Im MT 002
Im MT 002
Im MT 002
1st calibration: 15.April.2014 2nd Calibration: 1.October.2014 3rd calibration: 6 April 2015
Uncertainty: ±0.004mm Uncertainty: ±0.012mm Uncertainty: ±0.003mm
Description: - Description: - Description: -
Calibration Method : JIS B7502 Calibration Method : JIS B7502 Calibration Method : JIS B7502
Uncertainty Correction Uncertainty Correction Uncertainty Correction
No (mm) No (mm) No (mm)
+ - + - + -
1 0.004 -0.004 0 1 0.012 -0.012 0 1 0.003 -0.003 0
2 0.004 -0.004 0 2 0.012 -0.012 0 2 0.003 -0.003 0
3 0.004 -0.004 0 3 0.012 -0.012 0 3 0.003 -0.003 0
4 0.004 -0.004 0 4 0.012 -0.012 0.004 4 0.003 -0.003 0
5 0.004 -0.004 0 5 0.012 -0.012 0.004 5 0.003 -0.003 0.004
6 0.004 -0.004 0.002 6 0.012 -0.012 0.008 6 0.003 -0.003 0.008
7 0.004 -0.004 0.002 7 0.012 -0.012 0.010 7 0.003 -0.003 0.020
8 0.004 -0.004 0.010 8 0.012 -0.012 0.010 8 0.003 -0.003 0.020
9 0.004 -0.004 0.010 9 0.012 -0.012 0.010 9 0.003 -0.003 0.020
10 0.004 -0.004 0.010 10 0.012 -0.012 0.010 10 0.003 -0.003 0.020
11 0.004 -0.004 0.010 11 0.012 -0.012 0.010 11 0.003 -0.003 0.010
1st Calibration
0.012
0.01
0.008 Upper limit
0.006 Lower limit
0.004 Correction
0.002
0
-0.002 1 2 3 4 5 6 7 8 9 10 11
-0.004
-0.006
2nd calibration
0.015
0.010
Upper limit
0.005 Lower limit
Correction
0.000
1 2 3 4 5 6 7 8 9 10 11
-0.005
-0.010
-0.015
3rd calibration
0.025
0.020
Upper limit
0.015 Lower limit
Correction
0.010
0.005
0.000
1 2 3 4 5 6 7 8 9 10 11
-0.005
Calibration Frequency :
The calibration interval shall be 3 months. As per section 3.0 Calibration Report above that
this instrument is never use.
The status is : Not In Use
*) The status for this instrument is not in use, when needed to use this instrument will send
for calibration before use.
5.0 Reference
[1] API Specification 6D , Specification for Pipeline and Piping Valves 24th Edition, August
2014