Professional Documents
Culture Documents
Measurement Techniques
Measurement Techniques
1. Measuring instruments
Rulers
1. Metre rulers are used as vertical or horizontal scales in experiments.
2. To set a metre rule in a vertical position: use a plumb line and align the ruler parallel to it, or
use a set square perpendicular to the ruler and bench if the bench is known to be horizontal.
3. Look perpendicularly towards the scale and object you are measuring to avoid making a
parallax error. This is an error in reading an instrument due to the relative movement between the
object and scale marking when the object is viewed from different positions against the scale
markings.
Vernier callipers
http://surendranath.tripod.com/Applets/General/Vernier/Vernier.html
Micrometers
1. Micrometers give reading to within 0.01 mm.
2. Reading a micrometer:
Read the main scale including half divisions.
Add the thimble scale reading.
3. Precautions
make sure that the faces between the anvil and spindle are clean,
Do not over tighten, use the ratchet.
Check the zero error and account for it when taking a reading.
MEASUREMENT TECHNIQUES 2
http://www.upscale.utoronto.ca/PVB/Harrison/Micrometer/Flash/MicSimulation.html
Zero error
(a) The reading is low by 0.03mm. We have to add 0.03 mm to the reading we take with this micrometer.
MEASUREMENT TECHNIQUES 3
(b) The reading is too high by 0.03 mm. We have to subtract 0.03 mm to the reading we take with this
micrometer.
Timers
1. When using a stopwatch, the reaction time, about 0.2 s, is variable for any individual. The
precision of a single timing is unlikely to be better than 0.1 s
2. When timing oscillations, the timings should be repeated several times to get an average value.
3. Use a fiducial mark for accurate timings.
4. When measuring the time period of a pendulum, use the countdown method to start the
timings: 5, 4, 3, 2, 1, 0, and at zero, start the stopwatch when the bob passes the fiducial mark
moving in the same direction.
An oscilloscope
1. Check that the variable control is fully clockwise.
2. Check that the oscilloscope input is for d.c. measurements or for a.c. measurements.
3. When measuring an a.c. waveform, ensure
(i) the y-gain is adjusted so that the vertical height of the waveform is as large as possible from
top to bottom on the screen,
(ii) several cycles are displayed across the screen, and measure across as many cycles as
possible to reduce experimental uncertainty.
R to Y input
of CRO
1 cm
The wave shown is obtained when the time base is 10 ms cm-1 and Y sensitivity is
0.5 V cm-1.
What is the frequency and peak voltage?
Answer
The time taken for one cycle T
= 2 cm x 10 ms cm-1
= 20 ms = 0.02 s
1 1
f= = =50 Hz
T 0.02
A travelling microscope
1. The microscope can be aligned vertically or horizontally to measure a vertical or horizontal
distance. The platform has a vernier scale to measure distances to within ±0.1 mm
MEASUREMENT TECHNIQUES 4
2. Before use, the platform should be levelled horizontally using a spirit level, otherwise there
could be a systematic error in the measurements.
10 mm 20 mm
For example, with a rule graduated in mm, there is an error of ½ mm in judging the position of
each end of the body against the rule, so the maximum possible error is
2 x ½ mm = ± 1 mm.
The precision is ±1 mm.
A micrometer with a precision of 0.01 mm gives readings that each have a precision of 0.01 mm
3. For a reading of say (208 ± 1) mm:
the absolute error is 1 mm
(this has units)
1
the fractional error is
208
(this has no units)
1
the percentage error is x 100
208
¿ 0.48 % = 0.5 %
(this has no units)
The error is an estimate only and is usually required to one significant figure.
Combining errors
Consider two quantities:
a = (5.1± 0.1) cm, and
b = (3.2 ± 0.1) cm.
Q = the quantity required.
MEASUREMENT TECHNIQUES 5
1. Adding or subtracting
When two quantities are to be added or subtracted, add the absolute error.
Q = a + b = (8.3 ± 0.2) cm
Q = a – b = (1.9 ± 0.2) cm
2. Multiplying or dividing
When two quantities are to be multiplied or divided, add the % errors.
First the % errors for a and b are found:
Note: The answer is quoted with the same number of decimal places as the probable error.
3. Powers
For a quantity raised to a power:-
multiply the % error by the power.
Q = a2 = 5.12 (2 x 1.96%)
4% x 26.01
= 1.0404 = ± 1 cm2
MEASUREMENT TECHNIQUES 6
Q = (26 ± 1) cm2
Q = √b = b½
= √3.2 (½ x 3.13%)
= 1.78885 (± 1.565%)
= 1.78885 (± 2%)
2% x 1.78885
= 0.035777 = 0.04
Q = (1.78 ± 0.04) cm½
Example
If A = 10 ±1,
B = 12 ± 2,
C = 20 ± 1, calculate
1. (a) B – A,
(b) 2A + 3B
2. (a) A.B
(b) 3A.B
(c) A2 B
AB
3. A2B + 2
C
Answer
1. (a) B – A = 2 ± 3
(b) 2A + 3B = (20 ± 2) + (36 ± 6)
= 56 ± 8.
Note % error in A is the same as in 2A.
Example
The density of a ball is (300 ± 8) kg m-3.
MEASUREMENT TECHNIQUES 7
The ball is placed on a metre rule to find its diameter.
One end of a diameter is opposite the 35 cm mark and the other end is at 78 cm and the error in
each of these measurements is
± 1 cm.
Find the mass of the ball.
Answer (12 ± 2) kg
systematic error
To account for heat losses in heat experiments, the experiment can be repeated with a different
rate of heating or a cooling correction may be applied.
In radioactivity experiments, the background count rate is measured and subtracted from all
measurements to get a count rate due to the source alone.
When measuring the magnetic field inside a solenoid due to a current, the background field is first
measured without the current.
3. Random errors are reduced by repeating the measurements several times and taking the
average value, and by statistical methods.
Example
Suppose the readings of the diameter of a wire, using a micrometer screw gauge were:
Answer
The mean value = 5.9/7 mm
= 0.843 mm
The average of the differences between each reading and the mean, neglecting the sign, is:
= 0.077/7 = 0.011 mm
The diameter = (0.84 ± 0.01) mm
MEASUREMENT TECHNIQUES 9
repeat readings
small systematic error large random error
repeat readings
large systematic error small random error
3. A precise reading is a measure of the smallest measurement that can be made with an
instrument, e.g.
± 0.01 mm with a micrometer,
± 0.1 mm with a vernier callipers,
± 1 mm with a metre rule,
± 1° for a protactor.