Download as pdf or txt
Download as pdf or txt
You are on page 1of 31

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/259194998

Bulk analysis of volatiles in fluid inclusions

Chapter · January 2003

CITATIONS READS
26 2,134

2 authors:

Stefano Salvi Anthony E. Williams-Jones


Paul Sabatier University - Toulouse III McGill University
137 PUBLICATIONS 3,608 CITATIONS 362 PUBLICATIONS 15,844 CITATIONS

SEE PROFILE SEE PROFILE

All content following this page was uploaded by Stefano Salvi on 22 May 2014.

The user has requested enhancement of the downloaded file.


CHAPTER 10. BULK ANALYSIS OF VOLATILES IN FLUID INCLUSIONS

Stefano Salvi1 and Anthony E. Williams-Jones2


1
Laboratoire des Mécanismes de Tranfert en Géologie,
UMR 5563, Équipe de Minéralogie
39 Allées Jules Guesde, 31000 Toulouse, France
salvi@cict.fr
2
Department of Earth & Planetary Sciences,
McGill University
3450 University Street, Montreal,
Quebec, Canada H3A 2A7
willyj@eps.mcgill.ca

“...a specimen of the Goldie aplite...was split in vacuo. The liberated gas was
conducted to a ... mass spectrometer, which unfortunately showed nothing.
Rechecking of the specimen by subsequent splitting in air yielded the characteristic
odor, which led us to conclude that the nose is more sensitive than the machine.” 1

INTRODUCTION
Although H2O is the dominant 1993, Nivin et al. 1995, Potter et al. 1998,
component of most fluid inclusions, researchers Potter & Konnerup-Madsen 2003). In addition
have long realized that, in some environments, to reduced carbonic species, the fluid inclusions
a significant proportion of the fluid inclusion may also contain appreciable proportions of N2,
population consists of inclusions in which non- H2S, and H2. Oxidized and reduced carbonic
aqueous fluids such as CO2, SO2, N2, H2, H2S, species are also commonly observed in fluid
CH4 and heavier hydrocarbons predominate. inclusions in sedimentary (e.g., Newman et al.
Less well known is the fact that most aqueous 1996, Giggenbach 1997, Burruss 2003a) and
fluids trapped as inclusions contain small metamorphic rocks (e.g., Eugster & Skippen
proportions of these gases as solutes. This is 1967, Kreulen & Schuiling 1982, Samson &
not surprising because, unless the dissolved Williams-Jones 1991, Moine et al. 1994), in
gases are in sufficient concentration to exsolve hydrothermal mineral deposits and in rocks that
as a separate phase (or form a clathrate) that can have been hydrothermally altered (e.g.,
be distinguished optically, their presence will Giggenbach 1993, Roedder & Bodnar 1997,
be undetected. Wang et al. 1999, Graupner et al. 2001,
Next to H2O, oxidized carbonic species Ishibashi et al. 2002, Simoneit 2002). Higher
(CO, CO2) are the most important gases hydrocarbon-bearing fluid inclusions (aliphatic
escaping from active volcanoes (cf. Gerlach and aromatic species) are particularly common
1980, 1989) and are commonly preserved as in sedimentary environments (e.g., Burruss
primary fluid inclusions in felsic igneous rocks 1981, Jensenius & Burruss 1990, Goldstein
(e.g., Roedder & Coombs 1967, Holloway 2001, Munz 2001, Burruss 2003a) but have also
1976, Roedder 1984). Reduced carbonic been reported to occur in serpentinites (e.g.,
species are released in relatively minor Agafonov & Andreyeva 1972, Berndt et al.
quantities, except during the emplacement of 1996, Peretti et al. 1992, Normand 2001). As
alkaline and hyperalkaline magmas, when they in igneous settings, the carbonic species, and
are the dominant carbonic species and particularly the reduced varieties, may be
commonly comprise methane and a number of accompanied by significant proportions of N2,
other aliphatic and even aromatic hydrocarbons H2S and H2. Other volatile species that have
(e.g., Petersilie & Sørensen 1970, Ikorski 1981, been reported to occur in fluid inclusions
Konnerup-Madsen et al. 1981, Salvi & include O2, NH3, SO2, H2S, HCl, HF, He, N2,
Williams-Jones 1992 and 1997a, Ikorski et al. and Ar and other noble gases.

1
Heinrich, E. Wm. and Anderson, R.J., 1965. Carbonatites and alkalic rocks of the Arkansas River area,
Fremont County, Colorado. 2. Fetid gas from carbonatite and related rocks. Am. Mineral., vol. 50, pp. 1915-
1916.

10-1
The challenge for fluid inclusionists is containing them will expand if the inclusions
to develop methods that can be used to are breached. By contrast, the vapor bubble of
accurately determine the concentrations of the an aqueous fluid inclusion devoid of other
various volatile species trapped during a single volatile species will collapse and disappear, as
geological event. Ideally, these methods should the vapor pressure of H2O is only 0.03 bars at
permit analysis of the volatiles in single fluid 25ºC.
inclusions, and, although such methods have A crushing test is initiated by placing a
been developed based on laser Raman and small grain of the fluid inclusion-bearing
Fourier Transform infrared spectroscopy mineral on the lower glass plate of a crushing
(Burruss 2003b), they are not universally stage, covering the grain with a few drops of
applicable. Owing to the very high detection immersion oil, lowering the upper glass plate
limits, important gases present in significant onto the grain and fastening the two parts of the
concentrations can be overlooked (the assembly (Fig. 10-1). The crushing stage is
minimum detection limit is >1 mole %), and then placed on a microscope stage, where the
because of the large minimum size of the glass plates are gradually brought closer
inclusion that must be analyzed (the minimum together by slowly tightening the knurled
diameter is commonly >10 µm) it may be fastening knob while observing the sample. As
difficult to find suitable inclusions for analysis. the pressure on the sample is increased,
An additional limitation, in the case of laser fractures begin propagating in various
Raman spectroscopy, is that it cannot be used to directions and, in some cases, will be seen to
analyze inclusions in some minerals or intersect fluid inclusions. If a fluid inclusion
inclusions that contain cyclic and aromatic breached in this manner contains volatile
hydrocarbons because of laser-induced species other than or in addition to H2O, the
fluorescence. The alternative is to use methods vapor bubble will expand (unless the proportion
that involve extraction of the bulk fluid from of these volatile species is extremely low) and
large numbers of inclusions. These methods be released into the surrounding immersion
permit the analysis of a wide variety of liquid. The identity of the dominant volatile
inclusion gases to very low concentration, and species released in this manner can often be
characterization of a fluid entrapment event in a established by repeating the crushing test with
single analysis. Moreover, they are effectively immersion fluids in which the particular gases
independent of the size of the inclusions. On are known to be extremely soluble, e.g., a
the other hand, they can only be applied in barium chloride solution for CO2 or kerosene
cases where petrographic examination and/or for CH4 (the bubble will shrink rapidly and
microthermometric analyses have demonstrated
that the sample contains, or is dominated by, a
single population of fluid inclusions.
In this chapter, we start by describing
the crushing technique, a widely used
qualitative method for detecting fluid inclusion
gases, and then proceed to a comprehensive
review of the principal methods used to
quantitatively analyze the composition of the
bulk fluid.

TECHNIQUES AND METHODOLOGY


Crushing tests
A simple, inexpensive, yet very
efficient way of detecting the presence of
volatiles in fluid inclusions, and in some cases
identifying them and roughly estimating their
proportions, is by performing a crushing test.
This technique makes use of the fact that,
except for H2O, these volatiles have vapor
pressures considerably higher than atmospheric FIG. 10-1. A crushing stage (A) and a sketch of
pressure at room temperature (~25°C), and the crushing stage designed by Roedder (B;
consequently vapor bubbles in fluid inclusions modified after Roedder 1984).

10-2
disappear into the immersion liquid). The in containers of known volume, reheating them,
pressure of the gas species, and, in turn, its and measuring the corresponding gas pressures
mole fraction in the fluid inclusion (assuming at a fixed temperature (this assumes that H2O
there is one, overwhelmingly dominant species, and CO2 dominate their respective groups).
and that it can be identified) can be roughly Analyses for gases, such as N2, CH4, H2, CO,
estimated by measuring the diameter of the He and Ar, which are not condensed by liquid
bubble in the immersion liquid (the gas should N2, require additional steps. In the case of H2
be relatively insoluble in this liquid) and and CO, this involves oxidization to H2O and
comparing this diameter to that of the vapor CO2 (at a temperature low enough to prevent
bubble in the fluid inclusion before breaching. oxidation of CH4) and separation of the latter
From these data, the corresponding change in gases through a second series of CO2-
volume can be estimated and used to calculate acetone/alcohol and liquid nitrogen traps (cf.
the pressure in the fluid inclusion using the Piperov & Penchev 1973). Methane is treated
relationship P1V1 = P2V2, where V1 is the the same way by repeating the oxidation at
volume of the bubble in the unbreached fluid higher temperature, N2 is separated by reacting
inclusion, V2 is its maximum volume in the the gas with hot calcium, and Ar and He are
immersion liquid, P1 is the pressure in the fluid separated using activated charcoal (cf. Piperov
inclusion and P2 is the pressure of the bubble in & Penchev 1973, Moore et al. 2001). Although
the immersion fluid (assumed to be 1 bar). the methodology is relatively simple and
Knowing volume and pressure, the number of inexpensive, the analyses are semi-quantitative
moles of the volatile species can be estimated and the detection limits are high.
using the ideal gas law PV = nRT, where n is The main and very important
the number of moles and R is the gas constant. advantage of using MS is the extremely low
The mole fraction of this species is obtained by detection limits, on the order of 10-9 µmoles (cf.
comparing n to the mass of the liquid, which is Shepherd et al. 1985, Norman & Musgrave
calculated from its density by estimating the 1994), which allows analysis of gases
inclusion volume and subtracting the volume of comprising a very small proportion of the bulk
vapor (cf. Roedder 1984, for further inform- fluid, and/or very small numbers of fluid
ation on the crushing method). inclusions (cf. Graney et al. 1991, Landis &
Hofstra 1991, Moore et al. 2001). However,
Quantitative techniques species identification is difficult, partly because
The techniques routinely employed for peaks in the mass spectrum may represent more
quantitative bulk analysis of fluid inclusion than one species (e.g., N2, CO, and C2H4 have
gases are based on mass spectrometry (MS) the same molecular weight of 28), and partly
(e.g., Shepherd et al. 1985, Barker & Smith because gases fragment on ionization to yield
1986, Kesler et al. 1986, Norman & Sawkins particles with different masses (cf. Norman &
1987, Guha et al. 1990, Landis & Hofstra 1991, Sawkins 1987, Landis & Rye 1989). The latter
Sasada et al. 1992, Ruff 1993, Newman et al. is particularly problematic because the
1996), and/or gas chromatography (GC) (e.g., fragmentation may produce masses
Petersilie and Sørensen 1970, Cuney et al. corresponding to species not present in the gas
1976, Andrawes & Gibson 1979, Behar & or contribute to a peak in the mass spectrum for
Pineau 1979, Burruss 1987, Bray & Spooner a gas that is present, rendering the measured
1989, Salvi & Williams-Jones 1997b). amount of the latter gas anomalously high; the
However, the bulk composition of fluid apparent amount of the fragmented gas will be
inclusion gases can also be analyzed anomalously low. For example, MS analysis of
volumetrically. This involves the use of cold an inclusion gas consisting only of CO2 will
traps to separate the gases into groups based on produce a spectrum with peaks corresponding
their freezing points. For example, by passing to masses of 12, 16, 28 and 44. The peak for
the bulk gas through a trap containing a CO2- mass 44 would predict an anomalously low
acetone/alcohol mixture (-70°C) followed by amount of CO2 and the peak for mass 28 could
one containing liquid N2 (-196°C), it can be be misinterpreted to indicate that N2, and/or
separated into an H2O group, a CO2 group, and CO, and/or C2H4 are also present in the
a group of gases not condensed by liquid N2 (cf. inclusion. These problems can be greatly
Piperov & Penchev 1973, Shepherd et al. alleviated by separating the gases cryogenically
1985). The proportions of H2O and CO2 are before analysis into an H2O group, a CO2 group
determined by placing the corresponding solids and a group that is not condensed by liquid

10-3
nitrogen (see above), and comparing the spectra 1987, Norman & Musgrave 1994). Continuous
for each group with standard spectra for monitoring of real-time analyses (on-line
appropriate pure gases (see below). extraction, see below) can also help unravel
A problem that cannot be addressed some of these problems, but requires very
satisfactorily by the above method is the careful calibration of the MS (cf. Graney &
pyrolysis of hydrocarbons (due to the high Kesler 1995, Williams 1996).
temperature of ionization), which can produce a Gas chromatography does not suffer
large number of lighter hydrocarbons, thereby from the same problem of species identification
yielding an extremely complex mass spectrum as MS, and permits analysis of virtually all
that may be impossible to resolve. Moreover, gases likely to be encountered in nature. It is
even if pyrolysis is not a problem, the number also non-destructive in the sense that most GC
of hydrocarbons species initially present in the detectors do not destroy the gases detected, and
inclusion fluid may be too large to permit consequently other types of analyses can be
resolution of the resulting mass spectrum. A conducted on them, e.g., isotopic analyses of
related and potentially serious problem is that the component elements. However, GC is
peaks in these spectra may overlap with those considerably less sensitive than MS, with the
of non-hydrocarbon species and their associated lowest reported detection limits being on the
ion fragments, making reliable analysis of the order of 10-6 µmoles (Bray & Spooner 1992),
latter impossible. For example, Norman et al. and thus the sample selected for analysis must
(1991) concluded that their measured content of be relatively large.
CO2 in some hydrocarbon-bearing fluid From the above discussion, it is
inclusions was unreliable because equilibrium apparent that GC and MS are complementary,
calculations indicated that CO2 should not be the former providing the advantage of
present at the concentrations detected. They unambiguous species identification and
also concluded that the analyzed CO content applicability to all gases, and the latter the
was an artifact because it correlated positively advantage of extremely low detection limits. In
with hydrocarbon content, and therefore view of this, and in order to capitalize on the
probably reflected pyrolysis of organic non-destructive nature of gas chromatographic
compounds. analysis, a potential approach to the bulk
Other gases that are difficult to analyze analysis of fluid inclusion gases is GC followed
reliably with MS, include H2O, H2S, SO2, HCl by on-line MS of the eluted gases (GC-MS).
and H2. The problem with water is that MS As one of the principal objectives of fluid
data reduction involves standardization against inclusion analysis is to determine the complete
N2 (see below), and it is difficult to prepare elemental composition of the fluid, it is also
standards with known quantities of H2O and N2 recommended that bulk gas analysis be
(Graney et al. 1991). Hydrogen sulfide is combined with in situ bulk leaching and
problematic because it adsorbs onto the metal analysis of the non-volatile residues of the
of the ion source and inlet line (cf. Shepherd et crushed sample (see below). In this way the
al. 1985, Norman & Sawkins 1987), and reacts analyst can be confident that the determined gas
with any water introduced into the MS to composition and elemental composition of the
produce a variety of other sulfur species non volatile phases apply to the same fluid
(Newman et al. 1996). Sulfur dioxide and HCl inclusion population (cf. Channer & Spooner
are commonly overestimated as they may be 1994, Channer et al. 1999, Graupner et al.
produced during extraction of the inclusion 2001, Gleeson 2003).
gases into the vacuum (Graney & Kesler 1995),
due to the increased stability of these species Methods of gas extraction
with decreasing pressure (e.g., Shinohara 1992). Historically, gas analysis of fluid
Finally, H2 may be produced by the filament inclusions proceeded in three steps, i.e., the
(cf. Bray & Spooner 1992). These problems gases were first released, then collected, and
can be largely resolved by using glass lines, finally an aliquot of the captured gases was
keeping the gas frozen until analysis (Norman introduced into the instrument. Extraction
& Musgrave 1994) and ensuring that H2O is techniques involved stepwise decrepitation by
removed cryogenically prior to analysis. Water heating (e.g., Piperov & Penchev 1973),
may be analyzed by weighing the cryogenically decrepitation under vacuum (Behar & Pineau
separated ice or determining the pressure of the 1979, Welhan 1988) or grinding the sample in a
corresponding vapor (cf. Norman & Sawkins stainless-steel or ceramic ball mill under

10-4
vacuum (e.g., Abell et al. 1970, Petersilie & cracking of hydrocarbons or H2O. Bray et al.
Sørensen 1970, Roedder 1972, Ikorski & (1991) subsequently modified the design of
Voloshin 1982) (Fig. 10-2). Among these Andrawes & Gibson (1979), developing an on-
techniques, the decrepitation extraction method line GC system that permits analysis of up to
has proven to be the most problematic because three samples within one session (cf. Thomas et
of thermal decomposition and heat-induced al. 1988 and 1990, Bray & Spooner 1992, Salvi
chemical reactions, which can significantly & Williams-Jones 1997b for further descrip-
alter the original gas compositions. This tions of this system). Similar on-line crushing
produces unreasonably high blank methods have been developed for MS analysis
concentrations of common inclusion (e.g., Graney et al. 1991, Sloan 1992, Ruff
components such as CO2, CH4 and N2 (cf. 1993, Williams & Taylor 1996, Moore et al.
Norman & Sawkins 1987). However, all these 2001). The principal difference is that the
extraction techniques share another major sample chamber and extraction lines are under
weakness, namely that only a small proportion partial vacuum, because the MS operates at
of the released volatiles can be collected for reduced pressure. The released volatiles are
analysis. This significantly decreases the thus "sucked" into the MS. As is the case for
accuracy of the analysis and the ability to detect GC analysis, the extraction lines are heated to
trace components. minimize volatile adsorption (cf. Newman et al.
A substantially more efficient 1996, Parnell et al. 2001). In contrast to the GC
approach to the analysis of bulk inclusion gases
is to release the gases liberated by crushing or
decrepitation directly into the analytical device,
i.e., to extract the gases "on-line". Among the
first to conduct analyses of this type were
Andrawes & Gibson (1979), who designed an
on-line crushing device that they used in
conjunction with a gas chromatograph (see also
Andrawes et al. 1984, Barker & Smith 1986,
Burruss 1987). The device consists of a
piston/cylinder operated by a hydraulic press
and a crushing chamber where the inclusion
gases are released into a carrier gas, which
transports them to the GC (Fig. 10-3). The
sample chamber and all extraction lines are
heated to ~150ºC in order to reduce gas
adsorption and to induce immediate release and
dilution of the volatiles into the carrier gas.
This temperature is also low enough to prevent

FIG. 10-3. Schematic design of a piston-cylinder


crushing system for on-line GC analysis of
fluid inclusion volatiles (modified from
FIG. 10-2. A sketch of an off-line grinding Andrawes & Gibson 1979 and Bray &
system for fluid inclusion volatile extraction Spooner 1989; cf., Salvi & Williams-Jones
(modified after Petersilie & Sørensen 1970). 1997b).

10-5
method and because ionization involves heating All mass spectrometers consist of three
the sample to high temperature, thermal distinct parts, i.e., an ion source where ions are
decrepitation (typically at a temperature of created from gas molecules, a mass-selective
~1000°C) is also widely used for gas extraction analyzer (filter), which separates the ions
in MS analyses (e.g., Barker & Smith 1986, according to their m/e, and an ion detector,
Kesler et al. 1986, Norman & Sawkins 1987, which measures the quantity of ions for each
Guha et al. 1990, Moore et al. 2001) and offers m/e (Fig. 10-4). Some mass spectrometers, i.e.,
the advantage of nearly instantaneous transfer those with magnetic-sector, quadrupole, and
of the inclusion gases to the ionization time-of-flight designs, also have optical
chamber. However, as a result of the systems to facilitate extraction and transfer of
development of better vacuum systems, this ions from the source region to the analyzer.
advantage has largely been eliminated, and Quadrupole mass spectrometers (Fig. 10-5) are
crushing is now the preferred method of by far the most widely used in fluid inclusion
extraction for bulk analyses (J.N. Moore 2003, gas analysis because they have very high
pers. comm.). On the other hand, the thermal scanning speeds, and are therefore ideally
decrepitation method still offers one important suited to analyze gases introduced in very short
advantage, namely that the gas spectrum can be bursts as is the case during crushing or
continuously scanned as the sample is heated, decrepitation of fluid inclusions. Another
allowing gas compositions of sub-populations advantage of the quadrupole mass spectrometer
of inclusions (or even single inclusions) is the low voltage applied to the ion source, i.e.,
decrepitating at different temperatures to be the kinetic energy of the ions is on the order of
determined separately (Landis & Hofstra 1991). 5x10-10 eV, compared with several keV for a
Although on-line extraction techniques sector instrument. This eliminates problems
are preferred for releasing volatiles from fluid related to high voltage, and makes interfacing
inclusions, whether for GC or MS analysis to a GC easier (cf. March et al. 1989).
(e.g., Newman et al. 1996, Volk et al. 2000, Additional advantages of the quadrupole mass
Parnell et al. 2001), off-line extraction spectrometer are its superior transmission
techniques are still used by some researchers efficiency and wide acceptance angle, which
(e.g., Karlsen et al. 1993, Ruble et al. 1998, give it very high sensitivity, its small size, its
George et al. 1997, 2001, Munz 2001), relatively low cost and its ease of operation.
particularly for GC-MS analysis of petroleum Figure 10-6 shows the schematics of an on-line
inclusions. For example, Guilhaumou et al. system for fluid inclusion volatile analysis by
(2000) used micro-drilling under a microscope QMS. Time-of-flight mass spectrometers have
to extract oil from selected inclusions, thereby also been employed for fluid inclusion gas
avoiding the averaging effect of bulk methods. analysis (cf. Williams 1996) and make use of
Off-line crushing is recommended for heavy differences in transit time through a drift region
hydrocarbons, which are in liquid form and not to separate ions of different masses. An electric
readily volatilized, but requires the use of field accelerates all the ions into a field-free
organic solvents and precludes analysis of the drift region with a kinetic energy of qV, where
volatile fraction (cf. Ruble et al. 1998) unless q is the ion charge and V is the applied voltage.
relatively complex analytical protocols are As the ion kinetic energy is 0.5mv2, lighter ions
followed (e.g., leaching in a vacuum chamber,
cf. Munz 2001).

MASS SPECTROMETRY
A mass spectrometer creates charged
particles (ions) from molecules and then uses
the difference in mass-to-ionic charge ratio
(m/e) of the ionized species to separate them
from each other (e.g., Johnstone & Rose 1996,
Breci 2003). Mass spectrometry therefore
provides quantitative data on the amounts of
each molecular species as well as information
about their structures and other aspects of their FIG. 10-4. Flow chart depicting the principal
chemistry. steps involved in mass spectrometric analysis.

10-6
FIG. 10-5. Schematic design of a quadrupole MS (modified after Breci 2003).

FIG. 10-6. Design for MS analysis of fluid inclusion volatiles (adapted from Graney & Kesler 1995).

10-7
have a higher velocity than heavier ions and Fragmentation of larger molecules can
reach the detector at the end of the drift region proceed along several paths and lead to a large
sooner. Advantages of time-of-flight mass variety of smaller fragments. For example, the
spectrometers are that they can analyze a wider molecular ion of methanol may decompose
range of m/e than any other type of MS and do according to the reactions:
so more rapidly (e.g., Brunnee 1987).
[CH3OH] +.(molecular ion) →
However, instruments of this type are not well-
[CH2OH] +(fragment ion) + H+.
suited to fluid inclusion analysis because they +.
(or) [CH3OH] (molecular ion) →
operate in a pulsed mode requiring that ions be
[CH3] +(fragment ion) + [OH] + + .
produced or extracted in pulses.
(. is a radical electron)
Analysis of fluid inclusion gases by
MS takes place in the following steps regardless and additional fragments may be created by
of the type of mass spectrometer used: further breakdown of the larger fragments
produced by these two reactions (e.g., De
Extraction → Ionization → Mass Analysis →
Hoffmann et al. 1996, Johnstone & Rose 1996).
Ion Detection → Data Analysis
The fragments are actually charged ions with a
particular mass, and as most fragments have a
Ionization methods
charge of +1, the m/e usually represents the
The two most commonly used
molecular weight of the fragment.
techniques to ionize gases involve electron
bombardment (electron ionization) and
Chemical ionization (CI) is used mainly when
chemical reactions that remove non-bonding
there is a need to enhance the abundance of the
electrons (chemical ionization) (cf.,Breci 2003).
molecular ion, and involves reacting the analyte
The ionization takes place at reduced pressure
molecules with a reagent capable of converting
(<10-7 torr) in order to prevent collisions of ions
them to ions by either proton or hydride
with residual gas molecules in the analyzer
transfer:
during their flight from the ion source to the
detector.
MH + C2H5+ → MH2+ + C2H4
MH + C2H5+ → M+ + C2H6
Electron ionization (EI) is widely used in mass
spectrometry for relatively volatile samples that
where M is the molecule of interest. The
are insensitive to heat and have relatively low
reagent ions are produced by introducing a
molecular weight. It is the preferred ionization
large excess of methane (relative to the analyte)
method for fluid inclusion volatiles. An EI
into an electron impact (EI) ion source.
source uses a high-energy electron beam (70
Electron collisions produce CH4+ and CH3+,
eV), usually generated from a tungsten
which further react with methane to form CH5+
filament, to knock electrons off analyte
and C2H5+.
molecules to create ions (e.g., De Hoffmann et
The two ionization methods described
al. 1996, Johnstone & Rose 1996). EI usually
are limited to the molecular weight range 50 to
produces singly charged ions containing one
800 Dalton, although in rare cases it is possible
unpaired electron. A charged molecule that
to analyze samples of higher molecular weight
remains intact is called the molecular ion.
(Vidavsky 2003). The accuracy of the mass
However, energy imparted by the electron
measurement at low resolving power is ± 0.1
impact and instability in a molecular ion can
Dalton and at high resolution is ± 5 ppm (De
cause that ion to break (crack) into smaller
Hoffmann et al. 1996).
pieces (fragments). Thus, for example, while
Other ionization methods include
the water ionization reaction is:
electrospray (ESI), fast atom bombardment
H2O + 1 (fast) electron → [H2O] + + (FAB), matrix-assisted laser desorption
2 electrons (MALDI), laser ionization (LIMS), plasma-
desorption ionization (PD), resonance
the ionization process will also produce
ionization (RIMS), secondary ionization
fragments of [OH]+, O+, and H+. Consequently,
(SIMS), and spark source thermal ionization
the mass spectrum of water will have four
(TIMS) (cf. Breci 2003). However, EI is the
peaks corresponding to masses of 1 (H+), 16
preferred ionization technique for fluid
(O+), 17 ([OH] +), and 18 ([H2O] +).
inclusion volatile analysis by MS.

10-8
Ion sorting and is performed by introducing a well
Molecular and fragment ions are characterized compound into the instrument,
accelerated through the mass spectrometer by and "tweaking" the circuits until the amounts of
using electromagnets to manipulate the charged the molecular and fragment ions are reported
particles; uncharged molecules and fragments correctly.
are pumped away. The quadrupole mass Because the sensitivity of quadrupole
analyzer employs positive (+) and negative (-) mass spectrometers has increased so much over
voltages to control the path of the ions, which the past few years, the amount of material
travel along trajectories that are determined by needed for analysis is sufficiently small that it
their m/e ratios. As EI ionization mostly is now possible to determine the composition of
produces singly charged particles, ions having very small populations of fluid inclusions or, in
the same mass will follow the same path. The some cases, even a single fluid inclusion (e.g.,
quadrupole (a group of four electromagnets) is Barker & Robinson 1984, Barker & Smith
programmed by the computer to only direct 1986, Sommer & Gibson 1986, Moore et al.
fragments with a particular m/e through the slit 2001). Nonetheless, as discussed earlier,
at any given time; the rest bounce away. By complex procedures and detailed analytical
changing voltages continuously, ions of ever- protocols have to be followed in order to avoid
increasing mass can find paths through the or minimize problems related to peak overlap,
electromagnets to the detector. The cycle is adsorption etc. (cf. Graney & Kesler 1995).
then repeated rapidly; 102 – 104 cycles or scans
can be performed per second depending on the Data analysis
number of masses analyzed. The computer Owing to the fact that ionization
records a spectrum for each scan, which may be causes fragmentation of the molecules
printed as a graph with an x-axis representing introduced into the MS, the mass spectrum for a
the m/e ratios and a y-axis representing the fluid inclusion gas containing even a small
signal intensity (i.e., abundance) for each m/e number of species can be relatively complex
(Fig. 10-7) (e.g., Johnstone & Rose 1996). and include extensive overlap between
molecular and fragment ions. Consequently,
Detection the data are reduced using a matrix algebra
There are many types of detectors, but method in which the mass spectrum is
most operate by producing an electronic signal compared to spectra of reference gases that may
when struck by an ion. Timing mechanisms, be present in the inclusion. The latter are
which integrate these signals with the scanning analyzed to establish their fragmentation
voltages, allow the instrument to identify and patterns and determine the ionization efficiency
report the m/e of the fragment striking the of each gas relative to N2, i.e., the response of
detector. The mass analyzer sorts the ions the detector to equal amounts of different gases
according to m/e and the detector records the (Landis & Hofstra 1991; see also Graney &
abundance of each m/e. Regular calibration of Kesler 1995). These data are then used to
the m/e scale is necessary to maintain accuracy,

FIG. 10-7. Mass spectrum of CO2. M+ designates the molecular ion (adapted from Chiu & Muddiman
2001).

10-9
prepare a calibration matrix and sensitivity elute from the column and enter a detector. The
vector multiplier, which are combined into a time from when the injection is made (time
general matrixA. The composition of the zero) to when elution occurs is referred to as the
unknown gas is determined by solving the retention time (RT). The column is contained
matrix equation: in an oven that can be programmed to increase
temperature gradually (or in GC terms, ramp
Amn Pn = Im (1) temperature). As the temperature increases,
where m is the number of mass peaks, n is the compounds with low boiling points elute from
number of gas species, Amn is the general the column sooner than those that have higher
calibration matrix, Im is a vector containing boiling points. One or more detectors
the measured intensities of the different mass (connected in series or in parallel) analyze the
peaks, and Pn is a vector of the partial components as they exit the column and send
pressures of the unknown gases. The vector the data to a chart recorder or computer-based
data handling software (Fig. 10-8) (e.g.,
Pn is calculated by solving the equation using
standard methods involving an iterative least Jennings 1987).
squares procedure. The equation is solved While the instrument is in operation,
the computer generates a graph from the signal.
several times for different sets of gases that
This graph is called a chromatogram (see
might reasonably be predicted to be present in
the inclusion fluid (a gas is assumed to not be below). Each of the peaks in the chromatogram
present and excluded from the matrix if the represents the signal created when a compound
most abundant ion fragment for this gas is not elutes from the GC column into the detector.
detected) and the residuals for each mass The x-axis displays the RT, and the y-axis the
intensity are determined. The latter represent intensity of the signal (proportional to the
the difference between the measured intensity abundance of the species analyzed).
Gas chromatography has been
and that calculated by summing the percentage
employed for the analysis of volatile
contribution of each gas to the intensity,
weighted by its calculated partial pressure constituents of bulk samples of fluid inclusions
for over thirty years (e.g., Goguel 1963,
(Pn). The reported analysis is provided by
Petersilie & Sørensen 1970, Mironova et al.
that solution for which the average residual and
1972, Ypma 1974, Cuney et al. 1976, Behar &
individual residuals for each mass are
Pineau 1979). However, until recently, the high
optimized to minimum values. Results are
detection limit has restricted analysis to the
normally presented as ratios of partial pressures
major gases. With advances in gas chromato-
of the different gases, mole fractions or mole
graphic technology and the development of on-
percentages, and are considered to have a
line extraction techniques, the sensitivity of the
precision of 3 to 5% (Landis & Hofstra 1991).
method has improved greatly, and detection
limits now range from 10-4 to 10-6 µmoles,
GAS CHROMATOGRAPHY
depending on the detector used. This makes it
Any substance that can be introduced
possible to extend the analyses to include
in its gaseous state into a gas chromatographic
components comprising <0.01 ppm of the fluid.
column can be potentially analyzed by this
technique. The sample (generally consisting of
Choice of column
a mixture of components) is introduced through
Gas chromatograph columns are of
a heated injector, which causes its vaporization
two types: packed or capillary (Fig. 10-9).
(if not already a gas, e.g., permanent gases and
Packed columns are typically stainless steel
light hydrocarbons) and is immediately fluxed
coils filled with a stationary phase, whereas
in a stream of pressurized gas. Helium, H2, Ar,
capillary columns are thin fused-silica tubes
N2, CO2 can all be used as carrier gases, but He
coated with a polar or non-polar stationary
is best-suited for most applications. The gas
phase on the inner surface. The stationary
stream then passes through a stationary phase,
phase interacts with the species in the gas
which is contained in a tube of some sort,
mixture, causing their separation. In columns
referred to as a column (two parallel columns
coated with a non-polar phase, the extent to
are also used). The stationary phase is usually a
which species are retarded depends on their
chemical compound that can selectively attract
volatility, whereas in polar columns it depends
components in a sample gas mixture causing
almost entirely on interactions between
the different species to be retarded to differing
functional groups of sample components and
degrees. As the compounds are separated, they

10-10
FIG. 10-8. Flow chart depicting the principal steps involved in (A) regular gas chromatography and (B) on-
line gas chromatographic analysis of fluid-inclusion volatiles.
those contained in the stationary phase. gases most commonly present in fluid
Capillary columns provide higher separation inclusions, and are relatively inexpensive (e.g.,
efficiency than packed columns but are more Bray et al. 1991). On the other hand, capillary
easily overloaded by too much sample. Factors columns are better adapted to deal with the high
influencing retention times are the length and molecular weight species commonly present in
internal diameter of the column, the thickness hydrocarbon-bearing fluid inclusions, but until
of the coating, the oven temperature, and the recently could not be used for fluid inclusion
carrier-gas flow rate (cf. Kaliszan 1987). analysis because the allowed flow rates were
The ability of a chromatographic too low. This problem has been overcome by
system to separate complex mixtures depends the introduction of a megabore class of
almost entirely on the choice of column. A capillary columns, which have the added
packed column (Fig. 10-9A) typically consists advantage that they can be used to analyze H2O
of 0.5 m to 3 m of stainless-steel (Ni alloys and (see below).
silica glass are also used) tubing, generally has
an internal diameter (ID) between ~2 mm and Choice of detector
~5 mm, and is loaded with stationary phases, Detectors that have been used for fluid
porous polymers, or molecular sieves. inclusion volatile analysis include the thermal
Capillary columns (Fig. 10-9B) usually conductivity detector (TCD) (e.g., Behar and
comprise between 12 m and 100 m of coiled Pineau 1979, Shephard et al. 1985, Smith et al.
fused silica (stainless-steel is also used) tubing 1984, Bray & Spooner 1989), the flame
of very small ID (the usual range is from 0.18 ionization detector (FID) (e.g., Behar & Pineau
mm to 0.53 mm), and are coated with a µm- 1979, Shephard et al. 1985, Smith et al. 1984),
scale thickness of stationary phases or porous the helium ionization detector (HID) (e.g.,
polymers (recently developed), molecular Andrawes & Gibson 1979), the photoionization
sieves, aluminum oxide or carbon. There are detector (PID) (Bray & Spooner 1989) and the
over a thousand different columns on the flame photometric detector (FPD) (e.g.,
market today, designed to meet the demands of Andrawes & Gibson 1979). Of these, TCDs are
a wide variety of applications involving a large the most widely used for detection of inorganic
number of different gases. Packed columns are gases such as H2, O2, CO, N2, Ar, CO2, and
preferred by most fluid inclusion researchers H2O; FPDs are the detector of choice for sulfur-
because they are able to handle a broad range of and phosphorus-bearing compounds, but their
sample volumes, satisfactorily separate the light application is limited to these species; FIDs

10-11
C2H2, C2H4 and C2H6, which are more tightly
bonded and hence have higher ionization
potentials than heavier molecular weight
compounds. A substantially wider range of
molecular weights can be detected with the
higher energy, 11.7 eV lamp, but this lamp has
an expected lifetime of only ~500 hrs and is not
designed to be operated at temperatures greater
than 125°C. The window is constructed from
lithium fluoride in order to be transparent to the
low-wavelength emission lines of argon in the
11.7 eV lamp, and consequently undergoes
solarization (color center formation) with time,
particularly at higher temperatures (cf. Driscoll
& Duffy 1987). Methane ionizes at ≥12.98 eV
and therefore does not produce a linear
response even with this lamp. However, it can
be analyzed with the TCD down to masses
below 10-4 µmoles. Many inorganic gases
ionize at >12 eV (N2, CO, CO2, CH4, SO2 and
H2O) and cannot be detected with the PID. As
noted above, the TCD is the detector of choice
for the analysis of these gases.

Principles of detection
Micro-thermal conductivity detector (TCD)
A TCD measures the difference in
thermal conductivity between the effluent flow
from the column (carrier gas + sample
FIG. 10-9. Examples of A) a packed column and components) and a reference flow of carrier gas
B) a capillary column. alone. A voltage is produced that is proport-
owe their popularity to their wide dynamic ional to this difference, i.e., to concentration,
range (ability to deal with a wide range of and generates the output signal. In the original
concentrations), universal response to organics, design, one TCD analyzed the column effluent
and simple design, but have relatively high while a second received a reference flow from a
detection levels; HIDs have greater sensitivity separate column. The resistances of the two
than other detectors (1 x 10-6 µmoles for most sets were compared in a bridge circuit. The
compounds), but are difficult to use (for TCD employed in the system described below
example, Andrawes & Gibson (1979) had to uses a single filament, and examines the
add H2 to the carrier gas to provide positive reference and column effluent gas streams
responses from their HID), cannot analyze H2O, alternately (switching every 200 msec) (Fig. 10-
and are very costly; PIDs have a dynamic range 10). This is a convenient design but causes
similar to FIDs (Driscoll & Duffy 1987), lower sample dilution of ~50 % at the detector vent.
detection limits (cf. Table 2 in Bray et al. Therefore, it is preferable to connect this TCD
1991), and respond to all compounds with an model to the exhaust port of the PID.
ionization potential ≤12 eV.
An analytical approach that is suitable Photoionization detector (PID)
for many applications is to direct the gas flow Figure 10-11 shows the schematics of
sequentially through a PID and TCD (which is a PID, which consists of a lamp (Fig. 10-11B)
possible because the PID analysis is non- and an ionization chamber (Fig. 10-11C). The
destructive). PIDs can be equipped with lamp is filled with a low pressure noble gas (Ar
ultraviolet lamps having energies ranging from in the 11.7 eV lamp), which produces an
8.3 eV to 11.7 eV. The 10.2 eV lamp has the emission line of specific wavelength in the far
greatest sensitivity of the available light sources UV region of the spectrum when excited by
(cf. Driscoll & Duffy 1987), but cannot detect electrical discharge. The flowing gas is
low-molecular-weight compounds such as CH4, exposed to the UV light in the ionization

10-12
FIG. 10-10. Schematic cross-section of the micro-thermal conductivity detector used on the GC at McGill
University (designed by Hewlett Packard®, now Agilent®) and its external components. A solenoid
switching valve allows the detector to analyze alternately (every 200 msec) (A) the column effluent gas
and (B) a flow of carrier gas alone (from Salvi 1995).
chamber and is ionized by absorption of the UV
radiation according to the expression: Data collection and processing
+ - If the GC conditions (oven temperature
R + hv = R + e (2)
ramp, column type, etc.) are the same, a given
where R is the species to be ionized, and hv gas species will always elute from the column
(Planck’s constant times frequency) describes a at nearly the same RT and be accompanied by a
photon with energy greater than the ionization signal created by the detector, which is
potential of R. The ions become part of an recorded on a graph in form of a peak. The
electric field, and the resulting current, which is graph is a chromatogram displaying the RT on
proportional to concentration, is measured by the x-axis and the intensity of the signal
an electrometer and sent to the data collection (corresponding to the amount of the gas) on the
interface. y-axis.

10-13
FIG. 10-11 (A) Schematic cross-section of the photoionization detector used on the McGill system (PI-52-
02A model of HNU® Systems), and exploded views of (B) the lamp assembly and (C) the ionization
chamber (from Salvi 1995).
Historically, chromatograms were (cf. Bray & Spooner 1989). The gas extraction
printed on chart recorders, and peak integration part of the McGill system is based on the design
had to be done by hand. Nowadays, most of Andrawes & Gibson (1979) and employs
machines are connected to a computer and the crushing stages supplied by the University of
data is processed digitally by software Toronto. The GC is an HP® 5890 equipped
specifically designed for the purpose. with a PID and TCD, and a mega-bore capillary
Computer-based manipulation of spectra column for gas separation.
represents a major advance over hand treatment In the above system (Fig. 10-12), the
of chart records. For example, there is no carrier gas (ultra-high purity He) is routed to a
limitation on the scale to which specific three-way valve that directs it either to the
segments of a signal trace can be enlarged. internal flow controller of the gas
This allows recognition of small peaks that chromatograph, used with most capillary
previously would have gone undetected. columns (these require a split flow, thus the
Computer-manipulation also eliminates the injector is configured as in Fig. 10-13A), or to
danger of off-scaling, as discussed by Bray et an external regulator (0 to 400 kPa), which
al. (1991), who reported that they had to switch mimics the controller generally employed with
the TCD to low sensitivity before elution of packed columns (i.e., unsplit flow, injector
CO2 and H2O to prevent loss of data by peak configured as in Fig. 10-13B). Both flow
truncation. As long as the signal is not stronger controllers are connected to a second three-way
than 1 V, which is highly unlikely for the valve that directs the selected flow to a
amounts of gas typically extracted from fluid switching panel comprising four three-way
inclusions, the entire spectrum is stored by the valves (control valves) (Fig. 10-14). From
software. there the carrier gas can be directed to any of
the three crushers or to the injector by opening
Example of a system for on-line GC analysis the appropriate control valve to carrier gas
In the paragraphs below, we describe flow. All four injection ports are routed to a 6-
in some detail the GC system installed at port valve, which selects one of the entry ports
McGill University (Salvi & Williams-Jones while venting all others at a common exit (Fig.
1997b). A similar system can be found in the 10-14). The carrier gas passes through the
F. Gordon Smith Fluid Inclusion Laboratory, selected entry port where it is directed to the
Department of Geology, University of Toronto

10-14
10-15
FIG. 10-12. A schematic diagram showing the gas chromatographic system for analysis of fluid-inclusion volatiles at McGill University (after Salvi & Williams-Jones
1997b). The design is similar to that of the gas chromatographic system installed in the F. Gordon Smith Fluid Inclusion Laboratory at the University of Toronto.
FIG. 10-13. Schematic cross-section of the split/splitless gas chromatograph inlet depicting the gas flows in
(A) the split configuration typically used with capillary columns and (B) in the splitless mode, compatible
with packed columns (after Salvi & Williams-Jones 1997b).
oven of the GC and into the chromatographic second cylinder flows to two metering valves
column. mounted on the switching panel. From here the
A second tank of ultra-high purity He gas is directed to the other end of the four
(labeled ‘auxiliary gas’ in Fig. 10-12) is used to control valves allowing the crushers and/or
provide the confining pressure in the upper injector that are not on-line with the column to
portion of the crushers, and to supply He for be flushed with a stream of He at a flow rate
flushing the crushing chamber in preparation regulated by the metering valve. The He
for analysis (Fig. 10-12). The gas from the employed for flushing and that as the carrier

10-16
FIG. 10-14. A schematic diagram illustrating the operation of the 6-port valve used to direct one of the
sample inlets to the gas chromatographic column. As one port is selected to allow gas flow to the
column, all others continue purging to a common exit (after Salvi & Williams-Jones 1997b).
gas are isolated from each other but vent at a m x 0.53 mm ID particle trap, connected at the
common exit controlled by the 6-port selector start of the column.
valve. A shut-off valve fitted on this exit The mega-bore capillary column (0.53
allows the plumbing system to be closed and mm ID) provides the high capacity needed to
the pressure to build up as needed. The second separate volatile compounds as well as handle
metering valve adjusts the flow to the top the wide range of sample volumes that are
chambers of the crushers where shut-off valves intrinsic to fluid inclusion gas analysis. The
on the chamber outlets allow the system to be wide bore also allows for the fast flow rates
pressurized. required to deal with the large dead volumes in
In order to minimize absorption/ the extraction system, thereby obviating the
condensation of sample components between need to split the flow (cf. Fig. 10-13A),
the crushers and the column, gas lines and something which creates problems when there
connectors were constructed from Ni alloy is a crusher on line. The principal drawback of
stainless-steel and the valves with Hastelloy®. most mega-bore capillary columns is that they
In addition, all these parts are heated to ~120°C cannot be used in the presence of H2O.
during runs. To prevent rock particles and dust However, this problem is overcome by the
from damaging the selector valve and the recently developed PLOT (porous layer open
column, a silk-screen particle filter (~200 µm tubular)-type columns. The McGill University
pores) and a 1-µm frit filter are connected in system uses a fused-silica PoraPLOT® Q
series to the sample chamber outlets of the column manufactured by Chrompack Inter-
crushers. A similar pair of filters is fitted on the national, featuring a 20-µm-thick uniform layer
injector-line to keep flows equal. Additional of particles immobilized on the inner wall,
protection for the column is provided by a 2.5 which separates both polar and non-polar
volatile compounds, including H2O.

10-17
FIG. 10-15. Chromatograms showing the signals from (A) the micro-thermal conductivity (TCD) and (B)
photoionization (PID) detectors of fluid-inclusion volatiles released by crushing 1.5 g of pegmatite quartz
from the Strange Lake peralkaline complex (sample 55-A-1; cf., Salvi & Williams-Jones 1997a and
1997b). The analysis was carried out using the PoraPLOT®-Q capillary column. The label N2+ refers to
the fact that this peak is an unresolved composite N2 ± H2 ± CO ± Ar ± O2 peak.
Output from the detectors takes the these spectra and analysis of the gas
form of analog signals in the range -18mV to composition.
+1V, which, in the system described here, are
converted to digital data by an HP® 35900C Sample analysis
dual-channel interface and transmitted to an Before analysis, the sample is frag-
HP® 3365 Chromatographic ChemStation® (a mented to pieces between 0.75 mm and 2 mm
PC Microsoft® Windows™ operated computer in diameter, hand-picked, and cleaned with
code). This code displays the chromatographic doubly distilled H2O (~8 MΩ-cm; ethanol is
spectra on a computer monitor and provides the used in the case of sulfide samples). Between
algorithms necessary for the manipulation of ~0.5 g and 2 g of material are loaded into the
crusher, and the piston is lowered to just touch

10-18
the grains. The crusher and all lines are heated,
and both sample and confining chambers are
pressurized with a low flux of He (≤1 ml/min)
for a duration of at least 12 hours (overnight).
Flushing removes all air and H2O that may be
absorbed onto the sample. An oil pump is
connected to the ram by an on/off valve. Just
prior to analysis, this valve is in the “off”
position and the pump is loaded to about 30 to
35 MPa. Releasing this pressure by switching
the valve “on” causes a sharp and sudden stroke
of the ram on the piston, which crushes the
sample. The gas chromatograph run is started
synchronously.
During analysis with the capillary
column described above, carrier gas flow is set FIG. 10-16. Examples of calibration curves for
to 20 ml/min at 35°C (because of its high volatile species created by analyzing several
capacity, the 25 m x 0.53 mm ID PoraPLOT® Q gas mixtures of known composition (after
column operates best at high flow rates, which Salvi & Williams-Jones 1997b).
are compatible with the injector set-up in
unsplit mode), and reference gas flow to the composite peak with the TCD detector at 2.7
TCD at 30 ml/min. A convenient temperature min. This peak is commonly assumed to
program, i.e., one that enables separation of the represent only N2. However, it is possible to
light gases with relatively short total run times resolve this peak by changing the carrier gas, as
(longer retention times result in inferior peak discussed below. Methane is detected at 2.9
definition) is set up as follows: The oven is min and CO2 at 5.6 min. The PID detects C2H4,
cooled to -20°C before the analysis is started, C2H2 and C2H6 at 9.9 min, 10.3 min, and 11.6
and 6.5 min into the run, after elution of CO2 on min, respectively. Water is separated at 14.5
the TCD, the oven temperature is ramped at min, on the 35°C plateau of the oven
10°C/min to 35°C where it is held for 10 min to temperature program and its elution typically
allow H2O elution on a flat background (the lasts about 2 to 4 minutes, depending on the
TCD is much more sensitive to temperature amount. After elution of H2O, data collection is
fluctuations than the PID, as illustrated by the stopped on the TCD, as species not yet eluted
rise in background coinciding with the onset of are detected on the PID. It is worth noting that
the heating program on the TCD plot on Fig. if a species were to elute at the same time as
10-15A). The temperature is then increased at water, the retention time of H2O could be easily
5°C/min to 115°C and left unchanged for ~40 changed, and the two peaks resolved by simply
min, when the run is stopped. The total running holding the oven temperature below zero for a
time is about 80 min. slightly longer/shorter time.
Identification of peaks and standard- At the beginning of the second
ization is achieved by injecting different temperature program, the C3 hydrocarbon
amounts of known gas mixtures (available species, i.e., propane (C3H8), propylene (C3H6),
commercially) into the GC, except for water, propadiene (C3H4) and propyne (C3H4), elute on
which can be injected as saturated vapor at a flat background starting at 24.5 min.
known temperatures and pressures. Identific- Isobutane (C4H10) elutes at 34.5 min, and at
ation of species is accomplished by comparing 36.3 min n-butane exits the column followed
retention times. Calibration curves (Fig. 10- immediately by the ethanol peak. Isopentane
16), constructed by plotting the integrated peak and n-pentane (C5H12) elute after the oven has
areas against the injected amounts of a species reached 115°C, at 44.5 min and 46.8 min,
(known), are used to determine the amounts of respectively (Fig. 10-15). In samples of
the eluted gases. Blanks are run by simulating hydrocarbon-bearing fluid inclusions from the
crush action with an empty sample chamber, hyperalkaline Strange Lake granite (Salvi &
i.e., by just moving the piston, and by crushing Williams-Jones 1997b) peaks were also
inclusion-free Brazilian quartz. observed at 66.8 min, 69.2 min and 75.0 min.
The first species to exit the column are The last peak corresponds to n-hexane (C6H14),
N2 ± H2 ± CO ± Ar ± O2, which form a and, although the other two are uncalibrated,
based on their boiling points, they are

10-19
tentatively assigned to isohexane and the first by normalizing it against a gas detected
neohexane, respectively. The poorer shape in both analyses with the TCD, for example,
(width/height ratios ≥10-3) of the hexane signals CH4. Carbon dioxide should not be used in the
is due to the late elution of these isomers and normalization as it does not respond linearly
can be improved by running the samples with amount when N2 is the carrier gas. The
isothermally at 115°C, which shortens retention composite peak of the first analysis is used to
times of these species to ~30 min. After elution determine the content of N2 after correction to
of the hexanes, the run is stopped, at about 80 remove the contributions of H2 and Ar; except
min, and the oven is returned to -20°C. for fluids trapped at near magmatic
Because of the nature of its construction, this temperatures, the contribution from CO can be
column reaches thermal equilibrium with the assumed to be negligible.
oven almost immediately, which is convenient, The composite peak discussed above
especially when making sub-ambient can also be separated into its constituent gases
temperature runs. by cooling the column down to –70°C and
Heavier aliphatic and some aromatic subsequently heating it in a series of
hydrocarbons (e.g., benzene, chlorobenzenes, programmed steps to ambient temperatures
toluene) can be separated on this column at (Thomas et al. 1988 and 1990, Bray & Spooner
115°C, but their retention times are extremely 1992).
long, producing very wide peaks (»5 min) Some researchers have tried analyzing
which drastically limits detection. This could H2 using molecular sieve packed columns in
be improved by fitting the PID with a 10.2 eV order to solve the problem of peak overlap with
lamp, which would allow an oven temperature N2, Ar, CO, O2 (e.g., Andrawes & Gibson
of up to 300°C, and would dramatically reduce 1979). However, although H2 is detected at low
the elution times of species with boiling points concentrations with such columns, most other
higher than ~100°C (e.g., octane b.p. = 125°C). gases cannot be detected at the required levels.
It would also improve detector sensitivity for Consequently, the amount of H2 cannot be
aromatics by a factor of ten (cf. Driscoll & related to those of the remaining fluid inclusion
Duffy 1987). However, as mentioned earlier, gases, which must be analyzed on another
the 10.2 eV lamp cannot detect the C2 alkanes. column from a separate extraction.
H2S and SO2 have similar retention times to
C2H4 and H2O, respectively, but have not been GC-MS
analyzed on the McGill system. This system As mentioned earlier and as has been
may, however, not be well-suited for analysis of discussed above, GC and MS are
these gases as they are reported to absorb on the complementary, the former providing the
metal surfaces of the crusher and transfer lines advantage of unambiguous species
(C. Bray, pers. comm., 2003). identification and applicability to all gases, and
the latter the advantage of extremely low
Analysis for H2 and Ar detection limits. In principle, an approach to
Hydrogen and argon can be analyzed fluid inclusion gas analysis, which would
successfully using the PoraPLOT® capillary combine the different strengths of the two
column described earlier if separate aliquots of techniques (and is used routinely in other
the same sample are analyzed and the carrier analyses) is sequential on-line analysis with GC
gas is changed from He to N2 (Salvi & and then MS. This approach would employ the
Williams-Jones 1997b). As noted above, the GC to identify the various gas species (where
first analysis will yield a composite peak for N2 concentrations are sufficient for detection) and
± H2 ± CO ± Ar ± O2, using a TCD. By separate them physically (by their different
contrast, the peak for N2 will be eliminated in retention times) for subsequent MS analysis.
the second analysis and there will be no peak The MS would thus receive a continuous stream
for CO because of the very high detection limit of gas directly from the GC comprising the
for this gas (~5 x 10-2 µmoles) with N2 as the carrier and a sequence of fluid inclusion gases
carrier. Furthermore, peaks for H2 and Ar will separated in time. By thus allowing the MS to
be distinguishable because the latter peak has analyze individual gas species rather than
inverted polarity (Fig. 10-17). Oxygen can be mixtures, on-line GC-MS would greatly reduce
ignored because of its extremely low partial the complexity of the spectra that must be
pressure, even under the most oxidizing resolved (ionization of even simple mixtures
conditions. The second analysis is related to can create large numbers of ion fragments with

10-20
FIG. 10-17. Chromatograms showing the response of the TCD to gases released by crushing two aliquots of
quartz from the same sample of the Strange Lake peralkaline complex using (A) helium and (B) nitrogen
as carrier gases (sample TTR-22; cf., Salvi & Williams-Jones 1997b).
overlapping peaks), and thus significantly analytical sensitivity that might be afforded by
improve the accuracy of the analysis. Where the MS. In order to minimize sample loss
concentrations are too low for GC detection, the during flow splitting, a number of GC/MS
approach would also ease the difficulty of MS interfaces have been designed that make use of
species identification. devices such as molecular sieves or semi-
The main obstacle to the application of permeable membranes. However, the majority
GC-MS analysis to fluid inclusion gases is that of these designs have intrinsic drawbacks that
the MS can only handle very low rates of gas limit the quality of the analyses.
flow. Consequently, the flow rate from the GC In principle, the problem of interfacing the GC
needs to be reduced before introduction of the with the MS could be solved by the use of
fluid inclusion gases into the MS. In the case of narrow bore capillary columns in the former, as
packed columns, approximately 99 % of the these provide for flow-rates low enough to
flow has to be diverted to waste leaving only permit direct entry of the gas into the MS.
1% to enter the MS. This results in a two However, on-line extraction of fluid inclusion
orders of magnitude increase in the detection gases produces a wide range of sample volumes
limit, which largely negates any advantages in and large dead volumes, necessitating the use of

10-21
packed columns or megabore capillary columns
for GC analysis (see discussion of McGill GC
system). Consequently, most researchers who
use GC-MS systems to analyze fluid inclusion
gases do their extraction off-line and inject the
collected gas (typically after cryogenic
separation) directly into the GC-MS (e.g.,
Norman et al. 1991, George et al. 1997, 2001).
This re-introduces the problem (solved by on-
line crushing) that only a very small proportion
of the released fluid inclusion gases can be
collected. An alternative approach, which has
been developed recently and retains on-line
extraction, is to run a loop off the extraction
line into a N2 trap (inside the GC or just before
it, e.g., Volk et al. 2000, Ruble et al. 1998)
designed to collect all the extracted volatiles
(Fig. 10-18). By subsequently warming the trap FIG. 10-18. Sketch of a closed system for
along a programmed temperature gradient in analysis of inclusion volatiles by GC-MS
the GC oven, the gases are released sequentially (adapted from Volk et al. 2000).
and analyzed (after splitting the flow; see
above) with a small bore capillary column an eluting compound that can be identified by
(Ruble et al. 1998, report the use of a 25 m x interpretation of the mass spectra recorded for
0.32 mm fused silica column). However, this the peak. Finally, as shown in the lower part of
method does not permit real-time analysis of Figure 10-19, the intensity at a single m/e ratio
the released inclusion gases. over the course of a chromatographic run can
The above notwithstanding, GC-MS be displayed to yield a selected ion current
remains a valuable tool, particularly for the profile or mass chromatogram. This allows
analysis of complex molecules, e.g., those in components of interest to be found in a
petroleum inclusions (cf. Ruble et al. 1998, complex mixture without having to examine
Volk et al. 2000, Guilhaumon et al. 2000, each mass spectrum individually.
George et al. 2001), and can also be very useful
in analyzing complex mixtures involving GEOLOGICAL APPLICATIONS
common inclusion volatiles such as CO2, H2S Analyses of bulk fluid inclusion gases
N2, H2, CO, SO2, He, Ar, CH4 and light represent a valuable source of data for
hydrocarbons (e.g., Norman et al. 1991), interpreting a variety of geological phenomena
especially when these species are in low and processes, and are playing an increasingly
concentrations. important role in both mineral and petroleum
The data obtained in GC-MS analyses exploration. In this section we briefly review
consist of a series of mass spectra that are examples of some of these applications.
acquired sequentially in time. To generate this Two of the questions which commonly
information, the mass spectrometer scans the challenge researchers investigating hydro-
selected mass range repeatedly during the thermal systems are the nature of the fluid
chromatographic run. If a scan is taken every source(s) and how the fluid evolved temporally
second and the run is 30 minutes long, 1800 and spatially. Moore et al. (2001) have recently
spectra are recorded. This information may be used MS analyses of fluid inclusion gases to
displayed in several ways, as shown in Figure address these questions for The Geysers, one of
10-19. Firstly, the intensities of all the ions in the largest producing geothermal systems in the
each spectrum are summed, and this sum is world. On the basis of N2/Ar, 3He/4He and
plotted as a function of chromatographic CO2/CH4 ratios, they showed that the fluids
retention time to give a total ion chromatogram come dominantly from magmatic and meteoric
(TIC), the appearance of which is similar to the sources, that the latter fluid is important mainly
output of a conventional chromatographic in the central and southeastern parts of the field,
detector. Secondly, as shown in the diagonal and that its influx coincided with the change
display in Figure 10-19, any of the spectra can from liquid- to vapor-dominated conditions.
be displayed. Each peak in the TIC represents They also showed that episodes of boiling could

10-22
FIG. 10-19. Example of a spectrum produced by a GC-MS analysis (adapted from Chiu & Muddiman
2001).
be readily identified by changes in the N2/Ar application in assessing the maturity of crude
ratio, variable incondensable gas to H2O ratios oils (e.g., George et al. 2001), determining their
and systematic variations in the proportions of evolution during migration (e.g., Jensenius &
CH4, CO2 and H2. In the case of ore-forming Burruss 1990) and reconstructing the history of
hydrothermal systems, analyses of the bulk reservoir filling (e.g., Karlsen et al. 1993).
gases have also provided valuable information Recently, several authors, notably Barclay et al.
with which to interpret depositional mechan- (2000) and Parnell et al. (2001), have suggested
isms. For example, Graney et al. (1991), in a that fluid inclusion gas geochemistry could be
study of inclusion gases from Carlin-type gold used as a tool for estimating petroleum
deposits in Nevada, showed that samples from reserves. One of the key pieces of information
mineralized areas had higher H2S/CO2 ratios required for the latter is the location of the oil-
and more consistent H2O/CO2 and H2O/N2 water contact, which normally is interpreted
ratios than samples from unmineralized areas, from geophysical well logs (e.g., electrical
and used these relationships to argue resistivity, gamma ray spectrometry). Barclay
convincingly that the gold was transported as a et al. (2000) have proposed that this
bisulphide complex and deposited as a result of information can also be provided by fluid
boiling. Contrasting H2S/CO2 ratios are also inclusion stratigraphy (FIS) and in some cases
characteristic of mineralized and unmineralized more reliably so than by geophysical well-logs.
samples from Mississippi-Valley-Type lead- This new approach involves MS analysis of the
zinc deposits, and have been used by Landis & bulk fluid inclusion gases in large numbers of
Hofstra (1991) to support a fluid mixing model drill cuttings or core samples and developing a
for ore deposition. More generally, by helping stratigraphy for specific compositional
to identify fluid sources, evaluate metal indicators. As the oil-water contact represents a
transport potential (estimate ligand concent- zone where the composition of petroleum is
ration, e.g., H2S), establish the redox state (e.g., modified by large differences in the solubility
using CH4/CO2 and H2S/SO2), estimate of various hydrocarbon components in water,
temperature and pH, and record gradients in changes in the ratios of key species, e.g., a
fluid chemistry, bulk inclusion gas sharp increase in the benzene/toluene ratio
geochemistry represents a valuable data source (toluene is 10 times more soluble in water than
for genetic modeling of ore deposition and a benzene), might signal close proximity to the
potentially powerful tool for metallic mineral oil/water contact. Similarly, FIS could play an
exploration (cf. Kesler et al. 1986, Graney et al. important role in optimizing oil recovery by
1991, Landis & Hofstra 1991, Norman et al. locating stratigraphic barriers to fluid flow, e.g.,
1991, and Wang et al. 1999). shale horizons from up-well reversals in
In petroleum geology, fluid inclusion petroleum density (e.g., sharp increases in the
gas geochemistry has found extensive proportion of longer chain alkanes).

10-23
As a final example of the application bulk analytical approach is that for the results to
of bulk fluid inclusion gas geochemistry, we be meaningful a single population must be
turn to an interesting feature of hyperalkaline sampled, something that is commonly difficult
intrusions, namely that they commonly contain to achieve because of multiple episodes of fluid
abundant carbonic fluid inclusions dominated entrapment. Modern methods that have been
by methane and other more complex developed for bulk inclusion analysis are based
hydrocarbons (Petersilie 1962, Sobolev et al. on gas chromatography and quadrupole mass
1970, Konnerup-Madsen et al. 1981). In the spectrometry, and generally involve on-line
absence of detailed information on the nature extraction of the gases by crushing or thermal
and concentration of the different decrepitation directly into the analytical device.
hydrocarbons, the preferred explanation for The advantage of gas chromatography is that
their presence had been that they were the species analyzed can be easily identified,
components of a C-O-H fluid, which exsolved however, the detection limits are relatively
from the magma and evolved by cooling along high. By contrast, mass spectrometry is
a relatively reduced (QFM) rock-buffered redox extremely sensitive but species identification is
path (e.g., Gerlach 1980, Konnerup-Madsen et more difficult. The two methods are thus
al. 1985, Kogarko et al. 1986, Nivin et al. complementary. In principle, on-line extraction
1995). However, using the GC system of fluid inclusions into a gas chromatograph,
described above, Salvi & Williams-Jones followed by on-line analysis of the eluted gases
(1997a) were able to determine the by mass spectrometry, could provide an
concentrations of alkanes up to C6 for approach to bulk fluid inclusion gas analysis
inclusions in the hyperalkaline Strange Lake applicable to samples approaching the size of
granite, and show that those for C2 to C6 alkanes single inclusions. If this goal were realized, the
are far too high to support this simple magmatic method would combine most of the advantages
model. On the other hand, the proportions of of single inclusion analysis with the superior
the alkanes relative to methane are remarkably detection capability of bulk inclusion analysis.
similar to those resulting from Fischer-Tropsch In practice, however, this combination of gas
synthesis, an industrial process used to produce chromatography and mass spectrometry
synthetic fuels by reacting CO or CO2 with introduces technical problems related to the rate
hydrogen in the presence of a Group VIII metal of injection of the gases into the mass
(e.g., Fe or Ni) or its oxide (cf. Anderson 1984). spectrometer, which are still not fully resolved.
This suggested a natural Fischer-Tropsch Thus, while bulk fluid inclusion gas analysis is
synthesis, which Salvi & Williams-Jones an important complement to single inclusion
(1997a) proposed occurred as a result of the gas analysis, it cannot replace the latter.
reaction of orthomagmatic CO and CO2 with H2 Analyses of the gas chemistry of bulk inclusion
generated during alteration of arvedsonite to fluids have played an important role in
aegirine. Magnetite, which was a product of resolving a variety of geological questions
the alteration, was inferred to be the catalyst. A ranging from the controls of metal
similar hypothesis has been proposed to explain concentration in the Earth’s crust to the origin
the presence of hydrocarbons in serpentinites of hydrocarbons in extraterrestrial bodies, and
(e.g., Berndt et al. 1996) and extraterrestrial there is every reason to expect that such
bodies (e.g., Zolotov & Shock 2000), and has analyses will find even greater application in
even been proposed as the mechanism by which the future.
life originated (e.g., McCollom et al. 1999).
ACKNOWLEDGMENTS
SUMMARY AND CONCLUSIONS We are grateful to J. Moore, M.
In this chapter, we have shown that Malcomson, L. Beci, D. Hall and M. Paddison
bulk analysis of fluid inclusion volatiles for their help in answering some of the
provides an attractive alternative to individual technical questions raised during the
inclusion analysis for cases where the preparation of this chapter. Constructive
inclusions are too small and/or the gases of reviews by C. Bray, S. Kesler and I. Samson
interest are in concentrations too low for the helped improve the quality of the manuscript.
latter analytical approach. In other cases, bulk Finally, we would like to again thank C. Bray
analysis can provide a valuable complementary for his advice and assistance during the
data set to that provided by the analysis of installation of the gas chromatographic system
single inclusions. The major limitation of the

10-24
in the fluid inclusion laboratory at McGill detectors and digital peak processing [abs.].
University. 2nd Biennial Pan-Am. Conf. Res. Fluid
Inclusions, 2, 12-13.
REFERENCES
BRAY, C.J. & SPOONER, E.T.C. (1992): Fluid
ABELL, P.I, DRAFFEN, C.H., EGLINTON, G., inclusion volatile analysis by gas
HAYES, J.M., MAXWELL, J.R. & PILLINGER, chromatography with photoionization/micro-
C.T. (1970): Organic analysis of the Apollo 11 thermal conductivity detectors: Applications
lunar sample. Proc. Apollo 11 Lunar Sci. to magmatic MoS2 and other H2O-CO2 and
Conf. 1757-1773. H2O-CH4 fluids. Geochim. Cosmochim. Acta,
56, 261-272.
AGAFONOV, L.V. & ANDREYEVA, G.A. (1972):
Gases in the alpine-type ultramafic of the BRAY, C.J., SPOONER, E.T.C. & THOMAS, A.V.
Anadyr-Koryak fold system. Doklady Akad. (1991): Fluid inclusion volatile analysis by
Nauk SSSR 210, 232-234. heated crushing, on-line gas chromatography;
applications to Archean fluids. J. Geochem.
ANDERSON, R.B. (1984): The Fischer-Tropsch
Exploration 42, 167-193.
Synthesis, Academic Press, New York.
BRECI, L. (2003): Introduction to Mass Spectro-
ANDRAWES, F.F. & GIBSON, E.K. JR. (1979):
metry: http://www.chem.arizona.edu/massspec
Release and analysis of gases from geological
samples. Am. Mineral., 64, 453-463. BRUNNEE, C. (1987): The Ideal Mass Analyzer:
Fact or Fiction? Int. J. Mass Spectrom. Ion
ANDRAWES, F.F., HOLZER, G., ROEDDER, E.,
Proc. 76, 125-237.
GIBSON, E.K.JR. & ORO, J. (1984): Gas
chromatographic analyses of volatiles in fluid BURRUSS, R.C. (1981): Hydrocarbon fluid
and gas inclusions. J. Chromatogr. 302, 181- inclusions in studies of sedimentary
193. diagenesis. In Fluid inclusions, applications to
petrology. (L.S. Hollister & M.L. Crawford,
BARCLAY, S. A., WORDEN, R. H., PARNELL, J.,
eds.). Mineralogical Association of Canada,
HALL, D. L. & STERNER, S. M. (2000):
Short Course Handbook, 6, 138-156.
Assessment of fluid contacts and compart-
mentalization in sandstone reservoirs using BURRUSS, R.C. (1987): Crushing-cell, capillary
fluid inclusions: An example from the Magnus column gas chromatography of petroleum
Oil Field, North Sea. AAPG Bulletin 84, 489- fluid inclusions: Methods and application to
504. petroleum source rocks, reservoirs, and low
temperature hydrothermal ores. In Fluid
BARKER, C.G. & ROBINSON, S.J. (1984):
Inclusion Research (E. Roedder & A.
Thermal release of water from natural quartz.
Kozolowski, eds.) 20, p. 59 (abstr.).
Am. Mineral. 69, 1078-1081.
BURRUSS, R.C. (2003a): Petroleum fluid
BARKER, C.G. & SMITH, M.P. (1986): Mass
inclusions: An introduction. In Fluid
spectrometic determinations of gases in
Inclusions: Analysis and Interpretation (I.
individual fluid inclusions in natural minerals.
Samson, A. Anderson & D. Marshall, eds.)
Anal. Chem. 58, 1330-1333.
Mineralogical Association of Canada, Short
BEHAR, F. & PINEAU, F. (1979): Analyse de Course 32, xx-yy.
CO2, H2O, hydrocarbures des inclusions
BURRUSS, R.C. (2003b): Raman spectroscopy of
fluides par chromatographie en phase gazeuze:
fluid inclusions. In Fluid Inclusions: Analysis
application aux fentes alpines et aux roches
and Interpretation (I. Samson, A. Anderson &
métamorphiques. Bull. Minéral. 102, 611-
D. Marshall, eds.) Mineralogical Association
621.
of Canada, Short Course 32, xx-yy.
BERNDT, M.E., ALLEN, D.E. & SEYFRIED, W.E.
CHANNER, D.M.DER. & SPOONER, E.T.C.
JR. (1996): Reduction of CO2 during
(1994): Combined gas and ion chromato-
serpentinization of olivine at 300°C and 500
graphic and analysis of fluid inclusions;
bar. Geology 24, 351-354.
applications to Archean granite pegmatite and
BRAY, C.J. & SPOONER, E.T.C. (1989): Fluid pegmatite and gold-quartz vein fluids.
inclusion volatile analysis using heated Geochim. Cosmochim. Acta 58, 1101-1118.
crushing (~105°C)/gas chromatography with
in-series photoionization/thermal conductivity

10-25
CHANNER, D.M.DER, BRAY, C.J. & SPOONER, composition of carbon gases in high-heat flow
E.T.C. (1999): Integrated cation-anion/volatile sedimentary basins. Geochim. Cosmochim.
fluid inclusion analysis by gas and ion Acta, 61, 3763-3785.
chromatography; methodology and examples.
GLEESON, S. (2003): Bulk analysis of
Chem. Geol. 154 59-82.
electrolytes in fluid inclusions. In Fluid
CHIU, C.M. & MUDDIMAN, D.C. (2001): What is Inclusions: Analysis and Interpretation (I.
Mass Spectrometry? American Society for Samson, A. Anderson & D. Marshall, eds.)
Mass Spectrometry. http://www.asms.org/ Mineralogical Association of Canada, Short
whatisms Course 32, xx-yy.
CUNEY, M., PAGEL, M. & TOURET, J. (1976): GOGUEL, R. (1963): Die chemische
L’analyse des gaz des inclusions fluides par Zusammensetzung der inden mineralen einiger
chromatographie en phase gazeuse. Bull. Coc. Granite und ihrer Pegmatite engeschlossenen
Fr. Minéral. Crystallogr. 99, 169-177. Gase und Flüssigkeiten. Geochim.
Cosmochim. Acta 27, 155-181.
DE HOFFMANN, E., CHARETTE, J. & STROOBANT
V. (1996): Mass Spectrometry: Principles and GOLDSTEIN, R.H. (2001): Fluid inclusions in
Applications, Wiley, 352 p. sedimentary and diagenetic systems. Lithos
55, 159-194.
DRISCOLL, J.N. & DUFFY, M. (1987):
Photoionization Detector: A versatile tool for GRANEY, J.R. & KESLER, S.K. (1995): Factors
environmental analysis. Chromatography 2, affecting gas analysis of inclusion fluid by
21-27. quadrupole mass spectrometry. Geochim.
Cosmochim. Acta 59, 3977–3986.
EUGSTER, H.P. & SKIPPEN, G.B. (1967): Igneous
and metamorphic reactions involving gas GRANEY, J.R., KESLER, S.E. & JONES, H.D.
equilibria. In Researches in Geochemistry, (1991): Application of gas analysis of
Volume 2 (P.H. Abelson, ed.), Wiley, 492- jasperoid inclusion fluids to exploration for
520. micron gold deposits. J. Geochem. Explorat.
42, 91-106.
GEORGE S.C., KRIEGER F. W., EADINGTON P.J.,
QUEZADA RA., GREENWOOD P.F., EISENBERG GRAUPNER, T., BRAY, C.J., SPOONER, E.T.C. &
L.I., HAMILTON P.J. & WILSON M.A. (1997): HERZIG, P.M. (2001): Analysis of fluid
Geochemical comparison of oil-bearing fluid inclusions in seafloor hydrothermal
inclusions and produced oil from the Toro precipitates; testing and application of an
sandstone, Papua New Guinea. Organic integrated GC/IC technique. Chem. Geol. 177,
Geochemistry 26, 155-173. 443-470.
GEORGE, S.C., RUBLE, T.E., DUTKIEWICZ, A., GUHA, J., LU, H-Z. & GAGNON, M. (1990): Gas
EADINGTON, P.J. (2001): Assessing the composition of fluid inclusions using solid
maturity of oil trapped in fluid inclusions probe mass spectrometry and its application to
using molecular geochemistry data and the study of mineralizing processes. Geochim.
visually-determined fluorescence colours. Cosmochim. Acta 54, 553-558.
Applied Geochemistry, 16, 451-473.
GUILHAUMOU, N., ELLOUZ, N., JASWAL, T.M. &
GERLACH, T.M. (1980): Evaluation of volcanic MOUGIN, P. (2000): Genesis and evolution of
gas analyses from Kilauea Volcano. J. hydrocarbons entrapped in the fluorite deposit
Volcanol. Geoth. Res., 7, 295-317. of Koh-i-Maran, (North Kirthar Range,
Pakistan). Marine and Petroleum Geology 17,
GERLACH, T.M. (1989): CO2 from magma-
1151-1164.
chamber degassing; discussion. Nature. 337;
6203, p. 124. HOLLOWAY, J.R. (1976): Fluids in the evolution
of granitic magmas: Consequences of finite
GIGGENBACH, W.F. (1993): Redox control of
CO2 solubility. Geol. Soc. Am. Bull. 87, 1513-
gas compositions in Philippine volcanic-
1518.
hydrothermal systems. Geothermics, 22, 575-
587. IKORSKI, S.V. (1981): Hydrocarbon gases in alk-
aline intrusions. Geochem. Int. 28(6), 17-23.
GIGGENBACH, W.F. (1997): Relative importance
of thermodynamic and kinetic processes in IKORSKI, S.V. & VOLOSHIN, A.V. (1982):
governing the chemical and isotopic Typical gas-phase compositions in inclusions

10-26
in quartz from Kola Peninsula granite KONNERUP-MADSEN J., ROSE-HANSEN J. &
pegmatites. I. Geochem. Int. 19(5), 173-179. LARSEN, E. (1981): Hydrocarbon gases
associated with alkaline igneous activity:
IKORSKI, S.V., NIVIN, V.A. & PRIPACHKIN, V.A.
Evidence from compositions of fluid
(1993): New information about natural gases
inclusions. Rapp. Grønlands geol. Unders.
of alkaline massifs. Isotopenpraxis 28, 289-
103, 99-108.
291.
KREULEN, R. & SCHUILING, R.D. (1982): N2-
ISHIBASHI, J-I., SATO, M., SANO, Y., WAKITA, H.;
CH4-CO2 fluids during formation of the Dôme
GAMO, T. & SHANKS, W.C. (2002): Helium
de l’Agout, France. Geochim. Cosmochim.
and carbon gas geochemistry of pore fluids
Acta 46, 193-203.
from the sediment-rich hydrothermal system in
Escanaba Trough. Applied Geochemistry 17, LANDIS, G.P. & HOFSTRA, A.H (1991): Fluid
1457-1466. inclusion gas chemistry as a potential minerals
exploration tool; case studies from Creede,
JENNINGS, W. (1987): Analytical Gas Chrom-
CO, Jerritt Canyon, NV, Coeur d'Alene
atography. Academic Press, Orlando, Florida,
District, ID and MT, southern Alaska
259 p.
mesothermal veins, and mid-continent MVT's.
JENSENIUS J. & BURRUSS R.C. (1990): Hydro- J. Geochem. Explor. 42, 25-59.
carbon-water interactions during brine
LANDIS, G.P. & RYE, R.O. (1989):
migration: Evidence from hydrocarbon
Reconnaissance gas chemistry of the Creede,
inclusions in calcite cements from Danish
Colorado, hydrothermal system. US Geol.
North Seas oil fields. Geochim. Cosmochim.
Survey Open-file Report 89-84.
Acta 54, 705-713.
MARCH, R.E., HUGHES, R.J. & TODD, J.F. (1989):
JOHNSTONE, R.A.W. & ROSE, M.E. (1996):
Quadrupole Storage Mass Spectrometry,
Mass Spectrometry for Chemists and
Chemical Analysis Series, 102, Wiley.
Biochemists, 2nd edition. Cambridge Univer-
sity Press, 2nd edition, 521 p. MCCOLLOM, T.M., RITTER, G. & SIMONEIT,
B.R.T. (1999): Lipid synthesis under
KALISZAN, R., (1987): Quantitative Structure-
hydrothermal conditions by Fischer-Tropsch-
Chromatographic Retention Relationships.
type reactions. Origins of Life and Evolution
Chemical Analysis Series 93 (J.D.
of the Biosphere 29, 153-166.
Winefordner & I.M. Kolthoff, eds.) Wiley-
Interscience, New York, N.Y., 303 p. MIRONOVA. O.F., NAUMOV, V.B. & SALAZKIN,
A.N. (1972): Chromatographic analysis of
KARLSEN, D.A., NEDKVITNE, T., LARTER, S.R. &
gases trapped in rocks (in Russian).
BJORLYKKE, K. (1993): Hydrocarbon
Geokhimiya 10, 1514-1521.
composition of authigenic inclusions;
application to elucidation of petroleum MOINE, B., GUILLOT, C. & GIBERT., F. (1994):
reservoir filling history. Geochim. Controls of the composition of nitrogen-rich
Cosmochim. Acta 57, 3641-3659. fluids originating from reaction with graphite
and ammonium-bearing biotite. Geochim.
KESLER, S.E., HAYNES, P.S., CREECH, M.Z. &
Cosmochim. Acta 58, 5503-5523.
GORMAN, J.A. (1986): Application of fluid
inclusion and rock-gas analysis in mineral MOORE, J.N., NORMAN, D.I. & KENNEDY B.M.
exploration. J. Geochem. Explor. 25, 201-215. (2001): Fluid inclusion gas compositions from
an active magmatic-hydrothermal system: A
KOGARKO, L.N., KOSZTOLANYI, C. &
case study of The Geysers geothermal field,
RYABCHIKOV, I.D. (1986): Geochemistry of
USA, Chem. Geol. 173, 3-30.
the reduced fluid in alkali magmas.
Geokhimiya 12, 1688-1986. Translated in MUNZ, I.A. (2001): Petroleum inclusions in
Geochem. Int. 24, 1987, 20-27. sedimentary basins; systematics, analytical
methods and applications. Lithos 55, 195-212.
KONNERUP-MADSEN J., DUBESSY, J. & ROSE-
HANSEN J. (1985): Combined Raman micro- NEWMAN, B.D., NORMAN, D.I., GUNDIMEDA, M.
probe spectrometry and microthermometry of & LEVY, S.S. (1996): Understanding the
fluid inclusions in minerals from igneous rocks genesis of nonmarine calcite deposits through
of the Gardar Province (South Greenland). quadrupole mass spectrometric analysis of
Lithos 18, 271-280.

10-27
fluid inclusion gases. Chem. Geol. 132, 205- Hydrocarbons in Crystalline Rocks (N. Petford
213. & K.J.W. McCaffrey, eds.) Geol. Soc. Lon.
Spec. Publ. 214, 151-173.
NIVIN, V.A., DEVIRTS, A.L. & LAGUTINA YE.P.
(1995): The origin of the gas phase in the POTTER, J., RANKIN, A.H., TRELOAR, P.J., NIVIN,
Lovozero massif based on hydrogen-isotope V.A., TING, W. & NI, P. (1998): A preliminary
data. Geochem. Internat. 32(8), 65-71. study of methane inclusions in alkaline
igneous rocks of the Kola igneous province,
NORMAN D.I. & MUSGRAVE, J. (1994): N2, Ar,
Russia: implications for the origin of methane
and He in fluid inclusions: Tracers of
in igneous rocks. Eur. J. Mineral. 10, 1167-
hydrothermal fluids. Geochem. Cosmochim.
1180.
Acta. 58, 119-1131.
ROEDDER, E. (1972): Composition of fluid
NORMAN, D.I. & SAWKINS, F.J. (1987): Analysis
inclusions. U.S. Geol. Surv. Prof. Pap. 440 JJ,
of volatiles in fluid inclusions by mass
164 p.
spectrometry. Chem. Geol. 61, 1-10.
ROEDDER, E. (1984): Fluid inclusions. Reviews
NORMAN, D.I., HARRISON, R.W., BEHR-ANDRES,
in Mineralogy 12, Mineralogical Society of
C. (1991): Geology and geochemical analysis
America, Washington.
of mineralizing fluids at the St. Cloud and U.S.
Treasury mines, Chloride mining district, New ROEDDER, E. & BODNAR, R.J. (1997): Fluid
Mexico. J. Geochem. Explor. 42, 61-89. inclusion studies of hydrothermal ore deposits.
In Geochemistry of Hydrothermal Ore
NORMAND, C. (2001): Experimental and Field
Deposits (H.L. Barnes ed.), Wiley, 657-697.
Investigations of Serpentinization and Roding-
itization, Ph. D. Thesis, McGill University, ROEDDER, E. & COOMBS, D.S., (1967):
Montréal, Canada. Immiscibility in granitic melts, indicated by
fluid inclusions in ejected granitic blocks from
PARNELL, J., MIDDLETON, D., HONGHAN, C. &
Ascension Island. J. Petrol. 8, 417-451.
HALL, D. (2001): The use of integrated fluid
inclusion studies in constraining oil charge RUBLE, T.E., GEORGE, S.C., LISK, M. &
history and reservoir compartmentation: QUEZADA, R.A. (1998): Organic compounds
examples from the Jeanne d'Arc Basin, trapped in aqueous fluid inclusions. Org.
offshore Newfoundland. Marine and Geochem. 29, 195-205.
Petroleum Geology 18, 535-549.
RUFF, R.K. (1993): Gas analysis of fluid
PERETTI, A. DUBESSY, J, MULLIS, J, FROST, B.R. inclusions; application toward precious metal
& TROMMSDORFF, V. (1992): Highly reduced exploration, Steeple Rock mining district,
conditions during Alpine metamorphism of the Grant County, New Mexico. M.Sc. Thesis,
Malenco peridotite (Sondrio, northern Italy) New Mexico Institute of Mining and
indicated by mineral paragenesis and H2 in Technology, Socorro, NM, United States.
fluid inclusions. Contrib. Mineral. Petrol.
SALVI, S. (1995): Magma Degassing and Wall
112, 329-340.
Rock Alteration in the Rare Metal-Rich
PETERSILIE, I.A. & SØRENSEN, H. (1970): Peralkaline Granite at Strange Lake,
Hydrocarbon gases and bituminous substances Quebec/Labrador, Ph. D. Thesis, McGill Uni-
in rocks from the Ilímaussaq alkaline versity, Montréal, Canada
intrusion, south Greenland. Lithos 3, 59-76.
SALVI, S. & WILLIAMS-JONES, A.E. (1992):
PETERSILIE, I.A. (1962): Origin of hydrocarbon Reduced orthomagmatic C-O-H-NaCl fluids in
gases and dispersed bitumens of the Khibina the Strange Lake rare-metal granitic complex,
alkalic massif. Geokhimiya 1, 15-29. Quebec/Labrador, Canada. Eur. J. Mineral. 4,
Translated in Geochem. Int. 1, 1962, 14-30. 1155-1174.
PIPEROV, N.B. & PENCHEV, N.P. (1973): A SALVI, S. & WILLIAMS-JONES, A.E. (1997a):
study on gas inclusions in minerals. Analysis Fischer-Tropsch synthesis of hydrocarbons
of the gases from micro-inclusions in allanite. during sub-solidus alteration of the Strange
Geochim. Cosmochim. Acta 37, 2075-2097. Lake peralkaline granite, Quebec/Labrador,
Canada. Geochim. Cosmochim. Acta 61, 83-
POTTER, J. & KONNERUP-MADSEN J. (2003) A
99.
review of the occurrence of abiogenic
hydrocarbons in igneous rocks. In

10-28
SALVI, S. & WILLIAMS-JONES, A.E. (1997b): pegmatites using solid-liquid-vapor inclusions
Fluid-inclusion volatile analysis by gas from the Tanco Pegmatite, S.E. Manitoba,
chromatography: Application of a wide-bore Canada. Trans. Royal Soc. Edinburgh: Earth
porous-polymer capillary column to the Sciences 79, 299-315.
separation of organic and inorganic
THOMAS, A.V., PASTERIS, J.C., BRAY, C.J. &
compounds. Can. Mineral. 35, 1391-1414.
SPOONER, E.T.C. (1990): H2O-CH4-NaCl-
SAMSON, I.M. & WILLIAMS-JONES, A.E. (1991): CO2 inclusions from the footwall contact of
C-O-H-N-salt fluids associated with contact the Tanco granitic pegmatite: Estimates of
metamorphism, McGerrigle Mountains, internal pressure and composition from
Quebec: A Raman spectroscopic study: microthermometry, laser Raman spectro-
Geochim. Cosmochim. Acta. 55, 169-177. scopy and gas chromatography. Geochim.
Cosmochim. Acta 54, 559-573.
SASADA, M., SAWAKI, T. & TAKENO, N. (1992):
Analysis of fluid inclusion gases from VIDAVSKY, I. (2003): MS ionization methods
geothermal systems, using a rapid-scanning and instrumentation tutorial. http://wunmr.
quadrupole mass spectrometer. Eur. J. wustl.edu/~msf/ionmethd.html
Mineral. 4, 895–906.
VOLK, H., MANN, U., BURDE, O., HORSFIELD, B.
SHEPHERD, T.J., RANKIN, A.H. & ALDERTON, & SUCHY, V. (2000): Petroleum inclusions
D.H.M., (1985): A Practical Guide to Fluid and residual oils; constraints for deciphering
Inclusion Studies, Blackie, London, 239 p. petroleum migration. J. Geochem. Expl. 69-
70, 595-599.
SHINOHARA, H. (1992): Does acid volcanic gas
represent magmatic discharge at depth? Rept. WANG, Y., SASAKI, M., SASADA, M. &, CHEN,
Geol. Surv. Japan 279, 152-155. C.-H. (1999): Fluid inclusion studies of the
Chinkuashih high-sulfidation gold-copper
SIMONEIT, B.R.T. (2002) (ed): Geochemistry of
deposits in Taiwan. Chem. Geol. 154, 155-
sediments drilled in the hydrothermal systems
167.
of the Northeastern Pacific Ocean. Applied
Geochemistry, Special Issue 17, 1379-1502 WELHAN, J.A. (1988): Methane and hydrogen in
mid-ocean-ridge basalt glasses: analysis by
SLOAN, R.C. (1992): Development and
vacuum crushing. Can. J. Earth Sci. 25, 38-
application of a mass spectrometric system to
48.
study volatile components of fluid inclusions.
M.Sc. Thesis, University of California, WILLIAMS, A.E. (1996): Mass spectrometric
Riverside, CA, United States. analysis of volatiles in fluid inclusions; aliquot
calibration valve to simulate inclusion rupture.
SMITH, T.J., CLOKE, P.L. & KESLER, S.E. (1984):
Chem. Geol. 131, 155-165.
Geochemistry of fluid inclusions from the
McIntyre-Hollinger gold deposit, Timmins, WILLIAMS, A.E. & TAYLOR, M.C. (1996): Mass
Ontario, Canada. Econ. Geol. 79, 1265-1285. spectrometric identification of boric acid in
fluid inclusions in pegmatite minerals.
SOBOLEV, V.S., BAZAROVA, T.Y., SHUGUROVA,
Geochim. Cosmochim. Acta 60, 3435-3443.
N.A., BAZAROV, L.SH., DOLGOV, YU.A. &
SØRENSEN, H. (1970): A preliminary examin- YPMA, P. (1974): Analysis of fluid inclusions by
ation of fluid inclusions in nepheline, gas chromatography. NATO advanced study,
sorensenite, tugtupite and chkalovite from the Nancy, N.P. Report.
Ilímaussaq alkaline intrusion, South
ZOLOTOV, M.Y. & SHOCK, E.L. (2000): An
Greenland. Medd. Grønl. 181, 1-32.
abiotic origin for hydrocarbons in the Allan
SOMMER, M.A. & GIBSON, E.K. JR. (1986): Hill 84001 Martian meteorite through cooling
Volatile determinations of individual fluid of magmatic and impact-generated gases.
inclusions within the 3.4 b.y. North Pole Meteoritics & Planetary Sciences 35, 629-638.
barites from the Warrawoona group,
Northwestern Australia [abs.]. Lunar Planet.
Sci. 17, 815-816.
THOMAS, A.V., BRAY, C.J. & SPOONER, E.T.C.
(1988): A discussion of the Jahns-Burnham
proposal for the formation of zoned granitic

10-29
10-30

View publication stats

You might also like