Professional Documents
Culture Documents
Spectroscopies Atomic Absorption/ Emission
Spectroscopies Atomic Absorption/ Emission
spectroscopies
Fundamentals
The population of different states is
given by the Boltzmann equation:
ΔE
N1 g1
e kT
N0 g0
N0: number of atoms in ground state
N1: number of atoms in excited state
g1/g0 : weighting factors
3
Atomic emission:
Flame spectroscopy
Observation Caused by...
Persistent golden- Sodium Lithium
yellow flame
Violet (lilac) flame Potassium,
cesium
carmine-red flame Lithium
Brick-red flame Calcium
Crimson flame Strontium Cesium
Fluoresc- AAS
AES Others
ence ICP-MS SIMS
Spectros-
copy
Others ICP-OES
Flame AAS GFAAS
5
Atomic spectroscopies
Common principles:
Sample introduction:
Nebulisation, Evaporation
Atomisation (and excitation or
ionisation) by flame, furnace,
or plasma
Spectrometer components:
Light source
Sample
Wavelength selctor
Detector
6
Atomic spectra vs molecular spectra:
Lines Bands
(nm)
LUMO
Vibrational substates rotational substates
HOMO
8
Instrument components
AAS Spectrometer
Read-
Light Monochro-
Sample Detector out/Data
Source mator
system
ICP-OES Spectrometer
Sample = Read-
Monochro-
light Detector out/Data
mator
source system
UV-Vis Spectrometer:
Read-
Light Monochro-
Sample Detector out/Data
Source mator
system
9
Light sources
Wavelength(nm) 100 200 400 700 2000 4000 7000 10,000 20,000 40,000
Ar lamp
Xe lamp
Continuum
D2 lamp
Tungsten lamp
Nernst glower (ZrO2 + Y2O3)
Nichrome wire
Globar (SiC)
Line
12
Monochromators
Consist of
Entrance slit
Collimating lens or mirror
Dispersion element (prism or grating)
Focusing lens or mirror
Exit slit
Czerny-Turner grating monochromator:
Mirrors
13
Bandwidth of a monochromator
14
Effect of slit width on peak heights
15
Detectors
Wavelength(nm) 100 200 400 700 2000 4000 7000 10,000 20,000 40,000
Photographic plate
Photomultiplier
Phototube
Photon
Photocell
detectors
Silicon diode
Charge-coupled device (170-1000)
Photoconductor
Thermal Thermocouple
detectors Golay pneumatic cell
Pyroelectric cell
16
Photomultiplier: detects one
wavelength at a time
Based on photoelectric
effect
Photocathode and
series of dynodes in an
evacuated glass
enclosure
Sample preparation
Interferences
Calibration
19
Crucial steps in atomic spectroscopies
and other methods
Laser ablation etc.
M+ Excitation
Ionisation
Ions
Excited
ICP-MS and Atoms
other MS methods
AAS and AES,
X-ray methods
Adapted from www.spectroscopynow.com (Gary Hieftje) 20
Sample Introduction: liquid samples
Often the largest source of noise
Sample is carried into flame or plasma as aerosol,
vapour or fine powder
Liquid samples introduced using nebuliser
21
Sample preparation for analysis in
solution: Digestion
22
Microwave digestion
Rotor
24
Atomic absorption
spectroscopy
25
Atomic Absorption Spectroscopy
Flame AAS has been the most widely used of all atomic
methods due to its simplicity, effectiveness and low cost
First introduced in 1955, commercially available since
1959
Qualitative and quantitative analysis of >70 elements
Quantitative: Can detect ppm, ppb or even less
H He
Li Be B C N O F Ne
Na Mg Al Si P S Cl Ar
K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
Rb Sr Y Zr Nb Mb Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe
Cs Ba La Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
Fr Ra Ac
26
Flame AA Spectrometer
Hollow cathode lamps with
characteristic emissions
Burner
I0 It
Light Source Monochromator Detector Amplifier
E.g. Hollow
Fuel (e.g. acetylene)
cathode lamp Atomiser Air
29
from: Skoog
Structure of a flame
Relative size of
regions varies with
fuel, oxidant and
their ratio
30
Electrothermal atomisation: GFAAS
Provides enhanced sensitivity
entire sample atomised in very
short time
atoms in optical path for a second
or more
31
Sensitivity and detection limits in
AAS
Sensitivity: number of ppm of an element to give 1%
absorption.
Limit of detection: dependent upon signal:noise ratio:
S/N=3.2
32
Interferences in AAS
Broadening of a spectral line, which can occur due to
a number of factors (Physical)
Spectral: emission line of another element or
compound, or general background radiation from the
flame, solvent, or analytical sample
Background correction can be applied
Chemical: Formation of compounds that do not
dissociate in the flame
Ionisation of the analyte can reduce the signal
Matrix interferences due to differences between
surface tension and viscosity of test solutions and
standards
35
Doppler Effect
Photon detector
36
Pressure broadening
Results from collisions of absorbing/emitting species
With analyte atoms or combustion products of fuel
Deactivates the excited state – shorter lifetime - broader
spectral lines
Increases with concentration and temperature
E.g. in flame, Na absorbance lines broadened up to 10-3 nm.
37
Background correction in AAS
High energy Deuterium background corrector
Detector
Hollow
cathode Beam
lamp combiner
Lamps are
Sample
pulsed out of
phase with Deuterium lamp
each other
38
Minimising the effect of
Matrix Interferences
The term "matrix" refers to the sum of all compositional
characteristics of a solution, including its acid
composition
Calibration standards
and samples must be
matrix-matched in
terms of composition,
total dissolved solids,
and acid concentration
of the solution
Also advisable for
ICP-OES and -MS
40
Calibration – some practical
aspects
41
Calibration in AAS
In theory, Beer’s law applies
for dilute solutions
In practice, deviation from
linearity is usual
Small dynamic range
Possible to use non-linear
curve fitting for calibration
Reasons: Self-absorption:
excited atoms emit light that
can also be absorbed instead of
that of source: on average, Linear range
less light per number of atoms
is absorbed
42
Alternative to matrix-matching:
Method of standard additions
Extensively used in absorption
spectroscopy, accounts for matrix effects
Several aliquots of sample
Sample (1): diluted to volume directly
Samples (2,3,4,5…): known amounts of analyte
added before dilution to volume
BUT: Only makes sense if the added
standard closely matches the analyte
present in the samples chemically and
physically
if simple, dissolved ions are analysed
43
Method of standard additions
If linear relationship exists between measured quantity and
concentration (must be verified experimentally) then:
kVx c x kVs cs
AT
VT VT
intercept = b = (єlVxcx) / VT
Limitations
• The calibration graph must be substantially linear since accurate
regression cannot be obtained from non-linear calibration points.
• Caution: The fact that the measured part of the graph is linear does not
always mean that linear extrapolation will produce the correct results
• It is also essential to obtain an accurate baseline from a suitable reagent
blank
45
Most simple version of standard
addition: spiking
Spiking means deliberately adding analyte to
an unknown sample
Involves:
preparation of sample and measurement of
absorbance
Addition of standard with known concentration,
measurement of absorbance
From difference in absorbance, calculate e
From reading of sample alone, calculate amount
of analyte
(use Beer’s law for calculations)
46
Other uses for spiking
Add spike at beginning of sample
preparation
Process sample with and without spike
Difference should correspond to amount
spiked
Deviation allows to calculate recovery
factor
47
Atomic emission
spectroscopy
48
Atomic emission spectroscopy
Historically, many techniques based on
emission have been used (See Table on p. 4)
Flame and electrothermal methods now
widely superseded by Inductively-Coupled
Plasma (ICP) method
Developed in the 1970s
Higher energy sources than flame or
electrothermal methods
49
ICP-AES/OES
Inductively coupled plasma-atomic emission spectroscopy
(or optical emission spectroscopy)
50
Modern ICP-OES spectrometer
52
Components for sample injection
and the ICP torch
Up to 7000°C
53
Atomisation / Ionisation
54
Advantages of plasma
Prior to observation, atoms spend ~ 2 sec at
4000-8000 K (about 2-3 times that of hottest
combustion flame)
Atomisation and ionisation is more complete
Fewer chemical interferences
Chemically inert environment for atomisation
Prevents side-product (e.g. oxide) formation
Temperature cross-section is uniform (no cool
spots)
Prevents self-absorption
Get linear calibration curves over several orders of
magnitude
55
Applications
ICP-OES used for quantitative analysis of:
Soil, sediment, rocks, minerals, air
Geochemistry
Mineralogy
Agriculture
Forestry
Fornensics
Environmental sciences
Food industry
56
Lab Experiment 3
Analyse a Chromium complex for [Cr] in three ways:
UV (absorbance & extinction coefficient)
Titration (moles Cr and charge)
AAS (Cr standard curve and unknown concentration)
57