BS 01747-9-1987 (2010) Iso 7996-1985
BS 01747-9-1987 (2010) Iso 7996-1985
BS 01747-9-1987 (2010) Iso 7996-1985
1987
ISO 7996:1985
Methods for
Measurement of air
pollution —
Part 9: Determination of the mass
concentration of nitrogen oxides in
ambient air: chemiluminescence
method
[ISO title: Ambient air — Determination of the mass
concentration of nitrogen oxides — Chemiluminescence method]
Confirmed
November 2010
UDC [551.510.42 + 614.72]:543.272.32:535:379
BS 1747-9:1987
British Coal
British Gas
British Steel Corporation
Cement Makers’ Federation
Department of Health and Social Security
Department of Trade and Industry (Laboratory of the Government Chemist)
Department of Trade and Industry (Warren Spring Laboratory)
Electricity Supply Industry in England and Wales
GAMBICA (BEAMA Ltd.)
Health and Safety Executive
Institute of Petroleum
Institution of Environmental Health Officers
Institution of Gas Engineers
Lead Development Association
National Industrial Fuel Efficiency Service
National Society for Clean Air
Society of Chemical Industry
The following bodies were also represented in the drafting of the standard,
through subcommittees and panels:
Contents
Page
Committees responsible Inside front cover
National foreword ii
1 Scope and field of application 1
2 References 1
3 Principle 1
4 Reactions 1
5 Apparatus 1
6 Procedure 2
7 Expression of results 5
8 Test report 5
Annex Typical performance characteristics 10
Figure 1a) — Schematic representation of a dual type
chemiluminescence analyser (two reaction chambers
and photomultipliers) 6
Figure 1b) — Schematic representation of a dual type
chemiluminescence analyser utilizing a double
reaction chamber and one photomultiplier 6
Figure 1c) — Schematic representation of a cyclic type
chemiluminescence analyser 7
Figure 2 — Schematic representation of arrangement
for checking analyser converter efficiency 8
Figure 3 — Schematic representation of a typical
calibration system using a nitrogen dioxide
permeation cell 9
Publications referred to Inside back cover
© BSI 11-1999 i
BS 1747-9:1987
National foreword
This Part of BS 1747, which has been prepared under the direction of the
Environment and Pollution Standards Committee, is identical with
ISO 7996:1985 “Ambient air — Determination of the mass concentration of
nitrogen oxides — chemiluminescence method” published by the International
Organization for Standardization. It was prepared by Technical Committee 146,
Air quality, of the International Organization for Standardization (ISO) with the
active participation and approval of the UK.
This Part of BS 1747 is one of a series relating to air quality arising from UK
participation in the work of ISO/TC 146. Methods for the determination of
particular constituents of ambient air will be published as further Parts of this
standard. Topics relating to air quality characteristics will be published as Parts
of BS 6069 “Methods for characterization of air quality”.
Terminology and conventions. The text of the International Standard has
been approved as suitable for publication as a British Standard without
deviation. Some terminology and certain conventions are not identical with those
used in British Standards; attention is drawn especially to the following.
The comma has been used as a decimal marker. It is current practice in British
Standards to use a full point on the baseline as the decimal marker.
Wherever the words “International Standard” appear, referring to this standard,
they should be read as “Part of BS 1747”.
The symbol “l” has been used to denote litre (and in its submultiples). In British
Standards it is current practice to use the symbol “L”. It is also current practice
to use the spelling “sulphur”, instead of “sulfur”.
Cross-references
International Standards Corresponding British Standards
BS 4559: Methods for preparation of calibration gas
mixtures
ISO 6142:1981 Section 1.1:1983 Mixtures containing components fully
vaporizable under ambient conditions
(Identical)
ISO 6144:1981 Part 3:1983 Static volumetric methods
(Identical)
ISO 6349:1979 Part 7:1981 Permeation method
(Identical)
The Technical Committee has reviewed the provisions of ISO 6711 and ISO 6879,
to which reference is made in the main body of the standard and in the Annex
respectively, and has decided that they are acceptable for use in conjunction with
this standard.
Textual errors. When adopting the text of the International Standard, the
textual errors given below were discovered. They have been marked in the text
with an asterisk and have been reported to ISO in a proposal to amend the text
of the International Standard.
In 6.2.2.3, line 3, “± 1 °C” should be “± 0,1 °C”.
In 6.3.1, equation (3), delete “+ ̽ (NO)” and the explanation of this term
thereunder.
In 6.3.2, lines 7 and 8, delete “and ̽ (NO) by ̽ (NO2) or ̽ (NOx)”.
In 6.3.2, note 3, the last sentence should be part of item b).
In Figure 2, the cylinder containing the NO source, should show “. 1” before
“ppm (V/V)”.
ii © BSI 11-1999
BS 1747-9:1987
Additional information. For clarification, the term “zero air”, which is used
in 6.2.1 and subsequent clauses, refers to air containing no nitrogen oxides.
In order to provide clarification notice is brought to users of this standard that the
equations given in 6.3.1 assume a converter efficiency of 100 % for nitrogen
dioxide conversion to nitrogen oxide. For very accurate determinations, the
appropriate conversion factor may have to be used and in such cases the attention
of users is also drawn to note 3 in 6.3.2, which describes alternative calibration
systems.
A British Standard does not purport to include all the necessary provisions of a
contract. Users of British Standards are responsible for their correct application.
Summary of pages
This document comprises a front cover, an inside front cover, pages i to iv,
pages 1 to 10, an inside back cover and a back cover.
This standard has been updated (see copyright date) and may have had
amendments incorporated. This will be indicated in the amendment table on the
inside front cover.
1 Scope and field of application Analysers may be of the dual or cyclic type. In the
dual type, the air flow is divided into two streams,
This International Standard specifies a
one passing directly to the reaction chamber and the
chemiluminescence method for the determination of
other via the converter. There are two reaction
the mass concentration of nitrogen oxides present in
chambers and one or two detectors respectively
ambient air. monitoring the nitrogen monoxide and nitrogen
The method is applicable to the determination of the oxides content of the air sample. In the cyclic type
mass concentration of nitrogen monoxide present in analyser, there is only one reaction chamber and
ambient air up to approximately 12,5 mg/m3 and of detector and the output alternates between nitrogen
nitrogen dioxide up to approximately 19 mg/m3 monoxide and total nitrogen oxides measurements,
respectively1) at 25 °C and 101,3 kPa. i.e. the air sample alternately bypasses or passes
For interferents, see 6.6. through the converter.
2 References 4 Reactions
ISO 6142, Gas analysis — Preparation of Luminescence is a characteristic of many
calibration gas mixtures — Weighing methods. substances from which light is emitted when they
ISO 6144, Gas analysis — Preparation of are excited. This phenomenon is called
calibration gas mixtures — Static volumetric chemiluminescence when produced as a result of a
methods. chemical reaction.
ISO 6349, Gas analysis — Preparation of This chemiluminescence method is based on the
calibration gas mixtures — Permeation method. reaction
ISO 6711, Gas analysis — Checking of calibration
gas mixtures by a comparison method.
3 Principle
Excited nitrogen dioxide emits radiation centred
Passage of the air sample through a filter around the near infra-red (1 200 nm).
(to prevent mal-operation of the chemiluminescence
analyser) at a constant flow rate into the reaction 5 Apparatus
chamber of the analyser, where it is mixed with an
excess of ozone. Filtering of the emitted radiation, Schematic diagrams of typical analysers are given
proportional to the amount of nitrogen monoxide in Figure 1a) and Figure 1b) (dual type) and
present in the air sample, by means of a selective Figure 1c) (cyclic type). An apparatus consists of the
optical filter and conversion of the filtered radiation following principal components (5.1 to 5.10).
into an electric signal by a photomultiplier tube. 5.1 Sample line
Measurement of nitrogen dioxide in the air sample The sample intake shall consist of a small inverted
by reducing it to nitrogen monoxide by passage funnel connected to the sample line — the latter
through a converter before it enters the reaction being as short as practically possible. Both shall be
chamber. made of material which is inert to nitrogen
The electric signal obtained is proportional to the monoxide and nitrogen dioxide. It is very important
total amount of nitrogen oxides, and the nitrogen not to use any materials made from copper based
dioxide contribution is calculated from the alloys. From experience, the best materials to use
difference between this value and that obtained for are polytetrafluoroethylene (PTFE) or
nitrogen monoxide only — when the air sample does perfluoro (ethylene/propylene) (FEP). In order to
not pass through the converter. avoid disturbance of the nitrogen
monoxide-nitrogen dioxide-ozone equilibrium
caused by reduced daylight intensity in the sample
line, the residence time in the sample line should be
as short as possible. The sample line may be
moderately heated to avoid any condensation.
1)
At the temperature and pressure given, the following conversion factors apply.
NO: 1 mg/m3 ç 0,81 ppm (V/V)
1 ppm (V/V) ç 1,23 mg/m3
NO2: 1 mg/m3 ç 0,53 ppm (V/V)
1 ppm (V/V) ç 1,88 mg/m3
© BSI 11-1999 1
BS 1747-9:1987
2 © BSI 11-1999
BS 1747-9:1987
Before calibrating the analyser, it is essential to If this efficiency falls below 95 %, replace or
ensure that the converter efficiency lies as close to regenerate the converter.
unity as possible. The determination of the 6.2 Calibration of the chemiluminescence
efficiency is based on the principle that the response analyser
of the apparatus to the total amount of nitrogen
oxides (NOx) does not change if analyses are made of 6.2.1 Principle
varying mixtures of nitrogen monoxide and nitrogen Generation of calibration gas mixtures containing
dioxide, but for which the total concentration of accurately known concentrations of nitrogen dioxide
nitrogen oxides is constant. by means of a permeation system (see ISO 6349).
A schematic diagram of a suitable arrangement is Dilution of the nitrogen dioxide emitted with zero
given in Figure 2 (see also 5.11). air of a relative humidity of about 50 % to obtain
nitrogen dioxide concentrations suitable for the
Ensure that the total flow rate of nitrogen monoxide
calibration of the nitrogen dioxide and total nitrogen
and oxygen (or air) is greater than the flow rate of
oxides channels. Calibration of the nitrogen
gas through the analyser — the excess being led
monoxide channel by using the same system in
away as shown in Figure 2.
conjunction with a converter, the efficiency of which
In each of the following steps, note both the nitrogen has been determined as in 6.1.
monoxide and total nitrogen oxides concentration
responses of the analyser. This then checks the level 6.2.2 Apparatus
of nitrogen dioxide being produced which should be A typical system for generating the required
in the range 10 to 90 % of the total nitrogen oxides concentrations of nitrogen monoxide and nitrogen
level. dioxide is shown in Figure 3.
a) With the ultra-violet lamp extinguished, note 6.2.2.1 Air flow rate controllers and flow meters,
the total nitrogen oxides concentration response, capable of maintaining and measuring flow rates
R1, and the nitrogen monoxide concentration to ± 1 %.
response, P1. 6.2.2.2 Drier, to remove moisture from the air
b) Switch on the ultra-violet lamp. Ozone is passing over the permeation cell. Satisfactory driers
formed which reacts with the nitrogen monoxide are permeation driers or suitable large tubes of
to produce nitrogen dioxide before the gases enter desiccants containing silica gel or similar.
the analyser. Note the total nitrogen oxides 6.2.2.3 Constant temperature chamber, capable of
concentration response, R2, and the nitrogen housing the nitrogen dioxide permeation cell and
monoxide concentration response, P2. maintaining its temperature to within ± 1 °C2). It
c) Vary the output of the ultra-violet lamp, and shall incorporate a temperature measuring device
note the total nitrogen oxides concentration having an accuracy of ± 0,05 °C.
responses (R3, R4, R5, etc.) and the nitrogen 6.2.2.4 Valves. Two valves — one to direct the
monoxide concentration responses (P3, P4, nitrogen monoxide (or nitrogen dioxide) flow into
P5, etc.). The ratios R2/R1, R3/R1, R4/R1, etc. the diluent zero air flow and one to allow the
should be as close to unity as possible (within the nitrogen dioxide to flow through or bypass the
range of repeatability of the analyser), i.e. the converter.
total nitrogen oxides concentration response
6.2.2.5 Output manifold, of an inert material and of
should be constant in each instance and
sufficient diameter to ensure an insignificant
independent of the ratio of the concentrations of
pressure drop at the analyser connection. The
nitrogen dioxide to nitrogen monoxide.
system shall have a vent designed to ensure
d) Determine the actual efficiency, expressed as a sufficient pressure within the manifold and so
percentage, of the converter from the formula prevent ambient air from entering it.
2)
See national foreword for details of textual error.
© BSI 11-1999 3
BS 1747-9:1987
4 © BSI 11-1999
BS 1747-9:1987
© BSI 11-1999 5
BS 1747-9:1987
6 © BSI 11-1999
BS 1747-9:1987
© BSI 11-1999 7
BS 1747-9:1987
8 © BSI 11-1999
© BSI 11-1999
Figure 3 — Schematic representation of a typical calibration system using a nitrogen dioxide permeation cell
BS 1747-9:1987
9
BS 1747-9:1987
10 © BSI 11-1999
BS 1747-9:1987
Publications referred to
© BSI 11-1999
BS 1747-9:
1987
ISO 7996:1985 BSI — British Standards Institution
BSI is the independent national body responsible for preparing
British Standards. It presents the UK view on standards in Europe and at the
international level. It is incorporated by Royal Charter.
Revisions
It is the constant aim of BSI to improve the quality of our products and services.
We would be grateful if anyone finding an inaccuracy or ambiguity while using
this British Standard would inform the Secretary of the technical committee
responsible, the identity of which can be found on the inside front cover.
Tel: 020 8996 9000. Fax: 020 8996 7400.
BSI offers members an individual updating service called PLUS which ensures
that subscribers automatically receive the latest editions of standards.
Buying standards
Orders for all BSI, international and foreign standards publications should be
addressed to Customer Services. Tel: 020 8996 9001. Fax: 020 8996 7001.
Information on standards
Copyright
Copyright subsists in all BSI publications. BSI also holds the copyright, in the
UK, of the publications of the international standardization bodies. Except as
permitted under the Copyright, Designs and Patents Act 1988 no extract may be
reproduced, stored in a retrieval system or transmitted in any form or by any
means – electronic, photocopying, recording or otherwise – without prior written