for liquids
Measuring range from 0,003l/h to 30 l/h
Features
• Suitable for conducting and nonconducting fluids
• High sensitiveness in low flow ranges
• Wide measuring ranges allow the detection of leakage and increases in
consumption
• Parallel measurement of the temperature of the medium
• Multiple signal outputs for standard model,
(Indication, set points, analog output, pulse output, RS232-interface)
Applications
• Recording of dosage of liquids
• Consumption measurement of valuable media
• Flow measurement for control purposes
• Min. and max. flow control
Issue 06/09/2006
connecting cable
with protection pipe clip with
process connection welding plate
process connection
additional parts
System overview
Analog
I/U output Converter electronic
type SF-586
Relay
S.P.C.O. contact
Keyboard
Interface
Sensor SF-6100
RS232
Interface
LC-indication
illuminated
Standard *1
in housing 150x150x57mm
12 Volt DC with 1x Pg7 and 2x Pg9
12 Volt DC Option *2
24 Volt DC in housing 200x150x79mm
or with 5x Pg9
230 Volt AC
3 x analog input
0-5 Volt
e.G. pressure
2
Construction
The unit consists of a compact cylindric sensor with a measuring pipe made of glass
and a separate computer assisted converter electronic contained in a plastic housing
for simple wall mounting.
The process connections of the sensor are placed axial on both ends.
Performance
Unit with indication of actual flow rate (volume flow, mass flow), totalizer and output signal
0-10V, or 0/4-20mA proportional to flow, pulse output (< 30x per minute) or set point and bi-
directional interface RS232.
Delivery on demand as customer requested: Indication of actual flow rate in terms of volume
or mass, totalized flow in terms of volume or mass, output signal (max. value), pulses, set
point, baudrate, address, mean value creation, off-set, min. quantity suppression.
If a keyboard is supplied, the customer can set the above values as well as starting,
stopping and resetting the totalizer. Analog and pulse output as well as set point can be
coordinated to each measured value via keyboard.
With the RS232-interface all settings and data transmitting can be done with a PC.
The software and a 1,5m cable for connection with the PC is part of the delivery.
Electrical connection
spring-cage connection
RS232
230V, 50/60Hz
RS232
RS232
3
Technical data of the sensors
Operating Principle heat dissipation principle, primary signal
proportional to mass flow
Medium liquids (Calibration with water)*
Measuring range 3 to 300ml/h 30 to 3.000ml/h 300 to 30.000ml/h
Accuracy +/-3% of the measured value
Repeatability +/-1% of the measured value
Warming up time about 5 minutes after switching on
Operating Pressure maximal 10bar
Operating Temperature +10°C to +40°C
Surrounding Temperature 0 to 60°C
Installation Position horizontal
Process Connection, standard G1/8 inside, 10mm deep
optionally Pipe screw coupling for outer diameter-∅4mm, with
adapter for plastic pipes and flexible tubes
or Pipe screw coupling for outer diameter-∅6mm, with
adapter for plastic pipes and flexible tubes
or Pipe screw coupling for outer diameter-∅8mm, with
adapter for plastic pipes and flexible tubes
Wetted Parts glass or glass and PVDF or glass and PVC,
stainless steel
Protection Class IP65
Housing dimensions see drawings
Connecting Cable length 5 m
sensor to converter
Mounting pipe clip with welding plate
* The measuring ranges will vary when used for liquids with different heat transfer responses. In this
case another calibration run using the liquid to be measured and a balance as reference is
recommended at customer’s premises. Signal correction procedure is described in the operation
manual.
Sundries
All wetted parts can modified with special materias, for example in PTFE or PVC.
For the process connection many spezial realizations are available, for example NPT-thread.
In case of the modular construction of the sensor it is very easy for the customer to change
the connection parts for his specific application.
4
Sensor dimensions (standard)
wire end sleeve
G1/8 G1/8
5
Technical data of the converter
*** internal terminal for PC-connection and internal connection for option keyboard.
Housing dimensions
standard option
6
Cleaning
7
Type key / specification for ordering SF-6100
8
Contact
softflow.de GmbH
Dorfstr. 34
D-15834 Gross Machnow