK@RADSOURCE.NET
INDICATE CIRCUIT BREAKER TERMINAL LUG DATA
SURFACE COVER NEMA/EEMAC TYPE 1 8121 N.W. 97TH TERR
CIRCUIT WIRE SIZE KANSAS CITY, MISSOURI 64153
BREAKER AMPERE CONDUCTORS (AWG/KCMIL)
TYPE
1.25 CATALOG NO. RATING PER PHASE

2.98 .25 PREFIX MAX MIN


PROJECT
5.95 C,E,F,G,H HD, HG
15-150 1 #3/0 #14 AL/CU
HJ, HL, HR
J250S FROM
MTG 2.08 3.10
31.43 C
G SURFACE 150-175 1 #4/0 #4 SAMSUNG GC85
JD, JG
.08 JJ, JL, JR
AL/CU
200-250 1 350 #3/0
J250F C
G SIOUX FALLS VA MEDICAL
27.00 25.30 CJ250NS AND H150 COVER OPENING
32.50 CENTER
SEE COVER
OPENING DETAILS
31.00 FIELD INSTALLABLE SOLID NEUTRAL ASSEMBLIES AND GROUND KIT
WIRE SIZE 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
.25 .08 CATALOG
MAXIMUM
NUMBER OF
CONDUCTORS
(AWG/KCMIL)
AMPERE PER TYPE
NUMBER TERMINALS
13.50 13.50 RATING TERMINAL MAX MIN
.28 DIA 1.00 4.00
2.08 #1/0 #14 CU
[7] SN100FA 100A 2 1
(2 HOLES) #1/0 #12 AL
HAS ARCHITECT:
2.18 OFFSET 4.25 2
1 #600 #1 AL OR CU N/A
SN400LA 250A OR 2 #250 #1 AL OR CU p.N/A
2 1 #300 #4 AL OR CU N/A
PK0GTJ250 250A 2 1 #300 #6 AL OR CU
J250 COVER OPENING ELECTRICAL CONTRACTOR:
11.90 1.23 2.85 PK0GTH150 150A 2 1 #2 #14 AL OR CU XXX
WITH TRIP SETTING ACCESS
FOR 200% NEURALS USE SN400LA (AMP RANGE 100 TO 250, p.XXX-XXX-XXXX
J250F C,E,F,G,H CU WIRE ONLY). m.XXX-XXX-XXXX
15.42 FIELD INSTALLABLE CIRCUIT BREAKER DATA .NET/.COM
UL LISTED INTERRUPTING RATING
CIRCUIT RMS SYMMETRICAL AMPERES GENERAL CONTRACTOR:
BREAKER WIRING AMPERE N/A
C,E,F,G,H CATALOG NO DIA RATING AC VOLTAGE DC VOLTAGE
1.82 p.XXX-XXX-XXXX
PREFIX 600Y/
240 480Y/277 480 600 125 250 500 m.XXX-XXX-XXXX
347 N/A
SEISMIC NOTES: HD A,B 15-150 25K 18K 18K 14K 14K 20K 20K -
USE (4) 1/4" DIA GRADE 5 STEEL MOUNTING BOLTS @ HOLES AA JD A,B 150-250 25K 18K 18K 14K 14K 20K 20K -
.25 9.25 2.37 MAX CONFIGURED WEIGHT 41 LB FOR THE PURPOSE OF DETERMINING HG A,B 15-150 65K 35K 35K 18K 18K 20K 20K - PROJECT MANAGER FOR CUSTOMER:
SEISMIC ANCHORAGE REQUIREMENTS. FOR ALL OTHER APPLICATIONS, JG A,B 150-250 65K 35K 35K 18K 18K 20K 20K - KERRY NILSON
CONTACT SCHNEIDER ELECTRIC. p.XXX
HJ A,B 15-150 100K 65K 65K 25K 25K 20K 20K -
14.00 JJ A,B 150-250 100K 65K 65K 25K 25K 20K 20K - KERRY.NILSON@VA.GOV
HL A,B 15-150 125K 65K 65K 25K 25K 20K 20K -
JL A,B 150-250 125K 65K 65K 25K 25K 20K 20K - IT
DUAL DIMENSIONS: INCHES
HR A,B 15-150 125K 65K 65K 25K 25K 20K 20K - TROY MERGEN
MILLIMETERS
JR A,B 150-250 125K 65K 65K 25K 25K 20K 20K - p. XXX
TROY.MERGEN@VA.GOV

INFECTIOUS DISEASE
ERIN HEWER
p.XXX
A A ERIN.HEWER@VA.GOV
Main
X-RAY
Circuit
Contractor Supplied B B Generator
480V ac, 3 Phase + Digital
C Breaker
GND. C X-RAY REVISION: DATE:
ST24DC
50/60Hz PDU
GND GND

AC 3
Phase
DC Type
2
Ethernet
In Use Sign
SPD 24VDC DRAWN BY: GENE N.
CHK'D BY: A.W.
CUSTOMER :
Imaging
Class 2 Interface
Internal Class 2
Cabinet SCALE: NOTED
Emergency 24VDC Controls 24VDC DATE:
(Dry Contact)
Power Local 6/16/2017 10:43:52 AM
Off(EPO) Indication
FILE NAME:
Sioux Falls F1 06-16-2017.dwg
MAIN DISCONNECT ( NEA-43080-4W-ST24DC-SPD-WLC-FC)
DRAWING SHEET #

E1.3
CABINET CEILING
GENE NOWACZYK
MEP
WIRING : WALL AND CEILING SITE PLANNING &
PROJECT DEVELOPMENT

OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR

7.85 m
KANSAS CITY, MISSOURI 64153
POWER
DISTRIBUTION PROJECT

SAMSUNG GC85
SIOUX FALLS VA MEDICAL
CENTER
CABINET 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105

CABINET WORK-
STATION ARCHITECT:
N/A
p.N/A
N/A
WIRING : FLOOR
CEILING ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
.NET/.COM

GENERAL CONTRACTOR:

13.1 m N/A
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
WALL STAND EXAM ROOM N/A
TABLE
PROJECT MANAGER FOR CUSTOMER:
KERRY NILSON
p.XXX
KERRY.NILSON@VA.GOV

IT
TROY MERGEN
p. XXX
TROY.MERGEN@VA.GOV

INFECTIOUS DISEASE
ERIN HEWER
p.XXX
POWER
DISTRIBUTION CABINET WALLSTAND CABINET TABLE CABINET ERIN.HEWER@VA.GOV

WIRING : WALL or FLOOR WIRING : FLOOR WIRING : FLOOR


WORK STATION REVISION: DATE:

& CIB

3.9 m 8.5 m 11.2 m


WALL-
WALLSTAND CEILING TABLE STAND
DRAWN BY:
CHK'D BY:
GENE N.
A.W.
CUSTOMER :

WIRING : FLOOR WIRING : FLOOR


SCALE: NOTED
DATE:
6/16/2017 10:43:52 AM

13.65 m 13 m CONTROL ROOM


FILE NAME:
Sioux Falls F1 06-16-2017.dwg

DRAWING SHEET #

QE1
21.97
95.73 31.73

GENE NOWACZYK

35.98
MEP
MAX SITE PLANNING &

21.46
PROJECT DEVELOPMENT

86.44
OFFICE: 816-741-5558
CELL: 816-223-6378
FAX: 816-587-2423
67.52
10.71
MIN MAX EMAIL: GNOWACZYK@RADSOURCE.NET
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153

70.53
PROJECT

MIN °
90
FRONT VIEW SIDE VIEW 20° SAMSUNG GC85

11.12
SIOUX FALLS VA MEDICAL
95.58 18.98
CENTER

2501 WEST 22ND ST, SIOUX FALLS, SD, 57105

5.40
FRONT VIEW 31.74 SIDE VIEW

30.69
ARCHITECT:
N/A

21.94
p.N/A
N/A

ELECTRICAL CONTRACTOR:
XXX
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
.NET/.COM

TOP VIEW TOP VIEW GENERAL CONTRACTOR:


N/A
p.XXX-XXX-XXXX
m.XXX-XXX-XXXX

TABLE DETAIL DRAWING WALL STAND DETAIL DRAWING N/A

PROJECT MANAGER FOR CUSTOMER:


KERRY NILSON
p.XXX
KERRY.NILSON@VA.GOV

IT
TROY MERGEN
p. XXX
TROY.MERGEN@VA.GOV
36.31

INFECTIOUS DISEASE
49.30

ERIN HEWER
Rotation Conter p.XXX
ERIN.HEWER@VA.GOV

Focal Spot Focal Spot


REVISION: DATE:

FRONT VIEW SIDE VIEW 19.29 29.53

29.53
118.50
42.72

42.72

17.72
DRAWN BY: GENE N.
CHK'D BY: A.W.
CUSTOMER :

TOP VIEW SCALE: NOTED

FRONT VIEW SIDE VIEW TOP VIEW DATE:


6/16/2017 10:43:52 AM

FILE NAME:

THU DETAIL DRAWING CABINET DETAIL DRAWING Sioux Falls F1 06-16-2017.dwg

DRAWING SHEET #

Q1
SPANNING MEMBERS: GENERAL STRUCTURAL NOTES
1 STEEL CHANNEL STRUT FLUSH WITH FINISHED
FLOOR SLABS ON WHICH EQUIPMENT IS TO BE INSTALLED MUST
CEILING. SPACED AS INDICATED ON SHEET "S1"
BE LEVEL TO 1MM IN 10.0'.
RECOMMENDED FABRICATION DETAILS ON SHEET
3'-10" "S2" LOADS SHOULD BE CALCULATED FROM SHEET DIMENSIONS ARE TO FINISHED SURFACES OF ROOM.
G1-1 (EQUIPMENT LEGEND ITEMS 4 & 5) GENE NOWACZYK
MEP
CUSTOMERS CONTRACTOR N/A MUST PROVIDE ALL SITE PLANNING &
AREA OF EXAM ROOM TO BE LEVELED FOR X-RAY
2 EQUIPMENT, CONTRACTOR TO VERIFY MINIMUM
PENETRATIONS IN POST TENSION FLOORS. PROJECT DEVELOPMENT

SPECIFICATION OF FLOOR SLABS ON WHICH OFFICE: 816-741-5558


CUSTOMERS CONTRACTORS N/AMUST PROVIDE AND INSTALL
EQUIPMENT IS TO BE INSTALLED MUST BE LEVEL TO CELL: 816-223-6378
ALL HARDWARE FOR "THROUGH THE FLOOR" ANCHORING
1MM IN 10.0'". AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS FLOORS. FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
CHEST STAND MOUNTING PLATE / ANCHOR BOLT THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRILLING THAT
8121 N.W. 97TH TERR
3 PATTERN CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUCTION
KANSAS CITY, MISSOURI 64153
ENCOUNTERED BY THE INSTALLER SUCH AS REBAR ETC...

TABLE MOUNTING BOLT PATTERN AND LOCATION IT IS THE RESPONSIBILITY OF THE CUSTOMER SIOUX FALLS VA PROJECT
4 MEDICAL CENTER/CONTRACTOR N/A TO PERFORM ANY FLOOR
OR WALL PENETRATIONS THAT MAY BE REQUIRED. THE
CUSTOMER SIOUX FALLS VA MEDICAL CENTER/CONTRACTOR
N/A IS ALSO RESPONSIBLE FOR ENSURING THAT NO SAMSUNG GC85
SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING OR ANY
OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT WORK OR
SIOUX FALLS VA MEDICAL
STRUCTURAL SUPPORTS E.G. POST TENSION CABLES OR REBAR
WILL INTERFERE OR COME IN CONTACT WITH SUBSURFACE CENTER
PENETRATION OPERATIONS, E.G. DRILLING AND INSTALLATION
OF ANCHORING PERFORMED DURING THE INSTALLATION 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
1

1'-021" PROCESS. TO ENSURE WORKERS SAFETY INSTALLERS WILL


ONLY PERFORM SURFACE PENETRATION OPERATIONS ONLY
AFTER THE CUSTOMERS CONTRACTOR HAS VALIDATED THE
SURFACE TO BE PENETRATED.

5"
ARCHITECT:
2

ALL CEILING MOUNTED FIXTURES SUCH AS AIR VENTS, N/A


11'-1043"
SPRINKLERS etc. TO BE FLUSH MOUNTED OR NOT TO EXTEND p.N/A

9"

9"
MORE THEN 1/4" BELOW THE FINISHED CEILING. N/A

ALL STEEL WORK AND PARTS NECESSARY TO SUPPORT CEILING ELECTRICAL CONTRACTOR:
XXX
4

MOUNTED EQUIPMENT IS TO BE SUPPLIED BY THE CUSTOMER


SIOUX FALLS VA MEDICAL CENTER OR THEIR CONTRACTOR N/A. p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
.NET/.COM
ALL UNITS THAT ARE WALL MOUNTED OR WALL SUPPORTED ARE
A 3

3
TO BE PROVIDED WITH SUPPORTS WHERE NECESSARY. WALL
6'-0"

SUPPORTS ARE TO BE SUPPLIED AND INSTALLED BY THE GENERAL CONTRACTOR:


A

A
CUSTOMER SIOUX FALLS VA MEDICAL CENTER OR THEIR N/A
721"

p.XXX-XXX-XXXX
CONTRACTOR N/A.
m.XXX-XXX-XXXX
N/A
721" 4'-141" 6'-1021" GENERAL STRUCTURAL CEILING SUPPORT REQUIREMENTS
4'-8"

PROJECT MANAGER FOR CUSTOMER:


KERRY NILSON
A WELDED OR FABRICATED STRUCTURAL SUPPORT IS REQUIRED FOR CEILING MOUNTED X-RAY p.XXX
2"

UNITS. IT'S TOP MEMBER IS FASTENED TO THE FLOOR SLAB (OR BUILDING STEEL) ABOVE THE EXAM KERRY.NILSON@VA.GOV
ROOM. THE BOTTOM OF IT'S LOWEST SPANNING MEMBER IS FLUSH WITH THE FINISHED CEILING TO
WHICH THE STATIONARY CEILING TRACKS ARE FASTENED. IT
GENERAL
TROY STRUCTURAL
MERGEN NOTES
2'-2" 2'-2" 2'-2" 2'-2" 2'-2" 2'-941" THE DETAILS PROVIDED HERE INDICATE A BASIC SUPPORT METHOD THAT PERMITS TRANSVERSE AND p. XXX
LONGITUDINAL ADJUSTMENT OF BOLT CENTERS AT THE INSTALLATION SITE. IT CAN BE ADAPTED TO FLOORTROY.MERGEN@VA.GOV
SLABS ON WHICH EQUIPMENT IS TO BE INSTA
ACCOMMODATE THE VARIOUS ARCHITECTURAL CIRCUMSTANCES ENCOUNTERED AT EACH BE LEVEL TO 1MM IN 10.0'.
STRUCTURAL LAYOUT SAMSUNG GC85 INSTALLATION.
INFECTIOUS DISEASE
SCALE: 3/8" = 1'-0" ERIN HEWER
DIMENSIONS ARE TO FINISHED SURFACES OF ROOM
p.XXX
RECOMMENDED CEILING HEIGHT 9'-4.0" THE SUPPORT MUST BE RIGID, SECURELY BRACED AND LEVEL IN BOTH DIRECTIONS AT THE FINISHED ERIN.HEWER@VA.GOV
CEILING LINE. LENGTH AND LOADING CAPACITY WILL DEPEND ON THE SELECTED CEILING UNIT AND CUSTOMERS CONTRACTOR
NUMBER OF CARRIAGE ASSEMBLIES TO BE BORNE. ACTUAL EQUIPMENT LOADS FOR EACH ROOM ARE PENETRATIONS IN POST TENSION FLOORS.
SPECIFIED ON SITE PLANNING DRAWINGS UNDER THE STRUCTURAL DETAIL SPECIFICATIONS.
CUSTOMERS CONTRACTORS
IT IS SUGGESTED THAT THROUGH BOLTING THE SUPPORT STRUCTURE TO BUILDING STRUCTURE ALL HARDWARE
REVISION:FOR "THROUGH THE FLOOR" ANCHO
DATE:
UPPER SLAB CUSTOMER OR CONTRACTOR'S PREPARATION OCCUR WHENEVER POSSIBLE. APPLICABLE CONDITIONS WILL ALSO ACCEPT THE APPLICATION OF AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS
HILTI EXPANSION ANCHORS (OR EQUAL) IN LIEU OF THROUGH BOLTING. ACTUAL METHODS AND THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRIL
CEILING SUPPORT UNISTRUT (CEILING SUPPORT STRUT) MATERIALS SHALL BE DETERMINED BY THE ENGINEER OF RECORD. ALL SUPPORT MEMBERS, CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUC
ANCHOR BOLTS AND HARDWARE SHALL BE PROVIDED AND INSTALLED BY THE ENCOUNTERED BY THE INSTALLER SUCH AS REBAR E
CUSTOMER OR CONTRACTOR'S PREPARATION CUSTOMER/CONTRACTOR. THE LENGTH OF BOLTS SHALL BE DETERMINED TO MEET SPECIFIC SITE
REQUIREMENTS BY THE ENGINEER OF RECORD. IT IS THE RESPONSIBILITY OF THE CUSTOMER
(FINISHED CEILING)
CENTERLINES OF THE TWO LONGITUDINAL RAILS HOLDING THE SUSPENDED WEIGHTS ARE SHOWN. OR WALL PENETRATIONS THAT MAY BE REQUIRED. TH
CEILING FINISH STEEL CHANNEL STRUT FLUSH FINISHED CEILING SHALL BE IDENTIFIED AS THE SPANNING MEMBERS.
SAMSUNG
THESE MEMBERS MUST BE LEVEL IN BOTH DIRECTIONS TO WITHIN 1MM AND BRACED TO PREVENT
(X-RAY CEILING RAIL INSTALL) MOVEMENT IN ANY DIRECTION. THEY MUST BE SECURED TO THE SLAB ABOVE AND/OR TO THE SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING O
DRAWN BY: GENE N.
BUILDING STRUCTURE TO HOLD THE DEAD WEIGHTS INDICATED. see EQUIPMENT LEGEND "SHEET A1" OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT W
X-RAY CHK'D BY:
STRUCTURAL SUPPORTS E.G. POST TENSION
A.W. CABLES
INSTALLATION OF THE SUPPORT MAY BE AFFECTED BY OTHER ARCHITECTURAL OR MECHANICAL WILL INTERFERE
CUSTOMEROR : COME IN CONTACT WITH SUBSU
PROVISIONS WHICH MAY BE LOCATED IN THE SAME AREA. IN THIS CASE, COORDINATION OF THE PENETRATION OPERATIONS, E.G. DRILLING AND INST
REPRESENTATIVE MUST BE CONSULTED BEFORE ANY DEVIATION FROM THE SITE PLANNING OF ANCHORING PERFORMED DURING THE INSTALLAT
DRAWINGS IS UNDERTAKEN. PROCESS. TO ENSURE WORKERS SAFETY
SCALE: NOTED INSTALLER
ONLY PERFORM
DATE: SURFACE PENETRATION OPERATION
1. CUSTOMER SIOUX FALLS VA MEDICAL CENTER CAN ALSO UTILIZE AN EXISTING SUPPORT 6/16/2017CONTRACTOR
AFTER THE CUSTOMERS 10:43:52 AM HAS VALIDAT
SECTION A-A STRUCTURE, ONLY IF THE STRUCTURE IS CERTIFIED BY A SURFACE TO BE PENETRATED.
TABLE LICENSED/PROFESSIONAL STRUCTURAL ENGINEER.
FILE NAME:
ALL CEILING MOUNTED FIXTURES SUCH AS AIR VEN
2. SUPPORT STRUCTURE MUST MEET ALL REQUIRED EQUIPMENT SUPPORT NEEDS FOR WIDTH, Sioux Falls F1 06-16-2017.dwg
FLOOR FINISH DEPTH, STRENGTH, SPRINKLERS etc. TO BE FLUSH MOUNTED OR NOT T
ETC. MORE THEN 1/4" BELOW THE FINISHED CEILING.
DRAWING SHEET #
3. CUSTOMER SIOUX FALLS VA MEDICAL CENTER IS RESPONSIBLE FOR MEETING SITE CONDITIONS

S1
FOR STATIC LOADS, LEVELNESS, ETC. ALL STEEL WORK AND PARTS NECESSARY TO SUPPO
MOUNTED EQUIPMENT IS TO BE SUPPLIED BY THE CU
SIOUX FALLS VA MEDICAL CENTER
SPANNING MEMBERS: GENERAL STRUCTURAL NOTES
1 STEEL CHANNEL STRUT FLUSH WITH FINISHED
FLOOR SLABS ON WHICH EQUIPMENT IS TO BE INSTALLED MUST
CEILING. SPACED AS INDICATED ON SHEET "S1"
BE LEVEL TO 1MM IN 10.0'.
RECOMMENDED FABRICATION DETAILS ON SHEET
3'-10"
"S2" LOADS SHOULD BE CALCULATED FROM SHEET DIMENSIONS ARE TO FINISHED SURFACES OF ROOM.
G1-1 (EQUIPMENT LEGEND ITEMS 4 & 5) GENE NOWACZYK
MEP
CUSTOMERS CONTRACTOR N/A MUST PROVIDE ALL SITE PLANNING &
AREA OF EXAM ROOM TO BE LEVELED FOR X-RAY
2 EQUIPMENT, CONTRACTOR TO VERIFY MINIMUM
PENETRATIONS IN POST TENSION FLOORS. PROJECT DEVELOPMENT

SPECIFICATION OF FLOOR SLABS ON WHICH OFFICE: 816-741-5558


CUSTOMERS CONTRACTORS N/AMUST PROVIDE AND INSTALL
EQUIPMENT IS TO BE INSTALLED MUST BE LEVEL TO CELL: 816-223-6378
ALL HARDWARE FOR "THROUGH THE FLOOR" ANCHORING
1MM IN 10.0'". AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS FLOORS. FAX: 816-587-2423
EMAIL: GNOWACZYK@RADSOURCE.NET
CHEST STAND MOUNTING PLATE / ANCHOR BOLT THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRILLING THAT
8121 N.W. 97TH TERR
3 PATTERN CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUCTION
KANSAS CITY, MISSOURI 64153
ENCOUNTERED BY THE INSTALLER SUCH AS REBAR ETC...

TABLE MOUNTING BOLT PATTERN AND LOCATION IT IS THE RESPONSIBILITY OF THE CUSTOMER SIOUX FALLS VA PROJECT
4 MEDICAL CENTER/CONTRACTOR N/A TO PERFORM ANY FLOOR
OR WALL PENETRATIONS THAT MAY BE REQUIRED. THE
CUSTOMER SIOUX FALLS VA MEDICAL CENTER/CONTRACTOR
N/A IS ALSO RESPONSIBLE FOR ENSURING THAT NO SAMSUNG GC85
SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING OR ANY
OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT WORK OR
SIOUX FALLS VA MEDICAL
STRUCTURAL SUPPORTS E.G. POST TENSION CABLES OR REBAR
WILL INTERFERE OR COME IN CONTACT WITH SUBSURFACE CENTER
PENETRATION OPERATIONS, E.G. DRILLING AND INSTALLATION
1'-021"

1
OF ANCHORING PERFORMED DURING THE INSTALLATION 2501 WEST 22ND ST, SIOUX FALLS, SD, 57105

PROCESS. TO ENSURE WORKERS SAFETY INSTALLERS WILL


ONLY PERFORM SURFACE PENETRATION OPERATIONS ONLY
AFTER THE CUSTOMERS CONTRACTOR HAS VALIDATED THE
5"

SURFACE TO BE PENETRATED.

2
11'-1043"
ARCHITECT:
ALL CEILING MOUNTED FIXTURES SUCH AS AIR VENTS, N/A
9"

9"

SPRINKLERS etc. TO BE FLUSH MOUNTED OR NOT TO EXTEND p.N/A


MORE THEN 1/4" BELOW THE FINISHED CEILING. N/A

ELECTRICAL CONTRACTOR:
4

ALL STEEL WORK AND PARTS NECESSARY TO SUPPORT CEILING


MOUNTED EQUIPMENT IS TO BE SUPPLIED BY THE CUSTOMER XXX
SIOUX FALLS VA MEDICAL CENTER OR THEIR CONTRACTOR N/A. p.XXX-XXX-XXXX
m.XXX-XXX-XXXX
.NET/.COM
A 3

ALL UNITS THAT ARE WALL MOUNTED OR WALL SUPPORTED ARE


6'-0"

TO BE PROVIDED WITH SUPPORTS WHERE NECESSARY. WALL


A

GENERAL CONTRACTOR:
A
SUPPORTS ARE TO BE SUPPLIED AND INSTALLED BY THE
721"

CUSTOMER SIOUX FALLS VA MEDICAL CENTER OR THEIR N/A


p.XXX-XXX-XXXX
CONTRACTOR N/A.
m.XXX-XXX-XXXX
721" 4'-141" 6'-843" N/A
4'-8"

GENERAL STRUCTURAL CEILING SUPPORT REQUIREMENTS


PROJECT MANAGER FOR CUSTOMER:
KERRY NILSON
2"

A WELDED OR FABRICATED STRUCTURAL SUPPORT IS REQUIRED FOR CEILING MOUNTED X-RAY p.XXX
UNITS. IT'S TOP MEMBER IS FASTENED TO THE FLOOR SLAB (OR BUILDING STEEL) ABOVE THE EXAM KERRY.NILSON@VA.GOV
ROOM. THE BOTTOM OF IT'S LOWEST SPANNING MEMBER IS FLUSH WITH THE FINISHED CEILING TO
WHICH THE STATIONARY CEILING TRACKS ARE FASTENED. IT
2'-941" 2'-2" 2'-2" 2'-2" 2'-2" 2'-2" 3'-11" GENERAL
TROY STRUCTURAL
MERGEN NOTES
THE DETAILS PROVIDED HERE INDICATE A BASIC SUPPORT METHOD THAT PERMITS TRANSVERSE AND p. XXX
LONGITUDINAL ADJUSTMENT OF BOLT CENTERS AT THE INSTALLATION SITE. IT CAN BE ADAPTED TO FLOORTROY.MERGEN@VA.GOV
SLABS ON WHICH EQUIPMENT IS TO BE INSTA
ACCOMMODATE THE VARIOUS ARCHITECTURAL CIRCUMSTANCES ENCOUNTERED AT EACH BE LEVEL TO 1MM IN 10.0'.
STRUCTURAL LAYOUT SAMSUNG GC85 INSTALLATION.
INFECTIOUS DISEASE
SCALE: 3/8" = 1'-0" ERIN HEWER
DIMENSIONS ARE TO FINISHED SURFACES OF ROOM
p.XXX
RECOMMENDED CEILING HEIGHT 9'-4.0" THE SUPPORT MUST BE RIGID, SECURELY BRACED AND LEVEL IN BOTH DIRECTIONS AT THE FINISHED ERIN.HEWER@VA.GOV
CEILING LINE. LENGTH AND LOADING CAPACITY WILL DEPEND ON THE SELECTED CEILING UNIT AND CUSTOMERS CONTRACTOR
NUMBER OF CARRIAGE ASSEMBLIES TO BE BORNE. ACTUAL EQUIPMENT LOADS FOR EACH ROOM ARE PENETRATIONS IN POST TENSION FLOORS.
SPECIFIED ON SITE PLANNING DRAWINGS UNDER THE STRUCTURAL DETAIL SPECIFICATIONS.
CUSTOMERS CONTRACTORS
IT IS SUGGESTED THAT THROUGH BOLTING THE SUPPORT STRUCTURE TO BUILDING STRUCTURE ALL HARDWARE
REVISION:FOR "THROUGH THE FLOOR" ANCHO
DATE:
UPPER SLAB CUSTOMER OR CONTRACTOR'S PREPARATION OCCUR WHENEVER POSSIBLE. APPLICABLE CONDITIONS WILL ALSO ACCEPT THE APPLICATION OF AND/OR ANY BRACING UNDER SLAB AND/OR ACCESS
HILTI EXPANSION ANCHORS (OR EQUAL) IN LIEU OF THROUGH BOLTING. ACTUAL METHODS AND THE CONTRACTOR MUST ALSO PROVIDE FLOOR DRIL
CEILING SUPPORT UNISTRUT (CEILING SUPPORT STRUT) MATERIALS SHALL BE DETERMINED BY THE ENGINEER OF RECORD. ALL SUPPORT MEMBERS, CAN NOT BE COMPLETED BECAUSE OF AN OBSTRUC
ANCHOR BOLTS AND HARDWARE SHALL BE PROVIDED AND INSTALLED BY THE ENCOUNTERED BY THE INSTALLER SUCH AS REBAR E
CUSTOMER OR CONTRACTOR'S PREPARATION CUSTOMER/CONTRACTOR. THE LENGTH OF BOLTS SHALL BE DETERMINED TO MEET SPECIFIC SITE
REQUIREMENTS BY THE ENGINEER OF RECORD. IT IS THE RESPONSIBILITY OF THE CUSTOMER
(FINISHED CEILING)
CENTERLINES OF THE TWO LONGITUDINAL RAILS HOLDING THE SUSPENDED WEIGHTS ARE SHOWN. OR WALL PENETRATIONS THAT MAY BE REQUIRED. TH
CEILING FINISH STEEL CHANNEL STRUT FLUSH FINISHED CEILING SHALL BE IDENTIFIED AS THE SPANNING MEMBERS.
SAMSUNG
THESE MEMBERS MUST BE LEVEL IN BOTH DIRECTIONS TO WITHIN 1MM AND BRACED TO PREVENT
(X-RAY CEILING RAIL INSTALL) MOVEMENT IN ANY DIRECTION. THEY MUST BE SECURED TO THE SLAB ABOVE AND/OR TO THE SUBSURFACE UTILITIES i.e. ELECTRICAL, PLUMBING O
DRAWN BY: GENE N.
BUILDING STRUCTURE TO HOLD THE DEAD WEIGHTS INDICATED. see EQUIPMENT LEGEND "SHEET A1" OTHER FORM OF WIRING, CONDUITS, PIPING, DUCT W
X-RAY CHK'D BY:
STRUCTURAL SUPPORTS E.G. POST TENSION
A.W. CABLES
INSTALLATION OF THE SUPPORT MAY BE AFFECTED BY OTHER ARCHITECTURAL OR MECHANICAL WILL INTERFERE
CUSTOMEROR : COME IN CONTACT WITH SUBSU
PROVISIONS WHICH MAY BE LOCATED IN THE SAME AREA. IN THIS CASE, COORDINATION OF THE PENETRATION OPERATIONS, E.G. DRILLING AND INST
REPRESENTATIVE MUST BE CONSULTED BEFORE ANY DEVIATION FROM THE SITE PLANNING OF ANCHORING PERFORMED DURING THE INSTALLAT
DRAWINGS IS UNDERTAKEN. PROCESS. TO ENSURE WORKERS SAFETY
SCALE: NOTED INSTALLER
ONLY PERFORM
DATE: SURFACE PENETRATION OPERATION
1. CUSTOMER SIOUX FALLS VA MEDICAL CENTER CAN ALSO UTILIZE AN EXISTING SUPPORT 6/16/2017CONTRACTOR
AFTER THE CUSTOMERS 10:43:52 AM HAS VALIDAT
SECTION A-A STRUCTURE, ONLY IF THE STRUCTURE IS CERTIFIED BY A SURFACE TO BE PENETRATED.
TABLE LICENSED/PROFESSIONAL STRUCTURAL ENGINEER.
FILE NAME:
ALL CEILING MOUNTED FIXTURES SUCH AS AIR VEN
2. SUPPORT STRUCTURE MUST MEET ALL REQUIRED EQUIPMENT SUPPORT NEEDS FOR WIDTH, Sioux Falls F1 06-16-2017.dwg
FLOOR FINISH DEPTH, STRENGTH, SPRINKLERS etc. TO BE FLUSH MOUNTED OR NOT T
ETC. MORE THEN 1/4" BELOW THE FINISHED CEILING.
DRAWING SHEET #
3. CUSTOMER SIOUX FALLS VA MEDICAL CENTER IS RESPONSIBLE FOR MEETING SITE CONDITIONS

S2
FOR STATIC LOADS, LEVELNESS, ETC. ALL STEEL WORK AND PARTS NECESSARY TO SUPPO
MOUNTED EQUIPMENT IS TO BE SUPPLIED BY THE CU
SIOUX FALLS VA MEDICAL CENTER
STRUT CHANNEL SUPPORT FRAME NOTES
RECOMMENDED UNIVERSAL STEEL STRUT
CHANNEL SUPPORT FRAME.

ALLOWS FOR TRANSVERSE AND LONGITUDINAL GENE NOWACZYK


ADJUSTMENT AT TIME OF EQUIPMENT MEP
SITE PLANNING &
INSTALLATION. DETERMINATION OF THE
PROJECT DEVELOPMENT
ACTUAL DESIGN SHALL BE COORDINATED AND
APPROVED BY THE ENGINEER OF RECORD. ALL OFFICE: 816-741-5558
STEEL STRUT MEMBERS SHALL BE PROVIDED CELL: 816-223-6378
AND INSTALLED BY CUSTOMER / THEIR FAX: 816-587-2423
CONTRACTOR. EMAIL: GNOWACZYK@RADSOURCE.NET
4
8121 N.W. 97TH TERR
KANSAS CITY, MISSOURI 64153
SPANNING MEMBERS: STEEL CHANNEL STRUT
1
FLUSH WITH FINISHED CEILING. SPACED AS
INDICATED ON SHEET "S1" PROJECT
LONGITUDINAL EQUIPMENT RAILS: FASTENED
2
TO UNDERSIDE OF SPANNING MEMBERS.
SAMSUNG GC85
COLUMNS: STEEL CHANNEL STRUT ATTACHED
3 TO SPANNING MEMBERS AS SHOWN ON SIOUX FALLS VA MEDICAL
GENERAL CEILING DETAIL. NUMBER AND CENTER
LOCATION OF VERTICAL COLUMNS SHALL VARY
DEPENDING ON BUILDING CONSTRUCTION.
2501 WEST 22ND ST, SIOUX FALLS, SD, 57105
1
BRACES: STEEL CHANNEL STRUT ATTACHED TO
3 4 VERTICAL COLUMNS AS SHOWN ON DETAIL.

THE ATTACHED DRAWINGS ARE FOR LAYOUT


ARCHITECT:
PURPOSES AND GENERAL MEANS OF
N/A
FABRICATION. THE CUSTOMER'S ENGINEER OF p.N/A
RECORD OR AGENT SHALL PREPARE N/A
CONSTRUCTION DOCUMENTS.
ELECTRICAL CONTRACTOR:
THESE DRAWINGS INDICATE PLACEMENT OF XXX
THE PURCHASED EQUIPMENT, AS WELL AS p.XXX-XXX-XXXX
LONGITUDINAL EQUIPMENT RAIL m.XXX-XXX-XXXX
2 DESCRIBE THE STRUCTURAL REQUIREMENTS
FOR THAT EQUIPMENT. RADSOURCE WILL NOT .NET/.COM
SPACING AS SPECIFIED ON PLANS.
BE RESPONSIBLE FOR OTHER DESIGNS AND
CONSTRUCTION. GENERAL CONTRACTOR:
N/A
p.XXX-XXX-XXXX
EVERY EFFORT HAS BEEN MADE TO ASSURE
m.XXX-XXX-XXXX
THAT THE EQUIPMENT DEPICTED ON THE
SEE GENERAL N/A
ATTACHED RECOMMENDED LAYOUT,
CONFIGURES THE LAYOUT PLAN TO ALLOW
CONNECTION DETAIL PROJECT MANAGER FOR CUSTOMER:
OPTIMUM OPERATION OF THE EQUIPMENT. THE
KERRY NILSON
2'-2" CUSTOMER'S ENGINEER OF RECORD AND / OR p.XXX
AGENT SHALL BEAR SOLE RESPONSIBILITY FOR KERRY.NILSON@VA.GOV
COMPLIANCE WITH APPLICABLE CODES ..
IT
O.C. TYPICAL N/A SHALL PROVIDE AND INSTALL ALL TROY MERGEN
MATERIALS ALONG WITH ANY OTHER FEATURES p. XXX
SCHEMATIC / ISOMETRIC - SUSPENSION STRUCTURE CALLED OUT IN THE PLANS. TROY.MERGEN@VA.GOV
NO SCALE INFECTIOUS DISEASE
THE CONTRACTOR N/AIS RESPONSIBLE FOR
ERIN HEWER
ANY FINAL PAINT OR TOUCH-UP WORK WHICH
SPANNING MEMBER PROVIDED p.XXX
& INSTALLED BY CUSTOMER/CONTRACTOR SHALL BE COMPLETED AFTER THE ERIN.HEWER@VA.GOV
VERTICAL COLUMN 3 INSTALLATION OF THE PURCHASED SYSTEM.

4 EXISTING SERVICES THAT WILL NOT BE USED


4 (PLUMBING, JUNCTION BOXES, FLOOR DRAINS
FINISHED
CEILING ETC.) MUST BE CAPPED OR COVERED PRIOR TO REVISION: DATE:
LINE COMMENCEMENT OF EQUIPMENT
BRACE REQUIRED FITTING INSTALLATION.

MINIMUM CEILING HEIGHT: MIN 9'-4"


REQUIRED FITTING / MAX 9'-6.0"

LONGITUDINAL
EQUIPMENT RAIL REQUIRED FITTING

SPANNING MEMBER DRAWN BY: GENE N.


1 CHK'D BY: A.W.
CUSTOMER :
BOTTOM OF SPANNING MEMBER FINISHED CEILING LINE
FLUSH WITH FINISHED CEILING LINE
EACH MEMBER 2'-2" O.C. SCALE: NOTED

MOUNTING SURFACE OF DATE:


LONGITUDINAL EQUIPMENT RAIL 6/16/2017 10:43:52 AM

FILE NAME:
Sioux Falls F1 06-16-2017.dwg
GENERAL CEILING DETAIL GENERAL CONNECTION DETAIL
NO SCALE NO SCALE DRAWING SHEET #

S2