Download as pdf or txt
Download as pdf or txt
You are on page 1of 11

JC35DN Linear Actuator

Application field - Medical Line


Preferably in medical bed, nursing bed, operating
bed and other fields.
JC35DN is widely used with high performance and strong design. It has
following specifications: JC35DN-6000N(push), JC35DN-4000N(push),
JC35DN-3000N(push), JC35DN-2000N(push), JC35DN-1500N(push),
JC35DN-1000N(push), JC35DN-750N(push).
JC35DN is characterized by large load capacity, strong design, low
noise and high IP grade.

General Features
Load in push (N): 6000 | 4000 | 3000 | 2000 | 1500 | 1000 | 750

Motor rated voltage Input Input (VDC): 24


Installation dimensions: L=225 (S≤50) | L=S+175 (50﹤S﹤400) | L=S+195 (400≤S﹤600)

Stroke length: 6000N: 50-400mm (in steps of 4mm) | 4000N: 50-400mm (in steps of 4mm)
3000N: 50-600mm (in steps of 4mm) | 2000N: 50-600mm (in steps of 4mm)
1500N: 50-600mm (in steps of 4mm) | 1000N: 50-600mm (in steps of 4mm)
750N: 50-600mm (in steps of 4mm)
Options: Hall sensor, Only push, Reed switch

Colour: Black (RAL 9005) / Grey (RAL 7035)

Limit switch: Built-in (electrical)

Weight: About 1.5kg


(different stroke length and installation dimensions with different weight)
Duty cycle: 10%, max. 2min. continuous use

Max IP grade: IP66

Noise leve: ≤48dB (environmental noise≤40dB)

Static bending moment: Lateral load is not allowed

Flame retardant grade: UL94-HB

Standards: IEC60601-1: 2012


EN 60601-1: 2006+A12:2014
ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and
A2:2010/(R)2012

1 / 11
Dimension Drawing
Top clevis D, end clevis D
83
20 27 25
15 11

25
31.5

162.5
L=225(S≤50) S
L=S+175(50﹤S≤200)
L=S+185(50﹤S≤300)
Φ10.1 Φ10.1

Top clevis A, end clevis D


83
20 27 25
15 11
25
31.5

162.5

L=225(S≤50) S
L=S+175(50﹤S≤200)
L=S+185(50﹤S≤300)
Φ10.1 Φ10.1

2 / 11
Technical parameters

Input Spindle Max.Load Self Speed Speed Current


voltage pitch capacity locking no load full load full load
(VDC) (mm) (N) (N) (mm/s) (mm/s) (A)

24 4 6000 6000 5.5 3 4.5

29 4 6000 6000 6.7 3.8 4.5

24 5 4000 4000 6.8 4 4

29 5 4000 4000 8.2 5.2 4

24 7.5 3000 3000 10 6 4

29 7.5 3000 3000 12 7.5 4

24 5 2000 2000 13 9 4

29 5 2000 2000 16 11 4

24 7.5 1500 1500 20.5 13 4.5

29 7.5 1500 1500 25 6.5 4.5

24 12 1000 1000 33 22 4

29 12 1000 1000 40 26 4

24 16 750 750 44 25 4

29 16 750 750 53 36 4

Remark: the above data are typical values at 25 "C

3 / 11
Stroke length versus load
8
7

Load capacity ( KN )
6
5
4
3
2
1

0 100 200 300 400 500 600

Stroke length (mm)

Characteristic
Typical speed
Min. speed
Typical current
Max. current

6000N

12 12
11 11
Current ( A ) or Speed ( mm/s )

Current ( A ) or Speed ( mm/s )

10 10
9 9
8 8
7 7
6 6
5 5
4 4
3 3
2 2
1 1
0 0
1 2 3 4 5 6 1 2 3 4 5 6

Load capacity ( KN ) Load capacity ( KN )

4000N

12 12
11 11
Current ( A ) or Speed ( mm/s )

Current ( A ) or Speed ( mm/s )

10 10
9 9
8 8
7 7
6 6
5 5
4 4
3 3
2 2
1 1
0 0
1 2 3 4 5 6 1 2 3 4 5 6

Load capacity ( KN ) Load capacity ( KN )


4 / 11
Typical speed
Min. speed
Typical current
Max. current

3000N

12 12
11 11
Current ( A ) or Speed ( mm/s )

Current ( A ) or Speed ( mm/s )


10 10
9 9
8 8
7 7
6 6
5 5
4 4
3 3
2 2
1 1
0 0
0.5 1 1.5 2 2.5 3 0.5 1 1.5 2 2.5 3

Load capacity ( KN ) Load capacity ( KN )

2000N

24 24
22 22
Current ( A ) or Speed ( mm/s )

Current ( A ) or Speed ( mm/s )

20 20
18 18
16 16
14 14
12 12
10 10
8 8
6 6
4 4
2 2
0 0
0.5 1 1.5 2 2.5 3 0.5 1 1.5 2 2.5 3

Load capacity ( KN ) Load capacity ( KN )

1500N

24 36
22 33
Current ( A ) or Speed ( mm/s )

Current ( A ) or Speed ( mm/s )

20 30
18 27
16 24
14 21
12 18
10 15
8 12
6 9
4 6
2 3
0 0
0.3 0.6 0.9 1.2 1.5 1.8 0.3 0.6 0.9 1.2 1.5 1.8

Load capacity ( KN ) Load capacity ( KN )

5 / 11
Typical speed
Min. speed
Typical current
Max. current

1000N

48 48
44 44
Current ( A ) or Speed ( mm/s )

Current ( A ) or Speed ( mm/s )


40 40
36 36
32 32
28 28
24 24
20 20
16 16
12 12
8 8
4 4
0 0
0.2 0.4 0.6 0.8 1 1.2 0.2 0.4 0.6 0.8 1 1.2

Load capacity ( KN ) Load capacity ( KN )

750N

60 60
55 55
Current ( A ) or Speed ( mm/s )

Current ( A ) or Speed ( mm/s )

50 50
45 45
40 40
35 35
30 30
25 25
20 20
15 15
10 10
5 5
0 0
0.15 0.3 0.45 0.6 0.75 0.9 0.15 0.3 0.45 0.6 0.75 0.9

Load capacity ( KN ) Load capacity ( KN )

6 / 11
Ordering Key

Take JC35DN06324150/325HGDZD2022D0 for example

Model JC35DN

Optional features 0 0 = Standard / P = Only push / Y = Hall sensor / R = Reed switch

6 = 6000N / 4 = 4000N / 3 = 3000N / 2 = 2000N


Load in push 6
1D5 = 1500N / 1 = 1000N / 0D75 = 750N

3 = 3mm/s
4 = 4mm/s
6 = 6mm/s
Speed at full load 3 9 = 9mm/s
13 = 13mm/s
22 = 22mm/s
25 = 25mm/s

Input voltage 24 24 = 24 VDC

Stroke length/ Stroke length = XXX


150/325
lnstallation dimension Installation dimension = XXX

Clevis directions H H = H direction / V = V direction

Color G G = Grey (RAL7035) / B = Black (RAL9005)

End clevis specification D D = Slot 6.5, hole diameterΦ10.1

End clevis material Z Z = Zinc Aluminum alloy

A = Without slot, hole 10.1 with bushing


Top clevis specification D
D = Slot 6.5, hole 10.1 with bushing

0 = No brake
Brake specification 2
2 = Brake in push

02 = 4-pin DIN plug


Type of plug and cable 02
10 = 5-pin ET plug

Cable length 2D0 2D0 = 2m / 0D5 = 0.5m

7 / 11
Ordering Details

Dimension drawing of clevis

End clevis: D
20
15
Φ10.1
30

29
6.5 33 Φ30

Top clevis: D

26 27 11
Φ10.1
6.5
Φ28.5

31.5

28
Top clevis: A
27
11
Φ10.1
Φ28.5
28

8 / 11
Dimensions JC35DN with control box

JCB35Q install on JC35DN


145

120

255

L≥275
83

77

Motor wire

4-pin straight plug (standard) Pin Color Function


1 NC Not connected
Matching controller: JCB35Q/ 2 Brown When positive voltage power on,actuator will go down
JCB35T/ JCB35R/ JCB35R1/
3 Blue When positive voltage power on,actuator will go up
JCB35S/ JCB35K2
4 NC Not connected
Outer ring NC Not connected

60
22.5
Φ18.3

9 / 11
5 pin DIN plug Pin Color Function
(with Hall sensor)
1 Red 5V
Matching controller: JCB35Q/ 2 Brown When positive voltage power on,actuator will go down
JCB35T/ JCB35R/ JCB35R1/
3 Black GND
JCB35S/ JCB35K2
4 Blue When positive voltage power on,actuator will go up
5 Yellow Hall signal 1
Outer ring Orange Hall signal 2

60
22.5

Φ18.3

10 / 11
Strictly abiding by medical safety certifications
Passed EMC tests. To ensure our products pass all kinds of tests domestic or abroad.

Communication protocol development


Aiming at different applications, develop more extensive functions for control system.

Professional customization
Powerful R&D strength, fully satisfy customers’ needs. Development cycle is about 10
days, seizing more market opportunities

Intelligent Control of Bluetooth


Mobile phone as the remote, applying in home care. Market pioneer, Jiecang provides
better choices

Wireless Communication
Controllers adopt high-frequency wireless system. It can avoid interference mixing with
signal receivers, assisting in remote control over medical equipment.

Latest Weighing System


Designed for physiotherapy beds and medical care beds. Cantilever carrying weighing
sensor, it is applicable to different bedsteads dispensing with any adjustment

Notes for mounting and usage can be found in the


corresponding mounting manual

11 / 11