CN-6000/CN-3000, CN-6500/CN-3500

ASTM Host Interface


Specifications

Revision 2.05
Table of Contents

1. General ..................................................................................................................................................................... 2
2. Scope ......................................................................................................................................................................... 2
3. Revision History ....................................................................................................................................................... 3
4. General ..................................................................................................................................................................... 7
5. Terminology.............................................................................................................................................................. 7
6. Communication Specifications ................................................................................................................................ 8
6.1. Physical Layer (Hardware) .................................................................................................................................. 9
6.1.1. Serial Connection ....................................................................................................................................... 9
6.1.2. Ethernet Connection ................................................................................................................................ 10
6.2. Data Link Layer (Transmission Protocol) ......................................................................................................... 11
6.2.1. Communication status ............................................................................................................................. 11
6.2.2. Establishment phase ................................................................................................................................ 12
6.2.3. Transfer phase .......................................................................................................................................... 13
6.2.4. Termination phase.................................................................................................................................... 14
6.2.5. Time out .................................................................................................................................................... 14
6.3. Presentation Layer ............................................................................................................................................. 15
6.3.1. Messages, Record and Field ..................................................................................................................... 15
6.3.2. Communication Protocol .......................................................................................................................... 17
6.3.3. Details of Record ....................................................................................................................................... 19
7. Examples of Communication ................................................................................................................................ 60
7.1. Inquiry of Analysis Order (IPU => Host Computer) ........................................................................................ 60
7.2. Inquiry of Re-analysis Order (IPU=>Host) ....................................................................................................... 60
7.3. Analysis Order (Host Computer => IPU) .......................................................................................................... 60
7.3.1. Specify Replication ................................................................................................................................... 61
7.3.2. Specify Measurement Mode ..................................................................................................................... 61
7.3.3. Specify Measurement Time ..................................................................................................................... 61
7.4. Analysis Results (IPU => Host Computer) ....................................................................................................... 61
7.4.1. When analysis is performed 1 time ......................................................................................................... 61
7.4.2. For Multiple Parameter Analysis ............................................................................................................ 63
7.4.3. When there is the re-analysis .................................................................................................................. 67
7.4.4. When analysis is performed 2 times ....................................................................................................... 77
7.4.5. Item unit output (if there is a comment record) ..................................................................................... 79
7.4.6. Output the path name to the image data ............................................................................................... 81
7.4.7. When MDA analysis is performed ........................................................................................................... 81
7.4.8. Output the path name to the Delayed cross mixing image data ........................................................... 97
7.4.9. When the measurement of platelet aggregation is output .................................................................... 98
7.4.10. Output when multiple vial sets are measured (when output targets are “Final Data only”) ......... 105
8. Parameter Code List ............................................................................................................................................ 106
8.1. Coagulation parameter code list ...................................................................................................................... 106
8.2. Immuno parameter code list ............................................................................................................................ 110
9. Appendix ............................................................................................................................................................... 111
9.1. TCP/IP Communication.................................................................................................................................... 111
9.2. Re-analysis orders............................................................................................................................................. 112
9.3. Difference List of Transmission Content ........................................................................................................ 118

1/123
1. General

CN-series device (CN-6000/CN-3000, CN-6500/CN-3500)

2. Scope

This document is intended to supply the information that the CN-series device communicates with the
host computer using ASTM protocol.

ASTM (the American Society for Testing and Materials), one of the world largest volunteer non-profit
organization, founded in 1898 for the purpose of creating standard regulations for material, products and
system services.

This specification conforms to the following two standards:

・Specifications for low level protocols to transfer data between clinical laboratory instruments and
computer systems.
ASTM E1381-02

・Standard specifications for transferring data between clinical instruments and computer systems.
ASTM E1394-97

2/123
3. Revision History

Version Date Description


1.00 2018/11/01 Initial version
1.01 2019/01/16 Analysis Result Record
Content output to CSV file
- Change: Changed “Sample Comment” to
“Comment”.
Example of CSV file output
- Change: Changed “Sample Comment” to
“Comment”.
Stated the PC code for CN-3000 and CN-6000 with side
CP.
6.3.3. Details of Record
2) 7.1.5 Sender Name or ID
- PS Code:
9.3. Difference List of Transmission Content
- <CS-5100>
- <CS-2400/CS-2500>
1.02 2019/01/31 Content on M Model/ H Model was corrected
1.10 2019/02/08 The version number was updated to 1.10
1.11 2019/04/10 6.3.3.9. Comment Records (QC analysis information)
3) 11.1.4 Comment Text
Corrected incorrect statement of field names.
(Error) Calibration Curve ID
(Correct) Control Name
1.12 2019/05/28 7.4.7. When the measurement of platelet aggregation is
output
Changed the default setting for
[HOST_ASTM_OutputAggregationItemUnit] from
“Enabled” to “Disabled”.
1.13 2019/06/12 7.4.7. When the measurement of platelet aggregation is
output
Deleted content concerning output of result records for
Aggregation % for each analysis parameter included in
the sample.
1.14 2019/07/08 6.3.3.5 Result Record
Corrected content concerning the 1st digit of the
parameter code when image data is transmitted
- Changed to specific content
- Corrected content concerning the 1st platelet digit
- Corrected an incorrect value for the 1st digit of the
parameter code in the communications example
6.3.3.4 Test Order Record
- Corrected content concerning dilution ratio to “Always
fixed values for the first 2 decimal places”.
6.3.3.5 Result Record
- Added the following analysis errors
Switched to 0000.0000.0999 high gain
4032.0000.0000 Bump
0000.0000.8000 Coag% Changed
8. Parameter Code List
- Added the host ID for FM 69X
1.20 2019/07/10 Updated the version number to 1.20
1.21 2019/08/09 6.3.3.4 Test Order Record
Added the output of Sample Seq.

3/123
Version Date Description
1.22 2019/08/22 7.4 Analysis Result (IPU => Host Computer)
Added that the targets for host output are “Final Data
only”, and added examples of communication in cases
when multiple vial sets are specified and measured.
- 7.4.8 Output when multiple vial sets are measured
(when output targets are “Final Data only”)
1.23 2019/09/12 6.3.3.4 Test Order Record
- Changed text from “Sample Seq” to “Sample seq.”
- Added the fact that no value is output if there is no
Sample seq. information.
1.24 2019/10/03 6.3.3.4 Test Order Record
4) 9.4.5 Universal Test ID
Added the fact that [HostOrderCaution] is enabled for
[Other service settings] in order to notify users that the
parameter code 000 was received.
1.30 2019/10/16 Updated the version number to 1.30.
1.40 2019/12/03 6.3.3.3 Request Information Record
2) 12.1.3 Starting Range ID No.
7.4.2. When there is the re-analysis
Added 7.4.2.1. Transmitting analysis results
Added 7.4.2.2. Item unit output
Changed section names and composition. Also gave
examples of communications when analysis results are
output to the host on completion of first analysis.
9.3. Difference List of Transmission Content
Added the requirements that the output format for rack
numbers of transfer samples analyzed with Sysmex
transfer must be the same as for ordinary samples, and
that tube positions have 2 digits.
1.41 2020/03/25 Added specifications in which the transferred sample
number becomes L□□□□□ when [Service settings] –
[Other service settings] – [LAS_RackNo_Replace]
setting is “1”.
6.3.3.3 Request Information Record
2) 12.1.3 Starting Range ID No.
9.3 Difference List of Transmission Content
<CS-5100>
<CS-2400/CS-2500>
1.42 2020/05/22 Added HIL information to the error information of the
Analysis Result Record.
Added the fact that the error information of the Analysis
Result Record is output regardless of the setting value
for [HOST_ASTM_OutputHILFlags].
Deleted unused analysis errors.
6.3.3.5 Result Record
2.1) When test parameters are output
Corrected Table 6.3.3.5-2: Content of analysis errors,
analysis result abnormal flags and data values, and
responses.
Added *4 and *5.
Added Table 6.3.3.5-3: Content of analysis result
abnormal flags, data values, and responses according to
the HIL check conditions.
1.43 2020/06/15 Added output specifications to the host computer along
with the newly employed re-analysis function that can
be performed with MDA analysis results.

6.3.3.3. Request Information Record


5) 12.1.13 Requested information status code
Notes:
9.2. Re-analysis orders
Deleted “The analysis is not multi dilution analysis

4/123
Version Date Description
(MDA).” from transmission conditions for a re-analysis
order.
6.3.3.5. Result Record
Table 6.3.3.5-6: List of the information included in the
file
Added the fact that if the first analysis is MDA (there is
the re-analysis), “Average” (the type of image data) is
provided to the file name of the first analysis MDA
graph.
Added file name output examples (“ex6)” and “ex7)”).

7.4. Analysis Results (IPU→Host Computer)


Added examples of communication for MDA analysis
results.
Added: 7.4.6. When MDA analysis is performed
1.44 2020/06/16 Added specifications in which inquiry items differ when
“HOST_ASTM_OrderInquiry_AnalyzableAssay” setting
is enabled for “Service settings” - “Other service
settings”.
6.3.3.3. Request Information Record
Table 6.3.3.3-1: Details of Request Information Record
12.1.5 Universal test ID
9.3. Difference List of Transmission Content
<The function which can be used by making the
setting effective>
Added: Inquiry items
1.45 2020/07/28 Added assay groups for up to Wave E and specifications
of changeable units to “8. Parameter Code List”.
8. Parameter Code List
Changed Table 7.4.8.4 1: Parameter Code Lists
Changed Table 7.4.8.4 2: Parameter Code
Lists(Result)
Added Table 7.4.8.4 3: Number of digits in the unit
format
1.46 2020/08/03 Added examples of communication for analysis results in
cases when “HOST_ASTM_Divide_FinalData” setting is
changed for “Service settings” - “Other service settings”.
7.4. Analysis Results (IPU→Host Computer)
7.4.2.1. Transmitting analysis results
Added Example: Results of redilution analysis (1 sample
measurement for multiple parameters: when only first
analysis parameters exist)
2.00 2020/12/01 Defined the specification of CN-6500/CN-3500.
8. Parameter Code List
Chage the parameter code from numbers to
alphanumeric characters.
2.01 2020/12/10 Added example of analysis result output when multiple
parameters are analyzed.
7.4. Analysis Results (IPU→Host Computer)
Addition 7.4.2. For Multiple Parameter Analysis
Added example of analysis result output when research
parameters are analyzed.
Change 7.4.2. For Multiple Parameter Analysis
Change 9.3 Difference List of Transmission Content
<CS-5100>
<CS-2400/CS-2500>
2.02 2020/12/16 Added and modified parameter codes for platelet
aggregation.
8. Parameter Code List
8.1.1 Coagulation parameter codes

5/123
Version Date Description
2.03 2021/01/20 Added the output sequence for cases when records are
output for multiple parameters in the communications
protocol.
6.3.2.3. Analysis Results & QC Data (CN series device →
Host computer)
Change Table 6.3.2.3-1: Analysis result and QC data
Added footnote
2.04 2021/02/24 Added a specification that allows the format for
date/time appended to the image data file name to be
changed in “Service settings”.
6.3.3.5. Result Record
2.4) The path name to the image data is output only
when the parameters have the image data.
b) 10.1.4 Data or measurement values
Table 6.3.3.5 7: List of information included in the
file
Change 1 Test Date
9.3 Difference List of Transmission Content
<CS-5100>
Addition Image data file names
<CS-2400/CS-2500>
Addition Image data file names
<The function which can be used by making the
setting effective>
Change Image data output
Added research parameter output as a function that can
be used by enabling setting values.
9.3 Difference List of Transmission Content
<The function which can be used by making the
setting effective>
Added Research parameter output
Corrected incorrect statement of CS-5100 specifications.
9.3 Difference List of Transmission Content
<CS-5100>
Change Image data path output (Result record)
2.05 2021/02/26 Changed the error message for an evaluation error.
6.3.3.5. Result Record
Table 6.3.3.5 2: Content of analysis errors,
measurement result error flags and data values, and
responses
Code: 8000.0000.0040

6/123
4. General

This document is intended to supply the information that the CN-series Automated Blood Coagulation
Analyzer communicates with the host computer using ASTM protocol.

5. Terminology

The definition of the terminology used in this document is described in the following.

Numerics:
Indicates ASCII codes “0” (30h) through “9” (39h)

Alphabet:
Indicates ASCII codes “A” (41h) through “Z” (5Ah) and “a” (61h) through “z” (7Ah)

Alpha-numeric:
Indicates numerical or alphabetical character

7/123
6. Communication Specifications

Communication specifications are based on a layer protocol.

(1) Physical Layer


Specifies the sending and receiving of signals between the IPU and the host computer through physical
and electrical connections
See the section “6.1 Physical Layer (Hardware)”

(2) Data link layer


Specifies the sending and receiving of data by link connections and for each frame between the IPU and
the host computer
See the section “6.2 Data Link Layer (Transmission Protocol)”

(3) Presentation layer


Specifies the messages that are sent and received by the IPU and the host computer
See the section “6.3 Presentation Layer"

Presentation layer Specifies message specifications


Data link layer Specifies link connection and frame specifications
Physical layer
Specifies mechanical and electrical specifications

Notes
The CN-series device supports connection by Ethernet/Serial only. From the combinations between
the specifications conforming to two standards and connection types, the CN-series device supports
the following two modes.

1.Ethernet connection
The presentation layer conforms to ASTM E1394-97.
The data link layer conforms to ASTM E1381-02.
The physical layer conforms to IEEE802.3. (*1)

2.ASTM E1381-02 mode (serial interface)


The presentation layer conforms to ASTM E1394-97 mode.
The data link layer and the physical layer conform to ASTM E1381-02.

Serial connection
Ethernet connection ASTM E1381-02
mode
Presentation layer ASTM E1394-97 ASTM E1394-97
Data link layer ASTM E1381-02 ASTM E1381-02
Physical layer *1 IEEE802.3 ASTM E1381-02

*1: The IEEE802.3 specifications in the physical layer are not described in this document.

8/123
6.1. Physical Layer (Hardware)

6.1.1. Serial Connection

6.1.1.1. Connectors
Although the ASTM standard specifies a D-SUB 25-pin male connector as standard, a D-SUB 9-pin
male I/O connector, which is located on the rear of the CN-series device, is used to communicate.

Table 6.1.1.1-1: Connector pin assignment


Pin No. Signal name Signal direction
1 NC
2 Receive data RxD IN
3 Transmit data TxD OUT
4 Data terminal ready DTR OUT
5 Signal ground SG -
6 Data set ready DSR IN
7 Request to send RTS OUT
8 Clear to send CTS IN
9 NC
The control signals are not used with ASTM specifications.
For this reason, do not make connections to unused pins.

6.1.1.2. Signal identification level

Table 6.1.1.2-1: Signal identification level


Level Data signal Control signal
+3V or more Logic ”0”, start bit ON
-3V or less Logic ”1”, stop bit OFF

9/123
6.1.1.3. Connection cable
Please configure a cable with a D-SUB 9pin female adaptor for connecting to the instrument's D-SUB 9
male connector in accordance with the following connection chart.

IPU Host computer


DB-9 DB-9 DB-25

TxD 3 3 2 TxD

RxD 2 2 3 RxD

SG 5 5 7 SG

RTS 7 7 4 RTS

CTS 8 8 5 CTS

DTR 4 4 20 DTR

DSR 6 6 6 DSR

NC 1

NC 9

6.1.1.4. Interface parameters

Table 6.1.1.4-1: Interface parameters


Parameter Selection of settings
Baud rate 300, 2400, 4800, 9600*, 19200 (bps)
Data length 7bits , 8bits*
Stop bit 1bits*
, 1.5bits , 2bits
Parity None *
, Even , Odd
The value with asterisk mark (*) allows conformance with the ASTM standard.
[Note]
Seven-bit data lengths, even or odd parity and two stop bits are recognized by the ASTM
standard for use with special applications.

6.1.1.5. Standard specifications (ASTM E1381-02)


The physical layer of the IPU conforms to ASTM E1381-02 “5. Physical Layer", except for the connector
type. The IPU uses a D-SUB-9 pin male connector. (The ASTM standard specifies a 25-pin male
connector.)

6.1.2. Ethernet Connection


Ethernet connection conforms to IEEE802.3. The communication cable uses the UTP category 5 cable.
Prepare the connector and cable which apply to the Ethernet connector of the CN-series device.

10/123
6.2. Data Link Layer (Transmission Protocol)
The data link layer transfers data between systems using a character-based protocol in accordance with
ASTM E1381-02 “6. Data Link layer”.
This section briefly describes communication control procedures. For details, refer to ASTM E1381-02.
When the TCP/IP connection is intended to be used, the TCP connection is established in prior to the
communication. To establish the TCP connection, the host computer acts as a server and the CN-series
device acts as a client. The CN-series device establishes a connection by requesting a connection for the
IP address and the port number that are provided by the host computer.

6.2.1. Communication status


The data link layer consists of the following two states.
• Neutral Status
• Linked status

Transition to each status is accomplished through the following three phases.

Establishment phase
Establishes communication line and determines the direction of data transfer. In this way, the sender
and the receiver are identified, and the change is made from neutral status to linked status.

Transfer phase
The sender transmits messages to the receiver until all messages are transferred.

Termination phase
Releases the communication line. Changes the communication status both the sender and the receiver
from linked status to neutral status.

Establishment phase

Neutral status Linked status Transfer phase

Termination phase

11/123
6.2.2. Establishment phase
(1) The sender (IPU) sends an [ENQ] signal to the receiver (host computer). To respond to the sender,
the receiver performs the following action.
Returns an [ACK] signal when the communication is enabled.
Returns a [NAK] when the communication is disabled.

If the receiver response [NAK], the sender waits for at least 10 seconds before attempting to send
another [ENQ] signal.

Max. 15 sec. Max. 15 sec. 10 sec.


E or more E
E
N N N
IPU IPU Q Q
Q
N
A
A
Host C
Host K
Computer K
Computer

(2) When both the sender and receiver send [ENQ] signals, the host computer must yield control
authority to the CN-series device.
The CN-series device sends [ENQ] again after 1 second.
The host computer must wait for 20 seconds before sending [ENQ] again.

1 sec.
or more E
IPU N
E E
Q
IPU N N
Q Q E E
N N
E Host
Host Q Q
N Computer
Computer
Q
20sec. or more

12/123
6.2.3. Transfer phase
During the transfer phase, the sender sends messages to the receiver. The transfer phase continues
until all messages have been sent.

(1) Messages are sent in each record with multiple frames. Each frame contains a maximum of 64,000
characters (including frame overhead). If the record is longer than 63,993 characters, it is divided
into two or more frames.

(2) Multiple records cannot be included in a single frame.

(3) If the record contains the maximum number of characters or less, a frame with the following
structure will be transferred.

[STX] [F#] [Text] [ETX] [CHK1] [CHK2] [CR] [LF]

If the record is longer than the maximum number of characters, it is divided into two or more frames.
The intermediate frame text termination code is [ETB], and the final frame text termination code is
[ETX], as shown below.

[STX] [F#] [Text] [ETB] [CHK1] [CHK2] [CR] [LF]


[STX] [F#] [Text] [ETB] [CHK1] [CHK2] [CR] [LF]
……
[STX] [F#] [Text] [ETX] [CHK1] [CHK2] [CR] [LF]

Symbol Explanation
[STX] Start of a frame
[F#] Frame number
One of the numbers 0 to 7 is used, beginning with 1 and repeating
2,3,4,5,6,7,0.
In case of retransmission, the same frame number is sent.
[Text] ASTM E1394-97 records are used. (See the subsequent section “6.3
Presentation Layer”.)
For this reason, the codes below will not be used.
0x00-0x06,0x08,0x0A,0x0E-0x1F,0x7F,0xFF
[ETB] Control code indicating end of text (for intermediate frame)
[ETX] Control code indicating end of text (for the final frame)
[CHK1] [CHK2] Expressed by characters “0” - “9” and “A” - “F”.
Characters beginning from the character following [STX] and until
[ETB] or [ETX] (including [ETB] or [ETX]) are added in binary. The
2-digit numbers, which represent the least significant 8 bits in
hexadecimal code, are converted to ASCII characters “0” - “9” and “A” -
“F”. The most significant digit is stored in CHK1 and the least
significant digit in CHK2.
[CR] [LF] Control code indicating end of frame

(4) If the receiver has successfully received the frame, and is prepared to receive the next frame, the
receiver responds with [ACK]. After the sender receives [ACK], the sender advances the frame
number and either sends a new frame or transition to the termination phase.

(5) If the receiver fails to receive the frame and is prepared to receive the same frame again, the receiver
responds with [NAK]. After the sender receives [NAK], the sender sends the most recent frame again,
using the same frame number. If a total of 6 attempts to send the frame failed, the sender
transitions to the termination phase and must end sending of the message.

13/123
Max. 15 sec. Max. 15 sec. E
First attempt Sixth attempt O
IPU T
Text frame#1 Text frame#1
・・・
N N
・・・
A A
Host K K
Computer

(6) The CN-series device processes the response of [EOT] from the Host computer as [ACK]. (Response
of [EOT] from the receiver is usually a request to suspend a transmission to the sender. However,
the CN-series device does not support this function.)

6.2.4. Termination phase


During the termination phase, the status returns to neutral.
The sender sends the [EOT] to inform the receiver that the message transmission has been completed.
When the sender sends [EOT], the sender transitions to neutral status. When the receiver receives
[EOT], the receiver transitions to neutral status.

6.2.5. Time out


The timer is used to detect a failure to coordinate between the sender and the receiver. The timer is
used as a mean of the recovery for communication line and communication destination device failure.

(1) During the establishment phase, the timer is set when the sender sends [ENQ]. The time out results
if a response of [ACK], [NAK] or [ENQ] is not received within 15 seconds. After time out, the sender
transitions to the termination phase.

(2) During transfer phase, the 15-second timer is set when the sender sends the final character of a
frame. Time out results if no response is received within 15 seconds. After time out, the sender
transitions to the termination phase.
The receiver sets a 30-second timer when first entering the transfer phase or when responding (with
either [ACK] or [NAK]) to a frame. Time out results if the receiver does not receive a frame or [EOT]
from the sender within 30 seconds. After time out, the receiver discards the latest incomplete
message and transitions to the termination phase.

14/123
6.3. Presentation Layer

6.3.1. Messages, Record and Field

6.3.1.1.Messages
In the presentation layer, all data is transmitted using messages. Messages are composed of record
arrays that start with message header record (H) and end with message termination record (L).

6.3.1.2.Records
A record is a series of text, beginning with an ASCII alphabet characters referred to as the identifier.

Table 6.3.1.2-1: Records


Record Type Record Level Contents
identifier
Header Record H 0 Contains the sender and the receiver
information.
Patient information Record P 1 Contains the patient information
Inquiry record Q 1 Requires test order information requesting to
the host computer
Test Order Record O 2 Contains test order information
Test Result Record R 3 Contains test result information
Comment Record C 1-4 Contains comment information
Manufacturer Information Record M 1-4 Not used
Scientific Information Record S N/A Not used
Message Terminator Record L 0 Indicates the end of the message

• A smaller level number indicates a higher level.


• A higher-level record contains information that is common to all lower-level records.
• All levels other than 0 must be located after higher levels. However, the comment record can be inserted
at any level. They are considered to be one lower level than the preceding record.
• Example transmission
H->P->O->R->L… Correct
H->R->L…………. Incorrect, because P and O must be transmitted in prior to R

15/123
6.3.1.3.Field
A record is further divided into multiple fields by field delimiters.
A field is identified by its position within a record and has a variable length. The followings are used as
delimiters.

Table 6.3.1.3-1: Fields


Type Code Description
Field delimiter Vertical bar (|)(7Ch) Separates adjacent field within a record
Repeat delimiter Back slash (\) (5Ch) Separates multiple numbers of descriptors in a field.
Component delimiter Caret ^ Separates data elements within a field that has a hierarchical
(5Eh) or qualifier nature.
Escape delimiter Ampersand &(26H) Is used within a text field to identify special case operations.

Message Record 1 Record 2 … Record n

Field 1 Field 2 … Field n CR


Record identifier Delimiter"|" or "\"

Component1 Component2 … Component n

Delimiter"^"

16/123
6.3.2. Communication Protocol

6.3.2.1.Analysis Order inquiry (CN-series device -> Host computer)


This protocol is used for CN-series device to inquire to the host computer analysis information to obtain
the sample information.
Inquiry can be made with keyword of the sample ID Number in CN-series device.

Table 6.3.2.1-1: Analysis Order Inquiry


CN-series device Direction Host computer
ENQ →
← ACK
H (Header Record) →
← ACK
Q (Inquiry Record) →
← ACK
L (Message Terminator Record) →
← ACK
EOT →

6.3.2.2.Analysis information (Host computer -> CN-series device)


This protocol is used for the host computer to respond analysis information against the inquiry made by
the CN-series device.

Table 6.3.2.2-1: Analysis Information


CN-series device Direction Host computer
← ENQ
ACK →
← H (Header Record)
ACK →
← P (Patient Record)
ACK →
← O (Test Order Record)
ACK →
← L (Message Terminator Record)
ACK →
← EOT

17/123
6.3.2.3.Analysis Results & QC Data (CN-series device -> Host computer)
This protocol is used for the CN-series device to transmit the analysis result, the QC data in a real-time
transmission (QC sample Number is “QCXX” and transmitted as similar to the regular sample data),
and the QC data in the manual output.

Table 6.3.2.3-1: Analysis results and QC data


CN-series device Direction Host computer
ENQ →
← ACK
H (Header Record) →
← ACK
P (Patient Record) →
← ACK
O (Test Order Record) →
← ACK
R: Result Record

(Replicating the No. of parameters)
(Replicating the No. of parameters) ← ACK (Replicating the No. of
receiving parameters)
C: Comment record

(Replicating the No. of parameters)
(Replicating the No. of parameters) ← ACK (Replicating the No. of
receiving parameters)
L (Message Terminator Record) →
← ACK
EOT →
* Comment codes can be set to Enabled or Disabled under [Service settings] - [Other service settings] -
[HOST_ASTM_OutputCommentRecord].

* If there are multiple assay group parameters subject to output, the parameter output sequence for
result records and comment records is in ascending order of the Host ID registered in [Assay Group
Settings]. If there are any duplicated host IDs, the sequence is the display sequence (Seq.) in [Assay
Group Settings].
If there are research parameters in the assay group parameters subject to output, the parameter
output sequence for result records and comment records is reportable parameters then research
parameters.
For details, refer to "7.4 Analysis Results (IPU => Host Computer)".

18/123
6.3.3. Details of Record

6.3.3.1.Header Record
[Example of transmission]
 CN-series device -> Host computer
(For the CN-6000)
H|\^&|||CN-6000^00-00^11001^^^USERINSNO^XXXXXXXX||||||||E1394-97<CR>
(For the CN-3000)
H|\^&|||CN-3000^00-00^11001^^^USERINSNO^XXXXXXXX||||||||E1394-97<CR>
 Host computer -> CN-series device
H|\^&|||||||||||E1394-97<CR>

Table 6.3.3.1-1: Details of Header record


ASTM CN-series device -> Host -> CN-series Max. Remarks
Field Field Name Host device size
(Byte)
7.1.1 Record type H H 1 Fixed
7.1.2 Delimiter definition |\^& |\^& 4 Fixed
7.1.3 Message control ID Not used Not used -
7.1.4 Access password Not used Not used -
7.1.5 Sender name or ID Analyzer name^ Not used 8^
Software version^ 13^
Analyzer serial No.^ 5^
Interface version^ ^
Logical ID number ^ ^
Nickname^ 15^
PS code 8
7.1.6 Sender street address Not used Not used -
7.1.7 Reserved field Not used Not used -
7.1.8 Sender Telephone Not used Not used -
No.
7.1.9 Sender Not used Not used -
characteristics
7.1.10 Receiver ID Not used Not used -
7.1.11 Comment Not used Not used -
7.1.12 Processing ID Not used Not used -
7.1.13 ASTM Version No. ASTM Version No. ASTM Version No. 8 E1394-97
7.1.14 Date and Time of Not used Not used -
message

[Detailed explanation of the fields]


1) 7.1.2 Delimiter definition
"|\^&" is used as a fixed character string. No field delimiter is required between 7.1.1 and 7.1.2.

2) 7.1.5 Sender name or ID


Analyzer name:
(For the CN-6000)
Set to “CN-6000” as fixed.
(For the CN-3000)
Set to “CN-3000” as fixed.
(For the CN-6500)
Set to “CN-6500” as fixed.
(For the CN-3500)
Set to “CN-3500” as fixed.

Software version:
Refers to the software version loaded on the CN-series device.

19/123
Analyzer Serial No.:
Analyzer serial number is output.

Interface version:
Not used

Logical ID number:
Not used

Nickname:
Analyzer nickname is output.

PS code:
(For the CN-6000)
Output the CN-6000 Model PS code from “Table 9.3 1: PS Codes for Each Destination (CN-6000,
CN-3000)”.
(For the CN-3000)
Output the CN-3000 Model PS code from “Table 9.3 1: PS Codes for Each Destination (CN-6000,
CN-3000)”.
(For the CN-6500)
Output the CN-6500 Model PS code from “Table 9.3-2: PS Codes for Each Destination (CN-6500,
CN-3500)”.
(For the CN-3500)
Output the CN-3500 Model PS code from “Table 9.3-2: PS Codes for Each Destination (CN-6500,
CN-3500)”.

3) 7.1.13 ASTM Version No.


The version number is “E1394-97”.

Caution:
Refer to [9.3 Difference List of Transmission Content] as the difference in the specification
of “Analyzer name”, “PS Code” and “ASTM Version No.” between CN-series device and
each device.

20/123
6.3.3.2.Patient Information Record
[Example of transmission]
 CN-series device -> Host computer
P|1||||^Heisei^Jiro<CR>
 Host computer -> CN-series device
P|1|||100|^Heisei^Jiro<CR>

Table 6.3.3.2-1: Details of Patient Information Record


ASTM CN-series device Host -> CN-series Max. size
Field Field Name -> Host *1 device (Byte) Remarks
8.1.1 Record type P P 1 Fixed
8.1.2 Sequence No. Sequence No. Sequence No. 4 Sequence No. of
records
8.1.3 Practice assigned Not used Not used -
Patient ID
8.1.4 Laboratory Not used Not used -
assigned patient ID
8.1.5 Patient ID No. Patient ID Patient ID 16
8.1.6 Patient name ^First name ^First name ^20^20
^Last name ^Last name
8.1.7 Mother's maiden Not used Not used -
name
8.1.8 Birth date Not used Not used -
8.1.9 Patient sex Not used Not used -
8.1.10 Patient race Not used Not used -
8.1.11 Patient address Not used Not used -
8.1.12 Reserved Not used Not used -
8.1.13 Patient telephone No. Not used Not used -
8.1.14 Attending Not used Not used -
physician name
8.1.15 Special field 1 Not used Not used -
8.1.16 Special field 2 Not used Not used -
8.1.17 Patient height Not used Not used -
8.1.18 Patient weight Not used Not used -
8.1.19 Patient’s known or Not used Not used -
suspected diagnosis
8.1.20 Patient active Not used Not used -
medications
8.1.21 Patient diet Not used Not used -
8.1.22 Practice field 1 Not used Not used -
8.1.23 Practice field 2 Not used Not used -
8.1.24 Admission and Not used Not used -
discharge dates
8.1.25 Admission status Not used Not used -
8.1.26 Location Not used Not used -
8.1.27 DRG or AVG Not used Not used -
8.1.28 DRG or AVG2 Not used Not used -
8.1.29 Patient religion Not used Not used -
8.1.30 Marital status Not used Not used -
8.1.31 Isolation status Not used Not used -
8.1.32 Language Not used Not used -
8.1.33 Hospital service Not used Not used -
8.1.34 Hospital institution Not used Not used -
8.1.35 Dosage category Not used Not used -
*1 To output the QC data, all fields other than 8.1.1 “Record Type” and 8.1.2 “Sequence number” are not
used.

21/123
[Detailed Explanation of the fields]
1) 8.1.2 Sequence No.
The sequence number starts with 1 and indicates sequence position in which the record appeared in the
message. This number is reset to 1 when a higher-level record appears in the message.

2) 8.1.5 Patient ID Number


This is the ID that is unique to the patient. It can use 16 alphanumeric characters and hyphens “-”
(2Dh).

3) 8.1.6 Patient name


The first name and the last name may be with a maximum of 20 characters each with consisting of
alpha-numeric.

22/123
6.3.3.3.Request Information Record
[Example of transmission]
 CN-series device -> Host computer
Q|1|1^1^ 1234567890^B||^^^040^PT|0|20010905150000||||P||N<CR>
 Host computer -> CN-series device
Not used

Table 6.3.3.3-1: Details of Request Information Record


ASTM Field Name CN-series device -> Host Host -> Max. size
Field CN-serie (Byte) Remarks
s device
12.1 1 Record Type Q Not used 1
12.1.2 Sequence No. 1 Not used 4
12.1.3 Starting Range ID Rack No.^ Not used 6^
No. Tube position^ 2^
Sample ID No.^ 15^
Sample ID No. attribute 1^
12.1.4 Ending Range ID Not used Not used -
No.
12.1.5 Universal test ID ^^^Test code (see in Not used ^^^3 * Only parameters
Table 8.1-1: Parameter ^10 that can be measured
Code Lists.) are set as an inquiry
Test name item when
When requesting multiple [HOST_ASTM_OrderI
parameters, use the nquiry_AnalyzableAss
repeat symbol as below: ay] setting is enabled
Test parameter for [Service settings] –
information1 [Other service
\Test parameter settings].
information 2\Test
parameter information
3\Test parameter
information4\…
12.1.6 Nature of request 0 Not used 1 0: Request date
day limit Request date is not
defined in the ASTM.
12.1.7 Beginning request YYYYMMDDHHMMSS Not used 14
date and time
12.1.8 Ending request Not used Not used -
result date
12.1.9 Requesting Not used Not used -
Physician Name
12.1.10 Requesting Not used Not used -
Physician
Telephone
Number
12.1.11 User field No. 1 Platelet aggregation Not used (1)
analysis: P
Cross mixing (immediate
type) analysis: C
Other: Not used
12.1.12 User Field No. 2 Not used Not used -
12.1.13 Requested (Order type)* Not used (1)* * The first
Information measurement/re-analysis
Status Codes instruction only has a set
value when [System
Settings] - [Device
Settings] - [Host] -
[Inquire re-analysis]
setting is enabled.
23/123
[Detailed Explanation of the field]
1) 12.1.2 Sequence No.
The sequence number starts with 1 and indicates the sequence position in which the record appeared in
the message. The number is reset to 1 when a higher-level record appears in the message.

2) 12.1.3 Starting Range ID No.


Rack No.:
Assigned to the rack. STAT table, reagent table are described as below.
Normal sample: Consists of 6-digit alpha-numeric (The rack number will be fixed
6-digit number by aligning to the right and padding zero’s to the most significant
digits if the number is less than 6-digit.)
STAT sample: Describes with STATΔΔ (Δ indicates a space character.)
LAS sample: Describes with L□□□□□ (□ indicates a numeric character. The rack
number will be fixed 5-digit number by aligning to the right and padding zeros to
the most significant digits if the number is less than 5 digits.)*
* Transfer samples analyzed with the Sysmex transportation system (the connection
type under [System Settings] - [LAS Connection] is Type 6) are the same as
ordinary samples.
The number becomes L□□□□□ (□ indicates a 5-digit alphanumeric character) when
[Service settings] - [Other service settings] - [LAS_RackNo_Replace] is set to
“Enabled”.

Caution:
Refer to [9.3 Difference List of Transmission Content] as the difference in the specification
of “Rack No.” between CN-series device and each device.

Tube Position:
The sample rack position number within a rack, with consisting of 2-digit number.
Normal sample: 01 - 06
STAT sample: ΔΔ (Δ indicates a space character.)
LAS sample: Fixed to 01.*
* Transfer samples analyzed with the Sysmex transportation system (the connection
type under [System Settings] - [LAS Connection] is Type 6) are 01 - 10.

Caution:
Refer to [9.3 Difference List of Transmission Content] as the difference in the specification
of “Tube Position” between CN-series device and each device.

Sample ID number:
Consists of 15-digit of alpha-numeric. Depends on the direction for use, the hyphen “-” (2DH)
can be inserted between characters. “-” is included in 15-digit number.
The zero suppression function is not performed. If the number is less than 15-digit, it is aligned
to the right and spaces (20H) are padded to the most significant digit. If the sample barcode
cannot be read, the sample ID number beginning with "ERR000000000001” is sequentially
assigned to the sample number. The sample number beginning with “QC” is reserved to use for
the QC analysis.

Sample ID No. attribute:


M: Sample ID manually entered.
Sample ID No. is manually entered through the touch panel or the IPU keyboard.
A: Analyzer automatically assigned number.
This number is assigned by the automatic-increment function, and is used to set the
sample number that begins with "ERR", which is used when the ID Read error
occurred.
B: Barcode reader read number.
This is used when the sample ID number is read by the ID barcode reader.

3) 12.1.5 Universal test ID

24/123
Test code:
Suffixing 0 to the “Host ID” in the “Basic” in the "Assay Group Setting" makes the test code.
See the Table 8.1-1: Parameter Code Lists.
※Only for CN-6500/CN-3500
For immuno parameter, suffixing 0 to the “Host ID” in the “Basic” in the " Parameter info,
settings " in the “Immuno module” makes the test code.

Test name:
The test name is the “Displayed Name” in the “Basic” in the “Assay Group Setting”.
It is consisted of up to 10 characters. See the Table 8.1-2: Parameter Code Lists(Result).
※Only for CN-6500/CN-3500
For immuno parameter, the test name is the “Displayed Name” in the “Basic” in the "
Parameter info, settings " in the “Immuno module”.

Note:
The measurement of platelet aggregation does not correspond to orders from the host
computer. In case reagents of the measurement of platelet aggregation (ADP, collagen,
epinephrine, arachidonic acid, and ristocetin) are set, parameter codes of the
measurement of platelet aggregation are transmitted. Avoid ordering the measurement of
platelet aggregation from the host computer.

4) 12.1.7 Beginning request result date and time


The date format is fixed to “YYYYMMDDHHMMSS”.
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system (00-23),
MM the minute (00-59), SS the second (00-59).

5) 12.1.13 Requested information status code


Inquiry type:
N: Real time inquiry of the first analysis
C: Real time inquiry of re-analysis
* Either N or C is selectable only when the “Device Settings” - “Host” - “Inquire re-analysis” setting in
the “System Settings” is ON.

※Only for CN-6500/CN-3500


CN-6500/CN-3500 doesn’t support with real time inquiry of re-analysis for immuno parameter.

(Examples of Communication)
First measurement order: Q|1|000017^01^000000000000001^A||^^^040^PT1|0|20150116181548|||||
|N<CR>
Re-test order: Q|1|000017^01^000000000000001^A||^^^040^PT1|0|20150116181548||||||C<CR
>

Notes:
・Requested information record for the C(Inquiry of re-analysis) is transmitted only when
it satisfies the following conditions ;
・The measurement mode is “Standard”mode
・The “Device Settings” - “Host” - “Target” of “Automatic Output” in the System Settings
is “All Data”
・The analysis is not QC analysis nor Calib.curve analysis
・The instrument is in measurable mode.
・The analysis result data including the parameters for “Re-analysis” in the assay group
settings has been already output.
※Only for CN-6500/CN-3500
・First measurement order is not for only immuno parameter.

・Select 9600bps or more of Baud Rate with“Ethernet connection”or“Serial connection”,


because the communication data volume increases when the “Device Settings” - “Host” -
“Inquire re-analysis” setting in the“System Settings” is ON.
25/123
Notes:
When the “Inquire re-analysis” is used, set up auto re-analysis, auto re-dilution and reflex
in the assay group settings of CN-series device in the same conditions with the re-analysis
logic in the Host Computer (because aspiration volume increases with the re-analysis).

If the assay group settings of CN-series device are wrong, when the re-analysis is ordered
by Host Computer, the error “Insufficient Sample”may be shown.

26/123
6.3.3.4.Test Order Record
[Example of transmission]
 CN-series device -> Host computer
O|1||000001^01^123456789012345^B^E||R||||||N<CR>
 Host computer -> CN-series device
O|1|000001^01^123456789012345^B||^^^040^^100.00\^^^050^^100.00\・・・
\^^^060^^100.00|R|20010807101000|||||N<CR>

Table 6.3.3.4-1: Details of Test Order Record


ASTM CN-series device Host computer -> Max.
Field Field name -> Host computer CN-series device size Remarks
(Byte)
9.4.1 Record type O O 1 Fixed
9.4.2 Sequence No. Sequence No. Sequence No. 4 Sequence No. of records
9.4.3 Specimen ID Not used Rack No.^ 6^ Sample ID No.
Tube Position^ 2^ attribute is one of the
Sample ID No.^ 15^ followings:
Sample ID No. 1(^^1) M: Manual entry
attribute(^^Meas * A: Automatically
urement Mode)* assigned by analyzer
9.4.4 Instrument Rack No.^ Not used 6^ B: Barcode reader
specimen ID Tube Position^ 2^
Sample ID No.^ 15^ *Measurement Mode
Sample ID No. 1^ can be set only when
attribute^ 1 "Service settings" -
Extended order 1 "Other service settings" -
request flag (^1)* "HOST_ASTM_Specify
(^Measurement (^4)* MeasurementMode"
Mode)* setting is ON.
(^Sample seq.)* *Sample seq. is only
usable when the setting
of "Service settings" -
"Other service settings" -
"HOST_ASTM_Output
SampleSeq" is enabled.
9.4.5 Universal Test ID Not used ^^^Test code ^^^3 *Replication can be set
^^Dilution ^^6^3 only when "Service
ratio^Option(^Re (^1)* settings" - "Other
plication)* (^1)* service settings" -
(^Measurement "HOST_ASTM_Specify
Time)* Replication" setting is
ON.
When requesting
multiple *Measurement Time
parameters, use can be set only when
the repeat symbol "Service settings" -
as below: "Other service settings" -
Test parameter "HOST_ASTM_Switch
information1 MeasurementTime"
\Test parameter setting is ON.
information2\Test
parameter
information
3\Test parameter
information4\…
9.4.6 Priority Priority Priority 1 S(STAT, STAT sample)
R(Routine,
Normal sample)
9.4.7 Requested/order Not used YYYYMMDDHH 14
date and time MMSS

27/123
ASTM CN-series device Host computer -> Max.
Field Field name -> Host computer CN-series device size Remarks
(Byte)
9.4.8 - Specimen Not used Not used -
9.4.11 collection date and
time
through
Collector ID
9.4.12 Action code Action code Action code 1 N Normal sample
Q QC material
9.4.13 Danger code Not used Not used -
- through
9.4.31 Material
institution

[Detailed explanation of the field]


1) 9.4.2 Sequence No.
The sequence starts with 1 and indicates the sequence position in which the record appeared in the
message.
This number is reset to 1 when a higher-level record appears in the message.

2) 9.4.3 Specimen ID
Rack No.:
Up to 6-digit alpha-numeric assigned to the rack.
Return the same number that was inquired.

Tube Position:
The sample position number with in a rack with consisting of 2-digit number.
Return the same number that was inquired.

Sample ID number:
Consists of 15-digit of alpha-numeric and hyphen “-“(2Dh).
Return the same number that was inquired.

Sample ID No. attribute:


Indicates how the sample ID No.was registered.
Return the same number that was inquired.

Measurement Mode:
Specifies sampling mode.
N: Normal Mode
M: Micro-sampling Mode
Other than above value: Sample is aspirated according to the instrument setting.

Analysis mode cannot be ordered from host computer for QC Analysis or transfer samples.
Analysis mode instructions from the host computer for QC Analysis or transfer samples are
ignored.

If the measurement mode is changed from Assay Group setting on the instrument (manual
order registration or preset order), the sample is aspirated according to the specified setting
from the instrument prior to Host Computer specification.

If this block is missing, analyze according to instrument settings.

* Analysis modes can be set to Enabled or Disabled under [Service settings] - [Other service
settings] - [HOST_ASTM_SpecifyMeasurementMode].
The default setting is “Disabled”.

28/123
(Examples of Communication)
<Host computer→CN-series device>
When disabled:
O|1|000017^01^000000000000001^A||^^^040^^^^2|R|20140620183015|||||N|||||||||||||||
||||<CR>
When enabled:
O|1|000017^02^000000000000002^A^^M||^^^040^^^^2|R|20140620183015|||||N|||||||||||
||||||||<CR>

Note:
When the “Device Settings” - “Host” - “Inquire re-analysis” setting in the “System
settings”is ON and in the case of test order record for the inquiry of re-analysis, it is
impossible to specify Micro-sampling mode. (The mode is not reflected.)

3) 9.4.4 Instrument Specimen ID


Rack No.:
Assigned to the rack. In the reagent table, Rack No. is described as below.
Normal sample: 6-digit alpha-numeric (The rack number will be fixed 6-digit
number by aligning to the right and padding zero’s to the most significant digits if
the number is less than 6-digit.)
STAT sample: STATΔΔ (Δ indicates a space character.)
Reagent table: REAG00

* In the case of overlay graph information output for the platelet aggregation
parameter and cross mixing, information for the sample with the oldest
measurement date and time is output on the Graph screen from among the results
set as valid.

Caution:
Refer to [9.3 Difference List of Transmission Content] as the difference in the specification
of “Rack No.” between CN-series device and each device.

Note:
“Platelet aggregation parameter” means the parameter (Aggregation %) calculated from
the measurement results for PPP samples and PRP samples.

Tube Position:
The sample position number with in a rack with consisting of 2-digit number.
Normal sample: 01 - 06
STAT sample: ΔΔ (Δ indicates a space character.)
Reagent table: ΔΔ (Δ indicates a space character.)

* In the case of overlay graph information output for the platelet aggregation
parameter and cross mixing, information for the sample with the oldest
measurement date and time is output on the Graph screen from among the results
set as valid.

Caution:
Refer to [9.3 Difference List of Transmission Content] as the difference in the specification
of “Tube Position” between CN-series device and each device.

Note:
“Platelet aggregation parameter” means the parameter (Aggregation %) calculated from
the measurement results for PPP samples and PRP samples.

29/123
Sample ID number:
Consists of 15-digit alpha-numeric. Depends on the direction for use, the hyphen “-“(2DH) can
be inserted between characters. “-“is included in 15-digit number. The zero suppression
function is not performed. If the number is less than 15-digit, it is aligned to the right and
spaces (20H) are padded to the most significant digit. If the sample barcode cannot be read, the
sample ID number beginning with "ERR000000000001” is assigned sequentially to the sample
number.
The sample number beginning with “QC” is reserved to use for the QC analysis.

Sample ID No. attribute:


M: Sample ID manually entered.
Sample ID No. is manually entered through the touch panel or the IPU keyboard.
A: Automatically assigned number by the analyzer.
This number is assigned by the automatic-increment function, and is used to set the
sample number that begins with “ERR”, which is used when the ID Read error
occurred.
B: Barcode reader read number.
This is used when the sample ID number is read by the ID barcode reader.

* In the case of overlay graph information output for the platelet aggregation parameter and
cross mixing, information for the sample with the oldest measurement date and time is
output on the Graph screen from among the results set as valid.

Extended order request flag:


E: Extended order
Indicates the analysis is continued, based on the rules of the analyzer.
If the analysis is not continued, it is skipped.

* In the case of overlay graph information output for the platelet aggregation parameter and
cross mixing, information for the sample with the oldest measurement date and time is
output on the Graph screen from among the results set as valid.

Note:
The flag is not output if the “Device Settings” - “Host” - “Inquire re-analysis” setting
in the“System Settings”is ON.

Measurement Mode:
Indicates sampling mode.
N: Normal Mode
M: Micro-sampling Mode

* In the case of overlay graph information output for the platelet aggregation
parameter and cross mixing, information for the sample with the oldest
measurement date and time is output on the Graph screen from among the results
set as valid.

* Measurement Mode can be set only when “Service settings” - “Other service settings” -
“HOST_ASTM_SpecifyMeasurementMode” setting is ON. Default setting is OFF.

(Examples of Communication)
<CN-series device → Host computer>
When disabled: O|1||000017^01^000000000000001^A^^||R||||||N<CR>
When enabled: O|1||000017^01^000000000000001^A^^M||R||||||N<CR>

Sample seq.:
Output sequence numbers, which are numbered in sequence from 1 when an order for ordinary
samples or STAT samples is registered.
If the number of digits is less than 4, the number will be aligned to the right and padded with
zeros.

30/123
If the following Sample seq. information is missing, the values in the Sample seq. field are not
output.
- Quality control analysis
- Overlay graph information for platelet aggregation parameters and cross mixing

* Sample seq. output can be set to enabled or disabled under “Service settings” - “Other service
settings” - “HOST_ASTM_OutputSampleSeq”.
The default setting is “Disabled”.

(Examples of Communication)
<CN-series device → Host computer>
When disabled: O|1||000017^01^000000000000001^A^^M||R||||||N<CR>
When enabled: O|1||000017^01^000000000000001^A^^M^0001||R||||||N<CR>

4) 9.4.5 Universal Test ID


When an order is sent from the host computer to the CN-series device, set the Universal Test ID which
is sent in the “Universal Test ID” in the section “4.3.3.3 Details the Request Information Record”. If a
parameter code is received that is not sent with the “12.1.5 Universal Test ID”, measurement is not
performed for that parameter. If there are no parameters, it can be skipped.

To notify the user when the parameter code 000 (no measurement parameters) is received from the host
in the order inquiry for the first measurement, it is necessary to enable the setting for “Service settings”
- “Other service settings” - “HostOrderCaution”.

Note:
The measurement of platelet aggregation does not correspond to orders from the host
computer. The IPU does not measure the platelet aggregation orders even though
parameter codes of platelet aggregation are set. For this reason, maximum aggregation
rate and evaluation parameters are not calculated.

※Only for CN-6500/CN-3500


Note:
When the “Device Settings” - “Host” - “Inquire re-analysis” setting in the “System se
ttings”is ON and in the case of test order record for the inquiry of re-analysis, it is
impossible to specify immuno parameter. (The mode is not reflected.)

Test code:
Suffixing 0 to the “Host ID” in the “Basic” in the “Assay Group Setup” makes the test code. See
the Table 8.1-1: Parameter Code Lists.
※Only for CN-6500/CN-3500
For immuno parameter, suffixing 0 to the “Host ID” in the “Basic” in the " Parameter info,
settings " in the “Immuno module” makes the test code. See the Table 8.2-1:Parameter Code
Lists.

Dilution ratio:
Specifies the dilution ratio used in analysis with percent. If it is not specified, follows the
instrument setting. The dilution ratio can be specified with including 2-digit of decimal
number.
The dilution ratio in the CN-series device which differs from the ordered dilution ratio by less
than 0.01 percent is used for analysis.
Note: If the MDA is set in the “Default Dilution Ratio” in the “Basic” in the “Assay Group
Setup”, the MDA analysis is performed in disregard of the dilution ratio from the host
computer (The same is true for cross mixing).
* If the dilution ratio was omitted in instructions, the dilution ratio set in [Assay Group Setup]
- [Basic] - [Default Dilution Ratio] is used to perform the measurement.
※Only for CN-6500/CN-3500

31/123
For immuno parameter, the dilution ratio set in [Immuno module] – [Parameter info, settings]
- [Basic] - [Dilution Ratio] is used to perform the measurement.

Warning:
The setting of the dilution ratio may have a critical impact on analysis results.
Set the standard setting of the dilution ratio to 100 percent. If other than 100
percent is set to the dilution ratio, make sure that it is the correct value.

Note:
The redilution, re-analysis and reflex test are performed according to the
instrument setting even if the analysis is specified at the dilution ratio different
from the instrument setting. If the redilution, re-analysis and reflex test are not
performed, specify the option from the host computer.

32/123
Option:
Sets the performance of the redilution, re-analysis and reflex test. More than 1 option can be
set with the combination of “D”, “R” and “F”. If there are no settings, follows the instrument
settings
D: No redilution
R: No re-analysis
F: No reflex test

Replication:
Sets the number of times the test is performed. “1” or “2” can be set.
If replication is specified “2” or over, the analysis is performed twice (2).
Other than above value: Sample is measured according to instrument setting.
In the case of QC sample, it is not possible to specify the replication from Host Computer.
If this block is missing, analyze according to instrument settings.
※Only for CN-6500/CN-3500
Sets the number of times the test is performed. “1” or “2” can be set.
If the sum of replication is specified “6” or over in the case of the order included some immuno
parameters for one sample, the analysis is performed fifth (5).

* Analysis modes can be set to Enabled or Disabled under [Service settings] - [Other service
settings] - [HOST_ASTM_SpecifyReplication].
The default setting is “Disabled”.

(Examples of Communication)
When disabled:
O|1|000017^01^000000000000001^A||^^^040^^^^|R|20140620183015|||||N|||||||||||||||
||||<CR>
When enabled:
O|1|000017^02^000000000000002^A||^^^040^^^^2|R|20140620183015|||||N|||||||||||||||
||||<CR>
Note:
When the “Device Settings” - “Host” - “Inquire re-analysis” setting in the “System
settings” is ON and in the case of the test order record for inquiry of re-analysis, it
is impossible to specify replication.

Measurement Time:
Set measurement time.If there are no settings, follow the instrument settings.
M: Measurement Time(Main)
S: Measurement Time(Sub)
* Measurement time can be set to Enabled or Disabled under [Service settings] – [Other
service settings] – [HOST_ASTM_SwitchMeasurementTime].
The default setting is “Disabled”.
※Only for CN-6500/CN-3500
For immuno parameter, the behavior of the measurement time is always “Disable” setting.

(Examples of Communication)
When disabled:
O|1|000017^01^000000000000001^A||^^^040^^^^|R|20140620183015|||||N|||||||||||||||
||||<CR>
When enabled:
O|1|000017^02^000000000000002^A||^^^040^^^^^M|R|20140620183015|||||N|||||||||||||
||||||<CR>

5) 9.4.6 Priority
S:STAT, STAT sample
R:Routine, Normal sample

6) 9.4.7 Requested/order date and time


Indicates the date and time of the analysis for the inquiry sample.
The date format is fixed to "YYYYMMDDHHMMSS".

33/123
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system (00-23),
MM the minute (00-59), SS the second (00-59).

7) 9.4.12 Action code


Indicates the contents of the result record to be sent.
N: Normal sample
Q: QC sample

34/123
6.3.3.5.Result Record
[Example of transmission]
 CN-series device -> Host computer
R|1|^^^041^PT sec^100.00^9^^^|10.2|sec||N||||||20110328135056<CR>
R|2|^^^042^PT %^100.00^9^^^| 99.4|%||N||||||20110328135056<CR>
……
Host computer -> CN-series device
Not used

Table 6.3.3.5-1: Details of the Result record


ASTM CN-series Host Max. size
Field Field name device -> Host computer (Byte) Remarks
computer ->
CN-series
device
10.1.1 Record type R Not used 1
10.1.2 Sequence No. Sequence No. Not used 4 Sequence No.
beginning with 1
10.1.3 Universal Test ID ^^^Test code^ Not used ^^^3^ Test code: Described
Parameter 24^6^ in "Table 8.1-1:
^Dilution ratio^ 1^ Parameter Code
Analysis result 1 Lists" below.
type^Extended ^1^1(^2) Parameter: Up to 24
order request^ * characters
Extended order
result^ *Replication can be
Reflex test output only when
request(^Repli "Service settings" -
cation)* "Other service settings" -
"HOST_ASTM_Specif
yReplication" setting
is ON.
10.1.4 Data or measurement Value Not used 6 (In case Up to 6 characters
value of the (In case of the path
path name to the image
name to data, up to 60
the image characters)
data, 60)
10.1.5 Units Units Not used 7
10.1.6 Reference ranges Not used Not used -
10.1.7 Result abnormal flags Result Not used 1^Not Evaluation Info. and
abnormal flags specified^ Instrument error
^Evaluation Not information are
Information^ specified output only when
Instrument "Service settings" -
Error "Other service settings" -
Information "HOST_ASTM_Outp
utErrorInformation"
setting is ON.
10.1.8 Nature of abnormality Not used Not used -
testing
10.1.9 Result status (Result Not used 1 *Result Status can be
Status)* output only when
"Service settings" -
"Other service settings" -
"HOST_ASTM_Outp
utItemUnit" setting
is ON.
10.1.10 Date of change in Not used Not used -
instrument normative
values

35/123
ASTM CN-series Host Max. size
Field Field name device -> Host
computer (Byte) Remarks
computer ->
CN-series
device
10.1.11 Operator identification Not used Not used -
10.1.12 Date/Time test started Not used Not used -
10.1.13 Date/Time test completed YYYYMMDDH Not used 14
HMMSS
10.1.14 Instrument identification Not used Not used -

[Detailed explanation of the field]


1) 10.1.2 Sequence No.
The sequence number starts with 1 and indicates the sequence position in which the record appeared in
the message. This number is reset to 1 when a higher-level record appears in the message.

2) 10.1.3 “Universal test ID” through 10.1.13 “Date time test completed”
Fields to be set will vary depending on the contents to be transmitted.
2.1) When test parameters are output
2.2) When evaluation parameters of platelet aggregation are output
2.3) When Sample Info. flags are output
2.4) When the path name to the image data is output
* Output of 2.2), 2.3) and 2.4) depends on the instrument setting.
* Result records from 2.1) to 2.4) are output continuously.

2.1) When test parameters are output: The parameters for which analysis orders are registered are
output.

a) 10.1.3 Universal Test ID


Test code:
Indicates the assay parameter.
Consists of the combination of the “Host ID” in the “Basic” in the “Assay Group Setup” and the
“Host ID” in the “Assay Parameter" in the “Assay Group Setup”. See the Table 8.1-1:
Parameter Code Lists.
※Only for CN-6500/CN-3500
For immuno parameter, consists of the combination of the “Host ID” in the “Basic” in the "
Parameter info, settings " in the “Immuno module” and “1: Count” or “2: Concentration”. See
the Table 8.2-1:Parameter Code Lists.

Parameter:
Consisted of up to 24 characters. The characters in the “Display Name” in the “Basic” in the
“Assay Parameter” in the “Assay Group Setup” are output.
See the Table 8.1-2: Parameter Code Lists(Result).
※Only for CN-6500/CN-3500
For immuno parameter, the test name is the “Displayed Name” in the “Basic” in the "
Parameter info, settings " in the “Immuno module”.
See the Table 8.2-2: Parameter Code Lists (Result).

Dilution ratio:
Outputs the dilution ratio used in analysis with percent. It is indicated with including 2-digit of
decimal number.
* If the MDA is set in the “Default Dilution Ratio” of the “Basic” in the “Assay Group Setup”,
the sub- field of the dilution ratio in the final report is skipped.

Analysis result type:


1: Normal (Auto output)
2: Average (Auto output)

36/123
3: Re-analysis (Auto output)
4: Average of re-analysis (Auto output)
5: Normal (Manual Output)
6: Average (Manual Output)
7: Re-analysis (Manual Output)
8: Average of re-analysis (Manual Output)
9: Final information (Auto output)
A: Final information (Manual Output)

Extended order request:


Indicates the redilution analysis and re-analysis are performed by the analyzer, based on the
rules of the analyzer.
Other than that, the extended order request is not added.
Only the First Measurement result that matches the Redilution Analysis and Perform
re-analysis conditions is output.
All assay parameters in the target assay group are flagged.
D: Redilution request
R: Re-analysis request

Extended order results:


Indicates the analysis result is one of the followings: the result of the redilution analysis,
reanalysis or reflex test. Other than that, the result is not output.
D: Results of redilution analysis
R: Results of re-analysis
F: Results of reflex test

Reflex test request:


Indicates the reflex test is performed by the analyzer, based on the rules of the analyzer. Other
than that, the extended order request is not added.
Only the first measurement result that matches the reflex analysis implementation condition is
output.
All assay parameters in the target assay group are flagged.
F: Reflex test request
Note:
The extended order and reflex test request are not output when the the “Device Settings” - “Host” -
“Inquire re-analysis” setting in the “System Settings”is ON.

Replication:
Indicates the number of times the test is performed.
* Replication can be set only when “Service settings” - “Other service settings” -
“HOST_ASTM_SpecifyReplication” setting is ON. Default setting is OFF.

(Examples of Communication)
When disabled: R|1|^^^041^PT sec^100.00^A^^^^|****.*|sec||A||||||20150116172743<CR>
When enabled: R|1|^^^041^PT sec^100.00^A^^^^1|****.*|sec||A||||||20150116172743<CR>

b) 10.1.4 Data or measurement value


Up to 6-digit number. Data format is described in the Table 8.1-2: Parameter Code Lists(Result).
If the analysis results cannot be obtained, the mask characters are input in part of the integral
number. The mask characters are as follows.
*: Analysis failure
/: Average calculation failure
+: Display Digit Overflow
-: Calculation failure
X: No validated calibration curve in the calculation
※Only for CN-6500/CN-3500

37/123
For immuno parameter, data format is the below. Data format is described in the Table 8.2-2:
Parameter Code Lists (Result)」.
Result: Concentration
Up to 10-digit number. Data is "integer part + decimal point "." (2eH) + decimal part". If
data is 11-digit number or over, round off from the minimum digit. And, if the minimum is
a decimal point, remove the decimal point.
The settings of display range and decial digits are in “Parameter info, settings”. When the
value is over the upper limit "999,999,999.0" in the device, "++++ (2bH is 4 characters)" is
output. If the value is out of the display range, a number will be output for this data and
the error information will be output with the result abnormal flag.
Result: Count (raw data with the device)
Set the count value of the measurement result. It consists of numbers only and does not
use commas (2ch). If the analysis results cannot be obtained, the mask characters are
input in part of the number.

c) 10.1.5 Units
Up to 7 characters. See the Table 8.1-2: Parameter Code Lists(Result).
In the CN-series device, the units set in the “Assay Parameter” in the “Assay Group Setup” are output.
(The ASTM describes the abbreviation of ISO standard should be used.)
See the Table 8.1-2: Parameter Code Lists(Result). If the unit is “Ratio” or “INR”, they are not output.
※Only for CN-6500/CN-3500
For immuno parameter, the unit is in the " Parameter info, settings " in the “Immuno module”.
See the Table 8.2-2: Parameter Code Lists (Result). If the unit is “Count”, it is not output.

d) 10.1.7 Result abnormal flag


Result abnormal flag:
L: Below the lower patient limit
H: Above the upper patient limit
<: Below lower report limit.
>: Above upper report limit
N: Normal test result
A: Analysis error

Evaluation Information:
Outputs a string of characters consisting of the combination of the evaluation error code and
error message, which is enclosed between square brackets “[]“. If there is more than one
evaluation information, they are separated by commas.
If there is no evaluation information, it is not output.
Refer to “Table 6.3.3.5-2: Content of analysis errors, measurement result error flags and data
values, and responses” about analysis errors.

Example of output) |A^[0008.0001.0000 Initial fluctuation drop],[0008.0002.0000 Coagulation Curve


Error : Sharp Drop]^|

* Evaluation information and instrument error information can be set to Enabled or Disabled under
[Service settings] - [Other service settings] - [HOST_ASTM_OutputErrorInformation].
The default setting is “Disabled”.
* For some analysis errors, the data value is output as either masked text or numeric values
depending on the content of the [Service settings] - [Other service settings] -
[HOST_ASTM_OutputCurveErrorInformation] setting. Refer to “Table 6.3.3.5-2: Content of
analysis errors, measurement result error flags and data values, and responses”.

Instrument Error Information:


Outputs a string of characters consisting of the error code and error message, which is enclosed
between square brackets “[]”.
If there is no Instrument error information, it is not output.

38/123
Example of output) |A^^[34422 Insufficient Reagent (Reagent Arm C Liquid Surface Not Detected)]|

* Evaluation information and instrument error information can be set to Enabled or Disabled under
[Service settings] - [Other service settings] - [HOST_ASTM_OutputErrorInformation].
The default setting is “Disabled”.

(Examples of Communication)
When disabled: R|1|^^^041^PT sec^100.00^A^^^^|****.*|sec||A||||||20150116172743<CR>
When enabled: R|1|^^^041^PT sec^100.00^A^^^^|****.*|sec||A^^[11050 Mechanical Stop button
is pressed.]||||||20150116172743<CR>

Table 6.3.3.5-2: Content of analysis errors, measurement result error flags and data values, and responses
Data value content
Analysis
The setting for The setting for
Code Messages result
error information error information
abnormal flag
output is OFF output is ON
Numeric values Numeric values
0000.0000.0000 [Dilution Ratio was Changed.] N
are output are output
Numeric values Numeric values
0000.0000.0340 [Wave Changed 340nm] N
are output are output
Numeric values Numeric values
0000.0000.0405 [Wave Changed 405nm] N
are output are output
Numeric values Numeric values
0000.0000.0575 [Wave Changed 575nm] N
are output are output
Numeric values Numeric values
0000.0000.0660 [Wave Changed 660nm] N
are output are output
Numeric values Numeric values
0000.0000.0800 [Wave Changed 800nm] N
are output are output
[Measured with a reagent without Numeric values Numeric values
0000.0000.1100 A
quality control analysis] are output are output
Masked values Masked values
0000.4000.0010 [Display Digit Overflow] N
are output are output
Masked values Masked values
0000.4000.0020 [Calculation Failure] N
are output are output
[Validated Calibration Curve is not Masked values Masked values
0000.4000.0030 N
found.] are output are output
Masked values Masked values
0000.4000.0040 [Extrapolation Boundary Over] N
are output are output
[Extrapolation Boundary Over (Upper
0000.4000.0041 > Upper limit value Upper limit value
Limit)]
[Extrapolation Boundary Over (Lower
0000.4000.0042 < Lower limit value Lower limit value
Limit)]
[Calib. Curve Boundary Over (Upper
0000.4000.0043 > Upper limit value Upper limit value
Limit)]
[Calib. Curve Boundary Over (Lower
0000.4000.0044 < Lower limit value Lower limit value
Limit)]
Numeric values Numeric values
0000.4000.0050 [Calculated later.] N
are output are output
Masked values Masked values
0000.5000.0010 [Formula calculation failed.] N
are output are output
Masked values Masked values
0000.5000.0020 [Can't perform formula calculation] N
are output are output
Masked values Masked values
0000.5000.0050 [Problem found in formula calculation] N
are output are output
0000.6000.0100 [Upper Report Limit Over] > Upper limit value Upper limit value
0000.6000.0200 [Lower Report Limit Over] < Lower limit value Lower limit value
Numeric values Numeric values
0000.6000.0300 [Upper Mark Limit Over] H
are output are output
Numeric values Numeric values
0000.6000.0400 [Lower Mark Limit Over] L
are output are output
Numeric values Numeric values
0000.6000.0500 [MDA Slope Ratio Over] A
are output are output
Numeric values Numeric values
0000.6000.0600 [Difference between replicated results] A
are output are output
Masked values Masked values
0000.7000.0020 [Mean Calculation Failure] N
are output are output

39/123
Data value content
Analysis
The setting for The setting for
Code Messages result
error information error information
abnormal flag
output is OFF output is ON
Masked values Masked values
0001.0000.0000 [Detector Block Temperature Error] A
are output are output
Numeric values Numeric values
0002.0000.0000 [Slight Coagulation] A
are output are output
Numeric values Numeric values
0004.0000.0000 [Analysis Time Over] A
are output are output
Masked values
0008.0001.0000 [Initial fluctuation drop] A ※1
are output
Masked values
0008.0002.0000 [Coagulation Curve Error: Sharp Drop] A ※1
are output
Masked values
0008.0008.0000 [Coagulation Curve Error: Jump Up] A ※1
are output
[Coagulation Curve Error: Stepping Masked values
0008.0016.0000 A ※1
Curve] are output
[Coagulation Curve Error: Fbg Curve Masked values
0008.0032.0000 A ※1
Error] are output
Masked values
0008.0064.0000 [Coagulation Curve Error: Terrace] A ※1
are output
Masked values Masked values
0008.0128.0001 [Early Reaction Error :Slow Reaction] A
are output are output
Masked values Masked values
0008.0128.0002 [Early Reaction Error :Start Angle 1] A
are output are output
Numeric values Numeric values
0008.0128.0004 [Early Reaction Error :Start Angle 2] A
are output are output
Masked values Masked values
0008.0128.0016 [Early Reaction Error :Early %] A
are output are output
Masked values Masked values
0008.0256.0000 [Noise] A
are output are output
Masked values Masked values
0016.0000.0000 [Turbidity Level Over] A
are output are output
Masked values Masked values
0032.0000.0000 [No Coagulation] A
are output are output
Masked values Masked values
0032.0002.0000 [Flat Curve] A
are output※2 are output
Flagged numeric Flagged numeric
0032.0002.0000 [Flat Curve] A
value*3 value
Masked values
0064.0000.0000 [Aged Sample] A ※1
are output
Masked values Masked values
0128.0000.0000 [Range Over] A
are output are output
Masked values Masked values
0256.0000.0000 [Trans Light High:Clot.] A
are output are output
Numeric values Numeric values
0000.0000.0999 [Switched to high gain] N
are output are output
Numeric values Numeric values
0000.0000.8000 [Coag% Changed] N
are output are output
Numeric values Numeric values
1100.4000.0000 [Defective Sample volume] A
are output are output
Masked values Masked values
4001.0000.0000 [Trans Light Low] A
are output are output
Masked values Masked values
4002.0000.0000 [Trans Light High:Chro.Imm.] A
are output are output
Masked values Masked values
4004.0000.0000 [No Linearity] A
are output are output
Masked values Masked values
4008.0000.0000 [Reaction Curve Error] A
are output are output
Masked values Masked values
4016.0000.0000 [Antigen Excess] A
are output are output
Masked values Masked values
4128.0000.0000 [No Polynomialadjustment] A
are output are output
Masked values Masked values
4256.0000.0000 [Range in non-linear] A
are output are output
Masked values Masked values
4032.0000.0000 [Bump] A
are output are output

40/123
Data value content
Analysis
The setting for The setting for
Code Messages result
error information error information
abnormal flag
output is OFF output is ON
Numeric values Numeric values
7010.0000.0000 [PRP_Low1] A
are output are output
Numeric values Numeric values
7030.0000.0000 [PPP_High1] A
are output are output
Masked values Masked values
7050.0000.0000 [Tube_Position] A
are output are output
Masked values Masked values
8000.0000.0010 [Exceeded the analysis range] N
are output are output
[Found an issue with the analysis count Masked values Masked values
8000.0000.0020 N
value] are output are output
Masked values Masked values
8000.0000.0030 [Failed to obtain the analysis data] N
are output are output
[A temperature error inside the reagent
Numeric values Numeric values
8000.0000.0040 cooling unit or within the immuno A
are output are output
function area occurred.]
[The concentration value has exceeded Numeric values Numeric values
8000.0000.0050 A
the setting range of Review] are output are output
[The reagent lot was switched during Masked values Masked values
8000.0000.0060 A
analysis] are output are output
[Analysis was performed with Numeric values Numeric values
8000.0000.0080 A
over-dilution] are output are output
Masked values Masked values
9999.0000.0000 [Analysis failed] A
are output are output
Masked values Masked values
9999.0000.9010 [TEST_RECV_ELSE_ERROR] N
are output are output
Masked values Masked values
9999.0000.9060 [ASSY_ANLY_NOPRISET_ERROR] N
are output are output
Masked values Masked values
9999.0000.9070 [ASSY_ANLY_NORESULT_ERROR] N
are output are output
Masked values Masked values
9999.0000.9080 [ASSY_ANLY_DILUTION_ERROR] N
are output are output
1000.1000.0000 [Hemolyzed Sample] ※4 ※5 ※5 ※5

1000.2000.0000 [Icteric Sample] ※4 ※5 ※5 ※5

1000.3000.0000 [Lipemic Sample] ※4 ※5 ※5 ※5


Masked values Masked values
9999.0000.9110 [TEST_REAGENTLOT_ERROR] N
are output are output
Numeric values Numeric values
Unknown9999 [Unknown Message] N
are output are output
*1 If HOST_ASTM_OutputCurveErrorInformation is enabled: numeric value. If OFF: Masked values are output
*2 If measurement time ≦ Check Time value
*3 If measurement time > Check Time value
*4 Analysis information is output regardless of the sample information flag output setting
([HOST_ASTM_OutputHILFlags]).
*5 Analysis result abnormal flags and data values are output according to the HIL check conditions set in [System
Settings]. The content of analysis result abnormal flags, data values and responses according to the HIL conditions is
described in the table below. If the HIL check conditions are separately set in each assay group in [System Settings],
output is performed according to that condition.

Table 6.3.3.5-3: Content of analysis result abnormal flags, data values and responses according to the HIL
conditions
HIL check Analysis result
Data value content
output method of analysis results abnormal flag
[Do not perform any operation.] N Numeric values
[Attach a low reliability flag.] A Numeric values
[Mask.] A Masked values

41/123
e) 10.1.9 Result Status
Indicates the status of analysis result.
S: Indicates the result is the part of the analysis results for the sample.
The analysis result is output when the analysis for one Assay Group is end.
F: Indicates the result is all analysis results for the sample.
The analysis result is output when the all analysis for the sample is end.

* Measurement result status can be set to Enabled or Disabled under [Service settings] - [Other
service settings] - [HOST_ASTM_OutputItemUnit].

(Caution)
The measurement results output in measurement parameter units are not the measurement results that are
validated on the instrument. Validate the output result of each Assay Group with analysis result of other
Assay Group on the Host Computer.

f) 10.1.13 Date/Time test completed


Indicates the date and time the test was completed.
The date format is fixed with "YYYYMMDDHHMMSS".
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system
(00-23), MM the minute (00-59), SS the second (00-59).

This date and time indicates when the all tests of the sample are completed.
In the case of the result which is one of the sample(the analysis result with 10.1.9 Result Status is
“S”) , this field is blank.

2.2) When evaluation parameters of platelet aggregation are output: Evaluation parameters are output
when platelet aggregation orders are registered. In order to output evaluation parameters of platelet
aggregation, the setting of “Service settings” - “Other service settings” -
“HOST_ASTM_OutputAggregationAnalysisInformation” on IPU is required to be turned ON. In order
to output platelet aggregation levels of ADP and COL, the setting of “Service settings” - “Other service
settings” - “Show_PlateletAggregationLevel” on IPU is required to be turned ON.

Caution:
Evaluation parameters of platelet aggregation is used for inspection in a clinical
laboratory only and not for examination of a patient. It is to let the operators know
a potential specific abnormality in evaluation parameters confirmed by the
analysis data.

Result records for platelet aggregation analysis output the result records for Aggregation % for each
measurement parameter included in the sample, and the measurement results for each parameter, on a
superimposed graph.

The measurement of platelet aggregation can output multiple evaluation parameters in one measurement. The
followings show output conditions for evaluation parameters of platelet aggregation.

Table 6.3.3.5-4: Evaluation parameter name and output conditions in measurement of platelet aggregation
Evaluation Description Output condition
parameter name
~Abs PPP PPP absorbance - This is always output in result records
for Aggregation.
- Overlay graph result records are output
according to the graph display settings.
~Abs PRP_s PRP absorbance before Same as above
reaction starts
~Abs PRP_e Absorbance after the end of Same as above
the PRP reaction
~Abs PRP_min Minimum value of absorbance Same as above

42/123
~Max Maximum aggregation rate Same as above
~Fin Final aggregation rate Same as above
~Lag Phase Lag Phase Time Same as above
~AUC Area under aggregation rate Same as above
curve
~VMax First maximum reaction speed Same as above
rate
~VMax Time First maximum reaction time Same as above
~VMax Angle First maximum reaction angle Same as above
~APAL Platelet aggregation level for - In platelet aggregation analysis, when
ADP combined measurement parameters for
calculating drug efficacy assessment
indices are displayed in overlay graphs.
Combination of measurement
parameters to display APAL
Activator type: ADP, concentration:
1uM
Activator type: ADP, concentration:
10uM

- Only output in overlay graph result


records.
~CPAL Platelet aggregation level for - In platelet aggregation analysis, when
COL combined measurement parameters for
calculating drug efficacy assessment
indices are displayed in overlay graphs.
Combination of measurement
parameters to display CPAL
Activator type: COL, concentration:
1ug/mL
Activator type: COL, concentration:
5ug/mL

- Only output in overlay graph result


records.

Note:
Additional evaluation parameters may be added in the future. In case evaluation parameters which
are not defined in Table 6.3.3.5-4: Evaluation parameter name and output conditions in
measurement of platelet aggregation are received, you need to use a program which can skip the data
with undefined evaluation parameters.

43/123
a) 10.1.3 Universal Test ID
In the measurement of platelet aggregation, a “Test code” is assigned to each measurement.
Evaluation parameters of platelet aggregation is output linking to the “Test code” of each
measurement. The host computer can specify the output data by the combination of “Test code” and
“Parameter”.

Test code: Test codes for the measurement of platelet aggregation. As for test codes,
refer to the Table 8.1-1: Parameter Code Lists. A test code is not set when
the “Evaluation parameter name” is either the “~APAL” and the “~CPAL”.

Parameter: “Evaluation parameter name” in the measurement of platelet aggregation is


output. As for “Evaluation parameter name”, refer to “Table 6.3.3.5-4:
Evaluation parameter name and output conditions in measurement of
platelet aggregation”.

Dilution ratio: Not used

Analysis result type: The type of analysis results is output. Either type listed below is output.
9: Final report(Auto output)
A: Final report(Manual output)
Extended order request: Not used
Extended order results: Not used
Reflex test request: Not used

b) 10.1.4 Data value


Data to an analysis parameter ID is output. The number of characters is fixed at 6 characters. The
followings describe formats of each evaluation parameter.

Table 6.3.3.5-5: Format definition in evaluation parameters of platelet aggregation


Evaluation parameter name Format
~Abs PPP ̺#####
~Abs PRP_s ̺#####
~Abs PRP_e ̺#####
~Abs PRP_min ̺#####
~Max ####.#
~Fin ####.#
~Lag Phase ####.#
~AUC ######
~VMax ̺#####
~VMax Time ̺̺####
~VMax Angle ###.##
~APAL ̺̺##.#
~CPAL ̺̺##.#
(̺:Half-width space)

In case evaluation parameter is not calculated, the masked character “-” is added to the numeric
part(#) except decimal point.

44/123
c) 10.1.5 Units
Unit to an analysis parameter ID is output. The followings describe units of each evaluation
parameter.

Table 6.3.3.5-6: Definition of unit in evaluation parameters of platelet aggregation


Evaluation parameter name Unit
~Abs PPP mOD
~Abs PRP_s mOD
~Abs PRP_e mOD
~Abs PRP_min mOD
~Max %
~Fin %
~Lag Phase sec
~AUC %*sec
~VMax %/sec
~VMax Time sec
~VMax Angle deg
~APAL None
~CPAL None

d) 10.1.7 Result abnormal flags


Result abnormal flags: Not used
Evaluation Information: Not used
Instrument Error Information: Not used

e) 10.1.9 Result status


Not used.

f) 10.1.13 Date/time test completed


The date and time the test was completed is output.
The format is fixed to “YYYYMMDDHHMMSS”.
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system
(00-23), MM the minute (00-59), SS the second (00-59).
* Calculated date and time of data value is indicated when “analysis parameter” for analysis
parameter ID is the “~APAL” and the “~CPAL”.

2.3) When Sample Info. flags are output: The parameters to which HIL flags or Sample Volume flag are
attached are output.
* Sample Info flags can be set to Enabled or Disabled under [Service settings] - [Other service settings].
- The HIL flag is set in [HOST_ASTM_OutputHILFlags].
- The sample volume flag is set in [HOST_ASTM_OutputSampleVolumeFlags].
The default setting is “Disabled”.
Parameters are output according to the HIL check conditions set in the System Settings.
If the HIL check conditions set are set for individual assay group units in the System Settings,
parameters are output according to those conditions.

(Caution)
Sample Info. flags are used only for inspection in a clinical laboratory and not for examination of a patient.
They notify an operator of the possibility of the specific abnormal sample confirmed by checking analysis
data.

a) 10.1.3 Universal Test ID


Test Code:
Not used.

Parameter:
Indicates the content of HIL flags, or sample volume flags
Hemolytic Sample: Sample suspected of having hemolysis.

45/123
Icterus Sample: Sample suspected of having icterus.
Lipemic Sample: Turbid sample affected by lipemia or other materials.
Defective Sample Volume: Sample suspected of defective sample volume.

Dilution ratio:
Not used.

Analysis result type:


Not used.

Extended order request:


Not used.

Extended order results:


Not used.

Reflex test request:


Not used.

b) 10.1.4 Data or measurement value


The HIL level value is output.
Check Level 5 : “5”
Check Level 4 : “4”
Check Level 3 : “3”
Check Level 2 : “2”
Check Level 1 : “1”
Check Level 0 : “0”
If the HIL detection value is an invalid value (-1) : “*”

* HIL levels can be set to Enabled or Disabled under [Service settings] - [Other service settings] -
[HOST_ASTM_OutputHILLevel].

c) 10.1.5 Units
Not used.

d) 10.1.7 Result abnormal flag


Result abnormal flag
A: Result with HIL flags or sample volume flags.
W: Above the analysis range of the HIL detector (Not used for sample volume flags)
N: Sample Volume check fell within the preset range (Not used for HIL flags)

Evaluation information:
Not used.

Instrument error information:


Not used.

e) 10.1.13 Date /time test completed


Indicates the date and time the test was completed.
The date format is fixed to "YYYYMMDDHHMMSS".
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system
(00-23), MM the minute (00-59), SS the second (00-59).

46/123
2.4) The path name to the image data is output only when the parameters have the image data.

(Caution)
The image data is used for a testing at a clinical laboratory only and not for examination of a patient. It is to
let the operators know a potential specific abnormality in a sample confirmed by the analysis data.

Caution:
Refer to [9.3 Difference List of Transmission Content] as the difference in the
specification of “Image Data Output” between CN-series device and each device.

a) 10.1.3 Universal Test ID


Parameter Code:
A value is set with a “0” appended to the right of the 2-digit ID set in [Assay Group Settings] -
[Basic] - [Host ID].

For platelets, the value that is set combines the 2-digit ID set in [Assay Group Settings] -
[Basic] - [Host ID] with the 1-digit ID number set in [Assay Group Settings] - [Assay Parameter
Settings] - [Basic] - [Assay Parameter Host ID].

Nothing is set for “Superimpose”.

Parameter:
Type of the image data is set.
Normal: Normal Clot Waveform
Average: Average Clot Waveform
MDA: MDA image data
Mixing: Cross mixing image data
Delayed Mixing: Delayed Cross mixing image data
Platelet: This is the image data for the aggregation curve for each
measurement parameter in platelet aggregation analysis.
Superimpose: This is the image data for the superimposed aggregation curves for
platelet aggregation analysis.
Dilution ratio:
Not used

Type of Analysis Results:


Not used

Extended order request:


Not used

Extended order result:


Not used

Reflex test request:


Not used

47/123
b) 10.1.4 Data or measurement values
The file path name to the image data is output. The character “\” used for the file path is converted to
the escape sequence “&R&”. The actual image data is written to the“C:¥Sysmex¥Shared¥PNG¥Date
folder”. The file path name is output after the“PNG”to the data value. The date folder is fixed
as“YYYYMMDD”.
The file name is attached depending on the date/time, sample No., image data, dilution ratio_test seq
No. and the type of extended order results.

Table 6.3.3.5-7: List of the information included in the file


No. Item Contents Remarks
1 Test Date Set analysis date (and time) in the form of When the type of image is
“YYYY_MM_DD_HH_MM_SS”. "Superimpose", the created date
If the set form is “YYYY_MM_DD_HH_MM”, it is and time is transmitted.
necessary to set [Service Setup] - [Other Service
Settings] -
[HOST_ASTM_GraphicData_TimeFormat(HH_M
M)] to Enabled (1).
2 Sample No. Sets the Sample No. -
3 Test Code Set a value with a “0” appended to the right of the No settings in Superimpose.
2-digit ID set in [Assay Group Settings] - [Basic] -
[Host ID].

For platelets, the value that is set combines the


2-digit ID set in [Assay Group Settings] - [Basic] -
[Host ID] with the 1-digit ID number set in [Assay
Group Settings] - [Assay Parameter Settings]
-[Basic] - [Assay Parameter Host ID].
4 Type of Sets the type of image data. Width and height of image data
image data Normal :Normal Clot Waveform describe as below.
Average:Average Clot Waveform Normal :500 x 500 pixel
MDA :MDA image data(*2) Average :500 x 500 pixel
Mixing :Cross mixing image data MDA :500 x 500 pixel
Delayed Mixing:Delayed Cross mixing image Mixing :500 x 500 pixel
data (superimposed graphs) Delayed Mixing
Platelet :Aggregation curve image :500 x 500 pixel
Superimpose :Overlapped aggregation curve Platelet :625 x 388 pixel
image(*1) Superimpose :625 x 388 pixel
(*1) If the overlay graph is “parameter display”.
“Superimpose_Activator type”
(*2) “Average” if the MDA graph is “first analysis
(with re-analysis)”.
5 Dilution Sets dilution ratio. (Optional) -
Ratio
6 Test Sets the sequence No. during analysis. (Optional) -
Sequence
No.
7 Extended Sets the information to identify whether the result -
order is re-analysis result or not. (Optional)
D :Result of re-dilution analysis
R :Result of re-analysis
F :Result of reflex test

"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_100_1.PNG"
In this case, the image data is saved in the “C:¥Sysmex¥Shared¥PNG¥20130930" folder with
the file name of “2013_09_30_12_00_1234567890_040_Normal_100_1.PNG”

Image data save folders are deleted after the IPU shuts down, if they are 3 days old, including
the date of creation.

48/123
Twice analyses
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_100_1.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_100_2.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Average.PNG"
MDA analysis
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_050_1.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_025_1.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_013_1.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_MDA.PNG"
Re-analysis
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_100_1.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_100_1_R.PNG"
MDA analysis (2 measurements)
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_100_1.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_100_2.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Average_100.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_025_1.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_025_2.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Average_025.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_MDA.PNG"
1 measurement + Redilution Analysis (MDA analysis)
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_100_1.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_100_1_D.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_050_1_D.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_025_1_D.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_MDA.PNG"
MDA analysis (1 measurement) + Re-analysis (MDA analysis)
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_100_1.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_050_1.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_025_1.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Average.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_100_1_R.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_050_1_R.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_Normal_025_1_R.PNG"
"PNG&R&20181001&R&2018_10_01_17_09_1_040_MDA.PNG"
MDA analysis (2 measurements) + Re-dilution analysis
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_100_1.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_100_2.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Average_100.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_025_1.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_025_2.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Average_025.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Average.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_050_1_D.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Normal_050_2_D.PNG"
"PNG&R&20181001&R&2018_10_01_17_05_1_040_Average_D.PNG"
Twice analyses + Re-dilution analysis
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_100_1.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_100_2.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Average.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_050_1_D.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Normal_050_2_D.PNG"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_040_Average_D.PNG"
The measurement of platelet aggregation - aggregation curve image for ADP% and
ADP2%
"PNG&R&20130930&R&2017_12_01_13_00_15_1234567890_671_Platelet.Png"
"PNG&R&20130930&R&2017_12_01_13_05_56_1234567890_672_Platelet.Png"
49/123
Platelet aggregation analysis - superimposed aggregation curves (parameter display)
If the 4 measurement parameters are ADP%, ADP2%, Col%, and Col2%, and the overlay
graph is “parameter display”:
This indicates that it is image data in activator units, with the “Superimpose_activator
type” format.
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_Superimpose_ADP.Png"
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_Superimpose_Col.Png"
Platelet aggregation analysis - superimposed aggregation curves (overall display)
If the 4 measurement parameters are ADP%, ADP2%, Col%, and Col2%, and the overlay
graph is “overall display”:
"PNG&R&20130930&R&2013_09_30_12_00_1234567890_Superimpose.Png"

c) 10.1.5 Units
Not used

d) 10.1.7 Result abnormal flag


Not used

e) 10.1.13 Date/time test completed


Indicates the date and time the analysis was completed.
The date format is fixed to "YYYYMMDDHHMMSS".
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system
(00-23), MM the minute (00-59), SS the second (00-59).

50/123
(Caution)
Output of Delayed Cross mixing image data

The delayed cross mixing image can be output to host computer manually. In case, the IPU sends Cross
mixing analysis information in CSV file format as well as PNG file. The CSV file is written to the
“C:¥Ssymex¥Shared¥CSV¥Date folder”.

10.1.4 Data or measurement values


The file path name to the image data is output.
The file path name is output after the“CSV”to the data value.
The file name conforms Table 6.3.3.5-7: List of the information included in the file.
The other items output the same information as the image data.

The contents of CSV file;


Item Contents Remarks
Sample No Sample No.
Date/Time The analyzed date and time. YYYY/MM/DD hh:mm:ss
Assay Group Assay Group Settings display
name
Sample Comment Sample No. 40 columns (20 full-width or 40
half-width characters)
Mixing Ratio Mixing Ratio. The dilution ratio is output as a
percentage, as the patient
plasma ratio.
Example) 1/4 → 25%

Repeat as many times as there


are measured mixing ratio
points, and output.
Immediate Type Result data of Immediate type Result data are specified with
in each mixing ratio. including 1-digit of decimal
number.
Output in parallel with the
MixingRatio.
Delayed Type Result data of Delayed type in Result data are specified with
each mixing ratio. including 1-digit of decimal
number.
Output in parallel with the
MixingRatio.
Reference Info. Title of Reference Info. Repeat the output as many times
as there are Reference Info.
items.
Value Value of Reference Info. Values of Reference Info. are
specified with including 1-digit of
decimal number.
Output in parallel with the
ReferenceInfo.

The format of CSV file;


Object Format
Character Code UTF-8
Control code <CR><LF>
Header Line Output
Delimiter Comma
Data output Double quote
(If there is no value, “” is
output. If double quotation
marks are present in the
value, the marks are
duplicated in the value).

51/123
- The following is an example of CSV file output.

"Sample No","1111"
"Date/Time","2017/11/13 18:11:30"
"Assay Group","APTT"
"Comment","CommentComment"

"Mixing Ratio","Immediate Type","Delayed Type"


"0%","10.2","10.2"
"50%","10.5","10.3"
"100%","10.7","10.6"

"Reference Info.","Value"
"ICA (Rosner Index)","2.8"
"ICA D","0.9"
"ICA D / ICA","33.6"
"ICA D - ICA","-1.9"

52/123
6.3.3.6.Comment record
Output 1-line comment records after the result record for each calculated parameter.

To output Comment records, set them to Enabled under [Service settings] - [Other service settings] -
[HOST_ASTM_OutputCommentRecord].

There are 3 types of comment record, and output proceeds in the sequence “calibration curve
information -> Reagent Lot Information -> QC Measurement Information”.
If any comment record is not output due to the output conditions for each comment record, that record
is skipped and output proceeds to the next one.

Refer to “6.3.3.7 Comment record (calibration curve information)” about comment records (Calibration
Curve Information).
Refer to “6.3.3.8 Comment record (Reagent Lot Information)” about comment records (Reagent Lot
Information).
Refer to “6.3.3.9 Comment Records (QC analysis information)” about comment records (QC
Measurement Information).

Comment records are attached to measurement parameter measurement result records, and to the
result record for the Final Data. They are not attached to average result records. Result records to
which comment records are attached are shown in “Table 6.3.3.6-1: Result Records to Which Comment
Records are Attached”.

Table 6.3.3.6-1: Result Records to Which Comment Records are Attached


Measurement result type of result record Comment record attachment
1 :Normal (Auto output) Attach
2 :Average (Auto output) Do not attach
3 :Re-analysis (Auto output) Attach
4 :Average of re-analysis (Auto output) Do not attach
5 :Normal (Manual Output) Attach
6 :Average (Manual Output) Do not attach
7 :Re-analysis (Manual Output) Attach
8 :Average of re-analysis (Manual Output) Do not attach
9 :Final information (Auto output) Attach
A Final information (Manual Output) Attach

6.3.3.7.Comment record (calibration curve information)


Output 1-line comment records (calibration curve information) after the result record for each
calculated parameter.
If the calculated parameter is a time or ratio, or if the calculated parameter is INR and the calculation
method is ISI input, the record is not output.
Comment records for calibration curve information are omitted if the standard curve ID is duplicated
in assay parameter units.

[Example of transmission]
 CN-series device -> Host computer
C|1|I|CAL^042^20140701^1^502501|I<CR>
 Host computer -> CN-series device
Not used

53/123
Table 6.3.3.7-1: Details of calibration curve information comment records
ASTM Field name CN-series device -> Host computer -> Max. size Remarks
Field Host computer CN-series device (Byte)
11.1.1 Record type C Not used 1
11.1.2 Sequence Sequence No. Not used 4 Sequence Number of
number records
11.1.3 Comment I Not used 1 Clinical instrument
Source system
11.1.4 Comment Information type^ Not used 3^
Text Computed 3^
parameter code^ 8^
Standard curve 7^
measurement 12
date^
Standard curve
ID^
Calibrator lot
11.1.5 Comment I Not used 1 Instrument
Type information comments

[Detailed explanation of the fields]


1) 11.1.2 Sequence No.
The sequence number starts at 1 and indicates the sequence position in which the record appeared in
the message.
This number is reset to 1 when a higher-level record appears in the message.

2) 11.1.3 Comment source


Fixed to “I”.
This means a clinical instrument system.

3) 11.1.4 Comment Text


Information type:
This is the information type for the Comment Text field. The standard curve information is
CAL.

Calculated parameter code:


This is the code No. that indicates the measurement procedure item.
This code is a combination of a 2-digit universal test code and the following 1-digit IDs.
See the Table 8.1-1: Parameter Code Lists.
1: Time
2: Activity percent/concentration
3: Ratio
4: INR
※Only for CN-6500/CN-3500
For immuno module, this code is a combination of a 2-digit universal test code and the
following 1-digit IDs.
See “Table 8.2-1:Parameter Code Lists”.
1: Count
2: Concentration

Standard curve analysis date:


Indicates the date and time at which calibration curve analysis was performed for the
measurement parameter in the result record.
The format is fixed to “YYYYMMDD”.
YYYY: Year, MM: Month, DD: Day
This is omitted if the standard curve analysis has not been performed.

Calibration curve ID:

54/123
Indicates the standard curve ID for the measurement parameter in the result record.
If the number of digits is less than 7, the number will be aligned to the right and padded with
zeros.
This is omitted if there is no calibration curve ID.

Calibrator lot:
Indicates the LOT number of the calibrator used for the standard curve ID for the
measurement parameter in the result record.
This is omitted if calibrator information was not set for the standard curve.

* There is 1 type of calibrator information which can be registered on the Reagent Settings screen for
the standard curve, so even if SHP and PT multi-calibrator is used when preparing the PT% standard
curve and PT-INR, only 1 type of information can be sent.

4) 11.1.5 Comment format


Fixed to “I”.
This means an instrument information comment.

55/123
6.3.3.8.Comment record (Reagent Lot Information)
Output 1-line comment records (Reagent Lot Information) after the result record for each
measurement parameter.
If there are multiple calculated parameters, the record is output after the final calculated parameter
result record.
Reagent Lot Information comment records are omitted if the combination of Transmission 1 reagent
LOT to Transmission 5 reagent LOT is repeated in assay group units.

[Example of transmission]
 CN-series device -> Host computer
C|1|I|LOT^040^527501|I<CR>
 Host computer -> CN-series device
Not used

Table 6.3.3.8-1: Details of Reagent Lot Information Comment Records


ASTM Field name CN-series device -> Host computer -> Max. size Remarks
Field Host computer CN-series device (Byte)
11.1.1 Record type C Not used 1
11.1.2 Sequence Sequence No. Not used 4 Sequence Number of
number records
11.1.3 Comment I Not used 1 Clinical instrument
Source system
11.1.4 Comment Information type^ Not used 3^
Text Parameter code^ 3^
Transmission 1 12,
reagent LOT, 12,
Transmission 2 12,
reagent LOT, 12,
Transmission 3 12
reagent LOT,
Transmission 4
reagent LOT,
Transmission 5
reagent LOT
11.1.5 Comment I Not used 1 Instrument
Type information comments

[Detailed explanation of the fields]


1) 11.1.2 Sequence No.
The sequence number starts at 1 and indicates the sequence position in which the record appeared in
the message.
This number is reset to 1 when a higher-level record appears in the message.

2) 11.1.3 Comment source


Fixed to “I”.
This means a clinical instrument system.

3) 11.1.4 Comment Text


Information type:
This is the information type for the Comment Text field. The Reagent Lot Information is the
LOT.

Parameter Code:
This is the code No. that indicates the measurement procedure item.
This is the value produced by appending a “0” to the right of the 2-digit Host ID set in [Assay
Group Settings].
※Only for CN-6500/CN-3500
56/123
For immuno parameter, suffixing 0 to the “Host ID” in the “Basic” in the " Parameter info,
settings " in the “Immuno module” makes the parameter code.

Transmission 1-5 reagent LOTs:


These are the LOT numbers of the diluent, factor-deficient plasma, and reagents used in
measurement.
The transmission sequence of output is the diluent and factor-deficient plasma registered in
the “Assay Group Settings, Test Protocol Settings”, and the reagents in the order of aspiration
procedure.
Reagents not under Reagent Lot Management are not output.
If there are 5 or more reagents subject to output, only the first 5 are output.
※Only for CN-6500/CN-3500
When the parameter code is for immuno paramter, these are the Lot numbeters of R1R2R3
reagent and R4R5 reagents used in measurement.
The transmission sequence of output is R1R2R3 reagent and R4R5 reagent.

4) 11.1.5 Comment format


Fixed to “I”.
This means an instrument information comment.

57/123
6.3.3.9.Comment Records (QC analysis information)
One-line comment records (QC analysis information) are output after the result record for each
measurement parameter.
If there are multiple computed parameters, the record is output after the final computed parameter
result record.
Comment records for QC measurement information are omitted if the combination of measurement
date and time and control LOT is duplicated in assay parameter units.
※Only for CN-6500/CN-3500
For immuno parameter, comment records for QC measurement information are omitted if the
combination of measurement date and control LOT is duplicated in parameter units.

[Example of transmission]
 CN-series device -> Host computer
C|1|I|QC^040^201407011530^^502701\QC^040^201407011600^^512601|I<CR>
 Host computer -> CN-series device
Not used

Table 6.3.3.9-1: Details of QC Measurement Information Comment Records


ASTM Field name CN-series device -> Host computer -> Max. size Remarks
Field Host computer CN-series device (Byte)
11.1.1 Record type C Not used 1
11.1.2 Sequence Sequence No. Not used 4 Sequence Number of
number records
11.1.3 Comment I Not used 1 Clinical instrument
Source system
11.1.4 Comment Information type^ Not used 3^ If multiple controls
Text Parameter code^ 3^ are used, they are
Measurement date 12^ separated by a repeat
and time^ 12^ symbol “\”.
Control name^ 12
Control LOT
11.1.5 Comment I Not used 1 Instrument
Type information comments

[Detailed explanation of the fields]


1) 11.1.2 Sequence No.
The sequence number starts at 1 and indicates the sequence position in which the record appeared in
the message.
This number is reset to 1 when a higher-level record appears in the message.

2) 11.1.3 Comment source


Fixed to “I”.
This means a clinical instrument system.

3) 11.1.4 Comment Text


Information type:
This is the information type for the Comment Text field. QC is QC measurement information.

Parameter Code:
This is the code No. that indicates the measurement procedure item.
This is the value produced by appending a “0” to the right of the 2-digit Host ID set in [Assay
Group Settings].
※Only for CN-6500/CN-3500
For immuno parameter, suffixing 0 to the “Host ID” in the “Basic” in the " Parameter info,
settings " in the “Immuno module” makes the parameter code.

58/123
Analyzed on:
This is the latest QC measurement date.

Control Name:
This is not used with the CN-series device.

Control LOT:
This is the LOT number of the control used in quality control analysis for the measurement
parameter in the result record.
If there are multiple QC files and the control information for quality control has been set, this
is the control LOT No. for the QC file with the newest measurement date and time.

4) 11.1.5 Comment format


Fixed to “I”.
This means an instrument information comment.

6.3.3.10.Message termination record


[Example of transmission]
 CN-series device -> Host computer
L|1|N<CR>
 Host computer -> CN-series device
L|1|N<CR>

Table 6.3.3.10-1: Message termination record


ASTM CN-series device -> Host computer -> Max. size
Field Field name Host computer CN-series device (Byte) Remarks
13.1.1 Record type L L 1 Fixed
13.1.2 Sequence No. 1 1 4 Always 1
13.1.3 Termination N N 1 N: Normal
cord termination

59/123
7. Examples of Communication

Examples of Communication are given for the cases when the serial connection (ASTM E1381-02 Mode) is
carried out

7.1. Inquiry of Analysis Order (IPU => Host Computer)


・When the order inquiry is performed from IPU

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>2Q|1|000001^01^ 1^B||^^^040^PT\^^^060^Fbg\^^^120^II\^^^150^V\^^^
170^VII\^^^180^VIII\^^^190^IX\^^^200^X\^^^210^XI\^^^300^AT\^^^310^APL|0|201103281
33318<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

7.2. Inquiry of Re-analysis Order (IPU=>Host)


* When the “Device Settings” - “Host” - “Inquire re-analysis” setting in the “System Setting”is ON

IPU <ENQ>
HOST <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97
<CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>2Q|1|000001^01^ 1^B||^^^040^PT\^^^060^Fbg\^^^120^II\^^^150^V\^^^
170^VII\^^^180^VIII\^^^
190^IX\^^^200^X\^^^210^XI\^^^300^AT\^^^310^APL|0|20110328133318||||||C<ETX><CH
K1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <EOT>

7.3. Analysis Order (Host Computer => IPU)


・When the host computer sends orders to IPU

Host <ENQ>
IPU <ACK>
Host <STX>1H|\^&|||||||||||E1394-97|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>2P|1|||100|^Heisei^Jiro<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>3O|1|000001^01^ 1^B||^^^040^^100.00^DF\^^^050^^100.00\^^^060^^10
0.00\^^^040^^100.00^DR|R|20110330123159|||||N|||||||||||||||||||<CR><ETX><CHK1
><CHK2><CR><LF>
IPU <ACK>
Host <STX>4L|1|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <EOT>

60/123
7.3.1. Specify Replication
*When "Service settings" - "Other service settings" - "HOST_ASTM_SpecifyReplication" setting is ON.
Host <ENQ>
IPU <ACK>
Host <STX>1H|\^&|||||||||||E1394-97|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>2P|1|||100|^Heisei^Jiro<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>3O|1|000001^01^ 1^B||^^^040^^100.00^DF^2\^^^050^^100.00^^1\^^^06
0^^100.00^^1\^^^040^^100.00^DR^2|R|20110330123159|||||N|||||||||||||||||||<CR><
ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>4L|1|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <EOT>

7.3.2. Specify Measurement Mode


*When "Service settings" - "Other service settings" - "HOST_ASTM_SpecifyMeasurementMode" setting is ON.
Host <ENQ>
IPU <ACK>
Host <STX>1H|\^&|||||||||||E1394-97|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>2P|1|||100|^Heisei^Jiro<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>3O|1|000001^01^ 1^B^^M||^^^040^^100.00^DF\^^^050^^100.00\^^^060
^^100.00\^^^040^^100.00^DR|R|20110330123159|||||N|||||||||||||||||||<CR><ETX><
CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>4L|1|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <EOT>

7.3.3. Specify Measurement Time


*3 When "Service settings" - "Other service settings" - "HOST_ASTM_SwitchMeasurementTime" is ON.
HOST <ENQ>
IPU <ACK>
HOST <STX>1H|\^&|||||||||||E1394-97|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
HOST <STX>2P|1|||100|^Heisei^Jiro<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
HOST <STX>3O|1|000001^01^ 1^B||^^^040^^100.00^DF^^S\^^^050^^100.00^^M\^^^
060^^100.00^^S\^^^040^^100.00^DR^^S|R|20110330123159|||||N|||||||||||||||||||<C
R><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
HOST <STX>4L|1|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
HOST <EOT>

7.4. Analysis Results (IPU => Host Computer)

7.4.1. When analysis is performed 1 time


IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
61/123
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^| 10.2|sec||N||||||20110328135056<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^^| 99.4|%||N||||||20110328135056<ETX><CHK1><C
HK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^^| 0.57|||N||||||20110328135056<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^^| 0.81|||N||||||20110328135056<ETX><CHK1><C
HK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^9^^^| 27.4|sec||N||||||20110328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^9^^^| 8.5|sec||N||||||20110328135056<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^9^^^| 588.2|mg/dL||N||||||20110328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|8|^^^^Hemolytic Sample^^^^^||||A||||||20110328135056<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>4R|9|^^^^Defective Sample Volume^^^^^||||N||||||20110328135056<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

* When the "Device Settings" - "Host" - "Inquire re-analysis" setting is ON


(The case of “analysis result (first analysis)” for “7.4.2.1 Transmitting analysis results” is the same)
IPU <ENQ>
HOST <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX>
<CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^^^| 10.2|sec||N||||||20110328135056<ETX><CHK1
><CHK2><CR><LF>
HOST <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^^^| 99.4|%||N||||||20110328135056<ETX><CHK1><
CHK2><CR><LF>
HOST <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^^^| 0.57|||N||||||20110328135056<ETX><CHK1><C
HK2><CR><LF>
HOST <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^^^| 0.81|||N||||||20110328135056<ETX><CHK1><
CHK2><CR><LF>
HOST <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^1^^^| 27.4|sec||N||||||20110328135056<ETX><CH
K1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^1^^^| 8.5|sec||N||||||20110328135056<ETX><CHK1
><CHK2><CR><LF>
HOST <ACK>

62/123
IPU <STX>2R|7|^^^062^Fbg C.^100.00^1^^^| 588.2|mg/dL||N||||||20110328135056<ETX><CH
K1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>3R|8|^^^^Hemolytic Sample^^^^^||||A||||||20110328135056<ETX><CHK1><CHK2>
<CR><LF>
HOST <ACK>
IPU <STX>4R|9|^^^^Defective Sample Volume^^^^^||||N||||||20110328135056<ETX><CHK1>
<CHK2><CR><LF>
HOST <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <EOT>

7.4.2. For Multiple Parameter Analysis


The following is an example of communication of output analysis result records, image data, and
comment records when multiple parameters have been analyzed.
Parameter output sequence is in ascending order of the Host ID registered in [Assay Group Settings].

Example: When outputting the following parameters


Assay Group Settings (Seq.) Host ID
PT THS(ISI) 04
AT BC 30
Fbg 06

IPU <ENQ>
HOST <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||
E1394-97<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CH
K2><CR><LF>
HOST <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^^|14.3|sec||N||||||2020112609564
7<CR><ETX>4B<CR><LF>
HOST <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^^^|8.1|%||N||||||20201126095647<C
R><ETX>F3<CR><LF>
PT THS(ISI)
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^^^|14.30|||N||||||20201126095647<
CR><ETX>8B<CR><LF>
HOST <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^^^|14.30|||N||||||20201126095647
<CR><ETX>F7<CR><LF>
HOST <ACK>
IPU <STX>0R|5|^^^061^Fbg sec^100.00^9^^^^|13.2|sec||N||||||2020112609564
7<CR><ETX>49<CR><LF>
Fbg HOST <ACK>
IPU <STX>1R|6|^^^062^Fbg C.^100.00^9^^^^|588.1|mg/dL||N||||||2020112609
5647<CR><ETX>F3<CR><LF>
HOST <ACK>
IPU <STX>2R|7|^^^301^AT dOD^100.00^9^^^^|0.000|dOD||N||||||2020112609
AT BC 5647<CR><ETX>E3<CR><LF>
HOST <ACK>
IPU <STX>3R|8|^^^302^AT %^100.00^9^^^^|10.0|%||<||||||20201126095647<
CR><ETX>28<CR><LF>
HOST <ACK>

63/123
IPU <STX>4R|9|^^^040^Normal^^^^^^|PNG&R&20201126&R&2020_11_26_09_56
_1_040_Normal_100_1.PNG|||||||||20201126095647<CR><ETX>AB<CR><LF
>
HOST <ACK>
IPU <STX>5R|10|^^^060^Normal^^^^^^|PNG&R&20201126&R&2020_11_26_09_5
6_1_060_Normal_100_1.PNG|||||||||20201126095647<CR><ETX>AD<CR><L
F>
HOST <ACK>
IPU <STX>6R|11|^^^300^Normal^^^^^^|PNG&R&20201126&R&2020_11_26_09_5
6_1_300_Normal_100_1.PNG|||||||||20201126095647<CR><ETX>A5<CR><L
F>
HOST <ACK>
IPU <STX>7C|1|I|CAL^042^20201124^0000001^502501|I<CR><ETX>1F<CR><LF
>
HOST <ACK>
IPU <STX>0C|2|I|CAL^062^20201126^0000005^502501|I<CR><ETX>27<CR><LF>
HOST <ACK>
IPU <STX>1C|3|I|CAL^302^20201124^0000002^502501|I<CR><ETX>23<CR><LF>
HOST <ACK>
IPU <STX>2C|4|I|LOT^040^505401|I<CR><ETX>B3<CR><LF>
HOST
IPU <STX>3C|5|I|LOT^060^527502|I<CR><ETX>AE<CR><LF>
HOST <ACK>
IPU <STX>4C|6|I|LOT^300^520101,520201|I<CR><ETX>FE<CR><LF>
HOST <ACK>
IPU <STX>5C|7|I|QC^040^202011260949^^W902|I<CR><ETX>39<CR><LF>
HOST <ACK>
IPU <STX>6C|8|I|QC^060^202011260949^^W902|I<CR><ETX>33<CR><LF>
HOST <ACK>
IPU <STX>7C|9|I|QC^300^202011260949^^W902|I<CR><ETX>34<CR><LF>
HOST <ACK>
IPU <STX>0L|1|N<CR><ETX>05<CR><LF>
HOST <ACK>
IPU <EOT>

64/123
・ Research parameters

Only reportable parameters are output when [Service Setup] - [Other Service Settings] -
[HOST_ASTM_OutputRUO] is disabled. Research parameter result records, image data, and comment
records, are not output.

Example: When [Service Setup] - [Other Service Settings] - [HOST_ASTM_OutputRUO] is disabled and
the following parameters were analyzed.
Assay Group Settings (Seq.) Host ID Research parameters / reportable parameters
PT THS(ISI) 04 Research parameters
AT BC 30 Reportable parameters
Fbg 06 Reportable parameters

IPU <ENQ>
HOST <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX|||||||
|E1394-97<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><C
HK2><CR><LF>
HOST <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^9^^^^|13.2|sec||N||||||202011260956
Fbg 47<CR><ETX>49<CR><LF>
HOST <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^100.00^9^^^^|588.1|mg/dL||N||||||202011260
95647<CR><ETX>F3<CR><LF>
HOST <ACK>
IPU <STX>6R|3|^^^301^AT dOD^100.00^9^^^^|0.000|dOD||N||||||202011260
AT BC 95647<CR><ETX>E3<CR><LF>
HOST <ACK>
IPU <STX>7R|4|^^^302^AT %^100.00^9^^^^|10.0|%||<||||||20201126095647<
CR><ETX>28<CR><LF>
HOST <ACK>
IPU <STX>0R|5|^^^060^Normal^^^^^^|PNG&R&20201126&R&2020_11_26_09_5
6_1_060_Normal_100_1.PNG|||||||||20201126095647<CR><ETX>AD<CR><
LF>
HOST <ACK>
IPU <STX>1R|6|^^^300^Normal^^^^^^|PNG&R&20201126&R&2020_11_26_09_5
6_1_300_Normal_100_1.PNG|||||||||20201126095647<CR><ETX>A5<CR><L
F>
HOST <ACK>
IPU <STX>2C|1|I|CAL^062^20201126^0000005^502501|I<CR><ETX>27<CR><LF
>
HOST <ACK>
IPU <STX>3C|2|I|CAL^302^20201124^0000002^502501|I<CR><ETX>23<CR><LF
>
HOST <ACK>
IPU <STX>4C|3|I|LOT^060^527502|I<CR><ETX>AE<CR><LF>
HOST
IPU <STX>5C|4|I|LOT^300^520101,520201|I<CR><ETX>FE<CR><LF>
HOST <ACK>
IPU <STX>6C|5|I|QC^060^202011260949^^W902|I<CR><ETX>33<CR><LF>
HOST <ACK>
IPU <STX>7C|6|I|QC^300^202011260949^^W902|I<CR><ETX>34<CR><LF>
HOST <ACK>
IPU <STX>0L|1|N<CR><ETX>05<CR><LF>
HOST <ACK>
IPU <EOT>
65/123
When [Service Setup] - [Other Service Settings] - [HOST_ASTM_OutputRUO] is enabled, research
parameters and reportable parameters are output, in the sequence of reportable parameters then
research parameters.

Example: When [Service Setup] - [Other Service Settings] - [HOST_ASTM_OutputRUO] is enabled and
the following parameters were analyzed.
Assay Group Settings (Seq.) Host ID Research parameters / reportable parameters
PT THS(ISI) 04 Research parameters
AT BC 30 Reportable parameters
Fbg 06 Reportable parameters

IPU <ENQ>
HOST <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||
||E1394-97<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
HOST <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><C
HK2><CR><LF>
HOST <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^9^^^^|13.2|sec||N||||||20201126095
Fbg 647<CR><ETX>49<CR><LF>
HOST <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^100.00^9^^^^|588.1|mg/dL||N||||||20201126
095647<CR><ETX>F3<CR><LF>
HOST <ACK>
IPU <STX>6R|3|^^^301^AT dOD^100.00^9^^^^|0.000|dOD||N||||||20201126
AT BC 095647<CR><ETX>E3<CR><LF>
HOST <ACK>
IPU <STX>7R|4|^^^302^AT %^100.00^9^^^^|10.0|%||<||||||20201126095647
<CR><ETX>28<CR><LF>
HOST <ACK>
IPU <STX>0R|5|^^^041^PT sec^100.00^9^^^^|14.3|sec||N||||||202011260956
47<CR><ETX>4B<CR><LF>
HOST <ACK>
IPU <STX>1R|6|^^^042^PT %^100.00^9^^^^|8.1|%||N||||||20201126095647<
CR><ETX>F3<CR><LF>
PT THS(ISI) HOST <ACK>
IPU <STX>2R|7|^^^043^PT R.^100.00^9^^^^|14.30|||N||||||20201126095647
<CR><ETX>8B<CR><LF>
HOST <ACK>
IPU <STX>3R|8|^^^044^PT INR^100.00^9^^^^|14.30|||N||||||202011260956
47<CR><ETX>F7<CR><LF>
HOST <ACK>
IPU <STX>4R|9|^^^060^Normal^^^^^^|PNG&R&20201126&R&2020_11_26_09_5
6_1_060_Normal_100_1.PNG|||||||||20201126095647<CR><ETX>AD<CR><
LF>
HOST <ACK>
IPU <STX>5R|10|^^^300^Normal^^^^^^|PNG&R&20201126&R&2020_11_26_09_
56_1_300_Normal_100_1.PNG|||||||||20201126095647<CR><ETX>A5<CR>
<LF>
HOST <ACK>
IPU <STX>6R|11|^^^040^Normal^^^^^^|PNG&R&20201126&R&2020_11_26_09_
56_1_040_Normal_100_1.PNG|||||||||20201126095647<CR><ETX>AB<CR>
<LF>
HOST <ACK>
IPU <STX>7C|1|I|CAL^062^20201126^0000005^502501|I<CR><ETX>27<CR><L
F>
HOST <ACK>

66/123
IPU <STX>0C|2|I|CAL^302^20201124^0000002^502501|I<CR><ETX>23<CR><L
F>
HOST <ACK>
IPU <STX>1C|3|I|CAL^042^20201124^0000001^502501|I<CR><ETX>1F<CR><L
F>
HOST <ACK>
IPU <STX>2C|4|I|LOT^060^527502|I<CR><ETX>AE<CR><LF>
HOST
IPU <STX>3C|5|I|LOT^300^520101,520201|I<CR><ETX>FE<CR><LF>
HOST <ACK>
IPU <STX>4C|6|I|LOT^040^505401|I<CR><ETX>B3<CR><LF>
HOST <ACK>
IPU <STX>5C|7|I|QC^060^202011260949^^W902|I<CR><ETX>33<CR><LF>
HOST <ACK>
IPU <STX>6C|8|I|QC^300^202011260949^^W902|I<CR><ETX>34<CR><LF>
HOST <ACK>
IPU <STX>7C|9|I|QC^040^202011260949^^W902|I<CR><ETX>39<CR><LF>
HOST <ACK>
IPU <STX>0L|1|N<CR><ETX>05<CR><LF>
HOST <ACK>
IPU <EOT>

7.4.3. When there is the re-analysis


Communications examples are presented for cases when re-test analysis is performed, based on
analyzer judgment rules.

7.4.3.1.Transmitting analysis results


Example: Results of redilution analysis

・ Analysis Results (The first analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^1^D^^^| 8.5|sec||N||||||20110328135407<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^100.00^1^D^^^| 588.2|mg/dL||N||||||20110328135407<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (The re-analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
67/123
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^50.00^3^^D^^| 8.5|sec||N||||||20110328135407<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^50.00^3^^D^^| 588.2|mg/dL||N||||||20110328135407<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (The final analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^50.00^9^^D^^| 8.5|sec||N||||||20110328135407<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^50.00^9^^D^^| 588.2|mg/dL||N||||||20110328135407<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

Example: Re-analysis results

・ Analysis Results (The first analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^R^^^| 10.2|sec||N||||||20110328135407<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^R^^^| 99.4|%||N||||||20110328135407<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^R^^^| 0.57|||N||||||20110328135407<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^R^^^| 0.81|||N||||||20110328135407<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>0L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

68/123
・ Analysis Results (The re-analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^3^^R^^| 10.2|sec||N||||||20110328135407<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^3^^R^^| 99.4|%||N||||||20110328135407<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^3^^R^^| 0.57|||N||||||20110328135407<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^3^^R^^| 0.81|||N||||||20110328135407<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>0L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (The final analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^R^^| 10.2|sec||N||||||20110328135407<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^R^^| 99.4|%||N||||||20110328135407<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^R^^| 0.57|||N||||||20110328135407<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^R^^| 0.81|||N||||||20110328135407<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>0L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

69/123
Example: Results of redilution analysis (1 sample measurement for multiple parameters: when only
first analysis parameters exist)
Analysis Parameters: PT (no re-analysis), APTT (no re-analysis), Fbg (redilution)

■ The final analysis results are collectively output if [HOST_ASTM_Divide_FinalData] setting is


disabled for [Service settings]-[Other service settings].

<Output data>
First analysis: Fbg
Re-analysis: Fbg
Final analysis: PT, APTT, Fbg

First analysis: IPU <ENQ>


Fbg Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E
1394-97<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^E^^||R||||||N<ETX><CHK1><C
HK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^1^D^^| 8.6|sec||N||||||201103281350
56<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^100.00^1^D^^| 9.00|g/L||>||||||201103281350
56<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>
Re-analysis: IPU <ENQ>
Fbg Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E
1394-97<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^E^^||R||||||N<ETX><CHK1><C
HK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^ 50.00^3^^D^| 8.1|sec||N||||||201103281350
56<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^ 50.00^3^^D^| 9.00|g/L||>||||||201103281350
56<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>
Final analysis: IPU <ENQ>
PT, Host <ACK>
APTT, IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E
Fbg 1394-97<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^E^^||R||||||N<ETX><CHK1><C
HK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^| 10.2|sec||N||||||20110328135056

70/123
<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^^| 99.4|%||N||||||20110328135056<
ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^^| 0.57|||N||||||20110328135056<E
TX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^^| 0.81|||N||||||20110328135056<
ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^9^^^| 27.4|sec||N||||||20110328135
056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^ 50.00^9^^D^| 8.1|sec||N||||||201103281350
56<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^ 50.00^9^^D^| 9.00|g/L||>||||||201103281350
56<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

■ The final analysis results are separately output into those with and without “re-analysis” if
[HOST_ASTM_Divide_FinalData] setting is abled for [Service settings] - [Other service settings].

<Output data>
First analysis: Fbg
Final analysis: PT, APTT Parameters without re-analysis
Re-analysis: Fbg
Final analysis: Fbg Parameters with re-analysis

First analysis: IPU <ENQ>


Fbg Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E
1394-97<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^E^^||R||||||N<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^1^D^^| 8.5|sec||N||||||201103281350
56<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^100.00^1^D^^| 30.0|mg/dL||<||||||2011032813
5056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>
Final analysis: IPU <ENQ>
PT, Host <ACK>
APTT IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E
1394-97<ETX><CHK1><CHK2><CR><LF>
* Parameters Host <ACK>
without IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
re-analysis Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^E^^||R||||||N<ETX><CHK1><
CHK2><CR><LF>

71/123
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^| 10.2|sec||N||||||20110328135056
<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^^| 99.4|%||N||||||20110328135056<
ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^^| 0.57|||N||||||20110328135056<E
TX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^^| 0.81|||N||||||20110328135056<
ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^9^^^| 27.4|sec||N||||||20110328135
056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>
Re-analysis: IPU <ENQ>
Fbg Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E
1394-97<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^E^^||R||||||N<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^400.00^3^^D^| 8.5|sec||N||||||201103281350
56<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^400.00^3^^D^| 30.0|mg/dL||<||||||2011032813
5056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>
Final analysis: IPU <ENQ>
Fbg Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E
* Parameters 1394-97<ETX><CHK1><CHK2><CR><LF>
with Host <ACK>
re-analysis IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^E^^||R||||||N<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^400.00^9^^D^| 8.5|sec||N||||||201103281350
56<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^400.00^9^^D^| 30.0|mg/dL||<||||||2011032813
5056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

72/123
7.4.3.2.Item unit output
* If [HOST_ASTM_OutputItemUnit] is enabled under [Service settings] - [Other service settings]

If [HOST_ASTM_OutputItemUnit] and [HOST_ASTM_OutputAfterFirstMeas] are enabled under


[Service settings] - [Other service settings], the first analysis results are sent to the host on completion
of first analysis, with an extended order request attached. In that case, the first analysis results are not
output when analysis for the re-test is completed.

Example: Results of redilution analysis

・ Fbg (The first analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
Host <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^1^D^^^| 8.5|sec||N||S||||<ETX><CHK1><CHK2><CR
><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^100.00^1^D^^^| 588.2|mg/dL||N||S||||<ETX><CHK1><CHK2>
<CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ Fbg (The re-analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^50.00^3^^D^^| 8.5|sec||N||S||||<ETX><CHK1><CHK2><CR>
<LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^50.00^3^^D^^| 588.2|mg/dL||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ Fbg (The final analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>

73/123
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^50.00^9^^D^^| 8.5|sec||N||S||||<ETX><CHK1><CHK2><CR>
<LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^50.00^9^^D^^| 588.2|mg/dL||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ DD (The first analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^611^DD dOD^100.00^1^D^^^| 0.0574|dOD||N||S||||<ETX><CHK1><CHK2
><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^612^DD C.^100.00^1^D^^^| 20.7|ug/mL||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ DD (The re-analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^611^DD dOD^50.00^3^^D^^| 0.1057|dOD||N||S||||<ETX><CHK1><CHK2>
<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^612^DD C. ^50.00^3^^D^^| 31.1|ug/mL||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ DD (The final analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><

74/123
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^611^DD dOD^50.00^9^^D^^| 0.1057|dOD||N||S||||<ETX><CHK1><CHK2>
<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^612^DD C.^50.00^9^^D^^| 31.1|ug/mL||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

Example: Re-analysis results

・ PT (The first analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^R^^^| 10.2|sec||N||S||||<ETX><CHK1><CHK2><CR
><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^R^^^| 99.4|%||N||S||||<ETX><CHK1><CHK2><CR><
LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^R^^^| 0.57|||N||S||||<ETX><CHK1><CHK2><CR><L
F>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^R^^^| 0.81|||N||S||||<ETX><CHK1><CHK2><CR>
<LF>
Host <ACK>
IPU <STX>0L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ PT (The re-analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^3^^R^^| 10.2|sec||N||S||||<ETX><CHK1><CHK2><CR
><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^3^^R^^| 99.4|%||N||S||||<ETX><CHK1><CHK2><CR><
LF>

75/123
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^3^^R^^| 0.57|||N||S||||<ETX><CHK1><CHK2><CR><L
F>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^3^^R^^| 0.81|||N||S||||<ETX><CHK1><CHK2><CR>
<LF>
Host <ACK>
IPU <STX>0L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ PT (The final analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^R^^| 10.2|sec||N||S||||<ETX><CHK1><CHK2><CR
><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^R^^| 99.4|%||N||S||||<ETX><CHK1><CHK2><CR><
LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^R^^| 0.57|||N||S||||<ETX><CHK1><CHK2><CR><L
F>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^R^^| 0.81|||N||S||||<ETX><CHK1><CHK2><CR>
<LF>
Host <ACK>
IPU <STX>0L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ APTT (The first analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|5|^^^051^APTT sec^100.00^1^R^^^| 27.4|sec||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ APTT (The re-analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><

76/123
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|5|^^^051^APTT sec^100.00^3^^R^^| 27.4|sec||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ APTT (The final analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E^^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|5|^^^051^APTT sec^100.00^9^^R^^| 27.4|sec||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

7.4.4. When analysis is performed 2 times


Example: Analysis result (The first analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^^^| 10.2|sec||N||||||20110328150948<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^^^| 99.4|%||N||||||20110328150948<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^^^| 0.57|||N||||||20110328150948<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^^^| 0.81|||N||||||20110328150948<ETX><CHK1><C
HK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^1^^^| 27.4|sec||N||||||20110328150948<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^1^^^| 7.9|sec||N||||||20110328150948<ETX><CHK1><
CHK2><CR><LF>

77/123
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^1^^^| 632.9|mg/dL||N||||||20110328150948<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

Example: Analysis result (The second analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^1^^^| 7.9|sec||N||||||20110328150948<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^100.00^1^^^| 632.9|mg/dL||N||||||20110328150948<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

Example: Analysis result (The final analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^| 10.2|sec||N||||||20110328150948<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^^| 99.4|%||N||||||20110328150948<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^^| 0.57|||N||||||20110328150948<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^^| 0.81|||N||||||20110328150948<ETX><CHK1><C
HK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^9^^^| 27.4|sec||N||||||20110328150948<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^9^^^| 7.9|sec||N||||||20110328150948<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^9^^^| 632.9|mg/dL||N||||||20110328150948<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>

78/123
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

7.4.5. Item unit output (if there is a comment record)


* If [HOST_ASTM_OutputErrorInformation] and [HOST_ASTM_OutputCurveErrorInformation] are
enabled under [Service settings] - [Other service settings]

* Comment records are output when the [Service settings] - [Other service settings] -
[HOST_ASTM_OutputCommentRecord] setting is enabled.

PT
IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX
><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^R^^| 10.2|sec||N||S||||<ETX><CHK1><CHK2><C
R><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^R^^| 99.4|%||N||S||||<ETX><CHK1><CHK2><CR>
<LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^R^^| 0.57|||N||S||||<ETX><CHK1><CHK2><CR><
LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^R^^| 0.81|||N||S||||<ETX><CHK1><CHK2><CR
><LF>
Host <ACK>
IPU <STX>0C|1|I|CAL^044^20110328^1^502501|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1C|2|I|LOT^040^527501|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2C|3|I|QC^040^201103281354^^502701\QC^040^201103281354^^512601|I<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

APTT
IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX
><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^051^APTT sec^100.00^9^^^| 27.4|sec||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>5C|1|I|QC^050^201103281354^^502701\QC^050^201103281354^^512601|I<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>

79/123
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

Fbg
IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX
><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^9^^^| 8.5|sec||N||S||||<ETX><CHK1><CHK2><CR
><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^100.00^9^^^| 588.2|mg/dL||N||S||||<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>6C|1|I|CAL^062^20110328^1^502501|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7C|2|I|LOT^060^538050,A2008|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0C|3|I|QC^060^201103281354^^502701\QC^040^201103281354^^512601|I<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

All Assay Groups


IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^R^^| 10.2|sec||N||F||||20110328135407<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^R^^| 99.4|%||N||F||||20110328135407<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^R^^| 0.57|||N||F||||20110328135407<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^R^^| 0.81|||N||F||||20110328135407<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^9^^^| 27.4|sec||N||F||||20110328135407<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^9^^^| 8.5|sec||N||F||||20110328135407<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^9^^^| 588.2|mg/dL||N||F||||20110328135407<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
80/123
IPU <STX>3C|1|I|CAL^042^20110328^1^502501|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4C|1|I|CAL^044^20110328^1^502501|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5C|2|I|LOT^040^527501|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6C|3|I|QC^040^201103281354^^502701\QC^040^201103281354^^512601|I<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7C|1|I|QC^050^201103281354^^502701\QC^050^201103281354^^512601|I<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0C|1|I|CAL^062^20110328^1^502501|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1C|2|I|LOT^060^538050,A2008|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2C|3|I|QC^060^201103281354^^502701\QC^040^201103281354^^512601|I<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

7.4.6. Output the path name to the image data


When "Service settings" - "Other service settings" - "HOST_ASTM_OutputGraphicData" setting is ON.
IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX>
<CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^| 10.2|sec||N||||||20110328135056<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^040^Normal^^^^^|PNG&R&20110328&R&2011_03_28_13_50_000001_040_Nor
mal_100_1.PNG|||||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

7.4.7. When MDA analysis is performed


Communications examples are presented for cases when MDA analysis is performed.

・ [System Settings] - [Device Settings] - [Host] - [Output Targets] setting: All


・ Setting for image data output ([Service settings] - [ Other service settings]
-[HOST_ASTM_OutputGraphicData]): Enabled

7.4.7.1.Example: When MDA analysis is performed 1 time


Dilution ratio used in analysis: MDA 3 points (dilution ratio: 1/1, 1/2, 1/4)

・ Analysis Results (MDA 1st point <dilution ratio: 1/1>)

IPU <ENQ>
Host <ACK>
81/123
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX>
1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^^^||R||||||N<CR><ETX>7F<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^^^^|13.2|sec||N||||||20200214084402<CR><ETX>34
<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^^^^|3.0|%||>||||||20200214084402<CR><ETX>C8<CR
><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^^^^|1.32|||N||||||20200214084402<CR><ETX>44<CR
><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^^^^|3.00|||>||||||20200214084402<CR><ETX>9D<C
R><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_44_0000000000000
01_040_Normal_100_1.PNG<CR><ETX>07<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (MDA 2nd point <dilution ratio: 1/2>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^^^||R||||||N<CR><ETX>7F<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^50.00^1^^^^|13.2|sec||N||||||20200214084402<CR><ETX>08
<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^50.00^1^^^^|6.0|%||>||||||20200214084402<CR><ETX>9F<CR
><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^50.00^1^^^^|1.32|||N||||||20200214084402<CR><ETX>18<CR
><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^50.00^1^^^^|3.00|||N||||||20200214084402<CR><ETX>81<
CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_44_000000000000
001_040_Normal_050_1.PNG <CR><ETX>0B<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (MDA 3rd point <dilution ratio: 1/4>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
82/123
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^^^||R||||||N<CR><ETX>7F<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^25.00^1^^^^|13.2|sec||N||||||20200214084402<CR><ETX>0A
<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^25.00^1^^^^|12.0|%||>||||||20200214084402<CR><ETX>CE<C
R><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^25.00^1^^^^|1.32|||N||||||20200214084402<CR><ETX>1A<CR
><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^25.00^1^^^^|3.00|||N||||||20200214084402<CR><ETX>83<
CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_44_000000000000
001_040_Normal_025_1.PNG<CR><ETX>0D<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (final analysis: MDA analysis results)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^^^||R||||||N<CR><ETX>7F<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^^9^^^^|****.*|sec||N||||||20200214084402<CR><ETX>59<C
R><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^^9^^^^|7.0|%||>||||||20200214084402<CR><ETX>B5<CR><L
F>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^^9^^^^|***.**|||N||||||20200214084402<CR><ETX>69<CR><
LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^^9^^^^|***.**|||N||||||20200214084402<CR><ETX>D5<CR
><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^MDA^^^^^^|PNG&R&20200214&R&2020_02_14_08_44_00000000000000
1_040_MDA.PNG||59<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

83/123
7.4.7.2.Example: When the re-dilution analysis is the MDA analysis
Dilution ratio used in analysis:
・ First analysis: 1/1
・ Re-dilution: MDA 3 points (dilution ratio: 1/1, 1/2, 1/4)

・ Analysis Results (first analysis – normal analysis <dilution ratio:1/1>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000008^01^000000000000006^A^E^^||R||||||N<CR><ETX>D0<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^D^^^|13.2|sec||N||||||20200214085814<CR><ETX>
83<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^D^^^|13.2|%||N||||||20200214085814<CR><ETX>5A
<CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^D^^^|1.32|||N||||||20200214085814<CR><ETX>93<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^D^^^|7.64|||N||||||20200214085814<CR><ETX>0
A<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_58_000000000000
006_040_Normal_100_1.PNG|||||||||20200214085814<CR><ETX>1C<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (re-dilution – MDA 1st point <dilution ratio:1/1>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000008^01^000000000000006^A^E^^||R||||||N<CR><ETX>D0<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^3^^D^^|13.2|sec||N||||||20200214085814<CR><ETX>
85<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^3^^D^^|13.2|%||N||||||20200214085814<CR><ETX>5C
<CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^3^^D^^|1.32|||N||||||20200214085814<CR><ETX>95<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^3^^D^^|7.64|||N||||||20200214085814<CR><ETX>0
C<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_58_000000000000
006_040_Normal_100_1_D.PNG|||||||||20200214085814<CR><ETX>BF<CR><LF>

84/123
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (re-dilution – MDA 2nd point <dilution ratio:1/2>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000008^01^000000000000006^A^E^^||R||||||N<CR><ETX>D0<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^50.00^3^^D^^|13.2|sec||N||||||20200214085814<CR><ETX>5
9<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^50.00^3^^D^^|26.4|%||N||||||20200214085814<CR><ETX>36<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^50.00^3^^D^^|1.32|||N||||||20200214085814<CR><ETX>69<C
R><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^50.00^3^^D^^|7.64|||N||||||20200214085814<CR><ETX>E0
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_58_000000000000
006_040_Normal_050_1_D.PNG|||||||||20200214085814<CR><ETX>C3<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (re-dilution – MDA 3rd point <dilution ratio:1/4>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000008^01^000000000000006^A^E^^||R||||||N<CR><ETX>D0<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^25.00^3^^D^^|13.2|sec||N||||||20200214085814<CR><ETX>5
B<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^25.00^3^^D^^|52.8|%||N||||||20200214085814<CR><ETX>3B<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^25.00^3^^D^^|1.32|||N||||||20200214085814<CR><ETX>6B<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^25.00^3^^D^^|7.64|||N||||||20200214085814<CR><ETX>E2
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_58_000000000000
006_040_Normal_025_1_D.PNG|||||||||20200214085814<CR><ETX>C5<CR><LF>
Host <ACK>

85/123
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (final analysis:MDA analysis results)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000008^01^000000000000006^A^E^^||R||||||N<CR><ETX>D0<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^^9^^D^^|****.*|sec||N||||||20200214085814<CR><ETX>A8<
CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^^9^^D^^|30.8|%||N||||||20200214085814<CR><ETX>48<CR><
LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^^9^^D^^|***.**|||N||||||20200214085814<CR><ETX>B8<CR>
<LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^^9^^D^^|***.**|||N||||||20200214085814<CR><ETX>24<CR
><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^MDA^^^^^^|PNG&R&20200214&R&2020_02_14_08_58_00000000000000
6_040_MDA.PNG|||||||||20200214085814<CR><ETX>6E<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

7.4.7.3.Example: When the first analysis is the MDA analysis (with re-analysis)
Dilution ratio used in analysis:
・ First analysis: MDA 3 points (dilution ratio: 1/1, 1/2, 1/4)
・ Re-dilution: 1/1

・ Analysis Results (first analysis - MDA 1st point <dilution ratio:1/1> first time)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^^^||R||||||N<CR><ETX>81<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^D^^^|13.2|sec||N||||||20200214084516<CR><ETX>
3A<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^D^^^|3.0|%||>||||||20200214084516<CR><ETX>CE<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^D^^^|1.32|||N||||||20200214084516<CR><ETX>4A<
CR><LF>
Host <ACK>

86/123
IPU <STX>7R|4|^^^044^PT INR^100.00^1^D^^^|3.00|||>||||||20200214084516<CR><ETX>A
3<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_45_000000000000
002_040_Normal_100_1.PNG|||||||||20200214084516<CR><ETX>10<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis - MDA 1st point <dilution ratio:1/1> second time)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^^^||R||||||N<CR><ETX>81<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^D^^^|13.2|sec||N||||||20200214084516<CR><ETX>
3A<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^D^^^|3.0|%||>||||||20200214084516<CR><ETX>CE<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^D^^^|1.32|||N||||||20200214084516<CR><ETX>4A<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^D^^^|3.00|||>||||||20200214084516<CR><ETX>A
3<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_45_000000000000
002_040_Normal_100_2.PNG|||||||||20200214084516<CR><ETX>10<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis - MDA 1st point <dilution ratio:1/1> average)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^^^||R||||||N<CR><ETX>81<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^2^D^^^|13.2|sec||N||||||20200214084516<CR><ETX>
3B<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^2^D^^^|3.0|%||>||||||20200214084516<CR><ETX>CF<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^2^D^^^|1.32|||N||||||20200214084516<CR><ETX>4B<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^2^D^^^|3.00|||>||||||20200214084516<CR><ETX>A
4<CR><LF>
87/123
Host <ACK>
IPU <STX>0R|5|^^^040^Average^^^^^^|PNG&R&20200214&R&2020_02_14_08_45_000000000000
002_040_Average_100.PNG|||||||||20200214084516<CR><ETX>23<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis – MDA 2nd point <dilution ratio: 1/2> first time)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^^^||R||||||N<CR><ETX>81<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^50.00^1^D^^^|13.2|sec||N||||||20200214084516<CR><ETX>0
E<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^50.00^1^D^^^|6.0|%||>||||||20200214084516<CR><ETX>A5<C
R><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^50.00^1^D^^^|1.32|||N||||||20200214084516<CR><ETX>1E<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^50.00^1^D^^^|3.00|||N||||||20200214084516<CR><ETX>87
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_45_000000000000
002_040_Normal_050_1.PNG|||||||||20200214084516<CR><ETX>13<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis - MDA 2nd point <dilution ratio: 1/2> second time)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^^^||R||||||N<CR><ETX>81<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^50.00^1^D^^^|13.2|sec||N||||||20200214084516<CR><ETX>0
E<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^50.00^1^D^^^|6.0|%||>||||||20200214084516<CR><ETX>A5<C
R><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^50.00^1^D^^^|1.32|||N||||||20200214084516<CR><ETX>1E<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^50.00^1^D^^^|3.00|||N||||||20200214084516<CR><ETX>87
<CR><LF>
Host <ACK>

88/123
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_45_000000000000
002_040_Normal_050_2.PNG|||||||||20200214084516<CR><ETX>14<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis - MDA 2nd point <dilution ratio: 1/2> average)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^^^||R||||||N<CR><ETX>81<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^50.00^2^D^^^|13.2|sec||N||||||20200214084516<CR><ETX>0
F<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^50.00^2^D^^^|6.0|%||>||||||20200214084516<CR><ETX>A6<C
R><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^50.00^2^D^^^|1.32|||N||||||20200214084516<CR><ETX>1F<C
R><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^50.00^2^D^^^|3.00|||N||||||20200214084516<CR><ETX>88
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Average^^^^^^|PNG&R&20200214&R&2020_02_14_08_45_000000000000
002_040_Average_050.PNG|||||||||20200214084516<CR><ETX>27<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis – MDA 3rd point <dilution ratio:1/4> first time)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^^^||R||||||N<CR><ETX>81<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^25.00^1^D^^^|13.2|sec||N||||||20200214084516<CR><ETX>1
0<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^25.00^1^D^^^|12.0|%||>||||||20200214084516<CR><ETX>D4<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^25.00^1^D^^^|1.32|||N||||||20200214084516<CR><ETX>20<C
R><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^25.00^1^D^^^|3.00|||N||||||20200214084516<CR><ETX>89
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_45_000000000000
002_040_Normal_025_1.PNG|||||||||20200214084516<CR><ETX>15<CR><LF>
89/123
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis - MDA 3rd point <dilution ratio:1/4> second time)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^^^||R||||||N<CR><ETX>81<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^25.00^1^D^^^|13.2|sec||N||||||20200214084516<CR><ETX>1
0<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^25.00^1^D^^^|12.0|%||>||||||20200214084516<CR><ETX>D4<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^25.00^1^D^^^|1.32|||N||||||20200214084516<CR><ETX>20<C
R><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^25.00^1^D^^^|3.00|||N||||||20200214084516<CR><ETX>89
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_45_000000000000
002_040_Normal_025_2.PNG|||||||||20200214084516<CR><ETX>16<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis - MDA 3rd point <dilution ratio:1/4> average)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^^^||R||||||N<CR><ETX>81<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^25.00^2^D^^^|13.2|sec||N||||||20200214084516<CR><ETX>1
1<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^25.00^2^D^^^|12.0|%||>||||||20200214084516<CR><ETX>D5<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^25.00^2^D^^^|1.32|||N||||||20200214084516<CR><ETX>21<C
R><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^25.00^2^D^^^|3.00|||N||||||20200214084516<CR><ETX>8A
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Average^^^^^^|PNG&R&20200214&R&2020_02_14_08_45_000000000000
002_040_Average_025.PNG|||||||||20200214084516<CR><ETX>29<CR><LF>
Host <ACK>

90/123
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis MDA analysis results)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1494-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^E^^||R||||||N<CR><ETX>C9<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^^2^D^^^|****.*|sec||N||||||20200214084826<CR><ETX>85<
CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^^2^D^^^|7.0|%||>||||||20200214084826<CR><ETX>19<CR><L
F>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^^2^D^^^|***.**|||N||||||20200214084826<CR><ETX>95<CR>
<LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^^2^D^^^|***.**|||>||||||20200214084826<CR><ETX>EE<C
R><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Average^^^^^^|PNG&R&20200214&R&2020_02_14_08_48_000000000000
002_040_Average.PNG|||||||||20200214084826<CR><ETX>DE<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (re-dilution – normal <dilution ratio:1/1> first time)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1494-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^E^^||R||||||N<CR><ETX>C9<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^3^^D^^|14.2|sec||N||||||20200214084826<CR><ETX>
84<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^3^^D^^|3.0|%||>||||||20200214084826<CR><ETX>18<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^3^^D^^|1.32|||N||||||20200214084826<CR><ETX>94<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^3^^D^^|3.00|||>||||||20200214084826<CR><ETX>E
D<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_48_000000000000
002_040_Normal_100_1_D.PNG<CR><ETX>BA<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>

91/123
Host <ACK>
IPU <EOT>

・ Analysis Results (re-dilution - normal <dilution ratio:1/1> second time)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1494-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^E^^||R||||||N<CR><ETX>C9<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^3^^D^^|14.2|sec||N||||||20200214084826<CR><ETX>
84<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^3^^D^^|3.0|%||>||||||20200214084826<CR><ETX>18<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^3^^D^^|1.32|||N||||||20200214084826<CR><ETX>94<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^3^^D^^|3.00|||>||||||20200214084826<CR><ETX>E
D<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_08_48_000000000000
002_040_Normal_100_2_D.PNG<CR><ETX>BB<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (re-dilution - normal <dilution ratio:1/1> average)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1494-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^E^^||R||||||N<CR><ETX>C9<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^4^^D^^|14.2|sec||N||||||20200214084826<CR><ETX
>85<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^4^^D^^|3.0|%||>||||||20200214084826<CR><ETX>19<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^4^^D^^|1.32|||N||||||20200214084826<CR><ETX>95<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^4^^D^^|3.00|||>||||||20200214084826<CR><ETX>E
E<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Average^^^^^^|PNG&R&20200214&R&2020_02_14_08_48_00000000000
0002_040_Average_D.PNG|||||||||20200214084826<CR><ETX>DE<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>

92/123
IPU <EOT>

・ Analysis Results (final analysis results)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1494-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000002^01^000000000000002^A^E^^||R||||||N<CR><ETX>C9<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^D^^|14.2|sec||N||||||20200214084826<CR><ETX
>8A<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^D^^|3.0|%||>||||||20200214084826<CR><ETX>1E
<CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^D^^|1.32|||N||||||20200214084826<CR><ETX>9A
<CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^D^^|3.00|||>||||||20200214084826<CR><ETX>F
3<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Average^^^^^^|PNG&R&20200214&R&2020_02_14_08_48_00000000000
0002_040_Average_D.PNG|||||||||20200214084826<CR><ETX>DE<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>03<CR><LF>
Host <ACK>
IPU <EOT>

7.4.7.4.Example: When the re-analysis is the MDA analysis


Dilution ratio used in analysis:
・ First analysis: MDA 3 points (dilution ratio:1/1, 1/2, 1/4)
・ Re-analysis: MDA 3 points (dilution ratio: 1/1, 1/2, 1/4)

・ Analysis Results (first analysis – MDA 1st point <dilution ratio:1/1>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^^^||R||||||N<CR><ETX>7F<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^R^^^|13.2|sec||N||||||20200214091002<CR><ETX
>34<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^R^^^|3.0|%||>||||||20200214091002<CR><ETX>C8<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^R^^^|1.32|||N||||||20200214091002<CR><ETX>44<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^R^^^|3.00|||>||||||20200214091002<CR><ETX>9
D<CR><LF>
Host <ACK>

93/123
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_09_10_000000000000
001_040_Normal_100_1.PNG<CR><ETX>07<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis – MDA 2nd point <dilution ratio:1/2>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^^^||R||||||N<CR><ETX>7F<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^50.00^1^R^^^|13.2|sec||N||||||20200214091002<CR><ETX>
08<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^50.00^1^R^^^|6.0|%||>||||||20200214091002<CR><ETX>9F<
CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^50.00^1^R^^^|1.32|||N||||||20200214091002<CR><ETX>18<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^50.00^1^R^^^|3.00|||N||||||20200214091002<CR><ETX>81
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_09_10_000000000000
001_040_Normal_050_1.PNG <CR><ETX>0B<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis – MDA 3rd point <dilution ratio:1/4>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^^^||R||||||N<CR><ETX>7F<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^25.00^1^R^^^|13.2|sec||N||||||20200214091002<CR><ETX>
0A<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^25.00^1^R^^^|12.0|%||>||||||20200214091002<CR><ETX>CE
<CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^25.00^1^R^^^|1.32|||N||||||20200214091002<CR><ETX>1A<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^25.00^1^R^^^|3.00|||N||||||20200214091002<CR><ETX>83
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_09_10_000000000000
001_040_Normal_025_1.PNG<CR><ETX>0D<CR><LF>
94/123
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (first analysis MDA analysis results)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^^^||R||||||N<CR><ETX>7F<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^^2^R^^^|****.*|sec||N||||||20200214091002<CR><ETX>59<
CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^^2^R^^^|7.0|%||>||||||20200214091002<CR><ETX>B5<CR><
LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^^2^R^^^|***.**|||N||||||20200214091002<CR><ETX>69<CR>
<LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^^2^R^^^|***.**|||N||||||20200214091002<CR><ETX>D5<C
R><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Average^^^^^^|PNG&R&20200214&R&2020_02_14_09_10_00000000000
0001_040_Average.PNG||59<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (re-analysis –MDA 1st point <dilution ratio:1/1>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^E^^||R||||||N<CR><ETX>D0<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^3^^R^^|13.2|sec||N||||||20200214091014<CR><ETX
>85<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^3^^R^^|13.2|%||N||||||20200214091014<CR><ETX>5
C<CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^3^^R^^|1.32|||N||||||20200214091014<CR><ETX>95<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^3^^R^^|7.64|||N||||||20200214091014<CR><ETX>0
C<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_09_10_000000000000
001_040_Normal_100_1_R.PNG|||||||||20200214091014<CR><ETX>BF<CR><LF>
Host <ACK>

95/123
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (re-analysis - MDA 2nd point <dilution ratio:1/2>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^E^^||R||||||N<CR><ETX>D0<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^50.00^3^^R^^|13.2|sec||N||||||20200214091014<CR><ETX>
59<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^50.00^3^^R^^|26.4|%||N||||||20200214091014<CR><ETX>36
<CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^50.00^3^^R^^|1.32|||N||||||20200214091014<CR><ETX>69<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^50.00^3^^R^^|7.64|||N||||||20200214091014<CR><ETX>E0
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_09_10_000000000000
001_040_Normal_050_1_R.PNG|||||||||20200214091014<CR><ETX>C3<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

・ Analysis Results (re-analysis - MDA 3rd point <dilution ratio:1/4>)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^E^^||R||||||N<CR><ETX>D0<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^25.00^3^^R^^|13.2|sec||N||||||20200214091014<CR><ETX>
5B<CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^25.00^3^^R^^|52.8|%||N||||||20200214091014<CR><ETX>3B
<CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^25.00^3^^R^^|1.32|||N||||||20200214091014<CR><ETX>6B<
CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^25.00^3^^R^^|7.64|||N||||||20200214091014<CR><ETX>E2
<CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^Normal^^^^^^|PNG&R&20200214&R&2020_02_14_09_10_000000000000
001_040_Normal_025_1_R.PNG|||||||||20200214091014<CR><ETX>C5<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>

96/123
Host <ACK>
IPU <EOT>

・ Analysis Results (final analysis: MDA analysis results on re-analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-11^01234^^^CN-6000^BP542727||||||||E1394-97<CR><ETX
>1F<CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<CR><ETX>EB<CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^000000000000001^A^E^^||R||||||N<CR><ETX>D0<CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^^9^^R^^|****.*|sec||N||||||20200214091014<CR><ETX>A8<
CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^^9^^R^^|30.8|%||N||||||20200214091014<CR><ETX>48<CR>
<LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^^9^^R^^|***.**|||N||||||20200214091014<CR><ETX>B8<CR>
<LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^^9^^R^^|***.**|||N||||||20200214091014<CR><ETX>24<C
R><LF>
Host <ACK>
IPU <STX>0R|5|^^^040^MDA^^^^^^|PNG&R&20200214&R&2020_02_14_09_10_00000000000000
1_040_MDA.PNG|||||||||20200214091014<CR><ETX>6E<CR><LF>
Host <ACK>
IPU <STX>1L|1|N<CR><ETX>04<CR><LF>
Host <ACK>
IPU <EOT>

7.4.8. Output the path name to the Delayed cross mixing image data
IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ETX
><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^051^APTT sec^100.00^A^^^| 10.2|sec||N||||||20110328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|1|^^^050^ DelayedMixing^^^^^|PNG&R&20110328&R&2011_03_28_13_50_000001
_050_DelayedMixing.PNG |||||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|1|^^^050^ DelayedMixing | CSV&R&20110328&R&2011_03_28_13_50_000001_05
0_DelayedMixing.csv | || ||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

97/123
7.4.9. When the measurement of platelet aggregation is output
“Analysis result (Maximum aggregation rate)”, “Evaluation parameter”, and “Aggregation curve” are
output as analysis result of the measurement of platelet aggregation. The setting of the IPU determines
whether or not to output the “Evaluation parameter” and the “Aggregation curve”.

Table 7.4.9-1: Output conditions of “Analysis result (Maximum aggregation rate)”, “Evaluation parameter”,
and “Aggregation curve” in the measurement of platelet aggregation
Service Settings
HOST_ASTM_Output
AggregationAnalysisInformation HOST_ASTM_OutputGraphicData Output data
(output platelet aggregation (output image data)
analysis information)
“Analysis Parameter (maximum
OFF ON/OFF
aggregation %)”
“Measurement parameter
ON OFF (maximum aggregation %)” +
“evaluation parameter”
“Measurement parameter
(maximum aggregation %)” +
ON ON
“evaluation parameter” +
“aggregation curve”

A graph of result records is output with measurement results for each parameter in the platelet
aggregation analysis result record superimposed.
The results for all measurement parameters for which graph display is enabled are output for the
overlay graph result records. Results for all measurement parameters are output in block units for
“measurement parameter (maximum Aggregation %)”, “evaluation parameter”, and “aggregation curve”.
The configuration and output sequence of result reports is shown in “Figure 7.4.9-1 Result Records for
Overlay Graphs”.

98/123
Output Target Output condition
Measurement Parameter 1 (maximum aggregation %) Output required
Repeat sufficiently ・
for the number of ・
measurement Measurement parameter n (maximum aggregation %)
parameters Analysis parameter 1 When outputting analysis
~Abs PPP parameters
~Abs PRP_s
~Abs PRP_e
~Abs PRP_min
~Max
~Fin
~Lag Phase
~AUC
~VMax
~VMax Time
~VMax Angle

Repeat sufficiently ・
for the number of Analysis parameter n
measurement ~Abs PPP
parameters ~Abs PRP_s
~Abs PRP_e
~Abs PRP_min
~Max
~Fin
~Lag Phase
~AUC
~VMax
~VMax Time
~VMax Angle
Analysis ~APAL - When outputting analysis
parameter Platelet Aggregation Level of ADP parameters
Platelet ~CPAL - When outputting Platelet
Aggregation Platelet Aggregation Level of COL Aggregation Level
Level
Repeat Aggregation curve 1 (image data: Platelet) When outputting aggregation
sufficiently for the ・ curves
number of ・
measurement Aggregation curve n (image data: Platelet)
Aggregation curve_activator 1 (image data: - When outputting aggregation
parameters
SuperImpose) curves
Repeat in activator ・ - When the overlay graph is
units ・ “parameter display”
Aggregation curve_activator n (image data:
SuperImpose)
Aggregation curve (image data: SuperImpose) - When outputting aggregation
curves
- When the overlay graph is
“overall display”

Figure 7.4.9-1 Result Records for Overlay Graphs

99/123
7.4.9.1.Communication example: Analysis result (Maximum aggregation rate)
The communication example is based on the following conditions.
Output of analysis result : ADP-1% and ADP-10%
Output of platelet aggregation level : Not output
Output of image data : Not output

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CS-5100^00-20^10000001^^^CS-5100^BY990757||||||||E1394-97<ETX
><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1|| ^ ^ 1234567890^M^||R||||||N<ETX><CHK1><CHK2><CR><L
F>
Host <ACK>
IPU <STX>4R|1|^^^671^ ADP-1%^100.00^A^^^| 5.7|%||N||||||20170328135056<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|1|^^^672^ ADP-10%^100.00^A^^^| 5.7|%||N||||||20170328135056<ETX>
<CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

7.4.9.2.Communication example: Analysis result (Maximum aggregation rate) + Evaluation


parameter
The communication example is based on the following conditions.
Output of analysis result : ADP-1% and ADP-10%
Output of evaluation parameter : Output
Output of platelet aggregation level : Not output
Output of image data : Not output

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CS-5100^00-20^10000001^^^CS-5100^BY990757||||||||E1394-97<ETX
><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1|| ^ ^ 1234567890^M^||R||||||N<ETX><CHK1><CHK2><CR><L
F>
Host <ACK>
IPU <STX>4R|1|^^^671^ ADP-1%^100.00^A^^^| 5.7|%||N||||||20170328135056<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|1|^^^672^ ADP-10%^100.00^A^^^| 5.7|%||N||||||20170328135056<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|2|^^^671^~Abs PPP^^A^^^| 67|mOD||||||||20170328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|3|^^^671^~Abs PRP_s^^A^^^| 496|mOD||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|4|^^^671^~Max^^A^^^| 39.6|%||||||||20170328135056<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
100/123
IPU <STX>1R|5|^^^671^~Fin^^A^^^| 7.3|%||||||||20170328135056<ETX><CHK1><CHK
2><CR><LF>
Host <ACK>
IPU <STX>2R|6|^^^671^~Lag Phase^^A^^^| 34.6|sec||||||||20170328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|7|^^^671^~AUC^^A^^^|37179|%*sec||||||||20170328135056<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|8|^^^671^~VMax^^A^^^| 65|%/sec||||||||20170328135056<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|9|^^^671^~VMax Time^^A^^^| 24|sec||||||||20170328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|10|^^^671^~VMax Angle^^A^^^| 61.61|deg||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|2|^^^672^~Abs PPP^^A^^^| 67|mOD||||||||20170328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|3|^^^672^~Abs PRP_s^^A^^^| 496|mOD||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|4|^^^672^~Max^^A^^^| 39.6|%||||||||20170328135056<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
IPU <STX>2R|5|^^^672^~Fin^^A^^^| 7.3|%||||||||20170328135056<ETX><CHK1><CHK
2><CR><LF>
Host <ACK>
IPU <STX>3R|6|^^^672^~Lag Phase^^A^^^| 34.6|sec||||||||20170328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|7|^^^672^~AUC^^A^^^|37179|%*sec||||||||20170328135056<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|8|^^^672^~VMax^^A^^^| 65|%/sec||||||||20170328135056<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|9|^^^672^~VMax Time^^A^^^| 24|sec||||||||20170328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|10|^^^672^~VMax Angle^^A^^^| 61.61|deg||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

7.4.9.3.Communication example: Analysis result (Maximum aggregation rate) + Evaluation


parameter + Aggregation curve
The communication example is based on the following conditions.
Output of analysis result : ADP-1% and ADP-10%
Output of evaluation parameter : Output
Output of platelet aggregation level : Not output
Output of image data : Output
Overlay graph : Overall display

IPU <ENQ>
Host <ACK>

101/123
IPU <STX>1H|\^&|||CS-5100^00-20^10000001^^^CS-5100^BY990757||||||||E1394-97<ETX
><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1|| ^ ^ 1234567890^M^||R||||||N<ETX><CHK1><CHK2><CR><L
F>
Host <ACK>
IPU <STX>4R|1|^^^671^ ADP-1%^100.00^A^^^| 5.7|%||N||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|1|^^^672^ ADP-10%^100.00^A^^^| 5.7|%||N||||||20170328135056<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|2|^^^671^~Abs PPP^^A^^^| 67|mOD||||||||20170328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|3|^^^671^~Abs PRP_s^^A^^^| 496|mOD||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|4|^^^671^~Max^^A^^^| 39.6|%||||||||20170328135056<ETX><CHK1><CHK
2><CR><LF>
Host <ACK>
IPU <STX>1R|5|^^^671^~Fin^^A^^^| 7.3|%||||||||20170328135056<ETX><CHK1><CHK
2><CR><LF>
Host <ACK>
IPU <STX>2R|6|^^^671^~Lag Phase^^A^^^| 34.6|sec||||||||20170328135056<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|7|^^^671^~AUC^^A^^^|37179|%*sec||||||||20170328135056<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|8|^^^671^~VMax^^A^^^| 65|%/sec||||||||20170328135056<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|9|^^^671^~VMax Time^^A^^^| 24|sec||||||||20170328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|10|^^^671^~VMax Angle^^A^^^| 61.61|deg||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|2|^^^672^~Abs PPP^^A^^^| 67|mOD||||||||20170328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|3|^^^672^~Abs PRP_s^^A^^^| 496|mOD||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|4|^^^672^~Max^^A^^^| 39.6|%||||||||20170328135056<ETX><CHK1><CHK
2><CR><LF>
Host <ACK>
IPU <STX>2R|5|^^^672^~Fin^^A^^^| 7.3|%||||||||20170328135056<ETX><CHK1><CHK
2><CR><LF>
Host <ACK>
IPU <STX>3R|6|^^^672^~Lag Phase^^A^^^| 34.6|sec||||||||20170328135056<ETX><CHK1
><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|7|^^^672^~AUC^^A^^^|37179|%*sec||||||||20170328135056<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|8|^^^672^~VMax^^A^^^| 65|%/sec||||||||20170328135056<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>

102/123
IPU <STX>6R|9|^^^672^~VMax Time^^A^^^| 24|sec||||||||20170328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|10|^^^672^~VMax Angle^^A^^^| 61.61|deg||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|1|^^^671^Platelet|PNG&R&20170328&R&2017_03_28_13_50_56_1234567890_671
_Platelet.Png|||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|2|^^^672^Platelet|PNG&R&20170328&R&2017_03_28_13_52_01_1234567890_672
_Platelet.Png|||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|3|^^^^Superimpose|PNG&R&20170328&R&2017_03_28_14_10_23_1234567890__S
uperimpose.Png|||||||||20170328141023<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

7.4.9.4.Communication example: Analysis result (Maximum aggregation rate) + Evaluation


parameter (including Platelet Aggregation Level) + Aggregation curve
The communication example is based on the following conditions.
Output of analysis result : ADP-1% and ADP-10%
Output of evaluation parameter : Output
Output of platelet aggregation level : Output
Output of image data : Output
Overlay graph : Parameter display

IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CS-5100^00-20^10000001^^^CS-5100^BY990757||||||||E1394-97<ETX
><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1|| ^ ^ 1234567890^M^||R||||||N<ETX><CHK1><CHK2><CR><L
F>
Host <ACK>
IPU <STX>4R|1|^^^671^ ADP-1%^100.00^A^^^| 5.7|%||N||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|1|^^^672^ ADP-10%^100.00^A^^^| 5.7|%||N||||||20170328135056<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|2|^^^671^~Abs PPP^^A^^^| 67|mOD||||||||20170328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|3|^^^671^~Abs PRP_s^^A^^^| 496|mOD||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|4|^^^671^~Max^^A^^^| 39.6|%||||||||20170328135056<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
IPU <STX>1R|5|^^^671^~Fin^^A^^^| 7.3|%||||||||20170328135056<ETX><CHK1><CHK
2><CR><LF>
Host <ACK>
IPU <STX>2R|6|^^^671^~Lag Phase^^A^^^| 34.6|sec||||||||20170328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|7|^^^671^~AUC^^A^^^|37179|%*sec||||||||20170328135056<ETX><CHK1><

103/123
CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|8|^^^671^~VMax^^A^^^| 65|%/sec||||||||20170328135056<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|9|^^^671^~VMax Time^^A^^^| 24|sec||||||||20170328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|10|^^^671^~VMax Angle^^A^^^| 61.61|deg||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|2|^^^672^~Abs PPP^^A^^^| 67|mOD||||||||20170328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|3|^^^672^~Abs PRP_s^^A^^^| 496|mOD||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|4|^^^672^~Max^^A^^^| 39.6|%||||||||20170328135056<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
IPU <STX>2R|5|^^^672^~Fin^^A^^^| 7.3|%||||||||20170328135056<ETX><CHK1><CHK
2><CR><LF>
Host <ACK>
IPU <STX>3R|6|^^^672^~Lag Phase^^A^^^| 34.6|sec||||||||20170328135056<ETX><CHK
1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|7|^^^672^~AUC^^A^^^|37179|%*sec||||||||20170328135056<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|8|^^^672^~VMax^^A^^^| 65|%/sec||||||||20170328135056<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|9|^^^672^~VMax Time^^A^^^| 24|sec||||||||20170328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|10|^^^672^~VMax Angle^^A^^^| 61.61|deg||||||||20170328135056<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|21|^^^^~APAL^^A^^^| 10.0|||||||||20170328135415<ETX><CHK1><CHK2>
<CR><LF>
Host <ACK>
IPU <STX>1R|1|^^^671^Platelet|PNG&R&20170328&R&2017_03_28_13_50_56_1234567890_671
_Platelet.Png|||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|2|^^^672^Platelet|PNG&R&20170328&R&2017_03_28_13_52_01_1234567890_672
_Platelet.Png|||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|3|^^^^Superimpose|PNG&R&20170328&R&2017_03_28_14_10_23_1234567890__S
uperimpose_ADP.Png|||||||||20170328141023<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

104/123
7.4.10. Output when multiple vial sets are measured (when output targets are “Final Data only”)
* Output final measurement results for the number of measured vial sets if the setting for “Service
settings” - “Instrument Settings” - “Host” - “Output Targets” is “Final Data only”, and the setting for
“Service settings” - “System Settings” - “HOST_ASTM_OutputMultipleVials” is enabled.

Example: When 3 vial sets were measured


- Measurement result (final: 1st set)
IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ET
X><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF
>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^| 10.2|sec||N||||||20110328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

- Measurement results (final: 2nd set)


IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ET
X><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF
>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^| 10.2|sec||N||||||20110328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

- Measurement results (final: 3rd set)


IPU <ENQ>
Host <ACK>
IPU <STX>1H|\^&|||CN-6000^00-03^10000001^^^CN-6000^XXXXXXXX||||||||E1394-97<ET
X><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF
>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^| 10.2|sec||N||||||20110328135056<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

105/123
8. Parameter Code List

Parameter Code uses alphanumeric characters.

8.1. Coagulation parameter code list

Table 8.1-1: Parameter Code Lists

Parameter Parameter
Assay Group Assay Group
Code Code
PT 04X Reserved 52X
APTT 05X PPP 53X
Fbg 06X ADP 54X
Reserved 07X Epi 55X
TTO 08X Col 56X
Hpt 09X Ris 57X
Reserved 10X Ara 58X
II 12X LA NR 59X
V 15X FDP 60X
VII 17X DDimer 61X
VIII 18X PFDP 62X
IX 19X UFDP 63X
X 20X Reserved 64X
XI 21X VWF Ag Med 65X
XII 22X VWF Ag Low 66X
Orgaran 23X ADP% 67X
Arixtra 24X Epi% 68X
PCcl 25X FM 69X
BXT 26X Col% 70X
LA1 27X Ris% 71X
LA2 28X Ara% 72X
PSAc 29X Reserved 73X
AT 30X Edoxaban Std 78X
APL 31X Edoxaban Low 79X
Plg 32X VWF Rco 80X
PC(Chrom) 33X AcR R 81X
Heparin 34X AcR Act 82X
Reserved 36X AcR OVB 83X
C1 Inh 37X FVL R 84X
VIIICh Low 38X FVL Act 85X
VIIICh 39X FVL OVB 86X
XIIIch 40X VWF Ac Low 87X
LA R 41X VWF Ac Med 88X
PCAT NR 42X VWF Rco Low 90X
PCAT 43X FPS 91X
PCAT0 44X Dabigatran Std 92X
FVPCAT NR 45X Dabigatran Low 93X
FVPCAT 46X Apixaban Std 94X
FVPCAT0 47X Apixaban Low 95X
IXCh High 48X Rivaroxaban Std 96X
IXCh Low 49X Rivaroxaban Low 97X
Reserved 50X Reserved 000
TT 51X Reserved 999

000: There is no analysis parameter for the inquired sample.


999: There is no information on the inquired sample.

When outputting measurement results, this is the 1-digit host ID set in [Assay Group Settings] - [Assay
Parameter Settings] - [Basic] - [Assay Parameter Host ID], attached to “X”.

106/123
In the case of order inquiry, order information texts and the analysis data format of formula, “X” is
replaced with 0.

* New analysis parameters may be added in the future. Please prepare a receive program that ignores
the data when receiving the codes other than the codes shown in the parameter code list .

By default, [Assay Group Settings] - [Assay Parameter Settings] - [Basic] - [Assay Parameter Host ID]
is set as follows:

1: Time/ dOD
2. Activity percent/ concentration
3: Ratio
4: INR

* By default the setting for Assay Parameter Host ID platelet aggregation parameters PPP and PRP is
1 or 2, and either one becomes the absorbance (mOD).
* The Assay Parameter Host ID set for the formula is fixed to “1”. (Except where platelet aggregation
level parameter code is 676-677, 706-707 *2)
* New analysis parameters may be added in the future. Please prepare a receive program that ignores
the data when receiving the codes other than the above-mentioned codes.

* When performing host orders for orders concerning formulas, the assay parameter host ID for the
formula parameter is not ordered from the host, and the order inquiry is made to the host with the
original parameter for the calculation.
In that case, the assay group output sequence is to output the parameters which can be measured
with the reagents placed on board, in ascending order of the assay parameter host ID registered in
the “Assay Group Settings”.
Example) When LA R = LA 1 / LA2,
If the reagents which allow measurement of “LA R” have been placed on board, the order is not made
with the assay parameter host ID for “LA R”, but with the assay parameter host IDs “LA 1” and “LA 2”.

Note:
Refer to ”9.3 Table of Output Content Differences” because the specifications for
“parameter output sequence” differ between models.

107/123
Table 8.1-2: Parameter Code Lists(Result)
Test code Test name Units Format Test code Test name Units Format
041 PT sec sec ####.# 381 VIIICh L dOD dOD #.####
042 PT % % ####.# 382 *1 VIIICh L % % ####.#
043 PT R ###.## 391 VIIICh dOD dOD #.####
044 PT INR ###.## 392 *1 VIIICh % % ####.#
045 *1 DFbg mg/dL ##### 401 XIIIch dOD dOD #.####
051 APTT sec ####.# 402 *1 XIIIch % % ####.#
061 Fbg sec sec ####.# 411 LA R ###.##
062 *1 Fbg C. mg/dL ####.# 421 PCAT NR ###.##
081 TTO sec Sec ####.# 431 PCAT sec sec ####.#
082 TTO % % ####.# 441 PCAT0 sec sec ####.#
084 TTO INR ###.## 451 FVPCAT NR ###.##
091 Hpt sec sec ####.# 461 FVPCATsec sec ####.#
092 Hpt % % ####.# 471 FVPCAT0sec sec ####.#
093 Hpt INR ###.## 481 IXCh dOD dOD #.####
121 II sec sec ####.# 482 *1 IXCh % % ####.#
122 *1 II % % ####.# 491 IXCh L dOD dOD #.####
151 V sec sec ####.# 492 *1 IXCh L % % ####.#
152 *1 V% % ####.# 511 TT sec sec ####.#
171 VII sec sec ####.# 531 PPP mOD #####
172 *1 VII % % ####.# 541 ADP_min mOD #####
181 VIII sec sec ####.# 542 ADP_s mOD #####
182 *1 VIII % % ####.# 551 Epi_min mOD #####
191 IX sec sec ####.# 552 Epi_s mOD #####
192 *1 IX % % ####.# 561 Col_min mOD #####
201 X sec sec ####.# 562 Col_s mOD #####
202 *1 X% % ####.# 571 Ris_min mOD #####
211 XI sec sec ####.# 572 Ris_s mOD #####
212 *1 XI % % ####.# 581 Ara_min mOD #####
221 X sec sec ####.# 582 Ara_s mOD #####
222 *1 X% % ####.# 591 LA NR ###.##
231 Orgaran dOD dOD #.#### 601 FDP dOD dOD #.####
232 Orgaran C. IU/mL #.## 602 FDP C. ug/mL ####.#
241 Arixtra dOD dOD #.#### 611 DD dOD dOD #.####
242 Arixtra C. ug/mL #.## 612 DD C. ug/mL ####.#
251 PC cl sec sec ####.# 621 PFDP dOD dOD #.####
252 *1 PC cl % % ####.# 622 PFDP C. ug/mL ####.#
261 BXT sec sec ####.# 631 U-FDP dOD #.####
271 LA1 sec sec ####.# 632 U-FDP C. ug/mL ###.##
281 LA2 sec sec ####.# 651 VWFAgM dOD dOD #.####
291 PS Ac sec sec ####.# 652 *1 VWFAgM % % ####.#
292 *1 PS Ac % % ####.# 661 VWFAgL dOD dOD #.####
301 AT dOD dOD ##.### 662 *1 VWFAgL % % ####.#
302 *1 AT % % ####.# 671 ADP 2% % ####.#
311 APL dOD dOD ##.### 676 *2 ADP 1% % ####.#
312 *1 APL % % ####.# 677 *2 ADP 10% % ####.#
321 Plg dOD dOD ##.### 681 Epi 5% % ####.#
322 *1 Plg % % ####.# 691 FM dOD dOD #.####
331 PC dOD dOD ##.### 692 FM C. ug/mL ####.#
332 *1 PC % % ####.# 701 Col 2% % ####.#
341 Hep dOD dOD ##.### 706 *2 Col 1% % ####.#
342 Hep C. IU/mL ###.## 707 *2 Col 5% % ####.#
371 C1 dOD dOD ##.### 711 Ris 1.2% % ####.#
372 *1 C1 % % ####.# 721 Ara 1% % ####.#

108/123
Test code Test name Units Format Test code Test name Units Format
801 VWFRCo dOD dOD #.#### 911 FPS dOD dOD #.####
802 *1 VWF RCo % % ###.## 912 *1 FPS % % ####.#
811 AcR R ###.## 921 Dab sec sec ####.#
821 AcRAct sec sec ####.# 922 Dab C. ng/mL ####.#
831 AcROVB sec sec ####.# 931 DabL sec sec ####.#
841 FVL R ###.## 932 DabL C. ng/mL ####.#
851 FVL Act sec sec ####.# 941 Api dOD dOD #.####
861 FVL OVB sec sec ####.# 942 Api C. ng/mL ###.##
871 VWFAcL dOD dOD #.#### 951 ApiL dOD dOD #.####
872 *1 VWFAcL % % ####.# 952 ApiL C. ng/mL ###.##
881 VWFAcM dOD dOD #.#### 961 Riv dOD dOD #.####
882 *1 VWFAcM % % ####.# 962 Riv C. ng/mL ###.##
901 VWFRCoL dOD dOD #.#### 971 RivL dOD dOD #.####
902 *1 VWFRCoL % % ####.# 972 RivL C. ng/mL ###.##

*1 Units can be changed. Check the unit in [Calculation Method] from [Assay Group Settings] - [Assay
Parameter Settings] setting and, if changed, set the unit format of the host computer according to the Table
7.4.8.4-3: Number of digits in the unit format.

Table 8.1-3: Number of digits in the unit format


Unit Format
% ####.#
mg/dL ####.#
g/L ###.##
IU/mL ##.###
IU/dL ####.#

*2 Platelet Aggregation Level

109/123
8.2. Immuno parameter code list
Only CN-6500/CN-3500 are applicable.
Table 8.2-1:Parameter Code Lists

Parameter Parameter
Parameter name Parameter name
Code Code
TM A0X PIC A2X
TAT A1X tPAI-C A3X

In the case of order inquiry and order information texts, “X” is replaced with 0.
When outputting measurement results, this is the 1-digit host ID corresponding to the unit of output, attached
to “X”. The 1-digit host ID is the below.
1: Count
2: Concentration

Table 8.2-2: Parameter Code Lists (Result)

Test Test
Test name Units Format Test name Units Format
Code Code
A01 TM ##### A21 PIC #####
A02 TM TU/mL TU/mL ####.# A22 PIC ug/mL ug/mL ##.###
A11 TAT ##### A31 tPAI-C #####
A12 TAT ng/mL ng/mL ####.# A32 tPAI-C ng/mL ng/mL ####.#

110/123
9. Appendix

9.1. TCP/IP Communication

A. Software

1) Data link/ network/ transport layers


The TCP port number used for communications with the host is 6000.
This setting can be changed on the setting screen of the CN-series device.
2) Session layer
The connection is established with the host computer as the server and the CN-series device as a client.
The CN-series device runs a connection check on startup. If connection fails, reconnection attempts will be
made at a set interval.
If the server goes down after the connection is established, the connection will be reattempted after the
server recovers.

111/123
9.2. Re-analysis orders

A. Conditions for a re-analysis order

- When the [System Settings] - [Device Settings] - [Host] - [Inquire re-analysis] setting is enabled.
- When the [System Settings] - [Device Settings] - [Host] - [Connect to host computer] setting is enabled, and
output subject is set to “All”.
- Measurement mode is “Normal”.
- Instrument status is one of “Processing”, “Int. Ready”, or “Int. Ready (flashing)”.
- For the settings for re-analysis, redilution analysis, and reflex, satisfy the condition for “Table 9.2-1:
Conditions Performing a Re-analysis Order for Re-analysis Settings”.

Re-analysis orders are not placed under the conditions below.


- When instrument status is “Not.Ready”
(Because measurement is not possible even if the re-analysis order is accepted. Also, as the results of tests in
progress are sent in a batch when an emergency stop occurs, this avoids increasing the communication time
by waiting for re-analysis instructions).
- For quality control analysis
- When analysis mode is “Micro-Sample Mode”
- When results are output manually
- When the Sample No. is “0”
※Only for CN-6500/CN-3500
- When first measurement order is not for only immuno parameter
(When first measurement order is a mixture of coagulation parameter and immuno parameter,
CN-6500/CN-3500 inquires about re-analysis for only coalguration parameter. And, if CN-6500/CN-3500
receives re-analysis order for immuno parameter, it ignores the designation of immuno parameter and does
not re-analysis.)

(Note) [System Settings] - [Operation Settings] - [Automatic Order Inquiry] - [Perform Automatic Order
Inquiry] is a setting for First Measurement, so it is unrelated to re-analysis orders.

Table 9.2-1: Conditions Performing a Re-analysis Order for Re-analysis Settings


Re-analysis
Conditions for a re-analysis order
settings
Repeat - In assay group settings, [Perform re-analysis] is checked and a condition
analysis expression is set. And, the analysis count is 1.
- When a first measurement order is performed, “R: Do Not Execute Re-analysis” is
not specified in the option settings from the host.
- When manual order registration is used, the check for [Re-analysis] has not been
removed.
Redilution - In assay group settings, [Perform redilution analysis] is checked and a condition
analysis expression is set. And, the dilution ratio for the first measurement differs from the
dilution ratio set by the condition expression.
- When a first measurement order is performed, “D: Do Not Execute Redilution
Analysis” is not specified in the option settings from the host.
- When manual order registration is used, the check for [Redilution] has not been
removed.
Reflex - In assay group settings, [Perform Reflex Test] is checked and a Reflex Condition
has been set that uses the parameters for which measurement results have been
output.
- When a first measurement order is performed, “F: Do Not Execute Reflex Test” is
not specified in the option settings from the host.
- When manual order registration is used, the check for [Reflex] has not been
removed.

112/123
Example flows of communication sequence and simulation results for the“Inquiry of Analysis Order”and the
“Analysis Order” when the “Device Settings” - “Host” - “Inquire re-analysis” setting in the “System Settings” is
ON are shown as follows;

B. When there is no re-analysis order (Batch transmission)

CN-series device HOST

Inquiry of the first analysis order (PT,APTT,Fbg)


Order Information (PT,APTT,Fbg)

Transmitting analysis results (PT,APTT,Fbg)

Restricted Stand-by time(*1)


transmission Inquiry of re-analysis order (PT,APTT,Fbg)
(*3)
No re-analysis order information (*2)

Transmitting final analysis results(PT,APTT,Fbg)

C. When there is a re-analysis order (Batch transmission)


1)When measurement order is for only coalgulation parameter

CN-series device HOST

Inquiry of the first analysis order (PT,Fbg)


Order Information (PT,Fbg)

Transmitting analysis results (PT,Fbg)

Inquiry of re-analysis order (PT,Fbg)

Order information (PT)


(Re-analysis)

Transmitting re-analysis results(PT)

Transmitting final analysis results(PT,Fbg)

113/123
※Only for CN-6500/CN-3500
2)When measurement order is a mixture of coagulation parameter and immuno parameter
①When the re-analysis order matches the inquiry of re-analysis order

CN-6500/CN-3500 HOST

Inquiry of the first analysis order (PT,TAT)


Order Information(PT,TAT)

Transmitting analysis results (PT,TAT)

(*6)
Inquiry of re-analysis order (PT)

Order Information(PT)(re- analysis)

Transmitting re-analysis results (PT)

Transmitting final analysis results (PT,TAT)

②When the re-analysis order does not matches the inquiry of re-analysis order
(When the re-analsysis order is for coagualation parameter and immuno parameter)

CN-6500/CN-3500 HOST

Inquiry of the first analysis order (PT,TAT)


Order Information(PT,TAT)

Transmitting analysis results (PT,TAT)

(*6)
Inquiry of re-analysis order (PT)

Order Information (PT, TAT)(re-analysis)

(*7)
Transmitting re-analysis results (PT)

Transmitting final analysis results (PT,TAT)

114/123
③When the re-analysis order does not matches the inquiry of re-analysis order
(When the re-analsysis order is for only immuno parameter)

CN-6500/CN-3500 HOST

Inquiry of the first analysis order (PT,TAT)


Order Informatio(PT,TAT)

Transmitting analysis results (PT,TAT)

(*6)
Inquiry of re-analysis order (PT)

Order Information(TAT)(re-analysis)

(*7)

Transmitting final analysis results (PT,TAT)

115/123
D. When there is a re-analysis order (For each assay group)
1) When measurement order is for only coagulation parameter

CN-series device HOST

Inquiry of the first analysis


Order Information (PT,Fbg)

Transmitting analysis results (PT)

Inquiry of re-analysis order

Order information (APTT)(Reflex)

Transmitting analysis results (Fbg)


(*4)

Re-analysis results (APTT)

Final analysis results(PT, Fbg, APTT)(*5)

※Only for CN-6500/CN-3500


2)When measurement order is a mixture of coagulation parameter and immuno parameter

CN-6500/CN-3500 HOST

Inquiry of the first analysis order


Order Information (PT,TAT)

Transmitting analysis results(PT)

Inquiry of re-analysis order

Order Information(APTT)(Reflex)

Transmitting analysis (TAT)

(*6)

Transmitting re-analysis results (APTT)

Transmitting final analysis results (PT,TAT,APTT)(*7)

116/123
(*1) Stand-by time:
Analysis results and Inquiry of re-analysis order. Default (System settings):1 sec
* Actual standby time may be longer then the set time.

(*2) No re-analysis order information


Please return“000” for the test code, if the HOST judges “do no perform re-analysis” for the inquiry
of re-analysis order.

(*3) Restricted transmission:


In this category, a separate automatic transmission will never interrupt, but manual transmission
may interrupt.

(*4) No Inquiry:
If the “Fbg” is not targeted for re-analysis, the inquiry of re-analysis order is not performed.

(*5) Final analysis results:


When the “Service settings” - “Other service settings” - “HOST_ASTM_OutputItemUnit” setting is
ON and “HOST_ASTM_OutputFinalDataAfterItemUnit” setting is OFF, the final analysis results
are not transmitted.

※Only for CN-6500/CN-3500:


(*6) No inquiry:
CN-6500/CN-3500 doesn’t iquire about re-analysis because TAT is immuno parameter.

(*7) Ignores the designation of re-analysis order:


CN-6500/CN-3500 ignores the designation of re-analysis order because TAT is immuno parameter.

117/123
9.3. Difference List of Transmission Content
The CN-series device outputs different content by the device specification from other analyzer.
Please observe the precautions for use and ensure the operation of sending and receiving systems works to use
the host computer.

<CS-5100>
Contents CS-5100 CN-series device Remarks
Analyzer name “CS-5100” is output. (For the CN-6000) Check that the host
(Header records) “CN-6000” is output. computer can receive
(For the CN-3000) CN-series device’s output.
“CN-3000” is output.
(For the CN-6500)
“CN-6500” is output.
(For the CN-3500)
“CN-3500” is output.
PS Code “BY990757” is output. (For the CN-6000) Check that the host
(Header records) The CN-6000 Model PS code from “Table 9.3-1: computer can receive the
PS Codes for Each Destination (CN-6000, “Table 9.3-1: PS Codes for
CN-3000)” is output. Each Destination
(For the CN-3000) (CN-6000, CN-3000)”
The CN-3000 Model PS code from “Table 9.3-1: and “Table 9.3-2: PS
PS Codes for Each Destination (CN-6000, Codes for Each
CN-3000)” is output. Destination (CN-6500,
(For the CN-6500) CN-3500)” PS codes.
The CN-6500 Model PS code from “Table
9.3-2: PS Codes for Each Destination
(CN-6500, CN-3500)” is output.
(For the CN-3500)
The CN-3500 Model PS code from “Table
9.3-2: PS Codes for Each Destination
(CN-6500, CN-3500)” is output.
Version number The following is output, “E1394-97” is output. -
(Header records) according to the format.
For E1394-97: “E1394-97”
For E1394-95: “1”
Rack number “L□□□□□” is output. “L□□□□□” is output. * Transfer samples
(Transfer sample) (Where □□□□□ is a 5-digit (Where □□□□□ is a 5-digit number. Zero analyzed with the
number. If the number of digits padding to the right if less than 5 digits.)* Sysmex transportation
is less than 5, the missing digits system (the connection
will be right padded with zeros). type under [System
Settings] - [LAS
Connection] is Type 6)
are the same as ordinary
samples. The number
becomes L□□□□□
(□ indicates a 5-digit
alphanumeric character)
when [Service settings] -
[Other service settings] -
[LAS_RackNo_Replace]
is set to “Enabled”.
Rack number "□□□□□□"is output. Not supported. -
(Platelet (“□” is a half-width space)
aggregation Data
Integration line)
Tube Position Regular rack: 01-10 Regular rack: 01-06 * Transfer samples
Number STAT sample: 01-05 STAT sample: □□ analyzed with the
Reagent Table: □□ Reagent Table: □□ Sysmex transportation
Transfer sample: Fixed to 01 Transfer sample: Fixed to 01* system (the connection
Platelet aggregation Data Platelet aggregation Data Integration line: Not type under [System
Integration line: □□ supported Settings] - [LAS
(“□” is a half-width space) (“□” is a half-width space) Connection] is Type 6)
are 01 - 10.
Test Code 3-byte alpha-numeric (000-999) 3-byte string (000-ZZZ) Check that the host
computer works when test
A00- is used when CN-6500/CN-3500 use code is a number or a
immuno parameter. character string.

118/123
Parameter output Output variably. The sequence is Output variably. -
sequence as follows: If there are multiple assay group parameters
subject to output, the parameter output
Output in ascending order of host sequence is in ascending order of the Host ID
ID registered in [Assay Group registered in [Assay Group Settings].
Settings]. If there are any duplicated host IDs, the
sequence is the display sequence in the Assay
Group Settings.
If research parameters + reportable
parameters are output, the sequence is
reportable parameters then research
parameters.
Data or Up to 6-byte Up to 10-byte Check that the host
measurement (In case of the path name to the (In case of the path name to the image data, computer receives a
value image data, up to 60 characters) up to 60 characters) maximum of 10-byte of
(Result Record) data when
Data value of 7 bytes or more is used when CN-6500/CN-3500 use
CN-6500/CN-3500 use immuno parameter. immuno parameter.
Image data path File paths to the image data are File paths to the image data files are output Check that the host
output output when output is enabled in when [Service settings] - [Other service computer receives the file
(Result record) [System Settings]. settings] - [HOST_ASTM_OutputGraphicData] path for image data, and
is enabled. can import images.
Image data file The format of the test date/time The format of the test date/time appended to -
name appended to the file names the file names.
YYYY_MM_DD_HH_MM_SS YYYY_MM_DD_HH_MM_SS
When enabled with [Service Setup] - [Other
Service Settings] -
[HOST_ASTM_GraphicData_Time
Format(HH_MM)], the following format is
used for output.
YYYY_MM_DD_HH_MM
Host order Only “Automatic order” is supported. -
method “Manual order” is not supported.
Automatic order: This order method queries the measurement parameter from
the Sample No. at the start of measurement.
Manual order: This order method queries the Sample No. and the measurement
parameters from the rack number and position.

119/123
<CS-2400/CS-2500>
Contents CS-2400/2500 CN-series device Remarks
Analyzer name “CS-2400” or “CS-2500” is output. (For the CN-6000) Check that the host
(Header records) “CN-6000” is output. computer can receive
(For the CN-3000) CN-series device’s output.
“CN-3000” is output.
(For the CN-6500)
“CN-6500” is output.
(For the CN-3500)
“CN-3500” is output.
PS Code “AR780981” or “BV981798” is (For the CN-6000) Check that the host
(Header records) output. The CN-6000 Model PS code from “Table 9.3-1: computer can receive the
PS Codes for Each Destination (CN-6000, “Table 9.3-1: PS Codes for
CN-3000)” is output. Each Destination
(For the CN-3000) (CN-6000, CN-3000)”
The CN-3000 Model PS code from “Table 9.3-1: and “Table 9.3-2: PS
PS Codes for Each Destination (CN-6000, Codes for Each
CN-3000)” is output. Destination (CN-6500,
(For the CN-6500) CN-3500)” PS codes.
The CN-6500 Model PS code from “Table
9.3-2: PS Codes for Each Destination
(CN-6500, CN-3500)” is output.
(For the CN-3500)
The CN-3500 Model PS code from “Table
9.3-2: PS Codes for Each Destination
(CN-6500, CN-3500)” is output.
Version number The following is output, “E1394-97” is output. -
(Header records) according to the format.
For E1394-97: “E1394-97”
For E1394-95: “1”
Rack number Not supported. “L□□□□□” is output. Check that the host
(Transfer sample) (□□□□□is a 5-digit number. If the number of computer can receive
digits is less than 5, the missing digits will be “L□□□□□”.
right padded with zeros).*
* Transfer samples
analyzed with the
Sysmex
transportation system
(the connection type
under [System
Settings] - [LAS
Connection] is Type 6)
are the same as
ordinary samples. The
number becomes
L□□□□□ (□ indicates a
5-digit alphanumeric
character) when
[Service settings] -
[Other service
settings] -
[LAS_RackNo_Replace]
is set to “Enabled”.
Rack number "□□□□□□" is output. Not supported. -
(Platelet (“□” is a half-width space)
aggregation Data
Integration line)
Tube Position Regular rack: 01-10 Regular rack: 01-06 * Transfer samples
Number STAT sample: 01-05 STAT sample: □□ analyzed with the
Reagent Table: □□ Reagent Table: □□ Sysmex transportation
Transfer samples: Not supported Transfer sample: Fixed to 01* system (the connection
Platelet aggregation Data Platelet aggregation Data Integration line: Not type under [System
Integration line: □□ supported Settings] - [LAS
(“□” is a half-width space) (“□” is a half-width space) Connection] is Type 6)
are 01 - 10.
Test Code 3-byte alpha-numeric (000-999) 3-byte string (000-ZZZ) Check that it works when it
is a number or a character
A00- is used when CN-6500/CN-3500 use string.
immuno parameter.

120/123
Parameter output Output variably. The sequence is Output variably. -
sequence as follows: If there are multiple assay group parameters
subject to output, the parameter output
Output in ascending order of host sequence is in ascending order of the Host ID
ID registered in [Assay Group registered in [Assay Group Settings].
Settings]. If there are any duplicated host IDs, the
sequence is the display sequence in the Assay
Group Settings.
If research parameters + reportable
parameters are output, the sequence is
reportable parameters then research
parameters.
Data or Up to 6-byte Up to 10-byte Check that the host
measurement (In case of the path name to the (In case of the path name to the image data, computer receives a
value image data, up to 60 characters) up to 60 characters) maximum of 10-byte of
(Result Record) data when
Data value of 7 bytes or more is used when CN-6500/CN-3500 use
CN-6500/CN-3500 use immuno parameter. immuno parameter.
Image data path File paths to the image data are File paths to the image data files are output -
output output when output is enabled in when [Service settings] - [Other service
(Result record) the System Settings. settings] - [HOST_ASTM_OutputGraphicData]
is enabled.
Image data file The format of the test date/time The format of the test date/time appended to -
name appended to the file names the file names.
YYYY_MM_DD_HH_MM_SS YYYY_MM_DD_HH_MM_SS
When enabled with [Service Setup] - [Other
Service Settings] -
[HOST_ASTM_GraphicData_Time
Format(HH_MM)], the following format is
used for output.
YYYY_MM_DD_HH_MM
Host order Only “Automatic order” is supported. -
method “Manual order” is not supported.
Automatic order: This order method queries the measurement parameter from
the Sample No. at the start of measurement.
Manual order: This order method queries the Sample No. and the measurement
parameters from the rack number and position.

Table 9.3-1: PS Codes for Each Destination (CN-6000, CN-3000)

Model Country/region Side CP PS Code


Yes BT988463
「Sysmex Japan」
CN-6000 No BP542727
Other than “Sysmex Japan” - CV339763
Yes AE136003
「Sysmex Japan」
CN-3000 No CP152931
Other than “Sysmex Japan” - CL971029

121/123
Table 9.3-2: PS Codes for Each Destination (CN-6500, CN-3500)

Model Country/region Side CP PS Code


Yes AX896795
「Sysmex Japan」
CN-6500 No CW756893
Other than “Sysmex Japan” - BH584790
Yes CK321362
「Sysmex Japan」
CN-3500 No BY296982
Other than “Sysmex Japan” - AX818921

<The function which can be used by making the setting effective>


The following function is the function which can be used by making the setting effective. Please ensure the host computer
is adaptable to the function to be validated.
Function Record CN-series device
Inquiry Item Details of Request To output the same items as those in CS-5100, CS-2400/CS-2500, enable
Information Record the [HOST_ASTM_OrderInquiry_AnalyzableAssay] setting for [Service
12.1.5 Universal test settings] – [Other service settings].
ID Disabled: Available assay group parameter settings are output.
Enabled: Only parameters that can be measured are output from available
parameter settings.*

* Parameters that can be measured vary depending on the setting for


[System Settings] – [Analysis in Case of Insufficient Reagent].
Analysis in case of Measurable parameters
insufficient reagents
Suspend Analysis The reagent used in analysis has been set
Suspend the test regardless of whether there is a remaining
amount or not.
Skip the test The reagent is set for use in analysis and there is
a remaining amount.
Inquiry of Request Information The following is output.
re-analysis Record N: Real time inquiry of the first analysis
12.1.13 Requested C: Real time inquiry of re-analysis
information status * Inquiry of re-analysis is performed only when the “Device Settings” -
code “Host” - “Inquire re-analysis” setting in the “System Settings” is ON.

Measurement Test Order Record The following is output.


Mode 9.4.3 Specimen ID N: Normal Mode
9.4.4 Instrument M: Micro-sampling Mode
specimen ID Other than above value: Sample is aspirated according to the instrument
setting.
* Measurement Mode can be set only when “Service settings” - “Other
service settings” - “HOST_ASTM_SpecifyMeasurementMode” setting is ON.
Replication Test Order Record Specifies replication.
9.4.5 Universal Test In case of test order record, “1” or “2” is selectable.
ID If replication is specified “2” or over, the analysis is performed twice (2).
Resulted record Other than the values above or no replication are specified: Sample is
10.1.3 Universal Test analyzed according to instrument setting.
ID * Replication can be set only when “Service settings” - “Other service
settings” - “HOST_ASTM_SpecifyReplication” setting is ON.
Measurement Test Order Record The following is output.
Time 9.4.5 Universal Test M:Measurement Time(Main)
ID S:Measurement Time(Sub)
Other than the values above or no measurement time are specified: Sample
is analyzed according to instrument setting.
* Measurement Time setting is selectable only when the “Service settings” -
“Other service settings” - “HOST_ASTM_SwitchMeasurementTime” setting
is ON.
Error Resulted record The following is output.
Information 10.1.7 Result Evaluation Information:Outputs a string of characters consisting of the
abnormal flags combination of the evaluation error code and error message, which is
enclosed between square brackets “[]”.
Instrument Error Information:Outputs a string of characters consisting of
the error code and error message, which is enclosed between square
brackets “[]”.

122/123
* The Evaluation Information and the Instrument Error Information are
output only if the setting of “Service settings” - “Other service settings” -
“HOST_ASTM_OutputErrorInformation” settings is turned ON.
There are the evaluation errors that are output by either the mask
characters or the numeric values
Sample Info Detailed result Sample information is only output if there is a HIL flag or sample volume
output records flag.
HIL flag * HIL flags are only output when set to Enabled under “Service settings” –
Sample Volume Flag “Other service settings” - “HOST_ASTM_OutputHILFlags”.
* Sample volume flags are only output when set to Enabled under “Service
settings” - “Other service settings” - “HOST_ASTM_OutputSampleVolumeFlags”.
Detailed result HIL levels are output for samples with HIL flags.
records * HIL levels are only output when “Service settings” - “Other service
HIL level settings” - “HOST_ASTM_OutputHILFlags” and
“HOST_ASTM_OutputHILLevel” are Enabled.
Image data Detailed result Output the file path for image data.
output records * Image data is only output when set to Enabled under “Service settings” -
Image data “Other service settings” - “HOST_ASTM_OutputGraphicData”.
* The format of the date/time appended to the file names is set to
YYYY_MM_DD_HH_MM_SS. The YYYY_MM_DD_HH_MM format is only
set when [Service Setup] - [Other Service Settings] -
[HOST_ASTM_GraphicData_TimeFormat(HH_MM)] is enabled.
Parameter Detailed result Display the status of measurement results.
unit output records S: Indicates that result is for a measurement parameter unit.
10.1.9 Analysis result F: Indicates that result combines the parameters ordered for the sample.
status * Item units are only output when set to Enabled under “Service settings” -
“Other service settings” - “HOST_ASTM_OutputItemUnit”.
Comment Comment Record Output 1 line comment records after the result record for each
record output measurement parameter.
* Comment records are only output when set to Enabled under “Service
settings” - “Other service settings” - “HOST_ASTM_OutputCommentRecord”.
Research Result records Output the result record and comment record for the research parameters.
parameters Comment records * Result records and comment records for the research parameters are only
output output when set to “enabled” under [Service Setup] - [Other Service
Settings] - [HOST_ASTM_OutputCommentRecord].

End of Documents