Calibration Report

REPORT FOR IM-MT-002 (NOT IN USE)


Details of the IMTE
Instrument's name: Analog Outside Micrometer
Serial Number: 00000315
Manufacturer: MC

1.0 Calibration History


Actual Reading = Instrument Reading + Correction
Correction = Actual Reading - Instrument Reading
Correction
Uncertainty = Instruments Re ading x100

1st calibration: 15.April.2014 2nd Calibration: 1.October.2014 3rd calibration: 6 April 2015
Uncertainty: ±0.004mm Uncertainty: ±0.012mm Uncertainty: ±0.003mm
Description: - Description: - Description: -
Calibration Method : JIS B7502 Calibration Method : JIS B7502 Calibration Method : JIS B7502
Uncertainty Correction Uncertainty Correction Uncertainty Correction
No (mm) No (mm) No (mm)
+ - + - + -
1 0.004 -0.004 0 1 0.012 -0.012 0 1 0.003 -0.003 0
2 0.004 -0.004 0 2 0.012 -0.012 0 2 0.003 -0.003 0
3 0.004 -0.004 0 3 0.012 -0.012 0 3 0.003 -0.003 0
4 0.004 -0.004 0 4 0.012 -0.012 0.004 4 0.003 -0.003 0
5 0.004 -0.004 0 5 0.012 -0.012 0.004 5 0.003 -0.003 0.004
6 0.004 -0.004 0.002 6 0.012 -0.012 0.008 6 0.003 -0.003 0.008
7 0.004 -0.004 0.002 7 0.012 -0.012 0.010 7 0.003 -0.003 0.020
8 0.004 -0.004 0.010 8 0.012 -0.012 0.010 8 0.003 -0.003 0.020
9 0.004 -0.004 0.010 9 0.012 -0.012 0.010 9 0.003 -0.003 0.020
10 0.004 -0.004 0.010 10 0.012 -0.012 0.010 10 0.003 -0.003 0.020
11 0.004 -0.004 0.010 11 0.012 -0.012 0.010 11 0.003 -0.003 0.010

QA/QC Department Page 1 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-002

2.0 Calibration Graph

1st Calibration
0.012
0.01
0.008 Upper limit
0.006 Lower limit
0.004 Correction
0.002
0
-0.002 1 2 3 4 5 6 7 8 9 10 11
-0.004
-0.006

2nd calibration
0.015

0.010
Upper limit
0.005 Lower limit
Correction
0.000
1 2 3 4 5 6 7 8 9 10 11
-0.005

-0.010

-0.015

3rd calibration
0.025

0.020
Upper limit
0.015 Lower limit
Correction
0.010

0.005

0.000
1 2 3 4 5 6 7 8 9 10 11
-0.005

QA/QC Department Page 2 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-002

3.0 Calibration Report

Based on Calibration Certificate Result :


As shown in section 2.0 above, shows 3 graph which construct by using data taken from
calibration certificate from 15.April.2014, 1.October.2014, and 6.April.2015. The graph
shows that in the first calibration most of the point is inside the upper limit and lower limit
except for the point from 8th until 11th. This may due to some error present when
calibration is being conduct. For 2nd calibration, the point is within the upper limit and
lower limit but for 3rd calibration it show that most of the point is outside the upper limit.
But the data from the next calibration which is on 1.10.2014 cannot be use due to the
interval between the date is exceed 3-4 month period. This may effect the calibration
analysis for 3 times data for 3 month each. Although the instrument may use by
production/ QC team but the instrument never use, so the instrument does not need to send
out for calibration. It will affect the cost effective.
Calibration intervals shall be established for calibrations based on repeatability and degree
of usage, as mention API 6D. Based on the IMTE usage monitoring on December 2015, the
instrument is never use during December 2015.

QA/QC Department Page 3 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-002

4.0 Calibration Frequency

Calibration Frequency :
The calibration interval shall be 3 months. As per section 3.0 Calibration Report above that
this instrument is never use.
The status is : Not In Use
*) The status for this instrument is not in use, when needed to use this instrument will send
for calibration before use.

QA/QC Department Page 4 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-002

5.0 Reference

[1] API Specification 6D , Specification for Pipeline and Piping Valves 24th Edition, August
2014