Neulog PH Meter Sensor Guide
Neulog PH Meter Sensor Guide
Prior to initial use the Neulog pH Sensor needs to be calibrated, please review the Tablet, smart phone device
“Sensor Calibration” section of this document starting on page 2 for detailed
instructions. Connect the pH Sensor to a USB-200 Module. The USB-200 Module Materials needed:
then connects to a computer via USB to mini USB cable. Please note that you cannot
NUL-206 PH Sensor
plug the pH probe directly to the computer. WIFI-201 WIFI module
BAT-200 Battery
Resident PC software or browser base software can be downloaded for free at
www.neulog.com/dowload as well as a full software user guide. Make sure not to Prior to initial use the Neulog pH Sensor needs to be calibrated, please review the
download and install both types of software as they will conflict on the computer. “Sensor Calibration” section of this document for detailed instructions. Your pH
Sensor needs to be connected to a WIFI-201 Module. The WIFI-201 Module will
Procedure: create a closed Neulog wifi network which will stream the Neulog data to a device of
your choosing. Once your device is wirelessly connected to the Neulog network you
1. Remove pH Sensor probe from the plastic storage solution bottle. can run experiments and collect data through a browser of your choosing.
2. Gently rinse tip of pH Sensor’s probe thoroughly with distilled water. Be
careful not to physically come in contact with the glass bulb. Procedure:
3. Calibrate the pH Sensor. (Follow the ‘Sensor Calibration” guide in this
document) 1. Remove pH Sensor probe from the plastic storage solution bottle.
4. Install the Neulog software 2. Rinse tip of pH Sensor’s probe thoroughly with distilled water. Be careful
5. Connect the USB-200 Module to the PC or Mac not to physically come in contact with the glass bulb.
6. Connect the pH Sensor to the USB-200 Module (they directly plug together). 3. Calibrate the pH Sensor. (Follow the ‘Sensor Calibration” guide in this
7. Open the Neulog software. document)
8. Once a pH Sensor logo appears on the left side of the screen the probe has 4. Connect the pH Sensor directly to the left side WIFI-201 (no wires required).
been automatically identified and you can begin experimentation. 5. Connect BAT-200 to the right side of the WIFI-201
9. If the pH Sensor is not automatically identified then click the search sensors 6. Although not required, we recommend plugging the BAT-200 to an outlet
icon to find the sensor. using a USB to micro USB charger (such as a typical cell phone charger).
10. Select the “On-line experiment”, this will open a graph below. The wifi201 will run for X minutes (depending on the sensor) without being
11. Click on the “Module setup” button located on the pH Sensor icon in the plugged in.
Module window to change the pH probe settings if need be. 7. For further WIFI-201 instructions or the WIFI-201 quick start guide please
12. Click on the experiment set up button to change the experiment settings if visit www.Neulog.com/download.
need be (Experiment duration for example). 8. The WIFI-201 indicator lights will flash; take no action until the LED to the
13. The pH Sensor will give a live reading in the box to the left of the screen far left turns blue.
when running. 9. Take your tablet or smart phone and go to Wi-Fi settings and select the
14. To run an experiment and collect data click “Run experiment”. Neulog network which matches the WIFI Module ID found on the back of the
15. To end data collection early, click “Stop experiment” WIFI-201 device. (NeuLogXXXX for example)
10. Give your device 1-2 minutes to connect to the WIFI-201.
NEULOG PH LOGGER SENSOR GUIDE
11. Once the device is connected go to your browser and type the website 6. The VIEW-101 will auto detect the sensor and you will see it appear on the
wifi201.com into the URL, then wait for 30-60 seconds. left side of the screen.
12. You will see a “Control mode” icon in the browser, click on this icon. 7. Once the sensor appears it will be monitoring data in real time.
13. The browser will then load a new screen and begin to auto detect the 8. To run an experiment and collect data click the run button (little green
sensors, this can take a few minutes. person)
14. If the browser does not auto detect, select “Search for Sensors”
15. Once the sensor is found you will see an icon on the left side of the screen Off-line experiments
for the pH Sensor. The icon will display real time data.
(Off-line experiments are for when you do not have a sensor connected directly to a
16. To set the pH Probe settings click on the “Module setup” button located on
computer, tablet, smartphone, or NeuLog viewer).
the pH Sensor icon on the left side of your screen (Graph settings for
example.) Materials needed:
17. To change the experimental settings click on the “Experiment settings”
button (Experiment duration for example). NUL-206 PH Sensor
18. To run an experiment and collect data select “On-Line” experiment, then BAT-200 Battery
select run.
Materials needed to configure your offline experiment:
Operation with NeuLog viewer
USB-200 USB Module or WIFI-201 WIFI module or VIEW-101 Viewing Module
A USB to mini USB cable (which comes with the USB-200)
Prior to initial use the Neulog pH Sensor needs to be calibrated, please review the
“Sensor Calibration” section of this document for detailed instructions. Procedure:
Materials Needed: 1. Connect the pH sensor directly to the left side of a charged BAT-200
module.
NUL-206 pH Sensor 2. When ready to collect data press the “Start/Stop” button on the pH sensor, a
VIEW-101 Viewing Module red light will turn on for the duration of the experiment.
BAT-200 Battery 3. To change the experiment settings first connect to your chosen device (PC,
Mac, tablet, smart device) and select the “Module setup” button. (For more
Procedure: information on how to change the experiment settings view quick start
procedure section for your chosen device in this document.)
1. Remove pH Sensor probe from the plastic storage solution bottle. 4. After the experiment has concluded (when the “Start/Stop” button is pressed
2. Rinse tip of pH Sensor’s probe thoroughly with distilled water. Be careful again or the data collection period ends) connect the sensor to your chosen
not to physically come in contact with the glass bulb. device. (full instructions in each device’s quick start procedure section)
3. Calibrate the pH Sensor. (Follow the ‘Sensor Calibration” guide in this 5. Open NeuLog software.
document) 6. Click the “Off-line experiment” button.
4. Connect the pH Sensor to the left side of the viewer. 7. Click the “Load data from sensors” button.
5. Connect the BAT-200 to the right side of the viewer. 8. Select which experimental data to upload. (5 experimental runs can be
stored at one time on the NUL-206 pH sensor)
NEULOG PH LOGGER SENSOR GUIDE
The philosophy behind Neulog’s plug and play technology is based on each sensor’s
ability to store its own data due to an internal flash memory chip and micro-controller
in each plastic Neulog body. This technology allows the sensor to collect and then
store the digital data in the correct scientific units (oC, oF, Lux, %, ppm, for example).
Flexible, simple, fast, forward thinking.
Building the conversion formula of the pH sensor is done by measuring at least 3 W: www.neulog.com
E: info@neulog.com
different pH level buffers. The sensor is pre-calibrated at the factory. The sensor
A: 850 St Paul Street, Suite 15, Rochester, NY 14605
needs to be calibrated to pH 7 prior to initial use. Please review the “Sensor P: 1.866.553.8536
Calibration” guide in the beginning of this document for detailed instructions.
The built-in software in the logger can be upgraded at any time using software.
The pH electrode houses a sensitive glass bulb which detects the H+ concentration in
the fluid being tested which is then translated to a voltage reading and passed along a
V04302013
silver-chloride wire. The voltage reading is compared with the second component, the
reference electrode to compare voltage differences. The voltage reading then is
calculated and translated back into pH.
Warranty:
We promise to deliver our sensor free of defects in materials and workmanship for a
period of 3 years from the date of purchase. Our warranty does not cover damage of
the product caused by improper use, abuse, or incorrect storage. Sensors with a shelf
life such as ion selective proves have a warranty of 1 year. Should you need to act
upon the warranty please contact your distributor. Your sensor will be repaired or
replaced.