Professional Documents
Culture Documents
SIO2 Manual Eng-V08 AIC
SIO2 Manual Eng-V08 AIC
INSTRUCTION MANUAL
Version 2.08
Warnings
This instrument conforms to DIN57411 part 1 / VDE 041 part 1 "protective measures for electronic
measuring instruments" and has left the factory in faultless condition. To maintain this condition
and to guarantee harmless operation of the instrument, the user has to observe all warnings and
directives of the manual.
Interruption of the protective grounding line or loosening the ground connection inside or outside
the instrument may lead to dangerous situations. Disconnecting the ground is prohibited.
Disconnect power whenever electronic service is required. Care should be taken when opening or
removing parts of the instrument; connectors may be under tension. Service should be performed
by authorized personnel only.
If proper operation is not possible anymore, the instrument has to be disconnected from all power
lines and measures should be taken to prevent inadvertent operation.
Contents
1. Introduction
1.1 Priciple of operation 4
1.2 Description of the measuring system 5
2. On-line operation
2.1. Frontpanel functions 5
2.2. Display - submenu 100 : Measurement 6
3. Installation
3.1. On-site requirements 6
3.2. Sample requirements 6
3.3. Electrical installation 6
8. Maintenance 17
9. Electrical Wiring
9.1 Wiring diagram for user 18
1. Introduction
1.1 Principal of operation
The analyzer has a microprocessor based display/control unit and an analytical part.
The heart of the analytical part has a measuring cell, in which a magnetic stirrer to mix the reagents
with the sample and, a 860 nm LED light source with a light detector to measure the intensity of the
formed color (absorbance).
The sample enters the analyzer through a needle inlet valve and an electromagnetic inlet valve to
the measuring cell from where it overflows into the drain. When the measuring cell is rinse out by
the sample, the electromagnetic inlet valve closes and about 25 ml of sample is trapped inside the
cell.
Before the first reagent is added, the light intensity is measured and set to zero absorbance. In this
way sample turbidity and color and light source aging is automatically compensated.
Reagent 1; Ammonium-molybdate reacts with reactive silica and phosphate to form a weak
yellow color. The reaction time is 240/300 seconds.
Reagent 2: Oxalic acid is added to decomplex the phosphate-molybdate compound and to stop
further reaction of the molybdate with the silica. For measurements in de-ionized
water, boiler feedwater without phosphate, condensate and low levels of silica this
reagent is not wanted. (optional)
Reagent 3: Reducing agent converts the weak yellow color in a more intense blue color, which
is very suitable for low ppb measurements. The reaction time is 300/360 seconds.
The sample cell is flushed at the end of the analysis with fresh sample and the analyzer is ready for
a new measurement.
For the calibration another electromagnetic valve just after the needle inlet valve is activated to
rinse and fill the measuring cell with a standard solution (recommended 250 ppb). The absorbance
measured with the standard solution is than equaled to its concentration value. The calculated
conversion factor is shown on the display and used to calculate the concentrations in the sample.
.
Manual silica Prolytics 2030 4
___________________________________________________________________________________________
Note : This is a schematic drawing. The flow detector can also be an overflowvessel at the
top, near the standard reagent bottle
Manual silica Prolytics 2030 5
___________________________________________________________________________________________
2. On-line operation
On the frontpanel one can see four LED 's with the following functions :
ANALYSIS, CALIBRATION, HIGH ALARM, FAIL ALARM
Select/Return The SELECT-key is used to enter and leave a „highlighted“ program menu.
In a program menu the SELECT-key is used to enter and leave „highlighted“
program sub-menu parameters.
In the sub-menu parameters the SELECT key is used to confirm a parameter
change.
ARROW keys The ARROW (up and down) keys are used to move the „highlight“ bar to the
different menu items. When the „highlight“ bar is not on the display, it is on
the corresponding one level up.
The ARROW-keys are also used to change parameters (confirm with Select-
key)
By selecting the submenu 100 : Measurement one gets the principal display menu
3. Installation
Parameter
Warning !
For viewing the main menu with the program possiblities, one has to press the
SELECT/RETURN key.
Press SELECT/RETURN
Par. Autoanalyse : Selecting the frequency of analysis, the washing time for
an analysis, On/Off and STORE
Manual test : For activating or de-activating the valves, pumps and stirrer.
4-20 mA output : Programming the analog output range. Shows the current
output in mAmps.
Trend To view the last 100 measured values update every 15 min
Time parameters To program the pump and reaction times, these values are
entered in the factory. Do not change them unless you are
specifically instructed to do so.
Time schedule To view the remaining times for the next calibration/analysis
Manual silica Prolytics 2030 8
___________________________________________________________________________________________
Select in the "level 1 data" menu 101 : Cal/Phot value and the following screen will be
viewed.
This is only a display menu and none of the parameters can be changed here.
The actual photometer output is measured in mV and monitoring helps for solving problems.
The cal(ibration) values are the concentration of the standard solution, which can be changed
in the submenu 103 : Par. autocal.
The blanc is value of impurities coming from the reagent in ppb. The blank value is
measured in 0.0 ppb sample. (Change under 103 : Par. autocal)
The factor is an indication for the response of the photometer. This value is calculated
after each calibration. The cal factor must be between 0.7 -1.3.
For changing parameters, move the cursor („highlight“ bar) to the line of the parameter and
press SELECT/RETURN. The line starts blinking and value is changed using the ARROW
keys. Press the SELECT/RETURN key to confirm the new value.
The standard frequency time for an analysis is set to 20 min, but can be changed from 1 to
900 min.
The status indicates if the auto analysis function is active or not. With the ARRROW keys
Manual silica Prolytics 2030 9
___________________________________________________________________________________________
After changing the parameters, the changes have to be confirmed in the STORE function.
Press SELECT/RETURN on the STORE line, the line starts blinking and select with the
ARROW keys YES/OK and press SELECT/RETURN.
To return to the main menu, move the cursor line off the display and press the
RETURN/SELECT key.
Select the 103 : Par. autocal submenu and the following screen will be shown.
The status indicates if the auto calibration function is active or not. With the ARRROW
keys and the SELECT/RETURN key ON or OFF can be selected.
After changing the parameters the changes have to be confirmed inthe STORE function.
Press SELECT/RETURN on the STORE line, the line starts blinking and select with the
ARROW keys YES/OK and press SELECT/RETURN.
To return to the main menu, move the cursor line off the display and press the
RETURN/SELECT key.
In the 104 : Start analysis submenu an analysis of the sample is started. Select START
ANALYSIS with the SELECT/RETURN key and to choose YES/OK with the ARROW
keys and press the SELECT/RETURN key.
Manual silica Prolytics 2030 10
___________________________________________________________________________________________
The reset function is used to stop an analysis. Select the RESET line and choose
YES/OK with the ARROW keys and press SELECT/RETURN.
NOTE : After RESET the inlet valve is activated according to the rinsing time
In the 105 : Start calibration submenu the user can start a calibration by selecting START
CALIBRATION with the SELECT/RETURN key and to choose YES/OK with the
ARROW keys and again pressing the SELECT/RETURN key.
The reset function is used to stop a calibration. Select RESETl and choose
YES/OK with the ARROW keys and press SELECT/RETURN.
The measured concentration in the standard solution, based on the previous calibration
factor is shown in the last line. This value has been used to calculate the new calibration
factor.
NOTE : With the RESET the inlet valve will be activated according to the rincing time
.
With the 106 : Manual test submenu one can activate/desacitvate pumps, valves and stirrer.
In the start-up procedure (paragraph 6) these operation functions are used.
With the SELECT/RETURN and the ARROW keys one can change :
Inlet valve off into Inlet valve on
Reagent 1,2 and 3 off into Reagent 1,2 and 3 on (start pump 1,2 and 3)
Cal. Valve off into Cal valve on
Stirrer off into Stirrer on
CONCENTRATED REAGENT.
WARNING: MAKE SURE THAT ALL THE FUNCTIONS ARE RESET TO OFF !
The 107 : Service submenu views the photometer outputs after the addition of the different
reagents. The difference between Eo and E3 is used to calculate the silica concentration.
The data are also the base for error messages.
Eo < 4000 mV Dirty cell
Eo < 50 mV Photom. error
Eo =/> E3 No Reagent
The first set of data from a 2 pump system gives 214 ppb. Other typical sets are next to it.
The analog output can be programmed through the submenu 108 : Analog output.
The start value is the value at 4 mA, the end value is the value at 20 mA.
With the SELECT/RETURN key and the ARROW keys, the default values of 0 and 500 ppb
are changed.
When a failure is detected, the right LED on the controller will lighten on and the fail relay is
activated.
In any display the top line will also give the message FAIL ALARM
The relay and fail message will remain activated unless the user acknowledge the fail message
in the submenu 109 : Fail alarms
The message is confirmed selecting the acknowledge line and with the SELECT/RETURN
key, go to Yes/ok and press SELECT/RETURN key.
All the messages are deleted until a next failure is detected.
Note : When the reagent tube(s) is|(are) blue, it is recommend to immerse the tube(s) in
5% ammonia solution. Start the reagent pump(s) and pump the ammonia solution in
the measuring cell for 15 minutes.
Manual silica Prolytics 2030 13
___________________________________________________________________________________________
An high alarm value (default 500 ppb) is available and is changed by using the
SELECT/RETURN key and the ARROW keys.
In case of an high alarm the LED on the controller will lighten up, the high alarm relay is
activated and in the top line of any display the message HIGH ALARM will be shown.
In the submenu 111: Trend the user can look at the last 100 measurements updated every 15
minutes.
In the submenu 112 : Time parameters, the pump and chemical reaction times can be viewed
and changed. The values are already updated during factory calibration and the user should first
consult a specialist before changing these times.
After changing the time parameters, the entries have to be stored before they become active
(similar for calibration and analysis parameters).
Manual silica Prolytics 2030 14
___________________________________________________________________________________________
5.1 Reagent 1
The preparation of this reagent releases heat. Rinsing the bottle with cold water is necessary or add
the sulfuric acid in small portions.
It is recommended to designate 2 liter plastic bottles for each reagent for preparation.
1. Fill a 2 liter reagent storage bottle with about 1.2 liter ultra pure water
2. Add 200 gram sulfuric acid H2SO4 (100%, analytical grade)
in small portions to prevent overheating. (110 ml H2SO4 = 200g)
3. Add 80 grams of ammonium molybdate (Merck 1182)
slowly while keeping the solution in motion.
4. Keep shaking the solution until all is dissolved.
5. Fill up with pure water to 2 liter mark.
1. Fill a 2 liter reagent storage bottle with about 1.5 liter ultra pure water
2. Add 40 gram of oxalic acid slowly while keeping the solution in motion.
3. Fill up with pure water to 2 liter mark.
4. Keep shaking the solution until all is dissolved.
5.3 Reagent 3
1. Fill a 2 liter reagent storage bottle with about 1.5 liter ultra pure water
2. Add 160 gram of Potassium Disulfite (Merck 5057)
slowly while keeping the solution in motion.
3. Add 40 gram of 4-methylaminophenol sulfate
(Photo-Rex Merck 7299) slowly while keeping the solution in motion.
4. Fill up with pure water to 2 liter mark.
5. Keep shaking the solution until all is dissolved.
Our reagent packs are delivered without sulfuric acid. All other reagents are in powder form and
require only to be dissolved in the prescribed amount on the plastic container.
Manual silica Prolytics 2030 16
___________________________________________________________________________________________
6. START-UP.
During start-up, the reagent bottles first have to be filled and the analyzer has to be prepared to
analyze the sample.
Note :
- Rinsing time in PAR. AUTOANALYSE should be long enough for good washing !
- Always switch off all pumps, valves and stirrer when leaving the MANUAL TEST
menu.
A blank determination (SiO2 in the reagent) can be performed by using a 0.0 ppb SiO2 solution in
the standard bottle and start a grab sample. Measurements will be shown in the menu.
7 Operation
7. 1 Normal operation
Under normal operation the electromagnetic inlet valve and the electromagnetic calibration valve
are deactivated. The calibration valve is only activated during the first rinsing step of the
calibration cycle.
and the inlet valve is activated when an analyze or calibration cycle is started.
When the analyzer is in stand-by mode the sample flows through the 2 electromagnetic valves into
the drain. To reduce sample consumption, the outlet from the valve to the drain may be closed.
It is possible to measure the silica content of a different sample point using the standard bottle.
1. Rinse the standard bottle well.
2. Fill the standard bottle with 1 liter of sample
3. Place the bottle in its place in the analyzer.
4. Go in menu to GRAB SAMPLE, select START change to YES/OK, cal valve will open
automatically. 3 values will be shown in the same menu for 1 liter of sample.
5. Remove the grab sample from the bottle and replace with standard.
8. Maintenance
Weekly: Check reagent levels, refill and fill the reagent tubing as explained in start-up
3 month Clean measuring cell and reagent tubing with 5% ammonia solution.
Place reagent tubing in ammonia solution and run the pumps for 15 minutes.
Open inlet valve to remove the ammonia from the measuring cell.
Perform star-up procedure
9. Electrical Wiring
1 Ground -
2 Neutral -
3 220 VAC -
6 IR led + brown) Amplifier (red) red
7 IR led - (white) Amplifier (red) blue
8 Detector + (brown) Amplifier yellow
9 Detector - (white) Amplifier green
13 Inlet valve - 24 VDC
14 Inlet valve Amplifier (black) blue
15 Calibration valve Amplifier (black) red
16 Calibration valve + 24 VDC
17 Pump 1 ( -24 VDC) -
18 Pump 1 (+ 24 VDC) -
19 + VDC stirrer (brown) -
20 - VDC Stirrer (white) -
21 + mA output Z40
22 - mA output Z39
24 Flow detector Z35
25 Flow detector -
27 Pump 3 (+24VDC) Z37
28 Pump 3 (-24 VDC) -
29 Potential free input start analysis + 24 VDC
30 Potential free input start analysis Z36
31 NC High value (250 VAC, 1A)
32 C High value (250 VAC, 1A)
33 NO High value (250 VAC,1A)
34 NO fail alarm contact (250 VAC, 1A)
35 C fail alarm contact (250 VAC, 1A)
36 NC fail alarm contact (250 VAC,1A) |
All menus with parameters which can be changed are protected by a password.
The factory password is 99.
Go back to 101 : Password and set the password to a value different from the password.
Go to :
Menu 107 : Cal menu
If you are measuring values below 50 ppb, we recommend you to calibrate the analyzer with 250 or
200 ppb. (Be aware 1000mg Si = 2140 mgSiO2)
Put the solution in the calibration bottle and start the calibration.
To verify the calibration one could start an analysis while activating the cal valve by hand so that
the cal solution enters the reaction chamber. The value which will be displayed will be within 5%
of the actual value.
Preparing a solution with half the concentration of the cal solution and repeat the same as explained
above will give a value which will be within 5% of the actual value.
If this is not the case then the reagents or standards should be checked or replaced.