Panoura 18s With Ceph Calibration and Alignment Quick Guide
Panoura 18s With Ceph Calibration and Alignment Quick Guide
Panoura 18s With Ceph Calibration and Alignment Quick Guide
914-592-6100
1. Computer set up
2. Tube head aging
3. Panoramic alignment
4. Panoramic sensor calibration
5. Ceph alignment
6. Ceph sensor calibration
7. Secondary collimater alignment
8. Ear rod alignment
9. Final Test Exposure
1. P18_CM.exe
2. P18_Launcher.exe
3. P18S_Setup.exe
4. Panoura18S.exe
1. Control Manager
2. Image Creator
1/11/2022
Alignment
Collimator
assembly
pan
slit
Place 20 mm aluminum block
ceph slit
over the pan collimator slit
Adjusting Pushing
Screws Screw
Note: The pushing screw only pushes, but does not pull. So turning the
push screw clockwise by one turn, will move the collimater approximately 1
mm to the left (when facing the collimater). However, to move in the opposite
direction, turning the screw counterclockwise one turn will create space, and
the collimater needs to be physically “tapped” to move it.
16
Note: vertical or rotational adjustments may be required. Please refer to
manual
ImageWorks Proprietary – Confidential Information
1/11/2022
3. Panoramic Mechanical Alignment
Turn off FL07 by pressing and holding down the MODE and FIX buttons on the
indicator panel simultaneously for about 2 seconds.
If arrows are green, tighten adjusting screws, and repeat steps for power cycling
the unit, reopening Control Manager and scanning.
Confirm that arrows remain green after tightening. If all is green, then click Close.
ImageWorks Proprietary – Confidential Information
18
1/11/2022
4. Panoramic Sensor Calibration
Remove all adapters such as chin rest block from irradiation path
of the x-ray. Push temple supports up.
Note: unit needs
to be powered on
Open Ajat Dental Imaging Software for about 5 min to
perform
calibration
Select Tools → Maintenance → Calibration
Click Yes
Make sure that secondary collimator and ear rods are removed
Press RESET
Press BEAM
Note: The push screw only pushes, but does not pull. So turning the push
screw clockwise by one turn, will move the collimator approximately 1 mm to
the right (when facing the collimater). However, to move in the opposite
direction, turning the screw counterclockwise one turn will create space, and
the collimater needs to be physically “tapped” to move it.
Remove the FL07 exposure lock by pressing and holding down the MODE
and FIX buttons on the indicator panel simultaneously for about 2-3
seconds.
Initiate scan.
If arrows are green, tighten adjusting screws and locking nuts, and repeat
steps for power cycling the unit, reopening Control Manager and
scanning.
Confirm that arrows remain green after tightening. If all is green, then
click Close.
Click Yes
Press RESET
Press BEAM
Tip: when installing ear rods, before tightening bolts, first allow gravity to
get them to hang plum (vertical), and then tighten.
Note: the side with the larger ear ring is the side
farthest from the sensor
Tip: after first exposure, the handswitch for each following exposure can be released
as soon as ear rings are revealed in the image. This can help avoid cool down wait
times.
Note:
Because the test scan does not
have patient anatomy, it is possible
that lines or artifacts may be seen
in other parts of the test image.
This will not affect quality of actual
dental images taken.