Professional Documents
Culture Documents
397 Full
397 Full
_l
FIG. 1. Arrangements of the circle system for laboratory and clinical studies. 1 = Wright respirometer;
2 = end-tidal gas sample site; 3 = expiratory unidirectional valve; 4 = inspiratory unidirectional valve;
5 = fresh gas supply; 6 = soda-lime cannister; 7 •= to paramagnetic oxygen analyser (clinical); 8 =
vented gas sample site (volunteer); 9 = shrouded Heidbrink valve; 10 = reservoir bag; 11 = patient.
In comparing figure 2 and figure 3 it is seen that of oxygen (4-5%) were attributable to the
the increase in argon fraction is closely comple- accumulation of argon, it would represent a
mented by the decrease in oxygen fraction at each doubling of the argon fraction in the fresh gas
fresh gas flow setting. However, when the circle supply.
was fully closed the decrease in oxygen fraction One may speculate on the use of nitrous oxide
was greater than the increase in argon fraction—a in low flow anaesthetic techniques with concen-
difference approaching 8% after 80 min; the trator oxygen. The margin of safety is greatly
discrepancy is accounted for by the observed reduced by the adoption of a baseline gas
increase in nitrogen fraction from 1 % to 9 % over composition which includes perhaps 33 % oxygen,
this period. This represents an accumulation of when a degree of argon accumulation, which
about 700 ml of nitrogen and falls within the might be unimportant whilst using oxygen as the
range of results reported by previous workers in sole carrier gas, would threaten a serious decrease
man [7] and the dog [8]. in oxygen fraction. However, if the fresh gas used
It was noted in the laboratory study that there for the maintenance of anaesthesia consisted of
were consistent small discrepancies between the oxygen 500 ml min"1 and nitrous oxide 500 ml
fractions of argon and oxygen in end-tidal and min"1, then the absolute degree of argon accumu-