Professional Documents
Culture Documents
Present As I
Present As I
Present As I
USER TRAINING
Location Germany
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
■ CARAT II pro may be used for life-sustaining ventilation and, when using a double-tube system, provides
continuous respiratory support and ventilation of patients without spontaneous respiration.
■ When using single-tube system, CARAT II pro may only be used for non-life sustaining ventilation and
provides intermittent respiratory support and ventilation of patients who demonstrate sufficient
spontaneous breathing.
■ Adults and children with a tidal volume of 50 ml and or more.
■ Can be used in home care, clinic and professional healthcare facilities.
■ CARAT pro has both pressure- and volume-controlled ventilation modes. Ventilation can be invasive (e.g.
tracheostoma) or non-invasive (mask)
■ CARAT II pro is technical able to run with a single line circuit with exhalation valve or with a double line
circuit.
■ Device can be connected to a low-pressure
source of oxygen, CARAT pro can be
combined with an external
humidifier.
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
■ Irrespective of the indications listed here, the device must always be used in accordance with the
physician's individual diagnosis.
● Severe hypotension
● Severe epistaxis
● Pneumothorax or pneumomediastinum
● Pneumoencephalus
● Cranial trauma
● Acute inflammation of the paranasal sinuses, middle ear infection or a perforated ear drum
● Aspiration hazard
The following undesired side effects may occur in connection with ventilation:
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
ON/OFF-Key
Function Action
Alarm-key
Function Action
Turn
Set parameters
Press briefly
Confirm parameter selection
Press briefly
Confirm modified parameter value
Open adjustment window for graphs and loops in the Press briefly
monitoring screen
Safe-key
Function Action
Home-key
Function Action
Escape-Key
Function Action
Battery-LED
Power-LED
Alarm-LED
Position Action
DC connection
The power supply plug is connected here.
O2 connection
During oxygen input the oxygen source is connected here. Use
the supplied oxygen connection adaptor for this purpose.
SD card slot
An SD card here can be inserted here
Data Storage
Statistics Yes No
O2 output
Oxygen monitoring:
Exit for oxygen from the oxygen valve when ventilation has
been turned off.
Pressure monitoring:
Excess oxygen pressure is exhausted here during oxygen
therapy.
Pressure ventilation:
Opens when the pressure is more than 1 hPa above the set
ventilation pressure setting.
Volume ventilation:
The valve also opens when the measured breathing volume is
more than 10% higher than the set value.
Filter cassette
The filter cassette contains the two air filters
(coarse and fine filter).
For information on how to replace and clean the filter.
Handle (pull-out)
The handle may be pulled out for device transport.
Bottom flap
Valve membrane (expiration) is located under the bottom flap.
PC connected via the Micro USB port Charge state of internal battery
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
Power supply
■ The ventilator may be supplied by three different power sources.
● Mains connection, via switched-mode power supply unit
● Internal battery
● External battery pack uni BASE and uni PLUS (optional accessory)
Mains operation
■ The device can remain permanently on the mains without any harm.
■ Internal battery is charged as soon as the power supply is plugged in.
■ Device boots up and performs the following hardware tests:
● Testing all 3 LEDs (white) during start-up.
● Display of detected error, every error must be confirmed by pressing the MFK
● Error message "System locked": Serious fault detected, device can not be put into service, service required.
● with internal battery & uni BASE & uni PLUS: approx. 20h
Note:
For further information on connecting and
handling the device, please refer to the
AKKUPACK uni BASE user’s manual.
E adaptor for bac. filter F pressure measuring tube G valve control tube H air outlet
Note:
When using bacterial filters
(recommended) follow the
manufacturer's instructions!
E adaptor for bac. filter F inspiration tube G valve control tube H external humidifier
E bacterial filter F adaptor for bac. filter G adaptor for bac. filter H bacterial filter
E B
D
E closed connector for proximal pressure measurement tube F adaptor with proximal pressure tube connection
E bacterial filter F adaptor for bac. filter G hydrophobic/bac. filter H external humidifier
I inspiration tube from chamber to patient J inspiration tube from ventilator to chamber
● Moistener
● Tubing system
● Mask
● FiO2-Sensor
● Water trap
● etc.
● Re-connect the tube system, alarm(s) should automatically turn off again
● Repeat the alarm test by disconnecting other parts within the tube system
Single tube system min. pressure, apnoea alarm, max. / min. tidal volume, min. SpO 2 min. pressure, apnoea alarm
Double tube system min. pressure, apnoea alarm, max. / min. tidal volume, SpO 2, leakage min. pressure, apnoea alarm
Connecting accessories
■ Connection of an in-house nurse's call system or the external alarm box is carried out at the socket “remote
alarm/nurse call”
■ Connection cable with RJ10 plug is required
Note:
You can order the cable as an accessory
You can order the alarm box and the set as an accessory
C
A
B
A alarm box set B nurse call connection cable C external alarm socket
Connecting accessories
■ Connect the SpO2 sensor to the device, then the following icon appears in the toolbar
Note:
You can order the extension cable as an accessory.
You can order the SpO2 sensor as an accessory.
Connecting accessories
■ Data evaluation and adjustment of the device parameters directly on the device or with the help of the PC
software “easySET"
■ easySET offers the following functions:
● Read out and adjusting the device parameters
Note:
The pre-terminated cable is available as an accessory
Connecting accessories
Note:
Additional SD-cards are available as an accessory
Connecting accessories
■ Using oxygen is possible in all ventilation modes
■ Changes in ventilation parameters, such as I:E, breathing frequency,
or applicate volume lead to change of the FiO2 content
■ Starting Initiation of oxygen: A
B
● Start ventilation, wait for some breathing cycles C
● Then leave ventilation activated for some breathing cycles. End ventilation
Note:
The O2 adaptors are available as an accessory
Connecting accessories
■ In the case of a fix oxygen flow (FlowO2), the oxygen concentration may not be constant
■ Inspiratory oxygen concentration (FiO2) varies depending on pressure, patient's breathing pattern, choice
of mask or amount of leakage
■ During the oxygen discharge, the oxygen concentration should therefore be measured with FiO2 sensor
■ The FiO2 sensor must be calibrated for
accurate measurement
■ Calibrate the FiO2 sensor:
● calibration is carried out using the ambient air
Note:
The O2 measuring-kit and the individual components
are available as an accessory
A connection cable B measuring cell C adaptor for measuring cell D T-adaptor
Connecting accessories
■ Use the functional bag for transporting the CARAT pro on wheelchairs, rollators or in transport vehicles
■ when operating in a functional bag, the following notes must be observed for safe and trouble-free
ventilation:
● Set the alarm volume to Level 3
● Make sure that all the necessary alarm messages can be read through the viewing window
● Make sure the ventilation openings of the bag are not blocked
● Use the bag so that the device is protected from overheating, dust and water
Note:
The function bag and the transport bag are available as an accessory
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
Note:
A minimum volume can be preset in the CARAT pro respirator
to ensure minimum ventilation, with the option of reaching
this via a pressure addition (IPAP + additional pressure)
● constant flow
● decelerating flow
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
Home Screen
■ From the home screen, you can activate the six individual screens:
● Monitoring
● Parameter
● Alarm log
● System
● Statistics
● Service
■ Monitoring → Graphs
● Display with active ventilation
■ Monitoring → Loops
● Graphical display of loops
● flow:volume
● volume:pressure
Parameter screen
■ For set the ventilation- and alarm parameters
■ Up to 3 sets with different ventilation modes and parameters
- Determine the number of sets in the system screen
■ Active sets (green) can not be deactivated under System
to avoid accidental deletion of the set
■ Active set (green) starts when the on / off button is pressed
■ The currently edited set is blue.
■ Thus, for example, set 1 can be edited
(blue) while set 2 (green) is actively
ventilating.
Note:
The ventilation and alarm parameters can
only be set in clinical mode
● Events
● Errors
■ View the last 50 incidents with the date and time of occurrence
■ Sorted chronologically
Service screen
■ PIN code protected
■ For service by authorized service companies Content:
● Calibrate flow sensor
● Calibrate the FiO2 sensor
● Default settings
● SW version
● Serial number
Statistics screen
■ Statistical evaluation of the ventilation parameters is performed for each ventilation set using percentiles
■ Percentiles are measures for the spread of statistical distribution of measured values
■ Statistics are available for the following ventilation parameters:
● minute volume
● frequency
● tidal volume
● I: E ratio
● leaks
System screen
■ Make basic device settings
■ Perform calibrations
■ See the device information
■ Some system settings can not be changed in Home mode
Night Screen
Key lock
■ To protect against accidental changes to device settings
■ The functions of all control elements are deactivated, except:
● ON / OFF button to start ventilation
● ON / OFF button + MFK to stop ventilation
● Alarm button
User modes
■ Device has two user modes: Clinic and Home
■ Activated user mode is displayed in the toolbar
■ User mode can be set in the system screen
■ Clinic mode:
User has full access to all device settings
■ Home mode:
ventilation and alarm
parameters can not be changed
■ switching from one set to another
remains possible
Before you discharge the patient into home ventilation, you should observe the following:
■ Limit the number of ventilation sets to those used by the patient
■ Adjust all the patient's parameters and settings for home ventilation
■ Finally, set the user mode to "Home"
● if necessary, additionally use key lock
● Hygiene, cleaning
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
The alarms related to the respective ventilation mode are distinguished in 3 alarm priorities:
■ HIGH
Immediate action is required. Monitor the patient and set the alarm off immediately
■ MIDDLE
Medium priority requires fast action. Clear the cause of the alarm
■ LOW
For low alarm priority the user's attention is required. Low priority alarms indicate a change in the
"normal" device operation. Clear the cause of the alarm
Volume of audio alarms
■ Audio alarm volume can be adjusted in the system screen
■ Level 1 and 2 are dynamically raised to level 3 if the alarm condition persists after 1 minute
■ If several alarms appear consecutively or simultaneously, then alarm with highest priority is displayed
■ If several alarms are displayed consecutively or simultaneously, then display the alarm with the highest
priority
© HOFFRICHTER 2018 |CARAT II pro | User | SW Ver. 3.105 | Rev. 3.105 EN | 84 |
Alarms
Forwarding of alarms
■ Alarms can be forwarded by means of a nurse call or an optional alarm box
■ Thus even better monitoring of the device in the home and hospital area. Especially when using multiple
ventilation devices in a room
■ All alarms are forwarded without delay to the nurse call system or alarm box
■ All alarms, including mains voltage failure, are transmitted.
■ Setting the alarm volume (SYSTEM) does not affect
external alarms
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
Note: Note:
The membrane change kit is available as an accessory Filter cassette or its filters are available as an
© HOFFRICHTER 2018 |CARAT II pro | User | SW Ver. 3.105 | Rev. 3.105 EN
accessory | 88 |
Cleaning and Maintenance
Clinical application
■ We recommend a non-alcohol desinfect, e.g. schülke mikrozid sensitive wipes
■ Similar disinfectant wipes are also acceptable as well
■ Device must be completely dried before commissioning
■ When used by several patients (in clinic), to protect the device from
contamination by human pathogenic germs, suitable bacterial filters
(e.g., Medisize Barr-vent S) throughout
■ Bacterial filters must be changed daily
Safety Checks
■ To maintain and verify the device functions
■ always before commissioning
■ By an authorized service technician
Maintenance
Battery care
■ In order to get the full capacity of the batteries, it is important to recharge the batteries on a regular basis
■ The number of charging cycles of lithium-ion batteries is limited, so batteries must be changed regularly
■ This is the case when the battery test at service menu is not passed, but at the latest after 2 years
Replace batteries
■ Exchange every 2 years by an authorized service technician
■ The procedure is described in the service manual of the CARAT pro
● Disconnect the device from the power supply and use it for 1h in battery operation
■ Passed:
● if after 1 hour the battery charge is > 10% and the alarm
"Internal battery low" has not occurred
■ Failure:
● If battery charge drops below 10% and the “low battery"
alarm has occurred, have the battery replaced by an authorized service center
■ Recharge batteries to 100%. To do this, operate the device on the power supply
1 Usage Purpose
2 Safety Information
4 Description of Device
5 Commissioning
6 Ventilation Modes
10 Practical Exercise
Exercises:
■ System assembly with filter or humidifier
■ Power on the unit
■ Setting of ventilation parameters and alarms
■ Activation of home or clinic mode and using the key lock
■ Suppression of alarms
■ Change of ventilation parameters and alarms
■ Change of coarse and fine filters as well
as the expiratory membrane