CS-2500 ASTM

Host Interface Specifications


for out of US

Revision 3.02

Revised on November 18, 2020

Sysmex Corporation
Sysmex is a registered trademark of SYSMEX CORPORATION in the USA, in Germany and other countries.

1/74
Table of Contents

1. General...........................................................................................4
2. Scope .............................................................................................4
3. Revision History...................................................................................5
4. Terminology ......................................................................................8
5. Communication Specifications ......................................................................9
5.1. Physical Layer (Hardware) ...................................................................... 10
5.1.1. Serial Connection .........................................................................10
5.1.2. Ethernet Connection ......................................................................12
5.2. Data Link Layer (Transmission Protocol).......................................................... 13
5.2.1. Communication status .....................................................................13
5.2.2. Establishment phase ......................................................................14
5.2.3. Transfer phase ............................................................................15
5.2.4. Termination phase ........................................................................16
5.2.5. Time out .................................................................................16
5.3. Presentation Layer.............................................................................. 17
5.3.1. Messages, Record and Field ................................................................17
5.3.2. Communication Protocol ...................................................................19
5.3.3. Details of Record ..........................................................................22
6. Examples of Communication ......................................................................47
6.1. Inquiry of Analysis Order (IPU => Host Computer) ................................................ 47
6.2. Inquiry of Re-analysis Order (IPU=> Host) ........................................................ 47
6.3. Analysis Order (Host Computer => IPU) .......................................................... 47
6.3.1. Specify Replication ........................................................................48
6.3.2. Specify Measurement Mode ................................................................48
6.3.3. Specify Measurement Time ................................................................48
6.4. Analysis Results (IPU => Host Computer) ......................................................... 48
6.4.1. When analysis is performed 1 time ..........................................................48
6.4.2. Transmitting Analysis Results (When there is the re-analysis).................................50
6.4.3. When analysis is performed 2 times .........................................................52
6.4.4. Output analysis result for each Assay Group .................................................54
6.4.5. Output the path name to the image data ....................................................55
6.4.6. Output the path name to the Delayed cross mixing image data ................................56
6.4.7. When the measurement of platelet aggregation is output ......................................57
6.4.8. Transmitting Comment records .............................................................61
7. Appendix ........................................................................................62
A. TCP/IP communication ........................................................................... 62
A.1 Software .................................................................................62
B. Parameter Code List ............................................................................. 63
B.1. Parameter codes in inquiry record and request record ..........................................63
B.2. Parameter codes in result record .............................................................65
C. Error Information ................................................................................ 67
D. Re-analysis order ................................................................................ 70
D.1 There is no re-analysis order (Batch transmission): .............................................70
D.2 There is a re-analysis order (Batch transmission): ..............................................70
D.3 There is a re-analysis order (For each assay group): ............................................71
E. Platelet aggregation .............................................................................. 72

2/74
F. Comment Record ................................................................................. 74

3/74
1. General

This document is intended to supply the information that the CS-2500 Automated Blood Coagulation
Analyzer communicates with the host computer using ASTM protocol.

2. Scope

This document describes the Data Communication Specifications for CS-2500 using ASTM
“E1394-97 and E-1381-02” and “ASTM E1394-95 and ASTM E1381-95”

ASTM (the American Society for Testing and Materials), one of the world largest volunteer non-profit
organization, founded in 1898 for the purpose of creating standard regulations for material, products and
system services. This specification conforms to the following two standards:

・Specifications for low level protocols to transfer data between clinical laboratory instruments and
computer systems.
ASTM E1381-02
ASTM E1381-95

In ASTM E1381-95, it is possible to communicate with the host computer conforming to ASTM
E1381-91.

・Standard specifications for transferring data between clinical instruments and computer systems.
ASTM E1394-97
ASTM E1394-95

In ASTM E1394-95, it is possible to communicate with the host computer conforming to ASTM
E1394-91.

4/74
3. Revision History

Revision Date Major Contents of Changes


1.00 May 27, 2016 The first edition is issued based on CS-2500 ASTM Host Interface
Specification Rev.1.20
2.00 April 12, 2017 Whole expression is changed according to "CS-5100 ASTM Host
Interface Specifications Revision 2.00".
Other revising points are following;

5.3.2.2.Analysis information (Host computer -> IPU)


Caution of order for "Analysis Information" and "Analysis Order
Inquiry" is added.

5.3.3.1.Header Record
2) 7.1.5 Sender name or ID
PS code is corrected only "BV981798".

5.3.3.3.Request Information Record


2) 12.1.3 Starting Range ID No.
Rack No.:
The description of rack No. with a reading error is added.

5) 12.1.13 Requested information status code


Inquiry type :
The note of Set up auto re-analysis, auto re-dilution and reflex is
changed.

5.3.3.4.Test Order Record


3) 9.4.4 Instrument Specimen ID
Rack No.:
The description of rack No. with a reading error is added.

4) 9.4.5 Universal Test ID


Measurement Time:
Default set value (OFF) is added.

5.3.3.5.Result record
2.1) When test parameters are output: The parameters for which
analysis orders are registered are output.
a) 10.1.3 Universal Test ID
Replication:
Default set value (OFF) is added.

2.3) The path name to the image data is output only when the
parameters have the image data.
b)10.1.4 Data or measurement values
The description for time of detail is added.

Table 18:List of the information included in the file


Test Date
The analyzed date format is corrected from
“YYYY_MM_DD_HH_MM” to “YYYY_MM_DD_HH_MM_SS”

Caution
Output of Delayed Cross mixing image data
The format of CSV file;
The Character Code is corrected from Shift-JIS to UTF-8.

Table 23:Parameter Code Lists


The tables is changed according to version 01-68.
The description of code "000" and "999" is added.

5/74
Table 24 Parameter Code Lists(Result)
The table is changed according to version 01-68.

Table 25:Analysis error, Result Abnormal Flags, and Display


method of analysis results
"Detector Block Temperature Error" is deleted:

E. Difference List of Transmission Content


The all description of this section is deleted.
And related descriptions are deleted on each chapter.
2.01 July 9, 2018 1) Apixaban, Rivaroxaban, Edoxaban and IX Ch are added to
Parameter Code Lists
2) 5.3.1.2 Added that “IPU needs [CR] at the end of the record
regardless of System Settings.”
3) Table 18:List of the information included in the file
Test Date
The analyzed date format is corrected from
“YYYY_MM_DD_HH_MM_SS” to “YYYY_MM_DD_HH_MM”
4) The description for new analysis parameters is duplicated. It is
omitted.
3.00 June 28, 2019 ■Handling of platelet aggregation function
1) As for inquiry record, the measurement of platelet aggregation
does not correspond to orders from the host computer so that the
note is added in 5.3.3.3 – 3).
2) As for request record, it is added that a request record is handled
as a rack order and calculation for maximum aggregation rate is
not performed because the measurement of platelet aggregation
does not correspond to orders from the host computer. (Added
5.3.3.4 – 4) )
3) In order to correspond to the host output regarding evaluation
parameters in the function of platelet aggregation analysis, the
specification regarding transmission sequence of result records is
added in 5.3.3.5 - 2).
4) As for result record, the specification regarding output of
evaluation parameters in the measurement of platelet aggregation
is added in 5.3.3.5 - 2.2).
5) As for result record, in case a path for image data is output, the
specification regarding individual aggregation curves and
superimposed aggregation curves is added in 5.3.3.5 - 2.4).
6) The specification regarding width and height of image data is
added in 5.3.3.5 - 2.4) - Table 18.
7) Communication examples of the measurement of platelet
aggregation are added in 6.4.7.
8) As for the description of parameter codes in Appendix B, the
section is divided into 2 fields, B.1 and B.2, according to the
difference of use. The definition of parameter codes in inquiry
records and request records is described in “Appendix B.1.
Parameter codes in inquiry records and request records”.
9) The definition of test codes in result records is described in “B.2.:
Parameter codes in result records”. The specification regarding the
measurement of platelet aggregation is also added in the “B.2.:
Parameter codes in result records”.
10) The incorrect description in the extension of image data is
corrected.
(Incorrect description: .PNG Correct description: .Png)
11) Error codes regarding the measurement of platelet aggregation
([PPP Low1], [PPP High1], and [Tube_Position]) are added in
“Table 23: Analysis error, Result Abnormal Flags, and Display
method of analysis results” of “Appendix C. Error information”.
■Result abnormal flags in 10.1.7 of Table 14 of Result record is
changed.
1) “Evaluation Information” and “Instrument Error Information”
are added in “5.3.3.5.Result record”.

6/74
■Image data is converted to YYYY_MM_DD_HH_MM_SS format.
1) As for the communication example of the image file name, the
format of transmission date and time is converted to
YYYY_MM_DD_HH_MM_SS format. (Changes: Table 18,
Section 6.4.5 and 6.4.6)
■Incorrect description in communication example is corrected.
1) Incorrect descriptions of a frame number, a sample number, and
a file extension regarding image data output in the 6.4.5 and the
6.4.6 are corrected.
■Explanation of dilution ratio in image data file is corrected.
1) As for explanation of dilution ratio in image data file, it is
corrected that “Dilution ratio is set in case of Normal or
Average”. (Table 18)
■Incorrect transmission example in case of specifying Measurement
Time (Sub) is corrected.
1) Incorrect transmission example in case of specifying
Measurement Time (Sub) is corrected in “6.3.3.”.
■Some misspellings are corrected.
■Calibration curve information and reagent lot information are
output to the host.
1) 5.3.2.3. “Comment record” is added as record to be transmitted
from IPU to Host computer.
2) 5.3.3.6. Definition of comment record is added.
3) 6.4.8 Transmission example of comment record is added.
4) Appendix F. for explanation of transmission order for comment
record is added.
■Appendix C. Error codes ([Coag% Changed]) is added in Table 25.
3.01 March 14, 2020 1) The following description is added in 5.3.3.4.
When “Rack Assignment” setting is ON, a specified setting of
Measurement Mode may be ignored.
2) The following description is added in 5.3.3.4.
The procedure which an error dialog is displayed in IPU when
order information is not registered in the Test Order Record from
the host computer is added.
3) The following description is added in 5.3.3.5.
Output priority of result abnormal flags depends on a setting in
IPU.
4) The following description is added in B.1.
The description regarding the operation of IPU when receiving
“000” from the host computer is corrected.
5) The following description is added in B.2.
Parameter Code of XIII ch dOD and XIII ch % are corrected.
6) The following description is added in C.
“Display method of analysis results” is changed depending on
Measurement Time when Flat Curve Error is occurred.

3.02 November 18, ■Appendix B.1.


2020 The redundant description is revised.

■Appendix B.2.
1) The Ratio Assay parameters is added to each of PT and APTT.
2) “5:Ratio” is added to default Host ID
3) The incorrect table number is corrected.

7/74
4. Terminology

The definition of the terminology used in this document is described in the following.

Numerics:
Indicates ASCII codes “0” (30h) through “9” (39h)

Alphabet:
Indicates ASCII codes “A” (41h) through “Z” (5Ah) and “a” (61h) through “z” (7Ah)

Alpha-numeric:
Indicates numerical or alphabetical character

8/74
5. Communication Specifications

Communication specifications are based on a layer protocol.

(1) Physical Layer


Specifies the sending and receiving of signals between the IPU and the host computer through physical
and electrical connections
See the section “5.1 Physical Layer (Hardware)”

(2) Data link layer


Specifies the sending and receiving of data by link connections and for each frame between the IPU and
the host computer
See the section “5.2 Data Link Layer (Transmission Protocol)”

(3) Presentation layer


Specifies the messages that are sent and received by the IPU and the host computer
See the section “5.3 Presentation Layer"

Presentation layer Specifies message specifications


Data link layer Specifies link connection and frame specifications
Physical layer Specifies mechanical and electrical specifications

Notes
The information processing Unit (IPU) of the CS-2500 supports connection by the Ethernet and
RS-232C.
From the combinations between the specifications conforming to two standards and connection
types, CS-2500 supports the following three modes.

1. TCP/IP connection
The presentation layer conforms to ASTM E1394-97.
The data link layer conforms to ASTM E1381-02.
The physical layer conforms to IEEE802.3. (*1)

2.ASTM E1381-02 mode (serial interface)


The presentation layer conforms to ASTM E1394-97 mode.
The data link layer and the physical layer conform to ASTM E1381-02.

3.ASTM E1381-95 mode (serial interface)


The presentation layer conforms to ASTM E1394-95.
The data link layer and the physical layer conform to ASTM E1394-95.

Serial connection
Ethernet connection ASTM E1381-02 ASTM E1381-95
mode mode
Presentation layer ASTM E1394-97 ASTM E1394-97 ASTM E1394-95
Data link layer ASTM E1381-02 ASTM E1381-02 ASTM E1381-95
Physical layer *1 IEEE802.3 ASTM E1381-02 ASTM E1381-95

*1: The IEEE802.3 specifications in the physical layer are not described in this document.

9/74
5.1. Physical Layer (Hardware)

5.1.1. Serial Connection

5.1.1.1. Connectors
Although the ASTM standard specifies a DB-25-pin male connector as standard, a DB-9-pin-male I/O
connector, which is located on the rear of IPU, is used to communicate.
Table 1: Connector pin assignment

Pin No. Signal name Signal direction


1 NC
2 Receive data RxD To IPU from host
3 Transmit data TxD From IPU to host
4 Data terminal ready DTR From IPU to host
5 Signal ground SG −
6 Data set ready DSR To IPU from host
7 Request to send RTS From IPU to host
8 Clear to send CTS To IPU from host
9 NC
The control signals are not used with ASTM specifications.
For this reason, do not make connections to unused pins.

5.1.1.2. Signal identification level

Table 2: Signal identification level

Level Data signal Control signal


+3V or more Logic ”0”, start bit ON
-3V or less Logic ”1”, stop bit OFF

10/74
5.1.1.3. Connection cable
Please configure a cable with a D-SUB 9pin female adaptor for connecting to the instrument's D-SUB 9
male connector in accordance with the following connection chart.

IPU Host computer


DB-9 DB-9 DB-25

TxD 3 3 2 TxD

RxD 2 2 3 RxD

SG 5 5 7 SG

RTS 7 7 4 RTS

CTS 8 8 5 CTS

DTR 4 4 20 DTR

DSR 6 6 6 DSR

NC 1

NC 9

5.1.1.4. Interface parameters

Table 3: Interface parameters

Parameter Selection of settings


Baud rate 600, 1200, 2400, 4800, *9600, 14400, 19200 (bps)
Data length 7bits , 8bits*
Stop bit 1bits* , 2bits
Parity None* , Even , Odd

The value with asterisk mark (*) allows conformance with the ASTM standard.
[Note] However, 7-bit data lengths, even or odd parity and two stop bits are recognized by the
ASTM standard for use with special applications.

5.1.1.5. Standard specifications(ASTM E1381-02)

The physical layer of the IPU conforms to ASTM E1381-02 “5. Physical Layer", except for the connector
type.
The IPU uses a D-SUB-9 pin male connector. (The ASTM standard specifies a 25-pin male connector.)

11/74
5.1.2. Ethernet Connection
Ethernet connection conforms to IEEE802.3. The communication cable uses the UTP category 5 cable.
Prepare the connector and cable which apply to the Ethernet connector of the IPU.

12/74
5.2. Data Link Layer (Transmission Protocol)
The data link layer transfers data between systems using a character-based protocol in accordance with
ASTM E 1381-02 “6. Data Link layer”.
This section briefly describes communication control procedures. For details, refer to ASTM E1381-02.
When the TCP/IP connection is intended to be used, the TCP connection is established in prior to the
communication. To establish the TCP connection, the host computer acts as a server and the IPU acts
as a client. The IPU establishes a connection by requesting a connection for the IP address and the port
number that are provided by the host computer.

5.2.1. Communication status


The data link layer consists of the following two states.
 Neutral Status
 Linked status

Transition to each status is accomplished through the following three phases.

(1) Establishment phase


Establishes communication line, and determines the direction of data transfer. In this way, the
sender and the receiver are identified, and the change is made from neutral status to linked status.

(2) Transfer phase


The sender transmits messages to the receiver until all messages are transferred.

(3) Termination phase


Releases the communication line. Changes the communication status both the sender and the receiver
from linked status to neutral status.

Establishment phase

Neutral status Linked status Transfer phase

Termination phase

13/74
5.2.2. Establishment phase
(1) The sender (IPU) sends an [ENQ] signal to the receiver (host computer). To respond to the sender,
the receiver performs the following action.
 Returns an [ACK] signal when the communication is enabled.
 Returns a [NAK] when the communication is disabled.

If the receiver response [NAK], the sender waits for at least 10 seconds before attempting to send
another [ENQ] signal.

Max. 15 sec. Max. 15 sec. 10 sec.


E or more
E E
N N N
IPU IPU Q Q
Q
N
A
A
Host C
Host K
Computer K
Computer

(2) When both the sender and receiver send [ENQ] signals, the host computer must yield control
authority to the IPU.
 The IPU sends [ENQ] again after 1 second.
 The host computer must wait for 20 seconds before sending [ENQ] again.

1 sec.
or more E
IPU N
E E
Q
IPU N N
Q Q E E
N N
E Host
Host Q Q
N Computer
Computer
Q
20sec. or more

14/74
5.2.3. Transfer phase
During the transfer phase, the sender sends messages to the receiver. The transfer phase continues
until all messages have been sent.

(1) Messages are sent in each record with multiple frames. Each frame contains a maximum of 64,000
characters (including frame overhead) in the ASTM E1381-02 mode. If the record is longer than
63,993 characters, it is divided into two or more frames. The maximum number of characters in each
record is set to 240 characters in the ASTM E1381-95 mode.

(2) Multiple records cannot be included in a single frame.

(3) If the record contains the maximum number of characters or less, a frame with the following
structure will be transferred.

[STX] [F#] [Text] [ETX] [CHK1] [CHK2] [CR] [LF]

If the record is longer than the maximum number of characters, it is divided into two or more frames.
The intermediate frame text termination code is [ETB], and the final frame text termination code is
[ETX], as shown below.

[STX] [F#] [Text] [ETB] [CHK1] [CHK2] [CR] [LF]


[STX] [F#] [Text] [ETB] [CHK1] [CHK2] [CR] [LF]
……
[STX] [F#] [Text] [ETX] [CHK1] [CHK2] [CR] [LF]

Symbol Explanation
[STX] Start of a frame
[F#] Frame number
One of the numbers 0 to 7 is used, beginning with 1 and repeating
2,3,4,5,6,7,0.
In case of retransmission, the same frame number is sent.
[Text] ASTM E1394-97 records are used. (See the subsequent section “5.3
Presentation Layer”.)
For this reason, the codes below will not be used.
0x00-0x06,0x08,0x0A,0x0E-0x1F,0x7F,0xFF
[ETB] Control code indicating end of text (for intermediate frame)
[ETX] Control code indicating end of text (for the final frame)
[CHK1] [CHK2] Expressed by characters “0” - “9” and “A” - “F”.
Characters beginning from the character following [STX] and until
[ETB] or [ETX] (including [ETB] or [ETX]) are added in binary. The
2-digit numbers, which represent the least significant 8 bits in
hexadecimal code, are converted to ASCII characters “0” - “9” and “A” -
“F”. The most significant digit is stored in CHK1 and the least
significant digit in CHK2.
[CR] [LF] Control code indicating end of frame

(4) If the receiver has successfully received the frame, and is prepared to receive the next frame, the
receiver responds with [ACK]. After the sender receives [ACK], the sender advances the frame
number and either sends a new frame or transition to the termination phase.

(5) If the receiver fails to receive the frame and is prepared to receive the same frame again, the receiver
responds with [NAK]. After the sender receives [NAK], the sender sends the most recent frame again,
using the same frame number. If a total of 6 attempts to send the frame failed, the sender
transitions to the termination phase and must end sending of the message.

15/74
Max. 15 sec. Max. 15 sec. E
First attempt Sixth attempt O
IPU T
Text frame#1 Text frame#1
・・・
N N
・・・
A A
Host K K
Computer

(6) The IPU processes the response of [EOT] from the Host computer as [ACK]. (Response of [EOT] from
the receiver is usually a request to suspend a transmission to the sender. However, the IPU does not
support this function.)

5.2.4. Termination phase


During the termination phase, the status returns to neutral.
The sender sends the [EOT] to inform the receiver that the message transmission has been completed.
When the sender sends [EOT], the sender transitions to neutral status. When the receiver receives
[EOT], the receiver transitions to neutral status.

5.2.5. Time out


The timer is used to detect a failure to coordinate between the sender and the receiver. The timer is
used as a mean of the recovery for communication line and communication destination device failure.

(1) During the establishment phase, the timer is set when the sender sends [ENQ]. The time out results
if a response of [ACK], [NAK] or [ENQ] is not received within 15 seconds. After time out, the sender
transitions to the termination phase.

(2) During transfer phase, the 15-second timer is set when the sender sends the final character of a
frame. Time out results if no response is received within 15 seconds. After time out, the sender
transitions to the termination phase.
The receiver sets a 30-second timer when first entering the transfer phase or when responding (with
either [ACK] or [NAK]) to a frame. Time out results if the receiver does not receive a frame or [EOT]
from the sender within 30 seconds. After time out, the receiver discards the latest incomplete
message and transitions to the termination phase.

16/74
5.3. Presentation Layer

5.3.1. Messages, Record and Field

5.3.1.1.Messages
In the presentation layer, all data is transmitted using messages. Messages are composed of record
arrays that start with message header record (H) and end with message termination record (L).

5.3.1.2.Records
A record is a series of text, beginning with an ASCII alphabet characters referred to as the identifier,
and ending with [CR].
[CR] is not output at the end of the record if the setting of “Add a [CR] at the end of a record“ in [System
Settings]-[Host Computer]-[Detailed setting] is turned OFF. When IPU receives the record, it needs
[CR] at the end of the record regardless of this setting.
Table 4: Records

Record Type Record Level Contents


identifier
Header Record H 0 Contains the sender and the receiver
information.
Patient information Record P 1 Contains the patient information
Inquiry record Q 1 Requires test order information requesting to
the host computer
Test Order Record O 2 Contains test order information
Test Result Record R 3 Contains test result information
Comment Record C 1−4 Not used.
Manufacturer Information Record M 1−4 Not used
Scientific Information Record S N/A Not used
Message Terminator Record L 0 Indicates the end of the message

 A smaller level number indicates a higher level.


 A higher-level record contains information that is common to all lower-level records.

 All levels other than 0 must be located after higher levels. However, the comment record can be inserted
at any level. They are considered to be one lower level than the preceding record. However, a
consecutive comment records are not allowed.
 Example transmission
H->P->O->R->L… Correct
H->R->L………… Incorrect, because P and O must be transmitted in prior to R

17/74
5.3.1.3.Field
A record is further divided into multiple fields by field delimiters.
A field is identified by its position within a record and has a variable length. The followings are used as
delimiters.
Table 5: Fields

Type Code Description


Field delimiter Vertical bar (|)(7Ch) Separates adjacent field within a record
Repeat delimiter Back slash (¥) (5Ch) Separates multiple numbers of descriptors in a field.
Component delimiter Caret ^ Separates data elements within a field that has a hierarchical
(5Eh) or qualifier nature.
Escape delimiter Ampersand &(26H) Is used within a text field to identify special case operations.

Massage Record 1 Record 2 … Record n

Field 1 Field 2 … Field n CR


Record identifier Delimiter"|" or "¥"

Component1 Component2 … Component n

Delimiter"^"

18/74
5.3.2. Communication Protocol

5.3.2.1.Analysis Order inquiry (IPU-> Host computer)


This protocol is used for CS-2500 to inquire to the host computer analysis information to obtain the
sample information.
Inquiry can be made with keyword of the sample ID Number in CS-2500.

Table 6: Analysis Order Inquiry

IPU Direction Host computer


ENQ →
← ACK
H (Header Record) →
← ACK
Q (Inquiry Record) →
← ACK
L (Message Terminator Record) →
← ACK
EOT →

5.3.2.2.Analysis information (Host computer -> IPU)


This protocol is used for the host computer to respond analysis information against the inquiry made by
the IPU.

Table 7:Analysis Information

IPU Direction Host computer


← ENQ
ACK →
← H (Header Record)
ACK →
← P (Patient Record)
ACK →
← O (Test Order Record)
ACK →
← L (Message Terminator Record)
ACK →
← EOT

19/74
Caution:
Host computer must send the "Analysis Information" in accordance with order of receiving "Analysis
Order Inquiry".
Host computer must send "Analysis Information" within 15 seconds from receiving "Analysis Order
Inquiry".
Host computer can never send the "Analysis Information", if it cannot send it within 15 seconds.

IPU

Analysis Order
Analysis Order
Max Inquiry Max
Inquiry
15sec 2nd Sample 15sec
1st Sample

Analysis Analysis
Host Information Information
computer 1st Sample 2nd Sample

20/74
5.3.2.3. Analysis Results & QC Data (IPU-> Host computer)
This protocol is used for the IPU to transmit the analysis result, the QC data in a real-time
transmission (QC sample Number is “QCXX” and transmitted as similar to the regular sample data),
and the QC data in the manual output.
Table 8:Analysis results and QC data

IPU Direction Host


computer
ENQ →
← ACK
H (Header Record) →
← ACK
P (Patient Record) →
Process sequences ← ACK
repeatedly according O (Test Order Record) →
← ACK
to the number of
R:Result Record →
parameters.
← ACK
C:Comment Record (*) →
← ACK
L (Message Terminator Record) →
← ACK
EOT →

(*)Comment record may be transmitted following the result record depending on the contents of the
result record.

21/74
5.3.3. Details of Record

5.3.3.1.Header Record
[Example of transmission]
 IPU -> Host computer
H|¥^&|||CS-2500^00-00^11001^^^USERINSNO^BV981798||||||||E1394-97<CR>
 Host computer -> IPU
H|¥^&|||||||||||E1394-97<CR>

Table 9:Details of Header record

ASTM IPU -> Host Host -> IPU Max. Remarks


Field Field Name size
(Byte)
7.1.1 Record type H H 1 Fixed
7.1.2 Delimiter definition |¥^& |¥^& 4 Fixed
7.1.3 Message control ID Not used Not used -
7.1.4 Access password Not used Not used -
7.1.5 Sender name or ID Analyzer name^ Not used 8^
Software version^ 13^
Analyzer serial No.^ 5^
Interface version^ ^
Logical ID number ^ ^
Nickname^ 15^
PS code 8
7.1.6 Sender street address Not used Not used -
7.1.7 Reserved field Not used Not used -
7.1.8 Sender Telephone Not used Not used -
No.
7.1.9 Sender Not used Not used -
characteristics
7.1.10 Receiver ID Not used Not used -
7.1.11 Comment Not used Not used -
7.1.12 Processing ID Not used Not used -
7.1.13 ASTM Version No. ASTM Version No. ASTM Version No. 8 E1394-97
or 1
7.1.14 Date and Time of Not used Not used -
message

[Detailed explanation of the fields]


1) 7.1.2 Delimiter definition
"|¥^&" is used as a fixed character string. No field delimiter is required between 7.1.1 and 7.1.2

2) 7.1.5 Sender name or ID


Analyzer name: Fixed as "CS-2500".
Software version: Refers to the software version loaded on the CS-2500
Analyzer Serial No.: Analyzer serial number is output.
Interface version: Not used
Logical ID number: Not used
Nickname: Analyzer nickname is output.
PS code: "BV981798" is output .

3) 7.1.13 ASTM Version No.


If the presentation layer is E1394-97, the version number is “E1394-97”.
If the layer is E1394-95, the version number is “1”.

22/74
5.3.3.2.Patient information Record
[Example of transmission]
 IPU -> Host computer
P|1||||^Thomas^Johnson <CR>
 Host computer -> IPU
P|1|||100|^Thomas^Johnson||20010820|M|||||^Dr.2||||||||||||^^^EAST<CR>

Table 10:Details of Patient Information Record

ASTM IPU -> Host *1 Host -> IPU Max. size


Field Field Name (Byte) Remarks
8.1.1 Record type P P 1 Fixed
8.1.2 Sequence No. Sequence No. Sequence No. 4 Sequence No. of
records
8.1.3 Practice assigned Not used Not used -
Patient ID
8.1.4 Laboratory Not used Not used -
assigned patient ID
8.1.5 Patient ID No. Not used Not used -
8.1.6 Patient name ^First name ^First name ^20^20
^Last name ^Last name
8.1.7 Mother's maiden Not used Not used -
name
8.1.8 Birth date Not used Not used -
8.1.9 Patient sex Not used Not used -
8.1.10 Patient race Not used Not used -
8.1.11 Patient address Not used Not used -
8.1.12 Reserved Not used Not used -
8.1.13 Patient telephone No. Not used Not used -
8.1.14 Attending Not used Not used -
physician name
8.1.15 Special field 1 Not used Not used -
8.1.16 Special field 2 Not used Not used -
8.1.17 Patient height Not used Not used -
8.1.18 Patient weight Not used Not used -
8.1.19 Patient’s known or Not used Not used -
suspected diagnosis
8.1.20 Patient active Not used Not used -
medications
8.1.21 Patient diet Not used Not used -
8.1.22 Practice field 1 Not used Not used -
8.1.23 Practice field 2 Not used Not used -
8.1.24 Admission and Not used Not used -
discharge dates
8.1.25 Admission status Not used Not used -
8.1.26 Location Not used Not used -
8.1.27 DRG or AVG Not used Not used -
8.1.28 DRG or AVG2 Not used Not used -
8.1.29 Patient religion Not used Not used -
8.1.30 Marital status Not used Not used -
8.1.31 Isolation status Not used Not used -
8.1.32 Language Not used Not used -
8.1.33 Hospital service Not used Not used -
8.1.34 Hospital institution Not used Not used -
8.1.35 Dosage category Not used Not used -
*1 To output the QC data, all fields other than 8.1.1 “Record Type” and 8.1.2 “Sequence number” are not
used.

23/74
[Detailed Explanation of the fields]
1) 8.1.2 Sequence No.
The sequence number starts with 1 and indicates sequence position in which the record appeared in the
message. This number is reset to 1 when a higher-level record appears in the message.

2) 8.1.5 Patient ID No.


The patient ID is a unique patient identification and may contain a maximum of 16 digits of
alpha-numeric and a hyphen "-"(2Dh).

3) 8.1.6 Patient name


The first name and the last name may be with a maximum of 20 characters each with consisting of
alpha-numeric.

Patient name is encoded/decoded to the character code shown in Table 11 depending on the IPU settings
of Language.

Table 11:Character Code


Character set IPU “Language “ Setting
ISO 8859-1 English, German, French, Italy, Spanish
Shift-JIS Japanese
GB2312-80 Chinese
windows-1251 Russian

24/74
5.3.3.3.Request Information Record
[Example of transmission]
 IPU-> Host computer
Q|1|1^1^ 1234567890^B||^^^040^PT|0|20010905150000<CR>
 Host computer -> IPU
Not used

Table 12: Details of Request Information Record

ASTM Field Name IPU -> Host Host -> Max. size
Field IPU (Byte) Remarks
12.1 1 Record Type Q Not used 1
12.1.2 Sequence No. 1 Not used 4
12.1.3 Starting Range ID Rack No.^ Not used 6^
No. Tube position^ 2^
Sample ID No.^ 15^
Sample ID No. attribute 1^
12.1.4 Ending Range ID Not used Not used -
No.
12.1.5 Universal test ID ^^^Test code (See the Not used ^^^3
Table 23(Appendix B.1)) ^8
Test name
When requesting
multiple parameters, use
the repeat symbol as
below:
Test parameter
information1
¥Test parameter
information 2¥Test
parameter information
3¥Test parameter
information4¥…
12.1.6 Nature of request 0 Not used 1 0: Request date
day limit Request date is not
defined in the ASTM.
12.1.7 Beginning request YYYYMMDDHHMMSS Not used 14
date and time
12.1.8- Ending request Not used Not used -
12.1 12 result date
through User field
No.2
12.1.13 Requested (Inquiry type)* Not used (1)* *The first analysis/
information status Re-analysis is
code selectable only when
the “Inquire
re-analysis”setting in
the “System Settings”
is ON.

25/74
[Detailed Explanation of the field]
1) 12.1.2 Sequence No.
The sequence number starts with 1 and indicates the sequence position in which the record appeared in
the message. The number is reset to 1 when a higher-level record appears in the message.

2) 12.1.3 Starting Range ID No.


Rack No.: Assigned to the rack. STAT table, reagent table are described as below.
Normal sample: Consists of 6-digit alpha-numeric (The rack number will be
fixed 6-digit number by aligning to the right and padding zero’s to the most
significant digits if the number is less than 6-digit.)
STAT sample: Describes with STATΔΔ (Δ indicates a space character.)
Reagent table: Describes with REAG00
LAS sample: Describes with LXXXXX.. (X indicates a numeric character. The
rack number will be fixed 6-digit number by aligning to the right and padding
zeros to the most significant digits if the number is less than 6 digits.

(*) If the rack barcode is duplicated, together with a reading error, the IPU is to conduct
automatic numbering as follows:
“R” + “Time 2 digits” + “Serial number 2 digits” + “*”

Tube Position : The sample rack position number within a rack, with consisting of 2-digit
number.
Normal sample: 01 - 10
STAT sample: 01- 05
Reagent table: ΔΔ (Δ indicates a space character.)
LAS sample: Fixed to 01.

Sample ID number: Consists of 15-digit of alpha-numeric. Depends on the direction for use, the
hyphen “-“(2DH) can be inserted between characters. “-“is included in
15-digit number. The zero suppression function is not performed. If the
number is less than 15-digit, it is aligned to the right and spaces (020H) are
padded to the most significant digit. If the sample barcode can not be read,
the sample ID number beginning with "□□ERR0000000001" or
"ERR000000000001” is sequentially assigned to the sample number. (The
symbol “□□” indicates a space character.) The sample number beginning with
“QC” is reserved to use for the QC analysis.

Note: In the System Settings screen, Sample No. in Barcode Reading Error is
selectable from 13-digit ("□□ERR0000000001") or 15-digit ("ERR000000000001")
to output. (The symbol “□□” indicates a space character.)

Sample ID No. attribute: M Sample ID No. is manually entered through the touch panel or the IPU
keyboard.
A Analyzer automatically assigned number. This number is assigned by the
automatic-increment function, and is used to set the sample number
that begins with "ERR", which is used when the ID Read error occurred.
B Barcode reader read number. This is used when the sample ID number is
read by the ID barcode reader.

26/74
3) 12.1.5 Universal test ID
Test code: Suffixing 0 to the “Host ID” in the “Basic” in the "Assay Group Setting" makes the
test code. See the Table 23(Appendix B.1).

Test name: The test name is the “Abbreviation” in the “Basic” in the “Assay Group Setting". It is
consisted of up to 8 characters. See the Table 23(Appendix B.1).

Note:
The measurement of platelet aggregation does not correspond to orders from the host
computer. In case reagents of the measurement of platelet aggregation (ADP, collagen,
epinephrine, arachidonic acid, and ristocetin) are set, parameter codes of the
measurement of platelet aggregation are transmitted. Avoid ordering the measurement of
platelet aggregation from the host computer.

4) 12.1.7 Beginning request result date and time


The date format is fixed to "YYYYMMDDHHMMSS".
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system (00-23),
MM the minute (0-59), SS the second (0-59).

5) 12.1.13 Requested information status code


Inquiry type:
N: Real time inquiry of the first analysis
C: Real time inquiry of re-analysis
* Either N or C is selectable only when the“Inquire re-analysis”setting in the “System Settings” is
ON.

Notes:
・Requested information record for the C(Inquiry of re-analysis) is transmitted only when
it satisfies the following conditions ;
・The measurement mode is “Standard”mode
・The “Target” of “Automatic Output” in the System Settings is “All Data”
・The analysis is not QC analysis nor Calib.curve analysis
・The instrument is in measurable mode.
・The analysis is not multi dilution analysis(MDA).
・The analysis result data including the parameters for “Re-analysis” in the assay group
settings has been already output.

・Select 9600bps or more of Baud Rate with“Ethernet connection”or“Serial connection”,


because the communication data volume increases when the “Inquire
re-analysis”setting in the“System Settings” is ON.

Notes:
・Set up auto re-analysis, auto re-dilution and reflex in the assay group settings of
CS-2500 in the same conditions with the re-analysis logic of Host Computer, if the
“Inquire re-analysis” is selected (because aspiration volume increases with the
re-analysis).
・Please be noted that the analysis settings are different from those with no inquiry of
re-analysis in the following conditions ;
Same parameters with the parameters used for the first analysis are selected as the
reflex test parameters.

Re-analysis is not performed when the CS-2500 condition is used (that does not inquire
re-analysis). If the analysis is performed by the order of Host Computer, the error
“Insufficient Sample”might occur.

27/74
5.3.3.4.Test Order Record
[Example of transmission]
 IPU -> Host computer
O|1||000001^01^123456789012345^B^E||R||||||N<CR>
 Host computer -> IPU
O|1|000001^01^123456789012345^B||^^^040^^100.00¥^^^050^^100.00¥・・・
¥^^^060^^100.00|R|20010807101000|||||N<CR>
Table 13: Details of Test Order Record

ASTM IPU-> Host computer -> Max.


Field Field name Host computer IPU size Remarks
(Byte)
9.4.1 Record type O O 1 Fixed
9.4.2 Sequence No. Sequence No. Sequence No. 4 Sequence No. of records
9.4.3 Specimen ID Not used Rack No.^ 6^ Sample ID No.
Tube Position^ 2^ attribute is one of the
Sample ID No.^ 15^ followings:
Sample ID No. 1(^^1) M: Manual entry
attribute(^^Meas * A: Automatically
urement Mode)* assigned by analyzer
9.4.4 Instrument Rack No.^ Not used 6^ B: Barcode reader
specimen ID Tube Position^ 2^
Sample ID No.^ 15^ *Measurement Mode
Sample ID No. 1^ can be set only when
attribute^ 1(^1)* "System
Extended order Settings"-"Specify
request Measurement Mode"
flag(^Measureme setting is ON.
nt Mode)*
9.4.5 Universal Test ID Not used ^^^Test code ^^^3 *Replication can be set
^^Dilution ^^8^3 only when "System
ratio^Option(^Re (^1)* Settings"-"Specify
plication)* (^1)* Replication" setting is
(^Measurement ON.
Time)*
When requesting *Measurement Time
multiple can be set only when
parameters, use "System
the repeat symbol Settings"-"Switch
as below: Measurement Time"
Test parameter setting is ON.
information1
¥Test parameter
information2¥Test
parameter
information
3¥Test parameter
information4¥…
9.4.6 Priority Priority Priority 1 S(STAT, STAT sample)
R(Routine,
Normal sample)
9.4.7 Requested/order Not used YYYYMMDDHH 14
date and time MMSS
9.4.8 - Specimen Not used Not used -
9.4.11 collection date and
time
through
Collector ID
9.4.12 Action code Action code Action code 1 N Normal sample
Q QC material

28/74
ASTM IPU-> Host computer -> Max.
Field Field name Host computer IPU size Remarks
(Byte)
9.4.13 Danger code Not used Not used -
- through
9.4.31 Material
institution

[Detailed explanation of the field]


1) 9.4.2 Sequence No.
The sequence starts with 1 and indicates the sequence position in which the record appeared in the
message. This number is reset to 1 when a higher-level record appears in the message.

2) 9.4.3 Specimen ID
Rack No.: Up to 6-digit alpha-numeric assigned to the rack. Return the same number
that was inquired.

Tube Position: The sample position number with in a rack with consisting of 2-digit number.
Return the same number that was inquired.

Sample ID number: Consists of 15-digit of alpha-numeric and hyphen “-“(2Dh). Return the same
number that was inquired.

Sample ID No. attribute: Indicates how the sample ID No.was registered.


Return the same number that was inquired.

Measurement Mode: Specifies sampling mode.


N: Normal Mode
M: Micro-sampling Mode
Other than above value: Sample is aspirated according to the instrument
setting.

In the case of QC sample, it is not possible to specify measurement mode from


Host Computer. Specified measurement mode from Host Computer is not
reflected to QC sample.

If the measurement mode is changed from Assay Group setting on the


instrument (manual order registration or preset order), the sample is
aspirated according to the specified setting from the instrument prior to Host
Computer specification.

*Measurement Mode can be set only when “System Settings”-"Specify


Measurement Mode” setting is ON.
Note:
When the“Inquire re-analysis”setting in the “System settings”is ON and the test
order record is for the inquiry of re-analysis, it is impossible to specify
Micro-sampling mode. (The mode is not reflected.)

Note:
When rack number which is specified by the Rack No. in Test Order Record are
matched to the rack numbers with enabled “Micro mode (without cap-check)”
setting in “System Settings”-“Operational Settings”-“Rack Assignment”, samples
on the matched rack are analyzed in micro-sampling mode regardless of
Measurement Mode to be set.

29/74
3) 9.4.4 Instrument Specimen ID
Rack No.: Assigned to the rack. In the STAT table and reagent table, Rack No. is
described as below.
Normal sample: 6-digit alpha-numeric (The rack number will be fixed 6-digit
number by aligning to the right and padding zero’s to the most significant
digits if the number is less than 6-digit.)
STAT sample: STATΔΔ (Δ indicates a space character.)
Reagent table: REAG00
Integrated data of Platelet Aggregation: ΔΔΔΔΔΔ
(Δ indicates a space character.)

(*) If the rack barcode is duplicated, together with a reading error, the IPU is to conduct
automatic numbering as follows:
“R” + “Time 2 digits” + “Serial number 2 digits” + “*”

Note:
“Integrated data of Platelet Aggregation” means the data of the aggregation ratio
calculated from measurement result of PPP and PRP.

Tube Position: The sample position number with in a rack with consisting of 2-digit number.
Normal sample: 01 - 10
STAT sample: 01 - 05
Reagent table: ΔΔ (Δ indicates a space character.)
Integrated data of Platelet Aggregation: ΔΔ (Δ indicates a space character.)

Note:
“Integrated data of Platelet Aggregation” means the data of the aggregation ratio
calculated from measurement result of PPP and PRP.

Sample ID number: Consists of 15-digit alpha-numeric. Depends on the direction for use, the
hyphen “-“(2DH) can be inserted between characters. “-“is included in
15-digit number. The zero suppression function is not performed. If the
number is less than 15-digit, it is aligned to the right and spaces (020H) are
padded to the most significant digit. If the sample barcode can not be read,
the sample ID number beginning with "□□ERR0000000001" or
"ERR000000000001” is assigned sequentially to the sample number. (The
symbol “□□” indicates a space character.) The sample number beginning with
“QC” is reserved to use for the QC analysis.

Note: In the System Settings screen, Sample No. in Barcode Reading Error is
selectable from 13-digit ("□□ERR0000000001") or 15-digit
("ERR000000000001") to output. (The symbol “□□” indicates a space character.)

Note: For QC orders, "QCXXXXXXXYYYYYY" is automatically generated as a


sample number.
The selected control name is displayed in XXXXXXX by seven alphameric
characters aligned to the right. If the control name exceeds seven characters,
only the first seven will be displayed. If it is less than seven characters, the
text is aligned to the right and displayed by space padding.
The selected control lot number is displayed in YYYYYY by six alphameric
characters aligned to the right. If the lot number is less than six characters,
the text is aligned to the right and displayed by space padding.

Sample ID No. attribute: M: sample ID No. is manually entered through the touch panel or the IPU
keyboard.

30/74
A: Automatically assigned number by the analyzer. This number is
assigned by the automatic-increment function, and is used to set the
sample number that begins with “ERR”, which is used when the ID Read
error occurred.
B: Barcode reader read number. This is used when the sample ID number
is read by the ID barcode reader.

Extended order request flag:


E: Extended order
Indicates the analysis is continued, based on the rules of the analyzer. If the
analysis is not continued, it is skipped.
Note:
The flag is not output if the“Inquire re-analysis”setting in the“System Settings”is
ON.

Measurement Mode: Indicates sampling mode.


N: Normal Mode
M: Micro-sampling Mode

*Measurement Mode can be set only when "System Settings"-"Specify


Measurement Mode” setting is ON.

4) 9.4.5 Universal Test ID


When an order is sent from the host computer to the IPU, set the Universal Test ID which is sent in the
“Universal Test ID” in the section “4.3.3.3 Details the Request Information Record”. If there are no
parameters, it can be skipped.

Note:
The measurement of platelet aggregation does not correspond to orders from the host
computer. The IPU does not measure the platelet aggregation orders even though
parameter codes of platelet aggregation are set. For this reason, maximum aggregation
rate and evaluation parameters are not calculated.

Caution:
When IPU sends an order inquiry to the host computer and then order information is not
registered in the Test Order Record from the host computer, an error dialog box can be
displayed indicating “There is no order in Host Computer (error code:21009)” by enabling
the setting of “A warning message is output in case there is no parameter to be analyzed.”
in “System Settings”-“Operational Settings”-“Order Inquiry” and inserting “000” in the
parameter code in Test Order Record from the host computer.

Test code: Suffixing 0 to the “Host ID” in the “Basic” in the “Assay Group Setup” makes
the test code. See the Table 23(Appendix B.1).

Dilution ratio: Specifies the dilution ratio used in analysis with percent. If it is not specified, follows the
instrument setting. The dilution ratio can be specified with including 2-digit of decimal
number.
The dilution ratio in the CS-2500 which differs from the ordered dilution ratio by less
than 0.01 percent is used for analysis.
Note: If the MDA is set in the “Default Dilution Ratio” in the “Basic” in the “Assay Group
Setup”, the MDA analysis is performed in disregard of the dilution ratio from the host
computer.
* If the unregistered dilution ratio in the CS-2500 is ordered, the “Default Dilution
Ratio” in the “Basic” in the “Assay Group Setup” is used for analysis.

Warning:
The setting of the dilution ratio may have a critical impact on analysis results.
Set the standard setting of the dilution ratio to 100 percent. If other than 100
percent is set to the dilution ratio, make sure that it is the correct value.

31/74
Note:
The redilution, re-analysis and reflex test are performed according to the
instrument setting even if the analysis is specified at the dilution ratio different
from the instrument setting. If the redilution, re-analysis and reflex test are not
performed, specify the option from the host computer.

Option: Sets the performance of the redilution, re-analysis and reflex test. More than 1 option can
be set with the combination of “D”, “R” and “F”. If there are no settings, follows the
instrument settings
D: No redilution
R: No re-analysis
F: No reflex test

Replication: Sets the number of times the test is performed. “1” or “2” can be set.
Other than above value: Sample is measured according to instrument setting.

In the case of QC sample, it is not possible to specify the replication from Host Computer.
Specified replication from Host Computer is not reflected to QC sample.

*Replication can be set only when “System Settings”-"Specify Replication" setting is ON.
Note:
When the“Inquire re-analysis”setting in the “System settings”is ON and the test
order record is for inquiry of re-analysis, it is impossible to specify replication.
(Measurement is performed according to the replication set at the assay group
settings of CS-2500.)

Measurement Time: Set measurement time. If there are no settings, follow the instrument settings.
M:Measurement Time(Main)
S:Measurement Time(Sub)

* This is selectable only when the“Switch Measurement Time”in the“System Settings”


is ON. Default setting is OFF.

5) 9.4.6 Priority
S:STAT, STAT sample
R:Routine, Normal sample

6) 9.4.7 Requested/order date and time


Indicates the date and time of the analysis for the inquiry sample.
The date format is fixed to "YYYYMMDDHHMMSS".
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system (00-23),
MM the minute (0-59), SS the second (0-59).

7) 9.4.12 Action code


Indicates the contents of the result record to be sent.
N Normal sample
Q QC sample

32/74
5.3.3.5.Result record
[Example of transmission]
 IPU -> Host computer
R|1|^^^041^PT sec^100.00^9^^^|10.2|sec||N||||||20110328135056<CR>
R|2|^^^042^PT %^100.00^9^^^| 99.4|%||N||||||20110328135056<<CR>
……
 Host computer -> IPU
Not used
Table 14: Details of the Result record

ASTM IPU -> Host Host Max. size


Field Field name computer computer (Byte) Remarks
-> IPU
10.1.1 Record type R Not used 1
10.1.2 Sequence No. Sequence No. Not used 4 Sequence No.
beginning with 1
10.1.3 Universal Test ID ^^^Test code^ Not used ^^^3^ Test code: Described
Parameter 24 in Test code (Table
^Dilution ratio^ ^6^ 24 (Appendix B.2)):
Analysis result 1 Parameter: Up to 24
type^Extended ^1^ characters
order request^ 1^
Extended order 1(^2)* *Replication can be
result^ output only when
Reflex test "System
request(^Repli Settings"-"Specify
cation)* Replication" setting
is ON.
10.1.4 Data or measurement Value Not used 6 Up to 6 characters
value (In case of the path
name to the image
data, up to 80
characters)
10.1.5 Units Units Not used 7
10.1.6 Reference ranges Not used Not used -
10.1.7 Result abnormal flags Result Not used 1^Not Evaluation Info. and
abnormal flags specified^ Instrument error
^Evalution Not information are
Information^I specified output only when the
nstrument format is ASTM
Error E1381-02 and also
Information when it is set to
output error
information.
10.1.8 Nature of abnormality Not used Not used -
testing
10.1.9 Result status (Result Not used (1)* *Result Status can be
Status)* output only when
“Output each Assay
Group result” setting
is ON.
10.1.10 Date of change in Not used Not used -
instrument normative
values
10.1.11 Operator identification Not used Not used -
10.1.12 Date/Time test started Not used Not used -
10.1.13 Date/Time test completed YYYYMMDDH Not used 14
HMMSS
10.1.14 Instrument identification Not used Not used -

33/74
[Detailed explanation of the field]
1) 10.1.2 Sequence No.
The sequence number starts with 1 and indicates the sequence position in which the record appeared in
the message. This number is reset to 1 when a higher-level record appears in the message.

2) 10.1.3 “Universal test ID” through 10.1.13 “Date time test completed”
Fields to be set will vary depending on the contents to be transmitted.
Each output specification is described in the following pattern.
2.1) When test parameters are output
2.2) When evaluation parameters of platelet aggregation are output
2.3) When Sample Info. flags are output
2.4) When the path name to the image data is output

2.1) When test parameters are output: The parameters for which analysis orders are registered are output.

a) 10.1.3 Universal Test ID


Test code: Indicates the assay parameter.
Consists of the combination of “Host ID” in the “Basic” in the “Assay Group
Setup” and the “Host ID” in the “Assay Parameter" in the “Assay Group Setup”.
See the Table 24 (Appendix B.2).

Parameter: Consisted of up to 24 characters. The characters in the “Abbreviation” in the


“Assay Group Setup” are output. See the Table 24 (Appendix B.2).

Dilution ratio: Outputs the dilution ratio used in analysis with percent. It is indicated with
including 2-digit of decimal number.
*If the MDA is set in the “Default Dilution Ratio” of the “Basic” in the “Assay Group
Setup”, the sub- field of the dilution ratio in the final report is skipped.

Analysis result type:


1: Normal (Auto output)
2: Average (Auto output)
3: Re-analysis (Auto output)
4: Average of re-analysis (Auto output)
5: Normal (Manual Output)
6: Average (Manual Output)
7: Re-analysis (Manual Output)
8: Average of re-analysis (Manual Output)
9: Final information (Auto output)
A: Final information (Manual Output)

Extended order request: Indicates the redilution analysis and re-analysis are performed by the analyzer,
based on the rules of the analyzer. Other than that, the extended order request
is not added.
D: Redilution request
R: Re-analysis request

Extended order results: Indicates the analysis result is one of the followings: the result of the redilution
analysis, reanalysis or reflex test. Other than that, the result is not output.
D: Results of redilution analysis
R: Results of re-analysis
F: Results of reflex test

34/74
Reflex test request: Indicates the reflex test is performed by the analyzer, based on the rules of the
analyzer. Other than that, the extended order request is not added.
F: Reflex test request
Note:
The extended order and reflex test request are not output when the“Inquire
re-analysis” setting in the “System Settings”is ON.

Replication: Indicates the number of times the test is performed.


*Replication can be set only when "System Settings"-"Specify Replication"
setting is ON. Default setting is OFF.

b) 10.1.4 Data or measurement value


Up to 6-digit number. Data format is described in the Table 24 (Appendix B.2).
If the analysis results can not be obtained, the mask characters are input in part of the integral number.
The mask characters are as follows.

*: Analysis failure
/: Average calculation failure
+: Display Digit Overflow
-: Calculation failure
X: No validated calibration curve in the calculation

c) 10.1.5 Units
Up to 7 characters
In the CS-2500, the units set in the IPU are output. (The ASTM describes the abbreviation of ISO
standard should be used.) See the Table 24 (Appendix B.2). If the unit is “Ratio” or “INR”, they are not
output.

d) 10.1.7 Result abnormal flag


L: Below the lower patient limit
H: Above the upper patient limit
<: Below lower report limit.
>: Above upper report limit
N: Normal test result
A: Analysis error

* Appendix.C - Table 25: Analysis error, Result Abnormal Flags, and Display method of analysis results

Evaluation Information
Outputs a string of characters consisting of the combination of the evaluation error code and error
message, which is enclosed between square brackets “[]“. If there is more than one evaluation
information, they are separated by commas.
If there is no evaluation information, it is not output.

Example of output) |A^[0008.0001.0000 Initial fluctuation drop],[0008.0002.0000 Coagulation Curve


Error : Sharp Drop]^|

* Instrument Error Information is output only if the setting of “Output Error Information” in the
system settings is turned ON when the format is ASTM E1381-02.

*In some analysis errors, it varies if the data is output as a numeric value or masked data, depending
on the setting of “Output Error Information” in the system settings. See the Table 25.

35/74
Instrument Error Information
Outputs a string of characters consisting of the error code and error message, which is enclosed between
square brackets “[]”.
If there is no Instrument error information, it is not output.

Example of output) |A^^[34422 Insufficient Reagent (Reagent Arm Liquid Surface Not Detected)]|

* Instrument Error Information is output only if the setting of “Output Error Information” in the
system settings is turned ON when the format is ASTM E1381-02. See the Table 25 (Appendix C).

Note :
Evaluation Information and Instrument Error Information are encoded to Character Code
according to the setting of displaying language on IPU for output to Host Computer. The relation
of Character Code and display language setting is shown in Table 11.

e) 10.1.9 Result Status


Indicates the status of analysis result.
S:
Indicates the result is the part of the analysis results for the sample.
The analysis result is output when the analysis for one Assay Group is end.
F:
Indicates the result is all analysis results for the sample.
The analysis result is output when the all analysis for the sample is end.

(Caution) Validate the output result of each Assay Group with analysis result of other Assay Group on the
Host Computer.

*Result Status can be output only when “Output each Assay Group result” setting is ON.

f) 10.1.13 Date /Time test completed


Indicates the date and time the test was completed.
The date format is fixed with "YYYYMMDDHHMMSS".
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system (00-23),
MM the minute (0-59), SS the second (0-59).

This date and time indicates when the all tests of the sample are completed.
In the case of the result which is one of the sample(the analysis result with 10.1.9 Result Status is “S”) ,
this field is blank.

2.2) When evaluation parameters of platelet aggregation are output: Evaluation parameters are output when
platelet aggregation orders are registered. In order to output evaluation parameters of platelet
aggregation, the setting of “Output Evaluation Info on Platelet Aggregation.” in “ [System Settings] -
[Host Computer] - [Detailed setting] ” on IPU is required to be turned ON.

Caution:
Evaluation parameters of platelet aggregation is used for inspection in a clinical
laboratory only and not for examination of a patient. It is to let the operators know
a potential specific abnormality in evaluation parameters confirmed by the
analysis data.

The measurement of platelet aggregation can output multiple evaluation parameters in one measurement. The
followings show output conditions for evaluation parameters of platelet aggregation.

Table 15: Evaluation parameter name and output conditions in measurement of platelet aggregation

36/74
Evaluation Description Output condition
parameter name
~Abs PPP PPP absorbance To be output
~Abs PRP_s PRP absorbance before To be output
reaction starts
~Max Maximum aggregation rate To be output
~Fin Final aggregation rate To be output
~Lag Phase Lag Phase Time To be output
~AUC Area under aggregation rate To be output
curve
~VMax First maximum reaction speed To be output
rate
~VMax Time First maximum reaction time To be output
~VMax Angle First maximum reaction angle To be output
~APAL Platelet aggregation level for Output is carried out when an order of
ADP APAL-1%, APAL -10%, CPAL-1%, and
CPAL-5% is registered.
~CPAL Platelet aggregation level for Output is carried out when an order
COL either of APAL-1%, APAL-10%, CPAL-1%,
and CPAL-5% is registered.

Note:
Additional evaluation parameters may be added in the future. In case evaluation parameters which
are not defined in Table 15 are received, you need to use a program which can skip the data with
undefined evaluation parameters.

a )10.1.3 Universal Test ID


In the measurement of platelet aggregation, a “Test code” is assigned to each measurement.
Evaluation parameters of platelet aggregation is output linking to the “Test code” of each
measurement. The host computer can specify the output data by the combination of “Test code” and
“Parameter”.

Test code: Test codes for the measurement of platelet aggregation. As for test codes,
refer to the Table 24 (Appendix.B.2.). A test code is not set when the
“Evaluation parameter name” is either the “~APAL” and the “~CPAL”.
Parameter: “Evaluation parameter name” in the measurement of platelet aggregation
is output. As for “Evaluation parameter name”, refer to “Table15:
Evaluation parameters and output conditions in measurement of platelet
aggregation”.
Dilution ratio: Not used
Analysis result type: The type of analysis results is output. Either type listed below is output.
9 :Final report(Auto output)
A :Final report(Manual output)
Extended order request: Not used
Extended order results: Not used
Reflex test request: Not used

b)10.1.4 Data value


Data to an analysis parameter ID is output. The number of characters is fixed at 6 characters. The
followings describe formats of each evaluation parameter.

Table 16: Format definition in evaluation parameters of platelet aggregation


Evaluation parameter name Format
~Abs PPP ̺#####
~Abs PRP_s ̺#####
~Max ####.#
~Fin ####.#
~Lag Phase ####.#
37/74
~AUC ######
~VMax ̺ #####
~VMax Time ̺ ̺####
~VMax Angle ###.##
~APAL ̺ ̺##.#
~CPAL ̺ ̺##.#
(̺: Half-width space)

In case evaluation parameter is not calculated, the masked character “-” is added to the
numeric part(#) except decimal point.

c)10.1.5 Units
Unit to an analysis parameter ID is output. The followings describe units of each evaluation
parameter.

Table 17: Definition of unit in evaluation parameters of platelet aggregation


Evaluation parameter name Unit
~Abs PPP mOD
~Abs PRP_s mOD
~Max %
~Fin %
~Lag Phase sec
~AUC %*sec
~VMax %/sec
~VMax Time sec
~VMax Angle deg
~APAL None
~CPAL None

d)10.1.7 Result abnormal flags


Result abnormal flags : Not used
Evaluation Information : Not used
Instrument Error Information : Not used

e)10.1.9 Result status


Not used.

f) 10.1.13 Date/Time test completed


The date and time the test was completed is output.
The format is fixed to “YYYYMMDDHHMMSS”.
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system
(00-23), MM the minute (0-59), SS the second (0-59).
*Calculated date and time of data value is indicated when “analysis parameter” for analysis
parameter ID is the “~APAL” and the “~CPAL”.

2.3) When Sample Info. flags are output: The parameters to which HIL flags or Sample Volume flag are
attached are output.

Sample info flags are out put with defective sample volume flag “N” even if the sample volume is within
the specified range, when the setting of “Output Sample Volume flags only in case of defective sample volume.” is
turned OFF in [System Settings].

(Caution)
Sample Info. flags are used only for inspection in a clinical laboratory and not for examination of a patient.
They notify an operator of the possibility of the specific abnormal sample confirmed by checking analysis
data.

38/74
a) 10.1.3 Universal Test ID
Test Code: Not used.
Parameter: Indicates the content of HIL flags, or sample volume flags
Hemolytic Sample: Sample suspected of having hemolysis.
Icterus Sample: Sample suspected of having icterus.
Lipemic Sample: Turbid sample affected by lipemia or other materials.
Defective Sample Volume: Sample suspected of defective sample volume.*
Dilution ratio: Not used.
Analysis result type: Not used.
Extended order request: Not used.
Extended order results: Not used.
Reflex test request: Not used.

b) 10.1.4 Data or measurement value


HIL level values are output if the [Output HIL Level] settings - [Output HIL Flag] in the [System
Settings] is ON.
Check Level 5 : “5”
Check Level 4 : “4”
Check Level 3 : “3”
Check Level 2 : “2”
Check Level 1 : “1”
Check Level 0 : “0”
In case HIL detection value is False value(-1) : “*”

c) 10.1.5 Units
Not used.

d) 10.1.7 Result abnormal flag


A: Result with HIL flags or sample volume flags.
W: Above the analysis range of the HIL detector (Not used for sample volume flags)
N: Sample Volume check fell within the preset range (Not used for HIL flags)

e) 10.1.9 Result Status


Indicates the status of analysis result.
S:
Indicates the result is the part of the analysis results for the sample.
The analysis result is output when the analysis for one Assay Group is end.
F:
Indicates the result is all analysis results for the sample.
The analysis result is output when the all analysis for the sample is end.

*Result Status can be output only when “Output each Assay Group result” setting is ON.

f) 10.1.13 Date /Time test completed


Indicates the date and time the test was completed.
The date format is fixed to "YYYYMMDDHHMMSS".
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour system (00-23),
MM the minute (0-59), SS the second (0-59).

39/74
2.4) The path name to the image data is output only when the parameters have the image data.

(Caution)
The image data is used for a testing at a clinical laboratory only and not for examination of a patient. It is to
let the operators know a potential specific abnormality in a sample confirmed by the analysis data.

a)10.1.3 Universal Test ID


Test Code :Test codes of image data to be output is set. In case the type of image
data is superimposed aggregation curve image, no code is set.
Parameter :Type of the image data is set.
Normal : Normal Clot Waveform
Average : Average Clot Waveform
MDA :MDA image data
Mixing :Cross mixing image data
Delayed Mixing :Delayed Cross mixing image data
Platelet :Aggregation curve of each analysis parameter in the
measurement of platelet aggregation.
Superimpose:A graph overlapped all aggregation curves in the measurement of
platelet aggregation.
Dilution ratio: Not used
Type of Analysis Results:Not used
Extended order request:Not used
Extended order result:Not used
Reflex test request:Not used

b)10.1.4 Data or measurement values


The file path name to the image data is output. The character “¥” used for the file path is
converted to the escape sequence “&R&”. The actual image data is written to
the“C:¥shared¥PNG¥Date folder”. The file path name is output after the“PNG”to the data
value. The date folder is fixed as“YYYYMMDD”. The file name is attached depending on the
date and time, sample No., test code, image data, dilution ratio_test sequence No. and the
type of extended order results.

40/74
Table 18: List of the information included in the file

No. Item Contents Remarks


1 Test Date Sets the analyzed date (and time) in the form of Date/Time test started
“YYYY_MM_DD_HH_MM_SS” When the type of image is
"Superimpose", the created date
and time is transmitted.
2 Sample No. Sets the Sample No.
3 Test Code Sets the test code set in the request record.
4 Type of Sets the type of image data. Width and height of image data
image data Normal :Normal Clot Waveform describe as below.
Average:Average Clot Waveform Normal:500 x 500 pixel
MDA :MDA image data Average:500 x 500 pixel
Mixing :Cross mixing image data MDA:500 x 500 pixel
Delayed Mixing:Delayed Cross mixing image Mixing:500 x 500 pixel
data Delayed Mixing:500 x 500 pixel
Platelet :Aggregation curve image Platelet:625 x 388 pixel
Superimpose : Overlapped aggregation curve Superimpose:625 x 388 pixel
image
5 Dilution Sets dilution ratio. This is skipped when the types of
Ratio image data are other than Normal
and Average.
6 Test Sets the sequence No. during analysis. This is skipped when the types of
Sequence image data are other than
No. Normal.
7 Extended Sets the information to identify whether the This is skipped when the types of
order result is re-analysis result or not. (Optional) image data are other than
D :Result of re-dilution analysis Normal.
R :Result of re-analysis
F :Result of reflex test

"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_100_1.Png"

In this case, the image data is saved in the “C:¥shared¥PNG¥20030930”folder with the file
name of“2003_09_30_12_00_33_1234567890_040_Normal_100_1.Png”

If the folders saving the image data have passed 3 days including the day the IPU starts
up, they are deleted when the IPU is started.

ex1) Twice analyses


"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_100_1.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_100_2.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Average.Png"
ex2) MDA analysis
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_050_1.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_025_1.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_013_1.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_MDA.Png"
ex3) Re-analysis
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_100_1.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_100_1_R.Png"
ex4) Twice analyses + Re-dilution analysis
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_100_1.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_100_2.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Average.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_050_1_D.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Normal_050_2_D.Png"
"PNG&R&20130930&R&2013_09_30_12_00_33_1234567890_040_Average_D.Png"

41/74
ex5) The measurement of platelet aggregation - aggregation curve image for ADP% and
ADP2%
"PNG&R&20171201&R&2017_12_01_13_00_15_1234567890_671_Platelet.Png"
"PNG&R&20171201&R&2017_12_01_13_05_56_1234567890_672_Platelet.Png"
ex6) The measurement of platelet aggregation - superimposed aggregation curve
"PNG&R&20171201&R&2017_12_01_13_06_20_1234567890__Superimpose.Png"

c)10.1.5 Units
Not used

d)10.1.7 Result abnormal flag


Not used

e) 10.1.13 Date/Time test completed


The date format is fixed to "YYYYMMDDHHMMSS".
Here, YYYY indicates the year, MM the month, DD the day, HH the hour in the 24-hour
system (00-23), MM the minute (0-59), SS the second (0-59).
*Created date and time of image data is indicated in case the type of the image data is
“Superimpose”

(Caution)
Output of Delayed Cross mixing image data

The delayed cross mixing image can be output to host computer manually. In case, the IPU sends Cross
mixing analysis information in CSV file format as well as PNG file. The CSV file is written to
the“C:¥shared¥CSV¥Date folder”.

10.1.4 Data or measurement values


The file path name to the image data is output.
The file path name is output after the“CSV”to the data value.
The file name conforms Table 18.

The contents of CSV file;


Item Contents Remarks
Sample No Sample No.
Date/Time The analyzed date and time. YYYY/MM/DD hh:mm:ss
Assay Group The test code set in the
request record. (see in Table
24 of App.B.2)
Sample Comment Sample No. MAX 30 characters.
Mixing Ratio Mixing Ratio.
Immediate Type Result data of Immediate type Result data are specified with
in each mixing ratio. including 1-digit of decimal
number.
Delayed Type Result data of Delayed type in Result data are specified with
each mixing ratio. including 1-digit of decimal
number.
Reference Info. Title of Reference Info.
Value Value of Reference Info. Values of Reference Info. are
specified with including 1-digit of
decimal number.

The format of CSV file;


Object Format
Character Code UTF-8
Control code <CR><LF>
Header Line Output
Delimiter Comma
Data output Double quote

42/74
5.3.3.6.Comment Record
Comment record can be transmitted only when “System Settings”-“Output calibration curve and
reagent lot information” setting is ON. Default setting is OFF.

[Example of transmission]
 IPU -> Host computer
C|1|I|CAL^042^^1^|I<CR>
C|1|I|LOT^040^OVB^527701¥PT THS^505401|I<CR>
 Host computer -> IPU
Not used

Table 19: Details of Comment Record

ASTM IPU -> Host Host Max. size


Field Field name computer computer (Byte) Remarks
-> IPU
11.1.1 Record type C Not used 1
11.1.2 Sequence No. Sequence No. Not used 4 Sequence No.
beginning with 1
11.1.3 Annotation source I Not used 1 I:Clinical
equipment system
11.1.4 Annotation text Annotation text Not used See the Table See the Table 20
20 and Table and Table 21.
21.
11.1.5 Annotation format I Not used 1 I:Clinical
equipment system

[Detailed Explanation of the field]


1) 11.1.2 Sequence No.
The sequence number starts with 1 and indicates sequence position in which the record appeared in the
message. This number is reset to 1 when a higher-level record appears in the message.

2) 11.1.4 Annotation text


Parameter to be set in annotation text will vary depending on the content to be transmitted. Each
output specification is described in the following patterns.
2.1) When calibration curve information is transmitted
2.2) When reagent lot information is transmitted

43/74
2.1) When calibration curve information is transmitted

Table 20: Annotation text of Comment Record (Calibration curve information)

ASTM Host Max. size


Field Field name IPU -> Host computer computer (Byte) Remarks
-> IPU
11.1.4 Annotation text Information class ^ Not used 3^
Test code ^ 3^
Analyzed date for calib. 8^
curve ^
Calib. curve ID^ 7^
Calibrator Lot No. 12

a) 11.1.4 Annotation text


Information class: It is the Information class of the text field. Standard curve Information is
set to CAL.

Test code: Indicates the assay parameter.


Consists of the combination of “Host ID” in the “Basic” in the “Assay
Group Setup” and the “Host ID” in the “Assay Parameter" in the “Assay
Group Setup”. See Table 24 (Appendix B.2).

Analyzed date for calib. curve:Not used.

Calib. curve ID: Validation ID for calibration curve. When the data size of the calibration
curve ID is not less than the maximum size, the calibration curve ID is
transmitted without space padding.

Calibrator Lot No.: Not used.

44/74
2.2) When reagent lot information is transmitted
Table 21: Annotation text of Comment Record (Reagent lot information)

ASTM Host Max. size


Field Field name IPU -> Host computer computer (Byte) Remarks
-> IPU
11.1.4 Annotation text Information class ^ Not used 3^ The output order of
Test code ^ 3^ reagent name and
Reagent name^ 8^ reagent lot number is
Reagent Lot No. 6 according to the
order specified in the
When multiple reagents reagent list of test
are used in an analysis protocol setting.
parameter, a reagent
name and a reagent lot
number are displayed as
a pair. Each pair is
separated with repeat
delimiter “¥”. The
following is an example
of output.

Reagent name
1^Reagnet Lot
No.1¥Reagent
name2^Reagent Lot
No.2¥Reagent
name3^Reagent Lot
No.3…

a) 11.1.4 Annotation text


Information class: It is the Information class of the text field. Reagent lot information is set
to LOT.

Test code: Code that is suffixed to the “Host ID (two digits)” with 0 in the “Basic” in
the “Assay Group Setup”. See Table 23 (Appendix B.1).

Reagent name: Reagent name for buffer, factor deficient plasma, and normal regent that
are used for analysis. When the reagent used for analysis is not managed
by lot number, the reagent name is empty. When the data size of the
reagent name is less than the maximum size, the reagent name is
transmitted without space padding.

Reagent Lot No.: Reagent lot number used in analysis. When the reagent used for analysis
is not managed by lot number, the reagent lot number is empty. When
the data size of the reagent lot number is less than the maximum size,
the reagent lot number is transmitted without space padding.

45/74
5.3.3.7.Message termination record
[Example of transmission]
 IPU -> Host computer
L|1|N<CR>
 Host computer -> IPU
L|1|N<CR>
Table 22: Message termination record

ASTM IPU -> Host Host computer Max. size


Field Field name computer -> IPU (Byte) Remarks
13.1.1 Record type L L 1 Fixed
13.1.2 Sequence No. 1 1 4 Always 1
13.1.3 Termination N N 1 N: Normal
cord termination

46/74
6. Examples of Communication

Examples of Communication are given for the cases when the serial connection (ASTM E1381-02 Mode) is
carried out
6.1. Inquiry of Analysis Order (IPU => Host Computer)
・When the order inquiry is performed from IPU

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97
<CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2Q|1|000001^01^
1^B||^^^040^PT¥^^^060^Fbg¥^^^120^II¥^^^150^V¥^^^170^VII¥^^^180^VIII¥^^^
190^IX¥^^^200^X¥^^^210^XI¥^^^300^AT3¥^^^310^APL|0|20110328133318<ETX><CHK1><CH
K2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

6.2. Inquiry of Re-analysis Order (IPU=> Host)


* When the“Inquire Re-anaysis”setting in the “System Setting”is ON

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97
<CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2Q|1|000001^01^
1^B||^^^040^PT¥^^^060^Fbg¥^^^120^II¥^^^150^V¥^^^170^VII¥^^^180^VIII¥^^^
190^IX¥^^^200^X¥^^^210^XI¥^^^300^AT3¥^^^310^APL|0|20110328133318||||||C<ETX><CHK1><CHK
2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

6.3. Analysis Order (Host Computer => IPU)


・When the host computer sends orders to IPU

Host <ENQ>
IPU <ACK>
Host <STX>1H|¥^&|||||||||||E1394-97|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>2P|1|||100|^Thomas^Johnson||20010820|M|||||^Dr.2||||||||||||^^^EAST||||||
|||<CR><ETX><CHK1><CHK2>
<CR><LF>
IPU <ACK>
Host <STX>3O|1|000001^01^
1^B||^^^040^^100.00^DF¥^^^050^^100.00¥^^^060^^100.00¥^^^040^^100.00^DR|R|201103301
23159|||||N|||||||||||||||||||<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>4L|1|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <EOT>

47/74
6.3.1. Specify Replication
*When "System Settings"-"Specify Replication" setting is ON.
Host <ENQ>
IPU <ACK>
Host <STX>1H|¥^&|||||||||||E1394-97|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>2P|1|||100|^Heisei^Jiro||20010820|M|||||^Dr.2||||||||||||^^^EAST|||||||||<
CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>3O|1|000001^01^
1^B||^^^040^^100.00^DF^2¥^^^050^^100.00^^1¥^^^060^^100.00^^1¥^^^040^^100.00^DR^2|
R|20110330123159|||||N|||||||||||||||||||<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>4L|1|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <EOT>

6.3.2. Specify Measurement Mode


*When "System Settings"-"Specify Measurement Mode” setting is ON.
Host <ENQ>
IPU <ACK>
Host <STX>1H|¥^&|||||||||||E1394-97|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>2P|1|||100|^Heisei^Jiro||20010820|M|||||^Dr.2||||||||||||^^^EAST|||||||||<
CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>3O|1|000001^01^
1^B^^M||^^^040^^100.00^DF¥^^^050^^100.00¥^^^060^^100.00¥^^^040^^100.00^DR|R|2011
0330123159|||||N|||||||||||||||||||<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>4L|1|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <EOT>

6.3.3. Specify Measurement Time


*3 When the“Switch Measurement Time”setting in the“System Settings”is ON.
Host <ENQ>
IPU <ACK>
Host <STX>1H|¥^&|||||||||||E1394-97|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>2P|1|||100|^Heisei^Jiro||20010820|M|||||^Dr.2||||||||||||^^^EAST|||||||||<
CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>3O|1|000001^01^
1^B||^^^040^^100.00^DF^^S¥^^^050^^100.00^^^M¥^^^060^^100.00^^^S¥^^^040^^100.00^DR^^S
|R|20110330123159|||||N|||||||||||||||||||<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <STX>4L|1|<CR><ETX><CHK1><CHK2><CR><LF>
IPU <ACK>
Host <EOT>

6.4. Analysis Results (IPU => Host Computer)

6.4.1. When analysis is performed 1 time


IPU <ENQ>
Host <ACK>

48/74
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Thomas^Johnson <ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^|
10.2|sec||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^^|
99.4|%||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^^|
0.57|||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^^|
0.81|||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^9^^^|
27.4|sec||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^9^^^|
8.5|sec||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^9^^^|
588.2|mg/dL||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|8|^^^^Hemolytic
Sample^^^^^||||A||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|9|^^^^Defective Sample
Volume^^^^^||||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

* When the“Inquire re-anaysis”setting in the“System Settings”is ON


<6.4.2 “Transmitting Analysis Results (The first analysis)”is also the same.>
IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^^^|
10.2|sec||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^^^|
99.4|%||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^^^|
0.57|||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^^^|
0.81|||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

49/74
IPU <STX>0R|5|^^^051^APTT sec^100.00^1^^^|
27.4|sec||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^1^^^|
8.5|sec||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^1^^^|
588.2|mg/dL||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|8|^^^^Hemolytic
Sample^^^^^||||A||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|9|^^^^Defective Sample
Volume^^^^^||||N||||||20110328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

6.4.2. Transmitting Analysis Results (When there is the re-analysis)


・ Transmitting Analysis Results (The first analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Thomas^Johnson <ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^R^^|
10.2|sec||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^R^^|
99.4|%||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^R^^|
0.57|||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^R^^|
0.81|||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^1^^^|
27.4|sec||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^1^^^|
8.5|sec||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^1^^^|
588.2|mg/dL||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|8|^^^^Icterus
Sample^^^^^||||W||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|9|^^^^Lipemic
Sample^^^^^||||A||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|10|^^^^Defective Sample
Volume^^^^^||||A||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
50/74
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ Transmitting Analysis Results (The re-analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^E||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^3^^R^|
10.2|sec||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^3^^R^|
99.4|%||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^3^^R^|
0.57|||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^3^^R^|
0.81|||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^^Icterus
Sample^^^^^||||W||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^^Lipemic
Sample^^^^^||||A||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^^Defective Sample
Volume^^^^^||||A||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

・ Transmitting Analysis Results (The final analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Thomas^Johnson <ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^R|
10.2|sec||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^R|
99.4|%||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^R|
0.57|||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^R|

51/74
0.81|||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^9^^^|
27.4|sec||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^9^^^|
8.5|sec||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^9^^^|
588.2|mg/dL||N||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|8|^^^^Icterus
Sample^^^^^||||W||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|9|^^^^Lipemic
Sample^^^^^||||A||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|10|^^^^Defective Sample
Volume^^^^^||||A||||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>
6.4.3. When analysis is performed 2 times
Example: Analysis result (The first analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Thomas^Johnson <ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^1^^^|
10.2|sec||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^1^^^|
99.4|%||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^1^^^|
0.57|||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^1^^^|
0.81|||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^1^^^|
27.4|sec||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^1^^^|
7.9|sec||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^1^^^|
632.9|mg/dL||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

52/74
Example: Analysis result (The second analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Thomas^Johnson <ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^1^^^|
7.9|sec||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^062^Fbg C.^100.00^1^^^|
632.9|mg/dL||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

Example: Analysis result (The final analysis)

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CH
K1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Thomas^Johnson <ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^|
10.2|sec||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^^^|
99.4|%||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^^^|
0.57|||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^^^|
0.81|||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^9^^^|
27.4|sec||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^9^^^|
7.9|sec||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^9^^^|
632.9|mg/dL||N||||||20110328150948<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

53/74
6.4.4. Output analysis result for each Assay Group
*When “Output each Assay Group result” setting is ON.

PT
IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^R^^|
10.2|sec||N||S||||<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^R^^|
99.4|%||N||S||||<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^R^^|
0.57|||N||S||||<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^R^^|
0.81|||N||S||||<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

APTT
IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^051^APTT sec^100.00^9^^^|
27.4|sec||N||S||||<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

Fbg
IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^061^Fbg sec^100.00^9^^^|
8.5|sec||N||S||||<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

54/74
IPU <STX>5R|2|^^^062^Fbg C.^100.00^9^^^|
588.2|mg/dL||N||S||||<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

All Assay Groups


IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-68^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^R^^|
10.2|sec||N||F||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT %^100.00^9^R^^|
99.4|%||N||F||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^043^PT R.^100.00^9^R^^|
0.57|||N||F||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^044^PT INR^100.00^9^R^^|
0.81|||N||F||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^051^APTT sec^100.00^9^^^|
27.4|sec||N||F||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^061^Fbg sec^100.00^9^^^|
8.5|sec||N||F||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^062^Fbg C.^100.00^9^^^|
588.2|mg/dL||N||F||||20110328135407<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

6.4.5. Output the path name to the image data


IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^00-17^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT sec^100.00^9^^^|
10.2|sec||N||||||20150328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|1|^^^041^Normal|PNG&R&20150328&R&2015_03_28_13_50_56_1_041_Normal_100_
1.Png|||||||||20150328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
55/74
IPU <EOT>

6.4.6. Output the path name to the Delayed cross mixing image data
IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^00-17^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><
CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^051^APTT sec^100.00^A^^^|
10.2|sec||N||||||20150328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|1|^^^050^DelayedMixing|PNG&R&20150328&R&2015_03_28_13_50_56_1_050_Dela
yedMixing.Png|||||||||20150328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|2|^^^050^DelayedMixing|CSV&R&20150328&R&2015_03_28_13_50_56_1_050_Dela
yedMixing.csv|||||||||20150328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

56/74
6.4.7. When the measurement of platelet aggregation is output

6.4.7.1.Communication example: Analysis result (Maximum aggregation rate)


The communication example is based on the following conditions.
Output of analysis result : ADP%
Output of evaluation parameter : Not output
Output of platelet aggregation level : Not output
Output of image data : Not output

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^00-10^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1|| ^ ^ 1234567890^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^671^ADP%^100.00^A^^^|
5.7|%||N||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

6.4.7.2.Communication example: Analysis result (Maximum aggregation rate) + Evaluation


parameter
The communication example is based on the following conditions.
Output of analysis result : ADP%
Output of evaluation parameter : Output
Output of platelet aggregation level : Not output
Output of image data : Not output

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^00-10^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CHK1><
CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1|| ^ ^ 1234567890^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^671^ADP%^100.00^A^^^|
39.6|%||N||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^671^~Abs PPP^^A^^^|
67|mOD||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^671^~Abs PRP_s^^A^^^|
496|mOD||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^671^~Max^^A^^^|
39.6|%||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^671^~Fin^^A^^^|
7.3|%||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^671^~Lag Phase^^A^^^|
34.6|sec||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

57/74
IPU <STX>2R|7|^^^671^~AUC^^A^^^|37179|%*sec||||||||20170328135056<ETX><CHK1><CHK2><C
R><LF>
Host <ACK>
IPU <STX>3R|8|^^^671^~VMax^^A^^^|
65|%/sec||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|9|^^^671^~VMax Time^^A^^^|
24|sec||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|10|^^^671^~VMax Angle^^A^^^|
61.61|deg||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

6.4.7.3.Communication example: Analysis result (Maximum aggregation rate) + Evaluation


parameter + Aggregation curve
The communication example is based on the following conditions.
Output of analysis result : ADP% and ADP2%
Output of evaluation parameter : Output
Output of platelet aggregation level : Not output
Output of image data : Output

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^00-10^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1|| ^ ^ 1234567890^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^671^ADP%^100.00^A^^^|
39.6|%||N||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^672^ADP2%^100.00^A^^^|
98.3|%||N||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^671^~Abs PPP^^A^^^|
67|mOD||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|4|^^^671^~Abs PRP_s^^A^^^|
496|mOD||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^671^~Max^^A^^^|
39.6|%||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^671^~Fin^^A^^^|
7.3|%||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^671^~Lag Phase^^A^^^|
34.6|sec||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|8|^^^671^~AUC^^A^^^|37179|%*sec||||||||20170328135056<ETX><CHK1><CHK2><C
R><LF>
Host <ACK>
IPU <STX>4R|9|^^^671^~VMax^^A^^^|
65|%/sec||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|10|^^^671^~VMax Time^^A^^^|
24|sec||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|11|^^^671^~VMax Angle^^A^^^|
61.61|deg||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|12|^^^672^~Abs PPP^^A^^^|
58/74
67|mOD||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|13|^^^672^~Abs PRP_s^^A^^^|
491|mOD||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|14|^^^672^~Max^^A^^^|
98.3|%||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|15|^^^672^~Fin^^A^^^|
99.0|%||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|16|^^^672^~Lag Phase^^A^^^|
20.3|sec||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|17|^^^672^~AUC^^A^^^|53889|%*sec||||||||20170328135201<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>5R|18|^^^672^~VMax^^A^^^|
146|%/sec||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|19|^^^672^~VMax Time^^A^^^|
22|sec||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|20|^^^672^~VMax Angle^^A^^^|
77.23|deg||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|1|^^^671^Platelet|PNG&R&20170328&R&2017_03_28_13_50_56_1234567890_671_Platelet.
Png|||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|2|^^^672^Platelet|PNG&R&20170328&R&2017_03_28_13_52_01_1234567890_672_Platelet.
Png|||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|3|^^^^Superimpose|PNG&R&20170328&R&2017_03_28_14_10_23_1234567890__Superimpo
se.Png|||||||||20170328141023<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

6.4.7.4.Communication example: Analysis result (Maximum aggregation rate) + Evaluation


parameter (including Platelet Aggregation Level) + Aggregation curve
The communication example is based on the following conditions.
Output of analysis result : APAL-1% and APAL-10%
Output of evaluation parameter : Output
Output of platelet aggregation level : Output
Output of image data : Output

IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^00-10^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><CHK1>
<CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^^<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1|| ^ ^ 1234567890^M^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^676^APAL-1%^100.00^A^^^|
39.6|%||N||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^677^APAL-10%^100.00^A^^^|
98.3|%||N||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|3|^^^676^~Abs PPP^^A^^^|
67|mOD||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

59/74
IPU <STX>7R|4|^^^676^~Abs PRP_s^^A^^^|
496|mOD||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|5|^^^676^~Max^^A^^^|
39.6|%||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|6|^^^676^~Fin^^A^^^|
7.3|%||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|7|^^^676^~Lag Phase^^A^^^|
34.6|sec||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|8|^^^676^~AUC^^A^^^|37179|%*sec||||||||20170328135056<ETX><CHK1><CHK2><C
R><LF>
Host <ACK>
IPU <STX>4R|9|^^^676^~VMax^^A^^^|
65|%/sec||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|10|^^^676^~VMax Time^^A^^^|
24|sec||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|11|^^^676^~VMax Angle^^A^^^|
61.61|deg||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|12|^^^677^~Abs PPP^^A^^^|
67|mOD||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|13|^^^677^~Abs PRP_s^^A^^^|
491|mOD||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|14|^^^677^~Max^^A^^^|
98.3|%||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|15|^^^677^~Fin^^A^^^|
99.0|%||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|16|^^^677^~Lag Phase^^A^^^|
20.3|sec||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|17|^^^677^~AUC^^A^^^|53889|%*sec||||||||20170328135201<ETX><CHK1><CHK2><
CR><LF>
Host <ACK>
IPU <STX>5R|18|^^^677^~VMax^^A^^^|
146|%/sec||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6R|19|^^^677^~VMax Time^^A^^^|
22|sec||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|20|^^^677^~VMax Angle^^A^^^|
77.23|deg||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|21|^^^^~APAL^^A^^^|
10.0|||||||||20170328135415<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1R|1|^^^676^Platelet|PNG&R&20170328&R&2017_03_28_13_50_56_1234567890_676_Platelet.
Png|||||||||20170328135056<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2R|2|^^^677^Platelet|PNG&R&20170328&R&2017_03_28_13_52_01_1234567890_677_Platelet.
Png|||||||||20170328135201<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3R|3|^^^^Superimpose|PNG&R&20170328&R&2017_03_28_14_10_23_1234567890__Superimpo
se.Png|||||||||20170328141023<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>

60/74
6.4.8. Transmitting Comment records
* Comment records can be transmitted only when “System Settings”-“Output calibration curve and
reagent lot information” setting is ON.
IPU <ENQ>
Host <ACK>
IPU <STX>1H|¥^&|||CS-2500^01-70^10000001^^^CS-2500^BV981798||||||||E1394-97<ETX><C
HK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2P|1||||^Heisei^Jiro<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3O|1||000001^01^ 1^B^||R||||||N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>4R|1|^^^041^PT INN~sec^100.00^A^^^|
11.9|sec||N||||||20190619145609<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>5R|2|^^^042^PT INN~%^100.00^A^^^|
11.9|%||N||||||20190619145609<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>6C|1|I|CAL^042^^3^|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>7R|3|^^^043^PT INN~INR^100.00^A^^^|
1.00|||N||||||20190619145609<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>0R|4|^^^044^DFbg
INN~g_L^100.00^A^^^|----.-|g/L||<||||||20190619145609<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>1C|1|I|CAL^044^^3^|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>2C|2|I|LOT^040^PT Inn^505401|I<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <STX>3L|1|N<ETX><CHK1><CHK2><CR><LF>
Host <ACK>
IPU <EOT>

61/74
7. Appendix

A. TCP/IP communication

A.1 Software
1) Data link/ Network/ Transport layer
The IP address of the IPU for the CS-2500 is fixed. The default value is 192.168.37.252. When this value
is to be changed, please contact a Sysmex technical representative.
The IP address for the host computer can be set using the setting screen in the IPU of CS-2500. This IP
address may be selected other than “192.168.37.*” that is used to communicate with the IPU of the
analyzer.
The TCP port number of the IPU of CS-2500 for host communication is 5000. This value may be
changed in the IPU setting screen.

2) Session layer
To establish the TCP connection, the host computer acts as a server and the IPU acts as a client. The
IPU checks the communication on start-up. If the communication fails, the attempt to communicate is
performed at intervals. If the server is aborted after communication is established, reconnection is not
performed.

62/74
B. Parameter Code List

B.1. Parameter codes in inquiry record and request record


Parameter codes to be output in the inquiry record and the request record are shown as Table 23.
Table 23: Parameter Code Lists
Parameter Parameter
Assay Group Assay Group
Code Code
PT 04X Reserved 52X
APTT 05X PPP 53X
Fbg 06X ADP 54X
PT-extended 07X Epi 55X
Reserved 08X Col 56X
Reserved 09X Ris 57X
APTT-extended 10X Ara 58X
II 12X Reserved 60X
V 15X DDimer 61X
VII 17X Reserved 62X
VIII 18X Reserved 63X
IX 19X Reserved 64X
X 20X vWF-Ag med 65X
XI 21X vWF-Ag low 66X
XII 22X Reserved 67X
PCcl 25X Reserved 68X
BXT 26X Reserved 70X
LA1 27X Reserved 71X
LA2 28X Reserved 72X
PSAc 29X Edo 78X
AT3 30X EdoLow 79X
APL 31X vWF-RCo 80X
Plg 32X AcR R 81X
PC(Chrom) 33X AcR Act 82X
Hep 34X AcR OVB 83X
Reserved 36X FVL R 84X
C1 Inh 37X FVL Act 85X
VIII ch 39X FVL OVB 86X
XIII ch 40X INN vWF-Ac low 87X
LA R 41X INN vWF-Ac med 88X
PCAT NR 42X Reserved 89X
PCAT 43X Reserved 90X
PCAT0 44X INN FPS 91X
FVPCAT NR 45X Dabi 92X
FVPCAT 46X Api 94X
FVPCAT0 47X ApiLow 95X
IX Ch 48X Riv 96X
IX Ch L 49X RivLow 97X
Reserved 50X Reserved 000
TT 51X Reserved 999

000: There is no analysis parameter for the inquired sample.


999: There is no analysis parameter for the inquired sample.

If there is no information for the inquired sample, host computer sends either the parameter code “999”.
or “000” Once IPU receives “999” from the host computer, IPU displays “There is no order in Host
Computer” (error code: 21009). Once the host computer sends the parameter code “000” to IPU, IPU
displays an error dialog box when the setting of [System Settings]-[Operational Settings]-[Order
Inquiry]-[A warning message is output in case there is no parameter to be analyzed.] is turned ON. In
case the setting is OFF, IPU do not display the error dialog box.

63/74
“X” is replaced with a one-digit host ID defined in the Assay Parameter Definition in when transmitting
analysis results.
In the case of order inquiry, order information texts and the analysis data format of formula, “X” is
replaced with 0.
New analysis parameters may be added in the future. Please prepare a receive program that ignores the
data when receiving the codes other than the codes shown in the parameter code list .

64/74
B.2. Parameter codes in result record
Parameter codes to be output in the result record are shown as Table 24.
Table 24: Parameter Code Lists(Result)
Test Test name Units Format Test Test name Units Format
code code
041 PT sec sec ####.# 482 IX Ch % % ####.#
042 PT % % ####.# 491 IX Ch L dOD dOD #.####
043 PT INR ###.## 492 IX Ch L % % ####.#
044 DFbg g_L g/L ####.# 50X Reserved
045 PT R ###.## 511 TT sec sec ####.#
051 APTT sec sec ####.# 52X Reserved
055 APTT R ###.## 531* PPP mOD #####
061 Fbg sec sec ####.# 541* ADP_min mOD #####
062 Fbg C. g/L ###.## 542* ADP_s mOD #####
071 PT ext sec sec ####.# 551* Epi_min mOD #####
072 PT ext INR ###.## 552* Epi_s mOD #####
08X Reserved 561* Col_min mOD #####
09X Reserved 562* Col_s mOD #####
101 APTT ext sec sec ####.# 571* Ris_min mOD #####
121 II sec sec ####.# 572* Ris_s mOD #####
122 II % % ####.# 581* Ara_min mOD #####
151 V sec sec ####.# 582* Ara_s mOD #####
152 V% % ####.# 60X Reserved
171 VII sec sec ####.# 611 INN DD dOD dOD ##.###
172 VII % % ####.# 612 INN DD C. mg/L ###.##
181 VIII sec sec ####.# 62X Reserved
182 VIII % % ####.# 63X Reserved
191 IX sec sec ####.# 64X Reserved
192 IX % % ####.# 651 vWF-Ag med dOD dOD #.####
201 X sec sec ####.# 652 vWF-Ag med % % ####.#
202 X% % ####.# 661 vWF-Ag low dOD dOD #.####
211 XI sec sec ####.# 662 vWF-Ag low % % ####.#
212 XI % % ####.# 671 ADP% % ####.#
221 XII sec sec ####.# 672 ADP2% % ####.#
222 XII % % ####.# 673 ADP3% % ####.#
251 PC cl sec sec ####.# 674 ADP4% % ####.#
252 PC cl % % ####.# 675 ADP5% % ####.#
261 BXT sec ####.# 676 APAL-1% % ####.#
271 LA1 sec sec ####.# 677 APAL-10% % ####.#
281 LA2 sec sec ####.# 681 Epi% % ####.#
291 PS Ac sec sec ####.# 682 Epi2% % ####.#
292 PS Ac % % ####.# 683 Epi3% % ####.#
301 AT3 dOD, INN AT dOD dOD ##.### 684 Epi4% % ####.#
302 AT3 %, INN AT % % ####.# 685 Epi5% % ####.#
311 APL dOD dOD ##.### 701 Col% % ####.#
312 APL % % ####.# 702 Col2% % ####.#
321 Plg dOD dOD ##.### 703 Col3% % ####.#
322 Plg % % ####.# 704 Col4% % ####.#
331 PC dOD dOD ##.### 705 Col5% % ####.#
332 PC % % ####.# 706 CPAL-1% % ####.#
341 Hep dOD dOD ##.### 707 CPAL-5% % ####.#
342 Hep C. IU/mL ###.## 711 Ris% % ####.#
36X Reserved 712 Ris2% % ####.#
371 C1 dOD dOD ##.### 713 Ris3% % ####.#
372 C1 % % ####.# 714 Ris4% % ####.#
391 VIII ch dOD dOD #.#### 715 Ris5% % ####.#
392 VIII ch % % ####.# 721 Ara% % ####.#
401 XIII ch dOD dOD #.#### 722 Ara2% % ####.#
402 XIII ch % % ####.# 723 Ara3% % ####.#
411 LA R ###.## 724 Ara4% % ####.#
421 PCAT NR ###.## 725 Ara5% % ####.#
431 PCAT sec sec ####.# 781 Edo dOD dOD #.####
441 PCAT0 sec sec ####.# 782 Edo C. ng/mL ###.##
451 FVPCAT NR ###.## 791 EdoL dOD dOD #.####

65/74
461 FVPCAT sec sec ####.# 911 INN FPS dOD dOD #.####
481 IX Ch dOD dOD #.#### 912 INN FPS % % ####.#
792 EdoL C. ng/mL ###.## 921 Dabi dOD dOD ##.###
801 vWF-RCo dOD dOD #.#### 922 Dabi C. ng/mL #####
802 vWF-RCo % % ###.## 941 Api dOD dOD #.####
811 AcR R ###.## 942 Api C. ng/mL ###.##
821 AcR Act sec sec ####.# 951 ApiL dOD dOD #.####
831 AcR OVB sec sec ####.# 952 ApiL C. ng/mL ###.##
841 FVL R ###.## 961 Riv dOD dOD #.####
851 FVL Act sec sec ####.# 962 Riv C. ng/mL ###.##
861 FVL OVB sec sec ####.# 971 RivL dOD dOD #.####
871 INN vWF low dOD dOD #.#### 972 RivL C. ng/mL ###.##
872 INN vWF low % % ####.# 000 Reserved
881 INN vWF med dOD dOD #.#### 999 Reserved
882 INN vWF med % % ####.#
89X Reserved
90X Reserved - - - -
“X” is one-digit of parameter code for reserved.

The following shows the default Host ID for assay parameter defined in the Assay Group Settings.

1: Time, dOD
2: Activity %/Concentration
3: INR
4: DFbg
5: Ratio

・The test code of formula other than platelet aggregation is fixed by “1” to the third digit of the code. On the other
hand, “1” to “7” as unique number is given to the third digit of test code for platelet aggregation. The followings are
examples for test code of platelet aggregation: 671 to 677, 681 to 685, 701 to 707, 711 to 715, and 721 to 725. Refer
to Table 24 for details.
*A result record of absorbance (mOD) (parameter code: 531 to 582) for PPP and PRP in the measurement of
platelet aggregation is data obtained in calculation process of the maximum aggregation rate. Therefore, a host
output is not carried out in the software version after Ver.01-70.

66/74
C. Error Information

Table 25: Analysis error, Result Abnormal Flags, and Display method of analysis results

*1: If the setting of “Output Error Information” in the system settings is turned ON and also“The data is not output when the
coagulation curve error is found” is turned ON, the masked values are output.
*2 When“HILCheck Settings” of System Settings is set to “Output numeric values”.
*3 When“HILCheck Settings” of System Settings is set to “Output values with low-reliability flags”.
*4 When“HILCheck Settings” of System Settings is set to “Output masked values”.
*5 When samples are analyzed with a measurement time which is equal to or greater than the setting value of the “Check Time[sec]”
in the “Flat Curve Check” in the “Evaluation Check Parameter2” in the “Evaluation Preset” in the “Assay Parameter Setup” in the
“Basic” in the “Assay Group Setup”, numeric values are output.

Display method of analysis results


Result
Code Message abnormal
“Output Error “Output Error
flag Information” is OFF Information” is ON
Numeric values are Numeric values are
0000.0000.0000 [Dilution Ratio was Changed.] N
output output
Numeric values are Numeric values are
0000.0000.0340 [Wave Changed 340nm] N
output output
Numeric values are Numeric values are
0000.0000.0405 [Wave Changed 405nm] N
output output
Numeric values are Numeric values are
0000.0000.0575 [Wave Changed 575nm] N
output output
Numeric values are Numeric values are
0000.0000.0660 [Wave Changed 660nm] N
output output
Numeric values are Numeric values are
0000.0000.0800 [Wave Changed 800nm] N
output output
Numeric values are Numeric values are
0000.0000.1100 [Vial QC has not performed.] A
output output
Numeric values are Numeric values are
0000.0000.8000 [Coag% Changed] A
output *5 output *5
Masked values are Masked values are
0000.4000.0010 [Display Digit Overflow] N
output output
Masked values are Masked values are
0000.4000.0020 [Calculation Failure] N
output output
Masked values are Masked values are
0000.4000.0030 [Validated Calibration Curve is not found.] N
output output
Masked values are Masked values are
0000.4000.0041 [Extrapolation Boundary Over (Upper)] N
output output
Masked values are Masked values are
0000.4000.0042 [Extrapolation Boundary Over (Lower)] N
output output
Numeric values are Numeric values are
0000.4000.0050 [Calculated later.] N
output output
Masked values are Masked values are
0000.5000.0010 [Formula Calculation Failure] N
output output
Masked values are Masked values are
0000.5000.0020 [Formula Calculation Failure] N
output output
Masked values are Masked values are
0000.5000.0050 [Problem in Calculating Formula] N
output output
Numeric values are Numeric values are
0000.6000.0100 [Upper Report Limit Over] >
output output
Numeric values are Numeric values are
0000.6000.0200 [Lower Report Limit Over] <
output output
Numeric values are Numeric values are
0000.6000.0300 [Upper Mark Limit Over] H
output output
Numeric values are Numeric values are
0000.6000.0400 [Lower Mark Limit Over] L
output output
Numeric values are Numeric values are
0000.6000.0500 [MDA Slope Ratio Over] A
output output
Numeric values are Numeric values are
0000.6000.0600 [Difference between replicated results] A
output output
Masked values are Masked values are
0000.7000.0020 [Mean Calculation Failure] N
output output
Numeric values are Numeric values are
0002.0000.0000 [Slight Coagulation] A
output output

67/74
Display method of analysis results
Result
Code Message abnormal
“Output Error “Output Error
flag Information” is OFF Information” is ON
Numeric values are Numeric values are
0004.0000.0000 [Analysis Time Over] A
output output
Masked values are Numeric values are
0008.0001.0000 [Initial fluctuation drop] A
output output*1
Masked values are Numeric values are
0008.0002.0000 [Coagulation Curve Error: Sharp Drop] A
output output*1
Masked values are Numeric values are
0008.0004.0000 [Coagulation Curve Error: Dip] A
output output*1
Masked values are Numeric values are
0008.0008.0000 [Coagulation Curve Error: Jump Up] A
output output*1
Masked values are Numeric values are
0008.0016.0000 [Coagulation Curve Error: Stepping Curve] A
output output*1
Masked values are Numeric values are
0008.0032.0000 [Coagulation Curve Error: Fbg Curve Error] A
output output*1
Masked values are Numeric values are
0008.0064.0000 [Coagulation Curve Error: Terrace] A
output output*1
Masked values are Masked values are
0008.0128.0001 [Early Reaction Error :Slow Reaction] A
output output
Masked values are Masked values are
0008.0128.0002 [Early Reaction Error :Start Angle 1] A
output output
Numeric values are Numeric values are
0008.0128.0004 [Early Reaction Error :Start Angle 2] A
output output
Masked values are Masked values are
0008.0128.0008 [Early Reaction Error :Drift] A
output output
Masked values are Masked values are
0008.0128.0016 [Early Reaction Error :Early %] A
output output
Masked values are Masked values are
0008.0256.0000 [Noise] A
output output
Masked values are Masked values are
0016.0000.0000 [Turbidity Level Over] A
output output
Masked values are Masked values are
0032.0000.0000 [No Coagulation] A
output output
Masked values are Masked values are
0032.0002.0000 [Flat Curve] A
output output
Masked values are Numeric values are
0064.0000.0000 [Aged Sample] A
output output*1
Masked values are Masked values are
0128.0000.0000 [Range Over] A
output output
Masked values are Masked values are
0256.0000.0000 [Trans Light High:Clot.] A
output output
Numeric values are Numeric values are
N*2
output*2 output*2
Numeric values are Numeric values are
1000.1000.0000 [Hemolyzed Sample] A*3
output*3 output*3
Masked values are Masked values are
A*4
output*4 output*4
Numeric values are Numeric values are
N*2
output*2 output*2
Numeric values are Numeric values are
1000.2000.0000 [Icteric Sample] A*3
output*3 output*3
Masked values are Masked values are
A*4
output*4 output*4
Numeric values are Numeric values are
N*2
output*2 output*2
Numeric values are Numeric values are
1000.3000.0000 [Lipemic Sample] A*3
output*3 output*3
Masked values are Masked values are
A*4
output*4 output*4
Numeric values are Numeric values are
1100.4000.0000 [Defective Sample Volume] N
output output
Masked values are Masked values are
4001.0000.0000 [Trans Light Low] A
output output
Masked values are Masked values are
4002.0000.0000 [Trans Light High:Chro.Imm.] A
output output
Masked values are Masked values are
4004.0000.0000 [No Linearity] A
output output
Masked values are Masked values are
4008.0000.0000 [Reaction Curve Error] A
output output

68/74
Display method of analysis results
Result
Code Message abnormal
“Output Error “Output Error
flag Information” is OFF Information” is ON
Masked values are Masked values are
4016.0000.0000 [Antigen Excess] A
output output
Masked values are Masked values are
4032.0000.0000 [Bump] A
output output
Masked values are Masked values are
4128.0000.0000 [No Polynomialadjustment] A
output output
Masked values are Masked values are
4256.0000.0000 [Range in non-linear] A
output output
Numeric values are Numeric values are
7010.0000.0000 [PPP Low1] A
output output
Numeric values are Numeric values are
7030.0000.0000 [PPP High1] A
output output
Masked values are Masked values are
7050.0000.0000 [Tube_Position] A
output output
Masked values are Masked values are
9999.0000.0000 [Measurement Failure] A
output output
Masked values are Masked values are
9999.0000.9010 [TEST_RECV_ELSE_ERROR] N
output output
Masked values are Masked values are
9999.0000.9020 [ASSY_CALV_PARA_ERROR] N
output output
Masked values are Masked values are
9999.0000.9030 [ASSY_CALV_INF_ERROR] N
output output
Masked values are Masked values are
9999.0000.9040 [ASSY_ANLY_CALC_ERROR] N
output output
Masked values are Masked values are
9999.0000.9050 [ASSY_RECV_ELSE_ERROR] N
output output
Masked values are Masked values are
9999.0000.9060 [ASSY_ANLY_NOPRISET_ERROR] N
output output
Masked values are Masked values are
9999.0000.9070 [ASSY_ANLY_NORESULT_ERROR] N
output output
Masked values are Masked values are
9999.0000.9080 [ASSY_ANLY_DILUTION_ERROR] N
output output
Masked values are Masked values are
9999.0000.9090 [ASSY_CALY_NODATA_ERROR] N
output output
Masked values are Masked values are
9999.0000.9100 [ASSY_AVER_IRIGALRESULT_ERROR] N
output output
Masked values are Masked values are
9999.0000.9110 [TEST_REAGENTLOT_ERROR] N
output output
Numeric values are Numeric values are
Unknown9999 [Unknown Message] N
output output

69/74
D. Re-analysis order

The sample flow of communication sequence and simulation results for the“Inquiry of Analysis
Order”and the “Analysis Order” when the “Inquire re-analysis” setting in the “System Settings” is ON
is shown as follows;

D.1 There is no re-analysis order (Batch transmission):

CS-2500 HOST

Inquiry of the first analysis order


Order Information

Transmitting analysis results (PT,APTT,Fbg)

Restricted Stand-by time(*1)


transmission Inquiry of re-analysis order
(*3)
No re-analysis order information (*2)

Transmitting final analysis results(PT,APTT,Fbg)

D.2 There is a re-analysis order (Batch transmission):

CS-2500 HOST

Inquiry of the first analysis order


Order Information
(PT,APTT,Fbg)

Transmitting analysis results


(PT,APTT,Fbg)

Inquiry of re-analysis order

Order information (PT)


(Re-analysis)

Transmitting re-analysis results(PT)

Transmitting final analysis results(PT,APTT,Fbg)

70/74
D.3 There is a re-analysis order (For each assay group):

CS-2500 HOST

Inquiry of the first analysis


Order Information (PT,Fbg)

Transmitting analysis results (PT)

Inquiry of re-analysis order

Order information (APTT)


(Reflex)

Transmitting analysis results (Fbg)


(*4)

Re-analysis results (APTT)

Final analysis results(PT, Fbg, APTT)(*5)

(*1) Stand-by time: Analysis results and Inquiry of re-analysis order. Default (System settings):1 sec
(*2) No re-analysis order information: Please return“000”for the test code, if the HOST judges “do no perform
re-analysis” for the inquiry of re-analysis order.
(*3) Restricted transmission: No other transmission can interrupt during the period.
(*4) No Inquiry:If the “Fbg” is not targeted for re-analysis, the inquiry of re-analysis order is not performed.
(*5) Final analysis results:When the “Output each Assay Group result” setting of the “System Settings” is
ON and the “Output the final result” setting is OFF, the final analysis results are not transmitted.

71/74
E. Platelet aggregation
“Analysis result (Maximum aggregation rate)”, “Evaluation parameter”, and “Aggregation curve”
are output as analysis result of the measurement of platelet aggregation. The setting of the IPU
determines whether or not to output the “Evaluation parameter” and the “Aggregation curve”.

Table 26: Output conditions of “Analysis result”, “Evaluation parameter”, and “Aggregation curve” in the
measurement of platelet aggregation
IPU System setting
[Host Computer] Detailed setting
Output
Output data
Evaluation Info Output Reaction
on Platelet Curve
Aggregation.
OFF ON/OFF “Analysis result”
ON OFF “Analysis result” + “Evaluation parameter”
“Analysis result” + “Evaluation parameter” +
ON ON
“Aggregation curve”

The chart below describes the output sequence of the “Analysis result”, the “Evaluation
parameter”, and the “Aggregation curve” in the measurement of platelet aggregation. The “Sample
information flag” is output if the sample has sample volume flag.

72/74
IPU Host
At the time of output “Analysis results”
Analysis result (The first result)

Analysis result (The second result)


At the time of output “Evaluation parameters”

Evaluation parameter(PPP absorbance)

Evaluation parameter(PRP absorbance before reaction starts) Repeat


according to
: the number of
: analysis
parameters to
At the time of output“Platelet Aggregation Level” be set.
Evaluation parameter(~APAL)

Evaluation parameter(~CPAL)

At the time of output “Sample information flags”

Sample information flag (Defective Sample Volume)

At the time of output of “Aggregation curves”

Aggregation curve (The first curve)

Aggregation curve (The second curve)



Superimposed aggregation curve

After outputting all “Analysis results”, the “Evaluation parameters” are output according to the
number of the measurements to be set. When output of the “Evaluation parameters” is completed,
the “Sample information flags” are output. Finally, the “Aggregation curve” is output.

73/74
F. Comment Record
There are two types of information to be transmitted as comment record, “Calibration curve information”
and “Reagent lot information”. The following chart describes transmission order when each of the
comment records is transmitted. The comment record can be transmitted when “Output calibration curve
and reagent lot information” setting in “System Settings” is ON. In the case of platelet aggregation result
or formula result, the comment record is not transmitted.

IPU HOST

If a calibration curve is used for calculation of data value of analysis result: (*1)

Result Record: Analysis result

Comment record: Calibration curve information

If a calibration curve is not used for calculation of data value of analysis result:

Process sequences
Result Record: Analysis result repeatedly according
to the number of
“Comment record: Calibration curve information” is not assay groups.
transmitted to the host computer.

:
:

Comment record: reagent lot information (*2)

(*1) When a calibration curve is used for calculation of data value of analysis result, a record as comment
record for calibration curve information is transmitted following the result record of the analysis
result. The calibration curve information is transmitted when calibration type is “Normal calibration
curve” or “INR calibration curve” and the aalysis status of the result is other than “On Hold”.

(*2) After transmitting the result record of the analysis result, a record as comment record for reagent lot
information is transmitted. The reagent lot information is transmitted by assay groups. Thus, when
an assay group contains multiple assay parameters, a record as comment record for reagent lot
information is transmitted following the result record of the last assay parameter.

End of Document