CO2 Fractional Vaginal Tightenning Laser User's Manual

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 25

CO2 Fractional Laser

User’s Manual
1. Equipment Introduction

The CO2 fractional laser treatment instrument, a new generation of laser


skin resurfacing system, also has the functions of super pulse and laser
scanning output, which can quickly and accurately perform various fine laser
surgery, especially suitable for human body plastic surgery and facial
cosmetic surgery. The system is equipped with a high-speed graphics
scanner that can scan and output graphics of different shapes, and can
provide personalized treatment plans according to the needs of different
patients.
With the emergence of innovative treatments for microvascular
reconstruction with CO2 laser, CO2 laser provides a revolutionary set of
solutions for common problems in women such as vaginal relaxation, internal
vaginal environment disorder, poor sensitivity, or urinary incontinence.
Precise painless solution: no need for anesthesia, no pain throughout the
course, no holidays, 15 minutes to complete the treatment, can maintain the
firming effect for a long time, is the most popular new body shaping project for
women today

2. Operational safety regulations

2.1 Optical security

1.Burn

The CO2 laser scanning laser has a wavelength of 10.6 microns, which is a
spectral line in the far infrared range, which is invisible to the human eye. The
maximum laser power output by this instrument can reach 30 watts, which
can cause 3 degree burns even without focusing. Therefore, sufficient
attention should be paid.

2.The hazard of reflection and direct exposure to human eyes

The instrument outputs visible red semiconductor laser and invisible CO2
laser, both of which are harmful to the human body.

Do not look directly at the red semiconductor laser at any time, even if it
does not cause burns to the human eyes, it may cause damage to a certain
degree.

The damage distance of CO2 laser is very large, direct


exposure to human eyes will cause blindness, and it is
irrecoverable, operators should be extra careful.

Since the surface of the object, especially the surface of smooth metal or
glass, can form a mirror surface to reflect light, it is important to remove such
objects or surfaces that may reflect on the path of the laser. In other words, do
not irradiate laser light onto objects that reflect light to prevent the reflected or
scattered laser light from causing harm to the human body.

2.2 Flammable and explosive

Do not use this laser instrument on the site with flammable and explosive
materials. Do not place or store flammable and explosive materials around
the instrument. Flammable and explosive materials include: gasoline, alcohol,
some narcotics, some solvents, desiccants, ointments, synthetic resins, etc.

2.3 Safe operation steps

1. Before treatment ‘
1. Keep the surface of the instrument clean.
2. Get the instrument key.
3. Ask the surgeon how to arrange the position of the instrument and the
patient.
4. Stick a laser warning sign in a conspicuous position.
5. If you need to use general anesthetics, prepare a wet towel for the
patient.
6. Prepare all necessary equipment and tools.
7. Check whether the instrument is working properly.
8. After inserting the switch key into the instrument's key, power on and
turn on the instrument. The instrument should perform self-check first, and
then enter the standby working state after selecting the working mode.
9. If necessary, press the setting key to reset the working parameters.
10. Check the same optical path.
11. Press the set button to return to the standby working state or turn off
the instrument until the operation is turned on again.
2. After treatment
1. Turn off and remove the instrument.
2. Pull out the switch key and keep it in a safe place.
3. Place the light guide joint arm in the normal position.
4. If necessary, remove the optical accessories and other tools for
cleaning or disinfection.

3. Accessories list

CO2 main unit Guide light arm Power cord

Key Foot switch Fractional laser head

Vaginal tightening head pulse head


4. Principle:
The laser is emitted in a scanning lattice method to form a burning area
composed of laser action lattices and intervals on the epidermis. Each laser
action point is composed of a single or several high-energy laser pulses,
which can penetrate directly to the dermis layer in an instant It vaporizes the
tissues in wrinkles or scars and stimulates the proliferation of collagen, which
in turn initiates a series of skin reactions such as tissue repair and collagen
rearrangement. Under the action of the laser, collagen fibers shrink by about
one-third, small wrinkles are flattened, deep wrinkles become shallower and
thinner, and the skin becomes firm and shiny.
Laser through the CO2 laser's microvascular reconstruction effect, the
oxygen content of vaginal tissue will increase, mitochondria will release more
ATP, cell function will be more active, and then the secretion of vaginal
mucosa will be enhanced, the color will be lighter, and the lubrication will be
enhanced. At the same time, through restoration The vaginal mucosa
normalizes the PH value and flora, reduces the recurrence rate of infections,
and restores female reproductive tissues to a younger level. In addition, the
private laser completely subverts the traditional birth canal repair method:
painless and non-invasive to solve the problems of postpartum genital tract
leakage, relaxation, decreased sensitivity and lubricity, and repeated
inflammation.

5. Application
5.1 Vagina Function
Thermal radiation make Mucosa tissue contraction and collagen fibers
thereby improving and restructuring vaginal tightening.
1. Shrink vagina: rapid tightening, constriction, lasting firmness.
2. Beautify vagina: dilute pigment, pink labia.
3. Moist vagina: increase secretion, eliminate dryness.
4. Maintenance vagina: deep rejuvenation, prevent aging.
5. To improve sensitivity.
6. Private health, balance PH, improve the internal environment.
5.2 Fractional And Pulse Function
1. Various scars (surgical scars, burn scars, scald scars)
2. Get rid of pigmented lesions (freckles, sun spots, age spots, sunburns,
chloasma, etc.)
3. Get rid of stretch marks.
4. Complete skin rejuvenation (rejuvenation, firming, shrinking pores, nodular
acne
5. Vascular disease (capillary hyperplasia, rosacea)
6. Remove false and true wrinkles
7. Remove acne scars

6. Specification

Wavelength: 10600nm
Laser apparatus: Sealed off laser device simulated by direct current
Optical beam quality: TEM00
Output power: 30w
Spot size: 0.02~0.05mm2
Aiming beam: Red semiconductor laser (650nm)
Beam transport device: Articulated arm with 6-joint and balance weight
Working mode: Continuous operation
Continuous, single pulse, interval pulse and super
Output Mode:
pulse
Scan scope: Max 20mmx20mm Min 0.1x0.1
Circle, triangle, square, rectangle,hexagon, ellipse,
Scan graphics:
line
Scanning Mode: Random , Normal and Midsplit scan
Energy: 1- 60W
Distance: 0.1-2.6mm
Cooling system: circulating water cooling
Weight: 31kg
Dimension: 93*52*36cm
Power: 220V/110V
7. Direction for Use
7.1Structure description

1. Joints of six light guide arm


The laser beam transmission is completed by the six-joint light guide arm.
The role of the light guide arm is to transmit the laser to the patient's site
(target surface) to be operated on
2. Probe
Graphics scanner can scan output, rectangular, round, hexagonal,
triangular and other graphics, the largest area of about 20 * 20mm² scanning;
3. Main unit
It includes a high voltage power supply, a low-voltage power supply,
control system, cooling system, laser system and blowing system.
4. touch display panel
Mainly used for parameter setting
5. Emergency switch
When the instrument has an abnormal condition, you can press the
emergency stop switch to cut off the power supply system of the instrument
6.key switch
A key into the key switch, go to the "open" position, the instrument power on,
go to the "off” position to cut the power, turn off the instrument.

7.Affusion hole
8.Overflow hole
9. Drainage hole
10.Foot switch
11.The power input socket

7.2 Installation
1) Checking

1. Check the packing box before open, if no abnormal, you can open it.
Becase the machine is equipped with laser tube and precision optical system,
forbid pound and collision when open the box.
2. Take out the packing list after opening the box, then take out the host and
spare parts
according to the packing list, check every spare parts, wire connections,
confirmed no
abnormal or troubleshooting before installation, if have any problem, please
contact
manufacturer timely, so that resolve the problem Quickly.

2) Fill distilled water

1. Press the stainless steel sheet at the above of the overflow hole, pull the
water tube into the hole.
2. Put the tube into the affusion hole, do like as above said, fill distilled water,
it's ok when
the water overflow the overflow hole; after turn on the machine for 3 minutes,
the machine need fill water again.
Attention: you can't pull out the tube from the affusion hole when using the
machine. You need to change the water in every 60 days.
3) Installation of guide light arm

Guide light system was made up of guide light arm and connection
seat .connection
seat has fixed into the top of the shell. After check the guide light arm, connect
it to
connection seat, then fix the screws firmly.
Attention: guide light arm only can move to the arrow direction.
Move B 180 degree according to C, then press A according to the Arrow.
Then insert B to the hole on main unit, then screw D, fix the guide light arm
firmly.
4) Installation of the scan head
1. Whirl the scan head to clockwise direction.

2. Installation of control wire on the scan head, insert the control wire, then fix
the screws firmly.

5) Installation of power cord and foot switch


1. Insert the power cord to the connector on the back of the machine, then
check if firmly.
2. Connect the foot switch, aim at the gap, whirl to the clockwise direction,
check if firmly.
6) change water
Put “add water” and ”run off water” connect, ventage plug connect run off
water, then add
water again.

7.3. Direction for use and Operating procedure


1) Turn on the power of device
2) Appear to be the main screen as below when key switch is ON.
3) Select a mode and range of adjustment when the main screen comes up.
4) Operate it by stepping FOOT SWITCH after completing every procedure.
5) Detail instructions of each mode are as follows

7.4 Screen Instruction

It is used for facial treatment, such as facial rejuvenation, acne, shrinking


pores, etc.

It is used for cutting, such as the treatment of foreskin, hemorrhoids, cervical


polyps, etc. It is generally not recommended for those without skilled
operation experience.
It is generally used for the treatment of the body, such as stretch marks, scars,
etc.

It is generally used for private treatment, such as the treatment of internal and
external vulva.

Press this button, enter the treatment interface, you can adjust every
parameter.

1) Touch “LASER POWER “ box, then press” ” or “ ” ,the energy will


increase or decrease.
* Range of Adjustment: 1-30W.

2) Touch “ RESIDENCE” box, press “ ” or “ ” to adjust.


* Range of Adjustment: 0.1-10.0ms

3) Touch “POINTS” box, press “ ” or” ” to adjust.


* Range of Adjustment: 1-5000ms

4) Touch “ROW SPACING” box, press” ” or “ ” to adjust.


* Range of Adjustment: 0.1mm ~ 2.6mm

5)Touch “SCAN MODE” box, press “ ” or “ ” to adjust.


* Range of Adjustment: Midsplit, Sequence,Disorder.

6)Touch “SCAN TIMES” box, press “ ” or “ ” to adjust.


* Range of Adjustment: 1-5th.

7)Touch ,you can change different shape,like Square, round,


triangular, hexagonal, straight.

8) Touch “ “ or “ “,you can change the shape size , And preview if


the size and shape is suitable when press “preview”.

9) After adjust suitable parameters, press “ ”,then press the foot


switch,the machine will shoot laser beam, turn to Main screen when touch
“BACK” box at the top right corner of screen.

1. Instruction of 3D mode
The treatment screen of PULSE MODE appears when selecting 3D box at the
Main Screen.

1) Touch “POWER “ box, then press” ” or “ ” ,the energy will increase or


decrease.
* Range of Adjustment: 1-30W.

2) Touch “ MODE” box, press” ” or “ ”, adjust impulse, continue, single etc.


* Range of Adjustment:
3) Touch “POINT STAY” box, press” ” or “ ” to adjust.
* Range of Adjustment: 1-5000ms

4) Touch “PULSE WIDTH” box, press” ” or “ ”to adjust.


* Range of Adjustment: 1-10ms.

5)Touch “INTERVAL” box, press” ” or “ ” to adjust.


* Range of Adjustment: 1-10ms.

2. Instruction of Face and 4D mode

1) Touch “LASER POWER “ box, then press” ” or “ ” ,the energy will


increase or decrease.
* Range of Adjustment: 1-30W.

2) Touch “ RESIDENCE” box, press “ ” or “ ” to adjust.


* Range of Adjustment: 0.1-10.0ms

3) Touch “POINTS” box, press “ ” or” ” to adjust.


* Range of Adjustment: 1-5000ms

4) Touch “ROW SPACING” box, press” ” or “ ” to adjust.


* Range of Adjustment: 0.1mm ~ 2.6mm

5)Touch “SCAN MODE” box, press “ ” or “ ” to adjust.


* Range of Adjustment: Midsplit, Sequence,Disorder.
6)Touch “SCAN TIMES” box, press “ ” or “ ” to adjust.

* Range of Adjustment: 1-5th.

7)Touch ,you can change different shape,like Square, round.

8) Touch “ “ or “ “,you can change the shape size , And preview if


the size and shape is suitable when press “preview”.

9) After adjust suitable parameters, press “ ”,then press the foot


switch,the machine will shoot laser beam, turn to Main screen when touch
“BACK” box at the top right corner of screen.

3. Instruction of Vaginal mode

1) Touch “LASER POWER “ box, then press” ” or “ ” ,the energy will


increase or decrease.
* Range of Adjustment: 1-30W.

2) Touch “ RESIDENCE” box, press “ ” or “ ” to adjust.


* Range of Adjustment: 0.1-10.0ms
3) Touch “POINTS” box, press “ ” or” ” to adjust.
* Range of Adjustment: 1-5000ms

4) Touch “ROW SPACING” box, press” ” or “ ” to adjust.


* Range of Adjustment: 0.1mm ~ 2.6mm

5)Touch “SCAN MODE” box, press “ ” or “ ” to adjust.


* Range of Adjustment: Midsplit, Sequence,Disorder.

6)Touch “SCAN TIMES” box, press “ ” or “ ” to adjust.


* Range of Adjustment: 1-5th.

7)Touch ,you can change different shape,like Square, round,


triangular, hexagonal, straight.

8) Touch “ “ or “ “,you can change the shape size , And preview if


the size and shape is suitable when press “preview”.

9) After adjust suitable parameters, press “ ”,then press the foot


switch,the machine will shoot laser beam, turn to Main screen when touch
“BACK” box at the top right corner of screen.

8. Operation

8.1 Preparation:
Alcohol, cotton ball, sterile treatment head.

8.2 Operation Of Vaginal:


Insert the support to the vagina, then insert the 360 degree head to the
bottom of the support, pay attention to the number on the 360 degree head,
press the foot switch, then pull out 0.5cm, press the foot switch again, then
pull our 0.5cm again, press the foot switch again, do again, after the 360
degree head reach vulva, insert to the bottom of the support again, and do
like the first time, total 3 times.
8.3 Vulva treatment:

Use the head for vulva, and make the distance 5cm, treat vulva, press
the foot switch for many times, until vulva tighten up.
8.4 Face treatment

These parameters are for the face, adjust the shape to square, and
biggest(20*20mm), then adjust same parameters, press GET READY, stating
the treatment. You can increase or decrease the POWER according to the
customer’s skin.

8.5 Stretch mark treatment:


These parameters are for stretch mark. You can increase or decrease the
POWER according to the customer’s skin.

9. Postoperative care and Notes:


1. After using the instrument, it is recommended to return to the standby state
before shutting down, and then turn the key switch to the "off" position to turn
off the instrument.
2. The light guide arm is played back to a natural position without force to
maintain a good optical path.
3. Remove the cutter head and other tools, take it for cleaning and
disinfection.
4. When the instrument is not in use, remove the key switch for safekeeping
to prevent unauthorized personnel from using or operating the instrument.

10.Troubleshooting guide
If the instrument fails, refer to the following table to find the
possible cause of the failure according to its failure
phenomenon, and take appropriate measures to eliminate the
failure. If you cannot solve the problem by yourself, please
contact our company's after-sales service department.
Warn:
High voltage and CO2 laser radiation may be generated when the
instrument is working normally. A little carelessness will cause harm to
the human body. Therefore, you must be careful when maintaining the
instrument.

Table 8-1 is the fault information that can be displayed on the control panel
and is relatively easy to solve.

Table 8-2 shows the situations in which the fault information cannot be
displayed on the control panel. The table lists the more detailed fault causes
and troubleshooting methods.

Table 8-1 Maintenance guide that can display fault information

Accident Details Causes of failures and troubleshooting methods

1.The instrument is not connected to AC power

2.Check whether the power cord is plugged into


the power socket, whether the main power control
switch, emergency stop switch are turned on, and
whether the connection sockets are well
connected
Nothing is displayed when
the instrument is powered 3.Low-voltage switching power supply failure

on 4.Check whether the input, output socket, input


and output voltage of the low-voltage switching
power supply are normal

5.The control board is faulty

6.The LCD display is malfunctioning

7.Contact our after-sales service department

1.The foot switch is not connected properly


Foot switch failure 2.The foot switch is damaged, the foot switch
needs to be replaced
1. The high-voltage power supply is not energized
High-voltage power supply or the main SSR is faulty
failure 2.High-voltage power supply failure, high-voltage
power supply needs to be replaced

Table 8-2 Maintenance guide without fault information

Fault description Causes of failure and repair methods

1.Not connected to AC power

2.The emergency stop switch is not turned on


When the switch key is
3.The main control switch on the rear panel is not
turned to the "ON" position,
turned on
the instrument has no action
4.The low-voltage power supply is damaged or
the main control board fails

1.Rotate the aiming light adjustment knob


position

2.Too much dust has fallen on the lens of the


Aiming indicator light is too
light guide system, the dust needs to be removed
weak
or the light guide arm needs to be replaced

3.The semiconductor laser is damaged and the


semiconductor laser needs to be replaced

CO2 does not fall on the Same optical path offset


aiming light spot

1.The air pump is not connected to 220V AC

When the instrument is in 2.The trachea is not connected


the ready state, there is no 3.The air pump is damaged, the air pump needs
blowing to be replaced

4.Other electrical faults

When the foot switch is 1.The connection of the foot switch is not well
pressed, there is no CO2 connected
laser emission 2.The instrument is not in a ready state

3.Laser tube is damaged

4.High-voltage power supply failure

5.Other electrical faults

1.Damaged control panel

2.Same optical path misalignment

3.The laser tube has been used for too long


Power cannot be calibrated 4.The control board is not working properly

5.Damaged high-voltage power supply or laser


tube

6.Other electrical faults