ST-200Pro Service Manual (6 Channel) Final

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 46

SERVICE MANUAL

ST-200Pro Service Manual

ST-200Pro
Na/K/iCa/Li/pH/Cl Electrolyte Analyzer

SERVICE MANUAL

SENSA CORE MEDICAL INSTRUMENTATION PVT. LTD.


Plot No.3, Export Promotion Industrial Park,
Pashamylaram, Sangareddy (Dist), Hyderabad, Telangana, 502307.
www.sensacore.com, Tel: +08455 223400.
service@sensacore.com
1
ST-200Pro Service Manual

INDEX

I. RESPONSIBILITY DECLARATION 5

II. SAFETY PRECAUTIONS 5

III. ELECTRICAL FEATURES AND LIN DISTURBANCE 6

1. GENERAL DESCRIPTION 7

1.1. Foreword 7

1.2. Technical Specifications (Na/K/iCa/Li/pH/Cl) 7

2. INSTALLATION AND PRELIMINARY OPERATIONS 10


2.1. Introduction 10
2.2. Transportation, Storage and Unpacking 10
2.2. a. Composition of the Supply 11
2.3. Preparation of the Place of Installation 11
2.4. Electrode Staging 12
3. COMPONENTS OF ELECTROLYTE ANALYZER 14
3.1. Fluidic Circuit Parts 14
3.1.1. Electrode Housing 15
3.1.2. Valve Module 16
3.1.3. Sampler 16
3.1.4. Pump Module 19
3.2. Electronic Parts 21

3.2.1. Mother Board 21


3.2.1.i. Mother Driver Card 23
3.2.1.ii. Pre –Amplifier 23
3.2.2. RFID 23
3.2.3. Printer 24
3.2.4. Display 25
3.2.5. Sample Detector 26
3.2.6. Rear Panel 27
4. MAINTENANCE 28
4.1. External Cleaning of the Instrument 28
4.2. Cleaning Flow Path 28
5. TROUBLE SHOOTING 29

2
ST-200Pro Service Manual

5.1. Reagent Packs Checking 29


5.2. Electrode Housing 29
5.3 Cleaning Sampler Mechanism 30
5.4 Procedure to Clean Valve Tubing 31
5.5. Electrode Cleaning 35
6. ERROR MESSAGES 36
6.1. 234.16 mV or ± 435 mV’s Error 36
6.2. Drift Error 37
7. SOFTWARE UP-DATE 39
8 QUICK ACCESS SHEET 44

3
ST-200Pro Service Manual

FIGURE’S INDEX

Fig: 1 Front View of the Analyzer 9

Fig: 2 Electrode Arrangements 13


Fig: 3 Fluidic Circuit 14
Fig: 4 Electrode Housing 15
Fig: 5 Models of Electrolyte Analyzer 15
Fig: 6 Valve Module 16
Fig: 7 Sampler 17
Fig: 8 Probe Seal Replacement 18
Fig: 9 Probe Seal and Seal Holder 18
Fig: 10 Pump Module 19
Fig: 11 Pump tube installation procedure 20
Fig: 12 Mother Board 21
Fig: 13 Test points voltages 22
Fig: 14 Motor Driver Card 23
Fig: 15 Pre-Amplifier 23
Fig: 16 RFID 24
Fig: 17 Printer Driver 24
Fig: 18 Replacing Printer Paper Roll 25
Fig: 19 Paper Roll Installation 25
Fig: 20 Display 25
Fig: 21 Sample Detector 26
Fig: 22 Sample Detector Connector 26
Fig: 23 Rear Panel 27
Fig: 24 Flush 30
Fig: 25 Fluid control valves 31
Fig: 26 Reference housing 36
Fig: 27 MSP430 J-Tag 39
Fig: 28 Electrodes staging 43

4
ST-200Pro Service Manual

I. RESPONSIBILITY DECLARATION

Our Company declines all responsibility for damage due to any type of
modifications made on the Hardware and Software due to connection to
other instruments not carried out by our personnel or not previously
authorized by our firm.
In case the apparatus is damaged, do not switch it on until it had been
repaired by a technician from our company’s Service Department.
Any type of electric operation required to install or repair the instrument
must be carried out by our staff; do not install in any other way.
The instrument should be used under the environmental conditions and
according to safety norms given in points 2 and 3 of the following
chapter, in order to guarantee that the results obtained by the
instrument are equivalent to those given in the technical specifications.

WARNING

USE THE INSTRUMENT ONLY AFTER READING THIS MANUAL

II. SAFETY PRECAUTIONS

The ST-200Pro guarantees the maximum safety to the operator during


its operation. However, as there is internal voltage, the operator must
follow the general safety rules:

 Make sure that the Power Supply has the required voltage before
switching ON the instrument.

 The instrument must have a ground connection. In case of an


electric short circuit the ground connection eliminates the danger of
electric contact with the external metal parts.

 In case of a current surge (due to a short circuit or another


malfunction within the instrument or due to sudden line voltage
variation) the instrument must be disconnected and the Service
Engineer must be contacted.

 Do not use the commands or adjustments inside the instrument and


notify the service staff as soon as any anomaly occurs during
functioning.

 Before starting any of intervention inside the instrument switches it


OFF and disconnects the power cable. Unplug the power cable to
disconnect it; do not pull the cable.

 Any repair of the instrument must be carried out by the Technical


Staff.

 Any electric operation that is necessary to the installation must be


carried out by qualified staff.
5
ST-200Pro Service Manual

 If the instrument is damaged do not switch it ON until it has been


repaired by a Technician of the Factory Assistance Service.

 Disconnect the instrument if it left inactive for a long period.

III. ELECTRICAL FEATURES AND LINE DISTURBANCE

Disturbances on the power line may be caused by several factors such


as:
 Starting and stop any of electric motors.
 Atmospheric phenomena.
 Connection and disconnection of high line loads.
 Generator and square waves inverters.

We must stress that serious power supply line disturbances have


negative effects on the system.
An efficient ground connection is necessary to reduce line disturbances
and guarantee the maximum safety to the operator.

Distribution : 1 phase, neutral and ground

Power supply : 110 – 220 VAC 50 – 60 Hz

Consumption : 80 watt

6
ST-200Pro Service Manual

GENERAL DESCRIPTION

1. GENERAL DESCRIPTION

1.1. FOREWORD

The ST-200Pro is an automated, microprocessor-controlled analyzer for


measurement of Sodium, Potassium, Ionized Calcium, Lithium, pH and
Chloride in serum, plasma, whole blood, cerebrospinal fluid and diluted
urine.
To obtain accurate results, the ST-200Pro must be operated with
specially packaged calibrant.
The instrument is fitted with the flow system of Electrode housing
which ensures a low level of carry over even with limited sample volume.
The Pump guarantees a high level of precision in the aspirated
volume.
The instrument is having 7 inch high definition LCD screen with
capacitive touch to enter the settings requested on the screen. The LCD
screen shows instrument status and error or malfunctioning signals.

The test results are shown at the results screen and are printed on
heat-sensitive paper.

1.2. TECHNICAL SPECIFICATIONS (Na/K/iCa/Cl/Li/pH)

Sample : Whole heparin blood, Serum, Plasma,


Cerebrospinal fluid & Diluted Urine.

Sample size : 140 micro liters of Whole heparin blood,


Serum, Cerebrospinal fluid, Plasma & 500
Micro liters of diluted (1:5) Urine.

Method : Direct Measurement with Ion Selective Electrode


(ISE)

Parameter : This functionality allows the change of analysis.

Detection Range : Blood


Na+ : 20.0 – 250.0 mmol/L
K+ : 0.2 – 40.0 mmol/L
iCa++ : 0.0 – 10.0 mmol/L
Cl- : 20.0 – 200.0 mmol/L
Li+ : 0.0 – 5.0 mmol/L
pH : 5.0 – 9.0

Urine
Na+ : 35 – 1000 mmol/L
K+ : 50 – 200 mmol/L
Cl- : 25 – 500 mmol/L
7
ST-200Pro Service Manual

Performance: Blood, Serum and Plasma

Na+ : 20.0 – 250.0 mmol/L


K+ : 0.2 – 40.0 mmol/L
iCa ++ : 0.0 – 10.0 mmol/L
Cl- : 20.0 – 200.0 mmol/L
Li++ : 0.0 – 5.0 mmol/L
pH : 5.0 – 9.0

If iCa unit is in mg\dL, then the measurement range is 0-40 mg\dL.

Reproducibility (serum) Performed according to a modified NCCLS protocol EP5-T2

Reproducibility (expected) C.V. <1% C.V. <2% C.V. <2%


Day to day (n=20) C.V. <2% C.V. <2.5% C.V. <2.5%

URINE
Reproducibility Na+ : 35 – 1000 mmol/L
(Urine STD.) K+ : 50 – 200 mmol/L
Cl- : 25 – 500 mmol/L

Reproducibility (expected)
C.V<5% C.V<5% C.V<5%
Day to day (n=20)

Analysis Time : 65 sec. (Blood) / 80 sec. (Urine)

Speed : 55 samples/hour.

Data Storage : 2,00,000+ Patient results.

Conditioning : Temperature: 50C - 400C, <85% moisture

Display : 7 inch high definition LCD with capacitive


touch.

Print : 2 inch 24 column thermal printer.

Input voltage : 100/115~V AC, 50-60 Hz, or 220~V AC.

Dimensions : 15” W x 12” H x 7” D,


Weight: 4.3 Kg

Battery : Optional support for 10 hours.

OS : Android. (WiFi enabled application


up-gradable)

8
Front Door
Printer
Power Bubble
Button Detector

Sample
Tube
Display
Flip/Sampler

Reference Housing
Reference
Membrane
ST–200Pro Electrolyte Analyzer:

Pump Tube

Reagent Pack
ST-200Pro Service Manual

Pump Module

Fig: 1 Front View of the analyzer


Valve Module

9
ST-200Pro Service Manual

INSTALLATION AND PRELIMINARY OPERATION

2. INSTALLATION AND PRELIMINARY OPERATIONS


2.1. INTRODUCTION
The instrument is fitted for power supply of 110 – 220 V AC, 50 – 60 Hz
or 12 volt 3.3 amps DC battery.

2.2. TRANSPORATION, STORAGE AND UNPACKING


There are certain precautions to keep in mind during transportation,
storage and unpacking of the instrument in order to avoid damages also
due to storage in an improper environment.

TRANSPORTATION
In package form the instrument weighs about 14.5 kg and about 4.3 kg
without packaging,

No special equipment is therefore required for transportation, loading or


unloading the instrument in its box.

When handling the equipment the warning on the outside of the


packaging (TOP – BOTTOM, HANDLE WITH CARE) must be respected.

STORAGE
The equipment should be stored in its original packaging consisting of a
one box, and stored in respect of the warnings applied to the outside
(TOP – BOTTOM, HANDLE WITH CARE).

If during shipment the packaging is damaged it is necessary to check it


at least to make sure the instrument has not undergone evident damage;
the packaging should then be repaired before storage. In case of damage,
notify the SUPPLIER.

The type of packaging ensures reliable protection and insulation if the


crate is deposited in a suitable place; it is necessary to store it in a dry
place that is free of dust and vibration.

UNPACKING
Unpacking is a simple operation. Unpack carefully. Keep the cases, in
case the instrument should have to be shipped back. Lift the instrument
at the sides, placing your hands near the edges of the base plate. Keep
the instrument vertical at all times.

10
ST-200Pro Service Manual

2.2. a. COMPOSITION OF THE SUPPLY

Proceed to unpack the instrument. The following accessories are


supplied with it:

S.No. Description
1 Thermal Paper Rolls
2 Troubleshooting kit
3 Reagent Pack
Low, Normal, High QC Controls for blood & Level 1, Level 2
4
controls for Urine.
5 Urine Diluent
6 Daily Cleaner Set
7 Waste container
8 Reference Inner Filling Solution

2.3. PREPARATION OF THE PLACE OF INSTALLATION

The operative requirement is very important for the choice of the room
where the installation should be effected.
The ST-200Pro overall dimensions are 15” W X 12” H X 7.0” D in any
case, in order to make it easier the maintenance procedures, it would be
better to leave some space around it.
The primary activities for installation of the instrument are:

 Identifying a person to put in charge of the activities necessary for


installation, who will contact the supplier for any explanations that
may be necessary.
 Selection of the place of installation.
 Purchase of any accessories (supports, etc.) other than those
supplied with the instrument.
Although the Electrolyte analyzer has been designed with components
capable of operating in a hostile environment, it is a good idea to check
the environment conditions in order to ensure more reliable
performance.

 High temperatures can age the parts and produce temporary or even
permanent alterations.

 A particular dusty environment may cause an abrasive action on the


components and therefore reduce their life.

 Vibrations can cause errors in the measurement made by the


instrument.

11
ST-200Pro Service Manual

 High frequency and high intensity pulses generated by the


electronic devices or induced by surroundings may cause errors in
the system.
 Do not place the instrument near any heat source like radiators, hot
air tubes or in places directly lighted by sun rays.
 Avoid storing in places subject to sudden temperature changes.

A power outlet should be available in the place of installation. It should


have the following characteristics:

Distribution 1 phase, neutral and ground


Power Supply 110-220 VAC
Consumption 80 Watt

A cable with three wires and a plug with three contacts are supplied
with each instrument, for connection to the power mains and
grounding. Never connect the instrument to a socket without
grounding.

If the wiring is not grounded according to regulations, this may


constitute a hazard for the personnel and may be a source of
malfunction of the instrument.

Environmental features of the place of installation:

10 to 45ºC (Operating)
Temperature
0 to 50ºC(Non Operating)
20 to 85ºC (Operating)
Relative Humidity
0 to 90ºC (Non Operating)
Altitude <2000 m (Operating)

2.4. ELECTRODE STAGING:


Electrode staging plays very important part while installation. Steps to
do Electrode Staging as per the required model.

1. Remove Electrodes & Reference housing from Reference housing


kit.
2. Remove electrode seals from Troubleshooting kit which is in
Reference housing kit.
3. Check the model depend upon your requirement by doing below
steps, press
MENU

PARAMETERS

Check the model which you want to select depending on those models
stag electrodes, by using Electrode seals as shown in below fig

NOTE: Where * indicates Blank electrode


12
ST-200Pro Service Manual

1. K Na * * * *
2. K Na * * * Cl
3. K Na iCa * * *
4. K Na iCa * * Cl
5. K Na iCa Li * *
6. K Na iCa pH * *
7. K Na iCa Li pH *
8. K Na iCa pH * Cl
9. K Na iCa Li * Cl
10.K Na iCa Li pH Cl

Fig: 2 Electrode arrangements

13
ST-200Pro Service Manual

3. COMPONENTS OF ELECTROLYTE ANALYZER


In the following pages describes each part of Electrolyte Analyzer.

3.1. FLUIDIC CIRCUIT PARTS


The following figure shows the main parts constituting the fluidic circuit
for aspiration of a sample from the sample probe.

1. Electrode Housing
2. Valve module
3. Sampler
4. Pump Module

Fig: 3 Fluidic Circuit


14
ST-200Pro Service Manual

3.1.1. ELECTRODE HOUSING

All Electrodes are staging in Reference housing. It consist of Potassium


chloride (KCl) solution.

Electrode housing consists of Electrodes & Bubble detector staged by


using Electrode seals. The following figure shows the electrode housing.

Fig: 4 Electrode Housing

Electrodes should be staged as mentioned in below figure depending


upon the model.

Fig: 5 Models of Electrolyte Analyzer


15
ST-200Pro Service Manual

MODEL 1 -- K/ Na
MODEL 2 -- K/ Na /Cl
MODEL 3 -- K/ Na /iCa
MODEL 4 -- K/ Na /iCa /pH
MODEL 5 -- K/ Na /iCa /Cl
MODEL 6 -- K/ Na /iCa /Li
MODEL 7 -- K/ Na /iCa /Li/pH
MODEL 8 -- K/ Na /iCa /pH/Cl
MODEL 9 -- K/ Na /iCa /Li/Cl
MODEL 10 -- K/ Na /iCa /Li /pH/Cl

3.1.2. VALVE MODULE:


Valve module is mounted in ST-200Pro for the supply of Calibrant A, Calibrant
B & solutions as per requirement.

Valve 1 for A/B calibrant solution.


Valve 2 for AIR & output of Valve 1.

Fig: 6 Valve Module

Valve module handling:

For removing or replacement of Valve module should be done by


using proper star screw driver if not screw driver will damage the valve
holder.
While tightening valve’s holder screws there should not apply extra
pressure and it should not be super tight. Every time valves and its
holders should be handled gently.

2.5.2 Sampler
Sampler or sample probe is used to intake sample, which consist
of a hole at the bottom of the probe. It has a self probe wiping facility
with three different aspiration modes (sample tube, sample syringe
and capillary).

16
ST-200Pro Service Manual

Fig: 7 Sampler

Sampler seal replacement

 Lift the sampler cover towards an upward direction.


 Rotate the seal holder in an anti-clockwise direction till it opens.
 Remove the seal out from the probe and dispose.

17
ST-200Pro Service Manual

Fig: 8 Probe Seal Replacement

Fixing probe seal:


Similarly, insert the new probe seal into the probe with respective
direction and insert the seal holder into the probe and rotate to
clockwise direction till it tights.

Probe seal Seal holder

Fig: 9 Probe seal & Seal holder

Note:
 Improper alignment of probe seal and seal holder may cause flow
errors.

18
ST-200Pro Service Manual

3.1.4. PUMP MODULE:


The pump tube has two stoppers. One stopper (red) is connected at the
base of the electrode housing and the other stopper (blue) is connected
to the reagent pack socket i.e. to the waste.
Pump module is used to generate vacuum in pump tube for
aspirating liquid. Pump module consist of four rollers.

Front View Pump Assembly Modules

Fig: 10 Pump module

Remove pump tubing:

To remove old tubing follow this procedure:

 Remove pump tube from the bottom of the Reference Housing &
waste pipe.
 Remove the front cover, rollers and pump tubing from the pump
module base.

New Pump tubing installation procedure:

1. Clean the module base: It is fixed to the chassis through a lock at


the top of base as shown in the figure. We need to clean the module base
and then apply the grease to inner surroundings softly.

19
ST-200Pro Service Manual

2. Apply grease to rollers: Take a four rollers set and apply grease
gently to rollers. The rollers set consists a grove at its bottom and a knob
on the front side. Insert grove side of the rollers set to the module base
knob.
3. Apply grease to new pump tubing: Take a new pump tubing and
apply grease all around it and then insert it as per the direction by
holding the roller firmly as shown in the figure.
4. Module cover: Close the module with front cover. So that rollers and
pump tubing will lock. Ensure that the tag of pump tubing should be
above the front cover lock.

Fig: 11 Pump tube installation procedure


20
ST-200Pro Service Manual

3.1. ELECTRONIC PARTS:

Electrolyte analyzer consist of following electronic parts


1. Mother board
i. Daughter card
ii. Motor driver card
iii. Pre amplifier
2. RFID
3. Printer
4. Display
5. Keypad
6. Sample Detector
7. Optical Sensor
8. Rear Panel

3.2.1. MOTHER BOARD:

All the electronic circuits processing the analogical and digital signals
and driving the motors and sensors on boards.

Fig: 12 Mother board

Mother board gives proper supply to each and every component


mounted on PCB by using Voltage Regulators. Mother board takes 12V
DC power supply as input.

Trouble shooting:
 Check the input connector of main board for +12 VDC
 Check the voltages at IC when the regulators are giving lower than
the actual voltages.
21
ST-200Pro Service Manual

Specification of Fuses:
Mother board consist of 6 fuses to protect IC’s on the mother board when
fluctuations are occurred.

Respective
Voltage on
S.No: fuse Problems in fuse fails
Mother Board
specification
There will be no power supply
+3.3 VDC at for Micro controller so the
1. 2 Amp
TP2 (F1) application will not work & no
function occurs.
+5 VDC at TP3
2. 2 Amp Printer will not work.
(F2)
Bubble detector, RFID PCB,
+5 VDC at TP4
3. 5 Amp Pump motor, Pinch valves (V1
(F3)
& V2) will not work.
+12 VDC at
4. 5 Amp Display will not work.
TP5 (F4)
+12 VDC at Pump motor, Pinch valves (V1
5. 5 Amp
TP6 (F5) & V2) will not work.
Fan will work but instrument
6. +12 VDC (F6) 5 Amp
will be in OFF condition.

There are no serviceable/replaceable components on this board.

Fig: 13 Test points voltages

 Check the voltages on mother board at respective test points.

Depending upon the problem change the fuse or reinsert the fuse.
22
ST-200Pro Service Manual

3.2.1. i. MOTOR DRIVER CARD:


Motor driver card is used to operate the pump motor & Unislider, which
takes 12 V DC power supply.

Fig: 14 Motor driver card

3.2.1.ii. PRE AMPLIFIER:


Pre amplifier is used to measure the voltages near every channel
(electrode). Pre amplifier consist of 6 channels & 1 reference channel.
It takes 5V D.C power supply.

Fig: 15 Pre amplifier

3.2.2. RFID (RADIO FREQUENCY IDENTIFICATION SYSTEM):


RFID is used to detect the reagent pack inserted in Electrolyte analyzer.
It will take 5 VOLTS DC. Supply as input.
1) After giving input 5 volts RED LED presents on the circuit glows &
when label on the reagent pack is detected ORANGE LED glows.
2) Ground to R56:- Voltage should be 5 volts.

3) IC2 {m430f156}:- 3 volts on 1st pin.


23
ST-200Pro Service Manual

Fig: 16 RFID

Trouble shooting:
 Check the red light for checking + 5V supply.
 Check for orange light if reagent pack detected.

3.2.3. PRINTER:
Besides printing the analysis result and the slope, the printer also relays
a series of messages to the operator. These will be indicated in the
subsequent detailed explanation of Operating Methods. Printer takes 5V
D.C power supply.

Fig: 17 Printer Driver


Trouble shooting:
 Set printer OFF when not required.
 Thermal paper is must in printer mechanism even when printer is
not in use or in OFF condition.

REPLACING PRINTER PAPER


In order to open the printer door lift the left side lock upward direction.

24
ST-200Pro Service Manual

Fig: 18 Replacing Printer Paper roll


To install the roll of paper into the printer, cut off the beginning section
of paper that has tape or adhesives. After installing the paper, close the
cover on the top of the housing until the door gets locked.

Fig: 19 Paper roll installation

Printer mechanism should have thermal paper roll loaded compulsory


to increase life of printer even printer is not in use.

3.2.4. Display:
ST-200Pro analyzer consists of 7 inch high definition LCD with
capacitive touch. It takes +12VDC input supply from the voltage
regulator and communicates with the microcontroller via OTG.

Power button

Fig: 20 Display
25
ST-200Pro Service Manual

It guides user through all aspects of analyzer operation, including


analyzer setup, information entry and display, and testing. The screen is
powered-on by pressing the power button of the touchscreen panel.
Trouble shooting:
 Avoid loose connections.

3.2.5. SAMPLE DETECTOR:


Sample detector is used to check bubbles in the flow. It is very important
to check clear flow of electrolyte analyzer.

Fig: 21 Sample detector

Directly we can check Bubble detector by pressing


MENU

DIAGNOSTICS

HARDWARE

SAMPLE DETECTOR
Trouble shooting:
 Flush periodically with Daily cleaner.
 Avoid loose connections.

Sample detector connector:

Fig: 22 Sample detector connector

Trouble shooting:
 Check the connector direction correctly
 Avoid loose connections.
26
ST-200Pro Service Manual

3.2.6. REAR PANEL:

Air Vent
USB Ports

Power Button

RS-232 Port

LIS USB Port

Grounding Port

Power Socket

J-Tag Port

Fig: 23 Rear Panel

27
ST-200Pro Service Manual

MAINTENANCE

4. MAINTENANCE:
Periodic maintenance will keep the instrument in perfect working order.
We recommend the following procedures:

4.1. EXTERNAL CLEANING OF THE INSTRUMENT:


 Clean the instrument’s external surfaces, using a soft brush to
remove dust.
 Remove any traces of dirt using a slightly damp cloth.

Note: Do not use compressed air in cleaning procedures, or Solvent


which may damage the surfaces.

4.2. CLEANING FLOW PATH:


Perform daily cleaner everyday to maintain the flow path without any
deposition. Follow given procedure.

Press
MENU DIAGNOSTICS DAILY CLEANER

Maintenance:

All other maintenance, trouble shooting or Servicing is performed by


replacing the components as described in the schedule below. When you
wish to perform routine or any other maintenance function, Follow the
given procedure bellow.
Press
MENU DIAGNOSTICS MAINTENANCE

28
ST-200Pro Service Manual

TROUBLESHOOTING

5. TROUBLE SHOOTING:

In troubleshooting Flow Path Error is the main error mainly it will cause
due to Protein deposition.

By cleaning the flow step by step flow path will be clear.

We have to concentrate on below parts to identify and to clear the flow


path.

1) Electrode housing
2) Sampler i.e. probe mechanism
3) Reagent pack
4) Valve module

Follow below steps to clear the flow path

5.1. REAGENT PACKS CHECKING:

Perform priming if you found that liquid is not coming then check
Reagent Pack consumption.

MENU → DIAGNOSTICS → REAGENT PACK

If it shows Solution left as 0%, then change the Pack. Make a note that
Sometimes if Solution left shows <20% also then there will be a chance
of empty Reagent Pack.

5.2. ELECTRODE HOUSING:


If Reagent pack is not empty but still it liquid is not coming out then. Go
to Maintenance mode.

 Remove Sample tube from Bubble detector side & remove the Pump
tube from the bottom of Reference housing and directly connect
Sample tube to pump tube i.e we Bypassed Reference Housing as
shown in figure.
 If liquid is coming out then remove electrode housing and remove the
stag of electrodes along with sample detector from the reference
housing and flush it with distilled water 2 times or daily cleaner.
Don’t forget to flush air several times after distilled water or daily
cleaner. As shown in fig.

29
ST-200Pro Service Manual

Fig: 24 Flush

 Insert all staged electrodes into the reference housing after cleaning
inner silicon seals individually and repeat above procedure with air
flush and then connect the electrode housing in the instrument.

5.3. CLEANING SAMPLER MECHANISM

If you found that there is a flow problem in Sampler then enter in


to maintenance mode, remove the peristaltic tube from the bottom of
the sampler, remove the sample tube from the bubble detector,
connect it to the bold needle & flush it with syringe by using DI water
or Daily cleaner without lifting the flip and then change the probe
seal.

Removing probe seal:


 Lift the sampler cover towards upward direction.
 Rotate the seal holder in anti clock wise direction till it opens.
 Remove the seal out from the probe and dispose.

Fixing probe seal:


 Insert the new probe seal in to the probe.
 Insert the seal holder in to the probe and rotate to clock wise
direction till it tights.

Note:
Improper alignment of probe seal and seal holder may cause flow
errors.

 Close the sampler mechanism finally do reagent priming test as


follows.

Menu Diagnostics Flow Reagent priming Prime All

Still if you are not able to see the liquid in sample tube then clean
the Valve module tubing by using VALVE TEST option in menu.

30
ST-200Pro Service Manual

5.4. PROCEDURE TO CLEAN VALVE TUBING

Valve Module consists of two valves, valve 1 & valve 2 respectively.


Valve 1 is used to separate CAL A & CAL B flow and Valve 2 is used to
take air and valve 1 output

 First remove CAL-A & CAL-B tubes from thermal system.


 Take out whole valve tubing from valves (V1 & V2) slots.
 Take syringe with DI water.
 Flush whole tubing with DI water in order to clear all the salt
formation in flow path.
 Finally flush whole tubing with air and then reinsert in respective
positions at valves flow path tube slots.

Note:
Pinch valve tubing should be totally inserted in the valve towards
inner side. Partial or incomplete inserted tube can let solutions mix
up.

Fig: 25 Fluid control valves

The valves should control standards flow by using the close or


open movements of lock pins.

If getting standards milli volts equally or ‘Air in CAL-B’ error then


do the VALVE TEST by using below steps.

Press Menu Diagnostics Flow Valve test

By performing this test, we can identify the salt blockage in the


tubing and also we can identify the functioning of individual valves.

Note:
At an ideal state A (V1), Air (V2) slots are in open condition
and obviously B (V1) & A/B (V2) slots are in close condition.

31
ST-200Pro Service Manual

CAL-A flow:
Check Cal A flow through valves as per its respective
functionality. It activates Cal A flow and provides 300 seconds time
duration to check its operation. When the time gets completed, it
deactivates automatically or the user can off it by press the Stop
button.

Observe the lock pin movements of V2.


 V1 lock will not move because already the CAL-A slot is in open
condition and CAL-B slot is in close condition.
 V2 lock closes ‘Air’ slot and opens ‘A/B’ slot.
CAL-A flow checking:
 Take syringe with bold needle.
 Remove the pinch valve tubing end point present at the bottom of
sampler and connect it to needle/syringe.
 Pull the syringe and take out the solution (CAL-A) in to it before
counter reaches to ‘0’.
 If the CAL-A solution comes out freely in to the needle that means
the valves V2 is functioning and there is no blockage in CAL-A
flow path.
 If the CAL-A solution does not comes out and syringe goes
reverse flow then the reasons are either may valves V1 or V2 are
not functioning or there is a blockage in CAL-A flow path. Hence,
flush the tubing with DI water, Air and check the pinch valves
connectors (heater 1 & heater 2) on Motor driver card.
32
ST-200Pro Service Manual

CAL-B flow:
Check Cal B flow through valves as per its respective
functionality. It activates Cal B flow and provides 300 seconds time
duration to check its operation. When the time gets completed, it
deactivates automatically or the user can off it by press the Stop
button.

Observe the lock pins movements of both V1 & V2.


 V1 lock closes CAL-A slot and opens CAL-B slot.
 V2 closes ‘Air’ slot and opens ‘A/B’ slot.
CAL-B flow checking:
 Take syringe with bold needle.
 Remove the pinch valve tubing end point present at the bottom of
sampler and connect it to needle/syringe.
 Pull the syringe and take out the solution (CAL-B) in to it before
counter reaches to ‘0’.
 If the CAL-B solution comes out freely in to the needle that means
the valves V1 & V2 are functioning and there is no blockage in
CAL-B flow path.
 If the CAL-B solution does not comes out and syringe goes
reverse flow then the reasons are either may valve V2 is not
functioning or there is a blockage in CAL-B flow path. Hence,
flush the tubing with DI water, Air and check the pinch valves
connectors (heater 1 & heater 2) on Motor driver card.

CAL-A/B IN flow:
Check Cal A/B IN flow through valves as per its respective
functionality. It activates Cal A/B IN flow and provides 300 seconds
time duration to check its operation. When the time gets completed,
it deactivates automatically or the user can off it by press the Stop
button.

33
ST-200Pro Service Manual

Observe the lock pin movements of V2.


 V2 lock closes ‘Air’ slot and opens ‘A/B’ slot.
CAL-A/B IN flow checking:
 Take syringe with bold needle.
 Remove the pinch valve tubing end point present at the bottom of
fill port and connect it to needle/syringe.
 Pull the syringe and take out the solution in to it before counter
reaches to ‘0’.
 If the solution comes out freely in to the needle that means the
valves V1 & V2 are functioning and there is no blockage in flow
path.
 If the solution does not comes out and syringe goes reverse flow
then the reasons are either may valve V1 or V2 are not
functioning or there is a blockage in flow path. Hence, flush the
tubing with DI water, Air and check the pinch valves connectors
(heater 1 & heater 2) on Motor driver card.

Air flow:
Check Air flow through valves as per its respective functionality.
It activates Air flow and provides 300 seconds time duration to check
its operation. When the time gets completed, it deactivates
automatically or the user can off it by press the Stop button.

Observe the lock pins movements of V1 & V2. There are no


movements and are in ideal states.
34
ST-200Pro Service Manual

Air flow checking:


 Take syringe with bold needle.
 Remove the pinch valve tubing end point present at the bottom of
fill port and connect it to needle/syringe.
 Pull the syringe and take out Air in to it before counter reaches to
‘0’.
 If the Air comes out freely in to the needle that means the valves
V1 & V2 are functioning and there is no blockage in Air flow path.
 If the Air does not comes out and syringe goes reverse flow then
the reasons are either may valve V1 or V2 are not functioning or
there is a blockage in Air flow path. Hence, flush the tubing with
DI water, Air and check the pinch valves connectors (heater 1 &
heater 2) on Motor driver card.

5.5. ELECTRODE CLEANING:


After performing calibration if slopes are not in the range or giving ****
even liquid flow is perfect, then clean electrodes by using below
mentioned methods
 Run Daily cleaner.
 Then also it is not giving good results then flush electrode stag using
DI water.
 Separate each electrode individually and clean Na, K and iCa
electrodes with the edge needle (spring needle) provided in service
tool kit along with the instrument to rub protein which has grip on
the membrane of the electrode and plain needle in Cl, Li and pH
electrodes. Don’t use edge needle in Cl, Li and pH electrodes.

CAUTION: Don’t use the spring needle which is already used for Cl
electrode to any other electrodes. And don’t use spring needle several
times to clean the electrodes which may cause damage to electrode
membrane.

After doing every step please calibrate the instrument two times with
print mV is ON.

Calibration slope ranges: If it is in the following range

Na : Should be in between100 to 35 mV/decade


K : Should be in between 100 to 35 mV/decade
iCa: Should be in between 50 to 15 mV/decade
Li : Should be in between 100 to 30 mV/decade
pH : Should be in between 100 to 30 mV/decade
Cl : Should be in between 100 to 30 mV/decade

Then go for sample analysis or replace faulty electrode.

35
ST-200Pro Service Manual

6. ERROR MESSAGES:
6.1. 234.16 mV’s or ± 435 mV’s error:

If you are getting all CAL-B & CAL-A mV’s as 234.16 mV’s or ± 435 mV’s

Then look at left down corner of the reference slightly so that you can see
the reference hole having any bubbles. If bubbles are present in the pin
hole, knock it or slightly shake up side down the reference to remove the
bubbles from the hole. If bubbles are completely removed then calibrate
the instrument it will work.

Fig: 26 Reference housing

Also check any salt coming from silver pin side. If yes clean it & try to
place the reference in the instrument & check mVs.

See reference IFS (Inner Filling Solution) is leak from any side. If yes
don't use it.

 After trying all the above steps and then check reference Pogo pin.
And make sure that it should be tight
 Then also problem remain same, then Reinsert Pre amplifier board
on the mother board, and make sure that no loose connections on
the Pre amplifier board. Reinsert the connections on the Preamp.

36
ST-200Pro Service Manual

6.2. DRIFT ERROR:


Instrument will always crosscheck with previous calibration mV’s .If
the difference between previous calibration mV’s and recent calibration
is more than 10 it will flag DRIFT, repeat calibration.

Refer take care sheet provided with instrument.

Error Messages Cause Action

Liquid is there in
sample detector or Clean the sample
AIR NOT DETECTED
Sample detector is not detector and recheck.
connected properly.

Occurs at 85% pack


INSUFFICIENT PACK
usage. Pack must be Check reagent pack.
PLEASE REPLACE
replaced.
LIQUID NOT Liquid is not present in Clean the sample
DETECTED sample detector. detector and recheck.

PACK EXPIRED Reagent pack expired Check date & time

Sample not aspirated or


Insufficient sample is
Aspirate sufficient
Air in Sample aspirated or sample is
clear sample before
not aspirated in 60 sec.
60 sec.
or hazy sample is
aspirated.

Check RFID
RFID system is not
Connection error connectors or replace
working.
RFID board.
Aspirate sufficient
Cleaning solution is not
Air in Cleaner daily cleaning
detected correctly.
solution.
Check reagent pack
consumption and do
Calibration A solution is
Air In CAL A prime CAL A check
not detected correctly.
weather liquid is
coming out or not.
Check reagent pack
consumption and do
Calibration B solution is
Air In CAL B prime CAL B check
not detected correctly.
weather liquid is
coming out or not.

37
ST-200Pro Service Manual

milli Volts of Na electrode


differ from previous Calibrate
Drift Na
sample analyses milli
Volts.
milli Volts of K electrode
differ from previous Calibrate
Drift K
sample analyses milli
Volts.
milli Volts of iCa electrode
differ from previous Calibrate
Drift iCa sample analyses milli
Volts.
milli Volts of Li electrode
differ from previous Calibrate
Drift Li
sample analyses milli
Volts.
milli Volts of pH electrode
differ from previous
Drift pH sample analyses milli
Calibrate
Volts.
milli Volts of Cl electrode
differ from previous
Drift Cl sample analyses milli
Calibrate
Volts.
Bubble at Na
Bubble at K
Bubble at iCa
Knock membrane
Bubble at Li
Bubble is present at part of reference
Bubble at pH
membrane part of housing.
Depends on selected reference housing
model if instrument
Refer: Topic 6.1
shows bubble at all the
parameters

Any other error message rather then above messages arrives or


could not solve the problem refer to technical person or service
Center.

38
ST-200Pro Service Manual

7. SOFTWARE UP-DATE:

Software can be upgraded by changing the Daughter card on Mother


Board or by using FET-Pro430 Lite Software and J-Tag.

SOFTWARE UPGRADATION THROUGH JTAG:

STEP 1: Download and install FET-Pro430 Lite Software from below link.
Flash Programmers for TI's MSP430 MCU uses TI's FET Adapter.
http://www.elprotronic.com/download.html

STEP 2: Create a desktop shortcut file of FETPRO LITE Software.

STEP 3: Now connect JTAG 14 PIN male connector to motherboard


female pin and USB connection to PC (Computer) respectively.

Fig: 27 MSP430 JTAG

STEP 4: After connecting the device, switch on the instrument and open
the FETPRO Lite application by double-clicking on the shortcut key.
Ensure to observe the red & green led glows.
STEP 5: After opening the application you need to click on open source
code sub menu and browse the path for .HEX file provided by us.

STEP 6: Select respective controller GROUP (MSP430F6xx) and the


controller specified MSP430F6779 for 24-bit Microcontroller.

STEP 7: Now click on the Verify security fuse and wait for 5-6 seconds.
You will find the verify security fuse as passed. Then, click on the auto
program (AUTO PROG.) menu and wait for 2-3 minutes. After finishing
the software update it will show passed and completed syntax.

39
ST-200Pro Service Manual

DEFAULT SETTINGS

At the initialize stage instrument will have default settings as mentioned


below.

Sample Type Serum/Plasma


Calibration Frequency 4 Hours
Sipping Frequency 30 Minutes
Printer Status ON
Print mV’s Status ON
Auto Standby OFF
Patient ID Auto

Depending upon your requirement you can change settings. Once


settings have been changed it will not change until you made any
changes or 3.3V DC battery is removed on motherboard.

40
ST-200Pro Service Manual

Take Care

1. Once Display is initialized, open the ST-200Pro) application and


calibrate. Calibration slopes should be in following ranges.

Calibration Slope ranges at RT (Max and Min)

Sodium [Na]:Slope should be in between 100 to 35 mV/decade


Potassium [K]: Slope should be in between 100 to 35 mV/decade
Ionized Calcium [iCa]: Slope should be in between 50 to 15 mV/decade
Chloride [Cl]: Slope should be in between 100 to 30 mV/decade
Lithium [Li]: Slope should be in between 100 to 30 mV/decade
pH [pH]: Slope should be in between 100 to 30 mV/decade

If slopes are not in the range or Instrument Flag “*****” press Calibrate
key and repeat Calibration.

Note: 24-bit controller millivolts reading range is from -435 to +435.

2. Perform quality control’s every day before analysis.


 Quality Control kit should be stored at room temperature.

Normal Ranges of Serum/Plasma


Sodium [Na]: Max 150 to Min 135 mmol/L
Potassium [K]: Max 5.0 to Min 3.5 mmol/L
Ionized calcium [iCa]: Max 1.35 to 1.10 mmol/L
Lithium [Li]: Max 1.50 to 0.30 mmol/L
pH [pH]: Max 7.45 to 7.35
Chloride [Cl]: Max 110 to Min 94 mmol/L

If the iCa unit is in mg/dL then the range is Max 5.4 to Min 4.4 mg/dL.

Normal Ranges of 24 hrs urine:


Sodium [Na]: Max 220 to Min 40 mmol/L
Potassium [K]: Max 125 to Min 20 mmol/L
Chloride [Cl]: Max 250 to Min 100 mmol/L

Normal Ranges of Cerebrospinal fluid (CSF)


Sodium [Na]: Max 150 to Min 135 mmol/L
Potassium [K]: Max 3.0 to Min 2.6 mmol/L
Ionized Calcium[iCa]: Max 1.40 to Min 1.0 mmol/L
Chloride [Cl]: Max 130 to Min 115 mmol/L

QC Ranges are given in Performance certificate provided with the


Analyzer.

3. Analyze type:

 Serum /Plasma:: Always Serum, heparin Plasma, clear CSF & in


emergency whole heparin blood.
41
ST-200Pro Service Manual

 Urine mode: Diluted urine i.e. reconstitutes 1000 µL urine diluent


with 200 µL urine sample(1:5 dilutions).

 QC Type: Quality Controls provided with the analyzer or external 3rd


party control’s.

 Non Detectable Mode: Eccentric sample, Hemolysis sample,


Lipemic sample, Dialyses fluids etc.

4. Reagent pack replacement: After inserting Reagent Pack do the


following procedure.

MENU

NEW REAGENT PACK

5. Daily Cleaning Solution (store kit in refrigerator):


Execute mandatory DC solution everyday.

MENU

Diagnostics

Daily Cleaner

6. Trouble Shootings:
A. Flip Mechanism:
Flip seal has to be replaced every six months.

B. Screen Hanging:
In Case if Screen/Software get struck do follow remedy.

 Press task bar and clear the screen


 Finally ‘SWITCH OFF’ the analyzer and Restart.

C. Factors for Air in Cal A or Air in Cal B:


 Reagent pack is empty and needs to be replaced.
 Bubble detector is not detecting cal A or cal B.
 There is a leak or disengagement in the flow path.
 There is a blockage in waste line which is restricting the flow of
solution.
 There is a blockage in welco pump tubing or probe.
Note: Pinch valve tubing should be wholly inserted towards inner side in
the valve. Partial or incomplete inserted tube may let two solutions mix
up.

D. Welco pump tubing need to be released, flushed & apply silicone


grease every four months.

E. Inner filling solution of reference house has to be replace every six


months.

42
ST-200Pro Service Manual

7. Steps to Switch off the Instrument:


In case to switch off the analyzer please follow below procedure.
Press ‘Shutdown’ on the screen. Then device is safe to turnoff.

8) Electrodes Staging:
Position of electrodes into the electrode housing below sample
detector is as follows:-

a) 6th position down is Chloride electrode.


b) 5th position is pH electrode.
c) 4rd position is Lithium electrode
d) 3nd position is ionized Calcium electrode
e) 2st position is Sodium electrode
f) 1st position top is Potassium electrode

Fig: 28 Electrode staging

43
ST-200Pro Service Manual

8. QUICK ACCESS SHEET:

44
Manufactured by:
Sensa Core Medical Instrumenta on Pvt. Ltd.
Plot No.3, Export Promo on Industrial Park, Pashamylaram,
Sangareddy (Dist), Hyderabad- 502307, Telangana, INDIA.
www.sensacore.com, Tel: +08455 223400.
service@sensacore.com

You might also like