Professional Documents
Culture Documents
Manual PN 51852 Aquavaire Supplement
Manual PN 51852 Aquavaire Supplement
2019.03.0
MANUAL PN 51852 AQUAVAIRE SUPPLEMENT REV 02-27-19 4 09:46:16
-08'00'
DISCLAIMER and WARNING
Read the OPERATION MANUAL before operating this equipment.
DISCLAIMER
Algas-SDI International, LLC (ASDI) shall not be liable for technical
and/or editorial errors or omissions in this manual. ASDI makes no
warranties, express or implied, including the implied warranties of
merchantability and fitness for a particular purpose with respect to
this manual.
Read this Operation and Maintenance Manual before operating the QH, QV or
QM systems.
DISCLAIMER
Algas-SDI International, LLC (ASDI) shall not be liable for technical and/or
editorial errors or omissions in this manual. ASDI makes no warranties, express
or implied, including the implied warranties of merchantability and fitness for a
particular purpose with respect to this manual. In no event shall ASDI be liable
for any special or consequential damages including, but not limited to loss of
production time, or loss of profits.
Product names used in the manual are for the manufacturer or supplier
identification only and may be trademarks, or registered trademarks, of these
companies.
Waterbath vaporizers contain all the ingredients for metallic corrosion. In order
to minimize this corrosion, certain steps must be taken. Grounding the vaporizer
skid and adding a rust-inhibiting antifreeze and/or rust inhibitor to the heat
transfer fluid (HTF) will help minimize this corrosion.
______
Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852 1-1
Waterbath Freeze and Corrosion Protection
The QV/QM/QH is an unpressurized, closed-loop waterbath vaporizer, meaning its heat transfer section is vented
to atmosphere and requires slightly different maintenance than a sealed and pressurized closed-loop system.
This is accomplished by increasing the frequency of testing the heat transfer fluid (HTF) to once every 30 days.
The QV/QM/QH is cleaned at the factory with a liquid alkaline cleaner to remove loose mill scale, oil,
contaminants and corrosion that occurred during the manufacturing process. After testing, and prior to shipment,
the vaporizer was drained and a vapor phase corrosion inhibitor (VPCI), in powder form, was introduced into the
boiler section of the vaporizer. The vent cap of the vaporizer was then sealed. The VPCI will remain active for a
period of 24 months as long as it remains sealed. The VPCI will NOT need to be washed from the system before
adding the HTF products when the system is activated.
NOTE
The AQUAVAIRE QV/QM/QH system must have either corrosion
inhibitors or both corrosion inhibitors and freeze protection added
to the waterbath heat transfer fluid (HTF).
Distilled or de-ionized water is recommended for both initial fill and make up due
to evaporation. Tap water should not be used because of dissolved solids and
gases, which can interfere with corrosion inhibitor effectiveness and even
increase inhibitor depletion rate. If distilled or de-ionized water is not available,
water quality should meet the minimum quality requirements in the table below.
FREEZE PROTECTION
If freeze protection is required, Algas-SDI recommends DOWFROST HD as the
preferred freeze protection product. DOWFROST HD is a heavy duty industrial
inhibited PROPYLENE GLYCOL based product and is preferred because the low
acute toxicity is more environmentally friendly than an ethylene glycol-based
product. DOWFROST HD also contains corrosion protection chemicals so none
need to be added.
______
1-2 Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852Q
Calculating Freeze Protection
______
Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852 1-3
CORROSION PROTECTION WITHOUT FREEZE PROTECTION
When freeze protection is not required, corrosion protection must still be used.
Algas-SDI recommends the use of CH2O 6431 HYDRO-TREAT when only
corrosion protection is required. Add this material at the rate of 1 gal to 500
gallons of distilled or deionized water as described in RECOMMENDED WATER
SPECIFICATION on page 1-2.
Circulate waterbath for 24 hours and refer to page 1-11 of this supplement to
record corrosion protection concentrations in your waterbath maintenance
log.
______
1-4 Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852Q
ORDERING INFORMATION
DOW CHEMICAL
Phone: 1-800-447-4369
Web Site: http://www.dow.com/heattrans/gss/inhibited/
CH20
8820 Old Hwy 99 SE
Olympia, WA 98501
Phone: 1-800-562-6184
Fax: 1-360-352-4813
Web Site: www.ch20.com/ch20
Algas-SDI
Phone: 1-206-789-5410
Fax: 1-206-789-5414
Web Site: www.algas-sdi.com
______
Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852 1-5
Maintenance Checklist
Solenoid operation.
Check heat transfer fluid pH, freeze protection and inhibitor levels on a monthly
schedule.
Check heat transfer fluid for algae, fungi and bacteria on a yearly basis.
1. START AND STOP THE BURNER several times to insure proper operation.
Check for proper functioning of low-water cutoff, high limit and operating
control.
______
1-6 Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852Q
CAUTION
Whenever any component or fitting in the system is removed,
replaced or reinstalled, the complete system must be leak tested.
NO LEAKS ARE TOLERABLE!
In addition to maintaining proper heat transfer fluid (HTF) level, pH and corrosion
inhibitor control is also necessary. Water loss by evaporation is natural and
depends on system use and ambient conditions. Monitor water loss for a month
when in use to determine a pattern of loss. Also, check HTF level before using if
the vaporizer has been idle for over 30 days.
pH control is important because glycol can oxidize into acidic end products which
are very corrosive. Dow products contain additives to neutralize these acids
which may become depleted over time. pH should be checked with a pH monitor
or indicating paper. Test kits are provided with each QV/QM/QH, located in the
instrument panel box. Replacement kits can be ordered from Algas-SDI (See
Ordering section for address information).
Test the HTF after treatment and circulating 24 hr, once per week for the
first month and monthly thereafter following the test procedures in the
following section and record the data in the maintenance log shown on page 1-
11. If the pH, freeze protection or corrosion inhibitors (if appropriate to your HTF)
are out of specification, submit two samples to a lab to test for all active
ingredients, pH, conductivity (μS) and the presence of iron. Provide the test
results and a copy of your maintenance log to the supplier of your HTF to
determine the proper corrective action.
Dow provides a free fluid analysis to determine inhibitor condition for systems
containing 250 gallons or more of DOWFROST or DOWTHERM, and a small fee
for lesser amounts. The kit consists of several 4-ounce containers and labels to
fill with fluid samples to send to Dow. It is recommended that you order the kit in
time to take a sample after the vaporizer has been initially filled and circulated for
24 hours. The test results from Dow will include inhibitor and glycol levels and
any necessary recommendations.
Check your HTF for biological activity (i.e. bacteria, algae, and fungi) at least
once per year. A dip slide test kit is provided inside the control panel of your
QV/QM/QH vaporizer.
______
Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852 1-7
NOTE___________________________________________________
Test heat transfer fluid (HTF) before maintenance is performed. The test
strips should be used by the date on the packaging. These test strips are for
maintenance screening for HTF problems. If the HTF is out of acceptable
range, submit additional samples to your test lab to determine the proper
corrective action. This kit will test for products such as molybdate and nitrite
that are not present in DOWFROST HD but are present in CH2O 6431
HYDRO-TREAT. Determine what products are in your HTF before testing.
For best results:
• Start with clean, dry hands and utensils.
______
1-8 Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852Q
• Collect HTF samples from both the top and bottom of the waterbath
tank. HTF must be between 50° F (10° C) and 110° F (43° F) when tested.
Room temperature is preferred.
• Remove one strip from the bottle. DO NOT touch the pads on the end of
the strip. DISCARD STRIP IF NITRITE TEST PAD HAS TURNED BROWN.
• Dip strip in HTF sample for one second, remove, and shake strip briskly
to remove excess liquid.
• 45 SECONDS after dipping the strip, compare and record results in the
following order: (1) Compare FREEZEPOINT (end pad) to color chart and
record the result. (2) Next compare MOLYBDATE (middle pad) to color chart
and record the result which must be between 200 – 300 ppm. (3) Then
compare NITRITE test to color chart and record result which must be
between 250 – 350 ppm. (4) Finally compare pH test to color chart and
record result.
• For best results follow test times carefully. Use a stopwatch or clock with
a sweep second hand. Comparing the test strips to the color chart too soon
or too late may result in incorrect readings and improper treatment and could
result in heat exchanger tubing pitting and vaporizer damage.
• Test for biological activity at least once per year. Dip slides and
instructions are provided inside the control box of your QV/QM/QH vaporizer.
Spare test kits can be ordered from Algas-SDI.
______
Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852 1-9
TESTING FOR BIOLOGICAL CONTAMINATION
1. If ANY fungal or bacteria growth is present, consult HTF supplier for corrective action.
Unscrew the cap and remove the paddle from the vial. Be careful not to touch either of the
paddle’s surfaces.
2. Immerse the paddle into the sample for 2-3 seconds so that both surfaces are covered with
the liquid.
3. Allow excess fluid to drain from the paddle by touching the tip to an absorbent piece of paper
(such as a paper towel).
4. Empty vial of liquid, screw cap back on and incubate in a upright position between 77o – 86o F
(25 – 30o C) for 24 – 36 hr. At that time bacterial side can be read. Fungal growth, if heavy, can
be read at 36 hr. For sparse fungal growth, up to 5 days of incubation may be required.
5. Compare the growth on the paddle to the conversion chart below for a quantitative
interpretation of the results.
______
1-10 Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852Q
______
Algas-SDI Operations and Maintenance Manual Supplement – P/N – 51852 1-11
Algas-SDI International, LLC
151 South Michigan Street
Seattle, Washington 98108
USA
Ph.: 1.206.789.5410
Fax.: 1.206.789.5414
www.algas-sdi.com