Professional Documents
Culture Documents
1 s2.0 S0021967320309912 Main
1 s2.0 S0021967320309912 Main
1 s2.0 S0021967320309912 Main
Journal of Chromatography A
journal homepage: www.elsevier.com/locate/chroma
a r t i c l e i n f o a b s t r a c t
Article history: In this study, we introduce a simple three-step workflow for a universally applicable RI system, to be
Received 3 August 2020 used in GC×GC analysis of ignitable liquid residue (ILR) for arson investigations. The proposed RI system
Revised 7 November 2020
applies a combination of two well-established GC RI systems: non-isothermal Kovats (K) index in the first
Accepted 9 November 2020
dimension and Lee (L) index in the second dimension. The proposed KLI RI system showed very good
Available online 14 November 2020
correlations when compared with predicted values and existing RI systems (r2 = 0.97 in first dimension,
Keywords: r2 = 0.99 in second dimension) and was valid for a wide range of analyte concentrations and operational
GC×GC settings (coefficient of variance (CV) < 1% in first dimension, < 10% in second dimension). Using the
forensic science KLI RI, an ILR classification contour map was created to assist with the identification of ILR types within
fire debris analysis ASTM E1618. The contour map was successfully applied to neat fuels and a fire scene sample, highlighting
arson investigations the application to wildfire investigation. Standardizing the RI process and establishing acceptable error
margins allows the exploration and comparison of comprehensive data generated from GC×GC analysis
of ILRs regardless of location, time, or system, further enhancing comprehensive and tenable chemometric
analyses of samples. Overall, the KLI workflow was inexpensive, quick to apply, and user-friendly with its
simple 3-step design.
© 2020 The Author(s). Published by Elsevier B.V.
This is an open access article under the CC BY-NC-ND license
(http://creativecommons.org/licenses/by-nc-nd/4.0/)
https://doi.org/10.1016/j.chroma.2020.461717
0021-9673/© 2020 The Author(s). Published by Elsevier B.V. This is an open access article under the CC BY-NC-ND license
(http://creativecommons.org/licenses/by-nc-nd/4.0/)
N. Boegelsack, C. Sandau, D.W. McMartin et al. Journal of Chromatography A 1635 (2021) 461717
tivity and selectivity, to improve the detection and prosecution of riety in compound classes of interest complicate defining its re-
arson [7,8]. Particularly for wildfires, where the sample matrix con- spective zero value. N-alkanes are commonly used as baseline due
sists of different types of wood and soil, tree bark and other veg- to their non-polarity combined with their utility in GC RI [33,34].
etation, the background composition has similar physico-chemical One method uses the average n-alkane secondary retention time,
properties to the ILR target compounds and is present at greater whereas another approach requires a continuous evaluation of the
concentrations [9]. bracketing n-alkanes for each compound of interest. Meanwhile,
Comprehensive multidimensional gas chromatography coupled it could be argued that the “true” zero is the modulation period
with mass spectrometry (GC×GC-MS) is a powerful tool for the switch, which is often accompanied by a characteristic peak of sol-
analysis of complex sample mixtures, with a ten-fold or larger in- vent.
crease in separation potential over one-dimensional GC [10]. Appli- Defining zero also impacts wrap-around peaks. The aligning of
cations of the technique include analysis of petroleum [11,12,13], chromatograms to an arbitrary value in order to minimize or ex-
and of food and beverages [14,15], and use in medical and diag- clude the appearance of wrap-around will result in incongruen-
nostic testing [16,17], and in environmental and criminal forensics cies in comparing multiple samples. These initial studies [32,33,34]
[18,19,20]. Initial studies of ILR analysis using GC×GC coupled to a highlighting the need to widen the compound class coverage, and
flame ionization detector [7] and a quadrupole mass spectrometer standardize and simplify the RI process, to build comprehensive
[21] highlighted the enhanced potential of the method to improve databases for identification of unknown compounds.
resolution, peak detectability, and sensitivity. Analysis of ILR using A standardized RI system would allow for comprehensive anal-
GC×GC coupled with time-of-flight mass spectrometry (TOF-MS) ysis of all the compounds in ILRs, not only the traditional qual-
has only recently emerged in the literature [22,23,24,9], and em- itative compounds, and facilitate the development of more com-
phasizes the accessibility of the large amount of resultant data to prehensive ILR databases and libraries. In addition, the RI method
multivariate statistical analysis. would allow independent data verification by other scientists, and
While GC×GC can provide better separation and resolution of cross-referencing between applications. An ILR reference library
complex mixtures such as ILRs, there remains the challenge of cor- stemming from retention indices would allow comparison of data
rectly identifying separated compounds. The most reliable way to from an arson investigation to a variety of reference samples. These
identify a compound is based on co-injection of authentic stan- could originate from different general ILR classes, from locally ac-
dards, which can be both expensive and time consuming, as well quired ILR classes as part of the investigation, and from other po-
as not always being possible in the absence of existing or compre- tential arson investigations thought to be a result of the same ar-
hensive standards. Retention time indices, which convert retention sonist. The comprehensive analysis of all compounds in ILRs by
times of compounds into system-independent unitless constants, GC×GC analysis provides large datasets that may allow the correla-
have supported compound identification successfully in conven- tion of ILR to a fuel brand. Any reference library standardized to RI
tional GC, as well as GC×GC [14]. For example, the widely used would be comparable and allow the same chemometric approach
Kovats Retention Index converts the retention time of compounds for identification of ILRs. In turn, this would facilitate the inclusion
by normalizing the retention time to adjacent eluting n-alkanes. of large amounts of data with minimal logistics, making any statis-
These derived RI may be used to identify peaks by comparing mea- tical deductions more tenable.
sured retention indices with the tabulated published values, re- In the work described here, an alternate RI system and work-
ferred to as retention index libraries [25]. The continuous relevance flow suitable for GC×GC analysis of ILRs is presented. The RI
of this has been demonstrated recently in application to analy- system is universally applicable, allowing users analyzing simi-
sis of fragrance, drugs, and environmental pollutants [26,27,28]. lar chemical mixtures on different systems and analytical condi-
An example is the analysis of isomers and diastereomers, which tions to compare results directly, regardless of modulator type.
are indistinguishable based on their mass spectra, but which often The benefits of the proposed RI system are that it uses readily
have distinct retention indices (e.g. pentyl- and isopentylbenzene, available standards in a single injection, is directly comparable to
xylenes). This highlights the power of RI data to complement mass one-dimensional GC, and is pertinent to various applications with
spectral data for reliable compound identification [25]. petroleum hydrocarbons. The performance and robustness of the
Initial work has begun to develop two-dimensional indices, in- proposed system has been tested by comparison to theoretically
cluding the use of mathematical (thermodynamic models) [29,30] predicted values and other applicable RI systems, and error anal-
and instrumental (isovolatility curves) approaches [14,31,32]. Gard- ysis has been completed to test the tolerance of the proposed RI
ner et al. [33] combined the Kovats index in the first dimen- system to instrument changes. Lastly, this work confirms the func-
sion and the Lee index in the second dimension to profile food- tionality of the RI system to wildfire arson investigations by apply-
borne pathogen bacteria using fatty acid methyl esters (FAMES). ing it in the analysis of accelerants and ILR in fire debris samples,
And, while those compound groups (n-alkanes and FAME) are rel- in conjunction with the ASTM classification system.
evant to ILR analysis, particularly biodiesels, the complexity (car-
bon number and ring structure) needs to be extended to cover 2. Material & Methods
the range of ILR compound classes. Veenas & Haglund [34] used
n-alkanes and polyethylene glycols (PEGs) in their RI system to 2.1. Standards and reagents
introduce a regular polarity increase in the second dimension. RI
systems in one-dimensional GC can be converted between one an- A C7 -C30 saturated alkanes certified reference material (Cat#:
other if the correlation for the compound classes of interest is 49451-U), C8 -C40 Alkanes Calibration Standard (Cat#: 40147-U),
known [25]. This principle transfers over to the first dimension in PAH Mix 3 certified reference material (Cat#:861291), and PAH
GC×GC as the factors for calculation do not change. A similar study Calibration Mix (Cat#: CRM47940) were purchased from Sigma
for the second dimension is yet to be undertaken. Aldrich (Supelco, Bellefonte, PA, USA). Benzene (99.9%), carbon
One complication of the second dimension is the process of disulfide (CS2 , 99.9%), dichloromethane (DCM, 99.9%), methanol
defining its ‘zero’ value. This is a straightforward procedure in the (99.9%), and toluene (99.9%) were obtained from Fisher Scientific
first dimension as it refers to the absolute zero in time. Variabili- (Ottawa, ON, Canada). A polyethylene glycol mixture (mono to
ties exist in determining which data to exclude, e.g. dead time or deca) was purchased from Chiron AS (Trondheim, Norway).
first eligible standard, but retention times always relate to the mo- A 1ppm aromatic standard mixture containing alkanes (49451-
ment of injection. In the second dimension, wrap-around and va- U), benzene, PAHs (861291), and toluene in DCM was prepared for
2
N. Boegelsack, C. Sandau, D.W. McMartin et al. Journal of Chromatography A 1635 (2021) 461717
3
N. Boegelsack, C. Sandau, D.W. McMartin et al. Journal of Chromatography A 1635 (2021) 461717
KLI: Step 1: Define data boundaries Step 2: Apply KI1 to x values Step 3: Apply LI2 to y values
PEG: Step 1: Define data boundaries Step 4: Apply to x values Step 5: Define 0
Step 2: Extrapolate t2 PEGs (compounds of interest)
Step 3: Linear regression PEGs Step 6: Extrapolate new t2
Step 7: Apply regression
formula to y values
Fig. 1. Workflow summarizing steps required to generate retention indices using conventional one-dimensional RI systems, the PEG RI system, and the proposed KLI system.
Data was acquired and processed using the ChromSpace soft- 3.1. Retention index system workflow
ware (V 1.5.1, Sepsolve, Peterborough, UK) and Microsoft Excel (Mi-
crosoft, Redmond, WA). A deconvolution algorithm was used for The general workflow for conventional retention index systems
integration, with a minimum ion count 300, minimum absolute consists of two stages: data preparation, which is where data
area and height of 10 0 0, and peak merging at 10% overlap. boundaries are defined, and RI calculation. Extending this work-
The performance was compared between the proposed KLI flow to a GC×GC setup requires the additional normalization of
system, the PEG system, and the conventional van den Dool & second dimension retention time (Fig. 1). A new workflow for
Kratz system by evaluating simplicity of use and the total per- GC×GC is presented in Fig. 1 and compared with current 1D and
centage of area used. The area covered in both retention spaces existing GC×GC RIs.
was calculated in accordance with the Cordero method including The first stage, data preparation, defines data boundaries re-
wraparound peaks as outlined in Eq.s (7) and (8) [37]: quired to ensure that results between different physical systems
will be comparable. Van den Dool and Kratz [35] noted in their
A=l×w (7) research the importance of keeping a uniform temperature ramp.
They suggest that compounds eluting close to the initial and fi-
With l representing the length in minutes (first dimension) and w nal hold temperatures are affected by pressure changes associated
representing the width in seconds (second dimension) with the initiation and termination of temperature ramps. For their
RI calculation, any compound impacted by this effect is consid-
Acovered ered outside the data boundaries. While the same principles apply
%Aused = × 100 (8) to the first dimension in GC×GC, data processing tools are read-
A potential
ily available to correct for these effects and to align data points
With Acovered describing the area pertaining to the RI system and to their projected path. Defining data boundaries also requires de-
Apotential as the total area in the chromatogram as calculated by notation of which data points were acquired outside of the range
Eq. (7) respectively. bracketed by standard compounds. Any compounds eluting before
The KLI system was then evaluated for its overall robustness the first and after the last calibration compound will require ex-
by comparing standard deviations associated with a linear range trapolation to obtain RI results. Table 1, which displays calibration
of analyte concentration and alternative operational settings. For compounds used for the first dimension, denotes absolute reten-
comparison between units (minutes / seconds for absolute reten- tion times by tR 1 . The last calibration compound to elute is tria-
tion times and retention index units for retention indices), the co- contane, and the RI of any compounds with a subsequent tR 1 must
efficient of variation (CV), expressed as percentage, was calculated be extrapolated from the temporal relationship of nonacosane and
in addition to the standard deviation. triacontane. The same principle applies to any compounds exceed-
Finally, the KLI system was applied to a variety of ignitable liq- ing the absolute retention time of the latest calibration compound
uids to test the accuracy of tR 2 assumptions and evaluate suit- in the second dimension, tR 2 (Tables 1 and 2).
ability for classification under ASTM E1618 [4]. In addition, data An additional component of data preparation for multidimen-
from a fire debris sample collected in the 2018 wildfire season sional RI systems is to define the zero-value in the second dimen-
was used to investigate whether a direct comparison was possible sion, which is denoted by tR 2 corrected (tR 2 corr.; Tables 1 and
with ILRs analyzed under different conditions. Non-isothermal Ko- 2). The KLI system defines a universal zero across the entire chro-
vats (KI1 ) and Lee index (LI2 ) data from analysis of 2018 fire season matogram area, which allows for a single-step data preparation
were chosen as they had been analyzed with a different secondary stage. The PEG system defines a separate zero value for each com-
column (smaller i.d. and film thickness), and would simulate the pound and requires the extrapolation of an intermediate tR 2 cor-
potential changes in instrumental setup between suspect samples rected relative to the bracketing n-alkane. In addition, a linear re-
and ignitable liquid comparison samples. gression calibration curve based on the PEGs is required as foun-
4
N. Boegelsack, C. Sandau, D.W. McMartin et al. Journal of Chromatography A 1635 (2021) 461717
Table 2
Second-dimension retention indices for calibration compounds using Lee indices (LI2 ) for the KLI
system and PEG indices (PI2 ) for the PEGs system.
5
N. Boegelsack, C. Sandau, D.W. McMartin et al. Journal of Chromatography A 1635 (2021) 461717
Table 3 changes is smaller than the rounding factor, making this model ap-
Comparison of expected (expct.) and calculated (calc.) retention index results to
plicable in the second dimension.
structures of verification compounds.
The expected value for KI1 was determined by taking the boil-
KI1 LI2 ing point of each verification compound and comparing it against
Compound Structure
expct. calc. expct. calc. the boiling points of n-alkanes. Substituting retention times in the
non-isothermal Kovats index calculation with boiling points, de-
rived the expected RI value in Table 2. Light to medium range com-
Benzene 650 693 100 67 pounds (as defined for ILR analysis by ASTM E1618) are very close
and within rounding parameters to their predicted value, except
Naphthalene, 2-methyl- 1350 1332 200 185 for benzene. As benzene exceeded the data boundaries and was an
extrapolated result, it was excluded from the analysis for the pur-
Naphthalene, 1-methyl- 1350 1350 200 192 pose of this paper.
Heavy range compounds (anthracene and larger) display an in-
creasing discrepancy between their calculated result and their pre-
Acenaphthylene 1550 1497 250 250 dicted result, which is a known issue with n-alkane based reten-
tion indices [38]. As non-polar compounds, n-alkanes have a con-
Acenaphthene 1550 1530 250 250
sistent increase in boiling point with size, which is not affected by
surface interactions. Polar compounds have various surface interac-
Fluorene 1650 1636 250 233 tion effects influencing their boiling points, leading to larger com-
pounds having a boiling point range instead of a fixed point, since
Anthracene 1950 1862 300 283
larger (polar) surfaces allow more interaction effects to take place.
Another factor that may be leading to a consistently lower actual
Fluoranthene 2250 2155 350 340 retention index, as compared to the theoretical result, is the effect
of the average linear velocity on elution temperature of different
Pyrene 2500 2218 400 380 compound groups. As a theoretical approximation, linear velocity
is commonly considered stable in constant flow mode. However,
Benz(a)anthracene 2850 2584 400 414 fluctuations are observed in practice as the temperature program
increases the pressure and the flow is adjusted accordingly. This
Benzo(b)fluoranthene 3375 2912 450 429
becomes more obvious when directly comparing constant flow and
constant pressure configurations [38]. In this method, the inlet
pressure was regulated in constant flow mode.
Benzo(k)fluoranthene 3375 2924 450 429
RI results from other literature sources have several poten-
tial sources of variation. These include compound misidentifica-
tion, column attributes (length, inner diameter, film thickness,
and choice of stationary phase), and system operating parameters
compounds in the top right (benzo(a)pyrene) in Fig. 2. Heavy and [38,39,25]. Initial quality control of results in one-dimensional GC
non-polar compounds, such as heptacosane, remain at the lower is based on the existing RI-boiling point correlation [39,25]. Utiliz-
end of the y-axis. ing the same quality control approach to estimate how well this RI
Using calibration compounds, retention indices were calculated system performed, a regression analysis was applied to calculated
for verification compounds in the standard mixture. Calculated re- results, versus their boiling point in the first dimension (Fig. 3A)
tention indices for these compounds were compared against their and the number of aromatic rings present in the second dimen-
theoretical range based on boiling point and number of aromatic sion (Fig. 3B). The results are plotted in Fig. 3 with their respective
rings. An increase in expected column interaction, which fell be- 95% (dark grey shading) and 90% (light gray shading) upper and
tween aromatic rings, was generalized as an added value of 50 in lower confidence limits.
the expected LI2 . The regressions yielded r2 values of 0.970 for A and 0.988 for B,
Upon comparing structural information of the verification com- and standard errors of 21.1°C and 0.1 aromatic ring units, respec-
pounds to that of calibration compounds, verification compounds tively. Due to the very low standard error, less than 5 data points
fell into expected areas in the second dimension. Using the chem- lie just outside the 95% confidence interval. Using twice the stan-
ical structure, an expected LI2 was derived based on number of dard deviation as acceptance criteria, all compounds pass (except
aromatic rings. for benzene, which had been excluded from further analysis). This
For the majority of compounds, calculated values were com- demonstrates that the KLI system works well within a narrow error
parable with theoretical values (Table 3). However, compounds band. In addition, LI2 shows the capability to predict polarity be-
in the heavy range (as defined for ILR analysis by ASTM E1618) yond the number of aromatic rings, suggesting that further inves-
show more discrepancy between predicted and calculated re- tigation into the potential to differentiate other functional groups
sults, which is likely due to their surface polarity. The addi- may prove worthwhile.
tion of more aromatic rings means increased variety in their
surface shape (e.g. pyrene versus benz(a)anthracene), which has 3.2.1.1. Data boundaries and alignment feature. A visual indicator
a large impact on their surface polarity and therefore on for compounds exceeding physical data boundaries based on a uni-
where these compounds elute in the second dimension. Upon form temperature ramp can be seen in Fig. 2, where heptane, oc-
closer inspection, the largest observed discrepancy is 20 LI2 tane, and triacontane (highlighted in brackets) do not follow the
units (fluoranthene, benz(a)anthracene, benzo(b)fluoranthene and linear outline of remaining n-alkanes in the second dimension. In
benzo(k)fluoranthene), which is within rounding parameters as ex- these cases, RI results should be considered an approximation and
pected results were either considered full aromatic rings (100 LI2 be used with caution. This can present a more noticeable challenge
units) or “half” rings (50 LI2 units). Treating calculated results and in shorter runs. One of the ways to address this challenge is to ap-
predicted results the same, i.e. rounding to the nearest 50, they ply an alignment correction, which is a feature in the ChromSpace
align with their predicted LI2 since the effect of surface polarity software used for data analysis. For alignment, an algorithm is ap-
6
N. Boegelsack, C. Sandau, D.W. McMartin et al. Journal of Chromatography A 1635 (2021) 461717
600 5
500
400
3
300
2
200
100
A 1 B
0 0
500 1000 1500 2000 2500 3000 50 100 150 200 250 300 350 400 450
KI1 LI2
Fig. 3. KI1 -boiling point correlation (A) and LI2 -aromatic ring correlation (B) with their respective 95% (dark grey) and 90% (light grey) confidence limits.
7
N. Boegelsack, C. Sandau, D.W. McMartin et al. Journal of Chromatography A 1635 (2021) 461717
Table 4
ILR classification definitions and their respective translations into KLI coordinates based on ASTM E1618.
Condensed KLI
Class General Alkanes Cycloalkanes Aromatics Ring Aromatics Coordinates
40% 250 L
B
CV%
30% 200
LI2
20% 150 G
10% A D
100
0% C, E, F
50
0
500 1000 1500 2000 2500 3000
KI1
Fig. 6. KLI contour map of ASTM E1618 ILR classes with arrows for light (L),
Fig. 5. Temperature ramp CV% comparison of absolute retention times, as repre- medium (M) and heavy (H) ranges (A - Gasoline, B - Petroleum Distillate, C -
sented by solid black and grey bars for first and second dimension respectively, and Isoparaffinic Products, D – Aromatic Products, E – Naphthenic – Paraffinic Products,
KLI RI, as represented by striped black and dotted grey bars in first and second F – Normal Alkane Products, G – Oxygenated Solvents).
dimension respectively.
8
N. Boegelsack, C. Sandau, D.W. McMartin et al. Journal of Chromatography A 1635 (2021) 461717
A B
350 350
300 300
250 250
200 200
LI2
LI2
150 150
100 100
50 50
0 0
500 1000 1500 2000 2500 3000 500 1000 1500 2000 2500 3000
KI1 KI1
Fig. 7. KLI contour map overlay on Gasoline (A) and Petroleum Distillate (B) in light (black crosses) and heavy (grey dots) range.
Fig. 7A shows a KLI map derived from the class definition of 350
gasoline in ASTM E1618 overlaid on a 21-petroleum station com-
300
posite gasoline sample (black dots). Petroleum distillates can occur
in light, medium, and heavy range, as represented in Fig. 7B by 250
lighter fluid (black crosses), and a 4-petroleum station composite
diesel sample (grey dots), for light and heavy ranges respectively. LI2 200
Fig. 7 shows that the compounds fall into their expected ranges
150
for different classes. While the mapping alone is not always suffi-
cient to distinguish ILR classes, it can highlight additional distin- 100
guishing characteristics at a glance.
50
0
3.3.2. Application to wildfire investigations 500 1000 1500 2000 2500 3000
When investigating potential arson, fire debris samples on KI1
scene must be collected and analyzed promptly not only due to
Fig. 8. Application of KLI ILR classification contour map to a fire scene sample.
the volatile nature of ILRs but also to swiftly assist fire investiga-
tors with the case. However, determining a potential source of the
ILR can take time, and reference samples from a particular source Comparing a scene sample (Fig. 8) with neat ignitable liquids
(e.g. gas station, suspect’s property etc.) for court comparison may (Fig. 7) shows the complexity of sample matrix and pyrolysis ef-
not be analyzed at the same time as the scene samples. In other fects in the abundance of data points. Although petroleum distil-
instances, investigations may point to the same arsonist, and sam- late appears a better fit than gasoline upon first inspection, ad-
ples would benefit from retrospective analysis to compare differ- ditional means of compound identification, such as extracted ion
ent investigations. In rare circumstances, different labs may want profiling and target compound analysis, confirmed the presence of
to compare analysis which cannot be done to this level using rou- both gasoline and diesel in this sample.
tine analysis. The diversity and breadth of compounds in the wildfire sam-
With instrumental analysis, systematic changes regularly take ple (Fig. 8) shows the potential for GC×GC analysis to facilitate ILR
place over time, e.g. column and inlet maintenance, making di- analysis. Comprehensive analysis of all compounds in ILRs, in ad-
rect comparisons based on absolute retention times challenging. dition to the current target compound analysis, brings future po-
Fire debris analysis, including ILR identification, classification and tential to monitor within-class variation and source determination
source comparison, for wildfire investigation would therefore ben- based on marker compound identification within the currently un-
efit from the application of an RI system in several ways. This in- reported wealth of compounds. Using KLI RI values as part of a
cludes negating the effects of systematic changes, utilizing untar- data triangulation (RI coordinates, mass spectral data, and peak
geted chromatography results for chemometric analyses, and estab- abundance) in a reference library creates great potential for apply-
lishing a standardized reference library. ing chemometrics to identify and compare ILRs more readily.
As the application to neat ILRs was successful, the application
of the KLI system for ILR analysis to a scene sample encompassed 4. Conclusions
the next step in evaluating suitability. A previous comparison of a
fire scene sample from 2018 with the gasoline and diesel compos- Building upon the existing workflow of one-dimensional GC
ite samples depicted in Fig. 7 based on absolute retention times RI systems, we developed a simplified but robust RI system for
proved extremely challenging. Between these, a new column had GC×GC. The proposed KLI system demonstrated a very straight-
been installed with different internal diameter, which meant both forward yet successful approach. With a single injection of read-
first and second dimension had shifted substantially. The scene ily available standard compounds, KLI RI can be calculated regard-
sample from 2018 was confirmed to be positive for ILR presence less of modulator restrictions as internal or external standards. The
and the ILR was identified as a mixture of diesel and gasoline. composition of compounds allows a direct comparison to Kovats
Using the KLI system, Fig. 8 shows the overlay of a 2018 fire and Lee indices in one-dimensional GC and pertains to applications
scene sample with the ILR contour map. related to petroleum hydrocarbons. Overall, the KLI workflow was
9
N. Boegelsack, C. Sandau, D.W. McMartin et al. Journal of Chromatography A 1635 (2021) 461717
inexpensive, quick to apply, and user-friendly with its simple 3- [15] P.H. Stefanuto, K.A. Perrault, L.M. Dubois, B. L’Homme, C. Allen, C. Loughnane,
step design. N. Ochiai, J.F. Focant, Advanced method optimization for volatile aroma pro-
filing of beer using two-dimensional gas chromatography time-of-flight mass
The performance was comparable to, if not better than, exist- spectrometry, Journal of Chromatography A 1507 (2017) 45–52 21 7.
ing RI systems, and showed very small error for correlation to pre- [16] W. Ibrahim, M. Wilde, R. Cordell, L. Bryant, R. Free, B. Zhao, M. Richardson, A.
dicted values. System changes did not affect results beyond the Singapuri, N. Greening, P. Monks, C. Brightling, S. Siddiqui, “Comparative Fea-
sibility of Sputum and Exhaled Breath VOCs Sampling in Patients with Severe
limit of acceptability on standards or wildfire analysis samples, in- Exacerbations of Airways Disease,” 2019.
dicating good robustness and suitability for the intended purpose. [17] M. Caldeira, R. Perestrelo, A.S. Barros, M.J. Bilelo, A. Morête, J.S. Câmara,
Combining KLI RI and ILR classification as defined in ASTM S.M. Rocha, Allergic asthma exhaled breath metabolome: A challenge for com-
prehensive two-dimensional gas chromatography, Journal of Chromatography
E1618, the application to wildfire analysis was successful. We de-
A 1254 (2012) 87–97 7 9.
veloped an ILR classification contour map, which was used on neat [18] G.S. Frysinger, R.B. Gaines, C.M. Reddy, GC × GC—a new analytical tool for en-
fuels and a fire scene sample to aid identification and classification vironmental forensics, Environmental Forensics 3 (1) (2002) 27–34.
[19] B. Gruber, B.A. Weggler, R. Jaramillo, K.A. Murrell, P.K. Piotrowski, F.L. Dorman,
of ILRs present. The KLI standard composition allows for a wide
in: Comprehensive two-dimensional gas chromatography in forensic science:
range of applications related to VOC analysis for easy standard- A critical review of recent trends, 105, Elsevier B.V., 2018, pp. 292–301.
ization and comparison of samples, as well as the development of [20] C.A. Manzano, D. Muir, C. Marvin, Separation of thia-arenes and aza-arenes
GC×GC-based databases. from polycyclic aromatics in snowpack samples from the Athabasca oil sands
region by GC×GC/ToF-MS, International Journal of Environmental Analytical
Chemistry 96 (10) (2016) 905–920 8 8.
Declaration of Competing Interest [21] C.M. Taylor, A.K. Rosenhan, J.K. Raines, J.M. Rodriguez, An Arson Investigation
by using Comprehensive Two-dimensional Gas Chromatography-Quadrupole
Mass Spectrometry, Journal of Forensic Research 03 (09) (2012).
The authors declare that they have no known competing finan- [22] K. Nizio, J. Cochran, S. Forbes, Achieving a Near-Theoretical Maximum in
cial interests or personal relationships that could have appeared to Peak Capacity Gain for the Forensic Analysis of Ignitable Liquids Using
influence the work reported in this paper. GC×GC-TOFMS, Separations 3 (3) (2016) 26 1 9.
[23] A.A. Sampat, M. Lopatka, G. Vivó-Truyols, P.J. Schoenmakers, A.C. van Asten, To-
wards chemical profiling of ignitable liquids with comprehensive two-dimen-
CRediT authorship contribution statement sional gas chromatography: Exploring forensic application to neat white spirits,
Forensic Science International 267 (2016) 183–195 1 10.
[24] M. Lopatka, A.A. Sampat, S. Jonkers, L.A. Adutwum, H.G. Mol, G. van der Weg,
Nadin Boegelsack: Conceptualization, Methodology, Investi- J.J. Harynuk, P.J. Schoenmakers, A. van Asten, M.J. Sjerps, G. Vivó-Truyols, Local
gation, Formal analysis, Writing - original draft, Visualization. Ion Signatures (LIS) for the examination of comprehensive two-dimensional
Court Sandau: Conceptualization, Writing - review & editing. gas chromatography applied to fire debris analysis, Forensic Chemistry 3 (2017)
1–13 1 3.
Dena W. McMartin: Writing - review & editing, Funding acquisi- [25] V.I. Babushok, Chromatographic retention indices in identification of chemical
tion. Jonathan M. Withey: Writing - review & editing, Funding ac- compounds, TrAC Trends in Analytical Chemistry 69 (2015) 98–104.
quisition. Gwen O’Sullivan: Conceptualization, Methodology, Vali- [26] L. Lebanov, L. Tedone, M. Kaykhaii, M.R. Linford, B. Paull, Multidimensional Gas
Chromatography in Essential Oil Analysis. Part 2: Application to Characterisa-
dation, Resources, Writing - review & editing, Supervision.
tion and Identification, Chromatographia 82 (1) (2019) 399–414.
[27] J.T. Davidson, G.P. Jackson, The differentiation of
References 2,5-dimethoxy-N-(N-methoxybenzyl)phenethylamine (NBOMe) isomers using
GC retention indices and multivariate analysis of ion abundances in electron
[1] T. Schoennagel, J.K. Balch, H. Brenkert-Smith, P.E. Dennison, B.J. Harvey, ionization mass spectra, Forensic Chemistry 14 (2019) 100–160.
M.A. Krawchuk, N. Mietkiewicz, P. Morgan, M.A. Moritz, R. Rasker, M.G. Turner, [28] C. Stultz, F. Dorman, Characterization of 9 gas chromatography columns by Ko-
C. Whitlock, Adapt to more wildfire in western North American forests as cli- vats and Lee retention indices for dioxin analysis, Journal of Chromatography
mate changes, Proceedings of the National Academy of Sciences 114 (18) (2017) A (2019).
4582–4590 2 5. [29] J.S. Arey, R.K. Nelson, L. Xu, C.M. Reddy, Using Comprehensive Two-Dimen-
[2] S. Gauthier, P. Bernier, P.J. Burton, J. Edwards, K. Isaac, N. Isabel, K. Jayen, H. Le sional Gas Chromatography Retention Indices To Estimate Environmental Par-
Goff, E.A. Nelson, Climate change vulnerability and adaptation in the managed titioning Properties for a Complete Set of Diesel Fuel Hydrocarbons, Analytical
Canadian boreal forest, Environmental Reviews 22 (3) (2014) 256–285 28 3. Chemistry 77 (22) (2005) 7172–7182 15 11.
[3] R.P. Guyette, F.R. Thompson, J. Whittier, M.C. Stambaugh, D.C. Dey, Future Fire [30] Y. Zushi, Y. Yamatori, J. Nagata, D. Nabi, Comprehensive two-dimensional
Probability Modeling with Climate Change Data and Physical Chemistry, Forest gas-chromatography-based property estimation to assess the fate and behavior
Science 60 (5) (2014) 862–870 29 5. of complex mixtures: A case study of vehicle engine oil, Science of the Total
[4] American Society for Testing and Materials ASTM E1618 - 14, ASTM Interna- Environment 669 (2019) 739–745 15 6.
tional (2014). [31] R. Jaramillo, F.L. Dorman, Retention time prediction in thermally modu-
[5] E. Stauffer, Concept of pyrolysis for fire debris analysts, Science and Justice - lated comprehensive two-dimensional gas chromatography: Correcting sec-
Journal of the Forensic Science Society 43 (1) (2003) 29–40. ond dimension retention time modeling error, Journal of Chromatography A
[6] J. Baerncopf, K. Hutches, A review of modern challenges in fire debris analysis, 1581-1582 (2018) 116–124 Vols.21 12.
Forensic Science Internationa 244 (2014) e12–e20 1 11. [32] R.J. Western, P.J. Marriott, Methods for generating second dimension retention
[7] G.S. Frysinger, R.B. Gaines, Forensic analysis of ignitable liquids in fire debris by index data in comprehensive two-dimensional gas chromatography, Journal of
comprehensive two-dimensional gas chromatography, Journal of forensic sci- Chromatography A (2003).
ences 47 (3) (2002) 471–482 5. [33] J.Y. Gardner, D.E. Brillhart, M.M. Benjamin, L.G. Dixon, L.M. Mitchell, J.-M.D. Di-
[8] E. Stauffer, J.A. Dolan, R. Newman, Other Techniques of Analysis and the Future mandja, The use of GC×GC/TOF MS with multivariate analysis for the charac-
of Fire Debris Analysis, Fire Debris Analysis (2007). terization of foodborne pathogen bacteria profiles, Journal of Separation Sci-
[9] L.N. Kates, P.I. Richards, C.D. Sandau, The application of comprehensive two-di- ence 34 (2) (2011) 176–185 1.
mensional gas chromatography to the analysis of wildfire debris for ignitable [34] C. Veenaas, P. Haglund, A retention index system for comprehensive two-di-
liquid residue, Forensic Science International 310 (2020) 110256 1 5. mensional gas chromatography using polyethylene glycols, Journal of Chro-
[10] M.S. Klee, J. Cochran, M. Merrick, L.M. Blumberg, Evaluation of conditions of matography A 1536 (2018) 67–74 9 2.
comprehensive two-dimensional gas chromatography that yield a near-theo- [35] H. van Den Dool, P. Dec. Kratz, A generalization of the retention index sys-
retical maximum in peak capacity gain, Journal of Chromatography A 1383 tem including linear temperature programmed gas-liquid partition chromatog-
(2015) 151–159. raphy, Journal of Chromatography A 11 (C) (1963) 463–471.
[11] C. Duhamel, P. Cardinael, V. Peulon-Agasse, R. Firor, L. Pascaud, G. Se- [36] M.L. Lee, D.L. Vassilaros, C.M. White, Retention indices for programmed-tem-
mard-Jousset, P. Giusti, V. Livadaris, Comparison of cryogenic and differential perature capillary-column gas chromatography of polycyclic aromatic hydro-
flow (forward and reverse fill/flush) modulators and applications to the anal- carbons, Analytical Chemistry 51 (6) (1979) 768–773 5.
ysis of heavy petroleum cuts by high-temperature comprehensive gas chro- [37] C. Cordero, P. Rubiolo, B. Sgorbini, M. Galli, C. Bicchi, Comprehensive two-di-
matography, Journal of Chromatography A 1387 (2015) 95–103 27 3. mensional gas chromatography in the analysis of volatile samples of natural
[12] K.D. Nizio, T.M. McGinitie, J.J. Harynuk, Comprehensive multidimensional sep- origin: A multidisciplinary approach to evaluate the influence of second di-
arations for the analysis of petroleum, vol. 1255, 2012, pp. 12-23. mension column coated with mixed stationary phases on system orthogonal-
[13] C. Von Mühlen, C.A. Zini, E.B. Caramão, P.J. Marriott, Applications of compre- ity, Journal of Chromatography A 1132 (1-2) (2006) 268–279.
hensive two-dimensional gas chromatography to the characterization of petro- [38] H. Miyagawa, K. Nakagawa, K. Kadokami, Reproducibility of Programmed-Tem-
chemical and related samples, Journal of Chromatography A (2006). perature Retention Indices under Average Linear Velocity Carrier Gas Control
[14] M. Jiang, C. Kulsing, Y. Nolvachai, P.J. Marriott, Two-Dimensional Retention In- of GC and GC–MS, Chromatographia 73 (9) (2011) 953–963.
dices Improve Component Identification in Comprehensive Two-Dimensional [39] W.P. Eckel, T. Kind, Use of boiling point–Lee retention index correlation for
Gas Chromatography of Saffron, Analytical Chemistry 87 (11) (2015) 5753–5761 rapid review of gas chromatography-mass spectrometry data, Analytica Chim-
2 6. ica Acta 494 (1) (2003) 235–243.
10