Calibration Report

REPORT FOR IM-MT-001

Details of the IMTE


Instrument's name: Precision Level
Serial Number: 242543
Manufacturer: Riken Keisokuki

1.0 Calibration History


Actual Reading = Instrument Reading + Correction
Correction = Actual Reading - Instrument Reading
Correction
Uncertainty = Instruments Re ading x100

1st calibration: 9.June.2015


Uncertainty: ±0.011mm
Description: Left Scale&Right
scale
Calibration Method : LVDT SD068,
Inclination Table SD047
Uncertainty Correction
No (mm)
+ -
1 0.011 -0.011 -0.0068
2 0.011 -0.011 -0.0059
3 0.011 -0.011 -0.0028
4 0.011 -0.011 -0.0027
5 0.011 -0.011 -0.0075
6 0.011 -0.011 -0.0052
1 0.011 -0.011 -0.0062
2 0.011 -0.011 -0.0071
3 0.011 -0.011 -0.0028
4 0.011 -0.011 -0.0034
5 0.011 -0.011 -0.0012
6 0.011 -0.011 -0.0018

QA/QC Department Page 1 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-001

2.0 Calibration Graph

1st calibration (Left scale)


0.015

0.01
Upper limit
0.005 Lower limit
Correction
0
1 2 3 4 5 6
-0.005

-0.01

-0.015

1st calibration (right scale)


0.015

0.01
Upper limit
0.005 Lower limit
Correction
0
1 2 3 4 5 6
-0.005

-0.01

-0.015

QA/QC Department Page 2 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-001

3.0 Calibration Report

Based on Calibration Certificate Result:

As refer to the graph above, it shows that the correction line is within the upper limit
and the lower limit which represent by the uncertainty. But to come out with solid
analysis of the data provided, at least we need to look at history of three previous
calibration as mention by Allen Bare. Calibration intervals shall be established for
calibrations based on repeatability and degree of usage, as mention API 6D. Based on
the IMTE usage monitoring on December 2015, the instrument is never use during
December 2015.

To analyze an accurate data, 3 calibration history must be obtained. But due to the time
interval between the 1st calibration certificate until now (18.1.2016), the data obtained
from the 1st calibration cannot be used and the data collection need to be start again.

QA/QC Department Page 3 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-001

4.0 Calibration Frequency

Calibration Frequency :

The calibration interval shall be 3 months for the first 3 calibration.

According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom use.

The status is : On Hold

*) The status for this instrument is on hold, when needed to use this instrument will
send for calibration before use..

QA/QC Department Page 4 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-001

5.0 Reference

Reference:
[1] Bare A. (2006), Simplified Calibration Interval Analysis. U.S
[2] API Specification 6D , Specification for Pipeline and Piping Valves 24th Edition,
August 2014

QA/QC Department Page 5 PT Borneo Mandiri Investment