Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Br. J. Anaesth.

(1988), 61, 397-402

PREDICTED AND MEASURED OXYGEN


CONCENTRATIONS IN THE CIRCLE SYSTEM USING LOW
FRESH GAS FLOWS WITH OXYGEN SUPPLIED BY AN
OXYGEN CONCENTRATOR

C. J. R. PARKER AND S. L. SNOWDON

Oxygen concentrators are being used increasingly


to supply the requirements for medical oxygen, SUMMARY

Downloaded from http://bja.oxfordjournals.org/ by guest on April 12, 2016


and have been installed in a limited number of An oxygen concentrator based on a zeolitic
District General Hospitals [1]. The oxygen con- molecular sieve has been used to supply piped
centrator uses a zeolitic molecular sieve to separate oxygen in a District General Hospital; such
oxygen from nitrogen in atmospheric air; the oxygen contains 5% argon. If concentrator
affinity of the zeolite for argon and oxygen is oxygen is used in the closed circuit, argon
similar. Since the atmosphere contains 0.93% accumulates. The present investigation defines
argon, and purification of oxygen from air repre- the extent and time course of argon accumulation
sents a five-fold concentration, such oxygen in partially and fully closed breathing systems. A
inevitably contains about 5 % argon. In addition, theoretical model is presented which predicts the
should there be malfunction of one of the extent and rate of inert gas accumulation. The
concentrator modules, nitrogen would appear in predictions have been tested in a volunteer under
the product gas. In preliminary studies on the laboratory conditions and in five anaesthetized
concentrator installed at the Countess of Chester patients. It is recommended that concentrator
Hospital, up to 3.7 % nitrogen has been observed oxygen may be used for low flow anaesthesia of
in the product gas and, on occasion, the delivered indefinite duration, provided the fresh oxygen
oxygen fraction decreased to 92.9% [2]. Whilst flow to the circuit is at least twice the oxygen
such a degree of purity might be acceptable in consumption. If the circuit is totally closed,
most areas of clinical practice, this may not be the periodic opening of the circuit is necessary.
case during low flow and closed circuit anaes-
thesia, where selective oxygen uptake by the
patient will lead to the accumulation of argon in
the breathing system. A theoretical treatment is described in the
The present study addresses the problem of the Appendix.
accumulation of inert gas in the breathing system
SUBJECTS AND METHODS
with low fresh gas flows. Initial studies were
carried out in the laboratory, using a human All anaesthetic machines and circle systems used
volunteer to simulate oxygen consumption by the in the present study were tested for leaks on the
patient. Later studies were performed during day of use; the maintenance of a pressure of 40 cm
routine low flow anaesthesia to examine the effect HjO in the apparatus after cessation of gas input
of the accumulation of inert gas on oxygen was confirmed using an aneroid pressure gauge.
concentration in the breathing system in clinical The circle systems were arranged as shown in
practice. figure 1, so that gas was vented after carbon
dioxide absorption.
C. J. R. PARKER, M_A., M.B., B.CHIR., F.F.A.R.C.S.; S. L. SNOW-
DON, M.B., CH.B., F.F.A.R.CS. ; University Department of Laboratory study
Anaesthesia, Royal Liverpool Hospital, Prescot Street, P.O.
Box 147, Liverpool L69 3BX. Accepted for publication: Laboratory studies used an anaesthetic machine
March 14, 1988. with flowmeters calibrated volumetrically for the
398 BRITISH JOURNAL OF ANAESTHESIA

_l

FIG. 1. Arrangements of the circle system for laboratory and clinical studies. 1 = Wright respirometer;
2 = end-tidal gas sample site; 3 = expiratory unidirectional valve; 4 = inspiratory unidirectional valve;
5 = fresh gas supply; 6 = soda-lime cannister; 7 •= to paramagnetic oxygen analyser (clinical); 8 =
vented gas sample site (volunteer); 9 = shrouded Heidbrink valve; 10 = reservoir bag; 11 = patient.

Downloaded from http://bja.oxfordjournals.org/ by guest on April 12, 2016


gas mixture used. The circle system was a BOC measured using an electronic Wright respirometer
Mk3. Gas analysis was performed using a mass and displayed on a chart recorder.
spectrometer (BOC MS2), including a multi- After a period of 5 min when the subject
plexed output unit set to atomic mass units (u) 28, breathed through the circle system with a fresh
32 and 40. Electrical zero was established from the gas flow rate of 8 litre min"1, to diminish pul-
output at u 10 and a full calibration against monary nitrogen stores, the fresh gas flow was
certified gas mixtures of medical oxygen, " white reduced to a predetermined rotameter setting.
spot" nitrogen and argon was performed before Gas was sampled from the breathing system at
and after each study. Calibrations were verified three points. Intermittent samples were taken at
frequently during the study and correction made the mouthpiece and the fresh gas inlet for a period
for any observed drift in performance. The of about 30 s every 2-5 min, depending on the
observed drift was less than 5 % of the measure- expected rate of change of system gas composition.
ment range throughout the study. Gas samples The regular sampling of fresh gas facilitated
were collected at a flow rate of 30 ml min"1 correction of any performance drift of the mass
through a sample tube heated to a temperature spectrometer. At all other times vented gas was
greater than 40 °C to prevent condensation of sampled, except at the lowest fresh gas flow used
water vapour and, hence, maintain vapour pres- when the circle was closed. Gas samples were
sure. analysed for oxygen, argon and nitrogen and the
A cylinder containing 9.5% argon in oxygen results were displayed on a chart recorder.
was prepared by the decantation of argon at high The volume of the circle system and the
pressure into a partially empty cylinder of medical functional residual capacity of the subject were
oxygen; the process was controlled by regulation each measured by dilution, using argon as a
of the filling pressure, and the approximate marker gas.
composition guided by the partial pressures of the
component gases. The composition of the result- Clinical study
ing mixture was then measured accurately by the A Penlon IM500 anaesthetic machine, with a
use of the mass spectrometer. The composition Fluotec Mk2 Type MJ halothane vaporizer
was chosen to simulate the worst condition of (Cyprane) (calibrated 0-10%) placed outside the
oxygen dilution by inert gas likely to be found in circle system was used. Apparatus used for gas
clinical practice. The gas was used to supply the analysis comprised a paramagnetic oxygen ana-
fresh gas requirement of a conscious, postprandial lyser (Servomex type OA 101 Mk2) calibrated on
non-smoking human adult seated at rest and each day of the study and an infra-red absorption
breathing through a mouthpiece attached to a capnometer (Datex "Normocap"). The para-
circle system; a noseclip was used to eliminate magnetic oxygen analyser included the facility for
nasal breathing. The subject's tidal volume was drying of sampled gas by passage over silica gel
LOW FLOWS AND THE OXYGEN CONCENTRATOR 399
before entry to the analytical cell. The purity of 40i
the concentrator oxygen was confirmed on each
day of the study against a cylinder of pure
oxygen. 35
The investigation was approved by the Hospital
Ethics Committee. Patients scheduled to undergo
surgery to varicose veins were studied; all were
30
free from cardiac, respiratory and renal disease.
None had received halothane anaesthesia within
the previous 3 months.
25
Premedication was with diazepam 10 mg by
mouth. After the establishment of venous access,
and pretreatment with tubocurarine 3 mg, anaes-
thesia was induced with thiopentone and followed 20

by suxamethonium 100 mg. The trachea was


sprayed with 4% lignocaine solution and intu-

Downloaded from http://bja.oxfordjournals.org/ by guest on April 12, 2016


bated. Spontaneous ventilation was established 15
and the patient allowed to breathe halothane in
oxygen from the circle system, with a fresh gas
flow rate of 10 litre min"1. After a period of about 10
20 min, when the oxygen fraction in sampled gas
had stabilized, the fresh gas flow rate was reduced
to 500 ml min"1, and the vaporizer setting was
adjusted to maintain stable anaesthesia, an initial
value of 8% being typical [3]. The tidal volume
was measured with a Wright respirometer.
Gas sampling in the clinical study was confined 20 40 60 80 100 120
to two sites. A continuous sample was taken from Time (min)
near the reservoir bag to the oxygen analyser at a FIG. 2. Predicted ( ) and measured ( • A A O I ) argon per-
rate of 15 ml min"1; the deadspace of the sample centages in end-tidal gas in the laboratory studies. Fresh gas
tube was 4.5 ml. Intermittent samples were taken flows: • = 310mlmin-'; A = 500 ml min"1; A = 700 ml
from the catheter mount for the measurement of min"1; O = 900 ml min- 1 ; • = 2000 ml mirr1. Predictions
end-tidal carbon dioxide concentration—to facili- based on an estimated oxygen consumption of 310 ml min"1
ATPS.
tate routine monitoring of the patient.

functional residual capacity of the volunteer was


RESULTS
3250 ml.
Details of the volunteer and of the patient are
shown in table I. The volume of the circle system, Laboratory study
including the reservoir bag was 5630 ml; the The argon fractions in end-tidal gas are shown
in figure 2; at each fresh gasflowsetting the argon
fraction increased with time. Both the rate and
TABLE I. Details of sex, age, weight and tidal volume of the sub- extent of the increase in argon fraction were
jects studied
greater as the fresh gas flow was reduced. At the
Tidal lowest fresh gas flow, when the circle was fully
Age Weight volume closed, the increase in end-tidal argon fraction
Subject Sex (yr) (kg) (ml) was linear—about 0.3 % min"1. At all higher flow
Volunteer M 30 92 728 rates the increase had the form of a wash-in
Patient 1 F 37 57 370 exponential function.
Patient 2 F 66 90 — The argon fractions in end-tidal and vented gas
Patient 3 M 58 85 450 were similar. As the argon fraction approached its
Patient 4 F 45 66 340
Patient 5 F 22 58 390
steady value, the argon fraction in vented gas
exceeded that in end-tidal gas by 0.5-1 %.
400 BRITISH JOURNAL OF ANAESTHESIA
85 flow rates, the increase in nitrogen fraction was
much smaller and never exceeded 3 % of end-tidal
,•:
gas.
6 o o o
75 4 4 4 Clinical study
The measured purity of concentrator oxygen
ranged from 94.3 % to 95.1 % at the times of the
65 study.
The oxygen fractions measured in gas sampled
near the expiratory valve are shown in figure 4. In
each patient there was a decrease during the
55 period of measurement of between 3.8% and
5.1 %. There was a delay of about 5 min before
the rate of decrease in oxygen fraction was
maximal, at around 0.17% min"1. Thereafter, in
45
20 40 80 100 120 each patient, the rate of decrease in oxygen

Downloaded from http://bja.oxfordjournals.org/ by guest on April 12, 2016


Time (min) fraction decreased with time and the oxygen
FIG. 3. Measured oxygen percentage in end-tidal gas in the fraction approached a new steady value.
laboratory studies. Fresh gas flows: • = 310 ml min"1; A =
500 ml min" 1 ; A = 700 ml min"1; O = 900 ml min"1; B =
2000 ml min"1. DISCUSSION

95 The presence of small concentrations of argon in


inspired gas appears to be of no biological
consequence. Argon has anaesthetic properties at
a partial pressure of about 15 atmospheres in a
variety of species, and it is likely that this is the
case in man [4]. However, prolonged inhalation of
a mixture of 80 % argon-20 % oxygen at atmos-
pheric pressure in the rat appears to have no effea
90 [5], and exposure of experienced divers to a
6 similar mixture produced no detectable effect [6].
* » 4
». " 4 » Thus the major problem arising from the accumu-
lation of argon would seem to be simple encroach-
ment upon the partial pressure of oxygen or
anaesthetic gases.
Theoretical considerations detailed in the Ap-
85 pendix predict the extent of argon accumulation
10 20 30 40 50 60 in the breathing system. Of the determinants of
Time (min) argon fraction, the anaesthetist has the greatest
FIG. 4. Measured oxygen percentages in gas vented from the control over the fresh gas flow. It is predicted
circle system in clinical studies using a fresh gas flow of that, in the totally closed circuit, the argon fraction
500 ml min"1. • = Patient 1; A = pati e n t 2; • = patient 3;
O = patient 4; A = patient 5.
will increase linearly whereas, at low flows with
some gas vented from the circuit, the argon
End-tidal oxygen fractions are shown in figure fraction should follow a wash-in exponential
3. The oxygen fraction decreased with time, at a function.
rate and to an extent related to the fresh gas flow. Figure 2 details predictions of this model which
Oxygen fractions in vented and end-tidal gas parallel the experimental observations. Although
decreased in parallel; the fraction in end-tidal gas there is some discrepancy in the absolute values,
was consistently less by about 5 %. laboratory results follow the pattern of the
Only when the circle was fully closed did the predictions. It is reasonable to suppose that the
end-tidal nitrogen concentration increase through- discrepancies between the predictions and the
out the period of observation; it increased from an experimental observations reflect, to some extent,
initial 1 % to reach 9 % after 80 min. At higher limitations of the model or of its quantification.
LOW FLOWS AND THE OXYGEN CONCENTRATOR 401

In comparing figure 2 and figure 3 it is seen that of oxygen (4-5%) were attributable to the
the increase in argon fraction is closely comple- accumulation of argon, it would represent a
mented by the decrease in oxygen fraction at each doubling of the argon fraction in the fresh gas
fresh gas flow setting. However, when the circle supply.
was fully closed the decrease in oxygen fraction One may speculate on the use of nitrous oxide
was greater than the increase in argon fraction—a in low flow anaesthetic techniques with concen-
difference approaching 8% after 80 min; the trator oxygen. The margin of safety is greatly
discrepancy is accounted for by the observed reduced by the adoption of a baseline gas
increase in nitrogen fraction from 1 % to 9 % over composition which includes perhaps 33 % oxygen,
this period. This represents an accumulation of when a degree of argon accumulation, which
about 700 ml of nitrogen and falls within the might be unimportant whilst using oxygen as the
range of results reported by previous workers in sole carrier gas, would threaten a serious decrease
man [7] and the dog [8]. in oxygen fraction. However, if the fresh gas used
It was noted in the laboratory study that there for the maintenance of anaesthesia consisted of
were consistent small discrepancies between the oxygen 500 ml min"1 and nitrous oxide 500 ml
fractions of argon and oxygen in end-tidal and min"1, then the absolute degree of argon accumu-

Downloaded from http://bja.oxfordjournals.org/ by guest on April 12, 2016


vented gas. When the argon fraction in the lation would be no greater than in the present
breathing system had stabilized, it was 0.5-1 % clinical study. Indeed, it would be less, since the
higher in vented than in end-tidal gas, whereas use of nitrous oxide dilutes the argon fraction in
the oxygen fraction was typically 5 % higher in the fresh gas to a value typically around 2.5%.
vented gas. These discrepancies are not un- Furthermore, a substantial volume of nitrous
expected for, whilst end-tidal gas contains about oxide is vented after the first few minutes of its
5% carbon dioxide, the vented gas had been uptake, so that the total rate of gas venting from
subjected to carbon dioxide absorption; further- the circle system is higher than in the present
more, whilst vented gas is at least partially clinical study. The argon fraction in the circle
saturated with water vapour at ambient tempera- would be expected to remain less than 5 % and
ture, end-tidal gas collected through a heated this might represent an acceptable dilution of
sample tube can be expected to be significantly respirable gas.
more humid. In addition to the dilution of Certain recommendations are justified con-
measured end-tidal gas fractions by carbon di- cerning the use of concentrator oxygen in low flow
oxide and a greater degree of water vapour, end- anaesthesia. If nitrous oxide is omitted, serious
tidal gas represents a sample from which oxygen argon accumulation does not occur if the fresh gas
consumption by the subject ensures a lower flow is about twice the oxygen consumption, and
oxygen fraction than in the vented gas, which is the clinical study supports the use of fresh oxygen
more representative of the gas in the circle system flows of 500 ml min"1 in anaesthetized patients.
as a whole. Fully closed circuit anaesthesia may also be used,
In the clinical study, the accumulation of but argon will continue to accumulate indefinitely
nitrogen was minimized by a long period of until the circle is opened. In the laboratory study,
denitrogenation [9] and a fresh gas flow at which the argon fraction had quadrupled after about
laboratory studies had shown that a large increase 90 min and it would seem wise to open the circuit
in nitrogen fraction does not occur. Any contri- with at least this frequency. Whilst the present
bution of possible changes in carbon dioxide and experimental data do not extend to the use of a
water vapour fraction within the breathing system low flow anaesthetic technique incorporating
to the observed changes in vented oxygen fraction nitrous oxide with concentrator oxygen, the
was minimized by measuring oxygen fraction in accumulation of argon in the breathing system
gas vented after carbon dioxide absorption, and should not be hazardous, provided the fresh gas
dried before analysis. mixture includes at least 50% oxygen at a total
Assuming the oxygen consumption of the flow rate of no less than 1 litre min"1.
patients to have been about 250 ml min"1, that is
approximately 50 % the fresh gasflowrate, theory APPENDIX
would predict an increase in the fraction of argon Certain assumptions simplify the theoretical treatment. Simi-
in the breathing system to twice that in the fresh lar assumptions have been made elsewhere to facilitate analysis
gas. If all of the observed decrease in the fraction of the behaviour of the nearly closed circuit [10].
402 BRITISH JOURNAL OF ANAESTHESIA
(1) Argon is not taken up by the subject. In fact, the effective Therefore:
capacity of the body for argon is only about 3 litre at 1
atmosphere. The fat group of tissues accounts for 2/3 of this Vx dFarg
(Farg, x V,) - Farg, x (V, - Vot)
volume and equilibrates with a time constant of about 5 h [11], dt
and so body uptake remains a small fraction of argon
accumulation in the gaseous phase in the present studies. This has the solution:
Farg, = Farg l n ,-(Farg, n f -Farg,) x exp"*1
(2) Carbon dioxide is completely and instantaneously ab-
sorbed upon entry to the system. The circle systems used in Farg,xK,
the present study were arranged to vent gas after carbon =
dioxide absorption, so the equations for gaseous homeostasis
are unchanged.
(3) The functional residual capacity of the subject and the
circle system form a homogeneous unit, considered to exist at It can readily be seen that the accumulation of argon follows
ATPD. In effect, the gas in the circle and lungs will be mixed a wash-in exponential function [12].
over a time course much less than that of the accumulation of The predictions of this analysis, corrected to ATPS, are
argon; as has been discussed, the observed small differences presented, together with experimental observations, in figure
between argon fractions in the vented and end-tidal gas are the 2.
result of the presence of carbon dioxide and the greater
humidity of the latter.

Downloaded from http://bja.oxfordjournals.org/ by guest on April 12, 2016


REFERENCES
(4) The subject is initially totally denitrogenated. In the 1. Howell RSC. Oxygen concentrators. British Journal of
laboratory study, nitrogen accumulation was noted only when Hospital Medicine 1985; 34: 221-223.
the circuit was fully closed, and in clinical studies, denitro- 2. Chester Health Authority. An Evaluation of the Cascade
genation was prolonged. Oxygen Concentrator Installed to Supply Oxygen at the
Countess of Chester Hospital. Chester: Chester Health
Authority, 1984.
3. Mushin WW, Galloon S. The concentration of anaes-
SYMBOLS thetics in closed circuits with special reference to halo-
thane III. Clinical aspects. British Journal of Anaesthesia
V = Volume of the circle system + functional residual capacity.
1960; 32: 324-333.
Vot = Oxygen consumption by the subject.
4. Ikels KG, Adams JD. Molecular sieve oxygen generating
V, = Dry fresh gas input to the breathing system.
system: the argon question—a brief review. Aviation,
Farg, = Argon fraction in the dry fresh gas input.
Space and Environmental Medicine 1979; 50: 939-942.
Farg, = Argon fraction in the breathing system at time t.
5. Aldrete JA, Virtue RW. Prolonged inhalation of inert
dFarg = Change in argon fraction in the breathing system
gases by rats. Anesthesia and Analgesia 1967; 46: 562-565.
during time dr.
6. Behnke AR, Yarbrough DD. Respiratory resistance,
Farglnf = Argon fraction in the breathing system after infinite
oil-water solubility, and mental effects of argon compared
time.
with helium and nitrogen. American Journal of Physiology
t = time. 1939; 126: 409-415.
dr = a short time interval.
7. Barton F, Nunn JF. Totally closed circuit nitrous oxide/
k = rate constant for argon accumulation.
oxygen anaesthesia. British Journal of Anaesthesia 1975;
V,Vot,V, considered to exist at ATPD.
47: 350-357.
8. Shaw LA, Behnke AR, Messer AC, Thomson RM,
Motley EP. Equilibrium time of the gaseous nitrogen in
the dog's body following changes of nitrogen tension in
ANALYSIS the lungs. American Journal of Physiology 1935; 112:
Volume of gas vented in time dt = 545-553.
9. Morita S, Latta W, Hamtro K, Snider MT. Accumulation
of methane, acetone and nitrogen in the inspired gas
Volume of argon vented in time di = during closed-circuit anesthesia. Anesthesia and Analgesia
1985; 64: 343-347.
dtxFaig,x(V,-Vo,) 10. Holmes CMcK, Spears GFS. Very-nearly-closed-circuit
Volume of argon entering the system in time dt = anaesthesia—a computer analysis. Anaesthesia 1977; 32:
846-857.
dt x Farg, x V,
11. Bell CMJ, Leach MO. A compartmental model for
Volume of argon accumulating in time di = investigating the influence of physiological factors on the
rate of washout of l u X e and *'Ar from the body. Physics
dFarg x V
in Medicine and Biology 1982; 27: 1105-1117.
In unit time argon accumulating = argon entering 12. Nunn JF. Applied Respiratory Physiology, 2nd edn.
argon vented. London: Butterworths, 1977; 469.

You might also like