Professional Documents
Culture Documents
Cynosure Starlux 500 Palomar Technical Service Manual
Cynosure Starlux 500 Palomar Technical Service Manual
StarLux500™
TECHNICAL
SERVICE
MANUAL
Document Part Number 5500-0000 Revision X1
Manufactured by:
Palomar Medical
82 Cambridge Street
Burlington MA 01803 USA
Toll Free 800-725-6627
Tel. 781-993-2315
Fax 781-993-2340
Email: customer_service@palomarmedical.com
www.palomarmedical.com
Chapter 1 Safety
1.1 Safety Scope 1-2
1.2 Training
Extent of Service:
Palomar assumes no liability whatsoever for any damage or injury resulting from
an application of a product which is not in strict accordance with the instructions
provided with the product. Palomar also assumes no liability for any damage or
injury arising as a result of user error or mistake, including but not limited to
injury arising from user’s lack of qualification to administer the treatment in the
individual circumstances, or as a result of errors or mistakes committed by such
user who would otherwise be considered as having the necessary professional
skills to apply such treatment.
Read the Installation and Operation Instructions thoroughly before connecting the
base system and chiller unit to the main power connection prior to use. The
system must be set, regulated and used in accordance with the instructions for use
listed in the handpiece appendices. Failure to observe usual safety precautions
may present a risk of hazardous exposure to laser or pulse light radiation.
All persons operating the StarLux500 System or in the vicinity of the system must be
aware of the potential hazards of the system output and should take the proper
safeguards as described in the Operator’s Manual. Be certain all personnel carefully
review the safety precautions pertaining to the Palomar StarLux500 System. Only
authorized individuals with appropriate safety training and knowledge should operate,
assist in the operation of, or provide maintenance/service to the system. One person
should be assigned to be responsible for system controls during the procedure. High
voltages are present inside the StarLux500 System, and any pulsed-light or laser
emitting device can cause injury if used improperly. Personnel who work with
pulsed light and laser sources should always be aware of the possible dangers and
should take the proper safeguards as described in the operator’s manual.
• Delivering excessive energy to the treatment site may cause thermal damage to the skin,
resulting in superficial damage to the skin, crusting, skin redness, hypopigmentation or
hyperpigmentation.
• Consumer safety is assured by a well-trained staff.
• Consumer education is also important and information about the nature of the treatment
should be provided.
• Never look directly at the light coming from the distal end of the handpiece as this may
cause eye injury.
• Never direct the laser or pulsed light at anything other than the targeted area.
• Never point the handpiece so that it discharges into free space. Make sure it is pointed
at, and in full contact with, the treatment site during treatment or when performing test
spots.
Warning
Do not submerse the handpiece in any solution. Submersing the handpiece may cause a
hazardous condition for the user and possibly the consumer. Submersing may also cause the
system to malfunction. Apply disinfecting or cleaning solution to wipes or gauze pads to be
used on the handpiece. If the handpiece becomes submersed in fluid or leaks any cooling
water, discontinue use of the handpiece immediately and call Palomar for replacement.
Do not use the handpiece if it is cracked, the sapphire lens missing, cracked, or burned.
• The electrical safety hazard of any laser and pulsed light system cannot be ignored or taken
lightly. Although the Palomar StarLux500 System has many safety features in it’s design,
certain precautions must be observed.
• It is possible for the high voltage components to retain a charge even after the system
has been turned off. Therefore, no part of the exterior housing of the main system or
handpiece should be removed except by trained and authorized technicians.
• This may expose personnel to dangerous high voltages and potentially dangerous light
radiation. Only a qualified Palomar Service Representative must perform any service
requiring access to the interior of the instrument. To contact a Service Representative, dial
toll free 1-800-PALOMAR or 1-781-993-2315.
• Customer Service Email: customer_service@palomarmedical.com
When the Palomar StarLux500 System is used, the following precautions should be
taken:
• Anesthetics administered must be approved as non-flammable.
• Ensure that the system and ancillary equipment are working properly before beginning
a procedure.
• Never direct the light output at anything other than the intended treatment site,
especially at reflective objects such as metal surgical instruments.
• Be certain the footswitch is clean and working properly. Place the footswitch where it
will not be mistaken for another piece of equipment.
• Exercise particular care in the use of oxygen. The presence of oxygen can accelerate the
combustion of any flammable material.
• Avoid using combustible materials, such as gauze and drapes, in the treatment area. When
they are required, these materials may be made fire-retardant by keeping them moist with
water or saline. Clothing should be kept well away from the area of treatment.
• Know the location and operation of the nearest fire extinguisher.
• Always turn off the system and remove the key when ever possible. This will prevent
accidental light emissions.
Chapter 2 Overview
2.1 Introduction 2-2
2.1 Introduction
The Palomar StarLux™500 Pulsed Light and Laser System is designed, by attaching
multiple treatment handpieces, to deliver pulses of broadband incoherent light or laser light
to a predetermined target site. The Palomar StarLux500 System is specially developed for
permanent hair reduction, pigmented and vascular lesion treatment, leg vein treatment, acne,
fractional soft tissue coagulation, fractional skin tightening through soft tissue coagulation,
and other clinical applications.
The Palomar StarLux500 System is user friendly, allowing the treatment provider to select
output fluence (J/cm2) and pulsewidth (ms). This enables the user to safely perform
treatments on skin types I-VI as indicated in Starlux Treatment Guide.
The system is housed in a protective cabinet with includes the connection to an external
cooling unit. Delivery of the energy to the treatment site is through a handpiece that contains a
cooled sapphire window.
Light is directly administered to the treatment area by placing the handpiece window in
contact with the skin and pressing an attached footswitch in conjunction with pressing the
handpiece trigger button. There is also an option to deliver light pulses by pressing the
footswitch only.
The “Cooltip” supplies contact cooling to the skin during treatments, this minimizes the
potential for skin damage and allows for higher fluence during treatment.
Item Description
System Type Class I Equipment, Type BF, applied part (CE Approval)
Class II Medical Device (USA)
Pulsed Light Class IIb (EC Community)
Class IV Laser (Lux1064TM and Lux1540 TM Fractional Handpiece only-USA)
Dimensions Base Unit: 19” (48 cm) L x 11” (29 cm) H x 14” (36 cm) D
Cooling Unit: 10 “ (27 cm) L x 15” (38 cm) H x 16” (41 cm) D
500 Cart: 22 “ (58 cm) L x 30 “ (78 cm) H x 22 “ (58 cm) D
1. External Cooling Unit, Provides cooling for the CoolTip in the handpiece, monitors
the water flow and temperature of the lamp and Cooltip loops.
2. TEC module (3 Dual Chamber TEC’s) within the cooling unit generates cooling for
the Cooltip.
3. Display Module, Contains the Main Processor, interconnect board and display screen.
4. Power Supply Module, provides the DC voltage to operate the System, Lamp
Controller and Thermocontroller boards.
5. Handpiece, provides the light to the target tissue and contact cooling. All calibration
data stored in handpiece.
2.3.2.1 A power cord is connected to both the base unit and external cooling unit, a handpiece
is connected and in the holster, and the coolant reservoir is full. Turn the main power switch on
the cooling unit and then the base unit. AC voltage is then applied to the Power Factor
Controller Board. This converts the base unit input voltage to supply 390VDC to the capacitor
charging circuit and Thermocontroller board. This board supplies 12VDC and 6VDC to all
boards and the display module. A “Menu/Next” screen will be displayed. (screen 1)
Screen 1 Screen 2
The system will read the EEprom information from the handpiece, the software will
determine the type of handpiece installed and then load the proper parameters for the
handpiece. (screen 2) If a handpiece communication error is detected during this time, it
will be displayed on the screen. Otherwise, the system will proceed to key switch screen.
2.3.2.2 With the key switch in the off position, the ‘turn key’ symbol will be displayed on
the touch screen panel. Note that the colored band at the top of the screen is the same color
as the handpiece installed. (screen 3).
Screen 3 Screen 4
2.3.2.3 When the key is turned to the “ON” (clockwise) position, the microprocessor will
check all operating systems for voltages or other faults and the cooling module power will
then switch on. As the system continues to power up, checks for water flow and
temperature will also take place. The lamp will start simmering. (screen 4)
2.3.2.4 If the system does not have any immediate faults, the “Cool Down” screen with
the hourglass symbol will be displayed.(screen 5) The cooling module is now fully
operational and if there are any faults not recognized during the system startup test, they
will be displayed now. The coolant reservoir holds about 350cc of cooling fluid. It is 2
chamber tank separated for the Cooltip loop and the Lamp loop.
Lamp Loop Cooling: If the lamp loop temperature is 35˚C or higher, the fans run at
100% speed. If the lamp loop is under 30˚C, then the fans will run at 80% speed. Two
Thermisters (Lamp & Cooltip) monitor the water temperature and must be within 4˚C of
each other. Lamp Water flow is >1.9 LPM. The Pressure switch is rated at 65 PSI and
will open if the pressure rises above this point.
CoolTip Cooling: A TEC module provides the cooltip cooling. Cooltip Water flow is
>0.45 LPM. The Pressure switch is rated at 65 PSI with a bypass circuit to prevent damage
if water flow is blocked. The “hourglass” may be displayed above the thermometer symbol
on the left side of the display screen. During this mode, the cooltip temperature is being
monitored while the TEC module is cooling the water and the hourglass symbol will stay
on until the water temperature is 15˚C (59˚F). This should occur within 7 minutes. If the
room temperature is above 22˚C (72˚F) or the system has been operating for a long period
of time and then shut off, it may take longer for the cooltip temperature to reach 15˚C. The
system will not go into the “Ready” mode until the hourglass is gone from the screen.
(screen 5) The Starlux must be in Ready mode in order for a handpiece to pulse.
2.3.2.5 From the “Ready” screen, the user can select pulse width (ms), fluence (J/cm2), and
repetition rate (Hz) up to the maximum parameters for the handpiece. For questions about
handpiece or system information (ie. hp shot count or system software version), select the
question mark button and choose Handpiece or System. (screen 6 and 7) Select ‘Back’
button to return to your previous screen or to exit back to the Ready screen.
2.3.2.6 For complete system operation please see Chapter 4 of StarLux500 Operator’s Manual.
Chapter 3 Troubleshooting
3.1 Fault Codes 3-2
3.1.1 Non Error Code Faults 3-14
1. Replace handpiece.
2. Check/replace connection
1. Restart system. plate.
Internal Voltage 2. Try another handpiece. 3. Replace power supply module.
001- 134 loaded down. 3. Call Palomar service. 4. Replace display module.
1. Monitor PFC temperature
sensors. If temperature is
1. Let the system cool down. unstable or operating
PFC temperature 2. Restart the system. abnormally, replace power
001- 135 exceeds limits. 3. Call Palomar service. supply module.
001- 136 Not used. Not used. Not used.
001- 137 Not used. Not used. Not used.
1. Using System Debug
Software, check that the
handpiece temperature is
o o
1. Cool down system for 15 between 10 & 50 C and
o o
minutes. remains between 10 & 50 C
2. Restart system. for 20 minutes.
3. Try another handpiece. 2. If temperature is wrong, the
1064 HP board 4. Call Palomar service. handpiece board needs to be
001- 138 temperature. replaced.
1. Verify that software CF
card is the latest revision.
2. If system is over s/n 15-
1873, the user may be
trying to install a
handpiece that is not 1. Install upgrades required for
allowed to be operated on the operation with the new
the system. handpiece type.
3. Call Palomar service. 2. Contact Palomar service.
001- 139 HP not supported.
1. Restart system.
2. Try another handpiece.
3. If problem re-occurs with 1. Restart system.
IR lamp controller did IR handpiece, call Palomar 2. If problem, re-occurs with IR
not receive a signal service. handpiece, replace handpiece.
001- 140 from the handpiece.
1. Restart system. 1. Restart system.
2. Try another handpiece. 2. Try another handpiece.
3. Call Palomar service. 3. Contact Palomar service.
001- 141 IR handpiece data.
1. Let system cool down for 5
minutes.
1. Verify the IR handpiece 2. Restart system.
internal cooling fan is 3. Verify the IR handpiece
operating. internal cooling fan is
2. Restart system. operating.
IR handpiece over 3. Call Palomar service. 4. Return IR handpiece for repair.
001- 142 temperature.
L40(150)A-SH-V2
Min Max
5 2 7.3 8.9
20 5 18.3 22.3
60 6 21.9 26.8
100 11 40.2 49.1
1. Select the lowest fluence available for the pulse width selected.
2. Pulse the system with the handpiece toward the photodetector and record the pulse
width. Compare to chart below.
SETTING RANGE
10 ms 9 – 11 ms
20 ms 18 – 22 ms
100 ms 90 – 110 ms
To enter service mode, attach a Starlux500 LuxR test handpiece, select IPL 500 HP and
press System Controller Utility.
Step 4
Step 6 & 7
Step 5
Step 2
Step 3
4. Set Fan Speed % to 80 using the arrow buttons on left and then select Fan Set to run
fans.
5. Select Polling to ON. This will display the active Lamp and HP temperature
measurements as detected by the Starlux internal thermistors.
6. Set TEC Temp to 20o C using arrow buttons on left and let the system run for 1 - 2
minutes.
7. Set TEC Temp to 2o C and measure the time (in minutes), it takes to reach 5o C.
This time interval should be less than 5 minutes.
Step 1
Step 2
Step 3
Step 2
Step 1
Step 4 Step 5
Step 1
1. Press the System Error Log Utility button from the Main User Screen.
Step 2
2. Then select Display Error Log to display the error log data on the screen.
Please Note: After installing a new Display, you may see a default value of “Starlux” and
“P-0002” printed in the System Type and System Serial Number text fields respectively.
The Manufacture Date and Warranty Date may default to the current date. This is normal
until you enter the system information.
Chapter 5 Disassembly
5.1 Removing the Display 5-2
1. Turn system power off and disconnect power cord, cooling unit communication cable and
hose assembly from Starlux500.
2. Remove all four (10-32) screws that secure the Bezel Assy to the chassis. (see fig. 1 & 2)
Figure 3. (Bezel Assy lifted) Figure 4. (Keyswitch wiring) Figure 5. (Dock Sensor wires)
4. Remove the 2 screws that hold the Rear Vent Assy (fig. 6) then remove all chassis screws (fig. 7).
Figure 6. (Rear Vent Assy removal) Figure 7. (Top Lid mounting screws)
5. Lift the entire front panel forward to access the interior of the system (figure 8) and
disconnect all cables from the rear of the display. (figure 9)
6. Remove 3 phillips screws on the left securing the display module, then remove the 2 phillips
screws on the right and lift the display module out. (Figure 10 & 11)
2. Disconnect the High Voltage cables from the Lamp Controller Board. (Fig. 13) Put a jumper
across the +400V and -400V wires. (Fig. 14 ) CAUTION: Failure to short these wires prior to
removing cables, will cause the High Voltage capacitors to “charge in air”. This will cause a
potentially dangerous voltage on the capacitors!
3. Disconnect the two pin connector on the Charger Board that connects to the capacitors.
5. Disconnect the 4 pin cable from the Voltage Regulator Board located on the side of the power
supply.
HV Cables
Jumper Wires
Figure 14. (Capacitor Jumpers) Figure 15. (Power Supply mounting screws)
Figure 16. (removing Power Supply) Figure 17. (Power Supply removed)
6. Remove the 3 rear mounting screws of the Power Supply indicated by the black circle. (Fig. 15)
7. Pull the Power Supply from the rear and out of the Starlux rear chassis (Fig.16 and 17)
2. Remove the Power Supply as instructed in steps 1 thru 8 in Section 5.2. Disconnect the High
Voltage cables from the Lamp Controller Board. (Fig. 13) Put a jumper across the +400V and
-400V wires. (Fig. 14 ) CAUTION: Failure to short these wires prior to removing cables, will
cause the High Voltage capacitors to “charge in air”. This will cause a potentially dangerous
voltage on the capacitors!
3. Loosen and remove the two Phillips screws holding the front of the capacitor bracket sheet
metal to the chassis. (Figure 18)
4. Check the opposite end of the capacitor bracket. If there are screws holding the bracket
down, loosen them. If the sheetmetal bracket has tabs that slide under the main chassis,
there are no screws to loosen.
5. Pull up on the front edge of the capacitor bracket and gently pull the assembly forward and
up to remove it.(Figure 19) Take care not to damage the speaker assembly mounted under
the capacitors when removing the assembly.
Figure 18. (Capacitor Bank mounting screws) Figure 19. (Capacitor Bank removed)
HP
HP Pressure Switch
Secondary Heat Exchanger
Aquatec Pump