Thermal Oxidation Stability of Aviation Turbine Fuels: Standard Test Method For

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 26

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.

Contact ASTM International (www.astm.org) for the latest information

Designation: D3241 − 20c

Designation 323/20

Standard Test Method for


Thermal Oxidation Stability of Aviation Turbine Fuels1,2
This standard is issued under the fixed designation D3241; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

This standard has been approved for use by agencies of the U.S. Department of Defense.

1. Scope* 2. Referenced Documents


1.1 This test method covers the procedure for rating the 2.1 ASTM Standards:3
tendencies of gas turbine fuels to deposit decomposition D1655 Specification for Aviation Turbine Fuels
products within the fuel system. D4057 Practice for Manual Sampling of Petroleum and
1.2 The differential pressure values in mm Hg are defined Petroleum Products
only in terms of this test method. D4306 Practice for Aviation Fuel Sample Containers for
Tests Affected by Trace Contamination
1.3 The deposition values stated in SI units shall be re- E177 Practice for Use of the Terms Precision and Bias in
garded as the referee value. ASTM Test Methods
1.4 The pressure values stated in SI units are to be regarded E691 Practice for Conducting an Interlaboratory Study to
as standard. The psi comparison is included for operational Determine the Precision of a Test Method
safety with certain older instruments that cannot report pres- 2.2 ISO Standards:4
sure in SI units. ISO 3274 Geometrical Product Specifications (GPS)—
Surface texture: Profile method—Nominal characteristics
1.5 No other units of measurement are included in this
of contact (stylus) instruments
standard.
ISO 4288 Geometrical Product Specifications (GPS)—
1.6 This standard does not purport to address all of the Surface texture: Profile method—Rules and procedures
safety concerns, if any, associated with its use. It is the for the assessment of surface texture
responsibility of the user of this standard to establish appro- 2.3 ASTM Adjuncts:5
priate safety, health, and environmental practices and deter- Color Standard for Heater Tube Deposit Rating
mine the applicability of regulatory limitations prior to use.
For specific warning statements, see 6.1.1, 7.1, 7.3, 12.1.1, and 3. Terminology
Annex A6.
3.1 Definitions of Terms Specific to This Standard:
1.7 This international standard was developed in accor-
3.1.1 deposits, n—oxidative products laid down on the test
dance with internationally recognized principles on standard-
area of the heater tube or caught in the test filter, or both.
ization established in the Decision on Principles for the
3.1.1.1 Discussion—Fuel deposits will tend to predominate
Development of International Standards, Guides and Recom-
at the hottest portion of the heater tube, which is between the
mendations issued by the World Trade Organization Technical
30 mm and 50 mm position.
Barriers to Trade (TBT) Committee.
3.1.2 heater tube, n—an aluminum coupon controlled at
elevated temperature, over which the test fuel is pumped.
1
3.1.2.1 Discussion—The heater tube is resistively heated
This test method is under the jurisdiction of ASTM International Committee
D02 on Petroleum Products, Liquid Fuels, and Lubricants and is the direct
and controlled in temperature by a thermocouple positioned
responsibility of Subcommittee D02.J0.03 on Combustion and Thermal Properties.
The technically equivalent standard as referenced is under the jurisdiction of the
3
Energy Institute Subcommittee SC-B-8. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Dec. 1, 2020. Published December 2020. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1973. Last previous edition approved in 2020 as D3241 – 20b. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/D3241-20C. the ASTM website.
2 4
This test method has been developed through the cooperative effort between Available from International Organization for Standardization (ISO), 1, ch. de
ASTM and the Energy Institute, London. ASTM and IP standards were approved by la Voie-Creuse, CP 56, CH-1211 Geneva 20, Switzerland, http://www.iso.org.
5
ASTM and EI technical committees as being technically equivalent but that does not Available from ASTM International Headquarters. Order Adjunct No.
imply both standards are identical. ADJD3241. Original adjunct produced in 1986.

*A Summary of Changes section appears at the end of this standard


Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 1


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
inside. The critical test area is the thinner portion, 60 mm in 6.2 Heater Tube Deposit Rating Apparatus:
length, between the shoulders of the heater tube. Fuel inlet to 6.2.1 Visual Heater Tube Rater (VTR), the tuberator de-
the heater tube is at the 0 mm position, and fuel exit is at scribed in Annex A1.
60 mm. 6.2.2 Standardization of Metrology Requirements:
3.2 Abbreviations: 6.2.2.1 Number of Measured Points—1200 in the ratable
3.2.1 ΔP—differential pressure. area of the heater tube (between 5 mm and 55 mm above the
bottom shoulder of the heater tube).
4. Summary of Test Method (1) Circumferential Resolution—(number of points mea-
4.1 This test method for measuring the high temperature sured on the heater tube circumference), 24 points equally
stability of gas turbine fuels uses an instrument that subjects spaced.
the test fuel to conditions that can be related to those occurring (2) Longitudinal Resolution—(number of points measured
in gas turbine engine fuel systems. The fuel is pumped at a on the 50 mm ratable length of the heater tube), 50 points
fixed volumetric flow rate through a heater, after which it equally spaced.
enters a precision stainless steel filter where fuel degradation 6.2.2.2 Standard Spot—Thickest average deposit area de-
products may become trapped. scribed by either a 2×3 or 3×2 (longitudinal × circumferential)
4.1.1 The apparatus uses 450 mL of test fuel ideally during arrangement of adjoining thickness measurement points,
a 2.5 h test. The essential data derived are the amount of amongst the 1200 measured by the metrology techniques.
deposits on a heater tube, and the rate of plugging of a 17 µm 6.2.3 Interferometric (Heater) Tube Rater (ITR)—the tu-
nominal porosity precision filter located just downstream of the berator described in Annex A2.
heater tube. 6.2.4 Ellipsometric (Heater) Tube Rater (ETR)—the tubera-
tor described in Annex A3.
5. Significance and Use 6.2.5 Multi-Wavelength Ellipsometric (Heater) Tube Rater
5.1 The test results are indicative of fuel performance (MWETR)—the tuberator described in Annex A4.
during gas turbine operation and can be used to assess the level 6.3 Because jet fuel thermal oxidation stability is defined
of deposits that form when liquid fuel contacts a heated surface only in terms of this test method, which depends upon, and is
that is at a specified temperature. inseparable from, the specific equipment used, the test method
6. Apparatus shall be conducted with the equipment used to develop the test
method or equivalent equipment.
6.1 Aviation Fuel Thermal Oxidation Stability Tester6—
Five models of suitable equipment may be used as indicated in 7. Reagents and Materials
Table 1.
6.1.1 Portions of this test may be automated. Refer to the 7.1 Use methyl pentane, 2,2,4-trimethylpentane, or
appropriate user manual for the instrument model to be used n-heptane as a general cleaning solvent. General cleaning
for a description of detailed procedure. A manual is provided solvent shall be 95 mol % purity, minimum. This solvent will
with each test rig. (Warning—No attempt should be made to effectively clean internal metal surfaces of apparatus before a
operate the instrument without first becoming acquainted with test, especially those surfaces (before the heater tube test
all components and the function of each.) section) that contact fresh sample. (Warning—Extremely
6.1.2 Certain operational parameters used with the instru- flammable. Harmful if inhaled (see Annex A6).)
ment are critically important to achieve consistent and correct 7.2 Use a nylon bristle brush and trisolvent to clean internal
results. These are listed in Table 2. (working) surface of heater tube test section only. Trisolvent is
an equal mix of acetone (1), toluene (2), and isopropanol (3).
6
The following equipment, as described in Table 1 and RR:D02-1309, was used
All three components of trisolvent shall be 95 mol % purity,
to develop this test method. The following equipment, as described in Table 1 and minimum. (Warning—(1) Extremely flammable, vapors may
determined as equivalent in testing as detailed in RR:D02-1631, is provided by PAC, cause flash fire; (2) and (3) Flammable. Vapors of all three
8824 Fallbrook Drive, Houston, TX 77064. The following equipment, as described harmful. Irritating to skin, eyes, and mucous membranes.) Use
in Table 1 and determined as equivalent in testing as detailed in RR:D02-1728, is
provided by Falex Corporation, 1020 Airpark Dr., Sugar Grove, IL, 60554-9585. a nylon bristle brush that makes stiff contact with the inner
This is not an endorsement or certification by ASTM International. walls of the heater tube test section.

TABLE 1 Instrument Models


Instrument Model Pressurize With Principle Differential Pressure by
230A hydraulic syringe Transducer + Printout
240A hydraulic syringe Transducer + Printout
230 Mk IIIB hydraulic dual piston (HPLC Type) Transducer + Printout
F400C hydraulic dual piston (HPLC Type) Transducer + Printout
230 Mk IVD hydraulic single piston (HPLC Type) Transducer + Printout
A
See RR:D02-1309.
B
See RR:D02-1631.
C
See RR:D02-1728.
D
See RR:D02-1757.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 2


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
TABLE 2 Critical Operating Characteristics of D3241
Item Definition
Test apparatus Tube-in-shell heat exchanger as illustrated in Fig. 1 and dimen-
sions in Fig. A5.1.
Heater Tube:
Heater tube A, B, C, D, E Specially fabricated heater tube that produces controlled heated
test surface; new one for each test.

Heater Tube identificationE Each heater tube shall be physically identified with a unique serial
number, identifying the manufacturer and providing traceability to
the original material batch.

Heater Tube metallurgy 6061-T6 Aluminum, plus the following criteria


a) The Mg:Si ratio shall not exceed 1.9:1
b) The Mg2Si percentage shall not exceed
1.85 %

Heater Tube surface polish over circumference of center section Rotational cut buffing technique with polishing compound to
achieve mechanical surface finish.

Heater Tube dimensions: Dimension Tolerance


Heater Tube length,F mm 161.925 ±0.254
Center section length, mm 60.325 ±0.051
Outside diameters, mm
Shoulders 4.724 ±0.025
Center section 3.175 ±0.051
Inside diameter, mm 1.651 ±0.051
Total indicator runout, mm, max 0.013
Mechanical surface finish, nm, over circumference in center 50 ± 20
section in accordance with ISO 3274 and ISO 4288 using the
mean of four 1.25–measurements
Test filter 6 nominal 17 µm stainless steel mesh filter element to trap deposits;
new one for each test
Stainless Steel Mesh Twilled Dutch Weave, 304 Stainless Steel, 165 × 1400 Mesh (tol-
erance; 4 % on 1400 and 2 % on 165) with Warp Diameter =
0.0028 in. and Shute Diameter = 0.0016 in.
Instrument parameters:
Sample volume 600 mL of sample is aerated, then 450 mL ± 45 mL of this aerated
fuel shall be pumped during the heating phase for a valid test
Aeration rate 1.5 L ⁄min dry air through sparger
Flow during test 3.0 mL ⁄min ± 10 % (2.7 mL/min minimum to 3.3 mL/min maxi-
mum)
Pump mechanism positive displacement or piston syringe
Cooling bus bars fluid cooled to maintain consistent tube temperature pro-
file
Thermocouple (TC) Type J, Inconel sheathed, or Type K, Inconel sheathed
Operating pressure:
System 3.45 MPa ± 10 % on sample by hydraulically transmitted force
against control valve outlet restriction
At test filter differential pressure (ΔP) measured across test filter by electronic
transducer in mm Hg
Operating temperature:
For test as stated in specification for fuel
Uniformity of run maximum deviation of ±2 °C from specified temperature
Calibration Models 230 and 240T – Three point calibration including pure tin
at 232 °C, pure lead at 327 °C for high point and ice + water for
low point reference
All other models – Two point calibration using pure lead at 327 °C
for high point and ice + water for low point reference
A
D3241/IP 323 Thermal Stability is a critical aviation fuel test, the results of which are used to assess the suitability of jet fuel for aviation operational safety and regulatory
compliance. The integrity of D3241/IP 323 testing requires that heater tubes meet the regulations of D3241 Table 2 and give equivalent D3241 results to the heater tubes
supplied by the original equipment manufacturer (OEM).
B
The following equipment, heater tubes, manufactured by PAC, 8824 Fallbrook Drive, Houston, TX 77064, was used in the development of this test method. This is not
an endorsement or certification by ASTM International.
C
A test protocol to establish equivalence of heater tubes is on file at ASTM International Headquarters and may be obtained by requesting Research Report RR:D02-1550.
D
The following equipment, heater tube and filter kits, manufactured by Falex Corporation, 1020 Airpark Dr., Sugar Grove, IL, 60554-9585, was run through the test
protocol in RR:D02-1550 and determined as equivalent to the equipment used to develop the test method. This test is detailed in RR:D02-1714. This is not an endorsement
or certification by ASTM International.
E
An electronic recording device, such as a radio-frequency identification device (RFID), may be embedded into the heater tube rivet located at the bottom of the heater
tube. Tube identification data may be stored on an electronic recording device, such as a RFID, embedded into the heater tube.
F
Tube length measurements are only applicable to the aluminum portion of the heater tube. Additions, such as an RFID, do not contribute to the length measurement of
the heater tube.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 3


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
9.1.5 Fuel System Prefilter Element, filter paper of 0.45 µm
pore size.
9.1.6 Heater Tube Control Temperature, preset as specified
in applicable specification.
9.1.7 Fuel Flow Rate, 3.0 mL ⁄min 6 10 %.
9.1.8 Minimum Fuel Pumped During Test, 405 mL.
9.1.9 Test Duration, 150 min 6 2 min.
9.1.10 Cooling Fluid Flow, approximately 39 L ⁄h, or center
of green range on cooling fluid meter.
9.1.11 Power Setting, internally set for computer models.

FIG. 1 Standard Heater Tube Test Section, Essential to All D3241 10. Preparation of Apparatus
Test Instruments
10.1 Cleaning and Assembly of Heater Tube Test Section:
10.1.1 Clean the inside surface of the heater tube test section
using a nylon brush saturated with trisolvent to remove all
7.3 Use dry calcium sulfate + cobalt chloride granules (97 + deposits. Replace the nylon brush when it shows signs of wear
3 mix) or other self-indicating drying agent in the aeration (such as missing bristles) and no longer makes stiff contact
dryer. This granular material changes gradually from blue to with inner walls of heater tube test section.
pink color indicating absorption of water. (Warning—Do not 10.1.2 Check the heater tube to be used in the test for
inhale dust or ingest. May cause stomach disorder.) surface defects and straightness by referring to the procedure in
A1.10. Be careful, also, to avoid scratching the heater tube
8. Sampling shoulder during the examination, since the heater tube shoulder
8.1 General Requirements—The choice of construction ma- must be smooth to ensure a seal under the flow conditions of
terials is an important factor, particularly in the case of aviation the test.
turbine fuel, where thermal stability can be degraded by the 10.1.3 Assemble the heater tube test section using new
presence of very low concentration of copper. The use of items: (1) visually checked heater tube, (2) test filter, and (3)
copper or copper alloys shall be eliminated wherever possible three O-rings. Inspect insulators to be sure they are undam-
by the use of alternative materials such as stainless steel or aged.
aluminum. Zinc and cadmium are two other metals that
adversely affect product quality. Copper, copper alloys (such as NOTE 1—Heater tubes must not be reused. Tests indicate that magne-
sium migrates to the heater tube surface under normal test conditions.
brass), zinc-rich (galvanized) coatings, cadmium alloys, and Surface magnesium may reduce adhesion of deposits to reused heater
cadmium plating shall not be used in sampling aviation fuels tube.
for evaluation under this test method.
10.1.4 During assembly of the heater tube test section,
8.1.1 Sampling Containers—Sampling containers shall be
handle the heater tube carefully so as not to touch center part
in accordance with the requirements for thermal stability
of heater tube. IF THE CENTER OF HEATER TUBE IS
testing described in Practice D4306.
TOUCHED, REJECT THE HEATER TUBE SINCE THE
8.1.2 Sampling Equipment—Sampling equipment fabricated
CONTAMINATED SURFACE MAY AFFECT THE
from copper or its alloys shall not be used for sampling in
DEPOSIT-FORMING CHARACTERISTICS OF THE
accordance with this test method. Sampling equipment shall be
HEATER TUBE.
in accordance with the requirements for aviation fuel described
in Practice D4057. 10.2 Cleaning and Assembly of Remainder of Test Compo-
nents:
9. Standard Operating Conditions 10.2.1 Perform the following steps in the order shown prior
9.1 Standard conditions of the test method are as follows: to running a subsequent test.
9.1.1 Fuel Quantity, 450 mL minimum for test plus about
NOTE 2—It is assumed that the apparatus has been disassembled from
50 mL for system. previous test (see Annex A5 or appropriate user manual for assembly/
9.1.2 Fuel Pre-Treatment—Filter the fuel through a single disassembly details).
layer of “qualitative” quality, medium-flow, cellulose filter
10.2.2 Inspect and clean components that contact test
paper. Filter should be a pleated (folded) circular shape and
sample and replace any seals that are faulty or suspect.
fitted into a conical funnel to filter fuel. Filter paper grades 2V,
MN514, or better, are recommended. After filtration, aerate the 10.2.3 Install the prepared heater tube test section (as
fuel for 6 min at 1.5 L ⁄min air flow rate for a maximum of described in 10.1.1 – 10.1.4).
1000 mL sample using a coarse 12 mm borosilicate glass gas 10.2.4 Assemble pre-filter with new element and install.
dispersion tube. 10.2.5 Check thermocouple for correct reference position,
9.1.3 Fuel System Pressure, 3.45 MPa (500 psi) 610 % then lower into standard operating position.
gauge. 10.2.6 On Models 230 and 240, make sure the water beaker
9.1.4 Thermocouple Position, at 39 mm. is empty.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 4


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
11. Calibration and Standardization Procedure NOTE 6—When counting drop rate, the first drop is counted as drop 0,
and time is started. As drop 20 falls, total time is noted.
11.1 Perform checks of key components at the frequency
indicated in the following (see Annexes or user manual for 12.5 Test:
details). 12.5.1 Record filter pressure drop every 30 min minimum
11.1.1 Thermocouple—Calibrate a thermocouple when first during the test period.
installed and then at least every 6 months (see A5.2.8). 12.5.2 If the filter pressure drop begins to rise sharply and it
11.1.2 Differential Pressure Cell—Standardize once a year is desired to run a full 150 min test, a bypass valve common to
or when installing a new cell (see A5.2.6). all models must be opened in order to finish the test. See
11.1.3 Aeration Dryer—Check at least monthly and change appropriate User Manual for details on operation of the bypass
if color indicates significant absorption of water (see 7.3). system (see A5.2.2).
11.1.4 Metering Pump—Perform two checks of flow rate for 12.5.3 Make another flow check within final 15 min before
each test as described in Section 12. shutdown (see 12.4.3 and accompanying note). (See X1.5.)
12.6 Heater Tube Profile—If a heater tube temperature
12. Procedure profile is desired, obtain as described in X1.4.
12.1 Preparation of Fuel Test Sample: 12.7 Shutdown:
12.1.1 Filter and aerate sample using standard operating
12.7.1 Shut down the instrument; some models may do this
conditions (see A5.2.9). (Warning—All jet fuels must be
automatically.
considered flammable except JP5 and JP7. Vapors are harmful
12.7.1.1 For applicable models after shutdown, turn FLOW
(see A6.3, A6.6, and A6.7).)
SELECTOR VALVE to VENT to relieve pressure.
NOTE 3—Before operating, see Warning in 6.1.1. 12.7.1.2 For Models 230 and 240, the piston actuator will
NOTE 4—Test method results are known to be sensitive to trace retreat automatically.
contamination from sampling containers. For recommended containers,
refer to Practice D4306. 12.7.1.3 Measure the amount of spent fluid pumped during
the test.
12.1.2 Maintain temperature of sample between 15 °C and
(1) For the Models 230 and 240, measure effluent water in
32 °C during aeration.
drip receiver, then empty.
12.1.3 Allow no more than 1 h to elapse between the end of
(2) For other models, measure the fuel in the receipt
aeration and the start of the heating of the sample.
container.
12.2 Final Assembly: 12.7.1.4 If the amount of water or fuel measured is less than
12.2.1 Check all lines to ensure tightness. 405 mL, the test shall be rejected.
12.2.2 Recheck thermocouple position at 39 mm.
12.2.3 Make sure drip receiver is empty. 12.8 Disassembly:
12.8.1 Disconnect fuel inlet line to the heater tube test
12.3 Power Up and Pressurization: section.
12.3.1 Turn POWER to ON. 12.8.2 Disconnect the heater tube test section.
12.3.2 Inspect the system for leaks. Depressurize the system
12.8.2.1 Remove the heater tube from the heater tube test
as necessary to tighten any leaking fittings.
section carefully so as to avoid touching the center part of
12.3.3 Set controls to the standard operating conditions.
heater tube, and discard the test filter.
12.3.4 Use a heater tube control temperature as specified for
12.8.2.2 Flush the heater tube with recommended general
the fuel being tested. Apply any thermocouple correction from
cleaning solvent (see 7.1) from top down. If the heater tube is
the most recent calibration (see A5.2.8).
grasped from the top, do not wash solvent over gloves or bare
NOTE 5—The test can be run to a maximum tube temperature of about fingers. Allow to dry, return the heater tube to original
350 °C. The temperature at which the test should be run and the criteria for container, mark with identification and hold for evaluation.
judging results are normally embodied in fuel specifications.
12.8.3 Disconnect and remove any containers.
12.4 Start Up: 12.8.3.1 Discard water and fuel to waste disposal.
12.4.1 Use procedure for each model as described in the 12.8.4 Disconnect gas dispersion tube (i.e., sparger).
appropriate User Manual.
12.8.4.1 Flush gas dispersion tube (i.e., sparger) with rec-
12.4.2 Some instrument models may do the following steps
ommended general cleaning solvent (see 7.1). Allow solvent to
automatically, but verify that:
fully dry before using the gas dispersion tube (i.e., sparger) for
12.4.2.1 No more than 1 h maximum elapses from aeration
another test.
to start of heating.
12.4.2.2 The test filter bypass valve is closed as soon as the
13. Heater Tube Evaluation
heater tube temperature reaches the test level, so fuel flows
through the test filter (see A5.2.6). 13.1 Rate the deposits on heater tube in accordance with
12.4.2.3 DP transducer is set to zero once DP stabilizes (see Annex A1, Annex A2, Annex A3, or Annex A4 as directed by
A5.2.6). the specification referencing this method and record data.
12.4.3 Check fuel flow rate against Standard Operating 13.1.1 When a specification allows multiple rating
Conditions by timing flow or counting the drip rate during first techniques, the method providing deposit measurements in SI
15 min of test. (See X1.5.) units is preferred.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 5


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
13.1.2 When the rating techniques do not agree, the method 15. Precision and Bias
providing measurements in SI units shall be regarded as the 15.1 An interlaboratory study of oxidative stability testing
referee. was conducted in accordance with Practice E691 by eleven
13.2 Return the heater tube to original container and retain laboratories, using thirteen instruments including two models
as appropriate. with five fuels at two temperatures for a total of ten materials.
Each laboratory obtained two results from each material.7
14. Report 15.1.1 The terms repeatability and reproducibility in this
14.1 Report the following information: section are used as specified in Practice E177.
14.1.1 The heater tube control temperature. This is the test 15.2 Precision—It is not possible to specify the precision of
temperature of the fuel. this test method because it has been determined that test
14.1.2 Heater tube deposit rating(s). method results cannot be analyzed by standard statistical
14.1.3 Maximum pressure drop across the filter during the methodology.
test or the time required to reach a pressure differential of 15.3 Bias—This test method has no bias because jet fuel
25 mm Hg. thermal oxidative stability is defined only in terms of this test
14.1.4 If the normal 150 min test time was not completed, method.
for example, if the test is terminated because of pressure drop
failure, also report the test time that corresponds to this heater 16. Keywords
tube deposit rating.
16.1 differential pressure; fuel decomposition; oxidative
NOTE 7—Either the heater tube rating or the ΔP criteria, or both, are deposits; heater tube deposits; thermal stability; turbine fuel
used to determine whether a fuel sample passes or fails the test at a
specified test temperature.
7
Supporting data have been filed at ASTM International Headquarters and may
14.1.5 Spent fuel at the end of a normal test (see 12.7.1.3). be obtained by requesting Research Report RR:D02-1309. Contact ASTM Customer
14.1.6 Heater tube serial number may be reported. Service at service@astm.org.

ANNEXES

(Mandatory Information)

A1. TEST METHOD FOR VISUAL RATING OF HEATER TUBES

A1.1 Scope A1.3.1.1 Discussion—This refers to deposit colors such as


A1.1.1 This method covers a procedure for visually rating blues and grays that do not match the Color Standard.
the heater tube produced by Test Method D3241. A1.3.2 peacock—A multicolor, rainbow-like heater tube
A1.1.2 The final result from this test method is a heater tube deposit.
color rating based on an arbitrary scale established for this test A1.3.2.1 Discussion—This type of deposit is caused by
method plus two additional yes/no criteria that indicate the interference phenomena where deposit thickness exceeds the
presence of an apparent large excess of deposit or an unusual quarter wave length of visible light.
deposit, or both. A1.3.3 Heater Tube Rating—A ten-step discrete scale from
A1.1.3 This international standard was developed in accor- 0 to >4 with intermediate levels for each number starting with
dance with internationally recognized principles on standard- 1 described as less than the subsequent number.
ization established in the Decision on Principles for the A1.3.3.1 Discussion—The scale is taken from the five
Development of International Standards, Guides and Recom- colors—0, 1, 2, 3, 4—on the ASTM Color Standard. The
mendations issued by the World Trade Organization Technical complete scale is: 0, <1, 1, <2, 2, <3, 3,< 4, 4, >4. Each step is
Barriers to Trade (TBT) Committee. not necessarily of the same absolute magnitude. The higher the
number, the darker the deposit rating.
A1.2 Referenced Documents
A1.2.1 Adjunct:5 A1.4 Summary of Test Method
Color Standard for Heater Tube Deposit Rating A1.4.1 This test method uses a specially constructed light
box to view the heater tube. The heater tube is positioned in the
A1.3 Terminology box using a special heater tube holder. Uniformity of the new
A1.3.1 abnormal—a heater tube deposit color that is neither heater tube surface is judged under the optimum light condi-
peacock nor like those of the Color Standard. tions of the box. Color of the heater tube is judged under light

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 6


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
and magnification by comparing to the Color Standard plate Then examine the center (thinner area) of the heater tube
slid into optimum position immediately behind the heater tube. between 5 mm and 55 mm above the bottom shoulder using the
Tuberator. If a defect is seen, establish its size. If it is larger
A1.5 Significance and Use than 2.5 mm2, discard the heater tube. Fig. A1.1 provides an
A1.5.1 The final heater tube rating is assumed to be an illustration of defect areas equivalent to 2.5 mm2.
estimate of condition of the degraded fuel deposit on the heater A1.10.2 Examine the heater tube for straightness by rolling
tube. This rating is one basis for judging the thermal oxidative the heater tube on a flat surface and noting the gap between the
stability of the fuel sample. flat surface and the center section. Reject any bent heater tube.
A1.6 Apparatus
A1.11 Procedure
A1.6.1 Heater Tube Deposit Rating Apparatus—The colors
of deposits on the heater tube are rated by using a tuberator and A1.11.1 Set Up:
the ASTM Color Standard. A1.11.1.1 Snap the upper end of the heater tube into the
clamp of the holder for the heater tube.
A1.7 Test Samples (Heater Tube) A1.11.1.2 Push the heater tube against the stop of the holder
A1.7.1 Handle the heater tube carefully so as not to touch for the heater tube.
the center portion at any time. A1.11.1.3 Slide the holder with the heater tube over the
guide rod into the tuberator.
NOTE A1.1—Touching the center of the heater tube will likely contami-
nate or disturb the surface of the heater tube, deposit, or both, which must A1.11.1.4 Rotate the holder and position the heater tube
be evaluated in pristine condition. such that the side with the darkest deposit is visible.
A1.11.1.5 Insert the ASTM Color Standard into the tubera-
A1.8 Standard Operating Conditions tor.
A1.8.1 Inside of Light Box, opaque black. A1.11.2 Evaluation:
A1.8.2 Light Source, three 30 W incandescent bulbs, clear, A1.11.2.1 On completion of the test, compare the darkest
reflective type; all shall be working for optimum viewing. heater tube deposit color, between 5 mm and 55 mm above the
A1.8.3 Bulb Positions, one above, two below, each directed bottom shoulder, with the ASTM Color Standard. Only rate a
toward heater tube holder and color standard. deposit if the area is greater than 2.5 mm2 and the width of any
axial (that is, longitudinal) streak or spot is greater than
A1.8.4 Magnification, 2×, covering viewing window. 0.8 mm. Fig. A1.1 provides an illustration of spots or axial
A1.8.5 Evaluators—Use persons who can judge colors, that streaks with an area equivalent to 2.5 mm2.
is, they should not be color blind. A1.11.2.2 When the darkest heater tube deposit color cor-
responds to a color standard, that number should be recorded.
A1.9 Calibration and Standardization
A1.9.1 No standardization is required for this test apparatus,
but since the Color Standard is known to fade, store it in a dark
place.
NOTE A1.2—The lifetime of the Color Standard is not established when
continuously or intermittently exposed to light. It is good practice to keep
a separate Standard in dark (no light) storage for periodic comparison with
the Standard in regular use. When comparing, the optimum under the light
conditions are those of the heater tube rating box.
A1.9.2 Standardization of Rating Technique:
A1.9.2.1 In rating a heater tube, the darkest deposits are
most important. Estimate grades for the darkest uniform
deposit, not for the overall average color of the deposit area.
A1.9.2.2 When grading, consider only the darkest continu-
ous color that covers an area equal or larger than a circle of size
one-half the diameter of the heater tube.
A1.9.2.3 Ignore an axial (that is, longitudinal) deposit streak
that is less in width than one-quarter the diameter of the heater
tube regardless of the length of the streak.
A1.9.2.4 Ignore spots, axial (that is, longitudinal) streaks, or
scratches on a heater tube that are considered heater tube
defects. These will normally not be present, since the heater
tube is examined before use to eliminate defective heater tubes.
A1.10 Pretest Rating of Heater Tubes
A1.10.1 Examine the heater tube without magnification in
laboratory light. If a defect is visible, discard the heater tube. FIG. A1.1 Defect Areas

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 7


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
A1.11.2.3 If the darkest heater tube deposit color being A1.12.2 Examples:
rated is in the obvious transition state between any two A1.12.2.1 Example 1—A heater tube has a maximum de-
adjacent color standards, the rating should be recorded as less posit falling between Color Standard Codes 2 and 3 with no
than the darker (that is, higher number) standard. other colors present. The overall heater tube rating would be
A1.11.2.4 In the event the heater tube has deposits which do less than 3.
not match the normal Color Standard colors, use the following A1.12.2.2 Example 2—The darkest deposit on a heater tube
rules for rating. With reference to standard terms: matches a Code 3, but there is also a peacock deposit present.
(1) If the heater tube deposit is peacock color, rate this as The overall rating of the heater tube would be reported as 3P.
Code P, but also rate any heater tube deposit that shows normal A1.12.2.3 Example 3—A heater tube has a deposit that
deposit color; or matches Color Standard Code 1 and also has an abnormal
(2) If the heater tube deposit contains an abnormal color, deposit. The overall heater tube rating would be reported as
rate this as Code A, but also rate any heater tube deposit that 1A.
shows normal deposit color.
A1.13 Precision and Bias
A1.11.3 Remove the rated heater tube and return to its A1.13.1 Precision—The precision of the procedure in Test
original container. Method D3241 for measuring heater tube deposit rating by this
A1.12 Report method was evaluated by the subcommittee and is reported in
RR:D02-1786.8
A1.12.1 Report the numerical rating for the heater tube
A1.13.2 Bias—The procedure in Test Method D3241 for
deposit plus A or P, or both, with additional description, if
determining heater tube deposit rating has no bias because the
applicable.
value of heater tube deposit rating is defined only in terms of
A1.12.1.1 When reporting the overall rating, report the the test method.
maximum rating, and, if there are colors present that do not
match the Color Standard, report these also. 8
Supporting data have been filed at ASTM International Headquarters and may
A1.12.1.2 If there are only P or A, or both, deposits, report be obtained by requesting Research Report RR:D02-1786. Contact ASTM Customer
only these and do not attempt to estimate a numerical grade. Service at service@astm.org.

A2. TEST METHOD FOR THICKNESS DEPOSIT RATING OF HEATER TUBES—INTERFEROMETRIC METHOD

A2.1 Scope the same technique to establish a base line or condition of satisfactory
starting quality.
A2.1.1 This annex describes a procedure for the interfero-
metric thickness deposit rating in the range of 0 nm to 1200 nm A2.2 Terminology
of heater tubes produced by Test Method D3241—Thermal
Oxidation Stability of Aviation Turbine Fuels. A2.2.1 Definitions of Terms Specific to This Standard:
A2.1.2 The final result from this rating procedure is an A2.2.1.1 deposit—film of oxidized product deposited on the
absolute measurement of the thickness and volume of deposit test area of the heater tube after D3241 test procedure.
on the heater tube that provides a basis for judging the thermal A2.2.1.2 deposit profile—three-dimensional representation
oxidative stability of the fuel sample. For aircraft fuel systems of deposit thickness profile along and around the length of the
performance, deposit thickness and volume are useful param- heater tube test section.
eters.
A2.2.1.3 deposit thickness—the thickness of deposit present
A2.1.3 An interlaboratory study was conducted in October
on the heater tube substrate surface expressed in nanometers,
2011 (see ASTM Research Report RR:D02-17869 for support-
nm.
ing data) involving 8 interferometric instruments and 117
heater tubes tested in duplicate. The interferometric procedure A2.2.1.4 deposit volume—the volume of deposit present on
demonstrated objective rating. the test section of the heater tube expressed in mm3.
NOTE A2.1—The particular technique used for this test method is called A2.2.1.4.1 Discussion—The deposit volume is derived by
spectral reflectance. integration of the area under the deposit profile.
NOTE A2.2—If this procedure is to be used to rate the heater tube after
the thermal oxidation test, the new heater tube may also be examined by A2.2.1.5 interferometry—a technique used for measuring
the optical properties of surfaces (refractive index and absorp-
tion coefficient) based on studying the pattern of interference
9
created by their superposition. In the presence of a thin
Supporting data have been filed at ASTM International Headquarters and may
be obtained by requesting Research Report RR:D02-1786. Contact ASTM Customer transparent layer called film, interferometry can also be used to
Service at service@astm.org. provide film thickness information.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 8


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
A2.2.1.6 standard spot—the mean thickness of the six heater tube rotating system, optical probe displacement system
thickest points in a 2.5 mm2 area, as shown in Fig. A2.6, and computer-driven software for analyzing the interferometric
defined in section A1.11.2.1 of this test method. data.
A2.6.1.2 The instrument must be capable to precisely and
A2.3 Summary of Test Method automatically displace the optical probe with the resolution
A2.3.1 An interferometric apparatus, as shown in Fig. A2.1, defined in section A2.8.2.3.
is used to rate the deposit on the heater tube. The computer- A2.6.1.3 The instrument must be able to automatically
driven software analyzes the interferometric data. The deposit rotate the heater tube with the resolution defined in section
thickness and deposit volume are derived and displayed. A2.8.2.4.
A2.6.1.4 The instrument, with its optical probe, must be
A2.4 Significance and Use able to automatically detect the edge of one of the two
A2.4.1 The final heater tube rating is a direct thickness and shoulders of the heater tube; the distance between these two
volume measurement of the degraded fuel deposited on the shoulders is 60 mm.
heater tube. This rating is one basis for judging the thermal A2.6.1.5 The instrument can measure the thickness over the
oxidative stability of the fuel sample. whole length of the heater tube. However, this test method
describes a procedure to measure the deposit thickness between
A2.5 Reagents and Materials the 5 mm and the 55 mm points located between the two
A2.5.1 Reference Heater Tube10—with two reference de- shoulders of the heater tube, as defined in Test Method D3241
posits of known and traceable thickness made with silicon (Fig. A2.7).
dioxide on silicon (Si + SiO2). See Fig. A2.2. A2.6.1.6 For the calculation of the film deposit thickness,
the computer driven software must be able to automatically
A2.6 Apparatus select against wavelength the appropriate refractive index
A2.6.1 Deposit Rater:11 value (n) and absorption coefficient value (k) for the substrate
A2.6.1.1 Comprising of a suitable UVVIS light source and the deposit film. These values are indicated in the graphs
(200 nm to 1100 nm), reflected light probe capable to generate in sections A2.8.2.1 and A2.8.2.2.
a spot light of 200 µm diameter, detector of reflected light for
A2.7 Test Samples (Heater Tube )
measuring light interferences, heater tube handling assembly,
A2.7.1 Handle the heater tube carefully so as not to touch
the center portion at any time.
NOTE A2.3—Touching the center (thinner area) of the heater tube will
10
likely contaminate or disturb the surface of the heater tube, deposit, or
The sole source of supply of the reference heater tube known to the committee
both, which must be evaluated in pristine condition.
at this time is AD systems (www.adsystems-sa.com), available from AD systems,
P.A. Portes de la Suisse Normande, Allée de Cindais, 14320 Saint André sur Orne,
France. If you are aware of alternative suppliers, please provide this information to
A2.8 Apparatus Preparation
ASTM International Headquarters. Your comments will receive careful consider- A2.8.1 Install the apparatus in accordance with the manu-
ation at a meeting of the responsible technical committee,1 which you may attend
11
The sole source of supply of the deposit rater apparatus known to the
facturer’s instructions. If any malfunction is indicated refer to
committee at this time is AD systems (www.adsystems-sa.com), model DR 10 – the manufacturer’s instructions.
Deposit Rater, available from AD systems, P.A. Portes de la Suisse Normande, Allée NOTE A2.4—Malfunctions are checked automatically when switching
de Cindais, 14320 Saint André sur Orne, France. If you are aware of alternative on the instrument and during the operation. Any malfunction will be
suppliers, please provide this information to ASTM International Headquarters. automatically indicated.
Your comments will receive careful consideration at a meeting of the responsible
technical committee,1 which you may attend.

FIG. A2.1 Interferometric Apparatus—General Principle

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 9


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c

FIG. A2.2 Reference Heater Tube

FIG. A2.3 n and k Values of the Heater Tube

A2.8.2 Standard Operating Conditions: be used for instrument verification. Behenate films are anisotropic and do
A2.8.2.1 Heater Tube (Substrate)—refractive index (n) and not relate to deposits. They are sub-layered with metals separated by fatty
chains which creates multiple reflection for each sub layer. In other words,
absorption coefficient (k) values are automatically selected by the Lanmuir-Blodgett films of cadmium behenate do not create any
the computer software using the relationship shown in the interference with spectral reflectance technique when the emitted light is
graph below (Fig. A2.3). normal to the deposit surface which is the case with the DR 10.
NOTE A2.5—The value of n and k index are dependent on the A2.8.2.3 Circumferential Resolution—(number of points
metallurgy and surface roughness of the heater tube under examination. measured on the heater tube circumference), 24 points (Fig.
The values on the graphs here above are appropriate for heater tubes used A2.5).
in D3241.
A2.8.2.4 Longitudinal Resolution—(number of points mea-
A2.8.2.2 Deposit Film—refractive index (n) and absorption sured on a length of 50 mm), 50 points (Fig. A2.6).
coefficient (k) values are automatically calculated by the A2.8.2.5 Heater Tube Scanned Surface—the instrument
computer software using the relationship shown in the graph shall be set up to measure the thickness between the 5 mm and
below (Fig. A2.4). the 55 mm points. See Fig. A2.7.
NOTE A2.6—This represents the average value of refractive index of
heater tube deposit after D3241 test procedure.
A2.9 Baseline Calibration
NOTE A2.7—It is assumed that the film is optically transparent to the A2.9.1 Heater tube surface finishes vary slightly between
wavelength of UVVIS light used in the DR 10, currently 200 nm to brands due to variations in polishing techniques. Therefore, a
1100 nm.
NOTE A2.8—This method is not applicable to heater tubes coated with baseline calibration shall be performed using a new heater
Langmuir-Blodgett films of cadmium behenate. Consequently, heater tube. The heater tube shall have been visually inspected
tubes coated with Langmuir-Blodgett films of cadmium behenate cannot according to section A2.11.2. Baseline calibration shall be

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 10


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c

FIG. A2.4 n and K Values of the Heater Tube

FIG. A2.5 Circumferential Resolution

FIG. A2.6 Standard Spot

performed prior to initial use of the instrument and whenever A2.9.3 The typical reflected spectrum is calculated by
there is a change in heater tube brand. averaging the 12 points and it is saved as the baseline. This
A2.9.2 During this phase, the instrument records and baseline is then used to determine the interferences wave
memorizes the reflected light spectrum from 12 different points length generated by the deposit.
on the new heater tube (Fig. A2.8).

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 11


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c

FIG. A2.7 Description of Heater Tube Scanned Surface

FIG. A2.8 Baseline Calibration Points

A2.9.4 To perform this calibration, refer to the operation magnifies the serial number and takes a picture of the heater tube serial
manual delivered with the instrument, the duration of this number. Second, it positions the heater tube on the same point of the
circumference to ensure that the sequence of thickness measurements is
phase is approximately three minutes. started from the same point on the circumference. Consequently, for
verification, when the test is repeated on the same heater tube, it facilitates
A2.10 Apparatus Verification the comparison of the deposit profiles.
A2.10.1 On a regular basis, a minimum of once per week, A2.12.3 At the end of the measurement, the reported values
verify the operation of the apparatus by measuring a reference are displayed and recorded, as shown in Fig. A2.9.
heater tube (A2.5.1). To do so, refer to the operation manual
A2.12.4 Remove the rated heater tube and return it to its
delivered with the instrument. The results shall be within 65 %
original container.
of the known thickness. If the determined values are not within
the acceptable limits, return the apparatus to the manufacturer A2.13 Reporting
for recalibration.
A2.13.1 Report the following:
A2.11 Heater Tube Inspection A2.13.1.1 Standard spot deposit thickness to the nearest
1 nm.
A2.11.1 Handle the heater tube carefully and do not touch A2.13.1.2 Deposit volume to the nearest 0.00001 mm3 or
the center section at any time. “N/A” when the volume is not reported.
A2.11.2 Visually examine the heater tube for straightness by NOTE A2.10—In some cases, the apparatus is unable to precisely
rolling the heater tube on a flat dust-free surface and noting the determine the thickness at one or several of the 1200 measured points.
gap between the flat surface and the center section. Heater When this situation occurs, because safety is involved with D3241 test, the
tubes with a bent center section are not suitable for rating by apparatus does not extrapolate the thickness from the surrounding points.
the DR 10. It simply reports “N/A” (Non Applicable) for the point in question. If the
number of reported “N/A” points exceeds 10, then the Deposit Volume of
A2.11.3 Examine the heater tube, without magnification, in is reported as “N/A”. However, the apparatus still reports the standard spot
laboratory light for visible mechanical scratches on the centre thickness which is based on the average of the thickest points in a 2.5 mm2
section of the heater tube. If such defects are visible, discard area. This reported standard spot is calculated with all the points for which
the thickness was determined.
the heater tube. In some cases, when the apparatus reports a “N/A Deposit Volume”, the
heater tube in question is visually rated (VTR) as an abnormal deposit. No
A2.12 Procedure correlation can be established between “N/A Deposit Volume” reporting
A2.12.1 To start a test, refer to the operation manual and abnormal visual rating. It was observed with the results of the ILS
conducted in 2011 (Research Report RR:D02-1786) that when a “N/A
delivered with the instrument. Deposit Volume” was reported, it was associated with thick deposits and
A2.12.2 Once the jet fuel sample details in addition with the it is expected to be typical.
heater tube details are entered, the thickness measurement is
A2.14 Test Report
automatically initiated.
NOTE A2.9—The rotation and the positioning of the heater tube has two A2.14.1 The test report shall contain at least the following
functions. The first one for traceability of the test, the instrument information:

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 12


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c

FIG. A2.9 Reporting Example

A2.14.1.1 Sufficient details for complete identification of correct operation of Annex A2, exceed the values in Table A2.1
the fuel and the heater tube tested; only in one case in twenty, where x is the average of the two
A2.14.1.2 A reference to this standard; deposit measurements.
A2.14.1.3 The result of the test (see A2.12); A2.15.1.2 Deposit Volume:
A2.14.1.4 Any deviation, by agreement or otherwise, from (1) Repeatability Limit (r)—The difference between suc-
the procedure specified; cessive deposit measurements, obtained by the same operator
A2.14.1.5 The date of the test. with the same apparatus under constant operating conditions on
the same heater tube, would in the long run, in the normal and
A2.15 Precision and Bias correct operation of Annex A2, exceed the values in Table A2.2
A2.15.1 Precision—The following criteria provides infor- only in one case in twenty, where x is the average of the two
mation on the accuracy of the deposit measurements using this deposit measurements.
technique:12 (2) Reproducibility Limit (R)—The difference between two
A2.15.1.1 Deposit Thickness: single and independent deposit measurements, obtained by
(1) Repeatability Limit (r)—The difference between suc- different operators working in different laboratories on the
cessive deposit measurements, obtained by the same operator same heater tube, would in the long run, in the normal and
with the same apparatus under constant operating conditions on correct operation of Annex A2, exceed the values in Table A2.2
the same heater tube, would in the long run, in the normal and only in one case in twenty, where x is the average of the two
correct operation of Annex A2, exceed the values in Table A2.1 deposit measurements.
only in one case in twenty, where x is the average of the two A2.15.2 Bias—Since there are no accepted reference mate-
deposit measurements. rials suitable for determining the bias for this procedure, bias of
(2) Reproducibility Limit (R)—The difference between two deposit thickness and deposit volume measurements between
single and independent deposit measurements, obtained by Annex A2, Annex A3, and Annex A4 of this test method have
different operators working in different laboratories on the not yet been determined.
same heater tube, would in the long run, in the normal and
A2.15.3 The precision statements were derived from a 2011
intralaboratory study of deposit thickness and volume measure-
ments.12 One laboratory using 8 instruments tested 117 heater
12
Supporting data have been filed at ASTM International Headquarters and may tubes in duplicate for deposit thickness and deposit volume.
be obtained by requesting Research Report RR:D02-1786. Contact ASTM Customer Information on the heater tubes and their average deposit
Service at service@astm.org.
measurements are in the research report.12
TABLE A2.1 Precision of Deposit Thickness Measurement by TABLE A2.2 Precision of Deposit Volume Measurement by
Interferometric (Heater) Tube Rating (ITR) (nm) Interferometric (Heater) Tube Rating (ITR) (mm3)
Measurement Range Repeatability (r) Reproducibility (R) Measurement Range Repeatability (r) Reproducibility (R)
7.2 to 1189.6 0.0009857 (x+2000) 0.001996 (x+2000) 0.0004 to 0.1934 0.002274 (x+0.2) 0.006447 (x+0.2)

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 13


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c

A3. RATING OF HEATER TUBES—ELLIPSOMETRIC METHOD

A3.1 Scope A3.3 Summary of Test Method


A3.1.1 This annex describes a procedure for the ellipsomet- A3.3.1 An ellipsometric apparatus13 is used to rate the
ric rating of heater tubes produced by Test Method D3241— deposits on the heater tube. The computer-driven software
Thermal Oxidation Stability of Aviation Turbine Fuels. analyses the ellipsometric data and the maximum deposit
A3.1.2 The final result from this rating procedure is an thickness of an average 2.5 mm2 is derived and displayed.
absolute measurement of the deposit thickness on the heater
tube that can provide a basis for judging the thermal oxidative A3.4 Significance and Use
stability of the fuel sample. For aircraft fuel systems A3.4.1 The final heater tube rating is a direct measure of the
performance, deposit thickness is a useful parameter. thickness of the degraded fuel deposited on the heater tube.
NOTE A3.1—If this procedure is to be used to rate the heater tube after This rating is a basis for judging the thermal oxidative stability
the D3241 test, the new heater tube may be examined in an Ellipsometeric
(Heater) Tube Rater (ETR) to establish a base line or condition of of the fuel sample.
satisfactory starting quality.
A3.5 Reagents and Materials
A3.2 Terminology
A3.5.1 Ellipsometer NIST Traceable Reference Tube, a
A3.2.1 Definitions of Terms Specific to This Standard: special tube designed for holding two wafer pieces with known
A3.2.1.1 deposit profile—three-dimensional representation NIST Traceable deposits of SiO2 on Si in the ranges of 60 nm
of the deposit thickness along and around the length of the and 100 nm. Additional areas on the tube to align wafer pieces
heater tube test area. to laser beam (refer to Fig. A3.1). Each tube includes a
A3.2.1.2 deposit thickness—the thickness of deposit present certification of thickness of the spots for the user to compare in
on the heater tube substrate surface expressed in nanometers the verification process.14
(nm).
A3.2.1.3 ellipsometry—a technique used for measuring the
optical properties of surfaces (refractive index and absorption
coefficient) based on changes in the polarized state of light
upon reflection from the surface.
A3.2.1.3.1 Discussion—In the presence of a thin transparent
13
layer, with a known refractive index and absorption coefficient, Supporting data have been filed at ASTM International Headquarters and may
be obtained by requesting Research Report RR:D02-1774. Contact ASTM Customer
ellipsometry can also be used to provide film thickness Service at service@astm.org.
information. 14
The sole source of supply of the apparatus (reference tube kits) known to the
committee at this time is Falex Corporation, 1020 Airpark Dr., Sugar Grove, Ill
A3.2.1.4 maximum deposit thickness—the maximum thick-
60554. If you are aware of alternative suppliers, please provide this information to
ness of an average 2.5 mm2 deposit present on the heater tube ASTM International Headquarters. Your comments will receive careful consider-
surface, expressed in nanometers, nm. ation at a meeting of the responsible technical committee,1 which you may attend.

FIG. A3.1 Ellipsometer NIST Traceable Reference Tube

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 14


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
A3.6 Apparatus point combination, displays the maximum average and shows
A3.6.1 Ellipsometric (Heater) Tube Rater (ETR) is com- 15 the location of the maximum on the deposit graph.
prised of a suitable laser light source, optical detector for A3.8 Apparatus Verification
measuring the ellipsometric parameters, heater tube handling
assembly and computer-driven software for analyzing the A3.8.1 Verify the operation of the ellipsometric apparatus
ellipsometric data. It is designed to combine the ellipsometric by measuring Ellipsometer reference tubes (A3.5.1). The
principles with specific heater tube handling. results shall be within the 65 % of the known thickness and
61 mm of the known position. If the determined values are not
A3.6.2 The thickness measurement is performed over the within the acceptable limits contact the manufacturer for
length of the heater tube, from 5 mm to 55 mm of the total re-calibration service.
60 mm length, as designed by Test Method D3241.
A3.9 Heater Tube Inspection
A3.7 Apparatus Preparation
A3.9.1 Handle the heater tube carefully and do not to touch
A3.7.1 Install the apparatus in accordance with the manu- the center portion at any time.
facturer’s instructions. Ensure satisfactory function of the
ellipsometric device in terms of alignment and optical diag- A3.9.2 Visually examine the heater tube for straightness by
nostics. If any malfunction is indicated refer to the manufac- rolling the heater tube on a flat surface and noting the gap
turer’s instructions. between the flat surface and the center section. Heater tubes
with a bent center section are not suitable for rating by the
A3.7.2 Standard operating conditions. ETR.
A3.7.2.1 Substrate refractive index setting 2.65.
NOTE A3.2—The value of refractive index selected is dependent on the
A3.10 Procedure
metallurgy and surface roughness of the D3241 heater tube under A3.10.1 Run the instrument in accordance with the manu-
examination. The value 2.65 is appropriate for heater tubes used in D3241. facturer’s operating instructions.
See the ETR operating instructions for detailed requirements for other
metallurgies. A3.10.2 At the end of the measurement procedure, record
A3.7.2.2 Substrate absorption coefficient setting (k) – 9.40. the maximum average 2.5 mm2 deposit thickness in
nanometers, nm. When using the test resolution of 24 × 50
NOTE A3.3—The value of absorption coefficient is dependent on the points, the 2.5 mm2 area is defined 6 points in a row or any set
metallurgy of the D3241 heater tube under examination. The value of
minus 9.40 is appropriate for heater tubes used in D3241. See ETR of 3 consecutive pairs of points. The manufacturer includes an
operating instructions for detailed requirements for other metallurgies. algorithm that automatically identifies all 6 point
combinations, calculates the average thickness of each of these
A3.7.2.3 Deposit film refractive index setting 1.45.
combinations, selects the maximum average 6 point
NOTE A3.4—This represents the average value of refractive index based combination, displays the maximum average and shows the
on many organic films. location of the maximum on the deposit graph.
A3.7.2.4 Deposit film absorption coefficient 0.00.
A3.11 Reporting
NOTE A3.5—It is assumed that the film is optically transparent to the
wavelength of the laser light used in the ETR, currently 1550 nm. A3.11.1 Report the:
A3.11.1.1 The heater tube reference number;
A3.7.2.5 Resolution (number of circumferential points), 24
A3.11.1.2 The maximum average 2.5 mm2 deposit thick-
points.
ness in nm;
A3.7.2.6 Resolution (number of longitudinal points), 50
A3.11.1.3 The deposit volume in mm3;
points.
A3.11.1.4 Sufficient details for complete identification of
NOTE A3.6—The ETR has the capability to measure up to a maximum the fuel and the heater tube tested;
of 100 points along the length of the heater tube and up to 360 equally A3.11.1.5 A reference to this standard;
spaced points around the circumference of the heater tube.
A3.11.1.6 Any deviation, by agreement or otherwise, from
A3.7.2.7 The ETR records the maximum average 2.5 mm2 the procedure specified; and
deposit thickness in nanometers, nm. When using the test A3.11.1.7 The date of the test.
resolution of 24 × 50 points, the 2.5 mm2 area is defined 6
points in a row or any set of 3 consecutive pairs of points. The A3.12 Precision and Bias
manufacturer includes an algorithm that automatically identi- A3.12.1 Precision—The following criteria provides infor-
fies all 6 point combinations, calculates the average thickness mation on the accuracy of the deposit measurements using this
of each of these combinations, selects the maximum average 6 technique:16
A3.12.1.1 Deposit Thickness:
15
The sole source of supply of the ETR ellipsometric rater apparatus known to (1) Repeatability Limit (r)—The difference between suc-
the committee at this time is Falex Corporation, Model Falex 430 Ellipsometer cessive deposit measurements, obtained by the same operator
available from Falex Corporation, 1020 Airpark Dr., Sugar Grove, Ill 60554. A
Falex 430 Mk I/Mk II correlation report is available as RR:D02-1912. If you are
16
aware of alternative suppliers, please provide this information to ASTM Interna- Supporting data have been filed at ASTM International Headquarters and may
tional Headquarters. Your comments will receive careful consideration at a meeting be obtained by requesting Research Report RR:D02-1774. Contact ASTM Customer
of the responsible technical committee,1 which you may attend. Service at service@astm.org.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 15


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
with the same apparatus under constant operating conditions on A3.12.1.2 Deposit Volume:
the same heater tube, would in the long run, in the normal and (1) Repeatability Limit (r)—The repeatability of heater
correct operation of Annex A3, exceed the values in Table A3.1 tube deposit volume measurements by Ellipsometric (Heater)
only in one case in twenty, where x is the average of the two Tube Rating has not yet been determined.
deposit measurements. (2) Reproducibility Limit (R)—The reproducibility of
(2) Reproducibility Limit (R)—The difference between two heater tube deposit volume measurements by Ellipsometric
single and independent deposit measurements, obtained by (Heater) Tube Rating has not yet been determined.
different operators working in different laboratories on the A3.12.2 Bias—Since there are no accepted reference mate-
same heater tube, would in the long run, in the normal and rials suitable for determining the bias for this procedure, bias of
correct operation of Annex A3, exceed the values in Table A3.1 deposit thickness and deposit volume measurements between
only in one case in twenty, where x is the average of the two Annex A2, Annex A3, and Annex A4 of this test method have
deposit measurements. not yet been determined.
A3.12.3 The precision statements were derived from a 2013
TABLE A3.1 Precision of Deposit Thickness Measurement by intralaboratory study of deposit thickness measurements.16
Ellipsometric (Heater) Tube Rating (ETR) (nm) One laboratory using 6 instruments tested 10 heater tubes in
Measurement Range Repeatability (r) Reproducibility (R) duplicate for deposit thickness. Information on the heater tubes
6 to 112 0.00981 (x+73.12) 0.0261 (x+73.12) and their average deposit measurements are in the research
report.16

A4. RATING OF HEATER TUBES—MULTI-WAVELENGTH ELLIPSOMETRIC METHOD

A4.1 Scope A4.2.1.5.1 Discussion—In the presence of film deposits,


A4.1.1 This annex describes a procedure for the multi- with presumed refractive index and absorption coefficient,
wavelength ellipsometric rating of the heater tubes in Test ellipsometry can also be used to provide film thickness
Method D3241—Thermal Oxidation Stability of Aviation Tur- information.
bine Fuels. A4.2.1.6 maximum deposit thickness, n—the maximum
A4.1.2 This procedure covers fuel deposit thickness mea- thickness of an average 2.5 mm2 deposit present on the heater
surement in the range of 2 nm to 252 nm, and the precision has tube surface, expressed in nanometers, nm.
been estimated for fuel deposit thickness in the range of 2.8 nm A4.2.1.7 standard spot, n—the mean thickness of the six
to 234 nm. thickest points in a 2.5 mm2 area, as defined in section
A4.1.3 The multi-wavelength ellipsometer is capable of A1.11.2.1 of this test method.
deposit thickness measurement above 252 nm, however, pre-
cision has not been estimated. A4.2.2 Abbreviations:
A4.2.2.1 MWETR, n—multi-wavelength ellipsometric
A4.2 Terminology (heater) tube rater
A4.2.1 Definitions of Terms Specific to This Standard: A4.2.2.2 VTR, n—visual heater tube rater
A4.2.1.1 deposit, n—film of oxidized product deposited on
the test area of the heater tube after D3241 test procedure. A4.3 Summary of Test Method
A4.2.1.2 deposit profile, n—three-dimensional representa- A4.3.1 A multi-wavelength ellipsometric apparatus is used
tion in terms of deposit thickness along and around the length to rate the deposits on the heater tube. The computer-driven
of the heater tube test section. software analyzes the ellipsometric data and the maximum
A4.2.1.3 deposit thickness, n—the thickness of deposit pres- deposit thickness of an average 2.5 mm2 is derived and
ent on the heater tube substrate surface expressed in nanome- reported, known as a standard spot. The deposit volume is also
ters (nm). calculated and reported.
A4.2.1.4 deposit volume, n—the volume of deposit present A4.4 Significance and Use
on the test section of the heater tube expressed in mm3.
A4.2.1.4.1 Discussion—The deposit volume is derived by A4.4.1 The final heater tube rating is a measure of the
integration of the area under the deposit profile. thickness and volume of the degraded fuel deposited on the
heater tube. This rating is a basis for judging the thermal
A4.2.1.5 ellipsometry, n—a technique used for measuring oxidative stability of the fuel sample.
the optical properties of surfaces (refractive index and absorp-
tion coefficient) based on changes in the polarized state of light A4.4.2 For aircraft fuel systems performance, deposit thick-
upon reflection from the surface. ness and volume is a useful parameter.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 16


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
A4.5 Reagents and Materials value (n) and absorption coefficient value (k) for the substrate,
A4.5.1 Ellipsometer reference tubes, on which layers of oxide layer, and the deposit film. These values are indicated in
deposits of known thickness of 30 nm, 50 nm, 80 nm, 130 nm, section A4.7.
and 250 nm have been generated using atomic layer deposition A4.6.6 The rater shall be capable of precisely displacing the
of silicon dioxide upon aluminum. heater tube longitudinally and rotationally, in relation to the
NOTE A4.1—Reference tubes with deposits of known thickness are optical bench, with the resolution defined in A4.7.
available from the manufacturer of the multi-wavelength ellipsometer.
A4.6.7 The rater shall be capable of calculating and record-
A4.6 Apparatus ing the thickness in nanometers and the volume in cubic
A4.6.1 Multi-Wavelength Ellipsometric (Heater) Tube millimeters over the 24 by 50 points.
Rater (MWETR)17,18 as shown in Fig. A4.1 is comprised of A4.6.8 The rater shall be capable of calculating the mean
suitable LED light source(s), optical detector(s) for measuring thickness amongst all 24 points in each of the 50 rows. Subtract
the ellipsometric parameters, precise heater tube handling the minimum row mean value from each of the 1200 overall
assembly and computer-driven software for analyzing the points, and generate the final thickness report.
ellipsometric data. It is designed to combine the ellipsometric A4.6.9 The rater shall be capable of searching for any
principles with specific heater tube handling. cluster of six neighboring points with an average thickness of
A4.6.2 The thickness and volume measurement is per- greater than 85 nm or more and reporting status.
formed over the length of the heater tube, from 5 mm to 55 mm A4.6.10 The rater shall be capable of calculating and
of the total 60 mm length, as defined by Test Method D3241. reporting the total deposit volume, in cubic millimeters, of the
A4.6.3 The rater shall sequentially acquire ellipsometric thickness and surface area scanned.
data across four discrete wavelength bands centered at 466 nm,
525 nm, 596 nm, and 638 nm. A4.7 Apparatus Preparation
A4.6.4 The rater shall sequentially pass each LED wave- A4.7.1 Install the apparatus in accordance with the manu-
length on a point on the heater tube surface of approximately facturer’s instructions. Ensure satisfactory function of the
0.41 mm by 1.1 mm with the Brewster Angle set nominally at ellipsometric device in terms of alignment and optical diag-
69.8 degrees. nostics. If any malfunction is indicated refer to the manufac-
turer’s instructions.
A4.6.5 The rater analytical system shall be capable of
selecting against wavelength the appropriate refractive index A4.7.2 Standard operating conditions.
A4.7.2.1 Substrate refractive index setting (n)
n(466 nm) = 0.662
n(525 nm) = 0.856
n(596 nm) = 1.177
17
The sole source of supply of the multi-wavelength ellipsometric (heater) tube
n(638 nm) = 1.400
rater known to the committee at this time is PAC LP (www.paclp.com.com), model NOTE A4.2—The value of refractive indices selected were obtained
Opitireader™, available from PAC LP, 8824 Fallbrook, Houston, Texas, USA
77064. If you are aware of alternative suppliers, please provide this information to
ASTM International Headquarters. Your comments will receive careful consider-
ation at a meeting of the responsible technical committee,1 which you may attend.
18
Supporting data have been filed at ASTM International Headquarters and may
be obtained by requesting Research Report RR:D02-1872. Contact ASTM Customer
Service at service@astm.org.

FIG. A4.1 Multi-Wavelength Ellipsometer Apparatus—General Principle

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 17


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
from well-known reference literature.19 combination, displays the maximum average and total deposit
A4.7.2.2 Substrate absorption coefficient setting (k) volume, and shows the location of the maximum on the deposit
k(466 nm) = 5.659 graph.
k(525 nm) = 6.382
A4.8 Apparatus Verification
k(596 nm) = 7.21
k(638 nm) = 7.668 A4.8.1 Verify the operation of the multi-wavelength ellip-
sometric apparatus periodically by measuring ellipsometer
NOTE A4.3—The value of absorption coefficients selected were ob-
tained from well-known reference literature.19
reference tubes (A4.5.1). The results shall be within the 65 %
of the known thickness. If the determined values are not within
A4.7.2.3 Aluminum Oxide refractive index setting (n) the acceptable limits contact the manufacturer for re-
n(466 nm) = 1.778 calibration service.
n(525 nm) = 1.772 NOTE A4.9—A recommended interval for verification is at least once
n(596 nm) = 1.768 per week, or at any time that re-calibration or changes in condition are
n(638 nm) = 1.766 apparent.

NOTE A4.4—The value of aluminum oxide refractive indices selected A4.9 Heater Tube Inspection
were obtained from well-known reference literature.19
A4.9.1 Handle the heater tube carefully and do not touch the
A4.7.2.4 Aluminum Oxide absorption coefficient (k) center portion at any time.
k(466 nm) = 0
k(525 nm) = 0 A4.9.2 Visually examine the heater tube for straightness by
k(596 nm) = 0 rolling the heater tube on a flat surface and noting the gap
k(638 nm) = 0 between the flat surface and the center section. Heater tubes
with a bent center section are not suitable for rating by the
NOTE A4.5—The value of aluminum oxide absorption coefficients MWETR.
selected were obtained from well-known reference literature.19
A4.7.2.5 Deposit film refractive index setting (n) A4.10 Procedure
n(466 nm) = 1.478 A4.10.1 Operate the instrument in accordance with the
n(525 nm) = 1.471 manufacturer’s operating instructions.
n(596 nm) = 1.466
n(638 nm) = 1.464 A4.10.2 Insert the heater tube into the measuring system
with the serial number on bottom and opposing the operator (6
NOTE A4.6—The average value of deposit film refractive indices o’clock position).
selected were derived from many representative organic films measured
on multiple multi-wavelength instruments which had the substrate con- A4.10.3 Enter the heater tube details in the user interface
stants and aluminum oxide constants set to the stated literature values. and initiate the deposit thickness and volume measurements.
A4.7.2.6 Deposit film absorption coefficient (k) A4.10.4 At the end of the measurement procedure, record
k(466 nm) = 0.107 the maximum average 2.5 mm2 deposit thickness in nanome-
k(525 nm) = 0.082 ters (nm) as provided by the apparatus. The manufacturer
k(596 nm) = 0.064 includes an algorithm that automatically identifies all 6 point
k(638 nm) = 0.057 combinations, calculates the average thickness of each of these
NOTE A4.7—The average value of deposit film absorption coefficients combinations, selects the maximum average 6 point
selected were derived from many representative organic films measured combination, displays the maximum average and shows the
on multiple multi-wavelength instruments which had the substrate con- location of the maximum on the deposit graph.
stants and aluminum oxide constants set to the stated literature values.
A4.10.5 Record the deposit volume in cubic millimeters
A4.7.2.7 Resolution (number of circumferential points), 24
(mm3) as provided by the apparatus.
points.
A4.7.2.8 Resolution (number of longitudinal points), 50 A4.11 Reporting
points.
A4.11.1 Report:
NOTE A4.8—The rater has the capability to measure up to a maximum A4.11.1.1 The heater tube reference number;
of 5000 points along the length of the heater tube and up to 384 equally
spaced points around the circumference of the heater tube. A4.11.1.2 The maximum average 2.5 mm2 deposit thick-
ness to the nearest 1 nm;
A4.7.2.9 The rater records the maximum average 2.5 mm2 A4.11.1.3 The volume deposit to the nearest 0.00001 mm3;
deposit thickness in nanometers (nm). The manufacturer in-
A4.11.1.4 Sufficient details for complete identification of
cludes an algorithm that automatically identifies all 6 point
the fuel and the heater tube tested;
combinations, calculates the average thickness of each of these
A4.11.1.5 Reference to this standard, and annex;
combinations, selects the maximum average 6 point
A4.11.1.6 Any deviation, by agreement or otherwise, from
the procedure specified;
A4.11.1.7 Date of the test.
19
Palik, Edward D., Handbook of Optical Constants of Solids, Handbook of
Optics, 3rd edition, Vol. 4, 2009, McGraw-Hill. NOTE A4.10—A typical report is shown in Fig. A4.2.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 18


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c

FIG. A4.2 Multi-Wavelength Ellipsometer Apparatus Typical Reports 3D and 2D Color Maps (VTR 3)

TABLE A4.1 Precision of Deposit Thickness Measurement by


Multi-Wavelength Ellipsometric (Heater) Tube Rating (MWETR)
A4.12 Precision and Bias Measurement Range Repeatability (r) Reproducibility (R)
2.8 to 234 1.0627 (x0.2) 1.5251 (x0.2)
A4.12.1 Precision—The following criteria provides infor-
mation on the accuracy of the deposit measurements using this
technique:20
A4.12.1.1 Deposit Thickness:
(1) Repeatability Limit (r)—The difference between suc- (2) Reproducibility Limit (R)—The difference between two
cessive deposit measurements, obtained by the same operator single and independent deposit measurements, obtained by
with the same apparatus under constant operating conditions on different operators working in different laboratories on the
the same heater tube, would in the long run, in the normal and same heater tube, would in the long run, in the normal and
correct operation of Annex A4, exceed the values in Table A4.1 correct operation of Annex A4, exceed the values in Table A4.1
only in one case in twenty, where x is the average of the two only in one case in twenty, where x is the average of the two
deposit measurements. deposit measurements.
A4.12.1.2 Deposit Volume:
(1) Repeatability Limit (r)—The difference between suc-
20
cessive test deposit measurements, obtained by the same
Supporting data have been filed at ASTM International Headquarters and may
be obtained by requesting Research Report RR:D02-1872. Contact ASTM Customer operator with the same apparatus under constant operating
Service at service@astm.org. conditions on the same heater tube, would in the long run, in

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 19


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
the normal and correct operation of Annex A4, exceed the correct operation of Annex A4, exceed the values in Table A4.2
values in Table A4.2 only in one case in twenty, where x is the only in one case in twenty, where x is the average of the two
average of the two deposit measurements. deposit measurements.
(2) Reproducibility Limit (R)—The difference between two A4.12.2 Bias—Since there are no accepted reference mate-
single and independent deposit measurements, obtained by rials suitable for determining the bias for this procedure, bias of
different operators working in different laboratories on the deposit thickness and deposit volume measurements between
same heater tube, would in the long run, in the normal and Annex A2, Annex A3, and Annex A4 of this test method have
not yet been determined.
A4.12.3 The precision statements were derived from a 2016
TABLE A4.2 Precision of Deposit Volume Measurement by Multi- intralaboratory study of deposit thickness and deposit volume
Wavelength Ellipsometric (Heater) Tube Rating (MWETR) (mm3) measurements.20 One laboratory using 6 instruments tested 45
Measurement Range Repeatability (r) Reproducibility (R) heater tubes in duplicate for deposit thickness and deposit
0.00024 to 0.04827 0.001386 (x0.5) 0.005254 (x0.5) volume. Information on the tubes and their average deposit
measurements are in the research report.20

A5. EQUIPMENT

A5.1 Test Instrument A5.2.2.1 The heart of the test system is the tube-in-shell
heat exchanger, or heater tube test section Fig. A5.1, which
A5.1.1 The instrument described in this annex is the Avia-
holds the heater tube and directs flow of fuel over it. It is
tion Fuel Thermal Oxidation Stability Tester that is used to test
important for the heater tube to be aligned correctly in the
the thermal oxidation stability of turbine fuel. There are five heater tube test section as shown in Fig. A5.2. This component
instrument models which will be described. All provide a is critical to consistent results and is a common component in
means to pump the sample once through the test system across all models.
the heater tube and through a test filter. There are means to
A5.2.2.2 There are some other points regarding the fuel
control and measure heater tube temperature, system pressure,
system that deserve mention:
and pressure drop across the filter, and methods of control and
(1) Fresh fuel is filtered immediately out of the container
measurement vary with each instrument model. Mechanism for
containing the aerated fuel through 0.45 µm membrane filter
pumping is positive displacement using a gear pump or piston paper before entering the heater tube test section;
pump. (2) The heater tube is sealed in the heater tube test section
A5.2 Test Details by elastomer O-rings (Fig. A5.3);
(3) The test filter is of stainless steel of 17 µm rated
A5.2.1 General Description—This instrument uses a fixed porosity. Bypass of the filter occurs automatically when differ-
volume of jet fuel that has been filtered, then aerated to provide ential pressure reaches 250 mm Hg;
a sample saturated with air. During the test, fuel is pumped at (4) All models use two containers, one for fresh fuel and
a steady rate across a heater tube which is maintained at a one for spent;
relatively high temperature, typically 260 °C, but higher under (5) Flow of fuel in all models can be monitored by visually
some specifications. The fuel, saturated with oxygen from the counting drops of flow. Also allow volumetric measure of flow
aeration, may degrade on the hot heater tube to form deposits with time, which is considered the most accurate flow measure.
as a visible film. Also, the degraded materials of the fuel may
A5.2.3 Heating/Temperature Control System—The heater
flow downstream and be caught by the test filter. Both the
tube is resistively heated by the conductance of high amperage,
increase in differential pressure across the test filter and the
low voltage current from a transformer through the heater tube.
final heater tube rating are used to determine the oxidative
The heater tube is clamped to relatively heavy, water cooled
stability of the fuel. current conducting bus bars, which increase in temperature
A5.2.2 Fuel System—Freshly filtered and aerated fuel is relatively little.
initially placed in a container, then circulated once through the A5.2.3.1 The temperature controller in all instrument mod-
apparatus to a spent sample container. Motive force for the els serves as indicator and controller. In automatic mode, the
sample is a positive displacement pump that will maintain flow controller provides a source of steady heat during the test,
at 3.0 mL ⁄min and overcome any tendency of initial filter varying the power as necessary to maintain the target (setpoint)
blockage from affecting the flow rate. Deviation of 10 % in temperature. Temperature range of operation is from 180 °C to
flow rate is permitted. If filter blockage becomes severe, the 350 °C.
bypass valve located before the test filter can be opened in A5.2.3.2 Critical to temperature control is the thermocouple
order to finish the test. Then, any deposit on the heater tube can and its position. The thermocouple itself must be calibrated to
be evaluated based on a complete test. ensure acceptable accuracy. The position of the tip must be

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 20


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c

FIG. A5.1 Test Section Dimensions

carefully placed so the temperature reading during automatic A5.2.4 Cooling System—In the normal operation of the
control is the maximum (the hottest spot) for the heater tube. A instrument, some cooling is necessary to remove heat going
simple mechanical positioning system allows easy and accurate into the bus bars by conduction from the hot heater tube.
placement of the thermocouple. Cooling water is circulated through each bus bar using an
A5.2.3.3 A diagram of the basic heating system is shown in internally circulated and radiator cooled liquid system. The
Fig. A5.4. only precautions with these systems is to monitor them to be

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 21


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c

FIG. A5.4 Heater Tube and Temperature Control Schematic

FIG. A5.2 Alignment of Heater Tube

interfere with natural deposit formation. Therefore, the system


must be operated under a total pressure of about 3.45 MPa
(500 psi). This pressure level is accomplished by a hydraulic
piston pump (Models 230 and 240) to produce the high
pressure needed.
A5.2.5.1 A pressure transducer is used to measure and allow
monitoring of the total system pressure. The hydraulically
pressurized systems have a relief valve through which the fluid
passes in a constant leak throughout the test. For the relief
valve control to operate uniformly for any fuel, a displacement
cell is used where the spent fuel enters the top displacing water
out the bottom and through the relief valve (for Models 230
and 240 only). Since the valve sees only water, the valve works
consistently.
A5.2.6 Differential Pressure Measure—All instruments use
an electronic ΔP transducer.
A5.2.6.1 Proper use of this differential measuring device
requires two special actions: bypass and air bleeding. The first
allows the fuel flow to bypass the filter whenever that action
becomes necessary. The second is used to remove air that at
times may become trapped in the cell chambers.
A5.2.6.2 The original manometer system, by nature, in-
cluded a bias due to the presence of fuel instead of the usual air
over the mercury. This changed the value of pressure expressed
in terms of column height of mercury such that a result about
6 % higher than true occurs. A transducer is not subject to this
error, so in order to have manometer and transducer models
read the correctly a 6 % bias is added to the transducer so it
gives the same value as would a manometer.
A5.2.6.3 When operated, the ΔP measuring device em-
ployed must be zeroed under actual flow conditions at the start
of the test. This is because a small pressure drop is created
FIG. A5.3 Assembly Drawing of Heater Tube Test Section
across the system when fuel is flowing. Zeroing the transducer
at the beginning of the test compensates for the flow.
sure they are working and to avoid use of coolants that contain A5.2.7 Differential Pressure Measurement
contaminants or salts that may eventually foul the system. Standardization—The ΔP measurement accuracy can be
A5.2.5 Pressurization—At the temperature of a normal test, checked by a technique of reading the pressure created by a
jet fuel would typically boil at the temperature of the heater column of known density fluid on each side of the ΔP cell. The
tube. This would prevent accurate temperature control and details for doing this are included as part of each operating

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 22


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
manual for the particular model instrument. This standardiza-
tion is really a verification that the ΔP cell is operating
correctly and is not meant to be a true calibration of the cell.
Calibration must be done by the manufacturers of the cell if
such action is suggested based on the results of the standard-
ization.
A5.2.8 Thermocouple Calibration—It is important for the
thermocouple to be accurate. To ensure this value, a method of
calibration against known melting points is used and is
described in detail in the appropriate User Manual for each
instrument. Models 230 and 240 use two metals, pure tin at
232 °C and pure lead at 327 °C, to define two points surround-
ing the normal range used with the instrument. Also, an FIG. A5.5 Freezing Characteristics of Tin
ice-water mixture is used to establish a 0 °C low reference
point. Other models only use lead and an ice-water mixture.
A5.2.8.1 The principle used with melting point of metals is
to centrally immerse the thermocouple tip in the melted metal, A5.2.9 Fuel Aeration System—All models have means to
then allow the metal to cool. As the metal goes through its aerate the sample prior to testing. Without the presence of
freeze point, the temperature reading will hesitate momentarily oxygen in the sample, a proper test is not achieved. Filtered,
indicating the known point for the metal. dry air is metered through the sample at about 1.5 L ⁄min rate
A5.2.8.2 The difference between the known metal freezing for 6 min. This 9 L of air ensures 97 % saturation of the
value and the displayed temperature becomes a correction for sample.
setting test temperatures. For example, using tin with a known A5.2.10 Elapsed Time Measurement—There are various
freeze temperature of 232 °C (see Fig. A5.5), if the temperature methods of timing the test depending on the model. The
noted at the time the metal froze was higher than 232°, then elapsed time indicator is normally the basis used, but in some
this would indicate the thermocouple was reading high by the models, the timing of the ΔP data collection is done with a
difference indicated and the applied correction would be to different timer. Since these two timers may not be exactly the
lower any test temperature by this same amount. Where two same, the last data point may be lost if the test stops before the
metals and ice water (low point) are used, the principle is the last timed data point. The user manuals for the various
same but the correction is calculated and applied automatically instrument models cover techniques for avoiding loss of data
by the internal computer. points.

A6. WARNING STATEMENTS

A6.1 Acetone A6.2.7 Do not get in eyes.


A6.1.1 Keep away from heat, sparks, and open flame. A6.2.8 Can produce toxic vapors on contact with flames,
hot glowing surfaces, or electric arcs.
A6.1.2 Keep container closed. Use with adequate ventila-
tion. A6.3 Iso-propanol (2-propanol)
A6.1.3 Avoid buildup of vapors and eliminate all sources of A6.3.1 Keep away from heat, sparks, and open flame.
ignition, especially nonexplosion-proof electrical apparatus A6.3.2 Keep container away from heat, sparks, and open
and heaters. flame.
A6.2 Toluene A6.3.3 Keep container closed.
A6.3.4 Use with adequate ventilation.
A6.2.1 Avoid prolonged or repeated breathing of vapor or
spray mist. A6.3.5 Avoid buildup of vapors and eliminate all sources of
ignition, especially nonexplosion-proof electrical apparatus
A6.2.2 Use only with adequate ventilation. and heaters.
A6.2.3 Eye irritation and dizziness are indications of over- A6.3.6 Avoid prolonged breathing of vapor or spray mist.
exposure. A6.3.7 Avoid prolonged or repeated skin contact.
A6.2.4 Do not take internally.
A6.4 n-Heptane
A6.2.5 Swallowing may cause injury, illness, or death. A6.4.1 Keep away from heat, sparks, and open flame.
A6.2.6 Avoid prolonged or repeated contact with skin. A6.4.2 Keep container closed.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 23


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
A6.4.3 Use with adequate ventilation. A6.5.13 Do not use dented or damaged cylinders.
A6.4.4 Avoid prolonged breathing of vapor or spray mist. A6.5.14 For technical use only.
A6.4.5 Avoid prolonged or repeated skin contact. A6.5.15 Do not use for inhalation purposes.

A6.5 Compressed Gases (Nitrogen) A6.6 Aviation Turbine Fuel (Jet B, See Specification
D1655)
A6.5.1 Keep cylinder valve closed when not in use.
A6.6.1 Keep away from heat, sparks, and open flames.
A6.5.2 Do not enter storage areas unless adequately venti-
lated. A6.6.2 Keep container closed.

A6.5.3 Always use a pressure regulator. A6.6.3 Use with adequate ventilation.
A6.6.4 Avoid breathing vapor or spray mist.
A6.5.4 Release regulator tension before opening cylinder.
A6.6.5 Avoid prolonged or repeated contact with skin.
A6.5.5 Do not transfer to cylinder other than one in which
gas is received. A6.7 Aviation Turbine Fuel (Jet A or A-1, See Specification
A6.5.6 Do not mix gases in cylinders. D1655)
A6.7.1 Keep away from heat, sparks, and open flame.
A6.5.7 Never drop cylinder.
A6.7.2 Keep container closed.
A6.5.8 Make sure cylinder is supported at all times.
A6.7.3 Use with adequate ventilation.
A6.5.9 Stand away from cylinder outlet when opening
cylinder valve. A6.7.4 Avoid buildup of vapors and eliminate all sources of
ignition, especially nonexplosion-proof electrical apparatus
A6.5.10 Keep cylinder out of sun and away from heat. and heaters.
A6.5.11 Keep cylinder from corrosive environment. A6.7.5 Avoid breathing vapor or spray mist.
A6.5.12 Do not use cylinder without label. A6.7.6 Avoid prolonged or repeated contact with skin.

APPENDIXES

(Nonmandatory Information)

X1. INSTALLATION, MAINTENANCE, SPECIAL CHECKS

X1.1 Laboratory Installation Requirements X1.2.5 Remove and install the AutoCal Calibrator in up-
right position and refill with new metal. The proper amount of
X1.1.1 The tester should be placed on a level laboratory
tin for one filling is about 1.5 g to 1.9 g, and for lead about
bench, allowing a 200 mm to 300 mm wide bench area in front
3.3 g to 4.7 g, or within 1 mm of the top of the cup.
of the tester. Ready access to the rear of the tester should be
provided for routine maintenance and service requirements. X1.3 Thermocouple Replacement and Position Adjust-
Ensure any vents in the instrument cabinet are not obstructed ment
during installation or use. Adequate ventilation should be
X1.3.1 The thermocouple used for measuring and control-
provided, and proper procedures for handling solvents and
ling the temperature of the heater tube may have to be replaced
hydrocarbons should be used.
at intervals due to damage or failure.
X1.2 AutoCal Calibrator Metal Replacement X1.3.2 Install a new thermocouple reversing the steps used
to remove old thermocouple. Replace and tighten screws as
X1.2.1 The tin and lead in the well of the AutoCal Calibra- required. The tip of the thermocouple must be flush with top of
tor must be replaced whenever the quantity is below minimum upper fixed bus when position indicator is set at the reference
or when contaminated. mark.
X1.2.2 To remove the metal, install the AutoCal Calibrator X1.3.3 Check for proper thermocouple indexing under ac-
inverted between the upper fixed bus and the lower floating tual test operating conditions.
bus.
X1.4 Heater Tube Temperature Profile (for Models 230
X1.2.3 Place a damp paper tissue, cloth rag, or aluminum and 240 Only)
foil under the well to catch the molten metal.
X1.4.1 If it is desired to measure the heater tube temperature
X1.2.4 Apply power to the AutoCal Calibrator as during profile, do so after the first hour of the test or before significant
normal calibration, and at same time gently tap the well until ΔP occurs. Follow the procedure in the user manual for the
all molten metal has dropped out. particular model.

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 24


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
X1.5 Flow Rate Check X1.5.3 The volumetric method is the referee.
X1.5.1 Drip Rate Method—Check the flow by observing the NOTE X1.1—Drip rate counts start with zero (drop 0, drop 1, drop 2…
drip from the drip assembly and measuring the time for 20 drop 20) to get an accurate measure of the time required for 20 drops.
drops. Models 230 and 240, the time should be 19 s 6 1 s for
20 water drops. Models 230 Mk III, F400, and 230 Mk IV, the X1.6 Maintenance Manual
time should be 9 s 6 1 s for 20 fuel drops.
X1.6.1 A maintenance manual is available that provides
X1.5.2 Volumetric Method—Check the flow by measuring additional maintenance information, such as the electrical
the time it takes to collect 15 mL of fuel in a graduated
schematic. Complete details for operating the instrument mod-
cylinder. That time should be between 4 min, 33 s and 5 min,
els are contained in the user manuals for each instrument.
33 s.

X2. DETERMINATION OF BREAKPOINT

X2.1 Definition X2.2 Determination of Breakpoint


X2.1.1 Breakpoint—In Test Method D3241, breakpoint is X2.2.1 The breakpoint can be derived by conducting a
the highest control temperature at which the fuel meets heater series of tests of different control temperatures to arrive at the
tube rating and ΔP specification requirements. temperature, x °C, at which the fuel meets both heater tube
X2.1.1.1 Discussion—This definition of breakpoint de- rating and ΔP specification requirements, and where a test at a
scribes the highest pass temperature for a fuel. Note that some control temperature of (x + 5) °C would result in a fail (that is,
published papers have used the term breakpoint to describe the failure to meet heater tube rating or ΔP requirements). The
lowest fail temperature, which is the (x + 5) °C temperature temperature x °C would then be reported as the Test Method
referred to below. D3241 breakpoint.

SUMMARY OF CHANGES

Subcommittee D02.J0 has identified the location of selected changes to this standard since the last issue
(D3241 – 20b) that may impact the use of this standard. (Approved Dec. 1, 2020.)

(1) Added subsections 12.8.4 and 12.8.4.1. (3) Revised Table 2 and added Footnotes E and F.
(2) Revised subsections 7.1, 7.2, 10.1.1, 13.1, and 13.2.

Subcommittee D02.J0 has identified the location of selected changes to this standard since the last issue
(D3241 – 20a) that may impact the use of this standard. (Approved Sept. 1, 2020.)

(1) Revised subsections 7.1, 7.2, 7.3, 9.1.2, 10.2.2, and12.8.3. (5) Updated language throughout to ensure uniformity (replac-
(2) Revised Table 2. ing “coupon” with “heater tube,” adding “heater” to tube where
(3) Deleted previous subsections 12.2.1 and 12.2.2. missing, etc.).
(4) Replaced “reservoir” with “container” throughout.

Subcommittee D02.J0 has identified the location of selected changes to this standard since the last issue
(D3241 – 20) that may impact the use of this standard. (Approved July 1, 2020.)

(1) Revised Table 2. (2) Added Fig. A5.1.

Subcommittee D02.J0 has identified the location of selected changes to this standard since the last issue
(D3241 – 19b) that may impact the use of this standard. (Approved March. 1, 2020.)

(1) Deleted sections 1.6, 11.1.5, 12.2.3, 12.3.2, 12.3.3, 12.7.1, (2) Revised sections 6.1, 6.2.2.1(2), 9.1.11, 11.1.1, 12.1.2,
12.7.1.1, 12.7.1.2, 12.7.1.3, 12.8.3.1, A5.2.2.3, A6.8, X1.1.2, 12.2.3, 12.4.2.2, 12.4.2.3, 12.7.1, 12.7.1.1, 12.7.1.2, 12.7.1.3,
X1.5.1, X1.5.2, X1.6, X1.7 12.7.1.4, 12.8.1, 12.8.3.1, 14.1.3, 14.1.5, A5.2.2.2(3) (4) and

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 25


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
D3241 − 20c
(5), A5.2.3.1, A5.2.4, A5.2.5, A5.2.5.1, A5.2.6, A5.2.6.1, (3) Revised Table 1 and Table 2.
A5.2.6.2, A5.2.6.3, A5.2.8, A5.2.8.1, X1.1.1, X1.2.1, X1.2.3, (4) Deleted Fig. A5.3.
X1.2.5, X1.3.1, X1.3.2, X1.4, X1.5.1, X1.5.3, X1.6.1.

Subcommittee D02.J0 has identified the location of selected changes to this standard since the last issue
(D3241 – 19a) that may impact the use of this standard. (Approved Nov. 1, 2019.)

(1) Revised title of Table 2.

Subcommittee D02.J0 has identified the location of selected changes to this standard since the last issue
(D3241 – 19) that may impact the use of this standard. (Approved Aug. 1, 2019.)

(1) Added Precision and Bias information to Annex A2, Annex (3) Revised Table 2.
A3, and Annex A4. (4) Added new footnote to subsection A3.6.1.
(2) Added new Section 8.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

&RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG 6XQ-DQ*07 26


'RZQORDGHGSULQWHGE\
8QLYHUVLGDG,QGXVWULDOGH6DQWDQGHUSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG

You might also like