Specification: Agilent 4890D Gas Chromatograph
Specification: Agilent 4890D Gas Chromatograph
Specification
2
Detectors Flame Ionization Detector Electron Capture Detector
• Range to 450 °C • Range to 400 °C
Thermal Conductivity Detector • Grounded jet and current limited • A coaxial design with a 15 mCi
• Range to 400 °C design for operator safety source (555 MBq) of Ni–63
• Push-button flame ignition plated on the interior of the
• Single-filament (single-column) lower block
design has fluidic switching of • Fused silica columns insert
reference and analytical carrier within 2 mm of the jet tip • Constant current mode of opera-
flows; a passivated tungsten- tion features switch selection of
• Sensitivity: pulse parameters for using either
rhenium filament is used in a
3.5-µL cell operating at a con- – >18 mCoul/g carbon: nitrogen, hydrogen, helium, or
stant temperature difference rel- nitrogen carrier, 0.018-in. id jet argon/methane carrier gas
ative to the detector block – >15 mCoul/g carbon: • Minimum detectable level:
temperature helium carrier, 0.018-in. id jet <0.04 pg/sec lindane
• Minimum detectable: <400 pg/mL – >22 mCoul/g carbon: • Dynamic range: >104 for lindane
carrier—equivalent to <1 ppm of nitrogen carrier, 0.011-in. id
neon in 1 mL of air (may be capillary jet Conditions: detector 250 °C,
adversely affected by acoustic 60 mL/min nitrogen carrier
noise in the laboratory – >18 mCoul/g carbon:
environment) helium carrier, 0.011-in. id
capillary jet
• Linear dynamic range: <±5% over
105 range • Minimum detectable: <5 pg
carbon/sec, nitrogen carrier at
• Digital gain setting time–pro- S/N=2
grammable through the keyboard
• Linear dynamic range: <±10%
over a 107 range with 0.018-in. id
jet
Conditions: detector 100 °C,
45 mL/min switching and Conditions: column flow 50 mL/min,
30 mL/min analytical flow of 45 mL/min H2, 650 mL/min air,
helium, propane sample propane sample
3
Column Oven
• Usable volume:
– 11 in. × 12 in. × 6.5 in. +0.075 +0.21
(h × w × d)
– 279 mm × 305 mm ×
165 mm (h × w × d)
• Column span: 228.5 mm, +0.050 100 ˚C Oven +0.037 -0.40 300 ˚C Oven -0.46
9 in. (coil size)
• Automatic cooling under
processor control
• Operating range:
4 °C above ambient to 450 °C: 0.00 -0.58
80 to 450 °C with cryogenic
cooling
• Setpoint entry:
– 1 °C for temperatures Oven Heating Profiles
– 0.1 °C for program rates Temp (˚C) Temp (˚C)
• Programming: 118 Vac
300 300
240 Vac 108 Vac One Series 530-µ column, one
– Rates 0.1 to 70 °C* per min
injection port, and one detector
– 650 min maximum run time at 300 ˚C (oven held at 300 ˚C
200 200 for 1 hour prior to cooldown)
– Three ramps with 108 Vac Curve: One empty
initial/final holds 10-ft ss column; injection
port and detector unheated
100 118/220 Vac: One Series
Technical Performance 530- µ column; one injection 100
• Accuracy (true temperature rela- port, and one detector at 300˚C
50 50
tive to setpoint) 0 2 4 6 8 10 12 0 2 4 6 8
Specification: +1% (°K) from 4 °C Time (min) Time (min)
above ambient to 450 °C
• Stability (effect of ambient • Gradients (temperature varia-
change on actual temperature) tions within a column)
Specification: <0.01 °C for Specification: less than 2 °C
1 °C ambient change within a 9-in. coil anywhere
• Calibration (setting true temper- within the operating range
ature at a setpoint) *Achievable rates depend on zone
Oven can be recalibrated to temperature, voltage, and columns.
±0.01 °C with appropriate
instrumentation