Ticket PM 101236 W2-202110-0039

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

PT SABA INDOMEDIKA PM

National Diagnostic Distributors No: #101236


Date: 13/10/21

Customer Instrument

Name Meuraxa, PKM Type XP-100

Jl. Sultan Iskandar Muda, Gampong Blang Oi - Kec Meuraxa


Address Serial B4820
Banda Aceh

User Safrida / 081360818329 Contract Type Sold

Ticket

Ticket Type Request - PM On Site Open By Safrida

Number W2/202110/0039 Loging Call 10:42 - 13/10/21

Description maintenance Start 10:43 - 13/10/21

Division Engineer End 12:39 - 13/10/21

Support zamzami Status closed

Comment Sangat mudah dihubungi Ratting 5

Checklist PM Notes

A) Pneumatic Tube

☑ 1) Check and clean compressor, replace piston rings if necessary.


☑ 2) Check tubing's for dirt and damages, replace if necessary.
☑ 3) Check tubing nipples.
B) Main Unit Pneumatic System

☑ 1) Check and clean the pressure and vacuum lines, change if necessary.
☑ 2) Check bellows unit, needle valve, related tubing and connector, clean if necessary.
☑ 3) Clean pressure and vacuum nipples.
☑ 4) Check Pressure (0.5 ± 0.07kg/cm²), adjust if necessary.
☑ 5) Check Vacuum (250 ± 10mmHg ), adjust if necessary.
☑ 6) Clean Vacuum Trap Chamber and check for cracks.
☑ 7) Check and Clean Orifice.
C) Main Unit Hydraulic

☑ 1) Clean Waste Chamber


☑ 2) Clean Trap Chamber
Checklist PM Notes

☑ 3) Clean Rinse Cup assembly and mechanism.


☑ 4) Clean SRV assembly
☑ 5) Clean ASP Pipette.
☑ 6) Clean RBC/WBC Transducers and Chambers
☑ 7) Check on dirty and damaged tubing's.
☑ 8) Clean waste nipple (rear of analyzer)
D) Main Unit Adjustments

☑ 1) Clog adjustment for WBC and RBC Transducers


☑ 2) Adjustment of HGB Unit - Blank 2000 ± 100
☐ 3) Adjust Standard Sensitivity (W-MFV, R-MCV) -
E) General

☑ 1) Perform background check.


☑ 2) Measure "eightcheck" in "QC Mode" and Analyse normal human blood sample, adjust if necessary.
☑ 3) Reset counter for WC, detector and SRV if replaced.
☑ 4) Check Printer & Keyboard
F) Counter

☑ 1) Total Counter
☑ 2) SRV
Notes

Total Counter 2157

Engineer Approved By

Started By: zamzami Closed By: Safrida

Ended By: zamzami Closed At: 09:33 - 16/10/21

Note : This Service Report is electronically generated after closed in the system. No signature is required