Download as pdf or txt
Download as pdf or txt
You are on page 1of 84

Machine Translated by Google

MRL-7

data logger

User Manual
Manual Version: V01
2017-06-30

SOMMER GmbH
All rights reserved.
Machine Translated by Google

The copyrights for this manual are exclusively at the company

SOMMER GmbH
A-6842 Koblach

This manual may only be multiplied or passed on to third-parties with written permission of the company SOMMER. This applies also if
only excerpts of this manual are copied or passed on.
The same conditions apply for the passing on in digital form.

SOMMER GmbH
Strassenhaeuser 27
6842 Koblach
Austria
www.sommer.at

Distributed exclusively in the USA by


www.hydrologicalusa.com
Machine Translated by Google

CE compliance

This product is in conformity with the following standards


EN 60950-1
EN 61326-1
EN 61010
following the provision of directive R&TTE 1999/5/EC.

ÿ Information regarding images – image disclaimer


All images used in this manual are for illustrative purposes only. The actual product, the scope of
delivery or individual features may vary.

Safety information
Please read this entire manual before setting up or operating this equipment. Non-compliance
with this manual could result in damage to the equipment. Also in the case of non-compliance
injuries of individuals cannot be excluded totally.
To make sure that the protection provided of and by this equipment is not impaired, do not use or
install this equipment in any manner other than that specified in this manual.
Machine Translated by Google

Table of contents
1. Overview of the installation steps ........................................... ............................................ 1
2. Logger functions and interfaces .................................................. ....................................... 2
2.1. MRL-7 key functions and features ........................................... ........................................ 2
2.2. Measurement and data storage concept ........................................... ............................ 3
2.2.1. Performing measurements................................................ ............................................3
Analogue measurement example ..... .................................................. ................. 3
RS-485 measurement example ............................ ................................................ 3
SDI-12 measurement example ................................................. ........................... 4
2.2.2. Storing of measurement results and storage options................................................... .......5
2.2.3. Conditional storage................................................ .................................................. ..5
2.3. User interfaces ................................................ .................................................. .............. 6
2.3.1. Keyboard and display................................................. .................................................. 6
2.3.2. Display menu structure ................................................. ...................................................6
2.3. 3. Parameterisation and data transfer ................................................ ................................7
2.4. Sensor connections and sensor power supply ........................................... ..................... 7
2.4.1. Analogue interfaces ................................................ .................................................. .7
2.4.1. Digital interfaces................................................ .................................................. ......8
SDI-12 ........................................ .................................................. .................... 8
RS-485 ........................ .................................................. ....................................... 9
3. Hardware ................................................. .................................................. ........................... 10
3.1. How to open the housing ........................................... .................................................. 12
3.2. Mounting the MRL-7 ........................................... .................................................. ....... 13
3.3. Modem................................................. .................................................. ....................... 14
3.4. Micro SIM Card, Micro SD Card and Button Cell Battery ....................................... ........ 14
3.4.1. Insert / Exchange of Micro SIM Card................................................... ............................15
3.4.2. Insert / Exchange of Micro SD Card ........................................... ................................16
3.4.3. Exchange of button cells................................................ ................................................17
3.5. Pin description ................................................. .................................................. ............. 18
3.6. Spring Clips................................................ .................................................. ................. twenty-one
4.USB................................................ .................................................. .................................... 2. 3
4.1. Requirements for flash drive ................................................ ............................................ 23
4.2. Reading out data via USB ........................................... ................................................ 23 5
MRL-7 keyboard operation ........................................... .................................................. .. 25
5.1. Description of keyboard buttons, navigating the menu .......................................... ......... 25
5.2. Activating the display ................................................. .................................................. .... 26
5.3. Deactivating the display ................................................. .................................................. 26
5.4. Reading (current) measurement values ........................................... ............................ 26
5.5. Adjusting measurement values ............................................... ....................................... 26
5.6. Settings................................................. .................................................. ...................... 27
6. Parameterization ................................................ .................................................. ............... 29
6.1. Commander ................................................. .................................................. ................ 29
6.2. Defining a local connection with the MRL-7 ........................................... ........................... 29
6.2.1. RS-232................................................. .................................................. ....................30
6.2.2. Bluetooth................................................ .................................................. .................34
6.3. Commander: Establishing a connection ........................................... ............................ 39
6.4. Defining a station ................................................. .................................................. .......... 40
6.5. Establish a connection to a “Station”.................................................. ....................................
43 6.6. Downloading stored data ................................................. ................................................ 43
Machine Translated by Google

6.7. Remote connection via IP Call ........................................... ............................................ 44


6.8. Data transmission via HTTP/FTP.............................................. ....................................... 45
6.9. Time synchronization ................................................ .................................................. ... 45
7. Description of the parameters.............................................. ..................................................
46 8. Technical specifications ........................................... .................................................. ......74
9. Appendix................................................ .................................................. ............................ 75
9.1. Changing parameters via terminal.............................................. .................................... 75
9.2. Error messages................................................ .................................................. ............ 75
9.3. Index of tables ............................................... .................................................. .............. 77
9.4. Index of figures ............................................... .................................................. ............. 77
Machine Translated by Google
Machine Translated by Google

1. Overview of the installation steps


The following overview lists the most important steps for a full installation of the MRL-7 data
logger. The installation is divided into creating and establishing a connection with the MRL-7
and defining a station. Additionally, the most important settings and measurements to be
performed have to be defined.

1. Installation of “Commander” (chapter 6.1)


2. Defines “Connection” for being able to communicate with the MRL-7 (chapter 6.2)
3. Establish a “Connection” with the MRL-7 (chapter 6.3)
4. Define “Station” for easy data management (chapter 6.4)
5. Set the MRL-7 parameters as required
to. Set the language and decimal character (chapters JA and JB)
b. Set the station number and station name assigned to the MRL-7 (chapters A
and B)
c. Set the global measurement and storage interval (chapters C and D)
d. Defines the number of measurement values to be stored (chapter E)
and. Define measurements to be performed and/or values to be retrieved from
connected sensors (chapter G)
F. Configure the modem and data transmission parameters (chapters 6.7, 6.8 and
Description of the Parameters chapter I)
g. Adjust the advanced (“Technics”) settings as per your requirements (chapter J)
6. Download stored data (chapters 4.2 and 6.6)

Attention: You must not connect any voltages larger than 30 volts anywhere to the
logger or to any pin. Applying voltages above 30 volts you risk your health, physical
damages on the device or even a fire!

www.sommer.at page 1
Machine Translated by Google

2. Logger functions and interfaces


2.1. MRL-7 key functions and features

Fig. 1: MRL-7 key functions and features

www.sommer.at page 2
Machine Translated by Google

Advantages and main characteristics of the MRL-7 data logger

• Reliable and energy-saving: optimized to collect monitoring data in remote areas without
any infrastructure
• Logger can be operated fully independently (with batteries or solar-module), integrated
solar charge controller; no equipment cabinet required due to IP 67 rating
• Ideal integration and compatibility with complete SOMMER sensor family as well as
common sensor technology products
• Backlit display for easy data readout and logger parameterization
• Parameterisation via RS-232 and Bluetooth or via IP-call / modem connection
• Data readout with USB memory stick, Bluetooth-interface, RS-232 or via IP-call / modem
connection
• Minimum energy demand
• Solid, compact, weatherproof aluminum box
• Integrated remote data transmission through modem (2G / 3G and optionally 4G)

2.2. Measurement and data storage concept


2.2.1. Performing measurements
Analog measurement example
In this example three analogue inputs (An1, An2, An3) are measured. The diagram below
shows the first and second measurement intervals.

Fig. 2: Analogue measurement

Every measurement interval starts with the user defined holdback time (see JIE). Then An1 is
measured three times (see JIF and JIG) and so are An2 and An3.
After the last measurement is completed, the sensor supply voltage is switched off until the
next measurement interval starts again.

RS-485 measurement example


In this example three measurement values are requested from a sensor connected via the
RS-485 interface. The diagram below shows the first and second measurement interval.

www.sommer.at page 3
Machine Translated by Google

Fig. 3: RS-485 measurement


At the beginning of every measurement interval the MRL-7 activates the RS-485 interface and
issues three trigger commands to the connected sensor. As soon as all requested results are
received by the data logger, the RS-485 interface is switched off automatically.
If the requested results are not received within the “Trig, timeout” (see JGI) the RS-485 interface
is also switched off and the last values received are stored (again).

SDI-12 measurement example


In this example the measurement values are requested from three SDI-12 sensors connected.
The diagram below shows the first and second measurement intervals.

Fig. 4: SDI-12 measurement


At the beginning of every measurement interval the MRL-7 issues an M-command to the first
sensor which is responded with information about the number of measurement values available
and the time until the data is ready to be requested by the data logger. After eg one second
the MRL-7 issues a D-command to retrieve the measurement results from the sensor.
In this example the SDI-12 sensor addresses are “0”, “1” and “2”.

Table 1: SDI-12 example commands and answers

Example commands and answers


Command issued Answer received Comment
at 0M! 00013<CR><LF> 3 values are available in 1 second
b 0D0! 0+1.1+2.2+3.3><CR><LF> 3 values: “1.1”, “2.2”, “3.3”
c 1M! 10022<CR><LF> 2 values are available in 2 seconds
d 1D0! 1+4.4+5.5<CR><LF> 2 values: “4.4”, “5.5”
and 2M! 20031n<CR><LF> 1 value is available in 3 seconds
f 2D0! 2+6.6<CR><LF> 1 value: “6.6”

For detailed information about SDI-12 please see the official website www.sdi-12.org.

www.sommer.at page 4
Machine Translated by Google

2.2.2. Storing of measurement results and storage options


Measurement results are recorded in a separate storage interval.
ÿ The storage interval must always be equal or bigger than the set measurement interval.

The reason for using a separate storage interval is the advantage of being able to store
minimum, maximum and/or mean values of measurement results. These can be set with the
software Commander (see chapter G “Measurements, table”, column “Function”) or via the
terminal menu (see 9.1).

Fig. 5: Example storage interval


In above example the analog inputs An1, An2 and An3 are used to measure the outputs of
three sensors. Because the storage interval covers two measurement intervals, it is possible
to store either the mean value, the minimum or the maximum values measured within the set
storage interval. If none of the aforementioned options is chosen, the stored value will be the
last available measurement result(s) within the storage interval.

2.2.3. Conditional storage


Some measurement values are only stored when their value has changed since they were last
stored. For these measurement values an alternative storage interval can be defined (see
JDB) and the values can be reset to “0” once per day (see JDA).
The sensor types (“S-Typ”) using the conditional storage interval by default are “COUNT” and
“SYS” values (see chapter G “ Measurements, table” for more information).

www.sommer.at page 5
Machine Translated by Google

23. User interfaces


2.3.1. Keyboard and display
The MRL-7 display is being activated by pressing any of the buttons.

The cursor buttons , , and are used to navigate the on-screen menu.

Key Function(s)
Exit the current menu / Abort

Menu item down / Decrease value

Menu item up / Increase value

Enter the currently displayed menu item / Confirm /


Trigger measurement

2.3.2. Display menu structure


The following table shows the on-screen menu structure.
It shows an example setup consisting of a temperature and humidity measurement and a snow
depth measurement.

Table 2: Display menu structure

Start screen Displays the station number, current date and time.
* Measurements *
Main menu item containing all measured values
01 Temperature (eg -5.58 °C) *
02 Humidity (eg 62.35%) *
03 Snow depth (eg 436 mm) *
*** Stored V *** Main menu item containing all stored values
01 - 03 *
Temperature, Humidity, Snow depth

***Settings*** see chapter 5.6 „Settings“


**** Status **** Status information

1 Station No.

2 SOMMER ID

3 Station name

4 date

5 time

www.sommer.at page 6
Machine Translated by Google

6 Software version

7 ** IMSI

8 ** State

9 ** Type (Modem)
10 ** Signal (CSQ)

“*”
ÿ Values marked with are measured / stored values, all others are system values.
“**”
ÿ Menu items marked with relate to the logger modem.

2.3.3. Parameterisation and data transfer


Several interfaces are available to download stored data and/or to change MRL-7 parameter
settings:

Table 3: Interface types and features

Interface type data readout changing parameters

Keyboard and display

USB flash drive (see chapter 4) x

Bluetooth connection (see chapter 6.2.2) x x

Serial interface (see chapter 6.4) x x

IP call (see chapter 6.7) x x

automatic data transmission via x


HTTP/FTP (see chapter 6.8)

Time synchronization via time server –


NTP (see chapter 6.9)

2.4. Sensor connections and sensor power supply


2.4.1. Analogue interfaces

Table 4: Analogue inputs

Analog input type/range

An 1 0…2.5V

An 2 0…2.5V

www.sommer.at page 7
Machine Translated by Google

configurable:
- 0…2.5V
An 3
-NTC
- resistor > 2kÿ

configurable:
- 0…2.5V
- 0…0.3V
An 4
- resistor > 1k25ÿ
- resistor < 1k25ÿ
-PT1000

Wind – speed frequency measurement

Wind – direction resistor measurement

Counter 1 counts voltage pulses (min. 5V)

Counter 2 counts voltage pulses (min. 5V) up / down

Attention: Counter 2 is no additional input. It replaces wind direction (either/or)!

All analog sensors can be supplied via the internally provided voltage outputs. See technical
data of MRL-7 and sensor used for more details and specific requirements.

Table 5: Supply voltage outputs

Output maximum load

5V-Out 100mA (max. sum current over all sensors)

12V-Out 200mA (max. sum current over all sensors)

2.5V Reference 4mA

4 x switching outputs 1 x 1.10A


3 x 0.50A

max. total current: 2A

2.4.1. Digital interfaces


SDI-12
SDI-12 sensors can be connected to the MRL-7 via the “SDI12 – D” port. (See chapter 3.5 “Pin
description” for the port location.)

When data is requested from several SDI-12 sensors the sequence of requests starts with the
first sensor as specified in the “Measurements, table” (see item G).

www.sommer.at page 8
Machine Translated by Google

Example:

Fig. 6: SDI-12 example configuration

Attention: SDI-12 sensors must be supplied either separately or via the MRL-7 12V
output (max. 200mA). They must not be operated via switched supply lines. For this
reason the parameter setting of the sensor supply (see chapter JIB “12V sensor supply”)
has to be set to “always on”.

RS-485
Sensors manufactured by SOMMER with an RS-485 interface are connected to the MRL-7 via
the port “RS485 A”. (See chapter 3.5 “Pin description” for the port location.)
The RS-485 interface is always ready to receive measurement values when activated (see
chapter JG “RS485-2”). The polling and trigger procedure is explained in chapter 2.2.1.2.
“RS-485 measurement example”.

Attention: When more than one sensor is connected to the RS-485 bus, data
communication conflicts can occur if multiple sensors send their results to the MRL-7 at
the same time.

The prefix option (see chapter JG “RS485-2”) has to be active when multiple sensors are
connected so that recorded data can be assigned to the correct device.

www.sommer.at page 9
Machine Translated by Google

3. Hardware
The MRL-7 is available in two different configurations:
- MRL-7: as per standard configuration and without integrated battery
- MRL-7B: with a higher housing to accommodate an integrated battery

Fig. 7: MRL-7

Fig. 8: MRL-7B

www.sommer.at page 10
Machine Translated by Google

Drawings of MRL-7 housing:

Fig. 9: MRL-7 housing, dimensions (in mm)

Fig. 10: MRL-7B housing, dimensions (in mm)

www.sommer.at page 11
Machine Translated by Google

3.1. How to open the housing


Attention: Before the housing is opened, the power supply must be disconnected in
order not to risk an electric shock or an electrical short on the hardware.

• Remove the covering strips on the left and right side of the housing.

Fig. 11: MRL-7 covering strips

• Unscrew the screws using a Philips head no. 2 or a straight head screwdriver.

Fig. 12: MRL-7 screws

www.sommer.at page 12
Machine Translated by Google

• Open the housing by flipping the lid open.

Fig. 13: Opening the housing of the MRL-7

3.2. Mounting the MRL-7


After removing the covering strips the MRL-7 can be mounted on a backplane with four screws:

Fig. 14: Holes for Mounting

Compatible screws:
• M4 cheese-head screw (included within scope of supply)
• M5 hexagon socket head cap screw

www.sommer.at page 13
Machine Translated by Google

3.3. Modem
Data recorded by the MRL-7 can be transmitted to user-defined servers (HTTP and/or FTP)
via the integrated modem.
In order to minimize the power consumption the data logger controls the modem supply voltage.
Thus, the modem is only switched on when a data transfer has to be performed.
ÿ To activate the modem the MRL-7 parameter “Modem functionality” must be set to “on”.
For more details and all further modem configurations please see chapter I “Modem”.

3.4. Micro SIM Card, Micro SD Card and Button Cell Battery
The following picture shows the slots respectively the bracket where to find, insert and
exchange the micro SIM card, the micro SD card and the button cell of the MRL-7.
To insert and/or switch the micro SIM card, the micro SD card or the button cell, open the
MRL-7 housing as described in chapter 3.1 “How to open the housing”.

Attention: Before the housing is opened, the power supply must be disconnected in
order not to risk an electric shock or an electrical short on the hardware.

Micro SD card

button cell

Micro SIM card

Fig. 15: Lid of the MRL-7 from inside

www.sommer.at page 14
Machine Translated by Google

3.4.1. Insert/Exchange of Micro SIM Card


1. To insert / exchange the
micro SIM card gently
press on the covering
with your finger and shift
it a little to the side.

2. Then flip the cover open.

3. The micro SIM card can


now be inserted,
removed or exchanged.

Note the correct


orientation of the card to
be inserted (it will only fit
one way).

www.sommer.at page 15
Machine Translated by Google

4. Flip the cover back again


and lock it by sliding it
back to its old position.

ÿ The lock and un-lock position is also marked on the micro SIM card cover itself.
ÿ Note that the micro SIM card is not within the scope of supply.

3.4.2. Insert/Exchange of Micro SD Card


1. To insert / exchange the
micro SD card gently
press on the covering
with your finger and shift
it a little back.

2. Then flip the cover open.

www.sommer.at page 16
Machine Translated by Google

3. The micro SD card can


now be inserted,
removed or exchanged.

Note the correct


orientation of the card to
be inserted (it will only fit
one way).

4. Flip the cover back again


and lock it by sliding it
back to its old position.

ÿ The lock and un-lock position is also marked on the cover itself. Look closely.
ÿ Note that the SD card is not within scope of supply

3.4.3. Exchange of button cells


The MRL-7 is delivered with a button cell built in. This little battery ensures that the time and
date function of the logger is not interrupted during the time when the logger lacks any other
power supply.

www.sommer.at page 17
Machine Translated by Google

1. To exchange the button


cell use a screw driver
and gently push the cell
out of its holding bracket
from behind.

To insert a new cell


simply push it with your
fingers into the holding
bracket from the other
side.

3.5. Pin description

Fig. 16: MRL-7 pin description

www.sommer.at page 18
Machine Translated by Google

Table 6: MRL-7 pin assignment

connector Pin(s) function

Left-hand side connectors:

x1 - Battery connector (-)


[internal and/ or external battery] 1)

x1 + Battery connector (+), max. 14V


[internal and/ or external battery] 1)

x2 - Solar Supply voltage (-) or solar panel connector (-)

x2 NC Not connected / do not use!

x2 dis CR Disabling internal charge regulator

x2 + Solar Supply voltage (+) or solar panel connector (+)

X5 +RS485A RS-485A

X5 -RS485B RS-485B

X5 SDI12 SDI-12 sensor connector

X5 GND ground connector

X19 5V-Out 5 V output for sensor supply (max. 100 mA) *)

X19 Counter 2-b Counter 2 secondary input for encoders

X19 Counter 2-a Counter 2 input

X19 GND Ground connector

x10 Counter 1 Counter 1 input

x10 GND Ground connector

x12 2.5V Ref-Out 2.5V reference voltage output

x12 Wind-Dir. Wind direction potentiometer input

x12 Wind-Sp. Wind speed input

x12 GND Ground connector

Right-hand side connectors:

X8 AGND Analog ground connector

X8 AN1 Analog input 1

X8 5V-Out 5 V output for sensor supply (max. 100 mA) *)

X8 12V-Out 12 V output for sensor supply (max. 200 mA) **)

www.sommer.at page 19
Machine Translated by Google

X22 AGND Analog ground connector

X22 AN2 Analog input 2

X22 5V-Out 5 V output for sensor supply (max. 100 mA) *)

X22 12V-Out 12 V output for sensor supply (max. 200 mA) **)

X9 AGND Analog ground connector

X9 AN3 Analog input 3

X9 5V-Out 5 V output for sensor supply (max. 100 mA) *)

X9 12V-Out 12 V output for sensor supply (max. 200 mA) **)

X11 CurrL Differential current input (-)

X11 V.L. Differential voltage input (-)

X11 AN4/VH Differential voltage input (+)

X11 CurrH Differential current input (+)

X20 12Vbat OUT 12 V power supply (> max. 1 A)

X20 12Vbat OUT 12 V power supply (> max. 1 A)

X20 12Vbat OUT 12 V power supply (> max. 1 A)

X20 GND ground connector for 12 V power supply

X21 SW1 switching output 1, 12V

X21 SW2 switching output 2, 12V

X21 SW3 switching output 3, 12V

X21 GND ground connector for switching output, 12 V


1
) External batteries, if attached here, must always be of the same type as the internally used
one, if the internal and the external battery are used/attached together.
*) 100 mA sum current over all sensors
**) 200 mA sum current over all sensors

Attention: You must not connect any voltages larger than 30 volts anywhere to the
logger or to any pin. Applying voltages above 30 volts you risk your health, physical
damages on the device or even a fire!

www.sommer.at page 20
Machine Translated by Google

3.6. Spring Clips


The MRL-7 is delivered with an assortment of spring clips for connecting sensors and power
supply lines.

1. 4-pin spring clip

2. Use a flat head screw


driver to compress the
internal spring.

3. Push the screw driver in


to lock down the internal
spring and open the
connection terminal.

www.sommer.at page 21
Machine Translated by Google

4.Insert the wire.

5.Insert the wire


completely.

6. Pull out the screw driver


to release the spring and
fix the wire.

www.sommer.at page 22
Machine Translated by Google

4.USB
4.1. Requirements for flash drive
ÿ The used flash drive must be formatted as FAT32.
ÿ There must be no file (in the main directory) present from a previous readout of that specific MRL-7 (same Station
number).
ÿ The file created on the flash drive is named “Stationnumber.csv”.

4.2. Reading out data via USB


The following steps describe the data readout to a connected flash drive:

ÿ Remove (unscrew) the cap that protects the USB port.


ÿ Plug in a USB flash drive into the USB connector

Fig. 17: MRL-7 USB Connector

ÿ Simply press the button until the display shows the message “searching Stick”. The MRL-7 searches for a USB flash
drive for six seconds. Plug in a USB flash drive now (or also before) and all data are automatically copied to the
USB flash drive since the last readout.

This process may take up to three minutes. During the data download the message “Transfer running” is
displayed.
ÿ After the data transfer is finished, the message “data copied” is displayed.

ÿ The background light of the MRL-7 turns off during the data download to save energy.

www.sommer.at page 23
Machine Translated by Google

ÿ To transfer all stored data follow these steps:


o Press and hold the button until “searching for Stick” is displayed.

o Now additionally and promptly press the button , release the button again
and after that also release the button again.

Alternatively, all data can be copied to a connected flash drive via the “Settings” menu.
See chapter 5.6 for more information.

Attention: After reading out data to a flash drive, the USB port cap must be screwed on
again. When doing so, please take care that the threads of the cap and the USB port are
well aligned and the cap can be screwed on easily.

If screwing on the cap does not work smoothly, it maybe is not put on nicely so that the threads
do not fit well. Unscrew the cap and carefully try again.

www.sommer.at page 24
Machine Translated by Google

5.MRL-7 keyboard operation

Fig. 18: MRL-7 keyboard

5.1. Description of keyboard buttons, navigating the menu

The cursor buttons , , and are used to navigate the on-screen menu.

The button is used to search for a USB flash drive, to activate/search for a Bluetooth
connection or for special functions as described further below.

Key Function(s)
Exit the current (sub)menu / abort

Menu item down / decrease value

Menu item up / increase value

Enter the currently displayed menu item / confirm /


trigger measurement
Searching stick / connect Bluetooth/ special
functions

ÿ When Bluetooth connection is activated, any further manual access to the display menu
is denied. The message “BT activate no access!” is displayed.

www.sommer.at page 25
Machine Translated by Google

Attention: The button also aborts any activities on the logger, such as activate Bluetooth
connections, Modem test mode function, data downloads to USB flash drive etc.

5.2. Activating the display


The MRL-7 display is activated by pressing any button for at least one second and shows the main
screen (= station number, current date and current time). After four seconds without any activity
on the logger, the display switches and shows automatically the first measurement value.

5.3. Deactivating the display


The display will automatically be deactivated / go asleep again in order to minimize the logger's
energy consumption if no further button is pressed after activating the display for at least 10
seconds.
In case any further buttons are pressed after activating the display, the display will automatically
be deactivated / go asleep again only after 1 minute.

Alternatively, the display can also be deactivated by pressing and holding the button for at least 2
seconds. The message “Access end hold key!” is displayed. Keep pressing the button
and the display will be deactivated shortly thereafter.

5.4. Reading (current) measurement values


Activate the MRL-7 display by pressing any button. After 4 seconds the display will automatically
show the first measurement value.

Or, after activating the display, press any button (except the button first ) again to show the
measurement value in the list immediately.

To navigate the list use the buttons and .

To manually trigger a measurement of a single measurement value press the button.

5.5. Adjusting measurement values


Enter the measurement value list as described in the chapter before and navigate to the value to
be adjusted.

To adjust a single measurement value press and hold the Now for at least two seconds.
button enter the access code as described below.

To adjust the measurement value use the buttons and .

www.sommer.at page 26
Machine Translated by Google

Press the button to confirm the adjustment, or press to cancel the adjustment.

ÿ Access code: To enter the access code press the key sequence 2x followed by 2x

Attention: As long as the display is active during the ongoing session, the entered access
code unlocks all system settings of the logger.
ÿ Press and hold a button to increase or decrease the selected measurement value more
quickly (increase/decrease speed doubles every two seconds).

ÿ Instantly adjust a measurement value to “0” by pressing the button.

5.6. Settings
Some MRL-7 settings can be adjusted via the display menu.
To adjust the settings activate the MRL-7 display by pressing any button.

Press twice and select the menu “*** Settings ***”. Confirm your selection by pressing

and enter the access code as described above.


To change settings or to start a selected special mode navigate to the desired menu item and

press . To cancel inputting a value or abort a special mode press .

Table 7: ***** Settings *****

***** Settings ***** Example Value Description


1 Station No. 20160111 Station number
2 Date 06.04.2017 Current date
3 Time 09:26:28 Current day time
4 Exposure lock off
5 Copy all data
6 Erase all data
7 Continuous M turn on Continuous Measurement
8 Adjust Contrast weak Adjusts the display contrast
9 Reboot Device Reboots the MRL-7 without the
need to switch the supply
voltage.
10 Modem Testmode Only available, if a modem is
configured at all

4 Exposure lock

www.sommer.at page 27
Machine Translated by Google

When activated reading measurement values via the MRL-7 display is only possible after inputting the access code.
Also copying stored data to a connected flash drive or via a Bluetooth connection requires the user to input the access
code. (see also chapter JCE
“Exposure lock”).

This applies to all measurement values except for special observer inputs.

5 Copy all data

When selected all stored data are copied to a connected USB flash drive.

6 Erase all data

Attention: When selected, all data stored on the MRL-7 are deleted!

7 Continuous Measurement

Activates a special continuous measurement mode. Measurements are performed in the shortest possible interval.
This option is intended for testing setups.

The continuous measurement mode is deactivated automatically after three minutes.

8 Adjust Contrast

“Adjusts” the display contrast. There are four different levels available: base, weak, high and medium.

9 Reboot Device

Reboots the MRL-7 without the need to switch the supply voltage. This option is helpful for performing firmware
updates, which requires the device to be restarted.

10 Modem Testmode

A number of automatic tests are performed and related information is displayed:


ÿ The modem is initialized.

ÿ The SIM card IMSI number is displayed for a short amount of time.

To freeze the display press and hold the button.

ÿ The signal strength is tested and displayed.


ÿ The defined connections are established:

o The MRL-7 synchronizes its time with the defined time server.
o Data are sent to all defined servers.

o The MRL-7 IP call function is activated.

www.sommer.at page 28
Machine Translated by Google

6.Parameterization
6.1. Commander
The MRL-7 is parameterized with the Commander software or via the terminal menu (see
chapter 9.1).
To edit the parameters with the PC software Commander, the “Expert” mode must be active
(for more details see Commander manual):

Fig. 19: Setting the “Expert” mode in the “Commander” software

The PC software Commander highlights the edited parameters. After uploading the changes
to the MRL-7, the modified parameters are active (for more details see Commander manual).

ÿ The installation software of the Commander is delivered on a USB flash drive together
with the MRL-7.

6.2. Defining a local connection with the MRL-7


The connection between the MRL-7 and a computer is established via the USB/RS-232
converter (not within scope of supply) or via Bluetooth.

www.sommer.at page 29
Machine Translated by Google

6.2.1. RS-232

Fig. 20: USB/RS-232 converter Fig. 21: Connecting the USB/RS-232


converter

Plugin the USB connector into any USB port of your computer and use the Commander software
to define a connection. Then set the parameters as required using the “Communication assistant”
as described hereafter.

Install and start the


Commander software.
Launch the
“Communication
assistant” and follow
the given instructions.
Use the “Next” and
“Back” buttons to adjust
the different settings
from one step to
another or press
“Cancel” to abort
the communication assistant.

Select “Serial
connection” when using
the communication
cable.

www.sommer.at page 30
Machine Translated by Google

Make sure the MRL -7 is


supplied with power
(either internal batteries
and/or external power
supply), and the
communication cable is
plugged into the data
logger on one end and
the other end is
connected to a USB port
of the laptop running
“ Commander.”

Select “Logger (115200


Bd)”

Select the COM port


that was assigned to the
connection cable from
the drop down list.
If more than one COM
port is listed and you are
unsure which to select,
disconnect the
connection cable and
repeat the search. The
COM ports listed now
are not assigned to the
connection cable.
Plug in the converter
and repeat the search
once more. Now an
additional COM port will
be listed. This is the one
assigned to the
connection cable.

The communication
assistant dynamically
changes between 7 and
10 steps, depending on
whether an existing

www.sommer.at page 31
Machine Translated by Google

connection can be
selected or if a new
connection has to be
established (additional
steps to define specific
port and/or station
information).

Assign a name to the


connection for later use.

An automatic search for


connected devices is
performed. This step
can take up to two
minutes.

www.sommer.at page 32
Machine Translated by Google

If a station already
exists, it will be
recognized and
automatically selected.
It can be decided
whether or not to assign
the found device(s) to
the “Station”.

If no station is found, a
“Name” for the station
can be defined.
Select “Yes” to proceed.

The station created in


the step before is
selected automatically
and is ready to be used.

Alternatively and without using the “Communication assistant”, the communication parameters
can also be set manually within the “Connections” tab. There are different possibilities to display
the “Connections” tab:
ÿ by pressing “F8” on your keyboard
ÿ by selecting “Connections (F8)” in the main menu item “Tabs” (for temporary display only)
ÿ by selecting “Show connections tab” that can be found in the main menu item “Options”,
submenu item “Tabs” (for permanent display).

www.sommer.at page 33
Machine Translated by Google

ÿ For further information on how to define connections, please see the Commander
manual.

6.2.2. Bluetooth
Either an internal Bluetooth module or a USB Bluetooth dongle has to be available on the
computer used.

Use the “Commander” software to define a Bluetooth connection as described hereafter.

ÿ Before a Bluetooth connection can be established the MRL-7 Bluetooth transceiver must

be switched on. Press and hold the button until the message “hold for BT” is

displayed. Keep holding the button until the message “waiting for BT no access” is
displayed. Now a Bluetooth connection can be established with the MRL-7.
ÿ The MRL-7 Bluetooth ID is printed on a sticker on the MRL-7 housing.

www.sommer.at page 34
Machine Translated by Google

Install and start the


software “Commander” .
Launch the
“Communication
assistant”.
Use the “Next” and
“Back” buttons to adjust
the different settings
from one step to
another or press
“Cancel” to abort
the communication assistant.

Select “Bluetooth” for


type of connection and
follow the instructions of
the assistant.

The first time a


Bluetooth connection is
established with
Commander a “New
connection” needs to be
selected.
The communication
assistant dynamically
changes between 8 and
9 steps, depending on
whether an existing
connection can be
selected or if a new
connection has to be
established (additional
steps to define specific
port and/or station
information).

www.sommer.at page 35
Machine Translated by Google

This step might take a


few seconds.

www.sommer.at page 36
Machine Translated by Google

This step might take a


few seconds.

If no matching station
exists, a “Name” for the
station can be defined.

If a station already
exists, it will be
recognized and
automatically selected.

Select “Yes” to proceed.

www.sommer.at page 37
Machine Translated by Google

The station created in


the step before is
selected automatically
and is ready to be used.

Alternatively and without using the “Communication assistant”, the communication parameters
can also be set manually within the “Connections” tab. There are different possibilities to display
the “Connections” tab:
ÿ by pressing “F8” on your keyboard
ÿ by selecting “Connections (F8)” in the main menu item “Tabs” (for temporary display only)
ÿ by selecting “Show connections tab” that can be found in the main menu item “Options”,
submenu item “Tabs” (for permanent display).

www.sommer.at page 38
Machine Translated by Google

ÿ For further information on how to define connections, please see the Commander
manual.

6.3. Commander: Establishing a connection


To establish a connection to an MRL-7 either a “Connection” or “Station” (6.4 “Defining a station”)
can be selected. When a station is selected one of the assigned connections must be chosen.

The parameterization of the MRL-7 is performed with the Commander software. To establish a
connection with the device follow the steps described in the previous chapter.

To download the MRL-7


parameters switch to the
“Parameters (F2)” tab
and click “Download
parameters”.
When downloading the
parameters for the first
time, the process will
take approximately 2
minutes. When the
process is repeated, the
download is completed
in a shorter amount of
time.

Fig. 22: Download Parameters

www.sommer.at page 39
Machine Translated by Google

After the download has


been completed a list of
all parameters is
displayed.
All parameters are
described in chapter 7
“Description of the
parameters”.

Fig. 23: List of parameters

6.4. Defining a station


To download stored data, to comfortably manage several MRL-7 devices and to enable additional
functions (eg IP call), it is advised to create a “station” in Commander.

To create a station
select the “Stations (F7)”
tab and click “New
Station”.

Fig. 24: New Station

www.sommer.at page 40
Machine Translated by Google

Input the Station ID and


SOMMER ID set in the
MRL-7.
When delivered both IDs
are identical with the
device's serial number.
The MRL-7 serial
number (S/N) is printed
on the device housing.

Fig. 25: Station no. and SOMMER ID

Fig. 26: Station no. and SOMMER ID

Select which previously


defined “Connections”
(Serial, Bluetooth, IP
Call, …) to assign to the
“Station”.
The default
“Connection” for this
station can be set by
checking the circle next
to the connection type.

Fig. 27: Assigning connections to a station

www.sommer.at page 41
Machine Translated by Google

Depending on the
connection types
assigned to the station
additional information
must be inputted.
Please see chapter
6.2.2 “Bluetooth” and
chapter 6.7 “Remote
connection via IP Call”
for further details.

Fig. 28: Connection details

When data are


downloaded from a
connected MRL-7 they
are saved in an
“archive” file by default.
Each archive file
contains all data from a
year or month,
depending on the
chosen “Archive Type”.
When set to “None” all
data recorded by the
MRL-7 will be saved in
one file.
The default location of
downloaded data files is
“C:\Users\Public\Docum Fig. 29: Data archive
ents\Sommer\Data\”

To finish the session


press “Save station”.

Fig. 30: Saving a station

www.sommer.at page 42
Machine Translated by Google

6.5. Establish a connection to a “Station”

To connect to a station
select the
“Station” mode in the
“Communication” window.
Choose the station
name and select a
connection via which the
MRL-7 is reachable.
The settings are applied
immediately.
Click “Connect” now.
After the connection is
established successfully
it is now possible to
download parameters
from the MRL-7, to read Fig. 31: Connecting to a station
out measurement data
or to establish a terminal
connection.

6.6. Downloading stored data


To download data from
an MRL-7 select the
“Data (F4)” tab in
Commander. Choose
the “Station” and select
a connection via which
the MRL-7 is reachable.
Now click the
“Transfer data”
button to download all
stored data of the
station selected from the data logger.

Fig. 32: Downloading stored data

www.sommer.at page 43
Machine Translated by Google

Click “Download data


manually” to download
data stored since a
manually chosen date
and time. Use the slider
to select the timeframe.

Fig. 33: Downloading stored data manually

6.7. Remote connection via IP Call


To establish a
connection to an MRL-7
via an IP call, the MRL-7
has to be assigned to a
station which is set up
for IP call connection.
To establish a remote
connection, the SIM
card's IMSI number has
been entered in the
station settings.
The IMSI number of an
integrated SIM card can
be retrieved:

ÿ reading out the Fig. 34: IMSI number


device status via the
“Parameters” tab
(F2)”, “K Special
Functions” (see
chapter KA “Device
status”)
ÿ using the “Modem
Testmode” feature
(see chapter 5.6
“Settings”)
ÿ or reading the
menu 4
“Status” (item 7
“IMSI”) via the
logger display itself.

www.sommer.at page 44
Machine Translated by Google

Click “Save station” to


store the settings.

Fig. 35: Station no. and SOMMER ID

To establish the
connection select
“Station” as “Mode”,
choose the station you
want to connect to and
then select the “IP Call”
connection.
The settings are applied
immediately.
Now either parameters
can be downloaded from
the MRL-7 or a terminal
connection established
by clicking “Connect”.

Fig. 36: IP call connection

6.8. Data transmission via HTTP/FTP


When data are transferred to a server via HTTP or FTP protocol (see chapter IC “Data
transmission 1”), all data stored on the logger will be sent to the defined server(s).
The first transmission includes all data recorded so far. Further transmissions will include all
data recorded since the last successful transmission.
In case the data could not be transferred (eg because of bad cell reception) the “missing” data
will be sent on with the next successful transmission.

6.9. Time synchronization


The MRL-7 can synchronize its internal clock with a time-server (NTP-Server).
The time synchronization settings are described in chapters IF “NTP synchronization time”
through II “NTP timezone”.

www.sommer.at page 45
Machine Translated by Google

7. Description of the parameters


The MRL-7 settings are opened and adjusted with the SOMMER PC software "Commander".
Alternatively, all settings can be accessed and modified via a terminal program (see chapter
9.1 “Changing parameters via a terminal”).
The PC software “Commander” highlights the edited parameters. After uploading the changes
to the MRL-7, the modified parameters are active.

ÿ The installation software of the Commander is delivered on a USB flash drive together
with the MRL-7.

ÿ For further details, please refer to the user manual for the PC software “Commander”.

main menu
A Station ID
B Station name
C Measurement interval
D Storage interval
E Measurements, max. number
F Aux measurements, max. number G
Measurements, table
H Messages, table
I Modem
J Technics
K Special functions

The parameters are arranged in a main menu with submenus. The menu items are opened
and closed by double-clicking them.

A Station ID
The station ID is set to the MRL-7 serial number by default but can be adjusted by users.
Value range 0...99999999 00000000 (default)

B Station name
User defined name of the station (max. 32 characters).

C Measurement interval
With this setting, the measurement interval can be defined.

Value range 00:00:02…12:00:00 00:01:00 (default)

www.sommer.at page 46
Machine Translated by Google

The measurement interval defines at which times measurements are performed.

D Storage interval
With this setting, the storage interval can be defined.

Value range 00:00:10…12:00:00 00:01:00 (default)

The storage interval defines at which times the measured values are saved.

E Measurements, max. number


Up to 99 measurement values can be defined for the MRL-7.
Depending on the ordered components, SOMMER preconfigures some measurements.
However, the end user can define additional measurements. The configuration steps are
described in chapter G – see further below.

Value range 1…99 4 (default)

ÿ The maximum of 99 measurement values refers to the measurement values as set in chapter
“E” and “F” (measurements and auxiliary measurements) in total.

F Aux measurements, max. number


Auxiliary measurement values. Up to 99 auxiliary measurement values can be defined for the
MRL-7.
Same as the ordinary measurement values the auxiliary measurement values are displayed in
the measurement table (chapter G) however numbered starting from 99 downwards.
Auxiliary measurement values are set in order to be able to check additional/multiple limit
values.
In addition, an auxiliary measurement value allows defining an existing measurement value
with an altered setting (eg different unit, decimals, scale etc.).
Value range 1…99 1 (default)

ÿ The maximum of 99 measurement values refers to the measurement values as set in


chapter “E” and “F” in total.
Auxiliary measurement values are not stored and transmitted in the measurement data
file. They can only be read via the “Measurements, table” (chapter G) in the Commander
or the terminal window.

G Measurements, table
Column title Description
1…99 Unique number assigned to a measurement value.
This number specifies the position of the measurement value in the
MRL-7 output string (RS-232 and/or SDI-12). The same number is also
assigned to the measurement value when saved values are
downloaded from the data logger.

www.sommer.at page 47
Machine Translated by Google

function The function defines measurement which values are saved:


- off: No measurement is performed and no values are saved.
- current: The last measurement result within the active storage interval is
saved.
- meanval: The mean value of all measurements within the last
storage interval is saved.
- minimum: The lowest measured value within the last storage interval is
saved.
- maximum: The highest measured value within the last storage interval is
saved.
- sum: The sum of all measurement results within the last storage
interval is saved.
- intensity: The intensity (eg of the rainfall) is saved (only if the
counter input is used).
- diff.: The difference of the last stored value to the latest
measurement result is saved.
- custom1: currently not defined
- custom2: currently not defined
Identifier User defined name of the measurement value.

unit User defined unit of the measurement value.

Decimals You define -


1 … 5: 1 to 5 decimal places can be defined for the
the number measurement value.
of decimal places - none: No decimals are defined for the
assigned to the measurement value
measurement - as S: The source value's decimal place is
value. assumed for the MRL-7.

Scale User defined factor with which the retrieved measurement value is multiplied.

Attention: This option is only available when the “Decimals” are not set to “as S”.

Offset User defined offset that is applied to the retrieved measurement value.

adjustment By clicking this button, a measurement is performed and the result is outputted. In
case the outputted value is incorrect, the value can be adjusted by inputting the
correct value.

This adjustment only affects the “Offset” but not the “Scale”.
Test A measurement of the selected measurement value is performed and the result is
outputted.

S-Type Sensor type - AIN: analogue inputs


(source) AN1: analogue input 1
AN2: analogue input 2
AN3 (N): analogue input 3
AN4 (D): analogue differential input
- WIND: preconfigured for wind sensor
speed: wind speed
direct.: wind direction
- COUNT: counter input (registers and counts voltage
pulses at the counter input).
Counter 1
Counter 2a
Counter 2ab: quadrature encoder
- SDI12: a measurement value is retrieved from a
connected SDI-12 slave device.

www.sommer.at page 48
Machine Translated by Google

The sensor address is set via the “S -NUM” and


the measurement value position is defined via the
“S -MEA” (see details for the table “S -MEA” further
down).
- SBP: A measurement value is retrieved from a
connected SOMMER sensor that supports the
SBP protocol (via RS -485 connected to BUS
2/AB) .
The sensor address is set via the “S -NUM” and
the measurement value position is defined via the
“S -MEA” (see the following lines for more
information).
- MIO: A measurement value is retrieved from a
connected SOMMER sensor that supports the
MIO protocol.
The sensor address is set via the “S -NUM” and
the measurement value position is defined via the
“S -MEA” (see the following lines for more
information).
- SYS: A set of internally measured values is
available to be retrieved and stored:
key A: Counts the times the MRL - 7 is manually
activated via its keyboard.
key S: Counts the times the MRL - 7 is manually
activated via its keyboard with a special key
sequence.
manual: Observer input: Any time a manual
adjustment of this value is performed the newly set
value is stored unless it has not changed.
exep. A: Analysis value for SOMMER
exep. B: Analysis value for SOMMER
light: Ambient light measurement result
+Sup V: Supply voltage measurement
+ Bat V: Internal battery voltage
5 V sup: Internal 5V bus measurement
mo CSQ: Modem signal strength
mo ex cep.: Modem ex ception messages
mo srv1: Modem server 1
or srv2: Modem server 2
or srv3: Modem server 3
Switch: Current output switch status
Message: Number of successful message
transmissions since the last storage event.
ÿ The message number defined as per chapter “H
Messages, table” must be inputted in the
field “S -NUM”.
- RECYC: “Recycled” channel value
The source value is defined via the “S -NUM”
which corresponds with a (previously) defined
measurement value.
Example: Measurement value no. “03” is set to
measure analogue input 1 (AIN1) and to store the
“raw” value.
A RECYC -measurement value is defined as
measurement value no. “12”. The relating “S -
NUM” is set to “03” to retrieve the measured “raw”

www.sommer.at page 49
Machine Translated by Google

value.
Additionally the number of decimal places is set to
“2” to enable the option to set a “Scale” of “0.5”.
The stored value of measurement value no. “12”
will now always be half the measured value “03”
(AIN1).
S-NUM The sensor number assigned to the desired measurement source:
Examples:
- SDI-12 sensor address “3”
- RS-485 sensor address “2”
S-MEA The position in the input string of the measurement value to be
retrieved and stored.
Example: SDI-12 sensor output values= 0.0 + 1.1 + 2.2 + 3.3 + …
value position = “1” “2” “3” “4” …
To retrieve the value 1.1 the S-MEA has to be set to “2”.
S-ADD Additional - “_Cn”: SDI-12 C-command
commands to - “CCn”: SDI-12 C-command with CRC
be issued along - “_Mn”: SDI-12 M-command
a standard - “MCn”: SDI-12 M-command with CRC
request for data - _Rm”: SDI-12 R-command
to a source. - “RCm”: SDI-12 R-command with CRC

The different n … SDI-12 measurement cycle number


commands are m … SDI-12 measurement output line number
related to
- SW: “swap A/B” for S-TYP quadrature encoder
according - NR: “no reset” for function “sum”
sensor types - DD: for sum channels
(“S-TYP”) as
set. - SY: "sync" for async channels to store as
synchron
- AS: "async" for sync channels to store asynchron
- ST: "smaller than" for limits that fall below
- GS: "grenzwert status" to store the limit value
status of a specific channel
- TD: "trigger data" polling of data without triggering
the measurement
- TF: "trigger fake" something for S-TYP “MIO” add-
on
- SC: "subchannel" for S-TYP “SBP”

Example: The desired data is outputted when an “M2” command is


issued to the connected sensor. To ascertain the data
is transmitted correctly a CRC checksum is also
requested.
ÿ S-ADD = “MC2”

Limit User defined limit value that can be entered. The value format is the
same as for the measurement value itself. By default the measurement
value has to exceed the limit trigger condition. If the value shall be
decreed, the additional command “ST” (smaller than) must be set.
messages If a message for a specific measurement and its limit value is defined
(see chapter H “Measurements, table”), a box for the according
message number appears. Here the box of the message, that shall be
triggered once the limit value status is reached, must be ticked off.

www.sommer.at page 50
Machine Translated by Google

H Messages, table
The MRL-7 features a configurable messaging system. Messages are transmitted to the defined
containers whenever the “Limit” as set for a measurement value is exceeded or deceeded (according to
the message configuration).

Column title Description


1…16 Unique number assigned to a message. A maximum of 16 independent
messages can be set.
messages The message defines which kind of message is used:
- off: No message is sent
- Switch: A switching output is switched when the trigger condition is
fulfilled.
- E-mail: An E-Mail is sent to the defined recipient when the trigger condition
is fulfilled.
- text: A SMS text message is sent to the defined recipient when the trigger
condition is fulfilled.
Recipient User defined container
- Switch: setting not applicable / not available
- E-Mail: E-Mail address of the recipient
- text: Phone number of the recipient
Dialling codes can start with either “00” or “+”

Example: A text message (SMS) should be sent to a recipient in


Germany while the modem is installed in France.
The dialing code for Germany is either “0049” or “+49” followed by the
(national) phone number.
Subject Subject for email To be defined by user, max. 80 characters
messages. available.
(“Subject” is not
applicable / not Optional, message content codes can be used.
available for Switch See the following “Content Codes” table for detailed
and text.) description.
Content You define the content To be defined by user, max. 160 characters
of the message. available.
Optional, message content codes can be used.
See the following “Content Codes” table for detailed
description.
switch Defines which switching output is used (check boxes) – applicable only with
“Switch”.
hold You define the timeframe in seconds during which the switching output is
switched as soon as the trigger condition is fulfilled.

Example: Switching output number 2 shall be activated for 30 seconds after


the trigger condition is fulfilled:
Setting “switch” = “02” (via check box)
Setting “Hold” = “30”

Content Codes

The message content can be configured according to end user's requirements.

www.sommer.at page 51
Machine Translated by Google

To insert standard message content more easily a number of inline content codes are available.

code Description
%sid% Station ID assigned to the MRL-7
%sname% Station name assigned to the MRL-7
%cname% Name of the channel that triggered the message
%cunit% Unit assigned to the channel that triggered the message
%time% Time of the message transmission
%date% Date of the message transmission
%cval% Latest channel measurement result at the time of the message
transmission
%tval% Set limit value that was exceeded/deceeded
%rval” Measurement result that triggered the message output
%wnum% Message number as defined per MRL-7 setup (01 … 16)
%cp% Measurement result of a user defined channel.
Example:
Channel number “02” triggers an E-Mail message. This message shall
contain the channel name and value of the measurement result that
triggered the message and additionally the current result of channel
number “14”.
“Content” = %cname% %rval% Ch.14 m.value = %cp14%

I Modem
Modem unit
A modem functionality off
B Modem config
C Data transmission 1
D Data transmission 2
E Data transmission 3
F NTP synchronization time 02:08:43
G NTP server
H NTP port 123
I NTP timezone 3600 sec
JBackground function off
K Standby, start time 08:00:00
L Standby, duration 01:00:00
M IP Call server mds.sommer.at
N IP Call port 4647
O IP Call interval 00:01:00
P Email/SMTP

www.sommer.at page 52
Machine Translated by Google

IA Modem functionality
Setting Description
on The modem function is activated.
off (default) The modem function is deactivated.

IB Modem config

IBA SIM pin


Setting Description
-1 (default) Configure the SIM card PIN.
When set to “-1” no PIN is required for the SIM card.

ÿ It is not possible to change the SIM card's PIN with the MRL-7. Such changes have to be
performed with a different device (eg mobile phone).
Attention: If the SIM pin is set incorrectly, the SIM card will be locked after three
consecutively failing attempts to transfer data.

IBB Net type

Setting Description
auto (default) The network type is selected automatically.
3G Only 3G networks are used to transfer data.
2G Only 2G networks are used to transfer data.

ÿ To allow CSD calls the setting has to be “2G”.

IBC Foreign operator mcc&mnc


Setting Description
“blank” (default) If the field is blank, the setting is deactivated.
Enter the mcc&mnc code for your home network in case your provider
is a virtual one only. Make sure to enter the correct code!

IBD Operator select


Setting Description
auto (default) The network operator is selected automatically.
home Only the SIM card's home network is used.
Prefer home Preferably the set home network operator is chosen by default. If no
connection can be established with the home network, a search for all
available operators is performed and a different operator is selected
automatically.

www.sommer.at page 53
Machine Translated by Google

IBE APN address

Setting Description
“blank” (default) Configure the APN server address

ÿ The APN settings in menu items IBE, IBF and IBG can be set either manually or the settings
are chosen from a predefined list with the software Commander. The list opens by
clicking on “Select APN” in the menu “Special commands” on the left bottom side of the
window.

Fig. 37: APN list

IBF APN user name

Setting Description
“blank” (default) Configure the APN user name

IBG APN password


Setting Description
“blank” (default) Configure the APN password

IBH Custom command 1

Setting Description
“blank” (default) Configure special commands to be sent to the modem (AT).
This feature should be used by modem experts only.
All issued AT commands must be answered with “ok” by the connected
modem.

IBI Custom command 2

Setting Description
“blank” (default) Configure special commands to be sent to the modem (AT).
This feature should be used by modem experts only.
All issued AT commands must be answered with “ok” by the connected
modem.

www.sommer.at page 54
Machine Translated by Google

IC Data transmission 1
The measured and recorded data/camera pictures can be transmitted up to three separate servers.
The data can either be transferred via HTTP or FTP protocol.

ICA Target server type


Setting Description
off (default) No data transferred

http The data is sent to an HTTP server (HTTP settings are used)
ftp The data is sent to an FTP server (FTP settings are used)

ICB Data transmission interval


unit hh:mm:ss hh:mm:ss

Value range 00:01:00 … 23:59:59 00:10:00 (default)

ICC Data transmission offset

The data transmission offset is used to prevent several MRL-7 from sending their data to a server
at the same time by setting different transmission offsets for each data logger. When the
transmission offset is set to “00:00:00” the data will always be transmitted at “full” minutes (when
the transmission interval is also set to “full” minutes, eg “00:05:00”).
unit hh:mm:ss hh:mm:ss

Value range 00:00:00 … 23:59:59 00:01:09 (default)

ICD Content

Setting Description
data (default) To be set for data transmission
camera To be set for camera pictures transmission

ICE FTP server name

Setting Description
“blank” (default) You define the FTP server address / name to which the data are sent.
Host names or IP addresses can be used.

ICF FTP username

Setting Description
“blank” (default) You define the user name for the FTP server account.

ICG FTP password


Setting Description
“blank” (default) You define the password for the FTP server account.

www.sommer.at page 55
Machine Translated by Google

ICH FTP directory

Setting Description
“blank” (default) You define the FTP subdirectory where the data is saved.
When this setting is left empty, the data is saved in the FTP root directory.

ICI FTP port

Setting Description
“21” (default) You define the FTP server port.

ICJ FTP mode

Setting Description
“active” or Defines the FTP server mode (active or passive).
“passive” (default)

ICK HTTP server name

Setting Description
“blank” (default) You define the HTTP server address / name to which the data are sent.
Host names or IP addresses can be used.

ÿ In case a host name is used the network provider has to provide a DNS as standard.

ICL HTTP path

Setting Description
“blank” (default) You define the HTTP request/URL/script via which the data is
saved.

ICM HTTP port

Setting Description
“80” (default) You define the HTTP server port.

ICN HTTP cam chan

Setting Description
“1” (default) Camera channel – it defines the HTTP camera channel for the images of a
camera attached to the logger, which are sent to the SOMMER MDS server
(measured data server). The “cham chan” number has to match the image
channel in the SOMMER MDS.

ÿ The parameter “cam chan” is only applicable for HTTP servers.

ÿ To change this setting, the data transmission content (see ICD chapter) must be set to
“camera”.

ÿ The parameter “cam chan” is independent from the numbers of the measurement values
as in chapter G “Measurements, table”. Data (measurements) and camera images are handled
completely separately by the MRL-7.

www.sommer.at page 56
Machine Translated by Google

ID Data transmission 2
See chapter IC “Data transmission 1” for details.
The only difference is the default value for “Data transmission offset”.

IE Data transmission 3
See chapter IC “Data transmission 1” for details.
The only difference is the default value for “Data transmission offset”.

IF NTP synchronization time


You define the time at which the internal clock is synchronized with the timeserver time.
unit hh:mm:ss hh:mm:ss
Value range 00:00:00 … 23:59:59 02:08:43 (default)

IG NTP server
Setting Description
“blank” (default) Address of the NTP server (timeserver).

IH NTP port
Setting Description
“123” (default) Port of the NTP server (timeserver).

II NTP timezone
With this setting the local time zone can be defined via an offset related to the UTC time.
unit Seconds (sec.) Seconds (sec.)
Value range -43200 … 43200 3600 (default)

IJ Background function
Setting Description
off The MRL-7 is not set to accept remote connections
time window The MRL-7 is set to accept CSD calls (if 2G) in the set time frame (see
chapter IK and chapter IL).
time wi. +IP call The MRL-7 is set to accept CSD calls (if 2G) in the set time frame and
(default) checks periodically for IP call requests.

www.sommer.at page 57
Machine Translated by Google

Setting Description
Time wi. + socket The MRL-7 is set to accept CSD calls (if 2G) in the set time frame and
listens on its IP address with port 4646 – direct connection via internet.

IK Standby, start time


The standby launch time defines the daytime when the MRL-7 starts to be available for remote
communications (CSD calls, IP calls).
unit hh:mm:ss hh:mm:ss
Value range 00:00:00 … 23:59:59 08:00:00 (default)

IL Standby, duration
The standby duration time defines how long the MRL-7 is available for remote communications
(CSD calls, IP calls) and socket.
unit hh:mm:ss hh:mm:ss
Value range 00:00:00 … 23:59:59 01:00:00 (default)

IM IP Call server
Setting Description
IP call server address / name.
“mds.sommer.at” (default)

IN IP Call port
Setting Description
“4647” (default) The port via which the IP call server is reachable.

IO IP Call interval
You define the interval within the set standby time frame (see chapter IK and chapter IL), when
the MRL-7 shall check if there are any IP call requests to the logger.
unit hh:mm:ss hh:mm:ss
Value range 00:01:00 … 00:05:00 00:01:00 (default)

IP Email/SMTP
Settings of the MRL-7 to send out e-mails to be configured by the user according to its individual
e-mail server details.

www.sommer.at page 58
Machine Translated by Google

IPA E-Mail/SMTP

Setting Description
“blank” (default) Name / address of the e-mail server used by the MRL-7 to send out e-mails
(eg mail.samplename.at).

IPB sender

Setting Description
“blank” (default) E-mail address of the MRL-7 as the sender (eg sample@sample.at)

IPC username

Setting Description
“blank” (default) User name to access e-mail server account (if applicable/requested by
individual server settings)

IPD password

Setting Description
“blank” (default) Password to access e-mail server account (if applicable/requested by
individual server settings)

IPE Port

Setting Description
“25” (default) Port number of e-mail server port

J Technics
Technics unit
A Language/Sprache english/english
B Decimal character dot
C Additional settings
D Additional timings
E Com 1 protocol
F com 1
G RS485-2
H SDI-12 bus mastering
I ANALOG setup
J Camera

JA Language/Sprache
Setting Description
“german/deutsch” Select to set the device menu language to German.

www.sommer.at page 59
Machine Translated by Google

Setting Description
“english/english” Select to set the device menu language to English.
(default)

JB Decimal character

Setting Description
“comma” Select to set the “,” as decimal character.
“dot” (default) Select to set the “.” as decimal character.

JC Additional settings

JCA Summer ID
The SOMMER ID is used to define stations with the “Commander” software. For more
information, please see the Commander manual.
By default the SOMMER ID is set to the device's serial number.
SOMMER suggests not to change the SOMMER ID except if a damaged MRL-7 device is
replaced by a new MRL-7. In such a case it can be reasonable to change the SOMMER ID of
the replacement device to the SOMMER ID of the original logger in order to guarantee a trouble-
free archiving.

Setting Description
“serial number” SOMMER ID

JCB Block size, data load


The block size defines how much data is transferred with one communication block (only
applicable when transferring data via the serial interface).
unit Byte Byte
Value range 50…250 250 (default)

JCC Internal low volt. disconnect

The batteries of the MRL-7 are protected from deep discharge. In case the battery voltage
drops below the set limit, the MRL-7 is switched off automatically.
The set limit depends on the used batteries. See the battery data sheet for more information.
Attention An incorrect value in this parameter field can seriously affect the device
operation!
unit V V
Value range 5.0…11.0 9.5 (default)

JCD SommerXF starts with BOM

The BOM (Byte Order Mark) is a special character outputted when the MRL-7 storage is read
out.

www.sommer.at page 60
Machine Translated by Google

Setting Description
“on” (default) BOM is outputted.
“off” BOM is not outputted.

JCE Exposure lock


With this setting the option to display measured values on the MRL-7 device display can be
switched on and off.

Setting Description
“off” (default) Measured values are displayed.
“display” Measured values are displayed only after access code is entered.
“display + data” Measured values are displayed only after access code is entered.
Additionally, downloading data to a connected flash drive requires
entering the access code.

ÿ Access code: To enter the access code, press the key sequence 2x followed by 2x

Attention: As long as the display is active during the ongoing session, the entered access
code unlocks all system settings of the logger.

JCF LCD, Contrast


With this setting the display contrast can be set.

Setting Description
“weak” (default) Lowest contrast setting
"base" Base contrast setting
"medium" Medium contrast setting
“high” Highest contrast setting

JCG LCD, Brightness


The MRL-7 has an integrated ambient light sensor that is used to automatically adjust the display
brightness.
When the measured ambient light is less than specified in this setting the display backlight
brightness is adjusted automatically (three different levels) to a suitable level. When the measured
ambient light exceeds the setting value, the display backlight is switched off completely.

unit lm lm

Value range 0 … 3000 40 (default)

JCH Windspeed measurement duration


This value defines the period of time to measure the wind speed.

unit ms ms

Value range 500…2000 1000 (default)

www.sommer.at page 61
Machine Translated by Google

JD Additional timings

JDA Sum, reset time


You define the time at which accumulated sums are reset to “0”.

Attention: This applies to all measurement values using the “Sum” function.
unit hh:mm:ss hh:mm:ss

Value range 00:00:00 … 23:59:59 07:00:00 (default)

JDB Occasional storage interval


You define the storage interval of “special” measurement values.

This applies to all “SYS” measurement values.


unit hh:mm:ss hh:mm:ss

Value range 00:00:00 … 23:59:59 00:01:00 (default)

JDC Sync date (last)


Every time the device date is updated manually (via Commander) the current date is
automatically saved for the purpose of documentation.
unit YYYY-MM-DD YYYY-MM-DD

Value range 2000-01-01 … 2199-12-12 production date (default)

JDD Sync time (last)


Every time the device time is updated manually (via Commander) the current time is
automatically saved for the purpose of documentation.
unit hh:mm:ss hh:mm:ss

Value range 00:00:00 … 23:59:59 production time (default)

JDE min. Hold On message


The period of time set in this menu defines how long a measurement value's trigger condition
for the message has to be unfulfilled (the limit value has to be exceeded/deceeded) before the
trigger condition status ends again.
Example: “min. Hold On message” setting = 5 ÿ the trigger condition has to be exceeded/deceeded
for five consecutive minutes

unit Minutes (min.) Minutes (min.)


Value range 0…180 5 (default)

JDF max. Hold On message


By using this option, the MRL-7 can be configured to send multiple messages in a user defined
interval as long as a measurement value's trigger condition is still fulfilled.
Example: “max. Hold On message” setting = 120 ÿ a further message is sent after 120 minutes
from the beginning of the trigger condition.

www.sommer.at page 62
Machine Translated by Google

unit Minutes (min.) Minutes (min.)


Value range 0…120 0 (default)

ÿ Setting “0” minutes equals to deactivate this function.

JE Com 1 protocol
This submenu defines the protocol settings for the RS-232 interface of the MRL-7.

Com 1 protocol unit


Device number 0
B System key 0
C Output protocol (OP) type SOMMER
D Output (MO) time just per command
E MO wake-up sequence prefix
F MO prefix holdback 300 ms

JEA Device number

Device number assigned to the MRL-7 in the RS-232 bus.

Value range 0…98 0 (default)

JEB System key


System key assigned to the MRL-7 in the RS-232 bus.

Value range 0…99 0 (default)

JEC Output protocol (OP) type


Two different measurement output protocols are available for the RS-232 port.

Setting Description
SOMMER SOMMER Bus Protocol
(default)
standard Standard protocol

JED Output (MO) time


By default, the measurement results are outputted via the RS-232 interface and are available to
be retrieved via SDI-12 commands after a measurement was completed. However, it is possible
to configure the MRL-7 to output the results only when a command is issued or when a
measurement result is saved.

Setting Description
just per command Measurement results are only outputted when requested with
(default) commands (RS-232 or SDI-12).
measured Measurement results are outputted automatically after a measurement
automatic was completed.

www.sommer.at page 63
Machine Translated by Google

Setting Description
storage automatic Measurement results are outputted after a measurement result was
saved by the MRL-7.

JEE MO wake-up sequences

In case a connected device records the MRL-7 output strings it might be required for this device to receive
a sync, prefix or both before the measured values can be recorded.

Setting Description
off No wake-up sequence is used.
sync A sync is issued prior to the output string.
prefix (default) A prefix is issued prior to the output string.
prefix & sync A prefix and sync are issued prior to the output string.

JEF MO prefix holdback

unit ms ms

Value range 0 … 5000 300 (default)

JF Com 1
This submenu defines the connection settings for the RS-232 interface.

Com 1 protocol unit


Baudrate _ 115200

B Minimum response time 10 ms


C Flow control RTS-CTS

JFA Baudrate

Baud rate for the RS-232 interface.

Value range 1200 … 115200 115200 (default)

JFB Minimum response time


unit ms ms

Value range 0 … 2000 10 (default)

JFC Flow control

Setting Description
off No flow control is activated
XON-XOFF XON-XOFF blocking is active
RTS-CTS (default) RTS-CTS is active

www.sommer.at page 64
Machine Translated by Google

JG RS485-2
This submenu defines the connection settings for the RS-485 interface.

RS-485-2 unit
Baudrate _ 9600

B Parity, stop bits no stop, 1 stop


C Minimum response time 0 ms

D Transmitter warm-up time 10 ms


E Flow control XOFF-XON blocking
F Sending window 300 ms

G Receiving window 300 ms

H Trig, polling on
Yo
Trig, timeout 70 sec

J Trig, sleep while timeout on


K Network scan extension on

L Transparency to RS485-2

Attention: The settings of the MRL-7 must correspond with the settings of the connected sensor(s)
to guarantee smooth operation.

JGA Baudrate

unit bd bd

Value range 1200 … 115200 9600 (default)

JGB Parity, stop bits

Setting Description
no stop, 1 stop
(default)
no stop, 2 stop
even even, 1 stop
odd even, 1 stop

JGC Minimum response time


unit ms ms

Value range 0 … 2000 0 (default)

JGD Transmitter warm-up time


unit ms ms

Value range 0 … 2000 10 (default)

www.sommer.at page 65
Machine Translated by Google

JGE Flow control

Setting Description
Off Flow control is deactivated.
XOFF-XON Flow control is activated.
blocking (default)

JGF Sending window


You define the time span during which commands are sent via the RS485-2 interface.
unit ms ms
Value range 50…5000 300 (default)

JGG Receiving window


You define the time span during which commands are received via the RS485-2 interface.
unit ms ms
Value range 50…5000 300 (default)

JGH Trig, polling


Setting Description
off Continuous polling of measurement values – mode is deactivated.
on (default) Continuous polling of measurement values – mode is activated.

JGI Trig, timeout


Defines the time, the MRL-7 is waiting until expected commands/answers are received via the
RS485-2 interface.

unit sec sec


Value range 3…250 70 (default)

JGJ Trig, sleep while timeout


To save on power consumption the MRL-7 can switch off the RS-485 interface until a
measurement result is available. If set to “on” the connected sensor (eg RQ-30) must send a
“Sync” command to reactivate the RS-485 interface so that it can record the output string sent by
the sensor.

Setting Description
off
on (default)

JGK Network scan extension

When activated the network scan extension enables detecting connected SOMMER sensors (eg
RQ-30) with Commander. The parameters of detected SOMMER devices can be parameterized
with Commander.

Setting Description
off Detecting SOMMER devices connected via RS485-2 is deactivated.

www.sommer.at page 66
Machine Translated by Google

Setting Description
on (default) Detecting SOMMER devices connected via RS485-2 is activated.

JGL Transparency to RS485-2


This menu item is available only via the terminal menu (see 9.1 Changing parameters via a
terminal).
When this menu item is selected commands and responses can be sent and received directly
via the RS485-2 interface of the MRL-7.
This feature can be useful to adjust the device settings of a connected RS-485 sensor.

JH SDI-12 bus mastering


SDI-12 bus mastering unit
To Max. measurement duration twenty sec
B Sensor search
C Change sensor address
D Ask for a sensor address

JHA Max. measurement duration

Defines the maximum time available for requesting and receiving data from SDI-12 sensors. If
no response is received within this period of time, the logger stores an “empty” value instead of
the measured value.

unit sec. sec.


Value range 0 … 250 20 (default)

JHB Sensor search

This menu item is available via the terminal menu only. Please see chapter 9.1 for more
information.
When selected a search for all available SDI-12 sensors connected to the bus is performed and
the result is outputted (SDI-12 command “?!”)

JHC Change sensor address


This menu item is available via the terminal menu only. Please see chapter 9.1 for more
information.
When selected the address of a sensor in the SDI-12 bus can be changed (SDI-12 command
“aAb!”).

JHD Ask for a sensor address

This menu item is available via the terminal menu only. Please see chapter 9.1 for more
information.

JI ANALOG setup

ANALOG setup unit

www.sommer.at page 67
Machine Translated by Google

At 5V sensor supply off


B 12V sensor supply off
C Extended supply off
D Switch usage off
E Warm-up time 0 sec
F ADC - conv. rate 62Hz
G ADC filter off
H AN3 (N) type voltage 2.5
I AN4 (D) type voltage 2.5

JIA 5V sensor supply


With this setting the supply voltage provided at the analog inputs can be set.

Setting Description
“off” (default) No sensor supply voltage is outputted.
“switched” The sensor supply voltage is set to 5V and is outputted during
measurements only (max. 100mA).
“always on” The sensor supply voltage is set to 5V and is outputted at all times
(max. 100mA).

JIB 12V sensor supply


With this setting the supply voltage provided at the analog inputs can be set.

Setting Description
“off” (default) No sensor supply voltage is outputted.
“switched” The sensor supply voltage is set to 12V and is outputted during
measurements only (max. 200mA).
“always on” The sensor supply voltage is set to 12V and is outputted at all times
(max. 200mA).

JIC Extended supply


With this setting the extended supply voltage/current can be set.
This voltage/current is outputted when an analog measurement is performed. The set warm-up
time (see JIE) applies.

Setting Description
“off” (default)
“2.5V” 2.5V are outputted via the “2.5V Ref Out” pin (see 3.5 Pin description)
“0.5mA” 0.5mA are outputted via the “CurrH” pin (see 3.5 Pin description)
“2.5V + 0.5mA” Both aforementioned reference values are outputted

www.sommer.at page 68
Machine Translated by Google

JID Switch usage


With this setting connecting the sensor supply voltage of the MRL-7 through via the switches (SW1,
SW 2 and SW3) can be de/activated (see also chapter 3.5).

Setting Description
“off” (default) The sensor supply voltage is not connected through the “Switch Out”
pin.
"1" The sensor supply voltage is outputted via the SW1 (switch 1 pin)
during active measurements.
"2" The sensor supply voltage is outputted via the SW2 (switch 2 pin)
during active measurements.
"3" The sensor supply voltage is outputted via the SW3 (switch 3 pin)
during active measurements.

JIE Warm-up time


The “Warm-up time” is the period of time required to receive valid measurement results from
connected sensors.

unit sec. sec.

Value range 0 … 250 0 (default)

Example: A sensor connected to one of the analogue inputs needs to be powered on for 10s until
correct measurement values are outputted. The analog sensor's power is supplied via the 12V
sensor supply line from the MRL-7.
The “Warm-up time” is therefore set to (minimum) 10s.

JIF ADC – conv. rate

You define the sampling rate of the analogue inputs.


unit Hz Hz

Value range 2…250 62 (default)

JIG ADC filter

By default the ADC filter is set to “off” so one measurements per analogue input is performed and
outputted.
When set to any of the three remaining options three measurements of all analogue inputs are
performed in sequence. Depending on the ADC filter setting, the results are filtered and a single
value is outputted/stored.
ÿ When any of these options is selected please consider that getting measurement results will take
longer. The measurement interval/frequency is defined via the “ADC – conv. rate” (see
chapter JIF).

Setting Description
“off” (default) One measurement is performed and no filter is applied.
“minimum of 3” Three measurements of the analog inputs are performed and the lowest
measurement result is outputted.
“medium of 3” Three measurements of the analog inputs are performed and the
medium value of those measurement results is outputted.

www.sommer.at page 69
Machine Translated by Google

Setting Description
“mean of 3” Three measurements of the analog inputs are performed and the
mean value of those measurement results is outputted.

JIH An3 (N) type


Analogue input AN3 can be configured to measure resistors greater than 2000 ohm or a
connected NTC (resistor with Negative Temperature Coefficient).

Setting Description
“voltage Standard analogue input: Measures voltages of 0V … 2.5V.
2.5” (default)
“NTC” An NTC is measured and the resistance is outputted.
“R meas > 2k” A resistor greater than 2000 ohm is measured and the resistance is
outputted.

JII An4 (D) type


Analogue input 4 can be used for a set of predefined applications which are selected with this
setting.

Setting Description
“voltage 2.5” Standard analogue input: Measures voltages of 0V … 2.5V.
(default)
“voltage 0.3” High-resolution voltage measurement for up to 0.3V.
“R meas > 1k2”” A resistor with more than 1200 ohm is measured and the resistance is
outputted.
“R means < 1k2” A resistor with less than 1200 ohm is measured and the resistance is
outputted.
“PT1000” Preconfigured for connecting a PT1000 temperature sensor.

JJ Camera
A camera (web cam) is connected via the RS-485 interface. This chapter describes the available
camera parameter settings.
camera unit
to switch off
B Warm-up time 30 sec
C Night recognition on
D View

JJA switch

Setting Description
“off” (default) To be set, if the camera does not need a power supply by the MRL-7.
"1" Camera is connected to switching output 1.
"2" Camera is connected to switching output 2.

www.sommer.at page 70
Machine Translated by Google

Setting Description
"3" Camera is connected to switching output 3.

JJB Warm-up time


The time, before the camera is able to take a picture.
unit sec sec
Value range 0…120 30 (default)

ÿ The warm up time can only be set if the camera is supplied to a switching output.

JJC Night recognition


To safe battery, the MRL-7 can control the camera to prevent it from taking pictures in the
darkness (eg during the night).

Setting Description
“off” Function is deactivated. The camera can always take pictures.
“on” (default) Function is active. The camera does not take pictures during the night.

ÿ Only applicable if the MRL-7 is power supplied by solar panel.

JJD View

If this button is pressed, the camera is activated and an image (in JPG file format) is displayed
in the Commander. (In the terminal mode the image is shown as binary data code.) The image
is also stored automatically at the default location of downloaded data files (C:
\Users\Public\Documents\Sommer\Data\).

K Special functions
The special functions are either available by pressing the provided button (eg item A, B, H and
I) or via the terminal menu.
ÿ See chapter 9.1 “Changing parameters via a terminal” for more information.

Special features
Device status
BView setup
C Continuous measurement mode (tempo.)
D Inspection, nominal values
E Inspection, start process
F Set factory default
G Temp. load factory default
H Relaunch program
I Replace program

www.sommer.at page 71
Machine Translated by Google

KA Device status
Outputs the current device status.
Example of device status output:

Device status

Device: MRL-7
Software version: 1_19r00
Miu-OS version 4_06r00
Setup version 2_06r04
Serial number 00000000
System advice codes: 0x0000 0x0000 0x0000
0x0000
Date (YYYY,MM,DD) 2017-04-03
time 10:33:47
Modem, IMSI No.: 232018831215087
Modem, signal qual.: 32.0 2G
IP address:

KB View setup
Selecting this menu item causes a list of all device settings to be outputted. The list can be
recorded with Commander. For more information, please see the Commander manual.

KC Continuous measurement mode (tempo.)


This special measurement mode was implemented to recognize changes in measurement
values easier and more quickly when testing a setup at a measurement site. The continuous
measurement can also be activated via the MRL-7 keyboard (see chapter 5.6 for more
information).
Selecting this menu item activates a continuous measurement mode. When enabled
measurements are performed continuously in the shortest possible interval. In addition, the
sensor supply voltages are always outputted and saving asynchronous measurement values
(eg observer input etc.) is disabled.
ÿ The continuous measurement mode is disabled automatically after 3 minutes.

KD Inspection, nominal values


List of internal MRL-7 nominal values that are used for an automatic testing during the production
process of the logger.

KE Inspection, start process


Starts an automatic testing of the internally generated voltages and of the analogue inputs.

www.sommer.at page 72
Machine Translated by Google

To be able to perform this testing a specific testing-adapter has to be connected to the MRL-7.
This testing-adapter can be purchased from SOMMER Messtechnik at extra costs.

KF Set factory default


Menu item to set all MRL-7 device parameters to their default values (requires user confirmation).

KG Temp. load factory default


Menu item to temporarily set all MRL-7 device parameters to their default values. When the
MRL-7 is restarted or the terminal menu exited and then reopened the previously set
parameters are restored.
ÿ This feature is meant to be used as a reference of all default values in case no other
documentation (eg this manual) is available at a measurement site.

KH Relaunch program
With this menu item the MRL-7 firmware can be restarted. The boot process is the same as if
the device's supply was removed and reconnected.

KI Replace program
With this menu item the MRL-7 firmware can be restarted. The boot process is customized to
better support performing remote firmware upgrades (boot loader time extended).

www.sommer.at page 73
Machine Translated by Google

8. Technical specifications
Technical data are subject to change

Table 8: Technical specifications

Inputs 4 x analogue inputs 0...2.5 V (one differential input)


2 x Counter inputs
1 x Wind speed
1 x Wind direction (same as counter 2)
1 x RS-485 (SOMMER-Bus protocol)
1 x SDI-12 (MRL-7 as master)
1 x Observer input
Outputs 1 x RS-232: 1200...115200 Baud (ASCII protocol)
1 x USB
1 x Bluetooth
Modem 2G/3G (optionally 4G)
Total of 3 FTP/HTTP servers

Data retrieval and parameterization


Functions: IP call, fixed IP, time-synchronization
power supply Internal battery (MRL-7B only): 12 V / 4 Ah
Deep discharge protection for internal batteries
Supply voltage: 6 to 30 VDC
Reverse voltage protection
Overvoltage protection
Solar panel supply: max. 40W/12V
*
Energy consumption Standby: < 0.6 mA
Normal: < 23.0 mA

Area of application Operating temperature: -30°C to +60°C


Storage temperature: -40°C to +60°C
Storage Humidity: 10 to 90% rH
Housing Material: Anodized aluminum

Dimensions (W x H x D):
MRL-7: 180 x 150 x 60mm
MRL-7B: 180 x 150 x 90mm

Installation: four mounting holes


Protection type IP67

*
…Energy consumption measured with a supply voltage of 12 VDC. Sensors supplied via the
MRL-7 can increase power consumption.

www.sommer.at page 74
Machine Translated by Google

9. Appendix
9.1. Changing parameters via terminal
The main menu is opened by entering three question marks ??? in the terminal program.
The terminal menu is available via all connection types (serial connection, IP call, Bluetooth...).

Attention: Do not enter a “carriage return” after inputting ???.

Fig. 38: Main menu in terminal mode

The menu items are accessed by entering the menu key displayed left of the menu item. Either submenus are
opened or the specific parameter is displayed with its corresponding unit.
Changes are verified with Enter, editing is aborted with Esc. Sub menus are closed with X.
The menu is not case sensitive. When the main menu is closed, the sensor starts the measurement mode and
returns the message “Init done!”.

9.2. Error messages


PLEASE NOTE: Parameterconflict (view manual)!

This message is outputted when

ÿ …for example the parameter “Switch usage” in the menu “ANALOG setup” (see chapter JID) and the
parameter “switch” in the menu “Camera” (see chapter JJA) are set to the same switch number (eg “1 ").

When the above configuration example is identified, the parameter changed last will be set as defined by
the user and the conflicting parameter is automatically set to “off”.

www.sommer.at page 75
Machine Translated by Google

Reason: It is not possible to use the same switch number for two different parameters (purposes).

Switches can be selected (parameterised) at the different menu chapters “H Messages, table (when
selected message type “switch”), “JID Switch usage” for the analogue setup and “JJA switch” as of
the camera settings.

ÿ …the set “Measurement interval” (see chapter C) is shorter than the set “Trig, timeout” (see chapter
JGI) and/or the “Max. measurement duration” (see chapter JHA) and/or the “Warm-up time” (see
chapter JIE).
The measurement interval is automatically adjusted to the smallest possible value
taking into account the other mentioned parameters.
Reason: The setting of (the shortest possible) measurement and/or (!) storage interval is defined by
the measuring process of all sensors connected (via analogue, RS-485 and SDI-12 interfaces) and
their measurement duration needed to perform a measurement. Therefor the measurement (and/or
storage) interval can never be shorter than the aforementioned intervals of one or multiple sensors
(see also chapter “2.2 Measurement and data storage concept”).

Example: The supply voltage of a connected analogue sensor must be switched on 60 seconds
before a correct measurement result is outputted. The warm-up time is therefore set to 60 seconds.
The MRL-7 measurement interval was previously set to 20 seconds.

ÿ When the (changed) parameters are transferred (uploaded / synchronized) to the MRL-7, the
error message is outputted and the measurement interval is automatically set to 60 seconds.

DIRECTIVE: Please don't forget to change serial counterpart too!

This message is outputted when the baud rate set via parameter JFA Baudrate is changed.
Please note the new setting and adjust the previously defined “Connection” (see 6.2.1 Defining
a local connection with the MRL-7) accordingly or create a new “Connection”.

PLEASE NOTE: Maybe difficult interpretation of archive data!

This message is outputted when the „Identifier“ (see chapter G Measurements, table) and/or the unit of more
than one defined measurement is changed.
ÿ The reason for outputting this message is that because of such changes the ordering of measurement
values can change. This makes interpreting the archived data more difficult.
There are three options available to the user when this message is outputted:

ÿ Accepting the warning because the changes made have desired effects only (eg
correcting the “Identifier” of a measurement value).

ÿ Creating a new station (or changing the current station number) to make sure that the changes made
have no negative impact on previously created archives.
If this option is chosen SOMMER suggests deleting all data stored on the MRL-7 to prevent
previously archived data from being added to a new (and additional) archive.

ÿ Deleting the „old“ archived data (or shifting it to a new file storage location) locally on the used
computer and also erasing all data stored on the MRL-7 to reset the device to a “new” station.

PLEASE NOTE: An old archive pointer has been replaced!

www.sommer.at page 76
Machine Translated by Google

This message is outputted only when more than 16 different computers were used to archive data
collected with the currently connected MRL-7. In almost all cases this message can be ignored.

To keep the archive system “clean” SOMMER suggests appointing someone who is mainly responsible
for the station(s). This person can synchronize all previously used archives and then delete all archive
pointers (in Commander select “Extra”/”Data”/”Delete all data pointers).

Mistake: Please just enter valid values!

This message is outputted when characters are entered (eg in terminal mode) that are not allowed / not
applicable for the actual parameter or setting (eg a letter is entered although only numbers are possible).

Mistake: Please, just enter menu choice characters!

This message is outputted when characters are entered (eg in terminal mode) that are not part of the
contained menu choice of characters (eg the number “3” is entered although only the numbers “1” and
“2” are able to be entered as part of the menu choice).

9.3. Index of tables


Table 1: SDI-12 example commands and answers.............................................. ............................. 4
Table 2: Display menu structure................................................ .................................................. ...6
Table 3: Interface types and features ........................................... ................................................ 7
Table 4: Analogue inputs................................................ .................................................. ............. 7
Table 5: Supply voltage outputs ........................................... .................................................. ...8
Table 6: MRL-7 pin assignment................................................ .................................................. .... 19
Table 7: ***** Settings *****.................................... .................................................. .............. 27

Table 8: Technical specifications................................................ .................................................. .. 74

9.4. Index of figures


Fig. 1: MRL-7 key functions and features ....................................... ............................................ 2
Fig. 2: Analogue measurement.............................................. .................................................. ..... 3
Fig. 3: RS-485 measurement................................................ .................................................. .......... 4
Fig. 4: SDI-12 measurement .......................................... .................................................. .......... 4
Fig. 5: Example storage interval ........................................... .................................................. .... 5
Fig. 6: SDI-12 example configuration................................................ ................................................... 9
Fig. 7: MRL-7 ........................................... .................................................. ................................ 10
Fig. 8: MRL-7B................................................ .................................................. ............................ 10
Fig. 9: MRL-7 housing, dimensions (in mm) ................................... ......................................... eleven
Fig. 10: MRL-7B housing, dimensions (in mm).................................... ..................................... eleven
Fig. 11: MRL-7 covering strips................................................ .................................................. ....... 12

www.sommer.at page 77
Machine Translated by Google

Fig. 12: MRL-7 screws ........................................... .................................................. ................. 12


Fig. 13: Opening the housing of the MRL-7 ................... .................................................. ........ 13
Fig. 14: Holes for Mounting ........................................... .................................................. ......... 13
Fig. 15: Lid of the MRL-7 from inside ............................ .................................................. ......... 14
Fig. 16: MRL-7 pin description ............................ .................................................. ................ 18
Fig. 17: MRL-7 USB Connector ........................................... .................................................. ..... 23
Fig. 18: MRL-7 keyboard................................... .................................................. .................... 25 Fig.
19: Setting the “Expert” mode in the software “Commander”............ ................................ 29
Fig. 20: USB/RS-232 converter ....................................... .................................................. ...... 30
Fig. 21: Connecting the USB/RS-232 converter .............................. ........................................ 30
Fig. 22: Download Parameters ... .................................................. ............................................ 39
Fig. 23: List of parameters ........................................... .................................................. .......... 40
Fig. 24: New Station ................................. .................................................. ................................ 40
Fig. 25: Station no. and SOMMER ID ................................................. ........................................ 41
Fig. 26: Station no. and SOMMER ID ................................................. ........................................ 41
Fig. 27: Assigning connections to a station .................................................. ............................ 41
Fig. 28: Connection details ............. .................................................. .................................... 42
Fig. 29: Data archive ........................................... .................................................. .................. 42
Fig. 30: Saving a station................... .................................................. .................................. 42 Fig.
31: Connecting to a station ........ .................................................. .................................... 43
Fig. 32: Downloading stored data ........................................... .................................................. 43
Fig. 33: Downloading stored data manually................................................ ....................................... 44
Fig. 34: IMSI number ........................................... .................................................. ................. 44 Fig.
35: Station no. and SOMMER ID ................................................. ........................................ 45 Fig.
36: IP call connection .. .................................................. .................................................. .. Four. Five
Fig. 37: APN list ........................................... .................................................. ......................... 54 Fig.
38: Main menu in terminal mode ............... .................................................. .................... 75

Thanks for using a product made by


SOMMER Messtechnik GmbH.
www.sommer.at

Distributed exclusively in the USA


by www.hydrologicalusa.com

www.sommer.at page 78

You might also like