3 Spectrometric Performance of CdZnTe Ring Detectors

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

Nuclear Instruments and Methods in Physics Research A 633 (2011) S134–S136

Contents lists available at ScienceDirect

Nuclear Instruments and Methods in


Physics Research A
journal homepage: www.elsevier.com/locate/nima

Spectrometric performance of CdZnTe ring detectors


Anna Bulycheva a,n, Vladimir Kondratjev a, Vladimir Gostilo a, Victor Ivanov b
a
Bruker Baltic, Ganibu dambis 26, Riga, LV-1005, Latvia
b
Ritec, Aizkraukles 23, Riga, LV-1006, Latvia

a r t i c l e in f o a b s t r a c t

Available online 19 June 2010 Spectrometric performance of CdZnTe ring detectors based on the crystals with area 5  5 mm2 and
Keywords: thickness of 0.5–1.5 mm are presented. Fabricated detectors provided the energy resolution 0.58; 0.92;
CdZnTe ring detector 1.63; and 5.14 keV for energies 5.9; 59.6; 122; and 662 keV correspondingly. Peak/background ratio for
Spectrometric performance energy 122 keV was 56.1. The test results of detection unit with CdZnTe ring detectors demonstrate the
principal ability to create small dimension spectrometers with such detectors operating in energy range
of 5–662 keV at ambient temperature up to + 65 1C with high energy resolution and high temperature
stability.
& 2010 Elsevier B.V. All rights reserved.

1. Introduction shaping the detector field. The same was confirmed by experi-
mental testing of the detectors [3,4].
The ring structure of Si detectors was proposed by Kemmer
and Lutz [1] for significant improvement of the spectrometric
performance in low energy range based on the low capacity of 2. Detectors fabrication
collecting electrode and excellent collection of carriers in detector
crystal. Abbene et al. [2], Owens and Kozorezov [3], and For detector fabrication, crystals with an area of 5  5 mm2 and
Kozorezov et al. [4] considered that the ring structure of thickness of 0.5–1.5 mm were used. They had specific resistivity
electrodes can also provide for high performance of CdZnTe of (1.0–10)  10  10 O cm, and (mt)e value of (1.1–1.8)  10  3
detectors for X-ray registration in medical, XRF, and other cm2/V.
applications. They all investigated the electrophysical processes The fabrication was done using the standard technology of
in CdZnTe detectors with such contact structure [2–4]. In this multidetector structures with Au contact deposition [5]. The
work the authors pay more attention to the spectrometric fabricated detector crystal was placed on a hybrid microassembly
performance of such detectors in a wide range of energy and which included a FET 2N4416, a feedback resistor 10 GO, a
temperature conditions to demonstrate their possible utility in feedback capacitor and a temperature sensor PT-100 (Fig. 2a). The
different applications. microassembly with detector was installed on the two stages
In the design of the ring electrodes structure, proposed initially Peltier cooler that provided temperature up to 20 1C in a
by Abbene et al. [2], one side of the crystal includes solid contact standard case TO-8 (16 pin). The cap of the case had a 200-mm-
and the opposite side a ring electrode structure with a point thick Be window to allow low energy range measurements
central electrode for charge collection and two rings for shaping (Fig. 2b).
the electric field in the crystal. The rest of the crystal surface is
overlapped by the guard ring GR.
Such a detector structure can operate in two modes of charge 3. Results and discussions
collection—hemispherical (Fig. 1(a)) and drift (Fig. 1(b)). The
distribution of the force lines in the crystal for each mode are The spectrometric performance of the CdZnTe ring detectors
shown in Fig. 1(a) and (b). was investigated using the complete detection unit. The detection
The figures show that the operative volume is more in drift unit (Fig. 3) is a functional unit comprising preliminary amplifier,
charge collection mode than in hemispherical mode. But, in this HV converter with bias voltage circuits adjustment for detector’s
structure, the geometry of the rings becomes more important in electrodes, system for temperature stabilization of the detector
crystal, the outer cooling system with Peltier cooler that enables
operability of the detection unit at temperatures of up to 65 1C.
n
Corresponding author: Tel.: + 371 6738 3947; fax: +371 6738 2620. More than 10 detectors, each with an area of 5  5 mm2 and
E-mail address: a.bulycheva@bruker-baltic.lv (A. Bulycheva). thickness of 0.5–1.5 mm, were manufactured. The resistance

0168-9002/$ - see front matter & 2010 Elsevier B.V. All rights reserved.
doi:10.1016/j.nima.2010.06.147
A. Bulycheva et al. / Nuclear Instruments and Methods in Physics Research A 633 (2011) S134–S136 S135

Fig. 1. CdZnTe ring detector: (a) hemisphere mode and (b) drift mode.

Fig. 4. Am-241 spectrum of CZT ring detector (T¼  6 1C, shaping time 1.4 ms)
(Res. 59.6–0.92 keV).

Fig. 2. Ring CZT detector: (a) assembly and (b) detector in case TO-8.

Fig. 5. Co-57 spectrum of CdZnTe ring detector (hemispherical mode) (Res. 122–
1.63 keV).

Fig. 3. Ring CZT detector with electronics.


However, the charge collection conditions were not ideal in the
drift mode, and this leads to a low energy tail of the peak of total
absorption.
between the neighboring ring structure elements at 10 V was One reason for this is that the structure of rings is not optimal
higher than 55 GO for all the detectors. The detectors’ bias for the drift mode of detector operation. For more uniform drift
voltages were in the range 100–140 V and their optimum shaping charge collection, more rings are required to shape the field in the
time was within the range 0.8–1.5 ms. The optimum working crystal.
temperature of the detectors was + 20 1C on collection of charge It was also found that detector parameters inside detection
in middle and high energies 20–600 keV, and 0 to  20 1C in unit were stable in a wide range of temperature conditions. The
the low energy range (below 20 keV). Evidently, more cooling centroid peak shift at energy of 122 keV (the external temperature
was required at the registration of low energies to decrease the range 20–65 1C) was 0.2% or 0.0045%/C. The energy resolution
leakage current of detectors and thus to get better energy changes in this temperature range were not more than 3.75%.
resolution at these energies.
The best spectra of Fe-55, Am-241, Co-57 are presented in
Figs. 4 and 5. Energy resolutions 0.58, 0.92, 1.63, and 5.14 keV
were obtained for the detectors at energies of 5.9, 59.6, 122, and 4. Conclusions
662 keV, respectively.
The measurements show that the detectors’ energy resolutions The test results of CdZnTe detector with ring structure
are almost identical in hemispherical and drift modes. But, the demonstrate the feasibility of creating detectors that can operate
registration efficiency at 122 keV is about 1.5 times higher in the in the energy range 5–662 keV with acceptable resolutions of 0.6
drift mode than that in the hemispherical mode. and 6 keV. In the 20–662 keV range, detectors do not practically
The peak/background ratio was 56.1 for hemispherical mode require any cooling to give their best performance. At the same
and 20.8 for the drift mode. time, temperature stabilization of detectors with 20 W power
S136 A. Bulycheva et al. / Nuclear Instruments and Methods in Physics Research A 633 (2011) S134–S136

provides, in the external temperature range 20–65 1C, a centroid Topology optimization, using computer simulation of the optimal
peak displacement of 0.2% at 122 keV energy. The energy structure, should lead to improvement of the CdZnTe ring
resolution changes in this temperature range are less than 3.75%. detectors spectrometric characteristics in the drift mode.
The characteristic feature of the detector is its ability to
operate in both hemispherical and drift modes of charge References
collection. Although the energy resolution is virtually identical
in both cases, the charge collection in drift mode increases by [1] J. Kemmer, G. Lutz, Nucl. Instr. and Meth. A 235 (1987) 365.
almost 1.5 times as compared to that in hemispherical mode. [2] L. Abbene, S. Del Sordo, A. Cola, et al., Nucl. Instr. and Meth. A 583 (2007) 324.
However, the collecting terms in the case of the drift mode are not [3] A. Owens, A.G. Kozorezov, Nucl. Instr. and Meth. A 563 (2006) 31.
[4] A.G. Kozorezov, A. Owens, V. Gostilo, et al., J. Appl. Phys. 102 (2007) 054505.
ideal at the existing topology of rings, and that leads to the [5] V. Gostilo, V. Ivanov, S. Kostenko, et al., Nucl. Instr. and Meth. A 460 (1) (2001)
appearance of low energy tail of the peak of total absorption. 27.

You might also like