Professional Documents
Culture Documents
Apeksha Project Report
Apeksha Project Report
Dosimetry”
1. ABSTRACT
2. INTRODUCTION
The term “luminescence” itself was first introduced in 1888 by Eilhard Wiedemann to
describe the emission of light by processes that do not involve heat, as opposite to the
case of light emission by materials heated to elevated temperatures. But, in fact, the
thermoluminescence observations date from much earlier.
2.2 Application
Medical Dosimetry
Environmental Dosimetry
Environmental Dosimetry is used where it is likely that the environment will generate
a significant radiation dose. An example of this is radon monitoring. Radon is a
radioactive gas generated by the decay of uranium, which is present in varying
amounts in the earth's crust. Certain geographic areas, due to the underlying geology,
continually generate radon which permeates its way to the earth's surface. In some
cases the dose can be significant in buildings where the gas can accumulate. A
number of specialised dosimetry techniques are used to evaluate the dose that a
building's occupants may receive.
Other Applications:
Thermoluminescence dating
Thermoluminescent dosimeter
2.3 Types
Clinical dosimetry
The proton radiotherapy is one of the methods of treating eye-ball cancer. In this type
of therapy, a suitably formed proton beam is applied to irradiate the tumour. It is a
distinctive technique for the proton radiotherapy. Thanks to the so-called Bragg peak,
a suitable irradiation dose is precisely delivered to the treated tumour and at the same
time the irradiation of healthy tissues is minimal, what reduces side effects of the
treatment. Proton radiotherapy enables an effective destruction of tumours saving
critical organs such as optic nerve, yellow spot (macula) or lens. Therefore, the
patients can save their vision in the treated eye.
A part of IFJ is the Bronowice Cyclotron Centre, engaged in the application of proton
beam for scientific research and eye-tumor radiotherapy. The activities are based on
operation of two cyclotrons and two gantry devices. In medical applications the
requirements for the dose measurement accuracy are very high. Even small changes
of TLD sensitivity should be taken into account, so for dose mapping the prototype of
two – dimensional (2D) TL dosimetry systems were developed and realized. Some
examples of them are shown in Fig. 5.
Large-area (of a few cm2 ) TL detectors were developed using different techniques.
For example they are obtained by: attaching a thermoluminescent (LiF:Mg, Ti or
LiF:Mg, Cu, P) powder to 0.3 mm Al foil, or by pressing the powder and sintering.
Currently, the planar TL system is being applied to visualize the sub-millimeter. The
FIGURE 5
The cosmic radiation was discovered by Victor Franz Hess on 7-th August 1912 year
(Nobel Prize in 1936) and exactly 100 years later on 7-th August 2012 the rover
Curiosity has made the first measurement of radiation on Mars. Now the space
shuttles are able to penetrate up to 300 km from Earth.
• The gallactic component is coming from outside the solar system. It is a very
high energy (1020 eV), highly penetrating, difficult to shield, stimulating
secondary radiation. Its source are mainly nuclei of atoms ( 1H to 238
U) of high
ionisation density, giving strong biological effects
• The solar component exhibits a cyclic activity. It is a kind of solar wind with a
low energy, caused by solar explosions. It infects the galactic component.
• The radiation belts - Van Allen’s belts are the outer belts, consisting mainly of
electrons and inner belts of protons
The total duration of experiments was of 4705 days. The measurements were
performed inside and outside the space station. The MATRO-
SHKA model of the human body, designed to determine the spatial distribution of
radiation dose and exposure assessment of radiation astronauts, was equipped with a
set of active and passive radiation detectors. The Phantom was composed of 33 layers
in regular 2.5 cm lattice containing 1631 measurement sites with 5373 TLD detectors
(3140 from IFJ PAN) and 7 active detectors.
MATROSHKA - 2 KIBO
The three first of them measured the equivalent doses equal to: 0.695 mSv/d, 0.529
mSv/d and 0.569 mSv/d respectively. The personal dosimeters of the astronauts
showed respectively: 2.330 mSv/d, 0.620 mSv/d and 0.644 mSV/d, what means that
these dosimeters overestimated the radiation exposure by a factor of 3 outwards of
ISS and 13 – 18% inside the International Space Station (ISS).
The rapid assessment of the radiation dose after unexpected exposure is a task of
accidental dosimetry. In case of a radiological accident glasses originating from
mobile phone screens, placed usually near the human body could be used as
emergency thermoluminescent personal dosimeters.
The time between irradiation and TL readout is crucial and therefore preparation of
the mobile phone screen and their readout conditions should be optimized.
This preparation rest on removing the screen from the phone, removing foil, etching
in acid or a mixture of acids and cutting into sample 3×3mm 2. Then after a choice of a
filter the signal spectrum is possible to be read.
The glasses extracted from different brands of mobile phones have different
dosimetric properties but all of them give a luminescence signal, which may be used
to determine the absorbed radiation dose.
Another example with use of every day object, where TL signal is applied, is a
security system for banknotes. New banknotes that have entered into circulation in
Poland in the spring of 2014 have a high, own signal, so are less suited for
falsification.
The new activites, such as the micro imaging dose distribution, diagnostics of ion
beams, radiography (with resolution less than 1 µm) and observation of the single
traces of charged particles need LiF monocrystal detectors. Two techniques of
monocrystal groving: the Micro Pulling Down, and the Czochralski method are under
use and testing.
The application of optically stimulated luminescence (OSL) for dosimetry has been
reviewed previously [1]. Landauer has developed the application of aluminium oxide
• high sensitivity, improving low dose precision and enabling thin layer
dosemeters;
Landauer have recently developed a new system, called InLight ®, based on modified
Panasonic dosemeters and readers. In addition, we have developed a small, manual
reader, called microStar™, able to read-out small dots of our OSL detector material.
This extends our range of application into medical and other fields where single point
measurements are required.
This paper describes the microStar system, outlines the principal characteristics of the
system and some trials of the system.
The microStar system comprises dosemeters, which may take the form of
conventional personal monitoring dosemeters or individual dots, and a purpose
designed reader, as described below.
As its detectors, Landauer use aluminium oxide doped with carbon (Al 2O3:C). The
aluminium oxide, Al2O3:C, used in our dosemeters is produced at the Landauer
Crystal Growth Facility in Stillwater, USA. Al 2O3:C (corundum or sapphire) is used
primarily because of its high sensitivity to radiation (40 – 60 times that of LiF (TLD-
100)). It has a principal emission peak at 410 – 420 nm (blue). It also has excellent
properties for radiation dosimetry, including linear response up to 300 cGy. Al 2O3:C
powder is obtained by grinding crystals and sifting the powder to the desired size
range. This powder is mixed with a polyester binder and coated onto a roll of
polystyrene film. The aluminium oxide layer is approximately 0.2 mm thick
sandwiched between polyester foils 0.03 mm thick (top) and 0.1 mm thick (bottom).
The film roll is subsequently cut to the desired shape and size (for InLight dosemeters
and dots this is discs approximately 5 mm diameter). In commercial applications
these discs are either inserted into Panasonic type dosemeters or small holders for
individual dots.
Reader
A number of methods have been developed for analysis of the OSL from Al 2O3:C
detectors, including the pulsed optically stimulated luminescence method adopted for
Landauer’s Luxel badges. However, in the InLight system, Light Emitting Diodes
A number of readers are available, including fully automatic readers for reading up to
500 dosimetry badges in a single process.
The microStar reader is a small, portable reader, capable of reading out personal
monitoring dosemeters or dot detectors. This was designed for a variety of
applications, including medical dosimetry. The microStar measures approximately 30
x 20 x 10 cm and weighs approximately 15 kg. It uses a conventional electrical
power source (110 – 240 V, 1.5 amp, 50 – 60 Hz) and no other connections. It is
therefore a ‘plug-in and operate’ piece of equipment.
To control and monitor the operation, microStar has a dial and two indicator lights.
The reader connects via a USB cable to an external computer that controls the setup,
analysis and data recording. Quality control (QC) procedures to ensure accurate dose
results include: the measurement of a low intensity light photon source calibration
standard; dark current and LED intensity; the ability to calibrate to any set of
reference conditions; and the use of quality control dosimeters for periodic QC
testing. The frequency of quality control procedures is user defined and depends upon
the desired accuracy. For normal measurements (bias < 5%), calibration is needed
once a month or after transport.
In medical dosimetry, there are a wide range of clinical conditions (e.g. technique,
energy, angle, dose rate, etc.) which require the user to have a very thorough
understanding of their dosimetry system and how it performs in these various
environments. This translates into a series of correction factors which are typically
applied by hand after the measurement has been made. So, while electronic devices
such as diodes and MOSFETs provide instantaneous output, it is entirely possible that
an accurate measurement may require minutes or hours of post-measurement analysis,
or a knowledge of the exposure conditions that is not practical (e.g., irradiation angle
during an IMRT treatment).
These advantages will be described in further detail using data from published sources
as necessary.
Radiotherapy fields
For radiotherapy the energy range of interest is for 60Co (mean energy 1.250 MeV)
upward to 20 MeV using both photons and electrons. The on-phantom response at
these energies is about 3 times less that the response at 48 keV. Figure 3 gives the
energy response for the dots over the radiotherapy range. It can be seen that the
energy response typically varies by less than 4 % over this range. If one further
considers only the output from a linear accelerator (ignoring Co-60), this variation is
reduced to 2% or less. This allows a technician to use a single calibration for all
clinical conditions. In the event that the user does not record the exposure energy, the
readings are still valid. For those cases where even greater precision is desired, the
microStar is flexible enough to be individually calibrated to each particular field.
1.04
Relative response
1.02
1 Viamonte
0.98
(photons)
0.96
Jursinic
0.94
(photons)
0.92
0 5 10 15 20
Jursinic (electrons)
Schrembi (photons)
Schrembi (electrons)
Energy /MV
Shembri and Heijmen noticed a difference of 3.7 % in the response between the
photon and electron beams. There is also a significant difference for 18 MV photons
between the results for Viamonte et al [9] and Schembri and Heijman. The reasons for
these variations are not yet understood and indicate the need for a careful
understanding of the measurement protocol used before comparing the results.
With an effective Z of ~10, the Al 2O3:C detector material will over-respond to the x-
rays typically found in diagnostic applications. The expected energy range for these
exposures is from 20 kVp (mammography) to 120 kVp (Computed Tomography
(CT)). The energy response of the OSL dot detectors has been determined [10] both
for exposures on phantom (30 x 30 x 15 cm PMMA) and free in air for energies in
this range and is shown in Figure 4. To minimize the measurement bias of this effect,
Landauer supplies a set of calibrated dot dosemeters. These calibrates are created
using an 80 kVp Philips Industrial x-ray generator with a 2.5 mm filter of Al (2.9 mm
Al HVL) with the dosemeters exposed on a PMMA phantom. The energy response
given in Figure 4 is normalized to this exposure condition.
Relative response
1 On Phantom
Free air
0.8
0.6
0.4
0.2
0
0 20 40 60 80 100 120
Energy /keV
Using this set of reference conditions, a minimal correction factor (Table 1) can be
applied to measurements made for mammography, fluoroscopy, and CT. These
correction factors can be directly entered into the microStar, effectively creating a
separate calibration which can then be applied as necessary.
Again, the flexibility of the microStar allows the user to calibrate the device to
virtually any environment and achieve accurate results.
Angular response
Radiotherapy fields
At radiotherapy energies the angular response of the OSL dose detector is uniform
from 0 – 360o incident angles (Figure 5). This allows the technician to make a
measurement without knowing the exposure angle. This is important for a variety of
treatments in which the gantry angle is changed throughout the procedure.
Dose response
They find a linear response up to 600 cGy with a slight concavity at higher doses.
Similar results have been found by Jursinic, Viamonte et al and Schembri and
Heijman. The slight non-linear response above about 600 cGy is reproducible and
several authors have fitted the curve either with two linear equations covering
different regions or by a simple linear-quadratic equation.
Schembri and Heijmen also examined the dose rate response up to 6 Gy per minute
finding that all the results were within 1 %.
Jursinc found that the deviation from linearity became measureable at 300 cGy. This
response is predictable and the microStar provides its users with the ability to fit a
second-order polynomial to address this non-linear response. Additionally, a user
could simply calibrate the device over a more narrow range tailored to their clinical
conditions (e.g. 100-300 cGy for 200 cGy dose fractions).
Trials undertaken
Other existing methods used in radiotherapy and medical applications include the use
of ion chambers TLD, diodes and MOSFETs. A number of the trials undertaken with
the OSL dots have compared the properties of these devices and commented on the
relative advantages and disadvantages. Viamomonte et al note that the use of TLDs is
often considered as a complex read-out process involving heating and annealing
which precludes instantaneous or near instantaneous readout. Diodes, which are
commonly used for in vivo dosimetry, give an instantaneous readout, but their
calibration factor is dependent upon energy, angle, temperature, and dose-rate.
MOSFETs dosemeters are largely free of these drawbacks, but their life-time is
limited. Both diodes and MOSFETs and active devices and their wiring need to be
considered.
Energy dependence
As noted above (Figure 3), the energy response for the OSL dot detectors is relatively
independent of energy in radiotherapy fields. In contrast diodes have a reduction ion
response as a function of energy up to about 10 % in this energy range (see for
example Jursinic figure 9).
Angular dependence
The angular dependence of the OSL dot detectors for radiotherapy applications is
similarly independent of angle.
Dose response
Jursinic made some detailed measurements of the stability of the signal as a function
of time after irradiation. He found that (with 6 MV there is an initial drop in the
signal relative to that measured one minute after irradiation of about 30 %, but that
after 8 minutes the signal becomes very stable with a 2% decay from 10 to 3600
minutes.
Re-readability
Jursinic have noted that one particular advantage of the OSL dot dosemeters is that
they can be rtead repeatedly with very little loss in signal (0.05% per reading).
Combined with the very small long term fading of the signal, this gives the possibility
of reading the dosemeter and retaining it for future analysis or indeed as a permanent
record of the measured dose (for example attached to patient records).
New developments
To better accommodate the emerging needs of the medical community, Landauer has
recently released a new dosemeter, called the ‘nanoDot’. A picture of this device is
shown in Figure 8.
Figure 8: The new ‘nanoDot’ and adaptor for readout in the microStar reader.
The new dosemeter has the same detector material as previous dots and dosemeters,
but in a form factor 50% of the standard Dot measuring 10 mm x 10 mm x 2 mm.
This allows it to be placed in more restricted spaces such as the eyelid. As with the
previous dots an adaptor (also shown in figure 8) has been developed to allow it to be
easily read-out with the standard reader and all operations (exposure, readout and
storage) can be undertaken in ambient light conditions.
Concluding remarks
The OSL microStar system developed for medical applications could be an ideal tool
for radiotherapy applications. In comparison to TLD, the detectors are identified with
individual serial numbers the can be repeatedly read out enabling them to be archived
with the patient record as a permanent record of dose and can be readout within
minutes of exposure. In comparison with electronic detectors (Diodes and
MOSFETs), they require no cabling and are relatively cheap. For radiotherapy
applications the variation in response at different energies and angles is negligible and
a number of authors have noted their suitability for this application, particularly
entrance dose. The important corollary of these features is that the system doesn’t
2.4 Limitations
3. LITERATURE REVIEW
3.1 History
Robert Boyle (1627-1691) in 1668 has described a light emission from the diamond
heated to the human body temperature.
Alexander Stewart Herschel (1836 – 1907) in 1864 and 1873 observed a bright
green colour light emitted from the inner parts of two respective Geminid meteors.
In 1905 Maria Skłodowska Curie (1867 – 1934) has noticed and described the
thermoluminescent emission from the natural calcium fluoride (CaF 2) exposed to
226
Ra radiaton. She remarked that the natural calcium fluoride glows after heating and
that the emitted light intensity depends on the time of exposition. It was the first,
observed, connection of the thermoluminescence with the ionizing radiation.
Owing to the many years of investigations and the obtained achievements we review
in this paper primarily the research performed in IFJ on TL materials and their
application in the dosimetry.
Sensitivity to light
Advantages
High Sensitivity
High Precision
Size
Convenience
Readout Flexibility
Disadvantages
Sensitivity to light
X-ray diffraction (XRD) patterns of all doped samples were collected in order
to determine the crystal structure and purity and were compared to the XRD patterns
of the undoped samples previously reported. Table 2 and Figure 2 show the Rietveld
refinement results and the XRD patterns (bars/columns under data represent the
simulated XRD of undoped orthorhombic InVO4 and monoclinic InNbO4 and
InTaO4), respectively. All diffraction peaks correspond to the orthorhombic
InVO4 and wolframite InNbO4 and InTaO4 crystal structures. The host lattice was
barely affected by the small concentration of the dopants used. Similar results have
been also observed using other doping elements. In these structures, In3+ has 6-fold
octahedral coordination in all the samples, with an ionic radius of 0.8 Å. Considering
the same valence and coordination, Tb3+ and Yb3+ have ionic radii of 0.923 and 0.868
Å, respectively. Therefore, RE ions can be assumed to occupy the In 3+ sites in the
InVO4, InNbO4, and InTaO4 host lattices. Results from XRD and RS are consistent
with this hypothesis. Only some residual material from precursors have been observed
in the case of InTaO4 doped with Yb3+ (see Table 2). Even though RE ions possess
bigger ionic radii than In3+, it is observed that there is a small diminution of the lattice
parameters leading to a reduction of the unit cell volume less than 1%, contrary to
what would be expected (see Table 2). These variations of the lattice parameters could
be ascribed to a distortion of the octahedral units when a foreign RE element is
introduced into the crystal structure, which reduces the unit cell volume to
accommodate the RE ions. As we will see later, these modifications will be reflected
in the self-activated PL properties of the materials.
Figure 2. XRD patterns of InVO4, InNbO4, and InTaO4 doped samples with Tb3+ or
Yb3+. Bars/columns data represent the standard ICSD charts of the undoped
orthorhombic InVO4 (ICSD-237482) and the undoped monoclinic InNbO4 (ICSD-
257869) and InTaO4 (ICSD-72569), respectively. The height of the bars is
proportional to the theoretical intensity of the peaks. Tables indicating the index,
positions, and intensities of all reflections are included in the Supporting Information.
Table 2. Unit Cell Parameters and Goodness of the Rietveld Refinement for InVO4,
InNbO4, and InTaO4 Compounds and the Corresponding Doped Samples with Tb3+
or Yb3+ from our XRD Experiments and the Contribution of the R esidual Precursor
Materials Found on the InTaO4:Yb Sample
Raman measurements also support XRD observations. As can be seen in Figure 3, the
Raman signal of the doped samples is similar to that of the previously reported
undoped samples. (9,11,12) Although the dopants slightly modify the unit cell, no
appreciable shifts or broadenings of the peaks were observed on the results. This is
due to a small local disorder introduced in the crystalline network. Notice that if the
dopants were located at interstitial sites, that is, not substituting indium, more
important changes (likely with the appearance of new Raman modes) could have been
found in the Raman spectrum. Therefore, Raman measurements support the claim that
the RE atoms substitute In. Regarding the small changes in Raman frequencies, this
can be related to the small unit cell volume change associated with doping. The
reduction of the unit cell volume due to dopants can be seen as the effect of an
external applied pressure to the material equivalent to 0.64, 1.14, and 0.7 GPa for
InVO4, InNbO4 and InTaO4, respectively. (9−12) These pressures would shift the
Raman modes about 3–5 cm–1; however, we have observed all the shifts to be lower
than 2 cm–1, that is, within the instrumental resolution.
5. METHODOLOGY
The model describing the thermoluminescence is based on the quantum band structure
of solids. The energy of electron states in the crystal may be considered as being
derived from states which exist in the individual atoms or ions composing the crystal.
The narrow energy levels, which exist in ions in free space, are broadened in the
lattice into bands by the electric field created by the surrounding ions. In a perfect
crystal two bands allowed stands out (cf. Fig. 1):
The lower limit of the conduction band is located above the upper limit of the valence
band. The energy difference between these bands is called the band gap or energy
gap. By convention the crystals with a width of energy gap of more than 3 eV are
counted to dielectrics, and those with width of energy gap of less than 3 eV are
counted to semiconductors. If the width of the energy gap is equal zero, or the valence
band overlaps the conduction band the crystal is called metal.
FIGURE 1
range arise local levels called electron traps if they are close to the conduction band
and piercing traps (or simply holes) if they are close to the valence band. Electrons
raised by radiation from valence band into conduction band are captured by the
electron traps. Piercing traps, being recombination centers, during heating capture
electrons released from electron traps, which during this process give rise to a delayed
luminescence light emission.
In the second stage of the TL process due to heating the electrons from traps are
transferred to the conduction band, from which they recombine to the recombination
centers containing trapped carriers of opposite sign (trap / centers piercing). In the
process of recombination, the electrons release energy in the form of light, so that the
substance returns to its ground state (equilibrium). The energy needed to release
trapped carriers is at least equal to the difference between the level of the trap and
conduction band (if the “active carriers” are electrons). For typical thermoluminescent
materials (used in TL dosimetry), the difference between these layers should be
greater than the energy of thermal radiation at room temperature and lower than the
energy of thermal radiation at a temperature of several hundred degrees. Typical
emission spectra at different temperatures observed for LiF:MCP and LiF:MCP are
shown in Figs 2(a) and 2(b), respectively. As expected, an increase of the TL intensity
with increasing temperature is observed.
D = ITLCcalib (1)
The shapes of TL glow curves for LiF:MCP (a) and LiF:MT (b) detectors, obtained
by linear heating with rate 20C/s. Dose 1.5 mGy.
where ITL is the TL signal intensity and Ccalib is the calibration coefficient
As ITL the glow curve TL integrated over a selected temperature range is used. This
relationship can be used only in a limited range of doses and types of radiation
energy. In fact, in order to determine the absorbed dose it is necessary to use a number
of correction factors that take into account, inter alia, the non-linearity dose
characteristics, an amendment of the energy and a decrease of the signal over time
(fading).
The most commonly used TL material is the lithium fluoride LiF. There are several
detectors on its basis, especially such as: LiF:Mg,Ti (called MTS) and LiF: Mg,Cu,P
(called MCP) [9- 12]. For the production of TL detectors the natural lithium,
containing 92.5% of 7Li and 7.5% of 6Li is used, as well as lithium enriched or
depleted in the isotope 6Li. Thereby for each type of MCP three sub-types: MCP-N,
MCP-6 and MCP-7 are obtained. Some of the detectors developed at IFJ are shown in
Fig. 3.
MTS detectors based on LiF: Mg, Ti were discovered in the 40-s of the last century.
In 60-ties T. Niewiadomski with a team from IFJ has developed a technology for
production of LiF:Mg,Ti (12 ppm Ti, 200 ppm Mg) detectors in form of sintered
tablets having the diameter of 4.4 mm and the thickness of 0.4 to 0.9 mm. These
detectors are perfectly tested and became standard TL dosimetry measurers of
ionizing radiation. Signals of detectors MTS-N exhibit linear dependence on dose up
FIGURE 3
Different forms TL detectors developed and manufactured in IFJ. (a) MTS-N (LiF:
Mg,Ti) produced in form of solid pellets, chips or powder are environmental used
world –wide in radiation protection, environmental monitoring and clinical
dosimetry ,(b), EYE-DTM allows for precise measurements of radiation doses eye lens,
also in case when protective glasses are used, (c) ring holders enabling dose
measurement during operations with radiation, (d) TLD aluminium dosimetric cards,
with TL pellets or chips sealed in fluoropolymer (eg.PTFE), can be evaluated in all
compatible hot-gas readers. One to four detectors in arbitrary configurations and
sealed in fluoropolymer foil are mounted into the card. Different shape, thickness and
isotopic composition of detectors (7Li, 6Li, natural Li ) different colors and card
numbers are available on request.
The very important finding in MCP investigations was the appearance of a peak in the
high temperature emission of LiF: Mg,Cu,P detectors, heated to about
FIGURE 4
Changes of the glow-curve shape of MCP-N detectors exposed to ultra high doses: (a)
exposure below 50 kGy deconvoluted on 8 peaks, (b) exposures to doses above
50kGy: a new peak appears.
6000C after exposition ranging from 1kGy to 1 MGy [15-16]. With the increasing
dose its position shifts toward higher temperatures (cf. Fig. 4). The peak has
application for ultra-high-dose measurements. The new method in combination with
the traditional method of determining the dose based on the main dosimetric peak
makes the LiF: Mg,Cu,P detector unique capable of measuring doses ranging of 12
orders of magnitude from microgrey to megagrey.
The TLD method is the most often used technique in individual and environmental
dosimetry. Due to the small size, low price, good energy characteristic, a wide range
of measured doses, insensitivity to environmental factors TL detectors are readily
used in virtually all areas of dosimetry. But new materials and new TLD measurement
methods are permanently developed.
In addition to the thermally simulated TL detectors the optically (by light) stimulated
(OSL) ones are also developed. Similarly, the radio-photo-detectors, controlled by the
Mungsaji Maharaj Mahavidhyalaya, Darwha (2021-2022)
36
“Applications of Luminescence in Radiation
Dosimetry”
ionizing radiation and light exist too. In OSL detectors electrons are excited to the
conduction band by light. They migrate to the vicinity of hole centers and recombine
giving the light emission. OSL detectors are vulnerable to light so, in consequence,
they have to be stored in dark.
Comparing the measurement techniques TL and OSL, we see that they are based on
the same physical mechanism and differ only by the stimulating factors (light or heat).
The other, RPL and OSL techniques, differ by the physical mechanism and are similar
by the stimulating factor (light).
LiF:Mg,Cu,P, CaSO4:Dy, CaF2:Tm, Li2B4O7:Cu, Mg2SiO4 and others. For OSL one
uses Al2O3:C and BeO. For RPL the phosphate glasses as well as LiF for high doses
are applied.
More than 1 million of TL detectors from IFJ is actually used in nuclear power plants
as well as in dosimetry service clinics and laboratories in more than 40 countries.
In the individual dosimetry we usually have to deal with doses ranging from 0.1 mSv
to 1 Sv, evoked by radiation energies of 15 keV to 3 MeV during the exposure times
from 1 to 3 months. Here Sv is the Sievert radiation dose unit. The most common
The detectors used in the environmental dosimetry are highly sensitive. They detect
doses within a 3 orders of magnitude range, from 0.03 mSv to 1 Sv and are used for
monitoring the natural environment as well as the working environments.
The recently developed LiF: Mg,Cu,P detectors for measuring the ultra-high doses
were applied for testing various radiation fields at devices operating at Grand
Collisionneur de Hadrons (LHC), located in CERN (Organisation Européenne pour la
Recherche Nucléaire), Geneve, Suisse. More than one hundred TL dosimeters were
mounted in the infrastructure of the LHCb experiment. The dosimetry installed at the
International Thermonuclear Experimental Reactor in Cadarache, France, where
streams of fast neutrons are generated, was a particularly ambitious challenge.
6. FUTURE SCOPE
A comprehensive list of dosimeters used in FLASH studies and other high dose-rate
modalities is given in Table 1. The different columns represent some of the major
issues identified in section Dosimetric Aspects of FLASH That Need to be
Considered. The values are based on typical values and usage encountered in
literature. Exception to these values do exist; for example, radiochromic film is
categorized as a passive detector, but attempts are real-time dosimetry with film has
been made in the past. The “Measurement Type” column has bold entries in it, which
indicate the way those dosimeters were employed in FLASH studies. The time
resolution values are based on the underlying physics of the dosimeters, as explained
previously. This does not take in to account the available bandwidth of the read-out
method. Of course, the dead-time of the read-out electronics should be considered
while dealing with such dose-rates. While some of these issues are not necessarily
unique to FLASH, they nonetheless contribute to the overall dosimetric uncertainty.
The thermally stimulated luminescence is the most commonly used technique from all
measurement methods use in the passive dosimetry of ionizing radiation
Currently, the development of the dosimetry techniques turns toward the optical
stimulation as the light stimulation does not require a heating of the sample, or the use
of protective atmosphere. OSL detectors have a wide range of possibilities for tuning
of the stimulating and measurement wavelength and the possibility of multiple
readout once the irradiated samples. They may find as well, application in emergent
imaging and in micro-imaging of ion beams.
7. REFERENCES
[11] Y.S.. Horowitz, LiF:Mg,Ti versus LiF:Mg Cu,P: the competition heats
up, Radiat. Prot. Dosim, 47,135-141(1993).