Electrocardiogram and Phonocardiogram Mo

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 11

This article has been accepted for publication in a future issue of this journal, but has not been

fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

Electrocardiogram and Phonocardiogram


Monitoring System for Cardiac Auscultation
Shuenn-Yuh Lee, Senior Member, IEEE, Peng-Wei Huang, Student Member, IEEE, Jia-Ren Chiou,
Chieh Tsou, Student Member, IEEE, Yu-Yi Liao, and Ju-Yi Chen

Abstract—Heart-sound auscultation is a rapid and fundamental


technique used for examining the cardiovascular system. The main
components of heart sounds are the first and second heart sounds.
Discriminating these heart sounds under the presence of additional
heart sounds and murmurs will be difficult. To recognize these
Aortic area Pulmonic area
signals efficiently, this study proposes a monitoring system with Second intercostal space
phonocardiogram and electrocardiogram. This system has two key
points. The first is chip implementation, including capacitor coupled Tricuspid area
amplifier, transimpedance amplifier, high-pass sigma–delta Fourth intercostal space
modulator, and digital signal processing block. The chip in the Fifth intercostal space
system is fabricated in 0.18 µm standard complementary metal– Mitral area
oxide–semiconductor process. The second is a software application Right of the Left of the
on smartphones for heart-related physiological signal recording, sternum sternum
display, and identification. A wavelet-based QRS complex detection
Fig. 1. Four main regions of cardiac auscultation [2].
algorithm verified by MIT/BIH Arrhythmia Database is also
proposed. The overall measured positive prediction, sensitivity, and [2]. Given that heart sounds are caused by the closure of
error rate of the proposed algorithm are 99.90%, 99.82%, and different valves, heart murmurs indicate several valvular heart
0.28%, respectively. During auscultation, doctors may refer to these diseases such as valvular stenosis and regurgitation.
physiological signals displayed on the smartphone and
The heart sounds mainly comprise the first (S1) and second
simultaneously listen to the heart sounds to diagnose the potential
heart disease. By taking advantage of signal visualization and heart sounds (S2). Efficiently recognizing S1 and S2 under the
keeping the original diagnosis procedure, the uncertainty existing in influences of additional heart sounds and murmurs is difficult.
heart sounds can be eliminated, and the training period to acquire After the discrimination of S1 and S2, the diseases related to the
auscultation skills can be reduced. additional heart sounds and murmurs can be diagnosed.
Therefore, the accuracy of auscultation relies on the doctor’s
Index Terms—Bio-signal acquisition, QRS complex detection, experience. However, there exist some obstacles to the
phonocardiogram, electrocardiogram, system on chip, cardiac accumulation of experience and education. For example,
auscultation, wearable device individuals may have different auditory sensitivity. Moreover,
the distinction between various kinds of heart murmurs is hard
I. INTRODUCTION to describe.

CARDIOVASCULAR diseases (CVDs) are the leading


cause of death globally. According to reports [1] from the
To assist doctors in making diagnoses more rapid and
accurate, this study proposes a system capable of measuring the
World Health Organization, an estimated 17.9 million people phonocardiogram (PCG), heart-sound waveform, and
died from CVDs in 2016. It is a horrible amount because it electrocardiogram (ECG) simultaneously. The reason why
exceeded 30% of all global deaths. To prevent such a ECG measurement is adopted in the proposed smart
worldwide killer, regular physical examination is important. stethoscope system is because the systolic and diastolic phases
Typically, heart-sound auscultation is a rapid and common of the heart can be easily distinguished. Therefore, the
method used in examining the cardiovascular system. As recognition of S1 and S2 is more reliable. During auscultation,
illustrated in Fig. 1, the aortic, pulmonic, tricuspid, and mitral doctors may refer to these two signals displayed on the
areas are four main regions of interest for cardiac auscultation smartphone to diagnose the potential heart disease. The
This research was supported in part by the Taiwan Semiconductor smartphone will concurrently play the heart sounds for keeping
Research Institute and the Ministry of Science and Technology (MOST), the original diagnosis procedure. By taking advantage of signal
Taiwan, R.O.C., under Grants MOST 107-2218-E-006-034, MOST 108-
2622-8-006-004-TE2, and MOST 107-2514-S-006-008 and South Taiwan
visualization, the uncertainty of heart sounds can be eliminated,
Smart Biomedical Industrial Clusters under Grant AZ-13-05-28-107. and the training period to acquire auscultation skills can be
(Corresponding authors: Shuenn-Yuh Lee and Ju-Yi Chen) reduced. Furthermore, the measured signals can be stored on
S. Y. Lee, P. W. Huang, J. R. Chiou, C. Tsou, and Y. Y. Liao are with the
Electrical Engineering Department of National Cheng Kung University, No.
the smartphone and uploaded to a cloud server. These materials
1 University Rd., Tainan, 70101, Taiwan (e-mail: ieesyl@mail.ncku.edu.tw, can be utilized as educational resources to improve the learning
lookonthebrightside_j@hotmail.com). efficiency by replaying the recorded waveforms and heart-
J. Y. Chen is with National Cheng Kung University Hospital, Tainan, sound audio.
Taiwan (e-mail: juyi@mail.ncku.edu.tw).

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

the ECG band in clinical usage is between 0.05 Hz and 100 Hz.
CHIP To reduce the system design’s complexity, the acquisition
ECG CCIA HPSDM UART Clock bandwidth of the proposed system for heart-related signals is
Interface Divider Bluetooth set as 200 Hz. The sampling and transmission for both signals
Decimation Filter
Module are all synchronized. If the bandwidth and the sample rate are
PCG TIA HPSDM set differently for each signal, solving the problems of clock
Digital domain crossing and transmission will make the system more
Signal Acquisition Block Processing Block complicated. Moreover, the IP reuse in the same specification
of two HPSDMs can reduce the effort on system-level design.
Bluetooth
The ECG sensing channel can suppress the influences of DC
Software-based
User Module offset and common mode AC interference. The removal of
Signal
Processing
Interface baseline wander and motion artifacts are primarily implemented
in the software side.
Software Design Smart Phone The rest of the paper is organized as follows. Section II
describes the major concerns of on-chip circuit design. Section
Fig. 2. System diagram of the smart stethoscope.
III discusses the implementation of the whole system and the
The block diagram of the whole system is shown in Fig. 2. beat detection algorithm in software application. Section IV
In the chip implementation, it is a sensing circuit for the elaborates the results of algorithm testing and system
synchronization of ECG and PCG measurements. According to measurement. Finally, Section V draws the conclusions.
the integration of PCB system board, the real-time monitoring
function can be achieved with the wireless transmission of BLE II. ON-CHIP CIRCUIT DESIGN
module. At the software side, a wavelet-based beat detection The analog front end (AFE) in this system has two channels
algorithm with good performance is proposed. The mechanism that can acquire ECG and PCG signals simultaneously. The
of peak insertion and deletion based on the detection of the main difference between these channels is the front-end
possible P and T waves according to advice by cardiologist in amplifier. The first channel uses capacitor coupled
NCKU hospital is added in the proposed algorithm. In addition, instrumentational amplifier (CCIA) to magnify the voltage-type
the ultimate application scenario of the proposed system is ECG signal from the body appropriately. The second channel
different from the existing product. Combining the ECG uses transimpedance amplifier (TIA) to convert the heart-sound
sensing patch with the microphone-equipped stethoscope, the current-type signals received from microphone to voltage-type
reliability and quality of auscultation can be improved. Also, signals. These amplified signals will then be digitalized by an
detaching the stethoscope, the ECG patch itself can be applied analog-to-digital converter (ADC) and sent into the digital
to the patients who needs long-term monitoring. Since a low circuit for further processing. Chopper stabilization technique
power and wearable device is required for long-term healthcare, is applied on each of the amplifiers to prevent low-frequency
the chip implementation is beneficial to such applications. The noise during the complementary metal–oxide–semiconductor
miniaturization of system can make the sensing device easier to (CMOS) process. The whole chip can be organized as a CCIA,
wear. The reduction of power consumption accomplishes a TIA, high-pass sigma–delta modulator (HPSDM), decimation
better battery life. For the chip design in the proposed system, filter, and universal asynchronous receiver/transmitter (UART)
the 2-channel signal acquisition consumes about 800 µW. Also, transmission interface.
the power consumption is only 60 µW in the single-lead ECG
channel. In contrast with the chip implementation, ECG sensing A. Capacitor Coupled Amplifier
circuits integrated from discrete components often cost about The block diagram of the capacitor coupled amplifier
several mWs depending on the specification. These advantages employed in the proposed system is shown in Fig. 3(a). In this
gained from chip design help to enhance the feasibility of long- CCIA, the operational amplifier is realized in the form of
term healthcare and to minimize the inconvenience brought differential difference amplifier (DDA) illustrated in Fig. 3(b).
from the consecutive monitoring. For offline data collection, The DDA contains two inputs that can separate the input signal
switching off the BLE transmission function and adding a SD path from the feedback path; thus, the input of CCIA would not
card on the PCB system board can further extend the battery suffer from large loading caused by the feedback capacitor.
life. The two feedback capacitors C1 and C2 define the signal gain
For designing such a wearable monitoring system, of the CCIA, and the feedback resistor Mr in parallel with C2 is
establishing the appropriate and efficient specification is the used to determine the DC point of the feedback input pair.
first but fundamental step. It may directly influence the finished However, this feedback resistor results in a low-frequency pole
products in terms of performance, practicability, and battery and zero in the closed-loop transfer function. The transfer
selection. Hence, investigating the properties of input signals is function is derived as follows:
the primary task. The main frequency band of heart sounds
ranges between 20 Hz and 150 Hz [3]. Furthermore, on the basis  1 
Gm1  s  
of frequency distribution analysis in several types of heart  Rpseu  C1  C2  
ACL  s    (1)
sounds [4], most frequency components are under 200 Hz. On G C Gm 2
the other hand, the ECG bandwidth is around 0.01–250 Hz with Cc s 2  m 2 2 s 
C1  C2 Rpseu  C1  C2 
0.5–4 mV amplitude range [5]. Another analysis [6] shows that

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

C1 C2 On Chip
Rf
Pseudo
Resistor Rb
Cu

CME-1539-100LB
Mr Vout+
Mr
Vcm Vout-

C1 C2
Rf
Vout+ (a)
Vb1
Vin-
Vout-
+
Vin
Vin+ Vin-
(a)

Vb1 Vout+ Cc Cc Vout-


Vb2
Vi1+ Vi1- Vi2+ Vi2-
M1 M2 M3 M4

Vout+ Cc Cc Vout-
Vb2 Rf Rf

(b)
Fig. 5. (a) Block diagram of TIA, (b) schematic of TIA without chopper.
C2 C2
C1 band frequency range of CCIA is from 100 Hz to 60 kHz.
According to process corners simulation with five models, the
Rpseu Rpseu
simulated CMRR is higher than 68 dB and the PSRR is higher
(b) than 54 dB.
Fig. 3. (a) Block diagram of CCIA, (b) schematic of CCIA without B. TIA
chopper.
Unlike the ECG signal, the heart-sound signals are first
received by the microphone, which transforms the acoustic
|ACL(s)| signal into electronic signal. In this system, the electret
Gm1  C1  condenser microphone (CME-1538-100LB) fabricated by CUI
1  
Gm 2  C2  Inc. is utilized. The electret condenser microphone consists of
a microelectromechanical system (MEMS) microphone and a
Midband
Gm1 field-effect transistor (FET) impedance converter. Driven by
Gm 2 the MEMS microphone, the FET impedance converter extracts
f the current from output loading. Accordingly, a TIA should be
fz fp1 fp2
used to convert the current-type signals from the microphone
Fig. 4. Frequency response of CCIA. into voltage-type signals, so that the following HPSDM can
TABLE I operate normally.
POLE ZERO LOCATION OF CCIA The adopted microphone is single-ended; however, the
f zero , 𝑓 f pole ,low , 𝑓 f pole,high , 𝑓 conventional amplifier in CMOS process is in fully differential
1 1 1 1 1 Gm 2 C2 mode. Fig. 5(a) shows the implementation of the microphone
incorporated with TIA, where Rb is the bias resistor of
2 R pseu  C1  C 2  2 R pseu C 2 2 Cc C1  C2
microphone, Cu is the DC block capacitor, and Rf is the
feedback resistor of TIA. The conventional two-stage
where Gm1 and Gm2 are the transconductance of input pair M1,2
operational amplifier is used in this TIA as shown in Fig. 5(b).
and M3,4, respectively, Cc is the compensation capacitor of the
One of the input terminals is connected to the microphone, and
DDA, and Rpseu is the resistance of the feedback resistance. The
the other terminal is connected to the common mode reference
frequency response is shown in Fig. 4, and the location of poles
voltage. Similarly, a chopper is inserted between the
and zeros are listed in Table I. After chopping, the signal
microphone and the TIA to reject the low-frequency noise in
frequency is located within the mid-band region. The low-
TIA. Such connection can receive single-ended current signal
frequency pole–zero pair caused by the feedback resistor should
from the microphone by a fully differential TIA, and the gain is
be designed low enough to avoid affecting the signal band.
2Rf instead of Rf.
Therefore, the feedback resistor is implemented by pseudo
Fig. 6 shows the diagram of TIA for derivation of frequency
resistor [7] to realize extremely large resistor without using a
response, where Cp is the parasitic capacitor of TIA. Assuming
large area resistor. According to the simulation results, the mid

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

CP Rf Zf Second Stage

b1
Ix First Stage

Vout+ X(z) g1 H(z) g2 H(z) g3 H(z)

Iin A(s)
Vout- a1 a2 a3
Ix

CP Rf Zf Summing Stage
1-bit
fs=2fchopper
Fig. 6. Circuit diagram of TIA for derivation of frequency response. quantizer
DAC Y(z)

TABLE II
Fig. 7. System block of the SDM.
POLE LOCATION OF TIA
f p1 f p2 TABLE III
1 1 1 1 COEFFICIENTS OF EACH PATH
2 2Cp R f 2 (1  A0 )0 a1 2.5 g1 0.333
a2 1.25 g2 0.667
that the operational amplifier is designed appropriately, the a3 1.25 g3 0.179
open-loop transfer function is listed below:
b1 0.01
A0
Aopen  s  
s (2) CH
1 Φ2 Φ1
0
where A0 is the DC gain and ω0 is the dominant pole. The value Φ1 Φ2
of unity-gain bandwidth (UGB) ωt is the transconductance of -1
input pair divided by the compensation capacitor Cc. According CI
to the above assumption, the closed-loop transfer function of
TIA in Fig. 6 can then be derived as follows:
CS
  Φ1 Φ2
Vin
2R f  A 1  Vout
ACL  s    0  (3) Φ2 Φ1
1  s 2C p R f 1  A0 1  s 
 1  A0  0 
 Fig. 8. First-order high-pass integrator.
Because the feedback resistor is not in parallel with a capacitor,
the transfer function does not contain any zeros but has two effective number of bits increases by 0.5 bits when the OSR
poles fp1 and fp2 associated with parasitic capacitor and UGB of doubles. Except for the oversampling, the SDM behaves as a
the operational amplifier, respectively. The pole location is filter, so that the quantization noise can be suppressed in the
listed in Table II. Given that the TIA is operated in low signal band but increased in the out-of-signal band. In this way,
frequency and the compensation capacitor is usually larger than the in-band quantization noise would be exponentially inversely
parasitic capacitor, fp2 is considered as the dominant pole that is proportional to the OSR.
controlled by the UGB of operational amplifier to achieve the Fig. 7 shows the system block of the 3rd multi-feedforward
compensation effectively. In the simulation results, the cutoff SDM with a local feedback, and the noise transfer function is
frequency range is about 100 kHz which is much higher than listed below. The related coefficients of each path are illustrated
the chopper frequency 25.6 kHz. The PSRR can achieve greater in Table III.
than 135 dB under the single-ended TIA. 𝑌 𝑧 1 𝑏𝑔 𝑔 𝐻
𝑋 𝑧 1 𝑎 𝑔𝐻 𝑎 𝑔𝑔 𝑏𝑔 𝑔 𝐻 𝑔𝑔 𝑔 𝑎 𝑎 𝑏 𝐻
C. HPSDM
(4)
The sigma–delta modulator (SDM) is widely used in ADC
with high resolution. It is especially suitable for low- and mid- If the signal is chopped back to the original frequency before
band-frequency application. Moreover, the specifications of ADC, the offset and flicker noise of pre-amp would be
circuits such as operational amplifier and quantizer realized in modulated to the chopping frequency, and additional analog
the SDM can be much more relaxed. filter is required. Under the circumstances, the utility of signal
The ability of noise shaping is the main difference between bandwidth is inefficient. Therefore, the HPSDM is more
SDM and conventional Nyquist rate ADC. Given that the attractive in such application to digitalize the chopped signal
quantization noise is normally distributed between the half and retain bandwidth efficiency. The high-pass noise transfer
sampling frequency, the in-band quantization noise is linearly function is obtained by turning the variable in the low-pass
inversely proportional to the over sampling ratio (OSR), and the noise transfer function from z to −z. Thus, the central frequency
is shifted from 0 to fs/2. Accordingly, the low-pass integrator

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

614.4
CI Clock Divider
kHz

Cascaded
Cp CS 2-ch
Φ1 CS VX
Φ2 Chopper
Integrator- Compensation
IIR Filter Equalizer
SDM Data Comb Filter
Vin Filter
Vout
Decimation Filter
Φ2 kCS Φ2 Φ1
FIFO
UART Digital Biosignal
UART TX
Interface
Fig. 9. Capacitive-T integrator. Baud Rate (38400)

should be transformed into high-pass integrator [8] shown in Fig. 10. Block diagram of DSP block.
Fig. 8, and the transfer function is derived as follows:
noise transfer function is therefore much easily implemented
and still provides sufficient accuracy as compared with
𝐻 𝑧 (5) conventional AFE structure. The complete design flow [9] from
system stability to circuit implementation is introduced. Other
nonideal effects such as finite and nonlinear gain, bandwidth of
where the capacitance of CH and CI are identical in this design. operation amplifier, thermal noise of switch, and coefficients of
From Table III, the coefficient b1 is much smaller than other variation in the modulator should be carefully considered.
coefficients. This will lead to a very small capacitor to become Following the design flow, the modulator can be optimized by
severely influenced by the parasitic capacitors. Fig. 9 shows a applying the minimum specification of the operational
capacitive-T integrator that can implement a very small amplifier.
coefficient in an integrator and reduce the effect of parasitic
capacitor. The transfer function is shown below: D. Digital Signal Processing (DSP) Block
C 1  z  1 The proposed DSP unit mainly serves as the clock divider,
H  z   s  1  decimation filters, and UART package generator. The detailed
CI k  2  P  1  z 
C (6)
block diagram is shown in Fig. 10. The clock divider triggered
Cs
by the on-board voltage-controlled oscillator derives all the
where k is the capacitance ratio in the T capacitor network and clocks for the demand of AFE and DSP blocks. Given that the
Cp is the parasitic capacitor. This result reveals that the HPSDM in AFE plays the role of ADC, the digital circuit will
corresponding sampling capacitor Cs is reduced by k+2+Cp/Cs then start to receive the single-bit digital stream after the clock
times. divider is functioning. Two following decimation filters are
To sum up, the main features in the overall AFE architecture applied for different bio-signals: ECG and PCG. These two
are flicker noise suppression, low power consumption and high decimation filters are identical because the target specifications
bandwidth usage efficiency. First of all, in order to resist the for each application are the same. However, how do we
flicker noise, the chopper mechanism is adopted in the design implement these decimation filters considering that many
of AFE. In the conventional chopper-based architecture used in approaches are available? A cascade structure is adopted to
LPSDM system, a severe problem of ripple effect is achieve the goal of low power consumption. The four stages are
encountered when adding the output chopper stage of pre- as follows: a cascaded integrator–comb (CIC) filter, a
amplifier. This ripple may make the sensed bio-signal saturated. compensation filter, an infinite impulse response (IIR) filter,
In general, a well-designed filter needs to be inserted afterward and an equalizer [10]-[13]. A CIC filter is the first stage because
to solve this problem in the conventional structure. Such a it can fulfill the high decimation ratio in a simple concept. The
solution leads to extra efforts and power consumption. On the z-domain transfer function is listed below [10]:
other hand, if the LPSDM is implemented between the pre-amp
and output chopper stage, the extra requirement for its 1 𝑍
bandwidth is at least twice the frequency of chopper. To H z 2 , 𝑃 𝑀 𝑙𝑜𝑔 𝑅 (7)
1 𝑍
simplify the conditions mentioned above, several adjustments
are required for AFE structure optimization. In the proposed where M represents the order of CIC filter and R is the
system, the output chopper stage is implemented in digital decimation ratio. The whole down sampling ratio (DSR) of the
signal processing block to avoid the design of high-end analog proposed decimation filter is 128; thus, the decimation ratio (R)
filter. In addition, the HPSDM architecture is directly utilized of the CIC filter is set at 32. The remaining four DSRs are
to analyze the output signal of CCIA for better efficiency in accomplished by the following filter stages. To attain sufficient
bandwidth usage. Since the input bio-signal is chopped to high attenuation of stopband, the order M is set at 4. Although the
frequency to prevent the interference of flicker noise, the CIC filter can achieve high decimation ratio, the passband will
modulated-signal band is completely reserved by the HPSDM have some gain loss. The compensation filter is applied to
with the bandwidth of 200 Hz. Compared with LPSDM system, compensate the loss and make the passband much flatter. The
the output chopper stage and corresponding analog filter are no compensation filter is realized by equiripple finite impulse
longer needed in this architecture. The proposed AFE system response (FIR) filter. After placing the signal frequency much
can suppress the DC offset and flicker noise through a simpler closer to the destination, a sharp transition band filter should be
architecture. From the viewpoint of overall sensing channel, the constructed. Therefore, the third stage is an IIR filter. The IIR
chopper-based pre-amplifier with HPSDM providing high-pass filter has lesser coefficients and shorter delay time than the FIR

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

TABLE IV
1 0 LSB MSB 1 BLUETOOTH VERSION COMPARISON [15][16]
Bluetooth Bluetooth
Start Stop Low energy BR/EDR
8-bit Data
Bit Bit
Setup time <6 ms 100 ms
Data rate 125 Kb/s to 2 Mb/s 1 Mb/s to 3 Mb/s
Fig. 11. UART 8-N-1 format transmission diagram. ~0.01x to 0.5x
Power consumption 1 (reference value)
of reference
filter when achieving the same specification. However, the non- Point-to-point
linear phase is a problem for the IIR filter. The solution to this Network topologies Broadcast Point-to-point
Mesh
phase problem is to append an equalizer behind the IIR filter.
The equalizer compensates and minimizes the max difference
in group delay. The structure of the decimation filter is marked Rate/Enhanced Data Rate (BR/EDR), low power consumption
in the blue block of Fig. 10. After passing through several stages and short connecting latency are characteristic of BLE. Some
of signal down sampling and processing, the digitized ECG and key features of the two different technologies are listed in Table
PCG signals are finally acquired. IV [15][16]. The communication range is not a primary issue to
For the purpose of transmitting the physiological information this system because the distance between the proposed device
to the on-board Bluetooth module, the data should be and smartphone is only a few meters, and both technologies can
rearranged into a common communication standard. These 2- meet the need. Although BLE has lower data rate, its theoretical
channel 16-bit signals mentioned above are sampled value is still far beyond the least demand to keep persistence of
synchronously. After adding an 8-bit trailer, the whole data vision. In this case, the transmission period needs to be
length is 40 bits. The data rate from the output of decimation determined properly in order to correspond with the signal data
filter is 400 Hz; thus, the equivalent bit rate is 16 Kb/s. To rate, communication firmware standard, and acceptable user
minimize the number of total output pads, the UART approach
experience.
is selected to transmit the data. The UART package generator
A firmware operation routine and a custom BLE profile are
is implemented to divide the output data into a UART 8-N-1
implemented to serve as stable bridge of communication.
format [14], signifying that a start bit 0 and a stop bit 1 will be
added for each packet (byte). That is, the data rate is now Playing the role of a peripheral/server, the proposed hardware
increased by 25% and reaches 20 Kb/s. The UART system needs to push the processed data to the client. The
transmission diagram is shown in Fig. 11. Given that the notification operation is chosen and given that this application
common baud rates are 9600, 19200, 38400, 57600, and 115200 works at higher transfer rate. A BLE setting parameter called
Hz, it is adequate to choose 38400 in the proposed system in maximum transmission unit (MTU) size directly affects the
accordance with the preceding calculation. When waiting for data length of package per notification operation. The MTU
transmission, the data at each sample point needs to be reserved size remains at its default setting of 23 for wider device support.
temporarily. A circular first input first output (FIFO) buffer Each notification package length is 20 bytes. According to the
rather than shifting register is introduced to save the previous calculation, the physiological signals fill the package
unnecessary dynamic power dissipation. every 10 ms. However, this interval may not be an appropriate
The 2-channel equivalent data bit rates of HPSDMs and value for establishing every connection stably. To determine the
decimation filters are 102.4 and 16 Kb/s, respectively. Both suitable connection interval matching with data rate is the main
these digital signals can be packaged into UART format. Why concern of firmware implementation. The minimum connection
are the outputs of decimation filters eventually selected? If the interval varies from the adopted BLE chips in smartphones to
outputs of HPSDMs are adopted, the decimation filter can be the different mobile operating systems. Some mobile operating
removed. Under these circumstances, the power consumption systems have several listed rules [17][18] to assist in
of the chip is certainly lower. However, from the perspective of establishing and retaining the connection successfully. These
the system level, higher bit rate infers higher RF transmission
rules may be changed or updated at intervals. After many
power. The transmission power commonly dominates the
practical experiments, the minimum connection interval needs
power consumption of the overall system. It is a more
to be greater than 20 ms for the phones tested. A small buffer is
economical decision to the system when the decimation filter is
employed. Ignoring the 25% increment of transmission needed to hold the data and waits for the next notification. After
overhead, it can save up to 6.4 times the bandwidth than using converting the UART transmitter data into the corresponding
the output of HPSDMs. characteristic value attributes of custom BLE profile, the BLE
module can send all the information to smartphones by
III. APPLICATION IMPLEMENTATION complying with the setting suggestion of connection parameters.
Through smartphones, real-time bio-signal monitoring, offline
A. Firmware Design data storage, and further physiological condition analysis can
After setting the baud rate, the on-board Bluetooth module then be conducted.
successfully receives the ECG and PCG signals via UART
packages. Given that the total bit rate of the proposed system is B. Software Signal Processing and Beat Detection Algorithm
only up to 16 Kb/s, Bluetooth Low Energy (BLE) is the best Developing real-time mobile applications for wearable
option to this application. Compared with Bluetooth Basic sensors has become a popular topic since the concept of mobile

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

The time and frequency resolutions are affected by the


BLE
Receiver
Wavelet
Transform
decomposition level of wavelet. In the proposed system, the
ECG signal is decomposed by a 4-level Haar DWT operation.
Data Median
The output wavelet coefficient W4 is fed into a median filter for
Formatting Filter further noise suppression. Then, a window-based maximum
value searching is employed to select the R peak candidates.
PCG signal Filtering
Filtering
ECG signal Window-based
Max Searching
The window length is determined by the maximum heart beats
per minute (BPM) value of adults. For a digital ECG signal with
400 Hz sampling rate, a 109-point window is enough to detect
Heart Possible QRS
Sound Display
Beat
Detection
P/T Wave Configuration
R-R Interval
Analysis
the heartbeats under 220 BPM [19]. Given that the window-
Player Detection Extraction
based maximum value searching serves as the beat candidate
selection, the adaptive threshold mechanism is involved. The
Recording
/
QRS
Confirmation
location and value of the maximum larger than threshold are
Uploading
output as the features of the candidate. Only these selected
Software Main Functions Beat Detection Algorithm candidates can participate in the updating of adaptive threshold.
If two continuous windows both detect its local max value, the
Fig. 12. Block diagram of software application. comparison between these two candidates will then be triggered.
healthcare and telemedicine has been widely discussed and The former will be output if the interval is longer than a window
researched during recent years. However, it is never an easy length, and the output decision of the latter will be held over
task. First, signal monitoring with real-time data analysis is a until the next window is analyzed. However, if the interval
fundamental capability. Furthermore, noise interference is an between these two candidates is smaller than a window length,
unavoidable challenge to all portable and wearable applications. only the larger value will be output eventually. The detailed
The tolerable noise level determines the requirement of procedures are presented in our previous work [20].
suppression efforts. Both hardware circuits and mobile software The MIT-BIH arrhythmia database [21], a prevailing
have their limitations. Choosing the suitable system structures standard benchmark, is used to evaluate the proposed beat
and filtering methods can reach the balance economically. In detection algorithm. It is the excerpts of ambulatory ECG
addition, the operation complexity of signal discrimination is recordings. The overall detection results are passable in most
also an important index. Pursuing a simpler approach and better conditions when original algorithm is applied. But the
analysis results are the ultimate goal, but it often turns out to be performance decreases under the noisy coupling. In these
a trade-off. Last but not least, a good user experience including regions, if the noise can pass the threshold mechanism, the non-
friendly application interface, short operation latency, and R-peak noise will be recognized as R peak. So, the number of
robust analysis is the comprehensive score of the system. false positives (FP) increases. Therefore, the endurance of noise
The concise signal processing flow of software based on the and artifacts is more essential when applying to a patient who
mobile operating system is shown in the left side of Fig. 12. needs long-term monitoring. The proposed algorithm in this
After receiving the 2-channel heart-related physiological manuscript is based on the previous work and the mechanism
information, the data format is first rearranged into readable of peak insertion and deletion is added. Trying to find the
values. Then, low-pass and band-pass FIR filters are added to possible P or T wave between R-R intervals is the fundamental
retain the major frequency band of ECG and PCG signals, concept of the proposed method. Since the relations of P, R and
respectively, and reduce the interference of out-of-band noise. T waves can present the cardiac physiology, the duration and
The filter parameters are set adequately for signal monitoring interval reasonably need to follow some medical rules.
[6]. The cleaner ECG/PCG signals are displayed According to each output of the window-based maximum
simultaneously and synchronously on the screen for doctors to searching, the current P wave and the former T wave can be
read. These waveforms are stored in the smartphone’s file found between the current R peak and the former one in general
system, and the data files can be uploaded to a cloud server condition. Similarly, the following current T wave will appear
when a network is available. During monitoring, the PCG signal in a certain range. During the searching procedures, the
is written into the audio buffer and played; thus, the integration configuration of the current peak can be extracted. The peak
of hearing and vision is accomplished. At the same time, the configuration, including the locations and the values of start
proposed ECG beat detection algorithm is carried out. Once the point, peak, wave trough, and end point, is recorded.
QRS complex is detected, the systolic and diastolic phases of a Combining the foregoing features with several past RR interval
heart are distinguishable. The method of 2-channel cross- (RRI), the decision, on whether the peak is a real QRS complex,
reference is beneficial for a rapid diagnosis. is made. If the current chosen peak is really a true R peak, some
Given that R peak detection is fundamental to arrhythmia physiology-oriented requirements, such as PR interval, QRS
classification, a reliable and efficient discrimination method is duration or QT interval, should be satisfied. When the searching
essential to heart-related healthcare system. The block diagram on both P and T waves for current peak fails, this peak will then
of the proposed beat detection algorithm is shown in the right be removed because it is determined as a large non-R-peak
side of Fig. 12. The 1-D Haar discrete wavelet transform (DWT) artifact or a sharper T wave. Moreover, the missing R peak,
is adopted in the main preprocessing stage for its low filtered by the adaptive threshold in original algorithm, may be
complexity. DWT uses the scaling and wavelet filter to discovered during the P wave searching procedure. The
decompose the signal into low- and high-frequency components. mentioned procedure is basically implemented by peak
searching with threshold method. For each discovered peak,

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

TABLE V TABLE VI
MIT-BIH DATABASE TESTING RESULTS PERFORMANCE COMPARISON WITH RECENT BEAT DETECTION SYSTEMS
AND METHODS
Record P+ Se ERR
TP FP FN Method Se (%) P+ (%)
No. (%) (%) (%)
100 2273 0 0 100 100 0 Gutiérrez-Rivas et al. (2015) [23] 99.54 99.74
101 1863 2 2 99.89 99.89 0.21 Phukpattaranont (2015)* [24] 99.82 99.81
102 2187 0 0 100 100 0 Sahoo et al. (2016)* [25] 99.71 99.72
103 2084 0 0 100 100 0
104 2228 4 1 99.82 99.96 0.22
Li et al. (2017) [26] 99.60 99.77
105 2562 27 10 98.96 99.61 1.42 Hou et al. (2018) [27] 99.32 99.45
106 2022 2 5 99.9 99.75 0.34 Lee et al. (2018) [20] 99.44 99.40
107 2133 0 4 100 99.81 0.19 Bashar et al. (2019)* [22] 99.94 99.95
108 1760 10 3 99.44 99.83 0.73
The proposed algorithm 99.82 99.90
109 2532 0 0 100 100 0
111 2123 1 1 99.95 99.95 0.09
*
Pure algorithm development
112 2539 0 0 100 100 0
113 1795 0 0 100 100 0
114 1877 5 2 99.73 99.89 0.37
115 1953 0 0 100 100 0
116 2390 1 22 99.96 99.09 0.95
117 1535 0 0 100 100 0 CCIA HPSDM
118 2278 2 0 99.91 100 0.09

1.99 mm
119 1987 1 0 99.95 100 0.05
121
122
1863
2476
0
0
0
0
100
100
100
100
0
0 DSP
123 1518 0 0 100 100 0
124 1619 0 0 100 100 0
200 2599 3 2 99.88 99.92 0.19 TIA HPSDM
201 1949 0 14 100 99.29 0.71
202 2135 0 1 100 99.95 0.05
203 2925 12 55 99.59 98.15 2.24
205 2645 2 11 99.92 99.59 0.49
207 1847 0 13 100 99.3 0.7 2.44 mm
208 2945 3 10 99.9 99.66 0.44 Fig. 13. Microphotograph of the proposed system.
209 3005 1 0 99.97 100 0.03
210 2640 1 10 99.96 99.62 0.41
212 2748 0 0 100 100 0
213 3251 0 0 100 100 0
214 2262 1 0 99.96 100 0.04 IV. E XPERIMENT RESULT
215 3362 0 1 100 99.97 0.03
217 2201 1 7 99.95 99.68 0.36 A. Testing Results on MIT-BIH Arrhythmia Database
219 2154 0 0 100 100 0 After the testing on the first channel in MIT-BIH arrhythmia
220 2048 0 0 100 100 0
221 2423 1 4 99.96 99.84 0.21
database, the experiment results are listed in Table V. Of note,
222 2479 14 4 99.44 99.84 0.72 the regions of ventricular flutter wave in record no. 207 are
223 2604 0 1 100 99.96 0.04 excluded because they are not annotated [22]. The performance
228 2048 11 5 99.47 99.76 0.78 measures, such as p ositive prediction (P+), sensitivity (Se), and
230 2256 0 0 100 100 0
231 1570 0 1 100 99.94 0.06
error rate (ERR) are calculated using the following equations:
232 1778 6 2 99.66 99.89 0.45 P+ = TP/(TP+FP) (8)
233 3073 0 6 100 99.81 0.19
234 2753 0 0 100 100 0 Se = TP/(TP+FN) (9)
Total 109297 111 197 99.9 99.82 0.28
ERR = (FP+FN)/(TP+FP+FN) (10)
whether it is a P wave, T wave, missing R wave or even an
where TP represents true positive detection, FP means false
artifact, is estimated in accordance with its value and location.
positive detection, and FN is false negative detection. The
Since the proposed algorithm contains peak confirmation with
overall measured P+, Se, and ERR values of the proposed beat
insertion and deletion, the FP and false negative (FN) can be
detection algorithm are 99.90%, 99.82%, and 0.28%
further reduced to achieve better results as compared to our
respectively.
previous work [20]. Although the noise interference and
Table VI summarizes the performance measurements of
arrhythmia increase the difficulty in P/T wave detection, some
several different algorithms [20] and [22]-[27]. The proposed
conditions are added in practice to enhance the tolerance level.
algorithm shows a good result. Some algorithms can be
In addition, the above-mentioned features involve some extra
implemented in real-time detection systems and even in chip.
purposes. For example, the historical amplitudes of P/T waves
Although the proposed beat detection algorithm cannot achieve
may provide reference for reducing the influence of noise.
such high performance as the variable frequency complex
Furthermore, several types of arrhythmias can be classified by
demodulation-based ECG detection method [22], it is still an
linking up the configuration of QRS complex with historical
efficient approach that can be employed in real-time detection
RRI. The proposed algorithm is verified by MIT-BIH
systems. This method is currently realized in smartphones and
arrhythmia database and human study.

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

TABLE VII
CHIP PERFORMANCE SUMMARY
General specifications
Technology TSMC 0.18 µm 1P6M CMOS process
Chip Area 2.44×1.99 mm2
Supply voltage 1.2 V / 1.8 V / 3.3V
Analog front-end circuit (a)
Chopper frequency 25.6 kHz
Max input voltage 20 mVpp
CCIA Mid-band gain 16/20/27 dB
Input referred noise 1.84 µVrms
Max input current 40 µA
TIA Mid-band gain 800 Ω
Input referred noise 1.95 nArms
Bandwidth 200 Hz
HPSDM Sampling frequency 51.2 kHz
SFDR >72 dB
Power consumption 0.8 mW (b)
Digital signal processor
Signal bandwidth 200 Hz
Decimation filter
Word length 16 bits S1 S2 S1 S2 S1 S2 S1 S2 S1 S2 S1 S2 S1 S2
Power consumption 30 µW

has promising potential to be integrated in hardware given that R


peak
R
peak
R
peak
R
peak
R
peak
R
peak
R
peak
the mathematics used are not complex. (c)
B. Measurement Results of the Proposed System Fig. 14. Proposed smart stethoscope system. (a) Appearance of the PCB
board. (b) Measurement scenario. (c) Measured signals from Lead II ECG
The bio-signal acquisition chip is fabricated in the 0.18 µm and pulmonic area PCG.
standard CMOS process. The microphotograph of the chip is
shown in Fig. 13. The performance is also summarized in Table TABLE VIII
VII. The power consumption of the whole chip is COMPARISON WITH RECENT BIO-SIGNAL ACQUISITION SYSTEMS AND
approximately 0.8 mW. The TIA accounts for around 90% of PRODUCTS
the power in this chip. It is necessary to endure the current up [29] [30] [31]
[32] This
to 40 µA, which is as high as the total current of CCIA and [33] work
HPSDM; moreover, the linearity of TIA should be retained CMOS1
Tech CMOS1 COTS2 COTS2 -
COTS
under such severe condition. Even so, the overall chip power is 1.2 V4
still much lower than that of the BLE module. Supply
1.2 V 3.7 V 3 V3 - 1.8 V4
voltage
On the basis of the designed chip, the proposed smart 3.3 V4
stethoscope circuits comprise the required peripherals that are ECG
BioZ
integrated in a printed circuit board (PCB). A photograph of the Measured
GSR ECG PCG
ECG ECG
PCB is shown in Fig. 14(a). The PCB has an area of 2.6×1.85 signal PCG PCG
PPG
cm2 and weight of 1.65 g. Fig. 14(b) shows the practical GPA
measurement scenario. The wet ECG electrodes are attached to Signal
R- R- R-
analysis N/A S1/S2
the front side of the human body around the chest surface. The Peak Peak Peak
stethoscope is placed near the region of the second intercostal Power
26.6 25 87 45.54
space to the left of the sternum. The Lead II channel ECG and consumption
µW5 mW5 mW3,6
-
mW6
pulmonic area PCG from the volunteer subject are measured.
Wireless N/A HBC7 Bluetooth 4.0 BLE BLE
These two signals are displayed on the smartphone. When Battery life N/A N/A N/A 9h >24 h8
measuring the PCG signal, the subject needs to hold the breath Certification N/A N/A N/A FDA TFDA
for seconds to avoid the mixing of breathing sounds. Typically, 1: 0.18 µm 1P6M standard CMOS process
2. COTS: commercially available off-the-shelf
the closure of mitral and tricuspid valves is the cause of S1; 3: Using 3V to estimate the possible power
meanwhile, S2 results from the closure of the aortic and 4: 1.2 V for AFE, 1.8 V for digital core, 3.3 V for BLE module
pulmonic valves. According to the relationship between ECG 5: Average channel power (without wireless transmission power)
6: With wireless transmission power
and PCG stated in [28], the position of S1 is approximately 7: Human body communication
0.04–0.06 s after the start of the QRS complex, and S2 occurs 8: Using a 360 mAh Li-ion battery
near the end of the T wave. The measured signals reveal such a
estimation should be introduced into the proposed system to
cross-reference relation as illustrated in Fig. 14(c). To improve
filter out the unidentifiable regions.
the efficiency of ECG analysis in practical use, the noise level

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

REFERENCES
[1] World Health Organization. “Cardiovascular diseases (CVDs).”
www.who.int. https://www.who.int/en/news-room/fact-
sheets/detail/cardiovascular-diseases-(cvds). (accessed May 16, 2019)
[2] 3M Littmann. Cardiac Auscultation. (2017). [Online]. Available:
http://multimedia.3m.com/mws/media/1372905O/3mtm-littmannr-
stethoscopes-auscultation-posters.pdf
[3] K. Phua, J. Chen, T. H. Dat, and L. Shue, “Heart sound as a biometric,”
Pattern Recognit., vol. 41, no. 3, pp. 906-919, Mar. 2008.
[4] F. Arvin, S. Doraisamy, and E. S. Khorasani, “Frequency shifting
approach towards textual transcription of heartbeat sounds,” Biol. Proced.
Online, vol. 13, no. 7, Oct. 2011.
[5] J. D. Bronzino, The Biomedical Engineering Handbook, 2nd ed. Boca
Raton, FL: CRC, 2000.
[6] W. J. Tompkins, Biomedical digital signal processing. Englewood Cliffs,
N.J.: Prentice Hall, 2000.
Fig. 15. Ultimate application scenario of the proposed system.
[7] M. F. C. Monteiro, H. Klimach and S. Bampi, “High linearity and large
output swing sub-Hz pre-amplifier for portable biomedical applications,”
V. CONCLUSION in 2014 27th Symp. on Integr. Circuits and Syst. Des. (SBCCI), Aracaju,
2014, pp. 1-7.
This work proposes a PCG and ECG synchronously [8] V. T. Nguyen, P. Loumeau and J. -F. Naviner, “VHDL-AMS behavioral
monitoring system. The implemented chip comprises CCIA, modelling and simulation of high-pass delta-sigma modulator,” in Proc.
of the 2005 IEEE Int. Behavioral Model. and Simul. Workshop, San Jose,
TIA, HPSDMs, and DSP blocks. A software application on CA, USA, 2005, pp. 106-111.
smartphone for PCG and ECG signal recording, display, and [9] S. Y. Lee, C. Y. Chen, J. H. Hong, R. G. Chang, and M. P. H. Lin,
detection is developed. A wavelet-based QRS complex “Automated synthesis of discrete-time sigma-delta modulators from
detection algorithm is also proposed. The beat detection system architecture to circuit netlist,” Microelectronics J., vol. 42, no. 2,
pp. 347-357, Feb. 2011.
accuracy in terms of sensitivity and positive prediction on the [10] E. B. Hogenauer, “An economical class of digital filters for decimation
basis of the MIT-BIH arrhythmia database are 99.82% and and interpolation,” IEEE Trans. Acoust., Speech, Signal Process., ASSP-
99.90%, respectively. According to recent studies listed in 29, no.2, pp. 155-162, 1981.
Table VIII, heart-related signal sensing systems, including the [11] M. A. Nidhi, S. Jackuline, and J. Jeevitha, “A new architecture for the
generation of picture based CAPTCHA: A level constrained CSE low
multi-channel AFE for cardiovascular and respiratory signal power synthesis methodology for fixed point FIR filters,” in 2011 3rd Int.
recording [29], acquisition system with level crossing sampling Conf. Electron. Comput. Technol. (ICECT), vol. 6, Apr. 2011, pp. 386-
[30], and PCG sensing system [31], have been developed. Some 389.
works focus on chip implementation only, achieve the [12] P. P. Vaidyanathan, S. K. Mitra, and Y. Neuvo, "A new approach to the
realization of low-sensitivity IIR digital filters," IEEE Trans. Acoust.,
completeness of the system with large power consumption, or Speech, Signal Process., vol. 34, no. 2, pp. 350-361, Apr. 1986.
sacrifice the quality of signal to reduce power consumption and [13] S. K. Mitra, and K. Hirano, “Digital all-pass networks,” IEEE Trans.
chip area. The products [32][33] similar to the proposed system Circuits Syst., vol.21, no.5, pp. 688-700, Sep. 1974.
has been invented. It is integrated into a stick-type handheld [14] Maxim Integrated. Determining Clock Accuracy Requirements for UART
(2003). [Online]. Available:
device to improve the convenience when measuring, but the Communications.
http://www.eecs.umich.edu/courses/eecs373.w05/lecture/AN2141.pdf
signal quality may be affected because of the dry electrodes. [15] Bluetooth SIG. “Topology Options.” www.bluetooth.com.
Furthermore, it is hard to keep both PCG and ECG signals https://www.bluetooth.com/bluetooth-technology/topology-options/
measured at their standard positions and leads simultaneously. [16] Bluetooth SIG. “Radio Versions.” www.bluetooth.com.
https://www.bluetooth.com/bluetooth-technology/radio-versions/
Since the human body is not flat, the curved surface will make [17] J. Tosi, F. Taffoni, M. Santacatterina, R. Sannino, and D. Formica,
the measurement more difficult especially for females. Fig. 15 “Performance Evaluation of Bluetooth Low Energy: A Systematic
illustrates the ultimate application scenario of the proposed Review,” Sensors, vol. 17, no. 12, Dec. 2017. doi:10.3390/s17122898.
system. The ECG patch with wet electrodes can sense the ECG [18] Apple. Accessory Design Guidelines for Apple Devices Release R9.
(2019). [Online]. Available: Guidelines.pdf
signal stably at a specific lead. Meanwhile, the use of traditional [19] H. Tanaka, K. D. Monahan and D. R. Seals, “Age-predicted maximal
stethoscope head can retain the original habit of auscultation. heart rate revisited,” J. American College of Cardiology, vol. 37, pp. 153-
Through the operation of stethoscope, four main regions of 156, Jan. 2001.
cardiac auscultation can all be checked. After detaching the [20] S. Y. Lee, P. W. Huang, M. C. Liang, J. H. Hong and J. Y. Chen,
“Development of an Arrhythmia Monitoring System and Human Study,”
stethoscope, the ECG patch itself can be applied to the patients
IEEE Trans. Consum. Electron., vol. 64, no. 4, pp. 442-451, Nov. 2018.
for long-term monitoring. Therefore, our proposed system is [21] MIT-BIH Arrhythmia Database. [Online]. Available:
more complete and flexible, and the power consumption is http://www.physionet.org/cgi-bin/atm/ATM.
reasonable and relatively low. The high resolution of the system [22] S. K. Bashar, Y. Noh, A. J. Walkey, D. D. McManus and K. H. Chon,
“VERB: VFCDM-Based Electrocardiogram Reconstruction and Beat
is beneficial to further analysis and disease classification. In the Detection Algorithm,” IEEE Access, vol. 7, pp. 13856-13866, 2019.
future, doctors can use this system to improve the efficiency of [23] R. Gutiérrez-Rivas, J. J. García, W. P. Marnane, and Á. Hernández,
cardiac auscultation. By establishing a database containing “Novel real-time low-complexity QRS complex detector based on
PCG and ECG signals, the educational environment and adaptive thresholding,” IEEE Sensors J., vol. 15, no. 10, pp. 6036–6043,
Oct. 2015.
learning efficiency will be improved.
[24] P. Phukpattaranont, “QRS detection algorithm based on the quadratic
filter,” Expert Syst. Appl., vol. 42, no. 11, pp. 4867–4877, July 2015.
ACKNOWLEDGMENT [25] S. Sahoo, P. Biswal, T. Das, and S. Sabut, “De-noising of ECG Signal and
QRS detection using Hilbert transform and adaptive thresholding,”
The authors would like to acknowledge the support of Procedia Technol., vol. 25, pp. 68–75, 2016.
Taiwan Semiconductor Research Institute, Taiwan.

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TBCAS.2019.2947694, IEEE
Transactions on Biomedical Circuits and Systems

[26] P. Li et al., “A 410-nW Efficient QRS Processor for Mobile ECG Chieh Tsou (S’18) was born in Kaohsiung, Taiwan,
Monitoring in 0.18-μm CMOS,” IEEE Trans. Biomed. Circuits Syst., vol. in 1992. He received the B.S. degree and currently
11, no. 6, pp. 1356-1365, Dec. 2017. study for the Ph.D. degree in Electrical Engineering
[27] Z. Hou, Y. Dong, J. Xiang, X. Li and B. Yang, “A Real-Time QRS from National Cheng Kung University, Tainan,
Detection Method Based on Phase Portraits and Box-Scoring Calculation,” Taiwan, in 2014.
IEEE Sensors J., vol. 18, no. 9, pp. 3694-3702, May 2018. His research interests include the design of
[28] J. Wartak, Phonocardiology: Integrated Study of Heart Sounds and baseband modulation ASIC, MCU with peripheral
Murmurs. New York, NY, USA: Harper & Row, 1972. interface, memory verification, decimation filter,
[29] J. Xu et al., “A 36 μW 1.1 mm2 Reconfigurable Analog Front-End for bio-signal detection algorithm with hardware design,
Cardiovascular and Respiratory Signals Recording,” IEEE Trans. Biomed. SOC integration, software design, firmware design,
Circuits Syst., vol. 12, no. 4, pp. 774-783, Aug. 2018. biomedical system platform setup and PCB design,
[30] J. Xiang, Y. Dong, X. Xue and H. Xiong, “Electronics of a Wearable ECG and animal testing.
With Level Crossing Sampling and Human Body Communication,” IEEE Mr. Tsou awards and accepted papers include excellent work in 2014 cell-based
Trans. Biomed. Circuits Syst., vol. 13, no. 1, pp. 68-79, Feb. 2019. digital circuit design hold by CIC, Taiwan, excellent work in 2015 Silicon
[31] H. Ren, H. Jin, C. Chen, H. Ghayvat and W. Chen, “A Novel Cardiac Awards held by Macronix, Taiwan, attend 2015 ISBB, 2019 ISSCC, and 2019
Auscultation Monitoring System Based on Wireless Sensing for AICAS.
Healthcare,” IEEE J. Transl. Eng. Health Med., vol. 6, pp. 1-12, 2018.
[32] Eko. “DUO.” ekohealth.com. https://ekohealth.com/duo/
[33] C. Landgraf, P. Goolkasian and T. Crouch, “Wireless Cardiac Sensor,”
U.S. Patent Appl. 20180256061A1, Sept. 13, 2018. Yu-Yi Liao was born in Changhua, Taiwan, in 1993.
He received the B.S. degree from National Cheng
Kung University, Taiwan, in 2016. He is currently
Shuenn-Yuh Lee (M’98-SM’15) was born in pursuing the M.S. degree from the Institute of
Taichung, Taiwan, in 1966. He received the B.S. Electrical Engineering, National Cheng Kung
degree from the National Taiwan Ocean University, University, Tainan, Taiwan.
Keelung, Taiwan, in 1988, and the M.S. and Ph.D. His research interests include the design of digital
degrees from the National Cheng Kung University, integrated circuits, software development, and
Tainan, Taiwan, in 1994 and 1999, respectively. machine learning.
He is currently a Professor at the Department of
Electrical Engineering, National Cheng Kung
University, Tainan, Taiwan. He served as the
Technical Program Chair (TPC) of the 2014/2015
International Symposium on Bioelectronics &
Bioinformatics (ISBB), and the 2015 Taiwan and Japan Conference on Circuits
and Systems (TJCAS). From 2013 to 2016, he serves as the Chairman of IEEE Ju-Yi Chen was born in Tainan, Taiwan, in 1974.
Solid-State Circuits Society Tainan Chapter. From 2016 to 2017, he serves as He received the M.S. degree from the Chang Gung
the Vice Chairman of IEEE Tainan Section. From 2016, he serves as the University, Taoyuan City, Taiwan, in 1999, and the
Associate Editor of IEEE Transaction on Biomedical Circuits and Systems. Ph.D. degrees from the National Cheng Kung
His present research activities involve the design of analog and mixed-signal University, Tainan, Taiwan, in 2013.
integrated circuits, biomedical circuits and systems, low-power and low-voltage He is currently an Associate Professor from 2013, at
analog circuits, and RF front-end integrated circuits for wireless the Department of Internal Medicine, National
communications. Dr. Lee now is a member of Circuits and Systems (CAS) Cheng Kung University, Tainan, Taiwan.
Society, Solid-State Circuits Society, and Medicine and Biology Society of His present research activities involve the
IEEE. He is also a member of IEICE. cardiovascular diseases, including arrhythmias,
hypertension, arterial stiffness, and cardiac
implantable electric devices.

Peng-Wei Huang (S’19) was born in Tainan,


Taiwan, R.O.C., in 1990. He received the B.S.
degree from National Cheng Kung University,
Taiwan, in 2013. He is currently pursuing the Ph.D.
degree from the Institute of Electrical Engineering,
National Cheng Kung University, Tainan, Taiwan.
His research interests include the design of digital
integrated circuits, software development, and
biomedical circuits and systems.
His awards and accepted papers include excellent
work in 2014 cell-based digital circuit design hold
by CIC, Taiwan, excellent work in 2015 Silicon Awards held by Macronix,
Taiwan, and attended 2015 ISBB.

Jia-Ren Chiou was born in Kaohsiung, Taiwan,


R.O.C., in 1991. He received the B.S. degree from
National Cheng Kung University, Taiwan, in 2014.
He is currently pursuing the M.S. degree from the
Institute of Electrical Engineering, National Cheng
Kung University, Tainan, Taiwan.
His research interests include the design of
baseband analog front end and sigma-delta
modulator.

1932-4545 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.

You might also like