Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Rendered PDF File Page 1 of 5 DOC2099932, Rev:2

Electronic Signature Information

Name DOC2099932

Revision 2

Type Controlled Document

Title FMI 32058 Power Board Fuse Replacement Instructions

Originator 305008184_prashanth__gururaj

Release Date 03/15/2018 05:56:33 AM

Obsolete Date

File Name File Description File Size (Bytes)


2108779-001_FMI32058_Power Board Fuse FMI 32058 Power Board Fuse Replacement
301938
Instructions.pdf Instructions

Route Signer Function Status Comments Completion Date


R-9030467 305008184_prashanth__gururaj MIC IB LSI Approve Approved 15 Mar 2018 04:23:44 GMT

Periodic Review
There are no signatures or routes related to this business object.

Obsolesence Approval
There are no signatures or routes related to this business object.

* Printed versions are For Reference Only *


+ Indicates a task was reassigned from an original assignee
Released
Rendered PDF File Page 2 of 5 DOC2099932, Rev:2

Giraffe™ and Panda™ Warmers


FMI 32058 Power Board Fuse
Replacement Instructions

2108779-001 Rev A

© 2018 General Electric Company


All rights reserved.
Released
Rendered PDF File Page 3 of 5 DOC2099932, Rev:2

Purpose

GE Healthcare has recently become aware that the Power Board in


Giraffe and Panda Warmer is requiring early/unscheduled replacement
of Fuses (F7 & F8) in some machines.

This document provides all instructions needed to complete FMI 32058

Scope

This instruction and parts referenced are applicable only to unit’s that
are part of this FMI.

Reference Documents

• Panda™ and Giraffe™ Warmers Service Manual (M1128921)

Required Tools

• Standard Hand Tools (see 5.1 Common Hand Tools in the


Service Manual - M1128921)

Fuse FMI 32058 Kit (2108726-001) Contents

• SMD FUSE 2A, SOCKETED, LEAD-FREE NANO2 VERY FAST-


ACTING SUBMINIATURE, 2104203-001
• FMI 32058 POWER BOARD INSTALLATION INSTRUCTIONS
2108779-001
• FMI 32058 POWER BOARD FMI NOTICE/MEMO 2108780-001

FMI KIT Procedure

SENSITIVE TO ELECTROSTATIC DISCHARGE CAUTION


An Electrostatic Discharge (ESD) Susceptibility symbol is
displayed to alert service personnel that the part(s) are
sensitive to electrostatic discharge and that static control
procedures must be used to prevent damage to the
equipment.
Released

2 Field Installation Instruction 2108779-001 Rev A


Rendered PDF File Page 4 of 5 DOC2099932, Rev:2

1. Unplug unit from the wall before working on any part of the
unit.

Power Board Removal (Figure 1)

2. For wall mount models, remove the heater head from the wall
bracket.
3. Loosen the 2 captive screws at the top of the enclosure’s rear
panel, then tilt the panel down and lift the 2 tabs at the bottom
of the panel up out of the slots that they seat in.
4. Use a 5.5mm socket to remove 2 KEP nuts that secure the
observation light transformer and remove the transformer and
disconnect electrical connector
5. Disconnect the ribbon cable and the board’s 8 electrical
connectors.
6. Loosen the 2 captive screws that secure the power board, and
remove the board.

Figure 1. Power Board Panel


Released

Field Installation Instruction 2108779-001 Rev A 3


Rendered PDF File Page 5 of 5 DOC2099932, Rev:2

Fuse Removal/Installation Removal (Figure 2)

7. Using needle nose pliers remove fuse F7 and F8.


(See Figure 2 for fuse location)

NOTE : There are 4 Fuses (2 Sets of 2 each) are given in the FMI Kit .Use
only 2 Fuses . Additional 2 Fuses are given Just in case the Fuses are lost
during unpacking or Fixing at the sites. Since the Fuses are very small
and delicate to handle during replacements.

8. Once the fuses on the power board are replaced, re-install the
electronics into the enclosure in reverse order.

Figure 2. F7/F8 Fuse Location


Released

4 Field Installation Instruction 2108779-001 Rev A