Ticket PM 126378 W2-202207-0019
Customer Instrument
Ticket
Checklist PM Notes
A) Pneumatic Tube
☑ 1) Check and clean the pressure and vacuum lines, change if necessary.
☑ 2) Check bellows unit, needle valve, related tubing and connector, clean if necessary.
☑ 3) Clean pressure and vacuum nipples.
☑ 4) Check Pressure (0.5 ± 0.07kg/cm²), adjust if necessary.
☑ 5) Check Vacuum (250 ± 10mmHg ), adjust if necessary.
☑ 6) Clean Vacuum Trap Chamber and check for cracks.
☑ 7) Check and Clean Orifice.
C) Main Unit Hydraulic
☑ 1) Total Counter
☑ 2) SRV
Notes
Engineer Approved By
Note : This Service Report is electronically generated after closed in the system. No signature is required