Final Inspection Report

Vacuum Regulator Item NO.: A58-VR760-AS-DS PO/ PI Number PI-A221017P-1

INSPECTION INFO

PRODUCT NAME Vacuum Regulator


MODEL NUMBER A58-VR760-AS-DS
INSPECTION PERCENTAGE 100%
INSPECTION QTY 144 Sets

INSPECTION ITEMS MANUFACTURING REQUIREMENT Classfy TEST RESULT CONCLUSION

Apperance Surface smooth, uniform color, not be defective. C Visual surface color uniform, no defects Approved
Vacuum Gauge Accuracy not less than 2.5 magnitude mPa (0-0.1) B Accuracy 2.5 magnitude mPa (0-0.1) Approved
Should be light, loose and suitable, when closing Regulating valve is tight and no negative
Adjust Valve B Approved
regulator valve, there should be no output. pressure output when closing the valve
In atmospheric environments, should be able to Can be adjusted arbitrarily from -0.02MPa
Adjustment Range & adjust arbitrarily from -0.02Mpa (150mmHg) to (150mmHg) to -0.07MPa (525mmHg)
B Approved
Basic Error -0.07MPa (525mmHg), with a basic error of 2.5% with no error of more than 2.5%
of the maximum indication of the maximum indication value
Products should have a good sealing, the device
negative pressure output seal, the vacuum gauge
Gas Tightness to -0.07MPa negative pressure, close the valve port B Sealing meets standard requirements Approved
and seepage port, after 1 hour of pressure, vacuum
gauge indication should not be less than -0.05Mpa.
The device should be equipped with gas adapter for Equipped with specific adapter, connected
Connection hospital gas pipeline, shoud be sealed well after B well, no gas leakage Approved
connection, no gas leakage
Output Connector Maxiumn OD Φ9mmx1.5mm B Meets standard requirements Approved
Material Safety trap should be made of PC material B PC material Approved
Safety trap should be transparency for obserbing
Transparency B Safey trap body transparent Approved
the liquid capacity
Safety trap, suction bottle and suction tube won't
Pressure Resistance B No transshape Approved
transshape under -0.075mpa for 5 mins
Self-sealing Stop suction when the safety trap full of liquid B Meet the Requirement Approved
Flow Rate Max. flow should not less than 5L/min -0.075mpa B Flow rate 8.5/min under -0.075mpa Approved

ENCLOSED PICTURE:

INSPECTION PERSON CONCLUSION DATE