Professional Documents
Culture Documents
D-10 Dual Program Instruction Manual English
D-10 Dual Program Instruction Manual English
220-0201
D-10™
Dual Program
Instruction Manual
D-10™
Dual Program Reorder Pack
400
for HbA1c, or
200
for HbA2/F/A1c
• HbA1c Calibrator Set • HbA1c Calibrator/Diluent Set • HbA1c Calibrator Level 1 • HbA1c Calibrator Level 2
• HbA1c-Kalibratorset • HbA1c-Kalibrator-/Diluentset • HbA1c-Kalibrator Level 1 • HbA1c-Kalibrator Level 2
• Gamme étalon HbA1c • Gamme étalon / diluant HbA1c • Étalon de niveau 1 HbA1c • Étalon de niveau 2 HbA1c
• Juego de calibradores para • Juego de calibradores/ • Calibrador Nivel 1 para HbA1c • Calibrador Nivel 2 para
HbA1c diluyente para HbA1c • Calibratore livello 1 dell’HbA1c HbA1c
• Set del calibratore dell’HbA1c • Set del calibratore/diluente • Calibrador Nível 1 de HbA1c • Calibratore livello 2
• Conjunto calibrador de HbA1c dell’HbA1c • HbA1c Kalibrator, nivå 1 dell’HbA1c
• HbA1c Kalibratorset • Conjunto calibrador/diluente • HbA1c kalibrator niveau 1 • Calibrador Nível 2 de HbA1c
• HbA1c kalibratorsæt de HbA1c • Βαθμονομητής HbA1c επιπέδου 1 • HbA1c Kalibrator, nivå 2
• Σετ βαθμονομητών HbA1c • HbA1c Kalibrator/spädningsset • HbA1c kalibrator niveau 2
• HbA1c kalibrator/ • Βαθμονομητής HbA1c
fortyndingssæt επιπέδου 2
• Σετ βαθμονομητών/
αραιωτικού μέσου HbA1c
• HbA2/F/A1c Calibrator Set • HbA2/F/A1c Calibrator/Diluent • HbA2/F/A1c Calibrator Level 1 • HbA2/F/A1c Calibrator Level 2
• HbA2-/F-/A1c-Kalibratorset Set • HbA2-/F-/A1c-Kalibrator Level 1 • HbA2-/F-/A1c-Kalibrator
• Gamme étalon HbA2/F/A1c • HbA2-/F-/A1c-Kalibrator-/ • Étalon de niveau 1 HbA2/F/A1c Level 2
• Juego de calibradores para Diluentset • Calibrador Nivel 1 para HbA2/F/A1c • Étalon de niveau 2
HbA2/F/A1c • Gamme étalon/diluant • Calibratore livello 1 dell’HbA2/F/A1c HbA2/F/A1c
• Set del calibratore HbA2/F/A1c • Calibrador Nível 1 de HbA2/F/A1c • Calibrador Nivel 2 para
dell’HbA2/F/A1c • Juego de calibradores/ • HbA2/F/A1c Kalibrator, nivå 1 HbA2/F/A1c
• Conjunto calibrador de diluyente para HbA2/F/A1c • HbA2/F/A1c kalibrator niveau 1 • Calibratore livello 2
HbA2/F/A1c • Set del calibratore/diluente • Βαθμονομητής HbA2/F/A1c dell’HbA2/F/A1c
• HbA2/F/A1c Kalibratorset dell’HbA2/F/A1c επιπέδου 1 • Calibrador Nível 2 de
• HbA2/F/A1c kalibratorsæt • Conjunto calibrador/diluente HbA2/F/A1c
• Σετ βαθμονομητών de HbA2/F/A1c • HbA2/F/A1c Kalibrator, nivå 2
HbA2/F/A1c • HbA2/F/A1c Kalibrator/ • HbA2/F/A1c kalibrator
spädningsset niveau 2
• HbA2/F/A1c kalibrator/ • Βαθμονομητής HbA2/F/A1c
fortyndingssæt επιπέδου 2
• Σετ βαθμονομητών/
αραιωτικού μέσου HbA2/F/A1c
Table of Contents
Intended Use .......................................................................................................................................... 2
Summary and Explanation of the Test ................................................................................................... 2
Principle of the Procedure ...................................................................................................................... 3
Kit Components ...................................................................................................................................... 3
Additional Items, Available from Bio-Rad ............................................................................................... 4
Additional Required Items, Not Available from Bio-Rad ......................................................................... 4
Precautions/Warnings............................................................................................................................. 4
Specimen Collection and Handling ........................................................................................................ 5
Preparation and Storage of Reagents .................................................................................................... 6
Indications of Instability or Deterioration of Reagents ............................................................................ 7
Procedure ............................................................................................................................................... 7
Method Selection .......................................................................................................................... 7
Installing a New Reorder Pack Lot (Update Kit Floppy Diskette) ................................................. 7
Analytical Cartridge Priming Procedure ....................................................................................... 7
Calibration for Short (HbA1c) Program .......................................................................................... 7
Routine Run for Short (HbA1c) Program ....................................................................................... 8
Calibration for Extended (HbA2/F/A1c) Program ............................................................................ 8
Routine Run for Extended (HbA2/F/A1c) Program ......................................................................... 9
Certification/Traceability to Reference Material and Method ........................................................ 9
QC Requirements ....................................................................................................................... 10
Guidelines for the Interpretation of Results .......................................................................................... 10
Limitations of the Procedure................................................................................................................. 11
Reference Values.................................................................................................................................. 12
Performance Characteristics ................................................................................................................ 13
Precision ..................................................................................................................................... 13
Accuracy ..................................................................................................................................... 14
Linearity/Recovery ...................................................................................................................... 16
Interfering Substances ................................................................................................................ 18
Sample Report Format ......................................................................................................................... 18
Product Safety Information ................................................................................................................... 41
Trademark Information.......................................................................................................................... 41
References ........................................................................................................................................... 41
KIT COMPONENTS
220-0201, D-10 Dual Program Reorder Pack
The reorder pack contains supplies for 400 tests for HbA1c or 200 tests for HbA2/F/A1c :
Description
220-0210* Elution Buffer 1. Two bottles containing 2000 mL of a Bis-Tris/Phosphate buffer, pH 6.0.
Contains <0.05% sodium azide as a preservative.
220-0211* Elution Buffer 2. One bottle containing 1000 mL of a Bis-Tris/Phosphate buffer, pH 6.7.
Contains <0.05% sodium azide as a preservative.
220-0112* Wash/Diluent Solution. One bottle containing 1600 mL of deionized water with <0.05%
sodium azide as a preservative.
220-0128* HbA1c Calibrator/Diluent Set. One set consisting of 3 vials of Calibrator Level 1, 3 vials of
Calibrator Level 2, and 1 bottle of Calibrator Diluent. The calibrator vials contain lyophilized
human red blood cell hemolysate with gentamicin, tobramycin, and EDTA as preservatives.
Reconstituted volume is 7 mL per vial. Calibrator Diluent contains 100 mL deionized water
with <0.05% sodium azide as a preservative.
220-0218* HbA2/F/A1c Calibrator/Diluent Set. One set consisting of 3 vials of Calibrator Level 1,
3 vials of Calibrator Level 2, and 1 bottle of Calibrator Diluent. The calibrator vials contain
lyophilized human red blood cell hemolysate with gentamicin, tobramycin, and EDTA as
preservatives. Reconstituted volume is 7 mL per vial. Calibrator Diluent contains 100 mL
deionized water with <0.05% sodium azide as a preservative.
Description
220-0297 Sample Vial Adapter, 10 x 1.5 mL
220-0375 D-10 Thermal Paper, 10 rolls
740 Lyphochek® Diabetes Bi-level Control, 6 x 0.5 mL
553 Lyphochek® Hemoglobin A2 Bi-level Control, 4 x 1.0 mL
120 Lyphochek® Hemoglobin A1c Linearity Set (1 each of 4 levels), 4 x 0.5 mL
171 Liquichek™ Diabetes Control, Level 1, 6 x 1.0 mL
172 Liquichek™ Diabetes Control, Level 2, 6 x 1.0 mL
173 Liquichek™ Diabetes Control, Level 3, 6 x 1.0 mL
172X Liquichek™ Diabetes Control, Trilevel MiniPak, 3 x 1.0 mL
PRECAUTIONS/WARNINGS
1. For in vitro diagnostic use.
2. Wear personal protective equipment while handling all reagents and samples and while operating the
D-10 system.
3. Dispose of all waste in accordance with applicable national and/or local regulations.
4. Some reagents contain sodium azide, which may react with copper or lead plumbing to form explosive
metal azides. Use caution in disposing of these reagents. If disposing to drain, flush with large volumes
of water to prevent azide buildup.
5. Waste material containing patient samples or biological products should be considered biohazardous
when disposing or treating.
6. Chemical reagents should be handled in accordance with Good Laboratory Practices.
7. Clean up all spills immediately and thoroughly. Disinfect the area for any spills involving biohazardous
materials. Dispose of all contaminated materials appropriately.
8. Do not interchange vial or bottle caps and stoppers; this will lead to cross-contamination of reagents.
9. Never mix the contents from different bottles of the same reagent. Doing so may lead to reagent
contamination and compromise the performance of the product.
10. Each unit of whole blood used in the manufacture of the calibrators and whole blood primer was tested
by FDA accepted methods and found non-reactive for HIV-1, HIV-2, Hepatitis B (HBV), Hepatitis C
(HCV) and syphilis. No test method can offer complete assurance that products containing human
source materials will be absent of these and other infectious agents. In accordance with good laboratory
practice, all human source material should be considered potentially infectious for all infectious agents;
therefore, handle the calibrators and whole blood primer with the same precautions used with patient
specimens.
11. Adherence to the protocol specified herein is necessary to ensure proper performance of this product.
12. The calibrator vial stoppers contain dry natural rubber.
Specimen Storage
Whole blood specimens may be stored up to 4 days at 2–8 °C or 1 day at room temperature (15–30 °C).
Specimen Preparation
1. Allow sample tubes to reach room temperature (15–30 °C) before performing the assay. No sample
preparation is required. Mixing the tubes prior to loading is not necessary. The sample tubes are loaded
into the D-10 sample rack and placed in the D-10. Ensure that the sample barcodes are facing towards
the back of the instrument. Use special rack inserts for 12, 13, and 14 mm diameter tubes. Remove all
inserts for 16 mm diameter tubes. Tubes with a height of 75 mm to 100 mm are acceptable for use.
2. If the sample is in an abnormal size/type tube, or if there is less than 2.0 mL of sample in the tube, then
the sample must be prediluted. Before pipetting, thoroughly mix the sample by gently inverting the tube.
To predilute, pipet 1.5 mL of Wash/Diluent Solution into a labeled 1.5 mL vial, followed by 5 µL of the
whole blood sample. Cap the sample vial and mix thoroughly. Use a sample vial adapter for 1.5 mL vials.
NOTE: Studies indicate that the use of MONOJECT™ ( 8881311446, Covidien, Mansfield, MA, 02048 USA)
collection tubes on the D-10 Hemoglobin Testing System can result in elevated total area. In some cases, blood
collects under the cap and causes oversampling, resulting in high total area that exceeds the acceptable range.
To use this tube type, users must either manually dilute the sample or ensure that blood does not collect under
the cap before loading the sample on the D-10 Hemoglobin Testing System. Results with total area within the
acceptable range are reportable.
Specimen Shipping
All samples of human origin must be shipped in accordance with national and international transportation
regulations.
HbA2/F/A1c Calibrators
Reconstitute and store the HbA2/F/A1c Calibrators as directed in the Dual Program HbA2/F/A1c Calibrator/
Diluent Set Insert.
Extracted Standards
This HPLC method does not use extracted standards.
Controls
• Reconstitute and store the Diabetes and Hemoglobin A2 controls according to the manufacturer’s
package insert.
• Bio-Rad Lyphochek Diabetes Controls and Hemoglobin A2 Controls must be diluted 1:300 prior to
analysis. Pipet 1.5 mL of Wash/Diluent Solution into a labeled 1.5 mL vial, followed by 5 µL of the
reconstituted control. Cap each control vial and mix thoroughly.
• Bio-Rad Liquichek Diabetes Controls must be diluted 1:200 prior to analysis. Pipet 1.0 mL of Wash/
Diluent Solution into a labeled 1.5 mL vial, followed by 5 µL of the control. Cap each control vial and mix
thoroughly.
PROCEDURE
Method Selection
From the LOT INFO screen:
• Press METHOD on the touch screen.
• Select the desired method (HbA1c) or (HbA2/F/A1c).
• Press EXIT.
• Press YES to confirm the method change.
• Press EXIT.
• The selected method is indicated to the left of the status bar.
There is no need to perform a system flush unless you are installing a new lot of reagents.
Installing a New Reorder Pack Lot (Update Kit Floppy Diskette)
• Go to the LOT INFO screen.
• Press the UPDATE KIT button.
• Place the UPDATE KIT floppy diskette in the A:\drive.
• Follow the instructions on the screen to proceed with the Update Kit Procedure.
• Both the A1c and the A2/F/A1c parameters will be updated during this procedure.
• Remove the floppy diskette from the A:\drive once the procedure is completed.
Analytical Cartridge Priming Procedure
• The priming procedure must be performed once per cartridge and also after decontaminating the
D-10 System.
• The priming procedure can be performed while either the HbA1c or HbA2/F/A1c test is selected.
• Pipet 1 mL of reconstituted Whole Blood Primer into a sample vial. Put the sample vial into a sample vial
adapter labeled with a Primer barcode, then place the adapter into sample rack position 1.
• Start a run. When finished, the cartridge is ready for calibration.
IFCC NGSP
43 mmol/mol 6.1%
48 mmol/mol 6.5%
53 mmol/mol 7.0%
QC Requirements
Controls should be run for the reported analyte(s).
In keeping with good laboratory practice, low and high control specimens should be included in the run once per
24 hours. A repeat run is indicated when expected control values are not obtained.
When the result falls outside of the reportable range, the value is asterisked (*). HbA1c values that fall
outside of the reportable range should not be reported. To be consistent with the established reportable
ranges, HbA2 and HbF values that are lower than the reportable range should be reported as less than
1.5% for HbA2 or less than 0.8% for HbF. Very low levels of HbF may result in a negative value; these
values should be reported as less than 0.8%. Values in excess of the reportable ranges for HbA2 and
HbF should be reported as greater than 11.4% for HbA2 or greater than 16.5% for HbF.
6. Any sample with >15% or >140 mmol/mol HbA1c should be suspected of having a hemoglobin variant.15
7. Any sample with a combined area of ≥60% in the Variant (Short Program only), S, and C windows
should be suspected of having a homozygous variant or a variant–β-thalassemia phenotype. The HbA1c
result should not be reported for these samples.16
P3 Peak Resolution
The P3 peak may split due to improved resolution. When this occurs, an "Unknown" peak will be listed in the
peak table following the P3 peak (see Figure 8). The presence of this "Unknown" peak has no effect on the
HbA1c, HbA2, or HbF quantitation.
Hemoglobin Variants
HbA1c values determined using the D-10 Dual Short Program for HbS trait and HbC trait specimens showed no
clinically significant difference versus values determined by an NGSP certified boronate affinity method. Typical
chromatograms for HbS and HbC trait specimens are provided in Figures 10 and 11.
HbA1c values determined using the D-10 Dual Extended Program for HbE, HbD, HbS, and HbC trait specimens
showed no clinically significant difference versus values determined by an NGSP certified boronate affinity
method. Typical chromatograms for HbE, HbD, HbS, and HbC trait specimens are provided in Figures 18–21,
respectively.
In the rare homozygous forms (SS or CC), there is no HbA present; therefore, no HbA1c value can be
determined.
A degradation peak, which coelutes with the A2 peak, may appear in HbS samples which have not been stored
appropriately (see Figure 22). This degradation peak interferes with the quantitation of HbA2. Review the A2
peak shape for all HbS samples before reporting. Do not report results if the peak shape is abnormal.
A degradation peak that interferes with the HbA1c quantitation may appear in HbE and HbD samples which have
not been stored appropriately (see Figure 23). Do not report the HbA1c result if an “Unknown” peak is identified
between the A1c and P3 peaks.
The effect of other hemoglobin variants on the quantitation of HbA1c, HbA2, and HbF has not been fully
evaluated.
Chromatograms of 4 specimens containing hemoglobin variants (as presumptively identified by gel
electrophoresis), analyzed using the Extended Program, are provided in Figures 24–27.
REFERENCE VALUES
Hemoglobin A1c Ranges20
The following HbA1c ranges may be used for interpretation of results; however, factors such as duration of
diabetes, adherence to therapy, and the age of the patient should also be considered in assessing the degree
of blood glucose control. These values are for nonpregnant individuals. "Action Suggested" depends on
individual patient circumstances. Such actions may include enhanced diabetes self-management education,
co-management with a diabetes team, referral to an endocrinologist, change in pharmacological therapy,
initiation or increased self-monitoring of blood glucose, or more frequent contact with the patient.
Hemoglobin A1c (%) Degree of Glucose Control
>8 Action Suggested†
<7 Goal‡
<6 Non-Diabetic Level
† High risk of developing long-term complications such as retinopathy, nephropathy, neuropathy and cardiopathy.
Action suggested depends on individual patient circumstances.
‡ Some danger of hypoglycemic reaction in Type I diabetics. Some glucose intolerant individuals and "sub-clinical"
diabetics may demonstrate (elevated) HbA1c in this area.
PERFORMANCE CHARACTERISTICS
Precision
The precision of the D-10 Dual Program was evaluated based on the NCCLS EP5-T2 guideline (for the Short
Program) and EP5-A guideline (for the Extended Program), "Evaluation of Precision Performance of Clinical
Chemistry Devices." In these studies, 40 runs were performed over 20 working days. In each run, aliquots
of low and high specimens were analyzed in duplicate. The results of the precision study are summarized in
Tables 2–5.
Accuracy
D-10 Dual Extended vs Short Program
Hemoglobin A1c 14
12
intercept = −0.5416
4
R2 = 0.9953 4 6 8 10 12 14
D-10 %HbA1c
performed on 40 patient samples
analyzed in duplicate. See Figures 2 and
3.
slope = 0.9743
intercept = 0.3078
R2 = 0.9945
VARIANT II %HbA1c
14
D-10 Dual Extended Program %HbA 1c
12
10
slope = 0.9906
intercept = 0.431 8
R2 = 0.9843 6
4
4 6 8 10 12 14
VARIANT II %HbA 1C
Hemoglobin A2
Hemoglobin A2 results obtained from
D-10 Dual Program vs VARIANT II β-thalassemia Short Program (HbA 2)
the D-10 Dual Extended Program were 10
results. 4
slope = 1.0898 3
2
intercept = −0.2407
1
R2 = 0.9832 0
0 1 2 3 4 5 6 7 8 9 10
VARIANT II %HbA2
4
slope = 0.9497
2
intercept = −0.1785
0
R2 = 0.9959 0 2 4 6 8 10 12 14 16
-2
VARIANT II %HbF
Linearity/Recovery
To demonstrate linearity on the D-10 Dual Program throughout the reportable range, low and high specimens
were derived as follows:
HbA1c (Short Program)
Low: Whole blood from a normal patient sample was supplemented with Bio-Rad Lyphochek
Hemoglobin A1c Linearity Set Level 1 to yield relative HbA1c level of 3.8%.
High: Whole blood from a diabetic patient sample was supplemented with Bio-Rad Lyphochek
Hemoglobin A1c Linearity Set Level 4 to yield relative HbA1c level of 18.5%.
HbA1c (Extended Program)
Low: Whole blood from a normal patient sample was supplemented with Bio-Rad Lyphochek
Hemoglobin A1c Linearity Set Level 1 to yield relative HbA1c level of 3.7%.
High: Whole blood from a diabetic patient sample was supplemented with Bio-Rad Lyphochek
Hemoglobin A1c Linearity Set Level 4 to yield relative HbA1c level of 18.4%.
HbA2
Low: Whole blood with a HbA2 level of 1.5%.
High: Normal whole blood spiked with purified HbA2 to a level of 11.4%.
HbF
Low: Normal whole blood with HbF level of 0.8%.
High: Normal whole blood spiked with an elevated HbF sample to a level of 16.5%.
The high specimen was diluted with the low specimen in varying ratios to yield specific relative analyte
percentages (theoretical percent). These diluted samples were analyzed with the D-10 Dual Program
(observed percent). Percent recovery was determined by dividing the observed percent by the theoretical
percent and multiplying the result by 100. Results from the Linearity Study are shown in Tables 6–9.
% Contribution
Theoretical % HbA1c Observed % HbA1c % Recovery
Low High
100 0 3.8 3.8 100
80 20 6.6 6.6 100
67 33 8.6 8.5 98.8
50 50 11.0 11.0 100
33 67 13.5 13.2 97.8
20 80 15.4 15.2 98.7
0 100 18.5 18.5 100
Table 6: Results of Study on Linearity and Recovery for HbA1c, Short Program
% Contribution
Theoretical % HbA1c Observed % HbA1c % Recovery
Low High
100 0 3.7 3.7 100
80 20 7.1 6.7 94.4
67 33 9.2 8.8 95.7
50 50 11.7 11.4 97.4
33 67 14.1 13.7 97.2
20 80 15.9 15.7 98.7
0 100 18.4 18.4 100
Table 7: Results of Study on Linearity and Recovery for HbA1c, Extended Program
% Contribution
Theoretical % HbA2 Observed % HbA2 % Recovery
Low High
100 0 1.5 1.5 100
80 20 3.5 3.3 94.3
67 33 4.9 4.7 95.9
50 50 6.5 6.4 98.5
33 67 8.2 8.1 98.8
20 80 9.5 9.6 101.1
0 100 11.4 11.4 100
Table 8: Results of Study on Linearity and Recovery for HbA2, Extended Program
Concentration: % mmol/mol
A1c 5.1 32
Patient report
Bio-Rad DATE: 09/22/2010
D-10 TIME: 04:53 PM
S/N: #DA3G222606 Software version: 3.57
Sample ID: RACK03-7-19-20-9-2010
Injection date 09/20/2010 11:52 AM
Injection #: 19 Method: HbA1c
Rack #: 03 Rack position: 7
Concentration: % mmol/mol
A1c 9.2 77
Concentration:
% A1c 10.3
Control OK
Figure 8: Split P3 Peak, Short Program
Concentration:
% A1c 5.4
Patient report
Bio-Rad DATE: 09/23/2010
D-10 TIME: 02:57 PM
S/N: #DA3G222606 Software version: 3.57
Sample ID: RACKC2-1-24-17-9-2010
Injection date 09/17/2010 11:18 AM
Injection #: 24 Method: HbA1c
Rack #: C2 Rack position: 1
Concentration: % mmol/mol
A1c 6.3 46
Concentration: % mmol/mol
A1c 5.1 32
Patient report
Bio-Rad DATE: 10/14/2010
D-10 TIME: 07:35 PM
S/N: #DA2D000801 Software version: 3.60
Sample ID: NP
Injection date 10/14/2010 07:15 PM
Injection #: 23 Method: HbA2/F
Rack #: 05 Rack position: 3
Concentration: % mmol/mol
F < 0.8 * ---
A1c 4.8 29
A2 2.7 ---
Concentration: % mmol/mol
F < 0.8 * ---
A1c 9.1 76
A2 2.6 ---
Figure 13: Diabetic Sample with Elevated HbA1c Level, Extended Program
Concentration:
% A1c 4.9
% A2 2.8
Concentration:
%F < 0.8 *
% A1c 5.8
% A2 2.4
Figure 15: Patient Sample with Normal Hemoglobin F and A2, Extended Program
Concentration:
%F < 0.8 *
% A1c 5.7
% A2 6.8
Figure 16: Patient Sample with Elevated Hemoglobin A2, Extended Program
Concentration:
%F 6.1
% A1c 5.5
% A2 2.4
Concentration:
% A1c 4.9
% A2 22.9 *
Concentration:
% A1c 5.1
% A2 2.9
Patient report
Bio-Rad DATE: 10/12/2010
D-10 TIME: 04:17 PM
S/N: #DA3G222606 Software version: 3.57
Sample ID: RACKC1-1-29-12-10-2010
Injection date 10/12/2010 04:09 PM
Injection #: 29 Method: HbA2/F
Rack #: C1 Rack position: 1
Concentration: % mmol/mol
F < 0.8 * ---
A1c 6.5 48
A2 3.2 ---
Patient report
Bio-Rad DATE: 10/13/2010
D-10 TIME: 04:16 PM
S/N: #DA3G222606 Software version: 3.57
Sample ID: RACK08-4-47-13-10-2010
Injection date 10/13/2010 03:18 PM
Injection #: 47 Method: HbA2/F
Rack #: 08 Rack position: 4
Concentration: % mmol/mol
F < 0.8 * ---
A1c 5.5 37
A2 3.1 ---
Concentration:
%F < 0.8 *
% A1c 4.9
% A2 2.5
Concentration:
%F 1.2
% A1c 4.5
% A2 29.6 *
Figure 23: Patient Sample with Hemoglobin E trait, Extended Program: Degradation
Peak (Unknown) eluting between A1c and P3 peaks
Concentration:
%F 1.8
% A2 3.6
Concentration:
%F 3.9
% A1c 7.2
% A2 2.7
Figure 25: Patient Sample with Hemoglobin J-Baltimore trait, Extended Program
Concentration:
% A1c 5.4
% A2 3.1
Figure 26: Patient Sample with Hemoglobin O-Arab trait, Extended Program
Concentration:
% A1c 3.5 *
% A2 1.1 *
TRADEMARK INFORMATION
D-10, DIAMAT, VARIANT, and Liquichek are trademarks of Bio-Rad Laboratories, Inc.
Lyphochek is a registered trademark of Bio-Rad Laboratories, Inc.
MONOJECT is a trademark of Covidien.
REFERENCES
1. American Diabetes Association Home Page. http://www.diabetes.org (accessed Jan 2010).
2. Fowler, M. J. Microvascular and Macrovascular Complications of Diabetes. Clin. Diabetes 2008, 26 (2),
77–82.
3. Shaw, J. E.; Sicree, R. A.; Zimmet, P. Z. Global Estimates of the Prevalence of Diabetes for 2010 and
2030. Diabetes Res. Clin. Pract. 2010, 87, 4–14.
4. Forsham, P. H. Diabetes Mellitus: A Rational Plan for Management. Postgrad. Med. 1982, 71, 139–154.
5. Hollander, P. The Case for Tight Control in Diabetes. Postgrad. Med. 1984, 75, 80–87.
6. Baynes, J. W.; Bunn, H. F.; Goldstein, D.; Harris, M.; Martin, D. B.; Peterson, C.; Winterhalter, K. National
Diabetes Data Group: Report of the Expert Committee on Glucosylated Hemoglobin. Diabetes Care
1984, 7, 602–606.
7. Nathan, D. M.; Singer, D. E.; Hurxthal, K.; Goodson, J. D. The Clinical Information Value of the
Glycosylated Hemoglobin Assay. N. Engl. J. Med. 1984, 310, 341–346.
8. Mayer, T. K.; Freedman, Z. R. Protein Glycosylation in Diabetes Mellitus: A Review of Laboratory
Measurements and of Their Clinical Utility. Clin. Chim. Acta 1983, 127, 147–184.
9. Fairbanks, V. F. Thalassemias and Related Disorders. Hemoglobinopathies and Thalassemias; Brian C.
Decker: New York, 1980; pp 18–27.
10. Fairbanks, V. F. The “Common” Hemoglobin Variants. Hemoglobinopathies and Thalassemias; Brian C.
Decker: New York, 1980; pp 10–14.
11. Rohlfing, C. L.; Little, R. R.; Wiedmeyer, H. M.; England, J. D.; Madsen, R.; Harris, M. I.; Flegal, K. M.;
Eberhardt, M. S.; Goldstein, D. E. Use of GHb (HbA1c) in Screening for Undiagnosed Diabetes in the
U.S. Population. Diabetes Care 2000, 23, 187–191.
12. International Federation of Clinical Chemistry and Laboratory Medicine Home Page. http://www.
ifcchba1c.net (accessed Jan 2010).
13. Hoelzel, W.; Weykamp, C.; Jeppsson, J. O.; Miedema, K.; Barr, J. R.; Goodall, I.; Hoshino, T.; John, W.
G.; Kobold, U.; Little, R.; Mosca, A.; Mauri, P.; Paroni, R.; Susanto, F.; Takei, I.; Thienpont, L.; Umemoto,
M.; Wiedmeyer, H. M.; IFCC Working Group on HbA1c Standardization. IFCC Reference System for
Measurement of Hemoglobin A1c in Human Blood and the National Standardization Schemes in the
United States, Japan, and Sweden: A Method-Comparison Study. Clin. Chem. 2004, 50 (1), 166–174.
14. American Diabetes Association, European Association for the Study of Diabetes, International
Federation of Clinical Chemistry and Laboratory Medicine, and International Diabetes Federation.
Consensus Statement on the Worldwide Standardization of the Hemoglobin A1c Measurement. Diabetes
Care 2007, 30 (9), 2399–2400.
NOTES:
Australia, Bio-Rad Laboratories Pty. Ltd., Level 5, 446 Victoria Road, Gladesville NSW 2111 • Phone 61-2-9914-2800 • Telefax 61-2-9914-2888
Clinical Austria, Bio-Rad Laboratories Ges.m.b.H., Hummelgasse 88/3-6, A-1130 Vienna • Phone 43-1-877-8901 • Telefax 43-1-876-5629
Diagnostics Group Belgium, Bio-Rad S.A.-N.V. Begoniastraat 5, B-9810 Nazareth Eke • Phone 32-9-385-5511 • Telefax 32-9-385-6554
Brazil, Bio-Rad do Brasil, Praia de Botafogo, 440-3rd Floor, Botafogo, RJ CEP 22250-040, Rio de Janeiro • Phone 5521-3237-9400 • Telefax 5521-2527-3099
Canada, Bio-Rad Laboratories, Ltd., 2403 Guénette Street, Montréal, Québec H4R 2E9 • Phone 1-514-334-4372 • Telefax 1-514-334-4415
4000 Alfred Nobel Drive China, Bio-Rad Laboratories Shanghai Ltd., 3rd Floor, #18 Dong Fang Road, Bldg E, Poly Plaza, Pudong, Shanghai, PRC 200120 • Phone 86-21-64260808 • Telefax 86-21-64264988
Czech Republic, Bio-Rad spol. s r.o., Nad ostrovem 1119/7, 147 00 Prague 4 • Phone 420-241-430-532 • Telefax 420-241-431-642
Hercules, California 94547 Denmark, Bio-Rad Laboratories, Symbion Science Park, Fruebjergvej 3, DK-2100 Copenhagen East • Phone +45-4452-1000 • Telefax +45-4452-1001
Telephone (510) 724-7000 Finland, Bio-Rad Laboratories, Linnanherrankuja 16, FIN-00950 Helsinki • Phone 358-9-804-22-00 • Telefax 358-9-7597-5010
FAX (510) 741-6373 France, Bio-Rad, 3 boulevard Raymond Poincaré, 92430 Marnes-la-Coquette • Phone 33-1-47-95-60-00 • Telefax 33-1-47-41-91-33
www.bio-rad.com/diagnostics Germany, Bio-Rad Laboratories GmbH, Heidemannstrasse 164, D-80939 Munich • Phone +49-(0)89-318-840 • Telefax +49-(0)89-318-84100
Greece, Bio-Rad Laboratories M E.P.E, 2-4 Mesogeion Street, Fourth Floor 115 27 Athens • Phone 30-210-7774396 • Telefax 30-210-7774376
Hong Kong, Bio-Rad Pacific Ltd., Unit 1101, 11/F DCH Commercial Centre, 25 Westlands Road, Quarry Bay • Phone 852-2789-3300 • Telefax 852-2789-1290
Hungary, Bio-Rad Hungary Ltd., H-1082 Budapest, Futo street 47-53, Hungary • Phone +36-1-459-6100 • Telefax +36-1-459-6101
India, Bio-Rad Laboratories (India) Pvt. Ltd., Bio-Rad House, 86-87, Udyog Vihar Phase IV, Gurgaon, Haryana 122 015 • Phone 1-800-180-1224 • Telefax 91-124-2398115
Israel, Bio-Rad Laboratories Ltd., 14 Homa Street, New Industrial Area, Rishon Le Zion 75655 • Phone 972-3-9636050 • Telefax 972-3-9514129
Italy, Bio-Rad Laboratories S.r.l., Via Cellini 18/A, 20090 Segrate, Milan • Phone +39-02-216091 • Telefax +39-02-21609-398
Japan, Bio-Rad Laboratories K.K., Tennoz Central Tower 20F, 2-2-24 Higashi-Shinagawa, Shinagawa-ku, Tokyo 140-0002 • Phone 81-3-6361-7070 • Telefax 81-3-5463-8483
Korea, Bio-Rad Korea Ltd., 10th Floor, Hyunjuk Building, 832-41, Gangnam-gu, Seoul 135-080 • Phone 82-2-3473-4460 • Telefax 82-2-3472-7003
Mexico, Bio-Rad, S.A., Avenida Eugenia 197, Piso 10-A, Col. Narvarte, C.P. 03020 Mexico, D.F. • Phone +52 (55) 5488-7670 • Telefax 52 (55) 1107-7246
The Netherlands, Bio-Rad Laboratories B.V., Fokkerstraat 2-8, 3905 KV Veenendaal • Phone +31-318-540666 • Telefax +31-318-542216
New Zealand, Bio-Rad New Zealand, 189 Bush Road Unit B, Albany, Auckland • Phone 64-9-415-2280 • Telefax 64-9-415-2284
Norway, Bio-Rad Laboratories, Johan Scharffenbergs vei 91, N-0694 Oslo • Phone 47-23-38-41-30 • Telefax 47-23-38-41-39
Poland, Bio-Rad Polska Sp. z o.o., Nakielska Str. 3, 01-106 Warsaw • Phone 48-22-3319999 • Telefax 48-22-3319988
Portugal, Bio-Rad Laboratories, Lda., Edificio Prime, Ave. Quinta Grande, 53 – Fracção 3B Alfragide 26114-521 Amadora • Phone 351-21-472-7700 • Telefax 351-21-472-7777
Russia, Bio-Rad Laboratorii, Business Centre “West Bridge”, Leningradsky pr-t H.37A Bld. 14, 125167 Moscow • Phone 7-495-721-14-04 • Telefax 7-495-721-14-12
Singapore, Bio-Rad Laboratories (Singapore) Pte. Ltd., 27 International Business Park, #01-02 iQuest @IBP, Singapore 609924 • Phone 65-6415-3170 • Telefax 65-6415-3189
South Africa, Bio-Rad Laboratories (Pty) Ltd., 34 Bolton Road, Parkwood, Johannesburg 2193 • Phone 27-11-442-85-08 • Telefax 27-11-442-85-25
Spain, Bio-Rad Laboratories, S.A., C/ Caléndula, 95, Edificio M. Miniparc II, El Soto de la Moraleja, 28109 Madrid • Phone 34-91-590-5200 • Telefax 34-91-590-5211
Sweden, Bio-Rad Laboratories A.B., Vintergatan 1, Box 1097, S-172 22 Sundbyberg • Phone 46-8-555-127-00 • Telefax 46-8-555-127-80
Switzerland, Bio-Rad Laboratories AG, Nenzlingerweg 2, CH-4153 Reinach BL• Phone 41-61-717-95-55 • Telefax 41-61-717-95-50
Thailand, Bio-Rad Laboratories Ltd., 1st & 2nd Floor, Lumpini I Bldg., 239/2 Rajdamri Rd., Lumpini, Pathumwan, Bangkok 10330 • Phone 662-651-8311 • Telefax 662-651-8312
United Kingdom, Bio-Rad Laboratories Ltd., Bio-Rad House, Maxted Road, Hemel Hempstead, Herts HP2 7DX • Phone +44-(0)20-8328-2000 • Telefax +44-(0)20-8328-2550