REPORT
To : INTERNALTIONAL BREWERY PLC,
SAGAMU OGUN STATE .
Attention : MR SEGUN ADEYENI
From : Simon David
E-mail : info@bct-ng.com
simon.david@bct-ng.com
Direct dial : (+234) 70 693 444 85
Our ref. : 20200219/Eoa/intlbrwplc-sag/sla.38
No. of pages : 4
Date : 29 th March, 2022

Dear Sir,

Please find below the service report for the first biannual 2022 preventive maintenance.

FIRST BIANNUAL 2022 PREVENTIVE MAINTENANCE REPORT


S/No EQUIPMENT QTY COMMENT SERVICE REPORT

1 NIBEM FOAM STABILITY 1 Equipment - Comprehensive service/lubrication of the


TESTER. SN:3536 status ok. Mechanical part, and Calibration of Temperature,
Time and Distance were checked and confirmed Ok,
- Stepper motor current was less than 300mA.
2 INPACK 2000 SAMPLER 1 Equipment - Dismantling, Cleaning, lubrication of O-ring and re-
SN:ISD 5829 status ok. assembling of the equipment was carried out and
the equipment was confirmed ok.
3 INPACK 2000 FLASHER 1 Equipment - Disassembling, cleaning and re-assembling of
SN:2004 status ok. flasher orifice, flashing pins and spring, lubrication
of the guide O-rings and the equipment was
confirmed ok.
4 CO2/O2 GEHALTEMETER C- 2 Equipment - Cleaning of the plunger, throttle valve and the
DGM status was ok. plunger Plexiglas were carried out; O-rings were
SN:12945 lubricated;
SN;12860 - Calibration check of pressure transducer and
temperature sensor was also carried out and the
equipment was confirmed ok.
- However, it is advised that the equipment be
flushed with H2O after every use.
5 VOS ROTA 2.0 HAZE METER 2 Equipment - Cleaning of the sample bath holder and the control
SN: H-VR2-10064 status was ok. knob were carried out, The optic lenses were also
SN; H-VR2-10068 cleaned and confirmed ok,
- Calibration check with distilled water confirms the
accuracy, reproducibility & display stability.
6 REDPOST MONITOR 1 Equipment . this meter has a faulty battery/connector, and

....together to succeed!
(RPU 352) status Not ok needs to be replaced for proper functionality.
SN:98039
RT: 1483
7 REDPOST MONITOR 352 1 Equipment - Service and cleaning of the equipment was carried
SN::98040 status ok. out; Calibration and temperature check with
RT: 3521 Haffmans 60, k Plug, and communication test plug all
confirmed the equipment ok.
- The probe socket should be lubricated after
every 50 runs using the Haffmans contact
treatment grease.
8 REDPOST MONITOR 352 1 Equipment - Service and cleaning of the equipment was carried
SN::98245 status ok. out; Calibration and temperature check with
RT: 470 Haffmans 60, k Plug, and communication test plug all
confirmed the equipment ok.
- The probe socket should be lubricated after every
50 runs using the Haffmans contact treatment
grease.
9 ORBITAL SHAKER 1 Equipment - The speed regulator was checked for
SN:01014617110190 status ok. continuity and confirmed ok, the
- Electric motor bearings and belt were also
confirmed ok, a
A tachometer was used to confirm the accuracy of
the speed.
10 RECIPROCAL ORBITAL 1 Equipment - The speed regulator was checked for
SHAKER status ok. continuity and confirmed ok,
SN:01814127 - Electric motor bearings and belt were also
confirmed ok,
- A tachometer was used to confirm the accuracy of
the speed.
11 FORCED AIR DRYING OVEN 1 Equipment - The temperature accuracy was confirmed ok using
SN:BOV-V225F status ok. a reference thermometer.
- Safety thermostats, relay, cut-off temperature,
circulating fans and the earthling system were all
confirmed ok.
12 ELMA ULTRASONIC BATH Equipment - The temperature accuracy was confirmed ok using
S/N:100525048 status and a reference thermometer.
S/N:100526048 handling ok. - Safety thermostats, relay, cut-off temperature,
circulating fans and the earthling system were all
confirmed ok.
15 BOECO ANALYTICAL 5 Equipment - Service and calibration check with standard
BALANCE status ok weights of 10g, 50g & 100g respectively were carried
SN:553196/17 out and the equipment was confirmed ok.
SN:588227/17
SN:566539/17
SN:553192/17
SN:553195/17
15 HG-80 AUTOCLAVE 2 Equipment - The temperature accuracy was confirmed ok using
SN:32717100290 status ok a reference thermometer at the different points of
SN:32717100284 temperature till the sterilizing temperature.
- Earthling system, safety thermostat and the cut-
off temperature were all confirmed ok.

16 VIBRAC TORQUE METER 1 Equipment - Servicing and cleaning of clamp, sensor unit, drive
SN:18-0215 status ok. unit and column were carried out; earthling systems
were check and confirmed ok.
- Calibrations with known standards weight were
also carried out and the equipment was confirmed
ok.
17 AGR TOPWAVE HOT CUTTER 1 Equipment - The equipment was cleaned and the wiring and the
SN:N/A status ok. outings were checked, the hot cutter were checked
for tension and equipment was confirmed ok.

18 BIOBASE FUME CUPBOARD 1 Equipment - The equipment was cleaned and the wiring and
SN:FH18X1804008 status faulty. outings were checked, and errors corrected.
Control Panel is - The control panel is faulty making it beeping alarm.
weak
19 YAKOS LAMINAR FLOW 3 Equipments - The equipment was cleaned and wiring and
SN:AHA1803020 status ok. outings checked, alarm corrected also.
SN:AHA1803018
SN:AHA1803019
20 FLASH POINT TESTER 1 Equipment - The equipment was checked and cleaned and
SN:GA-1406-EN-2328001 status ok .but outings checked but the ignition is fault.
bad ignition.
21 AGR PGT800T 1 Equipment - The equipment was checked and cleaned and
SN:413 status ok. outings checked also and was confirmed ok.
22 PARR CALORIMETER 1 Equipment - The equipment was checked and cleaned and was
SN:PARR6100-1808-90852 status ok. confirmed ok.

23 MICROBIAL AIR SAMPLER 1 Equipment - Cleaning of the equipment was carried


SN:43125 status ok. out, the rotor, power source and the
printed circuit board were checked and
Confirmed ok.
- A reference tachometer
Was used to check the average speed of the sampler
and the equipment was confirmed ok.
24 ELIX-PROGARD TS2 1 Equipment - Equipment was cleaned through it internal
SN:PROGOTOS2 status ok. maintenance rules and was also cleaned using
chlorine .however it filter is a yearlong span.

25 MILLI-Q 1 Equipment - Equipment was cleaned doing the internal


SN:F8HA8035OA status ok but cleanings and was confirmed ok. but filter expired
filter expired. and expected to be changed.
With Kind Regards,
Blue Chip Technologies Ltd.

Simon David
Service Engineer