Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

Sheet1

PROJECT-BPCL-MANMAD, IRGUR HAVELI, MUZAFFARPUR

TECHNICAL DATASHEET

S.No. Description (HC) GAS ANALYSER


1 Tag Number AIT-HCI & HCO
2 Service HC Concentration Monitoring
GENERAL

3 Line Size & Class 1/2"


4 Instrument Type Vapour Reœvery Unit
5 Area Classification ZONE-1, IIA/IIB,T6
6 Ambient Temperature 45°C Max
7 Fluid Name HC
8 Phase Gasoline Vapours + Air
CONDITION
PROCESS

9 Process Pressure In Barg Min. Nor. Max. - 1 2


10 Process Temperature In °C Min. Nor. Max. - 20 60
11 Sample Point Nozzle/Spool Length In MM NA
12 Sample Point Nozzle/Spool In MM OD ID NA NA
13 Sample Point Nozzle/Spool Connection Size & Class Type 1/2"NPT(F)
14 Make Emerson
15 Model XEFD (X-STREAM SERIES)
16 Enclosure Class IP66
ANALYZER TRANSMITTER

17 Enclosure Material Aluminium


18 Ambient Temperature Rating 0° to 50°C
19 Humidity (Non-Condensing) <90% RH @ 20°C/<70% RH @ 40°C
20 Area Classification Approval ATEX/IECEx Ex db IIB + H2 T4, Gb
21 Front Panel LCD With Key Pad
1 x Ethernet With Modbus TCP & 1 x RS485
22 Communication
4 wires With Modbus RTU
23 Analog Outputs 2 x Isolated 4—20 mA Outputs(Rb s 500 D
24 Digital Outputs 4 x Relay Output Rated 1A @ 30VDC
25 Electrical Entry 4 x M20
26 Measured Components HC HC
27 Bench Wise Principle NDIR NDIR
28 Detection Limit ≤ 0.5% ≤ 0.5%
29 Linearity ≤ 1% ≤ 1%
30 Zero-Point Drift ≤ 1% Per Week ≤ 1%Per Week
31 Span (Sensitivity) Drift ≤ 1% Per Month 1% Per Month
32 Repeatability ≤ 0.5% ≤ 0.5%
ANALYZER TRANSMITTER

33 Response Time (T90) 4Sec < T90 < 7Sec 4Sec < T90 < 7Sec
Accuracy ≤ 1% of FS ≤ 1% of FS
34 Gas Path Connection Gas Path Tubing 1/4" OD SS Tubing
35 Gas Path Arrangement 2 Gas Path In Parallel
36 Sample Pressure ≤7 psig
37 Sample Flow 0.2-1.5 l/min
38 Component Unit Measurement Range
39 Mesurement HC g/m3 0 to 2590 g/m3 (Inlet)
40 HC g/m3 0 to 17.7 g/m3 (Vent)
41 Power Supply 220VAC, 50Hz
42 Mounting Frame
43 Tag Plate Emerson Standard Tag Plate
44 Dimension (H x W x D) 512 x 610 x 292 mm
45 Weight
46 Line Type Non Heated
47 Size and Material 1/4" OD, SS316
CONDITIONING

48 Length 3 Mtrs
SAMPLE

SYSTEM

49 Mounting Plate Mounted


Major Components on SCS Ball Valve, Needle Valve, Catch Pot, Filter
50
and Pressure Gauge
51 Sample Return Atmospheric Safe Vent
52 Major Sample System Wetted Part SS316
53 Sample Stream 2 Nos
Power Supply 220VAC, 50 Hz 1 phase, UPS Power
UTILITY

54 Supply for Analyser.

55 Instrument Air @ 3 Barg & 35°C

Page 1