User’s Manual
1. Equipment Introduction
1.Burn
The CO2 laser scanning laser has a wavelength of 10.6 microns, which is a
spectral line in the far infrared range, which is invisible to the human eye. The
maximum laser power output by this instrument can reach 30 watts, which
can cause 3 degree burns even without focusing. Therefore, sufficient
attention should be paid.
The instrument outputs visible red semiconductor laser and invisible CO2
laser, both of which are harmful to the human body.
Do not look directly at the red semiconductor laser at any time, even if it
does not cause burns to the human eyes, it may cause damage to a certain
degree.
Since the surface of the object, especially the surface of smooth metal or
glass, can form a mirror surface to reflect light, it is important to remove such
objects or surfaces that may reflect on the path of the laser. In other words, do
not irradiate laser light onto objects that reflect light to prevent the reflected or
scattered laser light from causing harm to the human body.
Do not use this laser instrument on the site with flammable and explosive
materials. Do not place or store flammable and explosive materials around
the instrument. Flammable and explosive materials include: gasoline, alcohol,
some narcotics, some solvents, desiccants, ointments, synthetic resins, etc.
1. Before treatment ‘
1. Keep the surface of the instrument clean.
2. Get the instrument key.
3. Ask the surgeon how to arrange the position of the instrument and the
patient.
4. Stick a laser warning sign in a conspicuous position.
5. If you need to use general anesthetics, prepare a wet towel for the
patient.
6. Prepare all necessary equipment and tools.
7. Check whether the instrument is working properly.
8. After inserting the switch key into the instrument's key, power on and
turn on the instrument. The instrument should perform self-check first, and
then enter the standby working state after selecting the working mode.
9. If necessary, press the setting key to reset the working parameters.
10. Check the same optical path.
11. Press the set button to return to the standby working state or turn off
the instrument until the operation is turned on again.
2. After treatment
1. Turn off and remove the instrument.
2. Pull out the switch key and keep it in a safe place.
3. Place the light guide joint arm in the normal position.
4. If necessary, remove the optical accessories and other tools for
cleaning or disinfection.
3. Accessories list
5. Application
5.1 Vagina Function
Thermal radiation make Mucosa tissue contraction and collagen fibers
thereby improving and restructuring vaginal tightening.
1. Shrink vagina: rapid tightening, constriction, lasting firmness.
2. Beautify vagina: dilute pigment, pink labia.
3. Moist vagina: increase secretion, eliminate dryness.
4. Maintenance vagina: deep rejuvenation, prevent aging.
5. To improve sensitivity.
6. Private health, balance PH, improve the internal environment.
5.2 Fractional And Pulse Function
1. Various scars (surgical scars, burn scars, scald scars)
2. Get rid of pigmented lesions (freckles, sun spots, age spots, sunburns,
chloasma, etc.)
3. Get rid of stretch marks.
4. Complete skin rejuvenation (rejuvenation, firming, shrinking pores, nodular
acne
5. Vascular disease (capillary hyperplasia, rosacea)
6. Remove false and true wrinkles
7. Remove acne scars
6. Specification
Wavelength: 10600nm
Laser apparatus: Sealed off laser device simulated by direct current
Optical beam quality: TEM00
Output power: 30w
Spot size: 0.02~0.05mm2
Aiming beam: Red semiconductor laser (650nm)
Beam transport device: Articulated arm with 6-joint and balance weight
Working mode: Continuous operation
Continuous, single pulse, interval pulse and super
Output Mode:
pulse
Scan scope: Max 20mmx20mm Min 0.1x0.1
Circle, triangle, square, rectangle,hexagon, ellipse,
Scan graphics:
line
Scanning Mode: Random , Normal and Midsplit scan
Energy: 1- 60W
Distance: 0.1-2.6mm
Cooling system: circulating water cooling
Weight: 31kg
Dimension: 93*52*36cm
Power: 220V/110V
7. Direction for Use
7.1Structure description
7.Affusion hole
8.Overflow hole
9. Drainage hole
10.Foot switch
11.The power input socket
7.2 Installation
1) Checking
1. Check the packing box before open, if no abnormal, you can open it.
Becase the machine is equipped with laser tube and precision optical system,
forbid pound and collision when open the box.
2. Take out the packing list after opening the box, then take out the host and
spare parts
according to the packing list, check every spare parts, wire connections,
confirmed no
abnormal or troubleshooting before installation, if have any problem, please
contact
manufacturer timely, so that resolve the problem Quickly.
1. Press the stainless steel sheet at the above of the overflow hole, pull the
water tube into the hole.
2. Put the tube into the affusion hole, do like as above said, fill distilled water,
it's ok when
the water overflow the overflow hole; after turn on the machine for 3 minutes,
the machine need fill water again.
Attention: you can't pull out the tube from the affusion hole when using the
machine. You need to change the water in every 60 days.
3) Installation of guide light arm
Guide light system was made up of guide light arm and connection
seat .connection
seat has fixed into the top of the shell. After check the guide light arm, connect
it to
connection seat, then fix the screws firmly.
Attention: guide light arm only can move to the arrow direction.
Move B 180 degree according to C, then press A according to the Arrow.
Then insert B to the hole on main unit, then screw D, fix the guide light arm
firmly.
4) Installation of the scan head
1. Whirl the scan head to clockwise direction.
2. Installation of control wire on the scan head, insert the control wire, then fix
the screws firmly.
It is generally used for private treatment, such as the treatment of internal and
external vulva.
Press this button, enter the treatment interface, you can adjust every
parameter.
1. Instruction of 3D mode
The treatment screen of PULSE MODE appears when selecting 3D box at the
Main Screen.
8. Operation
8.1 Preparation:
Alcohol, cotton ball, sterile treatment head.
Use the head for vulva, and make the distance 5cm, treat vulva, press
the foot switch for many times, until vulva tighten up.
8.4 Face treatment
These parameters are for the face, adjust the shape to square, and
biggest(20*20mm), then adjust same parameters, press GET READY, stating
the treatment. You can increase or decrease the POWER according to the
customer’s skin.
10.Troubleshooting guide
If the instrument fails, refer to the following table to find the
possible cause of the failure according to its failure
phenomenon, and take appropriate measures to eliminate the
failure. If you cannot solve the problem by yourself, please
contact our company's after-sales service department.
Warn:
High voltage and CO2 laser radiation may be generated when the
instrument is working normally. A little carelessness will cause harm to
the human body. Therefore, you must be careful when maintaining the
instrument.
Table 8-1 is the fault information that can be displayed on the control panel
and is relatively easy to solve.
Table 8-2 shows the situations in which the fault information cannot be
displayed on the control panel. The table lists the more detailed fault causes
and troubleshooting methods.
When the foot switch is 1.The connection of the foot switch is not well
pressed, there is no CO2 connected
laser emission 2.The instrument is not in a ready state