Professional Documents
Culture Documents
Capillary2 - Capillary2 FP - Dilution Issue
Capillary2 - Capillary2 FP - Dilution Issue
Capillary2 - Capillary2 FP - Dilution Issue
Table of contents
Introduction ................................................................................................................................................................... 2
Note :
- insufficient volume of serum or blood will generate a low O.D
- insufficient volume of Hemolysing Solution will generate a high O.D
To highlight a dilution issue, please run a SEBIA rack with a segment manually diluted :
Volume
aspirated
Serum / Blood Buffer Diluant/Hemolysing Water
Technique
Protein(e) 6 40 µL 160 µL not used not used
C.D.T 40 µL not used 40 µL 120 µL
Hemoglobin(e) 18 µL not used 90 µL not used
HbA1c 18 µL not used 90 µL not used
- If the issue is fixed when running the SEBIA rack, then we are facing a dilution problem.
It could be related to the mechanical play of the syringe (see Appendix2).
- If the issue is not fixed, we are facing another issue (capillaries, injection vacuum, optics, …etc).
Running samples :
Make sure there is a good vacuum inside the wash station (no dripping underneath).
In case of dripping, it could be a leak from the waste container, a dirty air filter, a weak pump
(make the pump PM), or even an issue with a liquid valve.
Verify that the needle is correctly positioned inside the wash station
Verify that the needle is correctly positioned upon the sample tubes and upon the dilution
segment. Refer to the mechanical calibrations (hereafter).
Expected range : 400 < Threshold < 800 4000 < Offset < 8000
Exceptionally when the detection is too sensitive (probe sensing something in the air before reaching
the serum), we can make a DCN calibration with 200 µl of distilled water instead of 100 µl into the
Eppendorf tube. It will decrease the sensitivity of the needle.
Heigh of the Needle Calibration : preferably, it should be done using the Protein(e) 6 technic.
Now, an empty IT segment is provided with the instruments. Fill the double comb with 380 µL of
distilled water and then place this segment the Maintenance rack (#100).
Launch the maintenance rack and select “Calibration of the sample probe”.
Make sure the extremity of the needle is correctly positioned inside the wash station.
From the Ready mode, adjust the sample probe extremity in the middle of the wash station with the
help of the 2 Allen screws on the top of the sample probe.
Verify the correct electrical continuity of the yellow wire (no bad contact on the top of the needle)
connector J5 (pin 1)
Note :
the external wall of the needle is Isolated.
An internal bad contact of the shielded cable is hardly visible, so it can be good to replace it.
Check all the grounding connections inside the instrument. Make sure the earth is good on the wall
socket, plug the power cord to another AC wall socket to eliminate any potential issue on this side.
The opto sensor (bubble sensor) located behind the needle assembly is not used.
The wash station of the needle must be clean and not obstructed. Otherwise, there is a bad
cleaning of the needle and it could remains wet when using the DCN sensitivity.
The cleaning is correct if there is a good aspiration (vacuum) inside the wash station.
The waste tubing line must never be pinched or obstructed. All along the way between the wash
station and the waste container
It’s the exact same principle on Capillarys2 Flex Piercing instruments, except that there are 2 liquid
valves on the wash station : EV4 and EV6.
The 4-way valve EV18 is replaced by a 6-way valve (or 8-way valve) EV18.
The liquid valve EV7 is replaced by a pinch valve EV7.
2016/01/29 Page 7/15
Make also sure that there is no air leaks on the waste container.
Note that the DCN PC board (P/N 10046223) is almost never responsible of a detection issue.
This cycle is not applicable on the releases 6.50, 6.51 and 8.50 (the graphs obtained are not usable -
significant). It can be performed on the releases 6.12, 8.51 and 8.60 and next.
Please note that when running Phoresis Core (release 8.60), the priming of the tubing line (after a
change of technic) is now improved, to avoid missing the first dilutions.
Using the test cycle C16 we can visualize the DCN level detection of the needle during a dilution.
When the needle touches the liquid, the signal detected by the needle is modified.
The more liquid there is on the tube, the more the difference of level (Threshold) will be important.
In our case (please refer to the Appendix1), the quantity of distilled water in the tubes is descending.
So, the threshold of the signal returned by the needle is more and more weak.
Check the correct grounding connections and the welding spots on the shielded cable
-----------------------------------------------------------------------------------------------------------
-----------------------------------------------------------------------------------------------------------
-----------------------------------------------------------------------------------------------------------
Using the test cycle C16 we can visualize the DCN level detection of the needle during a dilution
process. When the needle touches the liquid (serum or H2O), the signal detected by the needle is
modified. The more liquid there is on the tube, the more the difference of level will be.
In our case (attached pictures), the quantity of distilled water in the tubes is descending.
So, the threshold observed is increasingly weak.
The mechanical play indicates the correct functioning of the syringe (for the dilution ratio).
This value is saved on the EEPROM, at the address 400 (Length = 01).
The mechanical play shall be less than 64 (Hexa value). So 100 motor steps (decimal value).
If this value is forced (so different from 0), the measured value (first priming of the day) is not taken
into account. Only the forced value is applied.
Example : dilution issue, visually confirmed with different levels of dilution inside the dilution
segment. In such a case, read the measured mechanical play (address 400), and then force it
with P298 but enter a forced value lower than the measured one (10 steps less should be enough).
Important :
with time the mechanical play will increase, so the replacement of the piston will be necessary.
------------------------------------------
0 0x162=0
4 4x161 = 64
F Fx160 = 15x160 = 15x1 = 15 * LSB : Least Significant Bit - MSB : Most Significant Bit
------------------------------------------
Calculation :
0
16 =1
1
16 =16x1=16
2
16 =16x16=256
The draining line goes from the wash station of the needle to the external waste container.
Reminder : on the couplings of the waste container, any small air leak will generate a bad draining and
therefore a bad cleaning of the sample probe. Consequently the dilution cycle is disturbed.
In some locations (all along this draining line), a tubing could gets pinched, or obstructed.
It’s possible to modify the routing of the tubing lines, in order to prevent any pinching or plugging :
--------------------------
POSSIBLE MODIFICATION
OR
--------------------------