Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

The Humidity in a Dräger Primus Anesthesia Workstation

Using Low or High Fresh Gas Flow and With or Without a


Heat and Moisture Exchanger in Pediatric Patients
Gustavo P. Bicalho, MD,* Leandro G. Braz, MD, PhD,* Larissa S. B. de Jesus, MD,*
Cesar M. C. Pedigone,* Lídia R. de Carvalho, PhD,† Norma S. P. Módolo, MD, PhD,*
and José R. C. Braz, MD, PhD*

BACKGROUND: An inhaled gas absolute humidity of 20 mg H2O·L−1 is the value most con-
sidered as the threshold necessary for preventing the deleterious effects of dry gas on the
epithelium of the airways during anesthesia. Because children have small minute ventilation,
we hypothesized that the humidification of a circle breathing system is lower in children com-
pared with adults. The Primus anesthesia workstation (Dräger Medical, Lübeck, Germany) has
a built-in hotplate to heat the patient’s exhaled gases. A heat and moisture exchanger (HME) is
a device that can be used to further humidify and heat the inhaled gases during anesthesia. To
evaluate the humidifying properties of this circle breathing system during pediatric anesthesia,
we compared the temperature and humidity of inhaled gases under low or high fresh gas flow
(FGF) conditions and with or without an HME.
METHODS: Forty children were randomly allocated into 4 groups according to the ventilation
of their lungs by a circle breathing system in a Dräger Primus anesthesia workstation with low
(1 L·min−1) or high (3 L·min−1) FGF without an HME (1L and 3L groups) or with an HME (Pall
BB25FS, Pall Biomedical, East Hills, NY; HME1L and HME3L groups). The temperature and
absolute humidity of inhaled gases were measured at 10, 20, 40, 60, and 80 minutes after
connecting the patient to the breathing circuit.
RESULTS: The mean inhaled gas temperature was higher in HME groups (HME1L: 30.3°C ±
1.1°C; HME3L: 29.3°C ± 1.2°C) compared with no-HME groups (1L: 27.0°C ± 1.2°C; 3L: 27.1°C
± 1.5°C; P < 0.0001). The mean inhaled gas absolute humidity was higher in HME than no-HME
groups and higher in low-flow than high-flow groups ([HME1L: 25 ± 1 mg H2O·L−1] > [HME3L: 23
± 2 mg H2O·L−1] > [1L: 17 ± 1 mg H2O·L−1] > [3L: 14 ± 1 mg H2O·L−1]; P < 0.0001).
CONCLUSIONS: In a pediatric circle breathing system, the use of neither high nor low FGF pro-
vides the minimum humidity level of the inhaled gases thought to reduce the risk of dehydration
of airways. Insertion of an HME increases the humidity and temperature of the inhaled gases,
bringing them closer to physiological values. The use of a low FGF enhances the HME efficiency
and consequently increases the inhaled gas humidity values. Therefore, the association of an
HME with low FGF in the breathing circuit is the most efficient way to conserve the heat and the
moisture of the inhaled gas during pediatric anesthesia. (Anesth Analg 2014;119:926–31)

I
n normal breathing, inhaled air is warmed and humidi- lead to cilia and mucous gland destruction, reduced muco-
fied in the upper airways.1 During tracheal intubation and ciliary transport, and thickening of secretions.3–5 Preventing
mechanical ventilation, these mechanisms are partially dehydration is especially important in children because of
disrupted because the tracheal tube bypasses the upper air- the reduced caliber of the tracheal tube. Studies from animal
ways. In this situation, ventilation with dry and cold gases research3–5 and a recent review on the subject6 proposed that
leads to a considerable loss of water and heat from the respi- during anesthesia, the minimum humidity level of inhaled
ratory tract, unless appropriate means of humidification and gases should be 20 mg H2O·L−1.
heating are used.2,3 Dehydration of the respiratory tract can In anesthesia, the use of a circle breathing system with
low fresh gas flow (FGF) can help maintain the humidity of
From the *Department of Anesthesiology, Botucatu Medical School, and the inhaled gases.4,7,8 This maintenance occurs because the
†Department of Biostatistics, Institute of Biosciences, UNESP—Universidade humidified exhaled gases are partially rebreathed, and water
Estadual Paulista, São Paulo State, Brazil. is generated when the carbon dioxide (CO2) reacts with the
Accepted for publication May 9, 2014. soda lime. In adults, it has been shown that the use of a circle
Funding: Supported by grant 2011/13545-0, São Paulo Research Foundation breathing system with low FGF can provide humidity lev-
(FAPESP).
els of the inhaled gases just above 20 mg H2O·L−1.8–10 Because
The authors declare no conflicts of interest.
they have smaller minute ventilation, children generate less
This report was previously presented, in part, at the Annual Meeting,
American Society of Anesthesiologists, Washington, October 2012. CO2, water, and heat compared with adults. One may specu-
Reprints will not be available from the authors. late that the humidification of a circle breathing system with
Address correspondence to Gustavo P. Bicalho, MD, Department of Anesthe- low FGF in children may be lower than that in adults. Circle
siology, Botucatu Medical School, UNESP—Universidade Estadual Paulista, breathing systems are frequently used during pediatric anes-
­Distrito de Rubião Junior, P.O. Box 530, CEP: 18.618-970, São Paulo State,
­Brazil. Address e-mail to gpbicalho@gmail.com. thesia.11 Few studies have specifically examined the tempera-
Copyright © 2014 International Anesthesia Research Society ture and humidity of the inspired gases provided by a circle
DOI: 10.1213/ANE.0000000000000353 breathing system during anesthesia in children.12,13

926   
www.anesthesia-analgesia.org October 2014 • Volume 119 • Number 4
A heat and moisture exchanger (HME) is a device Orchard Park, NY) with the temperature initially set at 38oC.
designed to provide humidification and heating of inhaled The mattress was placed at the back of the patient from the
gases similar to that provided by the upper airways. It has mid-thoracic region and below, and a single cotton sheet
been shown that during anesthesia, the use of an HME can separated the back of the patient from the water mattress.
increase the temperature and humidity of inhaled gases in Oxygen-sevoflurane mask induction with high FGF
adults8–10 and children.14,15 (3 L·min−1) was performed in all patients. After the child
The Primus anesthesia workstation (Dräger Medical, was unconscious, an IV line was inserted with a 22-G cath-
Lübeck, Germany) has a built-in hotplate to heat the exhaled eter, and an infusion of lactated Ringer’s solution (4–6
gases and to prevent water condensation in the breathing mL·kg−1·h−1) at room temperature was initiated. An IV
system. The temperature and humidity of the inhaled gases bolus of fentanyl (5 µg·kg−1) and an IV bolus of cisatracu-
coming from this anesthesia machine have not yet been rium besylate (0.15 mg·kg−1) were given to facilitate tracheal
investigated in children. It may be expected that the hotplate intubation with a cuffed tracheal tube. The cuff was inflated
would warm the inhaled gases coming from this workstation. to ensure the absence of any air leaks. A pediatric HME
The aim of this study was to compare the mean tempera- (Pall BB25FS, Pall Biomedical, East Hills, NY) was placed
ture and humidity of the inhaled gases using low or high between the Y-piece of the breathing circuit and the tracheal
FGF with or without an HME in the circle breathing sys- tube in the HME groups. The lungs were mechanically ven-
tem of the Primus anesthesia workstation during pediatric tilated using a pressure-controlled mode. An inspiratory
anesthesia. pressure was set to maintain a tidal volume of 8 mL·kg−1.
The respiratory rate was adjusted to maintain a Petco2 close
METHODS to 35 mm Hg. The inspiratory and expiratory sevoflurane
This study was approved by the local Medical Ethics concentrations, oxygen concentration, end-tidal carbon
Committee (CEP 3877-2011) and registered at ensaiosclini- dioxide (Petco2), and ventilation variables were monitored
cos.gov.br (RBR-33bb7k). Written informed consent was with the Primus built-in monitor.
obtained from the parents of all children studied. Forty chil- After 5 minutes of high FGF (1.5 L·min−1 of O2 in 1.5
dren (10–30 kg), afebrile (T <37ºC) and with ASA physical L·min−1 of air), with an inspiratory concentration of sevoflu-
status I or II, who were scheduled for elective abdominal rane at the minimum alveolar concentration (2%), the FGF
or urological surgeries lasting 90 minutes or longer, were was reduced to 0.5 L·min−1 of O2 in 0.5 L·min−1 of air in the
enrolled in this study. patients randomly chosen to receive an FGF of 1 L·min−1
Patients were randomized into 4 groups (10 children (1L and HME1L groups), but not in the groups selected to
per group), according to the FGF used and the use of an receive an FGF of 3 L·min−1 (3L and HME3L groups). The
HME in the breathing circuit, as follows: 1L group, FGF of sevoflurane concentration was adjusted throughout the
1.0 L·min−1 without an HME; 3L group, FGF of 3 L·min−1 anesthesia to maintain a systemic arterial blood pressure
without an HME; HME1L group, FGF of 1.0 L·min−1 with and cardiac rate within ±20% of baseline. Anesthesia was
an HME; and HME3L group, FGF of 3 L·min−1 with an maintained with IV fentanyl (2 µg·kg−1) and cisatracurium
HME. An anesthesiologist who was not involved in the (0.02 mg·kg−1), if necessary.
perioperative management of the patients printed 40 group The intraoperative esophageal (core) temperature was
identification tags (10 for each group) and placed them in measured after tracheal intubation using a thermocou-
40 envelopes (1 identification tag per envelope). Then the ple sensor (90044 Mon-a-Therm®; Mallinckrodt Medical,
same anesthesiologist sealed, mixed, and listed the enve- Veracruz, Mexico). The thermocouple sensors for the esoph-
lopes sequentially, i.e., numbered in ascending order. Just agus and the OR temperature were attached to a 2-chan-
before each anesthesia induction, 1 envelope was opened nel electronic thermometer (4070, Mallinckrodt Medical, St.
following the sequential order. Louis, MO).
The temperature and relative humidity (RH) of the
Experimental Protocol gases were measured using a rapidly responding electronic
A Dräger Primus anesthesia workstation was used in all thermo-hygrometer (Kimo HD50, Marne La Vallée, France;
cases. The anesthetic system was equipped with clean and 90% response in <1 seconds). The RH and temperature sen-
dry silicone pediatric corrugated tubes (1.5 m in length and sor of the thermo-hygrometer was connected by a T-piece
10 mm in internal diameter; Dräger Medical) for each case. between the Y-piece of the breathing circuit and the tracheal
The CO2 canister (1.5 L) of the anesthesia machine was filled tube in the no-HME groups and between the HME and the
with fresh soda lime (Drägersorb 800 Plus, Dräger, Lübeck, tracheal tube in the HME groups; and it was connected to
Germany) before each case. the inspiratory limb outlet close to the anesthesia worksta-
None of the selected patients was excluded from the tion in all the groups. The thermo-hygrometer operates on
study. Patients received midazolam (0.3 mg·kg−1 by mouth) a capacitative principle and has a stated accuracy of ±2.5%
60 minutes before their admission to the operating room for RH and ±0.3°C for temperature. The RH and tempera-
(OR). Upon arrival to the OR, the patients received standard ture values fluctuated with the phases of the respiratory
clinical monitoring with an electrocardiogram (DII and V5 cycles and were lower during the inspiratory phase. We
leads), peripheral oxygen saturation (Spo2), and noninva- recorded the minimum values of the RH and temperature
sive arterial blood pressure measurements. over 10 respiratory cycles after 10, 20, 40, 60, and 80 min-
All patients received active conductive warming using a utes of connection between the patient and the breathing
pediatric circulating-water mattress from a warming device circuit. The absolute humidity (AH) values were calculated
(Medi-Therm III, Model MTA 6900, Gaymar Industries Inc., using the formula: AH = (3.939 + 0.5019T + 0.00004615T2 +

October 2014 • Volume 119 • Number 4 www.anesthesia-analgesia.org 927


Humidity of Inhaled Gases in Pediatric Anesthesia

0.0004188T)3 × RH/100, where AH is the absolute humidity (Windows Software, version 6.0; SPSS Inc., Chicago, IL).
(mg H2O·L−1), T is the temperature (oC), and RH is the rela- For all the analyses, P < 0.05 was considered statistically
tive humidity (%). significant.

Statistical Analysis RESULTS


The sample size of the groups was calculated based on pre- There were no statistically significant differences among the
vious studies regarding the humidification of gases during groups with regard to the patients’ characteristics (Table 1).
anesthesia, where the differences between groups who used There were no differences regarding hemodynamic vari-
an HME or not were >5.0 mg H2O·L−1.13,15,16 Thus, assum- ables (data not shown).
ing an AH difference of 5.0 mg H2O·L−1 among groups with There were no statistically significant differences among
an SD of ±3.0 mg H2O·L−1, a minimum of 10 patients in the groups with regard to the mean OR and core (esopha-
each group was necessary to detect this difference using a geal) temperatures (Table 2). The mean temperature of the
2-tailed test with the probability of a type I error (α) of 0.05 inhaled gas was higher in the HME groups compared with
and a type II error (β) of 0.05 (power of 95%). the no-HME groups (P < 0.0001; Table 3).
The normal distribution of the data was confirmed using The mean temperature in the inspiratory limb outlet of
the Lilliefors tests. Each of the 4 groups’ pooled data fol- the workstation was lower in the 1L group compared with
lowed a normal distribution (all Lilliefors P > 0.10). the HME groups (P = 0.003; Table 3). The mean RH and AH
The anthropometric variables were compared among values in the inspiratory limb outlet close to the anesthe-
groups by analysis of variance, and gender was compared sia workstation were higher in the groups using an FGF
among groups with the χ2 test. of 1 L·min−1 (1L and HME1L) compared with the groups
The AH, RH, and temperature values were compared who used an FGF of 3 L·min−1 (3L and HME3L; P < 0.0001;
among the 4 groups at different time points by a profile Table 4).
analysis. In this analysis, the hypotheses tested were the fol- The mean RH and AH values of the inhaled gases were
lowing: there was no interaction between groups and time significantly higher in the groups who used the HME
points, and there was no significant difference between (P < 0.0001). In the HME groups, the group with the low
mean groups over time. This analysis was followed by FGF (HME1L) had higher mean RH and AH values com-
Tukey test for pairwise comparisons. pared with the high FGF group (HME3L; P < 0.0001). In the
Pearson coefficient was used for the correlation analy- groups that did not use an HME, the group with the low
sis between the OR and inhaled gas temperatures in all the FGF (1L) had higher mean RH and AH values compared
groups. with the high FGF group (3L; P < 0.0001; Tables 5 and 6,
The data were expressed as the means ± SD and 95% respectively).
confidence interval. The statistical analysis was per- There was a significant and positive correlation between
formed using the Statistical Package for the Social Sciences the OR and inhaled gas temperatures in the no-HME groups
(1L: r2 = 0.52; P < 0.0001 and 3L: r2 = 0.31; P < 0.0001) but
not in the HME groups (HME1L: r2 = 0.0036; P = 0.66 and
Table 1.  Patient Characteristics Data (Means ± SD) HME3L: r2 = 0.014; P = 0.37).
in the Groups Studied None of the children had any surgical or clinical com-
Sexa plications, and all of the children were discharged from the
Groups N Age (y) Weight (kg) Height (cm) (M/F)
hospital, according to the guidelines and protocols estab-
1L 10 4.0 ± 2.4 18.4 ± 6.0 102.9 ± 18.6 8/2
lished for their particular surgical procedures.
3L 10 4.3 ± 2.6 19.3 ± 5.5 108.3 ± 18.6 10/0
HME1L 10 4.6 ± 1.9 19.5 ± 5.9 109.2 ± 18.0 9/1
HME3L 10 4.9 ± 2.5 19.4 ± 6.0 110.0 ± 13.0 10/0 DISCUSSION
P value 0.85 0.98 0.79 0.10 There are 3 main findings in the present study: (1) the
M = male; F = female. humidity of the inhaled gases in the Primus workstation
a
Data are frequency distributions. with a pediatric breathing circuit depends on the FGF and

Table 2.  Operating Room and Core Temperature Data (Means ± SD) in the Groups Studied
P value
Difference among Time × group
Temperature (ºC) Group Mean over time 95% CI groups interaction
Operating room 1L 22.6 ± 0.8 22.3–31.0 0.85 0.68
3L 22.8 ± 0.5 22.6–31.3
HME1L 22.8 ± 0.6 22.6–31.3
HME3L 22.7 ± 0.7 22.4–31.2

Core 1L 36.2 ± 0.5 36.0–49.9 0.18 0.04


3L 36.3 ± 0.4 36.1–50.0
HME1L 36.1 ± 0.4 36.1–49.8
HME3L 36.5 ± 0.5 36.3–50.3
Mean over time: mean of each parameter within each group from 10 to 80 min after connection of the patients to the breathing circuit. The SD listed is the SD
of the means among subjects.
CI = confidence interval.

928   
www.anesthesia-analgesia.org anesthesia & analgesia
Table 3.  Workstation Outlet Inspiratory Limb and Inhaled Gas Temperature Data (Means ± SD) in the
Groups Studied
P value
Difference among Time × group
Temperature (ºC) Group Mean over time 95% CI groups interaction
Workstation outlet 1L 30.2 ± 1.6b 29.5–41.4 0.003 0.54
inspiratory limb gas 3L 30.5 ± 1.3ab 30.0–41.9
HME1L 31.9 ± 1.0a 31.5–43.8
HME3L 31.8 ± 1.1a 31.4–43.7

Inhaled gas 1L 27.0 ± 1.2b 26.5–37.1 <0.0001 0.22


3L 27.1 ± 1.5b 26.6–37.2
HME1L 30.3 ± 1.1a 29.8–41.6
HME3L 29.3 ± 1.2a 28.8–40.3
Mean over time: mean of each parameter within each group from 10 to 80 min after connection of the patients to the breathing circuit. The SD listed is the SD
of the means among subjects. Means followed by different superscript letters are significantly different.
CI = confidence interval.

Table 4.  Relative and Absolute Humidity Data of the Gases in the Workstation Outlet Inspiratory Limb
(Means ± SD) in the Groups Studied
P value
Difference among Time × group
Group Mean over time 95% CI groups interaction
RH (%)
 Workstation outlet 1L 36 ± 5a 34–49 <0.0001 0.01
inspiratory limb gas 3L 15 ± 5c 13–19
HME1L 39 ± 3a 37–53
HME3L 20 ± 4b 18–27
AH (mg H2O·L−1)
 Workstation outlet 1L 11 ± 2b 10–15 <0.001 0.17
inspiratory limb gas 3L 5 ± 2d 4–6
HME1L 13 ± 1a 12–18
HME3L 7 ± 2c 6–9
Mean over time: mean of each parameter within each group from 10 to 80 min after connection of the patients to the breathing circuit. The SD listed is the SD
of the means among subjects. Means followed by different superscript letters are significantly different.
RH = relative humidity; AH = absolute humidity; CI = confidence interval.

Table 5.  Relative Humidity Data of the Inhaled Gases (Means ± SD) in the Groups Studied
Time (min)
Groups 10 20 40 60 80 Mean over time 95% CI P value
1L 65 ± 3 65 ± 4 65 ± 4 64 ± 4 65 ± 3 65 ± 4c 63–89 <0.0001
3L 56 ± 3 55 ± 5 55 ± 6 54 ± 3 54 ± 4 55 ± 4d 53–75
HME1L 80 ± 3 81 ± 2 81 ± 2 80 ± 1 81 ± 2 80 ± 2a 79–110
HME3L 77 ± 3 76 ± 4 76 ± 3 77 ± 3 76 ± 3 76 ± 3b 75–105
Mean over time: mean of each parameter within each group from 10 to 80 min after connection of the patients to the breathing circuit. In mean over time, the
SD listed is the SD of the means among subjects. There was no significant time × group interaction (P = 0.25). Means followed by different superscript letters
are significantly different.
CI = confidence interval.

Table 6.  Absolute Humidity Data of the Inhaled Gases (Means ± SD) in the Groups Studied
Time (min)
Groups 10 20 40 60 80 Mean over time 95% CI P value
1L 16 ± 1 17 ± 2 17 ± 1 17 ± 1 17 ± 1 17 ± 1c 16–23 <0.0001
3L 14 ± 1 15 ± 1 14 ± 1 14 ± 1 14 ± 1 14 ± 1d 14–19
HME1L 24 ± 2 25 ± 1 25 ± 2 25 ± 1 25 ± 1 25 ± 1a 24–34
HME3L 22 ± 2 22 ± 2 22 ± 2 22 ± 2 23 ± 2 22 ± 2b 22–31
Mean over time: mean of each parameter within each group from 10 to 80 min after connection of the patients to the breathing circuit. In mean over time, the
SD listed is the SD of the means among subjects. Means followed by different superscript letters are significantly different. There was significant time × group
interaction (P = 0.04).
CI = confidence interval.

is higher with a low FGF; (2) the minimum recommended HME increased the temperature and humidity of inhaled
inhaled gas humidity of 20 mg H2O·L−1 was not achieved gases, bringing them closer to physiological values, which
with either the high or low FGF; and (3) insertion of an were higher with the low FGF.

October 2014 • Volume 119 • Number 4 www.anesthesia-analgesia.org 929


Humidity of Inhaled Gases in Pediatric Anesthesia

There are 2 sources of humidity and heat in a circle the FGF than adults, thus allowing for proportionally less
breathing system. One source is the rebreathing of exhaled rebreathing.
gas, which contains water and heat released by the patient, The optimal AH values in the airways, especially in
and the other source is water vapor and heat released from children, are not always clearly defined in the literature.
the CO2 absorbent in an exothermic reaction.17 The effi- There is no minimum requirement for humidification per-
ciency of this humidity conservation depends on various formance in children in HME standards.19 Nevertheless, the
factors, including the FGF used, configuration of the breath- mean AH of the inhaled gases in the groups with an HME,
ing system, and patient characteristics. regardless of the FGF used, were always >20 mg H2O·L−1,
With a low FGF, the circle system permits a higher which is the value most frequently cited as the threshold
rebreathing of exhaled gases and, consequently, a higher necessary for preventing the deleterious effects of dry gas
humidity of the inhaled gases is expected. Our data con- on the epithelium of the airways during anesthesia.3,5,6 The
firmed this assumption, which is in accordance with another mean AH of the inhaled gases in the no-HME groups was
study that showed the effect of a low FGF in increasing the always <20 mg H2O·L−1.
humidity of inhaled gases during pediatric anesthesia.12 Few studies evaluated the efficiency of an HME dur-
In the breathing system of the Primus workstation, the ing pediatric anesthesia. A study in young children using
patient’s exhaled gases move through a hotplate and across an HME in the ventilatory system, with no rebreathing of
the soda lime once before mixing with the dry and cold FGF. exhaled gas, showed an inhaled gas AH of 22 mg H2O·L−1.15
Then the mixed gases are pulled by the ventilator plunger These results are in agreement with our data, because the
to fill the ventilator. When the inspiratory valve opens, the group with an HME and high FGF, which allowed for no
ventilator’s plunger sends the gaseous mixture into the or minimal rebreathing, had a mean inspiratory AH of 23
inspiratory limb of the breathing circuit.10 Anesthetic work- mg H2O·L−1. Another study was performed in children
stations, such as the Cicero and Cato (Dräger), also have using an HME in the breathing system during 140 minutes
built-in hotplates to heat the exhaled gases; however, unlike of ventilation.20 This study also did not allow rebreath-
the Primus, the exhaled gases mixed with a dry and cold ing. The authors showed a lower mean inhaled gas tem-
FGF pass through the CO2 absorber twice per breath.18 In perature (28.4ºC ± 1.5ºC) and a higher mean AH (26.2 ± 1.7
children, a study with a low FGF without an HME from mg H2O·L−1) compared with our data in the HME groups.
the Cicero workstation found higher inhaled gas humid- Differences between the HME used, the configuration of the
ity (22 mg H2O·L−1) when compared with our data.12 The breathing systems, and the time of ventilation may explain
difference in the humidity values is likely a result of the the different results between the studies. In addition, our
different breathing circuit configurations of the anesthesia data showed that the FGF has a significant correlation with
workstations. the HME efficiency and that humidity values are increased
The hotplate in the circle system of the Primus work- with low flow.
station kept the gases warm in the inspiratory limb close Two factors may explain the difference in the RH val-
to the inspiratory valve (approximately 30ºC–31ºC) in all ues between the gases in the inspiratory limb outlet (close
the groups. When an HME was not used, the inhaled gas to the anesthesia workstation) and the inhaled gases in the
temperatures were lower (approximately 3ºC) than the gas no-HME groups. First, the large difference between the OR
temperatures in the inspiratory limb close to the anesthesia and inspiratory limb outlet temperatures (approximately
workstation. A study showed that the temperature is not 8ºC) caused water condensation in the inspiratory limb of
maintained in the inspiratory limb of the circle breathing the breathing system. Second, a gas with a low RH has a
circuit during high- and low-flow anesthesia in infants.13 high thermal energy for absorbing the condensate water
According to these authors, the speed and amount of the along the inspiratory limb.21 In this way, the breathing sys-
heat loss are directly related to the heat transfer coefficient tem itself worked as a moisture exchanger, although with
of the material and the length of the inspiratory limb of the low efficiency, and it reached the humidity levels below
breathing circuit. These findings are in accordance with our the threshold believed to preserve the ciliary function of
data, i.e., a positive correlation between the OR and inhaled airways.
gas temperatures in the no-HME groups. The use of corru- Children undergoing anesthesia are prone to develop-
gated tubes with thermal insulation materials could have ing hypothermia because of their large surface area to body
maintained the temperature of the gases in the inspiratory weight ratio.20 The mean esophageal temperature values
limb of the breathing system. were similar and remained above 36ºC during the entire
In adults, a study using the Dräger Primus showed an procedure in all the groups. The use of an active warming
inhaled gas humidity of approximately 20 mg H2O·L−1 dur- system associated with adequate covering of patients with
ing anesthesia with low FGF.10 In our study, children with surgical drapes may have minimized the heat loss, thus
the low-flow breathing system of the Primus without an explaining the maintenance of normothermia during the
HME (1L group) had lower mean inhaled gas humidity (17 whole period studied. One study showed that the use of an
mg H2O·L−1). Two factors may explain the difference in the HME did not prevent intraoperative hypothermia in chil-
inhaled gas humidity. First, because children have lower dren, but it did minimize it.14
absolute minute volumes than adults, they provide less There are some limitations to our study. First, the study
CO2 to react with the soda lime, which results in the gen- was not blinded. The gas humidity and temperature mea-
eration of less heat and water in the circle system.13 Second, surements were made intermittently at predefined times.
children produce a smaller exhaled gas volume relative to Then it was necessary to place the thermo-hygrometer

930   
www.anesthesia-analgesia.org anesthesia & analgesia
probe inside the T-piece for each measurement. In the Name: José R. C. Braz, MD, PhD.
groups using an HME, the T-piece was connected between Contribution: This author helped design the study, conduct the
the HME and the tracheal tube. Thus, due to the proximity, study, analyze the data, and write the manuscript.
it was not possible to conceal the presence of an HME in Attestation: José R. C. Braz has seen the original study data,
the respiratory circuit during the measurements. Second, reviewed the analysis of the data, and approved the final
the humidity values were recorded until 80 minutes after manuscript.
the breathing circuit had been connected to the patients. This manuscript was handled by: Peter J. Davis, MD.
This recording period was compatible with typical pedi- REFERENCES
atric surgeries, which are often of short or medium dura- 1. Irlbeck D. Normal mechanisms of heat and moisture exchange
tion. A study using the Primus machine in adults verified in the respiratory tract. Respir Care Clin N Am 1998;4:189–98
an increase in the humidity of the inhaled gases after 90 2. Van Oostdam JC, Walker DC, Knudson K, Dirks P, Dahlby RW,
minutes of ventilation with low FGF with and without an Hogg JC. Effect of breathing dry air on structure and function
of airways. J Appl Physiol (1985) 1986;61:312–7
HME.10 Increases in the humidity of inhaled gases could 3. Martins RH, Braz JR, Defaveri J, Curi PR. Study of humidifica-
have occurred after the time period studied. Because of the tion and heating of inspired gases during mechanical ventila-
short duration of most pediatric surgeries, a device that tion in dogs. Rev Bras Otorrinolaringol 1996;62:206–17.
properly humidifies and warms the inhaled gases should 4. Branson RD, Campbell RS, Davis K, Porembka DT. Anaesthesia
be effective early in the procedure.15 The introduction of circuits, humidity output, and mucociliary structure and func-
tion. Anaesth Intensive Care 1998;26:178–83
an HME into the circuit produced an increase in humidity 5. Kleemann PP. Humidity of anaesthetic gases with respect to
and temperature of the inhaled gases beginning at the first low flow anaesthesia. Anaesth Intensive Care 1994;22:396–408
­measurement (10 minutes). 6. Wilkes AR. Heat and moisture exchangers and breathing sys-
tem filters: their use in anaesthesia and intensive care. Part
1—history, principles and efficiency. Anaesthesia 2011;66:31–9
CONCLUSIONS 7. Kleemann PP. The climatisation of anesthetic gases under
The use of neither high nor low FGF in the pediatric circle conditions of high flow to low flow. Acta Anaesthesiol Belg
breathing circuit of the Primus machine provided the mini- 1990;41:189–200
mum humidity level of the inhaled gases thought to reduce 8. Henriksson BA, Sundling J, Hellman A. The effect of a heat and
moisture exchanger on humidity in a low-flow anaesthesia sys-
the risk of dehydration of airways. Insertion of an HME
tem. Anaesthesia 1997;52:144–9
increases the humidity and temperature of the inhaled gases, 9. Yamashita K, Yokoyama T, Abe H, Nishiyama T, Manabe M.
bringing them close to physiological values. The FGF has a Efficacy of a heat and moisture exchanger in inhalation anes-
significant correlation with HME efficiency and increases thesia at two different flow rates. J Anesth 2007;21:55–8
humidity values when low flow is used. Therefore, the asso- 10. Castro J Jr, Bolfi F, de Carvalho LR, Braz JR. The tempera-
ture and humidity in a low-flow anesthesia workstation with
ciation of an HME with low FGF in the breathing circuit is and without a heat and moisture exchanger. Anesth Analg
the most efficient way to conserve the heat and moisture of 2011;113:534–8
the inhaled gas during pediatric anesthesia. E 11. Meakin GH. Low-flow anaesthesia in infants and children. Br J
Anaesth 1999;83:50–7
12. Igarashi M, Watanabe H, Iwasaki H, Namiki A. Clinical evalua-
DISCLOSURES tion of low-flow sevoflurane anaesthesia for paediatric patients.
Name: Gustavo P. Bicalho, MD. Acta Anaesthesiol Scand 1999;43:19–23
Contribution: This author helped design the study, conduct the 13. Hunter T, Lerman J, Bissonnette B. The temperature and humid-
study, analyze the data, and write the manuscript. ity of inspired gases in infants using a pediatric circle system:
Attestation: Gustavo P. Bicalho has seen the original study data, effects of high and low-flow anesthesia. Paediatr Anaesth
reviewed the analysis of the data, approved the final manu- 2005;15:750–4
14. Bissonnette B, Sessler DI. Passive or active inspired gas humidi-
script, and is the author responsible for archiving the study files. fication increases thermal steady-state temperatures in anesthe-
Name: Leandro G. Braz, MD, PhD. tized infants. Anesth Analg 1989;69:783–7
Contribution: This author helped conduct the study. 15. Monrigal JP, Granry JC. The benefit of using a heat and mois-
Attestation: Leandro G. Braz has seen the original study data ture exchanger during short operations in young children.
and approved the final manuscript. Paediatr Anaesth 1997;7:295–300
Name: Larissa S. B. de Jesus, MD. 16. Barra Bisinotto FM, Braz JR, Martins RH, Gregório EA, Abud
TM. Tracheobronchial consequences of the use of heat and
Contribution: This author helped conduct the study. moisture exchangers in dogs. Can J Anaesth 1999;46:897–903
Attestation: Larissa S. B. de Jesus has seen the original study 17. Foregger R. The regeneration of soda lime following absorption
data and approved the final manuscript. of carbon dioxide. Anesthesiology 1948;9:15–20
Name: Cesar M. C. Pedigone. 18. Wada H, Higuchi H, Arimura S. Temperature and humid-
Contribution: This author helped conduct the study. ity of the Dräger Cato anesthetic machine circuit. J Anesth
Attestation: Cesar M. C. Pedigone approved the final manuscript. 2003;17:166–70
19. International Standard Organization. Anaesthetic and
Name: Lídia R. de Carvalho, PhD.
Respiratory Equipment - Heat and Moisture Exchangers
Contribution: This author helped design the study and analyze (HMEs) Humidifying Respired Gases in Humans—part
the data. 1: HMEs for Use Minimum Tidal Volumes of 250 ml. ISO
Attestation: Lídia R. de Carvalho has seen the original study 9360–1. Geneva, Switzerland: International Organization for
data, reviewed the analysis of the data, and approved the final Standardization, 2000
manuscript. 20. Luchetti M, Pigna A, Gentili A, Marraro G. Evaluation of the
efficiency of heat and moisture exchangers during paediatric
Name: Norma S. P. Módolo, MD, PhD.
anaesthesia. Paediatr Anaesth 1999;9:39–45
Contribution: This author helped conduct the study. 21. Miyao H, Hirokawa T, Miyasaka K, Kawazoe T. Relative humid-
Attestation: Norma S. P. Módolo has seen the original study ity, not absolute humidity, is of great importance when using a
data and approved the final manuscript. humidifier with a heating wire. Crit Care Med 1992;20:674–9

October 2014 • Volume 119 • Number 4 www.anesthesia-analgesia.org 931

You might also like