Professional Documents
Culture Documents
1.HF-525Plus VET (32kW, 400ma, 230Vor110V, XR5)
1.HF-525Plus VET (32kW, 400ma, 230Vor110V, XR5)
Important!
Please read this manual thoroughly and use the system correctly before using the
system. This manual desires to keep in promised place to use always again after
reading.
Accordingly, when use this product should be thoroughly read the "Radiation Safety Control Regulation"
and then compliance with it in use. This product is serious injury's danger if do not keep use guideline
or regulation therefore, read and understood this manual before in use.
This product was designed by safety level (SL) presenting in Radiation Safety Control Regulation.
However, if do not keep using method and Radiation Safety Control Regulation, will not be able to
avoid serious injury's danger from radiation. Therefore, to prevent accident from radiation, everyone who
work connected with radioactive emission must understand radiation safety control regulation and
understand product manual fulfill thorough education and formality.
Our company recognizes enough about danger of excessive exposure to operator, installation people,
and maintenance people. Therefore, our company strongly advises that operator, installation people,
maintenance people must be educated about radiation safety control regulation and should be used this
product when understand fluently about all knowledge of operation this product.
Also, our company informs that operator, installation people, and maintenance people have a
responsibility to be educated fluently about radiation safety control regulation and understood about
product use method.
Also, our company has no responsibility about injury or calamity that occurs does not keep use method
of this product and radiation safety control regulation.
Therefore, our company strongly advises that to be used this product after that person in charge of
installation, maintenance and operator should understood radiation safety control regulation of domestic
or foreign.
States a direct danger that may cause death or serious injury if it is not avoided.
States an indirect or potential danger that may cause death or serious injury if it is not avoided.
State a danger that may cause slight or medium injury or may cause damage in equipment or fire
if it is not avoided.
NOTE ☞
CONTRAINDICATION!
1. The owner/user have a responsibility for management and maintenance of this equipment.
2. This equipment must be operated by only authorized personnel.
3. Contact authorized service engineer when repair or inspection of the inside of the equipment.
4. In general, modifications are strictly prohibited by the Regulations. Please contact our agency if it is
needed to modify the device.
5. Do not use this system in conjunction with the other machines because seriously danger could be
happened unexpectedly.
Periodic Inspection
1. Maintenance is required to maintain safety and performance of this system for a long time.
2. Maintenance must be done periodically with maintenance plan.
3. This system must be inspected from the inspection office per every 3 years.
Warranty
1. This warranty shall apply only to defects that arise within 12months of purchase.
2. This product could have a free of charge service if troubles are occurred during the warranty
period.
3. The warranty does not cover defects caused by abuse, mishandling, accidents or improper
installation.
4. The following will not be covered under the warranty.
4-1. Modifications to the product, or repairs made by unauthorized person but manufacturer or
authorized person by manufacturer.
4-2. Acts of God. Including fire, flood, storm, power or voltage surges.
4-3. Improper use or installation.
4-4. Defects that arise in conjunction with the use of other machines.
4-5. Use of improper parts or modification the circuit.
4-6. Improper use
5. In case of repaired during the warranted period, User shall advise the result of test to manufacture.
Precautions in Use
When using this equipment, please observe the following precautions for the safety of the operator
and the patient.
1. Improper use of the X-ray equipment might cause the operator or the patient to be accidentally
exposed to X-ray radiation.
2. During X-ray radiation, any person other than the subject the patient should not stay in the room.
3. X-ray room must have radiation protection facility by "Radiation Safety Control Regulation"
1. This system is made radiation and must shield completely X-ray room with lead shield. After that,
must be inspected by authorized inspection office and then have to use system.
2. In the radiation exposure, operator have to adjust exposure area with suitable exposure point and
try to expose it with not over exposure.
3. The most effective method to secure from first and second radiation defilades perfectly and
should be established x-ray room by lead etc... And, to minimize radiation exposure from
radiation exposure to operator, lead partition, lead gloves, apron, cover, Gonad Shields, Thyroid
collar, Ovary Shields must be used.
NOTE ☞
For the safety of operator and patient, opening the X-ray generator cover, X-ray tube,
Collimator and other equipment cover are strictly prohibited. Only qualified engineer
could open the cover.
Scope
1. This manual contains information of system operation, components, control keys, switches and
hazards and technical reference information.
2. This manual may not be reproduced in whole or in part without the written permission of ECORAY
CO., LTD.
3. The text of this manual is a translation based on the Korean alphabet (Hangŭl) original.
© January, 2007
EcoRay Co., Ltd.
60-10, Nanosandan 5-ro, Nam-myeon, Jangseong-gun, Jellanam-do, Korea
Revision History
Abbreviation
kV Kilo-voltage
mA milliampere
mAs milli-ampere Second mA*Sec
mS milli-second 1/1000 second
APR Automatic Programming Radiography
AEC Automatic Exposure Control Optional
Summary of Functions
Switching OFF
Switching ON
Error display
Reset switch
Table of Contents
Chapter 1
1.1 An Overview
This manual contains operation instructions for HF-525 PLUS VET and also product description,
safety and operation method.
This X-ray system was designed for diagnosis’s human beings and could get good regulations by
controlled digital value by microprocessor.
This X-ray system has a self-diagnosing function and then could improve service quality, save time
and decrease customers complain.
6. Specially, radiant ray can be accumulated in body and can be continued from several months to
several years. Therefore, safest method for operator avoids for the first, second beam.
7. Also, because the second radiate beam can be stronger than reach to film, must take safeguard by
this.
8. The most effective method to secure from first and second radiation defilades perfectly and should
be established x-ray room by lead etc... And, to minimize radiation exposure from radiation
exposure to operator, lead partition, lead gloves, apron, cover, Gonad Shields, Thyroid collar, Ovary
Shields must be used.
Electrical circuits inside the equipment may use voltages which are capable of causing
serious injury or death from electric shock. To avoid this hazard, never remove any of
the cabinet covers.
Observe the following safety procedures to avoid electric shock or serious injury to operators and
patients and to avoid system malfunction.
1. Do not remove any of the assembly covers.
2. Do not touch all electric circuit and components to be connected this x-ray system.
3. Do not bypass, jumper or otherwise disable the safety interlocks.
4. Do not place food or beverage containers on any part of the equipment. If spilled they can cause
short circuits and power off immediately.
5. Always remove power to the equipment before cleaning.
6. When cleaning the cover, use a slightly damp cloth or sponge for cleaning.
7. When cleaning the inside of system, remove the cover then remove the dust with duster. Only
qualified service engineer could clean the inside of system. (It is danger when touched electric
parts)
8. Only qualified service engineers are allowed to service or repair a system.
This X-ray system is not rated for water-tight operation. Water, soap or other
liquids which get inside the may cause short-circuits, electrical shocks or fire. If
liquid should enter the system, do not connect the unit to the power supply. If
liquids drip into the equipment, disconnect the power cord and do not operate the
system until it can be cleaned and inspected by a qualified service engineer (EcoRay
engineer or Representative’s engineer).
The operator / patient will be seriously body injured or death by potential radiation
hazards used x-ray system without maintenance plan, using procedure, operation
method.
1.5 Maintenance
1.5.1 Maintenance must be done by qualified engineer by manufacturer.
1.5.2 This X-ray system must be maintained periodically by local or state regulations.
1.6 Responsibilities
1.6.1 Ensure that all authorized person to use the equipment are aware of the danger of excessive
exposure to X-Ray radiation.
1.6.2 The responsibility herein described that is well acquainted with sufficient training of the
radiation safety control regulation and instruction of the equipment by definitely and accurately.
1.6.3 The manufacturer is not liable for injury and damage which may result from violation of the
radiation safety control regulation and instruction of the equipment.
1.6.4 The owner/operator must confirm that X-ray room must comply with relative regulations.
1.6.5 The owner/operator must ask relative regulations to be used correctly complied with regulations
to government office or independent institute. And the owner/operator has a responsibility
about this.
1.6.6 This X-ray system must be used in the suitable X-ray room to be complied with relative
regulations and technical requirements. It is the responsibility of owner/operator to keep this
rule
1.6.7. The owner of this system has a responsibility to control strictly exposed radiation of operator
and must order to operator to be used “Pocket Dose meter”.
The owner/operator must know that it is very dangerous if radiation over exposed to patient.
The manufacturers don’t have a responsibility that if seriously bodily injured or accident
occurred by over X-ray exposed.
If the system is operated together with other electrical devices from a multiple socket outlet,
current leakage can increase; the operator/user in this case is responsible for ensuring that the
requirements of IEC 60601-1-1/IEC 60601-1 are still met.
1.11 Contraindication
This x-ray system is not suitable for therapeutic applications.
1.13 Training
After installation, EcoRay engineer or EcoRay’s agent will carry our training of its use and if you
have any further questions on the system, please contact address as shown in this manual.
Obey the instructions of operation manual for both patient and operator’s continuously safety
and read and understood the operation manual and keep the manual in accessible place.
Radiation Safety plan must be established and advice strongly understand and observe the
radiation protection regulations
1.15 Specification
1.16 Option
1.17 Labels
1.17.1 Main parts of generator has a labels contains information as follows.
1.17.2 Labeling: Product Name, Model, Main power, manufactured date, Manufacturer & Address,
Serial No. are printed and attached at the generator.
1.18 Certifications
1.18.1 EcoRay have certificate as below about production for this product.
1. ISO 9001
2. ISO 13485
Troubles are could be occurred in the use or storage under unsuitable environment condition.
X-ray tube
Tube Stand
Collimator
Table
Console
X-ray
Generator
1.20.2 Function
1. This X-ray system is consisting of Tube Stand which supports X-ray tube, Console, X-ray gen
erator, table, and X-ray tube unit.
2. Tube Stand: Assembled with table, x-ray tube, collimator and operator’s console.
3. Console: Consist of each switch and could adjust exposure values and display it.
4. X-ray generator: Consists of control unit and high voltage transformer unit and PSU. And co
ntrol x-ray exposure values and make high voltages by console.
5. Table: Consists of tabletop and bucky devices and animal will lie down on the table.
6. X-ray tube unit: Consist of X-ray tube, Collimator.
7. X-ray radiation will be made by pushing of X-ray switch of console or pushing of hand switc
h or foot switch.
8. X-ray radiation will permeate animal from x-ray tube to bucky device and after permeation, c
ould archive x-ray images by x-ray film or digital detector.
1.23.1 Please observed the each national waste treatment regulations when disuse system or parts
If the unit has been powered off, but there are still indications that parts of the
system are still powered on, off the circuit breaker switch if want to completely
powered off.
● Push the Power ON of Console: Standby lamp on and off with yellow and green
● Push the Power OFF of Console: When push the Power OFF switch, green lamp light
up on the Power OFF, at the same time Standby lamp light up with yellow.
● In spite of the power off with circuit breaker, power supplied to system, call
EcoRay Service engineer and do not use the system until it has been
inspected by service engineer.
If used with another equipments, its 1. Do not use equipments not mentioned
cause of injury by unexpected danger in this manual or not mentioned in the
situation. EcoRay’s printed matters.
Equipment Safety 2. This system must be installed by
authorized personnel by EcoRay.
3. Radiation Safety
4. Electrical Safety
WARNING: Never any adjust the X-ray This x-ray tube is produced suitable for
generator and the X-ray tube. If adjust regulations. Never make any changes inside
against this warning, it makes cause of of circuit or parts. If needed change,
Unauthorized X-ray
serious injury of death. contact the EcoRay Service center before
tube change
change or modify it.
Fire from near or other equipments, it 1. Install X-ray system in good air flow
makes a same damage as like this condition.
product is fired. 2. If there is a fire in the area or system in
on fire, to do as follows.
Electrical Fire Switch the circuit breaker off, stop supply
electric power.
Escape all people near the system.
Ask assistance.
Put the fire out with fire extinguisher.
Almost any liquid spilled into electronic 1. This system is not water proof, it is very
equipment automatically becomes a dander if liquid is spilled on it.
conductor of electricity. This can cause 2. Do not put liquid on system.
short circuit or fire by this short circuit. 3. It must be switched off immediately if
liquid is spilled on it.
4. Do not switch the system on until the
liquid has dried completely.
5. Do not splash or spray cleaning
solution on the system.
Cleaning Safety
6. Clean the system by unplugging the
power cord from the wall outlet and using
a damp cloth or sponge to wipe the
system clean.
7. When cleaning the inside of system,
using duster after remove the cover. This
cleaning must be done by an authorized
service engineer. (Because, it is very
danger when mishandle.)
Chapter 2
2.1 Introduction
2.1.1 An Overview
1. This HF-525 PLUS VET manual contains of operation, safety and specification of this X-ray system.
2. This X-ray system was developed by high frequency inverter circuit base.
3. This X-ray system was designed for suitable for diagnosing human being with radiation.
4. This X-ray system was designed to control radiation with digital method and calibrated radiation
before delivery could checkable by radiation meter.
5. This X-ray system was designed to do self diagnostic by electric circuit and program and
therefore, it’s easy to check the system’s condition and it indicates the type of trouble on the
display.
14. This X-ray system has quick rise time as 1mSec and could have good radiation efficiency.
15 This X-ray system could use low electric condition with PSU (Power Storage Unit – Battery type
– Optional item).
4. kV / mA / Time
Display window
12. Anatomical
5. Bucky switch
mode
Reset switch
Small Focal
Large Focal
2. Power Switch
(1) This switch controls electric power and if push ON switch, power supplied to system and if OFF
switch, disconnect power. Green lamp lighted when
power off
Power ON
Power OFF
switch
mA adjustment mAs / Time adjustment switch
switch
Table Bucky
7. Selection of APR
(1) APR is abbreviation of Anatomical Programming Radiographic and it could memory exposure
value to the microprocessor of PCB and could use it with easily.
(2) Memorized exposure value of microprocessor could change by operator’s need.
(3) APR function will be activated automatically by Powering ON of Console.
Memory switch
8. Generator status display lamp
(1) Display Overload, Standby, Ready and X-ray working condition and could check system condition
by lighted lamp status.
Ready status display lamp
Power status
Circuit breaker
External terminal
Voltage meter
220VAC
Back side of generator
(2) Could check generator electric condition by Standby lamp lighting condition. (Standby: Yellow,
Supplied power: Green)
(3) External devices will be activated by Console power ON.
Power ON switch
Normal: Yellow
1. Introduction of Power ON
(1) To operator the system, push the power ON switch of Console.
(2) After power on, system will diagnosing automatically during 15 seconds.
● After power ON, Standby lamp lighting with yellow and green.
● In the normal condition, 100 will be displayed at the mA window and could
use X-ray system.
● In abnormal condition, error will be displayed at the mA window before 100 and
2. Procedure of Power ON
(1) Push power ON switch of Console
(2) Standby lamp lighting with yellow and green if system is normal green lamp will be lighting.
(3) ”Set 0 r 1.00” will be displayed at the Console window and if system is normal condition,
“Set 100 r 1.00” display after 15 seconds later.
(4) “Set 100 r 1.00” displayed at the window, could use X-ray system.
(5) If abnormal condition, Error code will be displayed at the Console window and then could not
use X-ray system.
(2) After power ON, will be changed to the Radiography mode after
15 seconds later.
2. In abnormal condition
(3)when system
If error troubled.
displayed, turn ON/OFF repeatedly, but error code not
되며 제품을
appeared, 사용할
system must수 be
없음checked by authorized engineer.
4. Memory switch
method
(7) Select Body point
● Body point is consists of 21 anatomical mode and 3 types of exposure method.
● If push switch and then selected and lamp lighting.
1 2
● mAs / Time select: If push No. 3 switch and then selected and selected
switch will be lighting.
● mAs / Time adjustment: Increase → No. 1, Decrease → No. 2
1 2 3
Display window
Adjustment switches
3. Select Bucky
(1) Bucky will be activated automatically after Console power ON.
(2) Bucky signal was designed to operate only EcoRay’s bucky devices. If want to use another
company’s bucky devices, please contact authorized distributor or manufacturer and ask services.
2. Select Chamber
1. AEC ON/OFF
position
4. Density lamp
switch
6. Density switch
5. X-ray Exposure
(1) Beep sound and lamp is will be activated during X-ray exposure to announce X-ray radiation to
operator.
(2) X-ray switch is located at the bottom of console and indicating lamp is located at the upper of
console.
(3) X-ray switch is designed to be assembled it at the external and use it. (Hand Switch)
(4) 1 second later after pushing the “Ready” switch, in normal condition, Ready lamp lighting and
then could use X-ray exposure.
(5) If push the "Ready" switch, "Ready" lamp will be lighting with green.
(6) If push the "X-Ray" switch, beep sound will be made and “X-ray” lamp will be lighting with yellow
during X-ray exposure.
(7) X-ray exposure values are different by body point and could adjust it by operator’s need.
(8) Radiation will be exposed by only setting exposure value and could not over expose by circuit
about the X-ray tube capacity.
Ready / X-Ray
Indicating lamp
X-Ray switch
Ready switch
2.3 How to use Touch Monitor (Optional item: If selected DR touch monitor)
⑩ ⑪ ⑧
①
⑨
③
Tube Arm
X-ray Tube
Tube Stand Body
Collimator
Combined part
with table
No Description
① X-Ray filed adjustable knobs
② Momentaly push switch of lamp
③ Dial for film size
④ Back panel
⑤ Lock Knobe
⑥ Mounting flange for X-Ray tube
⑧ Center Cross marked plate
X-ray radiation is dangerous for both patient and operator and try to be exposed in
minimum exposure area as soon as possible.
Cover
Lock Switch
Ready / Exposure
switch
Do not move the Table Top in locked condition, it cause of system troubles.
2. Push the arrow direction 2. Released condition of Lock lever 3. Lock condition of Lock lever
Chapter 3
Wire size
Generator Line Voltage
7.5 m 15 m 30 m
40 kVA 208-240 VAC 35 mm2 70 mm2 120 mm2
Chapter 4
Maintenance
4. Maintenance
4.1 An Overview
4.1.1. The purpose of this Periodic Maintenance is to assure continued safe performance of the X-
Ray generator, to increase serviceability, to reduce the cost (save time, repairs, etc.) and to
assure the safety (personal risk).
4.1.2 The following checks and maintenance procedures, together with the suggested intervals, are
the manufacturer’s recommendation for the most effective periodic maintenance schedule for
this generator.
4.1.3 Service task here described must be performed exclusively by service personnel specifically
trained on medical X-Ray generator.
Never attempt to clean or handle any part of the X-Ray generator when it is turn on.
Switch off the generator main disconnect before cleaning or inspecting.
Keep the external protection cover in place all the time, only remove them to perform service
operations by qualified engineer.
4.3.1 Check that all electrical connections are firm and secure and that all cable clamps and strain
4.3.2 Also check that connectors do not have exposed wires and check cable Sheaths for wear and
fraying.
4.4.2 Check correct operation of console switches, displays and indicators by performing the
following test:
2. Push the “Power On” button of console and supply electric power to generator.
5. Set over X-ray exposure value and check “Over Load” message display or not.
4.5.2 If oil leakage was found, have to check oil level with oil fill plug.
4.5.3 Put silicon oil to high voltage cable plug after cleaning with dry cloth.
4.5.4 Check arc evidence of high voltage cable plug and Receptacle Socket or not.
4.6.2 If oil leakage was found, have to refill or exchange X-ray tube.
4.6.3 Put silicon oil to high voltage cable plug after cleaning with dry cloth.
4.6.4 Check arc evidence of high voltage cable plug and Receptacle Socket or not.
4.7.1 Make a periodic maintenance plan to use X-ray system safety and operator has a responsibility
about this.
4.7.3 Periodic maintenance must be done after installation and must be inspected by authorized
Working 2. Check X-ray exposed or not when push the “X-ray” button. ○
Environment 2. Check around the X-ray system that dangerous factors for
○
patient to be or not.
Chapter 5
Easy Installation
5. Installation
5.1 An Overview
5.1.1 This installation depends on generator’s type and configuration and must be installed by
instruction of this manual.
Only qualified (must trained by EcoRay) engineer could install this X-ray system!
Check voltage
Generator
To fix tube stand, assemble(tighten some screws) two parts behind the main body.
After laying down the tube stand, remove the plate on top of the tube stand.
After fix the tube stand on articulated joint, tighten the screws using a wrench.
To fix OP console case to column of tube stand, Please tighten the 4 screws using the wrench like as
below picture.
For your information, the board is located in under the emergency switch.
Please bring the OP console cable to a hole on bottom of tube stand as like below picture.
And, do not pull the OP console cable too hard.
After removing the cap, place the tube on top of the tube stand as below picture.
Collimator plate!
Insert H/V cable to X-ray tube after using silicon. Fix H/V Cable with X-ray
tube with H/V nut and fixing bolt.
+ +
- -
SC 1
SM 2
SS 3
T-TEMP 5
T-TEMP 6
GND E
COL 8
COL 9
COL 8
COL 9
Fix Tube Arm cover & back side of Tube Stand cove
For assemble the foot switch, Please located two parts as below picture.
After assemble two parts, Please assemble the main body following the below steps.
After connecting the all cables, please move the generator into the main body.
■ Rayence Detector
Connection
X-ray Enable Cable
1 2 3
4 5
6 7
9 8
<Consist of Detector>
① Power Supply ………. 1ea
② Ethernet Cable (Connect PC with Detector) …………. 1ea
③ AC Power Supply Cable …. 1ea
④ Ethernet Cable ( Connect Console with ⑤ X-ray Enable Cable) ………… 1ea
⑤ X-ray Enable Cable (After connection of ④번 Ethernet Cable then connect with Detector) … 1ea
⑥ Install CD ……………….. 1ea
⑦ Menual ………………… 1ea
- Product Certificate ………….. 1ea
⑧ Detector Angle …………………….. 2ea
⑨ Grid ……………. 1ea
< Connection>
- Power Supply -
<Consist of Computer>
1 2 3 4
<LCD Monitor>
2 3 4 5 6
1 2 3
4 5
5
- Remove the detector cover. See below the table 4 screws, 2 in front, 2 in the back part.
After, pull up the cover as shown in the photo.
Caution!
When handling the detector pay attention not to drop it or make any scratches on it!
Position of the
detector port
- Once the frame is adjusted to the detector. Install the detector into the detector tray.
- After inserting the detector into the detector tray, fix it as shown in the photo(front and back)
■ Detector cables
2. Console
3. Detector
1. P C Controller
Detector Controller
■ Connecting cables
Foot Switch
Detector
Monitor
(Touch)
Communication
Cable
Dongle Key
Monitor
Console
PC
PC Power
Detector LAN
PC Console
Detector
Console Controller
PC
1 2
4 3
- Install the touch screen monitor on the stand and fix it with screws as shown in the photo.
- Connect power and data cables to the monitor.
■ Installig grid
- Place the grid on the frame and adjust it to the center as show in the photo(yellow arrow)
- After centering the grid lock it using the lockers in 6 places.
Caution!
When handling the grid, do not scratch the surface!
Never bend or drop the grid!
1. Open installation file and click “SETUP.EXE” then click right side mouse. And run as administrator.
2. Click “Next”
6. Check USB touch device connected. Click “OK” if InstallShield Wizard then click “Next”
10. Check “Create a eGalaxTouch Utility shortcut on desktop” then click “Next”
11. Installing...
16. On the Display tap, Click NO 2 monitor, then check “Enable Multiple Monitors” and click “Apply”.
Click “Yes”
20. Click red mark during 2 ~ 3 seconds within 15 seconds. 4 points calibration was completed.
3. Input resolution
- If Touch monitor: 12" (1024 * 768), Main monitor: 22"(1920 * 1080)
- If main monitor display left, touch monitor display right side (Pos X: 1920 Pos Y: 0)
- If main monitor display right, touch monitor display left side (Pos X:-1024 Pos Y: 0)
- Input values are changeable by monitor size.
4. How to check moniotor status.
- On the main display, click right button of mouse. Check monitor resolution.
5. ① Main monitor ② Touch monitor (Select ② and drag it. Could check numeric. (1920,0)
6. ① Main monitor ② Touch monitor (Select ②, and drag it then could check nemeric.(-1024,0)
1. Click Ecoview
- Operator: Guest
- Password : None
5. Input patient information, Control mode will be displayed on the touch monitor if go to
expose mode.
- Could adjust kVp, mA, mAs, ms and patient body size (S, M, L, XL) on the touch monitor.
- If displayed data were different main monitor with touch monitor or not operated with
touch monitor, have to reset PC and check cable wiring condition again.
Caution!
Be careful of unpacking not to
damage cables in the PSU
Detailed image
Caution!
Be careful of unpacking not to
damage cables in the PSU
Warning!
Extremely pay attention not to get
an electric shock from batteries
when you handle it!!!!
■ Move X-ray table into X-ray room where you exactly want to locate.
■ Assemble under part of tube stand into the table
■ Take off PSU battery pack from packing and be fastened behind of the table by screws.
Screws
Detailed image
Caution!
Need more than two people to carry
out with careful handling!!
Warning!
Extremely pay attention not to get
Cable connection
Communication cable
Power cable
Connection 2P cable
with connector
Chapter 6
Technical Information
6. Technical Information
6.1 An Overview
This technical information contains that maintenance history and calibration data.
Hospital Name
Model X-Ray
Model
Generator
Tube
S/N. S/N.
1 1
2 2
3 3
4 4
6.4.1 Enter the following information after each data modification in this book or Periodic
Maintenance Service
Chapter 7
Calibration
7. Calibration
7.1 An Overview
This manual contains that adjustment method of exposure condition and how to calibrate exposure
data. Calibration data will be memorized at memory chip at the ECO-XR5-Mio Board with digital
value.
7.2 Introduction
7.2.1 This calibration tool is specified program supplied by EcoRay and it will be operated at the
personal computer.
7.2.2 Control exposure value (Ex. kV, mA, etc)
7.2.3 This calibration tool must be operated by qualified engineer by EcoRay.
7.2.4 This calibration tool was designed by many experience data of manufacturer and this tool will
be operated by specified program supplied manufacturer.
1. Required PC specification
3. Connector between USB to RS232 Converter and Generator - Made by EcoRay Co., Ltd.
After finishing installation, it should be set up within Com5 in the Changes COM Port.
※ If it is set up above Com6, PC Console could not operate correctly.
■ Product Features
-
- Support the RS232 Serial interface
-
- Support remote wake-up and power management
■ System Requirements
- IBM PC 486DX4-100 MHz CPU or higher or compatible system
- Available USB port
- CD-ROM drive
- Windows 2000
- Click Next to continue. Select the “Search for a suitable driver for my device (recommend)“
- Windows will detect the driver and Click Next to continue installation.
- Click Finish to continue and let Windows copy the needed files to your hard disk.
- You can to check the Device Manager and see the Prolific USB to Serial Com Port)
Under Windows Me, Windows 2000, Windows XP, please kindly following the steps as below;
(1) Please click on the Device Manager\ Devices by connection \ ports (COM & LPT). (Images 1)
(Images 1)
(2) Doubly click on “Prolific USB-to- serial Comm port (COM3)” .And it could go to„ properties.
(Images 2)
(Images 2)
(3) Please choose "Port Settings ", and then click on "Advanced". Please kindly choose the needed
com, and click on" ok”. (Images 3)
(Images 3)
Notice:
If the Modem could not link with the new port, and run smoothly when changed the port, please
kindly unload Modem, and re-install it.
When changing the port, the peripheral equipment could not shift to the new port, and run
smoothly. Under this situation, please kindly unload the equipment and install again.
- Turn off your Modem. Plug the DB 9 connector of the USB-Serial cable into your Modem. Turn on
the Modem afterwards.
- Turn on your computer and plug in the USB connector of the USB-Serial cable into the USB port.
- If you have already installed a Modem device before, click Start, Settings, and Modems. Open
Modems Properties, select the Modem, and click Properties. In General tab, please change the Port
value to USB to Serial Port (COM3).
- After finishing these processes, you can use USB to Serial Bridge Cable to connect RS232 Serial 56K
FAX Modem and execute the Dial-up function to connect Internet, and send the FAX with the FAX
software (i.e.: Microsoft FAX) properly.
RUN Cancel
RUN Cancel
RUN Cancel
Finish
- After press the button of Start and Execution on the operating display line of Windows in
turns, input the following order.
It is ok
In the case that “Regsvr32.exe” file is not on the PC, please copy “regsv32a.exe” of “Util”
Folder on the provided Program into “C:\Windows\system32\” and run it.
Operation
Part
Setup
Part
2. Operation part
2.1 kV,mA,Time Set adjust
When choosing Fast in the case of kV, it would be changed into 10kV step.
2.4 AEC
When choosing Fast in the case of kV, it would be changed into 10kV step.
3. Setup part
If it is possible, don’t be changed into Factory set mode in the field voluntarily.
MIN MAX
S.kV Establishing Min kV of Small Focus Establishing Max kV of Small Focus
L.kV Establishing Min kV of Large Focus Establishing Max kV of Large Focus
S.mA Establishing Max mA of Small Focus Establishing Max mA of Small Focus
L.mA Establishing Max mA of Large Focus Establishing Max mA of Large Focus
4. Start
Connect converter as below.
PC Generator
4.3 Power ON
- If you choose Power ON, the figure of red area is going up and it will be
Power ON.
Power ON screen
4.4 Exposure Key
- Ready and EXP are utilized by keyboard and acceptable keys are like below
Tube 1 Tube 2
Radiography Radiography Fluoroscopy
Ready Ctrl Ctrl
Z
Exposure X X
7.7.4 Calibration
1. Mode entrance
- Choose “Show Calibration Window” on the right side of screen.
3. Calibration start
3.1 Data change
- If you choose the wanted condition of Calibration, color of window is changed and data is
input into “ADJ”
- Data change: change 740 750 of ADJ HV to change Data
- Data input: after inputting the data, if you choose Enter, the changed data would be saved.
Fluoroscopy mA Calibration VR
1. Save
1.1 Setup save
- If choose “Save Config”, new “Save as” screen will be appeared.
- For the next after service, use product serial no with new file name.
- Saved calibration file contains as like a below calibration data (see next page)
2. Load
- Data load method has a “Setup” and “Calibration data”. It has some different.
Chapter 8
Trouble Shooting
8. Trouble Shooting
8.1 An Overview
8.1.1 This generator has a self-diagnosing function and display error code on the console display if
system has an error.
8.2 Error type
8.2.1 Error type
No Error type Cause
1. Charging to Main Capacity is not normal.
Charge Circuit
1 2. Input voltage is out of boundary. (Too high or too low)
problem
(Standard input voltage: Single phase, 220VAC ±10%)
Tube stator 1. Cable connection to Tube stator is wrong.
2
Circuit problem 2. ECO-XR5-MiO in Stator circuit is out of order.
1. Tube Filament circuit connection is wrong.
Tube Filament
3 2. Filament circuit is out of order.
Circuit problem
3. Tube Filament is short.
HV-DRV-OLP (ECO-XR5-FnD) signal or
Over current to IGBT Driver IC may occur
circuit is wrong.
the error.
(See circuit diagram)
IGBT-OCP (ECO-XR5-FnD) signal or Over current to IGBT Module may occur the
Generator
circuit is wrong. (See circuit diagram) error
4 Protection Circuit
DRV-PWR-UVP (ECO-XR5-FnD) signal
problem Voltage from the IGBT Driver IC is lower
or circuit is wrong.
than standard.
(See circuit diagram)
FIL-OCP (ECO-XR5-FnD) signal or
Over current on the Tube Filament
circuit is wrong. (See circuit diagram)
Communication 1. Wrong connection of communication cable.
5
program 2. Communication circuit in the Main board or Console board is out of order.
1. Wrong connection of communication cable.
6 AEC problem 2. Wrong cable connection between AEC Board and Ion Chamber.
3. AEC Board or Ion Chamber is out of order.
Irregular Exposure 1. Actual exposure time is shorter than set time.
7
Time problem 2. Exposure switch is bad condition.
12. Error type 12: Displayed E r r roto at kVp, mA, Time section
14. Error type 14: Displayed E r r FILA at kVp, mA, Time section
15. Error type 15: Displayed E r r 232 at kVp, mA, Time section
16. Error type 16: Displayed E r r AEC at kVp, mA, Time section
17. Error type 17: Displayed E r r SEC at kVp, mA, Time section
18. Error type 18: Displayed E r r Pro----O at kVp, mA, Time section
19. Error type 19: Displayed E r r Pro--I-- at kVp, mA, Time section
20. Error type 20: Displayed E r r Pro---P- at kVp, mA, Time section
21. Error type 21: Displayed E r r Pro----F at kVp, mA, Time section
22. Error type 22: Displayed E r r Inte at kVp, mA, Time section
1. Inter Lock signal trouble and circuit trouble
2. Inter Lock is Door Open, Tube Temperature, Bucky start signal, Emergency Stop switch.
Connect (P805)
Fuse (F801)
Connect (P803)
Connect (P802)
Connect (P804)
Relay (RLY802)
L.Volt(VR801),C.Volt(VR802)
Test Point (TP802,803,804,805)
2. Check Procedures 1
(1) Board Name: ECO-XR5-CnD
(2) Functions: This board controls charge/discharge of Main Condenser.
(3) Followings are the check procedures for ECO-XR5-CnD board when charge/discharge trouble.
TP805 (12V).
Variable Resistor(VR701)
Diode(D707,709)
Test Point(TP106)
Connector(P706)
Test Point(TP105)
Variable Resistor(VR901)
Test Point(TP109)
Connector(P901)
LED(D906,911,912 Fuse(F901)
)
Variable Resistor(VR902)
3. Check procedure
(1) Board Name: ECO-XR5-Mio Board
(2) Functions: Controlling generator including Tube Stator movement.
(3) Under the tube stator trouble, check tube stator wiring condition and then check ECO-XR5-
Mio Board and check points are as follows;
Functioning well
ECO-XR5-FnD
3. Check procedure
(1) Board Name: ECO-XR5-FnDBoard
(2) Functions: Controlling Tube Filament
(3) Under Tube Filament circuit trouble, check high voltage and wiring condition of circuit and
then check ECO-XR5-FnD board.
Turn on the Main power and 1. Check Connector (P501) and measure the
voltage on it; 220VAC is normal
then turn on the Console Power
2. Check Fuse (F501).
3. Fix the Relay (RL501) position and check
LED (D516, 517, 518)
its functioning. Normal: Relay working.
ECO-XR5-FnD (Driver Board) Not lighting
4. Measure the voltage between Heat sink
Normal: All LED lighted
LED (D516, 517, 518) (U406) and Test point (M5V) Normal:
with green
점등상태확인 5VDC ±0.1V
5. Measure the voltage between Heat sink
정상: LED모두 녹색점등
(U406) and KA324 (U518 pin 4); 12VDC
Set exposure values
±0.1V is normal
6. Measure the voltage between Heat sink
(U406) and KA324 (U518 pin 11); -12VDC
Error when Ready ±0.1V is normal
Push the “Ready” button
7. In case above voltages are not good,
check input PWR of ECO-XR5-MiO (Main
Board).
8. Fix the Connector (P403) well.
8.3.4 IGBT Device signal and Protection circuit trouble (Check ECO-XR5-FnD Board)
1. Consist of part
ECO-XR5-FnD
ECO-XR5-FnD
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check high voltage cable, filament short condition of X-ray tube, wiring condition of circuit
when IGBT Device signal and protection circuit trouble and then check ECO-XR5-FnD Board
condition as bellows.
ECO-XR5-FnD
Displayed on console
ECO-XR5-FnD ECO-XR5-FnD
VR502
engineers.
Push X-ray button 5. In case IGBT_OCP circuit is out of order,
ECO-XR5-FnD ECO-XR5-FnD
VR502
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and check IGBT_OCP trouble when IGBT
Device signal and protection circuit troubles. And check procedures are as follows.
Displayed on console
8.3.8 DRV _ PWR _ UVP trouble or Error Pro - - P – error (ECO-XR5-FnD Board)
1. Consist of PCB
ECO-XR5-FnD
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and the check IGBT_OCP trouble when IGBT
Device signal and protection circuit trouble. And check procedures are as follows.
lamp ON)
8.3.9 DRV _ PWR _ UVP trouble or Error Pro --P- error (ECO-XR5-FnD Board)
1. Consist of PCB.
ECO-XR5-FnD
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and then check IGBT _ OCP troubles when
IGBT Device signal and protection circuit troubles. And check procedures are as follows.
Set exposure value 9. Check voltage between Heat sink (U504) and
Push X-ray button 13. Check voltage between Heat sink (U506) and
ECO-XR5-FnD
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and then check FIL _ OCP trouble when IGBT
Device signals and protection circuit trouble. And check procedures are as follows.
8.3.11 Communication error between generator and console or Err 232 error
(ECO-XR5-Mio Board & ECO-XR5-CON Board)
1. Consist of PCB
ECO-XR5-MiO
2. Check procedure
(1) Board Name: ECO-XR5-Mio Boar and ECO-XR5-CON Board
(2) Functions: Communication transfer between generator and console.
(3) Check communication cable condition when communication problem between console and
ECO-XR5-MiO Board. And check procedures are as follows.
2. Check procedure
(1) Check Ion Chamber connector condition.
(2) Chamber position and Bucky position must be same. (Same position)
(3) Check Number flickering or not on the Time display of console after X-ray exposure with
‘beep’ sound (3 times).
(Normal: Changed normal display condition after Number flickering on the Time display)
3. Consist of part 1
4. Check procedure 1
(1) Board Name: ECO-XR5-CON Board
(2) Functions: Controlling Ion Chamber
(3) Check communication cable condition and then check circuit when communication troubles
occurred between console and ECO-XR5-MiO Board and Ion Chamber. And check procedures
are as follows.
X-ray exposure
5. Consist of part 2
6. Check procedure 2
(1) Board Name: ECO-AEC Board
(2) Functions: Controlling Ion Chamber
(3) Check communication condition when communication trouble occurred between
input/output signal of Ion Chamber and circuit error. And check procedures are as follows.
Select AEC function 7. Check voltage U91 (Heat sink. GND) with TP91
Normal: 12VDC ±0.5V
X-ray exposure 10. Check voltage U91 (Heat sink. GND) with TP92
Normal: 5VDC ±0.3V
7. Consist of part 3
8. Check procedure 3
(1) Board Name: ECO-XR5-MiO Board
(2) Functions: Controlling Ion Chamber
1. Ready ON
2. X-ray exposure ON
3. Ion chamber reset
Test point (Q4 collector) &
U91 (Heat sink) – GND
4. X-ray ON
After ion chamber reset,
automatic exposure after
230mSec.
5. X-ray exposure signal
TP51 (AEC input) & U91
(Heat sink) –GND
6. Exposure stop
TP32 (stop) & U91 (Heat
sink) – GND
7. Exposure stop
Test point (stop) & U91
(Heat sink) - GND
Chapter 9
Circuit Diagram
9. Circuit Diagram
9.1 Block Diagram
9.6 GC board
9.10 Collimator