12.0 PSV Calibration Procedure

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 9

Doc.

No: INST-SOP-02
Rev: 0
PSV CALIBRATION PROCEDURE Date: 14-01-2020
(ISO 17025:2017) Approved by: Kenneth Ilouno

INSTRUTECH LIMITED
CONTROLLED COPY
1

PSV CALIBRATION
REVISION (ISO 17025:2017)
AND AUTHORISATION RECORD
REVISION: 3

INST-SOP-02
PREPARED REVIEWED APPROVED
BY BY BY
NAME EMMANUEL TORKA ILOUNO
UYAELUMUO CLEMENT KENNETH
SIGNATURE

DATE

Prepared by Reviewed by Approved by


LEAD AUDITOR Quality Manager Managing Director
AZUKA CHRISTOPHER EKEKE CLETUS ILOUNO KENNETH

This document belongs to Instrutech Limited. No corrections / amendments are to be made except by written authorization of the
Management.

1
Doc. No: INST-SOP-02
Rev: 0
PSV CALIBRATION PROCEDURE Date: 14-01-2020
(ISO 17025:2017) Approved by: Kenneth Ilouno

INSTRUTECH LIMITED
CONTROLLED COPY

AMENDMENT PAGE

Officer for Changes


S.N Page Item Changes Description
Name Sign.

2
Doc. No: INST-SOP-02
Rev: 0
PSV CALIBRATION PROCEDURE Date: 14-01-2020
(ISO 17025:2017) Approved by: Kenneth Ilouno

INSTRUTECH LIMITED
CONTROLLED COPY

TABLE OF CONTENTS
S/N DESCRIPTION PAGE
1.0 ABBREVIATIONS
2.0 PURPOSE
3.0 SCOPE
4.0 REFERENECE DOCUMENT
5.0 TEST EQUIPMENTs
6.0 SAFETY
7.0 PROCEDURE
7.1 SYSTEM OF IMPLEMENTATION
7.2 REVISION
7.3 DISTRIBUTION
8.0 APPENDIX

0.0 ABBREVIATIONS AND DEFINITIONS


The following abbreviations are used freely in this document
ABBREVIATION DEFINITIONS
ASME American Society of Mechanical Engineers
DEP Design and Engineering Practice
DP Design Pressure
MAWP Maximum Allowable Working Pressure
MAOP Maximum Allowable Operating Pressure
PEP Project Execution Plan
RPM Revolution Per Minute
SMYS Specific Minimum Yield Stress
STP Specified Test Pressure
INSTL INSTRUTECH LIMITED
ROG ROSTRUM OIL AND GAS
MPMP MUD PUMP MACHINE PROCEDURE
SCNL SAIPEM CONTRACTIG NIG LIMITED
SP SET PRESSURE
AS AS FOUND

3
Doc. No: INST-SOP-02
Rev: 0
PSV CALIBRATION PROCEDURE Date: 14-01-2020
(ISO 17025:2017) Approved by: Kenneth Ilouno

INSTRUTECH LIMITED
CONTROLLED COPY

1.0 Purpose
The purpose of this procedure is to provide guidelines for carrying
out calibration and certification of PSV for SPDC/SAIPEM.
.
The reason for the guidelines is to emphasis the method/ steps to
be followed during calibration and certification of this equipment.

2.0 Scope:

This procedure covers the calibration and certification of PSV for


SPDC/SAIPEM .

This work shall be carried out in accordance with the guidelines


stated in contract and ASME B31.1 section 137, ISO17025 and
DPR regulation for PSV calibration. The test medium will be
Nitrogen or Compressed air or distilled water or oil depending on
clients requirement and area of usage of the PSV.
All necessary documentation as indicated in the specifications
mentioned above shall be carried out at every stage of the activity.

3.0 REFERENCE DOCUMENTS:

The following document will be referred to in determining the


procedural requirement for carrying out this work.
 Nigerian Factories Decree 1987
• Maintenance Manuals Instrument Society of America
• NIST, ISO/IEC 17025(E), NIS of SON and DPR
• ASME B31.3
• Manufacturer’s manual

4
Doc. No: INST-SOP-02
Rev: 0
PSV CALIBRATION PROCEDURE Date: 14-01-2020
(ISO 17025:2017) Approved by: Kenneth Ilouno

INSTRUTECH LIMITED
CONTROLLED COPY

4.0 SAFETY PRECAUTION:

Considering the inherent dangers involved with this job, safety shall
be the major concern. The following steps in addition to the clients
procedures shall be followed..
Only personnel directly involved with the job shall be
allowed in the job area
An experienced Safety Officer will be available to ensure
safety at all times during pressure testing.
Ensure that a JHA has been prepared and is explained and
clearly understood by all parties involved in the work.
Check ALL personnel involved in the test have correct PPE
Activities of personnel performing this work should be carried
out within a well-established and management approved
framework. The frame work should be based on both
SPDC/SAIPEM and INSTRUTECH instructions and should
contain the following:

Standing Instructions -Both SPDC/SAIPEM


&INSTRUTECH

Authorization Procedures -Method of appointing the


competent person who is to
supervise or carry out the work`
Work Procedure - Detailed method of carrying out
the work (this procedure)
Only authorized personnel will be allowed to perform the
work. The safety officer will maintain vigilance throughout the
work and installation to ensure that the stipulated safety
precautions and procedures are being followed. Any
infringements will be reported without delay to the site
Coordinator.

5
Doc. No: INST-SOP-02
Rev: 0
PSV CALIBRATION PROCEDURE Date: 14-01-2020
(ISO 17025:2017) Approved by: Kenneth Ilouno

INSTRUTECH LIMITED
CONTROLLED COPY

5.0 METHOD STATEMENT

SN DESCRIPTION OF ACTIVITY RESPONSIBLE


PERSON

First, test the PRV to know if the Servicing


PRV will pop/relief at the tag set technician/equipment
pressure or the required pressure, operator
using the standard test gauge/ the
digital test gauge as reference.
If it pops/relieves at the set pressure Servicing
on the PRV tag, then, there will be technician/equipment
no need to reset or adjust, but if it operator
doesn’t, then calibration/adjustment
is needed.
Severe the tamper proof wire (with Servicing
cutter) to enable access to CAL technician/equipment
NUT. operator

Unscrew the CAP LOCK using Servicing technician


spanner.
Loosen the LOCK NUT (with Servicing technician
spanner)
The adjustment starts (using spanner Servicing technician
or pliers).
Open the isolation valve to let the Servicing technician
Nitrogen flow into the manifold.
Use the pressure regulator to Servicing technician
regulate pressure going into the
PRV.
Adjust the CAL NUT (calibration nut) Servicing technician
with spanner/pliers to get the
required pressure., referencing to the
standard test gauge, this is done till
the required pressure is achieved.

6
Doc. No: INST-SOP-02
Rev: 0
PSV CALIBRATION PROCEDURE Date: 14-01-2020
(ISO 17025:2017) Approved by: Kenneth Ilouno

INSTRUTECH LIMITED
CONTROLLED COPY

Put back a tamper proof wire on the Servicing technician


PRV to stop access to the CAL NUT.
Print the “SET DATE” and “SET Servicing technician
PRESSURE” of the PRV on a
metallic plate.
Put the calibration sticker on the Servicing
other side of the metallic plate. technician/equipment
operator
Tag the metallic plate on the PRV. Servicing
Prepare a Calibration certificate for technician/equipment
the PRV; this certificate will carry the: operator
Tag the metallic plate on the PRV. Servicing
technician/equipment
operator
Prepare a Calibration certificate for Servicing
the PRV; this certificate will carry the: technician/equipment
operator

THINGS THAT SHALL BE IN THE CERTIFICATE

 Calibration number
 Clients name
 Calibration date
 Due date
 Tractability
 Ambient Temperature
 Set pressure
 Relative humidity
 Method of calibration
 Manufacturer of the PRV
 Service
 Equipment name
 Part number/model number
 Tag number
 Serial number

7
Doc. No: INST-SOP-02
Rev: 0
PSV CALIBRATION PROCEDURE Date: 14-01-2020
(ISO 17025:2017) Approved by: Kenneth Ilouno

INSTRUTECH LIMITED
CONTROLLED COPY

SUGGESTION/CONCLUSION

If the PRV does not respond at all adjustments, both clockwise and anticlockwise
adjustments, then, its either the seat is pitted or the spring has lost its elasticity.
Any of these parts could be replaced.

5.1 TEST EQUIPMENTS

Mobile PSV workshop as attached(MS-RV Barbee PSV Test


Bench)
Instrument tools box
Mechanical tool box
Pressure test gauge(spare)
Spares (seal and diaphragm)
PRV CALIBRATION MANIFOLD(spare)
PIPE WRENCH
PRESSURE REGULATOR(spare)
ISOLATION VALVE(spare)
FITTINGS
NITROGEN BOTTLE(spare)

5.2 System of Implementation


 Select and assign personnel to be responsible for the work.
Selection to be based on the persons past experience with
the types of system being tested.
 Compile all the necessary documentation and make it
available to the personnel responsible for the work.
 Familiarize the responsible personnel with the tasks outlined
in this procedure and the equipment to be tested.
 Perform during calibration and certification must be in
accordance with SPDC/SAIPEM safety guidelines.
 Develop a maintenance plan and issue draft report to
SPDC/SAIPEM for review
 Incorporate comments and issue final report to
SPDC/SAIPEM.

8
Doc. No: INST-SOP-02
Rev: 0
PSV CALIBRATION PROCEDURE Date: 14-01-2020
(ISO 17025:2017) Approved by: Kenneth Ilouno

INSTRUTECH LIMITED
CONTROLLED COPY

5.3 Revision
The procedure can be revised by the Quality Manager upon
warranted request of any member of the staff.

5.4 Distribution
SPDC/SAIPEM Project Manager, QA Department and Work Site.

6.0 APPENDIX
Appendix 1: Typical calibration certificate